Thank you for purchasing the HIOKI "3285 CLAMP
ON AC/DC HiTESTER." To obtain maximum
performance from the instrument/ device/ product,
please read this manual first, and keep it handy for
future reference.
Request
We have tried to bring this manual as close to
perfection as we could achieve. If perchance you
find any unclear portions, mistakes, omissions, or
the like, we would be most obliged if you could
please notify us of them via any HIOKI agent, or
directly.
When the unit is delivered, check and make sure
that it has not been damaged in transit. In
particular, check the accessories, panel switches,
keys, and terminals.
If the unit is damaged, or fails to operate according
to the specifications, contact your dealer or HIOKI
representative.
Check the 3285 Unit and the Supplied
Accessories
Main unit
3285 CLAMP ON AC/DC HiTESTER
Supplied accessories
9345 CARRYING CASE 1
9207-10 TEST LEAD (red and black) 1
Hand Strap 1
6F22(006P) 1
Instruction manual 1
This instrument is designed to comply with IEC
61010 Safety Standards, and has been
thoroughly tested for safety prior to shipment.
However, mishandling during use could result in
injury or death, as well as damage to the
instrument. Be certain that you understand the
instructions and precautions in the manual
before use. We disclaim any responsibility for
accidents or injuries not resulting directly from
instrument defects.
This manual contains information and warnings
essential for safe operation of the instrument and for
maintaining it in safe operating condition. Before
using it, be sure to carefully read the following
safety precautions.
Thesymbol printed on the instrument
indicates that the user should refer to a
corresponding topic in the manual (marked
with thesymbol) before using the
relevant function.
In the manual, thesymbol indicates
particularly important information that the
user should read before using the instrument.
Indicates AC (Alternating Current).
Indicates DC (Direct Current).
Indicates both DC (Direct Current) and AC
(Alternating Current).
Indicates a device which is doubleinsulated.
Indicates that the instrument may be connected
to or disconnected from a live circuit.
f.s. (maximum display or scale value, or length of scale)
Signifies the maximum display (scale) value or the
length of the scale (in cases where the scale consists
of unequal increments or where the maximum value
cannot be defined).
In general, this is the range value (the value written
on the range selector or equivalent) currently in use.
rdg. (displayed or indicated value)
Signifies the value actually being measured, i.e., the
value that is currently indicated or displayed by the
measuring instrument.
dgt. (resolution)
Signifies the smallest display unit on a digital
measuring instrument, i.e., the value displayed when
the last digit on the digital display is "1".
This instrument complies with CAT III safety
requirements.
To ensure safe operation of measurement
instruments, IEC 61010 establishes safety standards
for various electrical environments, categorized as
CAT I to CAT IV, and called measurement
categories. These are defined as follows.
CAT I : Secondary electrical circuits connected to
an AC electrical outlet through a
transformer or similar device.
CAT II : Primary electrical circuits in equipment
connected to an AC electrical outlet by a
power cord (portable tools, household
appliances, etc.)
CAT III : Primary electrical circuits of heavy
equipment (fixed installations) connected
directly to the distribution panel, and
feeders from the distribution panel to
outlets.
CAT IV : The circui t from the service drop to the
service entrance, and to the power meter
and primary overcurrent protection device
(distribution panel).
Higher-numbered categories correspond to electrical
environments with greater momentary energy. So a
measurement device designed for CAT III
environments can endure greater momentary energy
thanadevicedesignedforCATII.
Using a measurement instrument in an environment
designated with a higher-numbered category than
that for which the instrument is rated could result in
a severe accident, and must be carefully avoided.
Never use a CAT I measuring instrument in CAT II,
III, or IV environments.
The measurement categories comply with the
Overvoltage Categories of the IEC60664 Standards.
Follow these precautions to ensure safe operation
and to obtain the full benefits of the various
functions.
DANGER
・ Use clamp testers only on power lines up to 600
Vrms, to avoid short-circuits and accidents that
could result in injury or death.
・ Always connect the clamp sensor to the
secondary side of a breaker. On the secondary
side of a breaker, even if the lines are shorted
the breaker can trip and prevent an accident.
On the primary side, however, the current
capacity may be large, and in the event of a
short-circuits there may be a serious accident.
・ When using an AC adapter, use only the
specified HIOKI model 9445-02 (SA10-0910N,
SINO-AMERICAN) or 9445-03 (for EU) (SA100910G, SINO-AMERICAN).
product to get wet, and do not use it when your
hands are wet.
・ To avoid electric shock when measuring live lines,
wear appropriate protective gear, such as insulated
rubber gloves, boots and a safety helmet.
・ During current measurement, to avoid an
electric shock accident, do not connect the test
leads to the product.
・ Do not input voltage exceeding 600 Vrms (1000
V max.).
・ To avoid electric shock when replacing the
batteries, first disconnect the clamp portion
from the object to be measured. Also, before
using the product after replacing the batteries,
replace the cover and screw.
・ When replacing the batteries, be sure to insert
them with the correct polarity. Otherwise, poor
performance or damage from battery leakage
could result.
・ To avoid the possibility of explosion, do not short
circuit, disassemble or incinerate batteries.
・ Handle and dispose of batteries in accordance
with local regulations.
・Do not use the instrument where it may be exposed
to corrosive or combustible gases. The instrument
may be damaged or cause an explosion.
・ This is a precision instrument: do not clamp any
foreign objects in the end of the clamp core, or
insert anything in the core gap.
・ To avoid damage to the product, protect it from
vibration or shock during transport and handling,
and be especially careful to avoid dropping. Do not
exert excessive pressure on the clamp sensor or
attempt to wedge the sensor into a tight spot for
measurement.
・ Before using the product the first time, verify that it
operates normally to ensure that the no damage
occurred during storage or shipping. If you find any
damage, contact your dealer or HIOKI
representative.
・ To avoid damage to the product, do not exceed the
maximum input current rating, which depends on
the frequency of the current being measured (see
Fig.4 of Chapter 3) Be careful about the evolution
of heat, when the input frequency is high.
・ Do not use the product if the battery is exhausted
(when the
sure to replace the exhausted battery with a new
one.
・ When replacing the battery, make sure that the
metal battery snap fitting is firmly connected. If the
metal fitting is loose, adjust it and recheck the
connection. If it isn't connected securely, the power
may not be turned on, and a power may be turned
off during the use.
power is turned off before plugging in or unplugging
the AC adapter.
・ This product is designed for indoor use, and operates
reliably from 0
・ Do not store or use the product where it could be
exposed to direct sunlight, high temperature or
humidity, or condensation. Under such conditions,
the product may be damaged and insulation may
deteriorate so that it no longer meets specifications.
・ Accurate measurement may be impossible in
NOTE
locations subject to strong external magnetic fields,
such as transformers and high-current conductors, or
in locations subject to strong external electric fields,
such as radio transmission equipment.
・ Gently wipe dirt from the surface of the unit with a
soft cloth moistened with a small amount of water
or mild detergent.
Do not try to clean the unit using cleaners
containing organic solvents such as benzine, alcohol,
acetone, ether, ketones, thinners, or gasoline. They
may cause discoloration or damage.
・ When not in use for a long time, to prevent possible
corrosion caused by battery leakage, remove the
batteries before storage.
The 3285 CLAMP ON AC/DC HiTESTER makes it
possible to measure DC, AC or AC+DC current in
live power lines without tapping into or connecting
the lines. Using a one-chip microprocessor, the
tester provides many functions, including an
automatic zero adjust function that changes a
troublesome task into a one-touch operation. An AC
adapter connection terminal and an output terminal
areequippedsothatyouareabletomeasureby
connecting to other instruments such as recorders.
A Current
hour 1 hour/segment (bar graph)
min 1 minute/segment (bar graph)
Input over (bar graph)
⑬ Output terminal
Connected to the optional 9094 OUTPUT CORD to
provide output during a current measurement or a
frequency measurement in a current mode.
⑭ AC adapter connection terminal
Connected to the optional 9445-02 or 9445-03 AC
ADAPTER to perform measurements for no battery
or a long term measurement.
⑮ Voltage measurement terminal (V and COM
terminals)
Connected to the 9207-10 TEST LEAD (red and
black, supplied with the unit) to measure voltage.
⑯ Slide knob
Slide up to use the voltage measurement terminal or
slide down to use the output terminal or AC adapter
connection terminal. Move until a click is heard.
1. Remove the rear cover and insert a battery. (Refer to
"Chapter 4 Battery Replacement".)
2. Press
3. The DC current measurement mode is activated.
【Low battery voltage detection function】
POWER
segments of the display light up briefly. Then the
model name is shown, and the bar graph indicates
the battery condition.
After the
below a certain level, the power goes off
automatically. When this occurs, bAtt and Lo are
displayed.
When power goes off after display of these marks,
replace the exhausted battery with a new one.
locations subject to strong external magnetic fields,
such as transformers and high-current conductors, or
in locations subject to strong external electric fields,
such as radio transmission equipment.
・ Make sure that only one conductor is clamped in
the center of the clamp sensor. If you clamp singlephase (2-wire) or three-phase (3-wire) lines together,
it will be impossible to measure.
・ The display may show a measured value bigger than
the actual value due to a magnet:c field interference.
The interference is less than 2 A during the
measurement.
1. Press
2. Switch between the auto range and the manual range
as necessary.
3. Press
(without clamping the measured conductor inside
the clamp sensor) with the clamp sensor firmly
closed. (see 2.5.1: Auto-zero-adjustment function).
ADJ
adjustment is complete. (If you make an auto-zeroadjustment in the auto range, two current ranges will
be adjusted in this mode.)
4. Open the top ends of the clamp core, orient the
current direction indicator on the clamp in the
current direction of the measured conductor, and
clamp the conductor so that it passes through the
center of the clamp core.
・ The 200 A range will display up to 250 A, however,
only the range from 10 A to 200 A can be displayed
with guaranteed accuracy.
・ At any range, gross errors may occur at 1% or
below of the range, whose accuracy is not
guaranteed, as a result of internal corrective
calculations.
2.2.2 MeasuringAC Current(AC A
1. Press
2. Switch between the auto range and the manual range
as necessary.
3. Open the top ends of the clamp core and clamp the
measured conductor so that it passes through the
center of the clamp core.
to display.
・ Just after suspension of input, or when modes are
NOTE
switched under no input, the counter would not
become zero for about 10 seconds. This is normal
and simply reflects the workings of the internal
circuit. But you can measure with guaranteed
accuracy before the counter becomes zero.
would not become zero under no input. If this
happens, perform a zero-cancel correction (2.5.2:
Zero-cancel correction function).
・ During a f.s. input, the measurement response speed
is about 250 ms during rise (0% to 90%) and about
500 ms (100% to 10%) during fall (2.2.5, Figs. 1
and 2).
・ The AC A mode does not allow measurement of DC
waveforms, full-wave rectification waveforms, halfwave rectification waveforms, or DC+AC
waveforms (See 1.5: Modes).
・ The 200 A range will display up to 250 A, however,
only the range from 10 A to 200 A can be displayed
with guaranteed accuracy.
・ At any range, gross errors may occur at 1% or
below of the range ,whose accuracy is not
guaranteed, as a result of internal corrective
calculations.
2.2.3 MeasuringAC/DC Current(AC+DC A
1. Press
2. Switch between the auto range and the manual range
as necessary.
3. Press
(without clamping the measured conductor inside
the clamp sensor) with the clamp sensor firmly
closed. (see 2.5.1: Auto-zero-adjustment function).
ADJ
adjustment is complete.
4. If the counter fails to become zero under no input,
press
perform a zero-cancel correction.
5. Open the top ends of the clamp core and clamp the
measured conductor so that it passes through the
center of the clamp core.
・ Just after suspension of input, or when modes are
NOTE
switched under no input, the counter would not
become zero for about 10 seconds. This is normal
and simply reflects the workings of the internal
circuit. But you can measure with guaranteed
accuracy before the counter becomes zero.
・ Depending on ambient temperatures, the counter
would not become zero under no input. If this
happens, preform a zero-cancel correction (2.5.2:
Zero-cancel correction function).
・ The polarity of the input is not displayed, even if
DC current is measured in this mode. If the clamp
sensor is reoriented, the measured values may
change, but the values are within the guaranteed
accuracy. (In case that you would like to measure a
DC current which doesn't have AC components, you
should make the measurement in DC A mode.)
・ During a f.s. input, the measurement response speed
is about 250 ms during rise (0% to 90%) and about
500 ms (100% to 10%) during fall (2.2.5, Figs. 1
and 2).
・ The 200 A range will display up to 250 A, however,
only the range from 10 A to 200 A can be displayed
with guaranteed accuracy.
・ At any range, gross errors may occur at 1% or
below of the range,whose accuracy is not guaranteed
as a result of internal corrective calculations.
・ When displaying the current measured value during
a frequency output, the auto-zero-adjustment is
unavailable for the current measurement.
4. Set a range based on the unit's measurement range
and other instruments, such as recorders. A
conversion table for ranges is provided below.
Range /DIV10 mV 20 mV 50 mV 0.1 V 0.2 V 0.5 V 1V
2000 A range20 A40 A 100 A 200 A 400 A 1000 A 2000 A
200 A range2A4A 10 A20 A40 A 100 A 200 A
※ The figures are current values per DIV of a measuring instrument,
WARNING
such as a recorder.
To avoid damage to the unit, do not short the
output terminal and do not input voltage to the
output terminal.
・ Before using the output function, be sure to press
NOTE
OUTPUT
, confirming that either
RECorMON
annunciator is on. Output is made even when both
annunciators are off, but the power will be
automatically off in approximately 10 minutes, since
the auto power-off function is activated.
・ If
OUTPUT
is pressed in an auto range (AUTO),
the current range is set as the key is pressed.
(AUTO annunciator is turned off.)
・ Errors may occur or output values unless an auto-
zero-adjustment is done in DC A mode.
・ The zero-cancel function does not affect output.
Thus, voltage would be generated because of an
ambient temperature, even if there is no input.
・ When the AC adapter is used and there is a large
outputs are analog outputs. The output
REC
response time during a f.s. input differs between rise
(0% to 90%, about 250 ms) and fall (100% to 10%,
about 500 ms). (Figs 2 and 3) As the measured
value is small to the range response time becomes
long.
measurement, frequency measurement, recording,
and data holding are possible when generating
current measurement output. (But output will be
changed during mode shift, range shift, shift to
voltage measurement mode, or auto-zeroadjustment.)
the unit to a recorder.
Low impedances will affect indicated values.
measurements, output is still for the frequency. To
obtain a current output, press
the frequency output, then make new settings.
9445-02 or 9445-03 AC ADAPTER.
amount of noise in the power line, the display may
show several counts or noise may be present in the
output. In such a case, connect the ground terminal
of the level recorder or the L side of the input to
ground.
2. Slide the slide cover up using the slide knob. Next,
insert the red test lead to V and the black test lead
to COM of the voltage measurement terminal.
3. Switch between the auto range and the manual range
as necessary.
4. If the counter does not show zero, switch to the range
you will use (manual range), and press
to perform an auto-zero adjustment.
annunciator lights to indicate that auto-zero-
ADJ
adjustment is complete.
5. Carefully contact the test leads to a circuit.
・ You can perform the auto-zero-adjustment up to 4%
NOTE
of the range.
・ When you switch ranges after auto-zero adjustment,
deviation of the adjustment value will prevent
accurate measurement. Always perform the auto-zero
adjustment after switching ranges. (Do not perform
auto-zero adjustment when using the auto range.)
・ A lit - annunciator indicates that potential is
higher at the black test lead than at the red test lead.
・ The DC V mode permits only DC voltage
measurements that does not include the AC
component (see 1.5: Modes).
・ Every range will display up to 125% of the range,
however, only the range from 10% to 100% can be
displayed with guaranteed accuracy.
・ At any range, gross errors may occur at 1% or below
of the range, whose accuracy is not guaranteed, as a
result of internal corrective calculations.
2. Slide the slide cover up using the slide knob. Next,
insert the red test lead to V and the black test lead
to COM of the voltage measurement terminal.
3. Switch between the auto range and the manual range
as necessary.
4. Carefully contact the test leads to a circuit.
・ Just after suspension of input, or when modes are
NOTE
switched under no input, the counter would not
become zero for about 10 seconds. This is normal
and simply reflects the workings of the internal
circuit. But you can measure with guaranteed
accuracy before the counter becomes zero.
・ Depending on ambient temperatures, the counter
would not become zero under no input, if this
happens, press
to perform a zero-cancel correction. (2.5.2: Zerocancel correction function)
・ During a f.s. input, the measurement response speed
is about 250 ms during rise (0% to 90%) and about
500 ms (100% to 10%) during fall (2.2.5, Figs. 1
and 2).
・ The AC V mode does not allow measurement of DC
waveforms, full-wave rectification waveforms, halfwave rectification waveforms, or DC+AC
waveforms (See 1.5: Modes).
・ Every range will display up to 125% of the range,
however, only the range from 10% to 100% can be
displayed with guaranteed accuracy.
・ At any range, gross errors may occur at 1% or
below of the range, whose accuracy is not
guaranteed, as a result of internal corrective
calculations.
2. Slide the slide cover up using the slide knob. Next,
insert the red test lead to V and the black test lead
to COM of the voltage measurement terminal.
3. Switch between the auto range and the manual range
as necessary.
4. If the counter does not show zero even after display
stabilizes, switch to the range you will use (a manual
range), press
perform a zero-cancel correction. (2.5.2: Zero-cancel
correction function)
5. Carefully contact the test leads to a circuit.
・ When you switch ranges after zero-cancel
NOTE
correction, deviation of the correction value will
prevent accurate measurement. (Do not perform a
zero-cancel correction when using the auto range.)
If you accidentally perform zero-cancel correction while
using the auto range, repeat zero-cancel correction after
turning the power off once and then on again.
・ Just after suspension of input, or when modes are
switched under no input, the counter would not
become zero for about 10 seconds. This is normal
and simply reflects the workings of the internal
circuit. But you can measure with guaranteed
accuracy before the counter becomes zero.
・ The polarity of the input is not displayed, even if DC
voltage is measured in this mode. If the connections
of test leads are moved, the measured values may
change, but the values are within the guaranteed
accuracy. (In case that you would like to measure
DC voltage which doesn't have AC components, you
should make the measurement in DC V mode.)
・ During a f.s. input, the measurement response speed
is about 250 ms during rise (0% to 90%) and about
500 ms (100% to 10%) during fall (2.2.5, Figs. 1
and 2).
・ Every range will display up to 125% of the range,
NOTE
however, only the range from 10% to 100% can be
29
displayed with guaranteed accuracy.
・ At any range, gross errors may occur at 1% or below
of the range, whose accuracy is not guaranteed, as a
result of internal corrective calculations.
2.3.4 Peak Hold Measurement
1. Press
measured circuit.
2. Slide the slide cover up using the slide knob. Next,
insert the red test lead to V and the black test lead
to COM of the voltage measurement terminal.
3. Set to PEAK. The measurement mode is switched
by
and select a measurement mode for the
SLOW/PEAK/Hz
as follow.
SLOWPEAKFAST
(The unit
symbol blinks.)
Hz
(Except
DC)
NORMAL
4. Switch between the auto and the manual range as
necessary. (If you are unable to estimate the peak
voltage value, start at the 600 V range.)
5. Before the measurement, press
0ADJ/RESET
to reset
the residual data.
・ The polarity of the input is not displayed during
NOTE
peak measurements. If the connections of test leads
are moved, the measured values may change, but the
values are within the guaranteed accuracy.
・ For peak measurements, internal resetting occurs
every 250 ms. This may cause a peak detection
failure, depending on the timing.
・ To check transitional peak value, press
MAX/MIN
to shift to the Instantaneous value (no annunciator).
・ The 10 Hz range or 100 Hz range will display up to
NOTE
125% of each range, however, only the range from
31
10% to 100% can be displayed with guaranteed
accuracy.
・
MAX/MIN
does not affect output values.
・ The frequencies, whose waveforms are special such
as inverters, would not be measurable, when the
carrier frequencies are lower than several kHz.
・ Full-wave rectification indicates twice the actual
value, due to an AC coupling in the internal circuit.
・ It would take time to stabilize the counter,
depending on the frequency range or the input
frequency.
2.4.2 FrequencyMeasurement in Voltage Mode
1. Press
the circuit to be measured.
2. If the voltage range of the measured circuit is
known, set the voltage range to the manual range.
2. Slide the slide cover up using the slide knob. Next,
insert the red test lead to V and the black test lead
to COM of the voltage measurement terminal.
4.
SLOW/PEAK/Hz
follows. Select Hz by pressing the key. (The unit
symbol V blinks, and a voltage value is displayed
on the bar graph.)
and select AC or AC+DC, depending on
switches the annunciators as
SLOWPEAKFAST
(The unit
symbol blinks.)
HzNORMAL
4. Switch between the auto range and the manual range
as necessary.
・ At the 100 Hz and 1000 Hz ranges, ---- appears on
NOTE
the counter when the frequency is lower than 10 Hz.
・ ---- appears on the counter, if the frequency is lower
than 1 Hz.
・ O. L. appears on the counter, if the frequency is
higher than 1 kHz.
・ If an input value is significantly lower than the
range, on accurate measurement may not be
achieved, resulting in ----, O. L. or display
fluctuations.
・ The 10 Hz range or 100 Hz range will display up to
125% of each range, however, only the range from
10% to 100% can be displayed with guaranteed
accuracy.
・ The frequencies, whose waveforms are special such
as inverters, would not be measurable, when the
carrier frequencies are lower than several kHz.
・ Full-wave rectification indicates twice the actual
value, due to an AC coupling in the internal circuit.
・ It would take time to stabilize the counter,
depending on the frequency range or the input
frequency.
2.4.3 Output Function For Frequenc
Frequency measurement output is available only in
current modes.
An output of DC 1 V is produced for 1000-count on
the full scale of the frequency range. An output is
produced twice per second, the same as display
refreshing. (The output waveform will be in step
form for a large frequency change, due to D/A
outputs.)
1. To make settings, refer to the frequency
measurement procedure in a current mode.
・ When the AC adapter is used and there is a large
amount of noise in the power line, the display may
show several counts or noise may be present in the
output. In such a case, connect the ground terminal
of the level recorder or the L side of the input to
ground.
2.5 Auto-Zero-Adjustment/
Zero-Cancel Correction Function
2.5.1 Auto-Zero-Adjustment Function
NOTE
The auto-zero-adjustment function is used to adjust
offsets in the internal circuit automatically that
result from temperature characteristics or clamp
sensor magnetization. The clamp core is
magnetized during a large DC current measurement,
or when a powerful magnet is placed close to the
clamp core.
1. Wait until the counter is stable under no input.
Then, press
・ You can perform the auto-zero-adjustment, if the
counter displays within ±45 A in a current mode.
・ When there is an input or the counter decreases, the
measurement accuracy will be spoiled by pressing
0ADJ/RESET as well as the accurate auto-zeroadjustment. If inaccurate auto-zero-adjustment is
performed, perform the correct procedure again.
・ It would take approximately 20 seconds to stabilize
the counter in AC+DC A mode.
・ Use the zero-cancel correction function if the
counter fails to revert to zero after correct auto-zeroadjustment in AC+DC A mode.
display and the digital display value and bar graph
display are maintained.
The data hold function is available for all
measurements.
To cancel the data hold function, press
If you press
RANGE
during the data hold function,
HOLD
again.
the bar graph display the present range.
2.7 Alteration of Counter Updates
The counter is updated twice per second when
powering on. The counter update may be altered
according to measurement conditions.
SLOW/PEAK/Hz
follows:
SLOWPEAK
FAST
(The unit
annunciator blinks.)
changes an annunciator as
Hz
(Except DC)
NORMAL
2.7.1 SLOW mode
If the counter fluctuates rapidly and is hard to read,
you can select a slower update rate (once every 3
seconds) by pressing
SLOW/PEAK/Hz
.
2.7.2 FAST mode
・ For current measurements and voltage measurements, the
counter is updated four times per second in FAST mode.
You can measure abrupt changes such as starting currents.
・ The unit symbol A or V blinks.
・ To facilitate reading when measuring a starting
current, use the record (REC) function to hold the
maximum value (MAX).
Use the recording function to hold the maximum
and minimum measured values and
maximum/minimum averages.
1. Measurement indicated value
Pressing the
current or voltage activates the recording function.
REC flashes and the product saves the maximum
value (MAX), minimum value (MIN), and average
value (AVE) in internal memory from the instant
you press the
MAX/MIN
switches the display as shown below. If MAX,
MIN, or AVE is not displayed, an instantaneous
value is assumed.
Data (MAX, MIN, AVE) remains displayed while
the display is switched. If maximum or minimum
data is updated in the meantime, however, the data
values will change.
With the recording function activated, the auto
power-off function remains disabled. (APS off.)
The average value (AVE) displayed is calculated by:
Average Value = [(Maximum value + Minimum
Value)/2].
If the recording function is activated and
Instantaneous value (no annunciator) selected after
you activate PEAK mode with the
key, you can see the fluctuation of the peak.
When you press the
recording function, the bar graph segments flash and
the elapsed time appears.
When "min" is shown in the right-hand corner of
the bar graph, each segment of the bar graph
corresponds to one minute. Every time one minute
elapses, one segment of the flashing bar graph goes
on. When all segments on the bar graph go on, the
elapsed time is 30 minutes.
When the elapsed time exceeds 30 minutes, one
segment of the flashing bar graph goes off every
time one minute elapses.
When the segments left of a flashing segment
remain on: the number of "on" segments represents
the elapsed time (0 ~ 29).
The illustration below shows when 20 minutes have
elapsed:
When the segments right of a flashing segment
remain on: the number of "off" segments (+30)
represents the elapsed time (30 ~ 59).
The illustration below shows when 50 minutes have
elapsed:
MAX/MIN
key to activate the
When digital display switches the average value
(AVE) to a instantaneous value when you press the
MAX/MIN
indicates hours. In this mode, each segment of the
bar graph corresponds to one hour. The way to read
the bar graph here is similar to reading it in
minutes. When all bar graph segments remain on,
the elapsed time is 29 hours.
The illustration below shows when one hour, 40
minutes have elapsed.
3. Deactivation of Recording Function
key deactivates the recording
Pressing the
function.
HOLD
HOLD
goes on,
stops flashing and
REC
goes on, and the elapsed time stops incrementing.
While the recording function is being deactivated,
data is not updated, even if the clamp sensor is
disconnected from the conductor.
Pressing the
HOLD
key again cancels
HOLD
display and activates the recording function again,
with
flashing again.
REC
4. Resetting of Recording Function
Push
0ADJ/RESET
key, in the case that data is
reset during the recording function action.
5. Cancellation of Recording Function
To cancel the recording function, press the related
function key (A or V) for the measurement in progress.
Once the recording function is canceled, the auto
power-off function becomes effective. (
APS
goes on.)
・ For a long term measurement, use the optional
NOTE
9445-02 or 9445-03 AC ADAPTER, or check how
much the battery power remains by pressing
OUTPUT
.
・ When starting the recording function (REC)inan
auto range, the range is set as the range of when
that pushed
data, make sure to activate the recording function
during measurement. If the function is activated
when there is no input, the minimum value will
remain zero. Also, when deactivating the recording
key to terminate
function, press the
HOLD
measurement once the minimum value and average
value data have been read. If you disconnect the
clamp or test lead from the circuit under
measurement without deactivating the recording
function beforehand, the minimum value will be
zero.
・ When the unit is turned off, accumulated data are
lost.
2.9 Auto Power-Off FunctionAPS
annunciator is displayed, the auto
・ When the
power-off function is active.
・ If no key is pressed for about 10 minutes, the unit
turns itself off automatically.
・ Immediately before turning off automatically,
annunciator blinks and a beep tone is heard for
about 30 seconds.
・ By pressing any key except
extend the powered state for another 10 minutes.
Procedure for disabling the auto power-off
function.
・ When this indication appears, the battery is depleted,
which can lead to inaccurate measurements. Replace
the battery to ensure accuracy.
・ To check remaining battery life, check the bar graph,
when powering on or by pressing
bar graph provides a rough approximation of
remaining battery life. Be careful for the battery life
especially before using an output function for a long
term or when using the REC function.
・ Batteries tend to increase in voltage somewhat when
left unused for a long period. Even if the battery
warning annunciator becomes temporarily off after
the period, replace the battery as soon as possible.
If the battery is not replaced at this point in time,
the annunciator may not light up immediately on the
next occasion. Replace a new battery before it ruins
a measurement or causes some other inconvenience.
(See Chapter 4: Battery replacement)
OUTPUT
.The
2.11 Beep Tone
To disable the beep tone, hold
turning the unit on by pressing
Power sourceOne 6F22 (006P) 9 V battery or 9445-02
Maximum power
consumption
Battery lifeApprox. 25 hours (continuous, no load)
External dimensionsApprox. 62W×260H×39D mm
MassApprox. 540 g
Accessories9207-10 TEST LEAD (red and black) 1
――――――――――――――――――――――――
EN61010-1:2001
Voltage input: Pollution level 2,
measurement category III (expected
transient overvoltage: 6000 V)
EN61010-031:2002
EN61010-2-032:2002
Type A current sensor
EN60529:1991 IP40 (protected against
access to hazardous parts with a wire)
EMC:
EN61326:1997+A1:1998+A2:2001
+A3:2003
EN61000-3-2:2000
EN61000-3-3:1995+A1:2001
φ55 mm max.
0to40℃ (32 to 104 ),
80%RH or less (no condensation)
In 0 to 40℃ range: 0.1×accuracy
specifications/℃
-10 to 50℃
(14 to 122
AC ADAPTER (SA10-0910N, SINOAMERICAN) or 9445-03 AC ADAPTER
(EU) (SA10-0910G, SINO-AMERICAN)
110 mVA
Approx. 2.44"W×10.24"H×1.54"D
Approx. 19.0 oz.
9345 CARRYING CASE 1
Hand Strap 1
6F22(006P) 1
Instruction manual 1
Fully insert the optional 9445-02 AC ADAPTER
into the AC adapter connection terminal.
・ The adapter may be used either with or without a
NOTE
――――――――――――――――――――――――
battery.
・ Use of a battery enables continuous measurement if
the AC power source temporarily becomes
unavailable, due to a blackout or some other reason.
・ When the power supply changes the battery to the
AC adapter such as back from a blackout, the
following would be shown,
1.
2. REC output waveform during frequency
3. Peak hold measurement:
output waveform during current
MON
measurement:
After the output shifts for approximately 10 ms,
the accurate output will be obtained.
It is recommended that
A mode should be used, when you would like to
make a long term measurement on DC A and a
blackout is likely to happen.
measurement:
After the output shifts for approximately 10 ms,
the accurate output will be obtained.
All the phenomena are likely to happen, in the case
the battery voltage is lower than AC adapter voltage
(typ. 9 V) . It is recommended that a new battery
should be used, when the blackout seems to be
short.
・ The battery will be preferred as the power source
when the battery voltage is higher than the voltage
that supplied by the AC adapter (typically the case
with a brand-new battery). When battery voltage
dips below the AC adapter, the source automatically
switches over to the AC adapter.
・ The remaining battery charge (indicated by the bar
graph), when used with the AC adapter, is based on
the AC adapter voltage. It is not a display for
battery life in this case.
・ The minimum stocking period for replacement parts
is five years after end of production.
・ For information regarding service, please contact
your dealer or the nearest HIOKI representative.
・ If the unit is not functioning properly, check the
battery. If a problem is found, contact your dealer
or HIOKI representative. Pack the unit carefully so
that it will not be damaged during transport, and
write a detailed description of the problem. HIOKI
cannot bear any responsibility for damage that
occurs during shipment.