December 2014 Revised edition 10 3285B981-10 14-12H
Page 2
Page 3
Contents
Introduction i
Verifying Package Contents
Safety Information
Operating Precautions
Preliminary Check
viii
xv
Chapter 1 Overview
1.1 Product Overview 1
1.2 Features
1.3 Names and Functions of Parts
1.4 Flowchart of button Operations
1.4.1 Current Measurements Mode 10
1.4.2 Voltage Measurements Mode
1.4.3 Frequency Measurements Mode
1.4.4 Resistance Measurements Mode
10
11
12
12
1.5 Modes 13
Chapter 2 Measurement 15
2.1 Preparations 15
2.1.1 Installing or Replacing the Battery 15
2.1.2 Attaching the Hand Strap
2.1.3 Turning the Power On and Off
2.1.4 Battery Low Warning (When
2.1.5 Pre-Operation Inspection
mark lights) 17
16
16
18
ii
iii
1
2
3
3285B981-10
Page 4
2.2 Current Measurement 19
2.2.1 Measuring DC Current (DC A) 19
2.2.2 Measuring AC Current (AC A)
2.2.3 Measuring AC/DC Current (AC+DC A)
2.2.4 Peak Hold Measurement
21
22
24
2.3 Voltage Measurement 26
2.3.1 Measuring DC Voltage (DC V) 26
2.3.2 Measuring AC Voltage (AC V)
2.3.3 Measuring AC/DC Voltage (AC+DC V)
2.3.4 Peak Hold Measurement
27
28
29
2.4 Frequency Measurement 31
2.4.1 Frequency Measurement in Current Mode 31
2.4.2 Frequency Measurement in Voltage Mode
32
2.5 Resistance Measurement 34
2.6 Continuity Check
36
Chapter 3 Useful Functions 39
3.1 Adjusting and Correcting Values 39
3.1.1 Auto-Zero-Adjustment Function 39
3.1.2 Zero-Cancel Correction Function
3.2 Pausing and Reading the Display
(Data Hold Function HOLD)
41
3.3 Alteration of Counter Updates
3.3.1 SLOW mode 42
3.3.2 FAST mode
3.4 Storing Values in Memory
(Recording Function REC)
3.5 Automatically T urning Power Off
(Auto Power-Off Function APS
)
47
3.6 Beep Tone
40
42
42
43
48
Page 5
Chapter 4 Specifications 49
4.1 Measurement Specifications 49
4.1.1 Current Measurement Specifications 49
4.1.2 Voltage Measurement Specifications
4.1.3 Resistance Measurement Specifications
4.1.4 Continuity Check
52
54
54
4.2 General Specifications 55
Chapter 5 Maintenanceand Service 59
5.1 Cleaning 59
5.2 After service
5.3 Troubleshooting
5.4 Error Massage
59
60
64
Page 6
Page 7
Introduction
Thank you for purchasing the HIOKI "3285-20
Clamp-on AC/DC HiTester."
To obtain maximum performance from the
instrument, please read this manual first, and keep it
handy for future reference.
Request
We have tried to bring this manual as close to
perfection as we could achieve. If perchance you
find any unclear portions, mistakes, omissions, or
the like, we would be most obliged if you could
please notify us of them via any HIOKI agent, or
directly.
When you receive the instrument, inspect it
carefully to ensure that no damage occurred during
shipping. In particular, check the accessories, panel
switches, and connectors.
If damage is evident, or if it fails to operate
according to the specifications, contact your dealer
or Hioki representative.
Main unit
3285-20 Clamp-on AC/DC HiTester
Supplied accessories
9345 Carrying Case 1
L9207-10 Test Lead (red and black) 1
Hand Strap 1
Battery 6F22(006P) 1
Instruction manual 1
61010 Safety Standards, and has been thoroughly
tested for safety prior to shipment. However,
mishandling during use could result in injury or
death, as well as damage to the instrument. Be
certain that you understand the instructions and
precautions in the manual before use. We disclaim
any responsibility for accidents or injuries not
resulting directly from instrument defects.
This manual contains information and warnings essential
for safe operation of the instrument and for maintaining
it in safe operating condition. Before using it, be sure to
carefully read the following safety precautions.
Thesymbol printed on the instrument
indicates that the user should refer to a
corresponding topic in the manual (marked with
the
symbol) before using the relevant
function.
In the manual, the
particularly important information that the user
should read before using the instrument.
Indicates both DC (Direct Current) and AC
(Alternating Current).
Indicates DC (Direct Current).
Indicates that dangerous voltage may be
present at this terminal
Indicates a double-insulated device.
Indicates a grounding terminal.
Indicates that the instrument may be connected
to or disconnected from a live circuit.
The following symbols in this manual indicate the
relative importance of cautions and warnings.
Indicates that incorrect operation presents
DANGER
extreme danger of accident resulting in
death or serious injury to the user.
Indicates that incorrect operation presents
WARNING
a significant hazard that could result in
serious injury or death to the user.
Indicates that incorrect operation
CAUTION
presents a possibility of injury to the
user or damage to the instrument.
Indicates advisory items related to
NOTE
performance or correct operation of the
instrument.
Accuracy
We define measurement tolerances in terms of f.s.
(full scale), rdg. (reading) and dgt. (digit) values,
with the following meanings:
f.s. (maximum display value or scale length)
The maximum displayable value or scale length.
This is usually the name of the currently selected
range.
rdg. (reading or displayed value)
The value currently being measured and indicated
on the measuring instrument.
dgt. (resolution)
The smallest displayable unit on a digital measuring
instrument, i.e., the input value that causes the
digital display to show a "1" as the least-significant
digit.Accuracy
This instrument complies with CAT III safety
requirements.
To ensure safe operation of measurement
instruments, IEC 61010 establishes safety standards
for various electrical environments, categorized as
CAT II to CAT IV, and called measurement
categories. These are defined as follows.
CAT II
CAT III
CAT IV
Using a measurement instrument in an environment
designated with a higher-numbered category than
that for which the instrument is rated could result in
a severe accident, and must be carefully avoided.
Use of a measurement instrument that is not CATrated in CAT II to CAT IV measurement
applications could result in a severe accident, and
must be carefully avoided.
Primary electrical circuits in equipment
connected to an AC electrical outlet by a
power cord (portable tools, household
appliances, etc.)
CAT II covers directly measuring
electrical outlet receptacles.
Primary electrical circuits of heavy
equipment (fixed installations) connected
directly to the distribution panel, and
feeders from the distribution panel to
outlets.
The circuit from the service drop to the
service entrance, and to the power meter
and primary overcurrent protection device
(distribution panel).
Follow these precautions to ensure safe operation
and to obtain the full benefits of the various
functions.
DANGER
To avoid short circuits and potentially lifethreatening hazards, never attach the clamp to
a circuit that operates at more than 600 Vrms.
The maximum rated voltage to earth is 600
Vrms. Attempting to measure voltage in excess
of the maximum rated voltage to earth could
destroy the instrument and result in personal
injury or death.
This instrument should only be connected to
the secondary side of a breaker, so the breaker
can prevent an accident if a short circuit
occurs. Connections should never be made to
the primary side of a breaker, because
unrestricted current flow could cause a serious
accident if a short circuit occurs.
To avoid electric shock, do not touch the
portion beyond the protective barrier during
use.
Never apply voltage to the test leads when the
Resistance measurement, Continuity Check
modes are selected. Doing so may damage the
instrument and result in personal injury.
To avoid electrical accidents, remove power
from the circuit before measuring.
Do not allow the instrument to get wet, and do
not take measurements with wet hands. This
may cause an electric shock.
To avoid electric shock when measuring live
lines, wear appropriate protective gear, such as
insulated rubber gloves, boots and a safety
helmet.
During current measurement, to avoid an
electric shock accident, do not connect the test
leads or temperature probe to the instrument.
Input voltage exceeding 600 Vrms (1000 V max.)
with respect to ground could damage the
instrument and result in personal injury.
Be sure to insert them with the correct polarity.
Otherwise, poor performance or damage from
battery leakage could result.
To avoid electric shock when replacing the
battery, first disconnect the clamp from the
object to be measured.
After replacing the battery, replace the cover
and screws before using the instrument.
To avoid the possibility of explosion, do not
short circuit, disassemble or incinerate
batteries.
Handle and dispose of batteries in accordance
with local regulations.
voltage of a power line use a test lead that
satisfies the following criteria:
・
Conforms to safety standards IEC61010 or
EN61010
・
Of measurement category III or IV
・
Its rated voltage is higher than the voltage to
be measured
The test leads provided with this instrument
conform to the safety standard EN61010.
Use a test lead in accordance with its defi ned
measurement category and rated voltage.
Do not use the instrument where it may be
exposed to corrosive or combustible gases.
The instrument may be damaged or cause an
explosion.
Do not store or use the instrument where it could
be exposed to direct sunlight, high temperature or
humidity, or condensation. Under such conditions,
the instrument may be damaged and insulation
may deteriorate so that it no longer meets
specifications.
This instrument is designed for use indoors. It can
be operated at temperatures between 0 and 40
without degrading safety.
Keep the clamp jaws and core slits free from
foreign objects, which could interfere with clamping
action.
For safety reasons, when taking measurements,
only use the Model L9207-10 Test Lead provided
with the instrument.
To avoid damage to the instrument, do not exceed
the maximum input current rating, which depends
on the frequency of the current being measured
(see Fig.3 of 4.1.1) Be careful about the evolution
of heat, when the input frequency is high.
The " " indicator lights up when the remaining
battery capacity is low. In this case, the
instrument's reliability is not guaranteed. Replace
the battery immediately.
When replacing the battery, make sure that the
metal battery snap fitting is firmly connected. If the
metal fitting is loose, adjust it and recheck the
connection. If it isn't connected securely, the
power may not be turned on, and a power may be
turned off during the use.
To avoid damage to the instrument, protect it from
physical shock when transporting and handling. Be
especially careful to avoid physical shock from
dropping.
Do not exert excessive pressure on the clamp
sensor or attempt to wedge the sensor into a tight
spot for measurement.
The protection rating for the enclosure of this
device (based on EN60529) is *IP40.
*IP40:
This indicates the degree of protection provided by
the enclosure of the device against use in
hazardous locations, entry of solid foreign objects,
and the ingress of water.
4: Protected against access to hazardous parts
with wire measuring 1.0 mm in diameter. The
equipment inside the enclosure is protected
against entry by solid foreign objects larger than
Removable sleeves are attached to the metal pins
at the ends of the test leads.
To prevent a short circuit accident, be sure to use
the test leads with the sleeves attached when
performing measurements in the CAT III
measurement category. In a CATII environment, if
the tips of the test leads do not reach the
measurement object, remove the rigid insulating
sleeve before measuring.
For details on measurement categories, see
“
Measurement categories”(page vi) in the
instruction manual.
When performing measurements with the sleeves
attached, be careful to avoid damaging the
sleeves. If the sleeves are inadvertently removed
during measurement, be especially careful in
handling the test leads to avoid electric shock.
To prevent an electric shock accident, confirm that
the white or red portion (insulation layer) inside the
cable is not exposed. If a color inside the cable is
exposed, do not use the cable.
Correct measurement may be impossible in the
presence of strong magnetic fields, such as near
transformers and high-current conductors, or in the
presence of strong electromagnetic fields such as
near radio transmitters.
To avoid corrosion from battery leakage, remove the
batteries from the instrument if it is to be stored for
a long time.
Hall elements are used in the current detection
portion of the clamp on AC/DC HiTester. As there
are two types of drifts in the Hall elements, drift
depending on time and drift depending on the
surrounding temperature, take note when conducting
continuous measurements.
The Hall elements lack uniformity, and it is
impossible to determine the amount of variation
which depends on time and the temperature. When
conducting continuous measurement or in an
environment where there are huge temperature
variations, it is recommended that the variation at
point 0 when nothing has been inputted should be
first noted. The variation at point 0 is the direct
current component.This does not affect the AC
mode.
・
Due to the offset peculiar to elements from a hall
effect, the display may sometimes not show 0
immediately after it is turned on, but this is not a
malfunction. The count remaining has an
instrumental difference of 0 to a few tens A.
・
When using in DC or AC/DC mode, press the
0ADJ/RESET key each time the power is turned
on to carry out auto zero adjustment.
When 0 is displayed after zero adjustment, the
measurement can be performed without the
interference of the count remaining immediately
after the power is turned on.
Before using the instrument the first time, verify
that it operates normally to ensure that the no
damage occurred during storage or shipping.
If you find any damage, contact your dealer or
Hioki representative.
WARNING
Before using the instrument, make sure that the
insulation on the test leads is undamaged and
that no bare conductors are improperly exposed.
Using the instrument in such conditions could
cause an electric shock, so contact your dealer or
Hioki representative for replacements. (Model
L9207-10 Test Lead).
The 3285-20 Clamp-on AC/DC HiTester makes it
possible to measure DC, AC or AC+DC current in
live power lines without tapping into or connecting
the lines.
And it is also possible to measure voltage, frequency
and resistance, and check circuit continuity.
Used to turn power on/off
To disable the auto power-off function, hold
button
POWER
, when you turn power on.
and press
2
HOLD
Switches current modes as follow.
DCAAC+DCAACA
3
RANGE
button
Switches between auto and manual ranges in
measurements of current, voltage, or frequency.
Switches manual ranges.
Displays a cursor on the bar graph to show the
selected range.
The current ranges are 200 A and 2000 A. The
voltage ranges are 30 V, 300 V and 600 V. The
frequency ranges are 10 Hz, 100 Hz and 1000 Hz.
Auto-ranging is used for resistance measurement.
HOLD
4
button
Used to suspend or inac tivate the screen-updating
function.
To disable the auto power-off function when
powering on, hold
Switches between resistance measurement and
continuity checking.
Ω
SLOW/PEAK/Hz
6
button
SLOW slows down screen updating speed (once per
three seconds).
FAST speeds up screen updating speed (four times
per second). There isn't an annunciator "
FAST
Instead, the unit symbol blinks.
PEAK measures peak values (Peak Hold).
Hz measures frequency (in AC or AC+DC mode).
Displays the maximum value (MAX), the minimum
value (MIN), or the average value (AVE) of the
maximum and minimum values for the recording
(REC) function.
MAX d isplays the maximum measured value after
the REC function is activated.
MIN displays the minimum measured value after the
REC function is activated.
AVE displays the average value of the maximum
and minimum measured values after the REC
function is activated.
The auto power-off function is disabled.
Not available during resistance measurement.
Performs auto-zero-adjustment in DC A, AC+DC A
and DC V modes.
Resets data when measuring peak values. Reset all
the data in a REC function.
If zero is not indicated under no input in the AC A,
AC+DC A, AC V or AC+DC V modes, press
HOLD
, then press
0ADJ/RESET
to perform a zero-
cancel correction.
10 Clamp sensor
To measure current, open the top ends of the clamp
sensor by gripping the lever 12. Then position the
conductor to be measured at the center of the clamp
sensor and firmly close the clamp sensor.
11 Protective barrier
Be careful to not touch the clamp sensor beyond the
safety barrier when clamping onto a conductor for
measurement.
Connected to the L9207-10 Test Lead (red and
black, supplied with the instrument) to measure
voltage or resistance.
15 Back case
To replace the battery, remove the two screws.
16 Hand strap
Attach to get a better grip on the instrument.
Resistance
Continuity
Frequency
Voltage
Wave peak value
True root mean square value
Current
1 hour/segment (bar graph)
1 minute/segment (bar graph)
Input over (bar graph)
Be careful to not touch the metal part of the test
lead beyond the safety barrier when contacting a
conductor for measurement.
2 Sleeve
Attach to the pins to prevent short circuit accidents.
CAUTION
・
Be sure to use the test leads with the sleeves
attached when performing measurements in the
CAT III measurement category. In a CAT II
environment, if the tips of the test leads do not
reach the measurement object, remove the rigid
insulating sleeve before measuring.
・
When performing measurements with the sleeves
attached, be careful to avoid damaging the
sleeves. If the sleeves are inadvertently removed
during measurement, be especially careful in
handling the test leads to avoid electric shock.
・
The tips of the metal pins are sharp, so take care
not to injure yourself.
■
Removing and attaching the sleeves.
Removing the
sleeves
Gently hold the bottom of the sleeves and
pull the sleeves off.Safely store the
removed sleeves so as not to lose them.
Explains how to attach the hand strap, for easy
handling of the instrument in the field.
2.1.3 Turningthe Power On and Off
Power ON
1. Press
POWER
to turn the instrument on. Verify that
all segments of the display light up briefly. Then the
model name is shown, and the bar graph indicates
the battery condition.
The " " indicator lights up when the remaining
battery capacity is low. In this case, the instrument's
reliability is not guaranteed. Replace the battery
immediately.
To check remaining battery life, check the bar
graph, when powering on. (See 2.1.3: Turning the
Power On and Off) The bar graph provides a rough
approximation of remaining battery life. Be careful
for the battery life especially when using the REC
function.
Batteries tend to increase in voltage somewhat when
left unused for a long period. Even if the battery
warning annunciator becomes temporarily off after
the period, replace the battery as soon as possible.
If the battery is not replaced at this point in time,
the annunciator may not light up immediately on the
next occasion. Replace a new battery before it ruins
a measurement or causes some other inconvenience.
(See 2.1.1: Installing or Replacing the Battery)
Low battery voltage detection function
After the
below a certain level, the power goes off
automatically. When this occurs, bAtt and Lo are
displayed.
When power goes off after display of these marks,
replace the exhausted battery with a new one.
Correct measurement may be impossible in the
presence of strong magnetic fields, such as near
transformers and high-current conductors, or in the
presence of strong electromagnetic fields such as
near radio transmitters.
Attach the clamp around only one conductor.
Single-phase (2-wire) or three-phase (3-wire) cables
clamped together will not produce any reading.
○
The display may show a measured value bigger than
the actual value due to a magnet:c field interference.
The interference is less than 2 A during the
measurement.
×
2.2.1 MeasuringDC Current(DC A
1. Pressbutton to display .
2. Switch between the auto range and the manual range
as necessary.
3. Please execute auto zero adjustment before
measuring. Press
zero-adjustment (without clamping the measured
conductor inside the clamp sensor) with the clamp
sensor firmly closed. (see 3.1.1: Auto-zeroadjustment function).
ADJ
annunciator lights to indicate that auto-zeroadjustment is complete. (If you make an auto-zeroadjustment in the auto range, two current ranges will
When 0 is displayed after zero adjustment, the
measureme nt can be performed without the
interference of the count remaining immediately after
the power is turned on.
4. Open the top ends of the clamp core, orient the
current direction indicator on the clamp in the
current direction of the measured conductor, and
clamp the conductor so that it passes through the
center of the clamp core.
Wrong
Current direction
Measured conductor
Current direction
indicator
NOTE
The DC A mode permits only DC current
measurements that does not include the AC
component (See 1.5: Modes).
The 200 A range will display up to 250 A, however,
only the range from 10 A to 200 A can be displayed
with guaranteed accuracy.
At any range, gross errors may occur at 1% or
below of the range, whose accuracy is not
guaranteed, as a result of internal corrective
calculations.
The value may sometimes not show 0 when the
power is switched on, but this is not an instrument
fault. Please execute auto zero adjustment before
measuring.
Just after suspension of input, or when modes are
switched under no input, the counter would not
become zero for about 10 seconds. This is normal
and simply reflects the workings of the internal
circuit. But you can measure with guaranteed
accuracy before the counter becomes zero.
Depending on ambient temperatures, the counter
would not become zero under no input. If this
happens, perform a zero-cancel correction (See
3.1.2: Zero-Cancel Correction Function).
During a f.s. input, the measurement response speed
is about 250 ms during rise (0% to 90%) and about
500 ms (100% to 10%) during fall (See: Figs. 1 and
2).
The AC A mode does not allow measurement of DC
waveforms, full-wave rectification waveforms, halfwave rectification waveforms, or DC+AC
waveforms (See 1.5: Modes).
The 200 A range will display up to 250 A, however,
only the range from 10 A to 200 A can be displayed
with guaranteed accuracy.
At any range, gross errors may occur at 1% or
below of the range ,whose accuracy is not
guaranteed, as a result of internal corrective
calculations.
2. Switch between the auto range and the manual range
as necessary.
3. Please execute auto zero adjustment before
measuring. Press
zero-adjustment (without clamping the measured
conductor inside the clamp sensor) with the clamp
sensor firmly closed. (see 3.1.1: Auto-zeroadjustment function).
ADJ
annunciator lights to indicate that auto-zeroadjustment is complete. (If you make an auto-zeroadjustment in the auto range, two current ranges will
When 0 is displayed after zero adjustment, the
measureme nt can be performed without the
interference of the count remaining immediately after
the power is turned on.
4. If the counter fails to become zero under no input,
press
HOLD
button and then press
0ADJ/RESET
button to perform a zero-cancel correction.
5. Open the top ends of the clamp core and clamp the
measured conductor so that it passes through the
center of the clamp core.
NOTE
Just after suspension of input, or when modes are
switched under no input, the counter would not
become zero for about 10 seconds. This is normal
and simply reflects the workings of the internal
circuit. But you can measure with guaranteed
accuracy before the counter becomes zero.
Depending on ambient temperatures, the counter
would not become zero under no input. If this
happens, preform a zero-cancel correction (See
3.1.2: Zero-Cancel Correction Function).
The polarity of the input is not displayed, even if
DC current is measured in this mode. If the clamp
sensor is reoriented, the measured values may
change, but the values are within the guaranteed
accuracy. (In case that you would like to measure a
DC current which doesn't have AC components, you
should make the measurement in DC A mode.)
During a f.s. input, the measurement response speed
is about 250 ms during rise (0% to 90%) and about
500 ms (100% to 10%) during fall (See 2.2.2
Measuring AC Current (AC A), Figs. 1 and 2).
The 200 A range will display up to 250 A, however,
only the range from 10 A to 200 A can be displayed
with guaranteed accuracy.
At any range, gross errors may occur at 1% or below of the
range,whose accuracy is not guaranteed as a result of
internal corrective calculations.
The value may sometimes not show 0 when the power is
switched on, but this is not an instrument fault. Please
execute auto zero adjustment before measuring.
2.2.4 Peak Hold Measurement
The peak value of measured current is retained on
the display to simplify reading. Peak fluctuations
canalsobeviewed.
1. Pressbutton and select a measurement mode for
the measured circuit.
2. In DC A and AC+DC A modes, make an auto-zeroadjustment by
Auto-Aero Adjustment Function).
3. Set to PEAK. The measurement mode is switched
SLOW/PEAK/Hz
by
0ADJ/RESET
button as follows.
button (See 3.1.1:
SLOWPEAKFAST
(The unit
symbol blinks.)
Hz
(Except
DC)
NORMAL
4. Switch between the auto and the manual range as
necessary. (If you are unable to estimate the peak
current value, start at the 2000A range.)
5. Before the measurement, press
0ADJ/RESET
button
to reset the residual data.
6. Open the top ends of the clamp core and clamp the
measured conductor so that it passes through the
center of the clamp core.
The polarity of the input is not displayed during
peak measurements. The measured values may
change if the clamp sensor is reoriented, but the
values are within the guaranteed accuracy.
You can capture from pulses with a width of about
more than 1ms.
Even after clamping, press
0ADJ/RESET
button to
reset the data as necessary.
In case that the counter doesn't become zero under
no input in peak measurement mode, even though
you pressed
0ADJ/RESET
button to reset the peak
data, the clamp sensor may be magnetized. Quit the
peak measurement mode, and perform the auto-zero
adjustment by
0ADJ/RESET
button . Then make
the settings again. (A few counts would remain,even
if you push
0ADJ/RESET
button .)
The hold value does not change, unless a larger
value is measured, but be careful to avoid accidental
loss of data resulting from the auto power-off
function. (See 3.5: Automatically Turning Power
off (Auto Power-Off Function APS))
Use the REC function to make measurements longer
than the auto power OFF time.
To check transitional peak value, press
MAX/MIN
button to shift to the instantaneous value (no
annunciator).
For safety reasons, when taking measurements, only
use the L9207-10 Test Lead provided with the
instrument.
2.3.1 MeasuringDC Voltage(DC V
1. Pressbutton to display .
2. Insert the red test lead to V/Ω and the black test
lead to COM of the voltage/resistance input
terminal.
3. Switch between the auto range and the manual range
as necessary.
4. In case that the counter isn't zero, press
0ADJ/RESET
adjustment.
auto-zero-adjustment is complete. (See 3.1.1: AutoZero-Adjustment Function)
5. Attach or remove the rigid insulating sleeve as
required by the measurement category.
6. Carefully contact the test leads to a circuit.
NOTE
Be sure to use the test leads with the sleeves
attached when performing measurements in the CAT
III measurement category. In a CAT II environment,
if the tips of the test leads do not reach the
measurement object, remove the rigid insulating
sleeve before measuring.
You can perform the auto-zero-adjustment up to 4%
of the range.
A lit - annunciator indicates that potential is higher
at the black test lead than at the red test lead.
The DC V mode permits only DC voltage
measurements that does not include the AC
component (See 1.5: Modes).
Every range will display up to 125% of the range,
however, only the range from 10% to 100% can be
displayed with guaranteed accuracy.
At any range, gross errors may occur at 1% or belo w
of the range, whose accuracy is not guaranteed, as a
result of internal corrective calculations.
2.3.2 MeasuringAC Voltage(AC V
1. Press
2. Insert the red test lead to V/Ω and the black test lead
to COM of the voltage/resistance input terminal.
3. Switch between the auto range and the manual range
as necessary.
4. Attach or remove the rigid insulating sleeve as
required by the measurement category.
5. Carefully contact the test leads to a circuit.
NOTE
Be sure to use the test leads with the sleeves
attached when performing measurements in the CAT
III measurement category. In a CAT II environment,
if the tips of the test leads do not reach the
measurement object, remove the rigid insulating
sleeve before measuring.
Just after suspension of input, or when modes are
switched under no input, the counter would not
become zero for about 10 seconds. This is normal
and simply reflects the workings of the internal
circuit. But you can measure with guaranteed
accuracy before the counter becomes zero.
Depending on ambient temperatures, the counter would
not become zero under no input, if this happens, press
HOLD
button and then press
0AD J /RESET
button to
perform a zero-cancel correcti on. (See 3.1.2: ZeroCancel Correction Function)
During a f.s. input, the measurement response speed
is about 250 ms during rise (0% to 90%) and about
500 ms (100% to 10%) during fall (See 2.2.2:
Measuring AC Current (AC A), Figs. 1 and 2).
The AC V mode does not allow measurement of DC
waveforms, full-wave rectification waveforms, halfwave rectification waveforms, or DC+AC
waveforms (See 1.5: Modes).
Every range will display up to 125% of the range,
however, only the range from 10% to 100% can be
displayed with guaranteed accuracy.
At any range, gross errors may occur at 1% or below
of the range, whose accuracy is not guaranteed, as a
result of internal corrective calculations.
2.3.3 MeasuringAC/DC Voltage(AC+DC V
1. Press
2. Insert the red test lead to V/Ω and the black test lead
to COM of the voltage/resistance input terminal.
3. Switch between the auto range and the manual range
as necessary.
4. If the counter fails to become zero under no input,
press
button to perform a zero-cancel correction. (See
3.1.2: Zero-Cancel Correction Function)
5. Attach or remove the rigid insulating sleeve as
required by the measurement category.
Be sure to use the test leads with the sleeves
attached when performing measurements in the CAT
III measurement category. In a CAT II environment,
if the tips of the test leads do not reach the
measurement object, remove the rigid insulating
sleeve before measuring.
Just after suspension of input, or when modes are
switched under no input, the counter would not
become zero for about 10 seconds. This is normal
and simply reflects the workings of the internal
circuit. But you can measure with guaranteed
accuracy before the counter becomes zero.
The polarity of the input is not displayed, even if DC
voltage is measured in this mode. If the connections
of test leads are moved, the measured values may
change, but the values are within the guarantee d
accuracy. (In case that you would like to measure DC
voltage which doesn't have AC components, you
should make the measurement in DC V mode.)
During a f.s. input, the measurement response speed
is about 250 ms during rise (0% to 90%) and about
500 ms (100% to 10%) during fall (See 2.2.2:
Measuring AC Current (AC A), Figs. 1 and 2).
Every range will display up to 125% of the range,
however, only the range from 10% to 100% can be
displayed with guaranteed accuracy.
At any range, gross errors may occur at 1% or below
of the range, whose accuracy is not guaranteed, as a
result of internal corrective calculations.
2.3.4 Peak Hold Measurement
The peak value of measured voltage is retained on the display
to simplify reading. Peak fluctuations can also be viewed.
4. Switch between the auto and the manual range as
necessary. (If you are unable to estimate the peak
voltage value, start at the 600 V range.)
5. Attach or remove the rigid insulating sleeve as
required by the measurement category.
6. Carefully contact the test leads to a circuit.
7. Before the measurement, press
0ADJ/RESET
to reset the residual data.
NOTE
Be sure to use the test leads with the sleeves
attached when performing measurements in the CAT
III measurement category. In a CAT II environment,
if the tips of the test leads do not reach the
measurement object, remove the rigid insulating
sleeve before measuring.
The polarity of the input is not displayed during
peak measurements. If the connections of test leads
are moved, the measured values may change, but
the values are within the guaranteed accuracy.
You can capture from pulses with a width of about
more than 1ms.
To check transitional peak value, press
button to shift to the instantaneous value (no
annunciato r).
The zero adjustment is not effective at peak hold
measurement mode.
2. If the current range of the measured circuit is
known, set the current range to the manual range.
SLOW/PEAK/Hz
3.
as follows. Select Hz by pressing the button. (The
unit symbol A blinks, and a current value is
displayed on the bar graph.)
SLOWPEAKFAST
(The unit
symbol blinks.)
4. Switch the auto range and the manual range as
necessary.
5. Open the top ends of the clamp core and clamp the
measured conductor so that it passes through the
center of the clamp core.
NOTE
At the 100 Hz and 1000 Hz ranges, ---- appears on
the counter when the frequency is lower than 10 Hz.
---- appears on the counter, if the frequency is lower
than 1 Hz.
O. L. appears on the counter, if the frequency is
higher than 1 kHz.
If an input value is significantly lower than the range,
an accurate measurement may not be achieved,
resulting in ----, O. L. or display fluctuations.
The 10 Hz range or 100 Hz range will display up to
125% of each range, however, only the range from
10% to 100% can be displayed with guaranteed
accuracy.
The frequencies, whose waveforms are special such
as inverters, would not be measurable, when the
carrier frequencies are lower than several kHz.
Full-wave rectification indicates twice the actual
value, due to an AC coupling in the internal circuit.
It would take time to stabilize the counter,
depending on the frequency range or the input
frequency.
2.4.2 FrequencyMeasurement in VoltageMode
CAUTION
For safety reasons, when taking measurements, only
use the L9207-10 Test Lead provided with the
instrument.
1. Press
button and select AC or AC+DC,
depending on the circuit to be measured.
2. If the voltage range of the measured circuit is
known, set the voltage range to the manual range.
3. Insert the red test lead to V/Ω and the black test
lead to COM of the voltage/resistance terminal.
SLOW/PEAK/Hz
4.
button switches the annunciators
as follows. Select Hz by pressing the button. (The
unit symbol V blinks, and a voltage value is
displayed on the bar graph.)
5. Switch between the auto range and the manual range
as necessary.
6. Attach or remove the rigid insulating sleeve as
required by the measurement category.
7. Carefully contact the test leads to a circuit.
NOTE
Be sure to use the test leads with the sleeves
attached when performing measurements in the CAT
III measurement category. In a CAT II environment,
if the tips of the test leads do not reach the
measurement object, remove the rigid insulating
sleeve before measuring.
At the 100 Hz and 1000 Hz ranges, ---- appears on
the counter when the frequency is lower than 10 Hz.
---- appears on the counter, if the frequency is lower
than 1 Hz.
O. L. appears on the counter, if the frequency is
higher than 1 kHz.
If an input value is significantly lower than the
range, on accurate measurement may not be
achieved, resulting in ----, O. L. or display
fluctuations.
The 10 Hz range or 100 Hz range will display up to
125% of each range, however, only the range from
10% to 100% can be displayed with guaranteed
accuracy.
The frequencies, whose waveforms are special such
as inverters, would not be measurable, when the
carrier frequencies are lower than several kHz.
Full-wave rectification indicates twice the actual
value, due to an AC coupling in the internal circuit.
It would take time to stabilize the counter,
depending on the frequency range or the input
frequency.
Never apply voltage to the test leads when the
Resistance measurement mode is selected. Doing
so may damage the instrument and result in
personal injury.
To avoid electrical accidents, remove power from
the circuit before measuring.
CAUTION
For safety reasons, when taking measurements, only
use the L9207-10 Test Lead provided with the
instrument.
1. Press
Ω/
button to display Ω. Auto-ranging is
used for resistance measurement.
2. Insert the red test lead to V/Ω and the black test
lead to COM of the voltage/resistance terminal.
3. Attach or remove the rigid insulating sleeve as
required by the measurement object.
4. Touch the tips of the red and black test leads
together and verify that 0 Ω is displayed.
5. Then connect the test leads to the test points for
measurement.
If voltage is applied inadvertently, the internal
circuitry is protected up to 600 VAC.
If voltage larger than approximately 2 or 3 V is
input, the Resistance measurement mode cannot
be selected.
If the display shows larger than 7 or 8 A at 200 A
range in the DCA mode, the Resistance
measurement mode cannot be selected. Press the
Never apply voltage to the test leads when the
Continuity check mode is selected. Doing so may
damage the instrument and result in personal
injury.
To avoid electrical accidents, remove power from
the circuit before measuring.
CAUTION
For safety reasons, when taking measurements, only
use the L9207-10 Test Lead provided with the
instrument.
1. Press
Ω/
button to display. Auto-ranging is
used for continuity checking.
2. Insert the red test lead to V/Ω and the black test lead
to COM of the voltage/resistance input terminal.
3. Attach or remove the rigid insulating sleeve as
required by the measurement object.
4. Touch the tips of the red and black test leads
together and verify that 0 Ω is displayed.
The auto-zero-adjustment function is used to adjust
offsets in the internal circuit automatically that
result from temperature characteristics or clamp
sensor magnetization. The clamp core is
magnetized during a large DC current measurement,
or when a powerful magnet is placed close to the
clamp core.
1. Wait until the counter is stable under no input.
Then, press
lights.
You can perform the auto-zero-adjustment, if the
counter displays within45 A in a current mode.
When there is an input or the counter decreases, the
measurement accuracy will be spoiled by pressing
0ADJ/RESET as well as the accurate auto-zeroadjustment. If inaccurate auto-zero-adjustment is
performed, perform the correct procedure again.
It would take approximately 20 seconds to stabilize
the counter in AC+DC A mode.
Use the zero-cancel correction function if the
counter fails to revert to zero after correct auto-zeroadjustment in AC+DC A mode.
In a current mode, if you press
0ADJ/RESET
button
again during the auto-zero-adjustment in the internal
circuit, the auto-zero-adjustment is canceled.
The auto-zero-adjustment is not effective at auto
range in DCV mode. If you perform the auto-zeroadjustment at manual range, the auto-zeroadjustment is invalid by changing the range.
3.1.2 Zero-Cancel Correction Function
Use the zero-cancel correction function when the
counter fails to become zero under no input in AC
A, AC+DC A, AC V, or AC+DC V mode.
NOTE
1. Press
2. Press
HOLD
button to display
0ADJ/RESET
button.
When there is an input or the counter decreases, the
measured values will be evaluated lower by pressing
0AJD/RESET
button.
HOLD
annunciator.
ADJ
annunciator blinks.
If the counter is zero, the zero-cancel correction
function does not work.
In AC+DC A mode, the zero-cancel correction
function does not work, unless auto-zero-adjustment
is complete.
The zero-cancel correction function is not effective
at auto range in DCV mode. If you perform the
zero-cancel correction at manual range, the zerocancel correction is invalid by changing the range.
The counter is updated twice per second when
powering on. The counter update may be altered
according to measurement conditions.
SLOW/PEAK/Hz
follows:
button changes an annunciator as
SLOWPEAK
FAST
(The unit
annunciator blinks.)
Hz
(Except DC)
NORMAL
3.3.1 SLOW mode
If the counter fluctuates rapidly and is hard to read,
you can select a slower update rate (once every 3
seconds) by pressing
SLOW/PEAK/Hz
button.
3.3.2 FAST mode
For current measurements and voltage
measurements, the counter is updated four times per
second in FAST mode. You can measure abrupt
changes such as starting currents.
The unit symbol A or V blinks.
To facilitate reading when measuring a starting
current, use the record (REC) function to hold the
maximum value (MAX).
3.4 Storing Values in Memory
(Recording Function REC)
Use the recording function to hold the maximum
and minimum measured values and
maximum/minimum averages. (Not available for
resistance measurements.)
1. Measurement indicated value
Pressing the
of current or voltage activates the recording
function. REC flashes and the instrument saves the
maximum value (MAX), minimum value (MIN),
and average value (AVE) in internal memory from
the instant you press the
Pressing the
function activated switches the display as shown
below. If MAX, MIN, or AVE is not displayed, an
instantaneous value is assumed.
Data (MAX, MIN, AVE) remains displayed while
the display is switched. If maximum or minimum
data is updated in the meantime, however, the data
values will change.
With the recording function activated, the auto
power-off function remains disabled. (APS off.)
The average value (AVE) displayed is calculated by:
Average Value = [(Maximum value + Minimum
Value)/2].
If the recording function is activated and
Instantaneous value (no annunciator) selected after
you activate PEAK mode with the
SLOW/PEAK/Hz
button, you can see the fluctuation of the peak.
2. Display of Elapsed Time
When you press the
MAX/MIN
button to activate
the recording function, the bar graph segments flash
and the elapsed time appears.
When "min" is shown in the right-hand corner of
the bar graph, each segment of the bar graph
corresponds to one minute.
Every time one minute elapses, one segment of the
flashing bar graph goes on. When all segments on
the bar graph go on, the elapsed time is 30 minutes.
When the elapsed time exceeds 30 minutes, one
segment of the flashing bar graph goes off every
time one minute elapses.
When the segments left of a flashing segment
remain on: the number of "on" segments represents
the elapsed time (0 ~ 29).
The illustration below shows when 20 minutes have
elapsed:
When the segments right of a flashing segment
remain on: the number of "off" segments (+30)
represents the elapsed time (30 ~ 59).
The illustration below shows when 50 minutes have
elapsed:
When digital display switches the average value
(AVE) to a instantaneous value when you press the
MAX/MIN
button, the right corner of the bar graph
indicates hours. In this mode, each segment of the
bar graph corresponds to one hour. The way to read
the bar graph here is similar to reading it in
minutes. When all bar graph segments remain on,
the elapsed time is 29 hours.
The illustration below shows when one hour, 40
minutes have elapsed.
3. Deactivation of Recording Function
Pressing the
recording function.
HOLD
button deactivates the
HOLD
goes on, REC stops
flashing and goes on, and the elapsed time stops
incrementing.
While the recording function is being deactivated,
data is not updated, even if the clamp sensor is
disconnected from the conductor.
Pressing the
HOLD
button again cancels
HOLD
display and activates the recording function again,
with REC flashing again.
To cancel the recording function, press the related
function button (
or) for the measurement
in progress. Once the recording function is
canceled, the auto power-off function becomes
effective. (
APS
goes on.)
NOTE
For a long term measurement, check how much the
battery power remains. (Shown on the bar graph at
power on.)
When starting the recording function (REC)inan
auto range, the range is set as the range of when
that pushed
MAX/MIN
button.
When you need minimum value and average value
data, make sure to activate the recording function
during measurement. If the function is activated
when there is no input, the minimum value will
remain zero. To deactivate the recording function,
press the
HOLD
button and read min imum value
and average value to terminate measurement. If you
disconnect the clamp or test lead from the circuit
under measurement without deactivating the
recording function beforehand, the minimum value
will be zero.
When the instrument is turned off, accumulated data
are lost.
annunciator is displayed, the auto
power-off function is active.
If no button is pressed for about 10 minutes, the
instrument turns itself off automatically.
Immediately before turning off automatically,
annunciator blinks and a beep tone is heard for
about 30 seconds.
By pressing any button except
POWER
button, you
will extend the powered state for another 10
minutes.
Procedure for disabling the auto power-off
function.
Press
POWER
button with holding down
HOLD
button, when you turn power on.
Use the recording function (REC) by pressing
To clean the instrument, wipe it gently with a soft cloth
moistened with water or mild detergent. Never use solvents
such as benzene, alcohol, acetone, ether, ketones, thinners
or gasoline, as they can deform and discolor the case.
Measurements are degraded by dirt on the mating surfaces
of the clamp-on sensor, so keep the surfaces clean by gently
wiping with a soft cloth.
To avoid corrosion from battery leakage, remove the batteries
from the instrument if it is to be stored for a long time.
5.2After service
The minimum stocking period for replacement parts is five
years after end of production.
If damage is suspected, check the "5.3 Troubleshooting"
section before contacting your dealer or Hioki
representative.
When sending the instrument for repair, remove the
batteries and pack carefully to prevent damage in transit.
Include cushioning material so the instrument cannot move
within the package. Be sure to include details of the
problem. Hioki cannot be responsible for damage that
occurs during shipment.
Check that it has sufficient remaining power. (See 2.1.4:
Battery Low Warning (When
mark lights) )
The counter doesn't become zero:
In DCA, AC+DCA, or DCV mode
Use the auto-zero adjustment function (See 3.1.1).
(If this occures in AC+DCA mode after performing the
auto-zero-adjustment, perform the zero cancel correction
function.)
In ACA, ACV, or AC+DCV mode
Use the zero-cancel correction function (See 3.1.2).
Resistance measurement or continuity check
If 0 Ω is displayed when the red and black test leads are
unconnected (open-circuit), internal circuitry may be
damaged, requiring repair service.
The measured value is larger than the estimated
value.
Current / Voltage / Frequency measurement:
Check that you're using the proper range.
Examine the waveform to confirm that no components
but the estimated frequency are being used.
Check that the counter has been reset with
0ADJ/RESET
button before peak measurement.
Look for magnetic fields, electrical fields or possible
noise sources near the instrument.
Check the waveform. Some special frequencies can't be
measured, such as those of inverters. (At frequency
measurement)
The measured value is smaller than the estimated
value.
Current / Voltage measurement:
Check that the clamp sensor is firmly closed. (At current
measurement)
Check that the test leads are fully connected. (At voltage
measurement)
Check that the frequency of the measured circuit is
within the range provided in the specifications. (A
smaller value will be displayed for a high inverter carrier
frequency.) *Items within parentheses apply to current
measurement only
Check that the proper steps have been taken, according to
procedure described in 3.1.1: Auto-zero adjustment
function and 3.1.2: Zero-cancel correction function.
Check that you're using the proper mode. (See 1.5:
Modes.)
Check that the peak value is below the circuit dynamic
value provided in the specifications. (To check
transitional peak values, see Notes in 2.2.4 and 2.3.4:
Peak Hold Measurement.)
Check that the crest factor (peak value / RMS value) is
below the circuit dynamic value provided in the
specifications.
Check that the battery warning annunciator is off.
Frequency measurement:
Check the waveform. Some special frequencies can't be
measured, such as those of inverters.
Check that the input value corresponds to 10% or more
of the range.
Check that the measured circuit is stable.
During voltage measurements, check that the test leads
are fully connected.
Check the waveform. Some special frequencies can't be
measured, such as those of inverters.
The displayed value is out of range O.L.
Resistance measurement or continuity check
If O.L. is displayed when the test leads are shorted
together, a test lead may be broken or internal circuitry
may be damaged.
Check for broken test leads by taking a voltage
measurement. If a test lead is broken, replace with
Model L9207-10 Test Lead.
Current measurment
o If nothing can be measured (no input detected) the
sensor element or internal circuitry may be damaged,
requiring repair service. If the damage is to internal
circuitry, the displayed value may appear over range
O.L. during voltage measurement.
Voltage measurement
If over range is indicated when nothing is being
measured, internal circuitry may be damaged.