28
Chapter 2 Measurement Procedure
NOTE
2.4.4 Current
1. Press the
I/
key to activate current display mode.
In current display mode, the unit will indicate an
effective value on Display 1, peak hold value on
Display 2, and frequency on Display 3.
2. Switch between Auto Range and Manual Range, as
needed. For details, see 2.3, "Range Setup."
3. Open the tip of the clamp core and clamp the
conductor roughly into the center of the clamp core.
4. Pressing the
I/
key in current display mode resets
the peak hold value.
Be sure to clamp one conductor only. Measurement
is not possible for single phase or three phases when
two or three conductors are respectively clamped at
the same time.
When only measuring current, there is no need to
connect the voltage cord.
Select the 1φ P meter, 1φ PF meter, or 3φ PF meter.
The unit does not display polarities in a peak
measurement.
The peak hold value will not vary, unless a large
value is entered in the unit. If the auto power-off
function is effective, the unit will be shut down in
about ten minutes, causing the data to be lost. (See
2.11, "Auto Power-Off Function.") One way to
prevent data from being lost is to disable the auto
power-off function (see 2.9, "SETUP Function") or
to use the recording function.
For measurement extending the auto power-off time,
use the recording function.
To check variations in a peak value, enable the
REC function by pressing the
MAX/MIN
key, then
activate peak value display mode by pressing the
I/
key.