Introduction 1
Notes on Safety 1
Inspection 3
Precautions 3
Chapter 1 3270 Current Monitor 5
1.1 Product Overview 5
1.2 Identification of Indicators 6
1.3 Product Specifications 6
1.4 Measurement Procedure 9
1.4.1 Preparation 9
1.4.2 Measurement 10
1.4.2.1 When Using the 9273 11
1.4.2.2 When Using the 9274 13
1.4.2.3 When Using the 9275 16
1.4.2.4 When Using the 9276 18
1.4.3 Explanation of Each Function 20
1.4.3.1 Connector 20
1.4.3.2 Output Terminals 20
1.4.3.3 Measurement ranges 21
1.4.3.4 Demagnetizing (DEMAG) 23
1.4.3.5 Zero adjustment (ZERO ADJ.) 23
1.4.3.6 Coupling (COUPLING) 24
1.4.3.7 Filter (FILTER) 24
1.4.3.8 Overload indicator (OVERLOAD) 25
1.5 Fuse Replacement 25
Chapter 2 9273 Clamp-On AC Sensor 27
2.1 Product Specifications 27
Chapter 3 9274 Clamp-On AC/DC Sensor 31
3.1 Product Specifications 31
Page 4
Chapter 4 9275 Clamp-On AC Sensor 35
4.1 Product Specifications 35
Chapter 5 9276 Clamp-On AC/DC Sensor 39
5.1 Product Specifications 39
Page 5
G
I
ntroduction
N
Thank you for purchasing this Hioki 3270 Current
Monitor. In order to use this product effectively and to
ensure that it enjoys a long operational life, read this
Instruction Manual carefully and then retain it for future
reference.
otes on Safety
1
WARNIN
This instrument is designed to prevent accidental
shock to the operator when properly used. However
no engineering design can render safe an instrument
which is used carelessly. Therefore, this manual
must be read carefully and completely before making
any measurement. Failure to follow directions can
result in a serious of fatal accident.
Page 6
2
R
G
N
Safety Symbols
This Instruction Manual provides information and
warnings essential for operating this equipment in a safe
manner and for maintaining it in safe operating
condition. Before using this equipment, be sure to
carefully read the following safety notes.
This symbol is affixed to locations on the equipment
where the operator should consult corresponding
topics in this manual (which are also marked with the
symbol) before using relevant functions of the
equipment.
In the manual, this mark indicates explanations
which it is particularly important that the user read
before using the equipment.
This symbol denotes a ground terminal.
This symbol denotes a fuse.
DANGE
WARNIN
CAUTIO
NOTE
The following symbols are used in this Instruction
Manual to indicate the relative importance of cautions
and warnings.
Indicates that incorrect operation presents extreme
danger of accident resulting in death or serious
injury to the user.
Indicates that incorrect operation presents
significant danger of accident resulting in death or
serious injury to the user.
Indicates that incorrect operation presents
possibility of injury to the user or damage to the
equipment.
Denotes items of advice related to performance of
the equipment or to its correct operation.
Page 7
N
I
nspection
P
When the unit is delivered, check and make sure that it
has not been damaged in transit. In particular, check the
accessories, panel switches, and connectors.
If the unit is damaged, or fails to operate according to the
specifications, contact your dealer or Hioki representative.
recautions
In order to ensure safe operation and to obtain maximum
performance from the unit, observe the cautions listed
below.
3
CAUTIO
• Do not subject the unit to vibrations or shocks during
transport or handling. Be especially careful to avoid
dropping the unit.
• Do not store the unit where it will be exposed to direct
sunlight, high temperature, high humidity, or
condensation. If exposed to such conditions, the unit
may be damaged, the insulation may deteriorate, and
the unit may no longer satisfy its specifications.
• Before using the unit, inspect it and check the operation
to make sure that the unit was not damaged due to
poor storage or transport conditions. If damage is
found, contact your dealer or Hioki representative.
Page 8
4
G
N
,
N
Current Monitor (3270)
WARNIN
CAUTIO
CAUTIO
This unit is designed to be grounded through the
three-wire power cord provided. In order to prevent
electric shock, etc., connect this power cord to a
grounded three-prong power receptacle.
Before using a three-prong adapter, be sure to
connect the green wire from the adapter to ground.
• When unplugging the power cord from the power
receptacle or from the unit, grasp the plug, not the cord
in order to avoid damaging the cable.
• Several volts are generated at the output terminal when
the power is turned on or turned off.
Clamp-On Sensor
• If there is any type of dust or dirt on the core contact
surfaces, measurements may be affected. Wipe it
away gently with a soft cloth.
• Do not apply any sort of mechanical impact to the core
section. Scratches on the core surfaces will make
accurate measurements impossible.
• Do not bend or pull the sensor cables (especially where
the cable connects to the sensor) in order to avoid
damaging the sensor cables.
• When the power for the 3270 is on, keep the core
section of the sensor closed, except when clamping
them onto the conductor to be measured. (The facing
surface of the core section can be scratched while it is
open.)
• Taking the sensor apart may damage the open-close
mechanism that could result in an inoperative unit.
Page 9
1
3270 Current Monito
r
1
.1Product Overview
This monitor is a seven-range amp designed especially for
the 9273, 9274, 9275, and 9276 clamp-on sensors. When
used in conjunction with any of the four sensors and
connected to a recorder, oscilloscope, etc., this unit can be
used to easily record and monitor current waveforms.
5
Chapter
Chapter 1 3270 Current Monitor
Page 10
6
g
r
r
)
r
h
h
l
r
h
1
1
.2Identification of Indicators
Filter
switc
ouplin
witch
Connecto
RangesRange switc
Output terminals
(Johnson terminals
Fr on t Rea
Power supply connecto
Overload
indicators
Demagnetizing
switch
Fuse holde
Power ON
Power switc
Power OFF
Zero
adjustment
dial
Ground terminal
Output termina
(BNC terminal)
.3Product Specifications
(Accuracy is guaranteed at 23°C± 3°° °C after the power has
been on for 30 minutes)
♦ Compatible sensors 9273, 9274, 9275, 9276
♦ Output voltage 1 V/range (front: Johnson
♦ Measurement ranges 0.1/0.2/0.5/1/2/5/10 A
(when using the 9273 or 9274)
1/2/5/10/20/50/100 A
Chapter 1 3270 Current Monitor
(when using the 9275 or 9276)
terminals; rear: BNC terminal)
Page 11
♦
Zero adjustment When the 9274 or 9276 is
connected
♦ Demagnetizing When the 9274 or 9276 is
connected
♦ Filter functions Switchable on/off (cutoff
frequency is 100 kHz± 10 %)
♦ Coupling function Switchable AC/DC
♦ Overload indicators Turn on for each range when
input is exceeded by factor of
5.5
♦ Amplitude accuracy ± 0.5 %rdg.± 0.05 %f.s. up to
100 % input for range (DC, 45
to 66 Hz)
± 1.2 %rdg. up to 200 % input
for range (DC, 45 to 66 Hz)
♦ Frequency band AC: 0.2 Hz to 10 MHz (-3 dB)
DC: DC to 10 MHz (-3 dB)
(see Figure 1)
♦ Noise 10 mVrms or less (when using
filter: 2 mVrms or less) (with a
voltmeter for up to 20 MHz)
♦ Temperature ± 0.1 %/° C or less
characteristics
♦ Crest factor 5.5 or less versus the range
value
7
♦ Output resistance 50Ω
♦ Operating temperature 0to40°C , 80 % RH or less
and humidity (no condensation)
♦ Storage temperature -10 to 50° C, 80 % RH or less
and humidity (no condensation)
♦ Dielectric strength 1500 V AC for one minute
(between power supply and case,
and between power supply and
output terminal)
♦ Insulation resistance 500 V DC, 100 MΩ or more
(between power supply and case,
and between power supply and
output terminal)
Chapter 1 3270 Current Monitor
Page 12
8
Figure 1 Frequency Band (example of characteristics)
♦ Power supply 100 V AC (50 /60 Hz) (120, 220,
and 240 V require specification)
♦ Power consumption 15 VA max.
♦ External dimensions Approx. 80(W)X 125(H)X
260(D) mm
♦ Mass Approx. 1,750 g
♦ Accessories Power cord, instruction manual,
spare fuse (F1.0 A/250 V,
20 mmX 5 mm dia., in the fuse
holder), 9177 Output cord
At range f.s. input
Coupling : DC
Output (dB, 0dB=1V)
Coupling : AC
Frequency (Hz)
Chapter 1 3270 Current Monitor
Page 13
N
1
.4Measurement Procedure
1
Before taking measurements, prepare this unit, the
sensor and the other measuring equipment such as a
recorder and an oscilloscope.
.4.1Preparation
9
CAUTIO
Before turning on the power, make sure that the voltage
of the power supply being used matches the supply
voltage indicated on the rear panel of the unit.
(1) Turn the power switch off and connect the power cord.
(2) Connect the sensor to be used to the connector on the
unit.
(3) Turn the power switch on and confirm that the power
indicator lights.
(4) Connect the output terminal on the unit and the input
terminal of the measuring equipment, such as a recorder.
(When using an oscilloscope, make the connections so that
the GND terminal for the probe is connected to the L
output terminal.)
Chapter 1 3270 Current Monitor
Page 14
10
R
N
1
.4.2 Measurement
DANGE
CAUTIO
To avoid short circuits and accidents that could
result in injury or death, use clamp testers only with
power lines carrying 600 V AC or less.
To avoid short circuits and accidents that could
result in injury or death, when the tip of jaws is
open, do not use on bare conductors. The core
and shield case are not insulated.
• Do not exceed the maximum allowable current input.
The maximum input range differs, depending on the
frequency of measured current. (Refer to specifications
of the sensor.)
• If current exceeds the maximum allowable input,
overheating of the sensor will trip the protective function
for the internal circuitry, so the unit will no longer
produce accurate output. If this happens, eliminate the
input immediately (either by removing the sensor from
the conductor being measured or by reducing the input
current to zero.) (Wait until the sensor has had
sufficient time to cool before resuming operation.)
• If the above situation occurs repeatedly, or if current in
excess of the maximum input range is input
continuously, the sensor could be damaged.
• Be sure to press only one range switch; avoid having
none of the LEDs that indicate the range lit, or having
two or more lit at the same time.
Chapter 1 3270 Current Monitor
Page 15
11
C
"OPEN" indication
Facing surface
Lever
Core section
Sensor connector
Sensor cable
Figure 2 9273 Identification of Indicators
1
NOTE
• Depending on the measured current value and the
frequency, however, some sound may be produced by
resonation from the sensor, it has no effect on
measurements.
• If a circuit carrying a large current is near to the sensor,
measurements may be affected by external magnetic
fields.
.4.2.1 When Using the 9273
urrent direction mark "UNLOCK" indication
NOTE
(1) Select both ranges of the 3270 and the measuring
equipment according to the input current value. (Refer to
Table 1 in Section 1.4.3.3.)
(2) Set the coupling switch to AC.
(3) Pull the lever of the sensor towards the "OPEN"
indication to open the core section. (Refer to Figure 2,
9273 Identification of Indicators.)
(4) Position the sensor so that the current direction mark on
the sensor part points in the direction of the load and so
that the conductor being measured is centered, and then
close the core section with the lever.
If the conductor being measured is not centered, the
results may be affected by its position.
Chapter 1 3270 Current Monitor
Page 16
12
N
,
D
s
Lock state
NOTE
(5) Press the lever forward until the "UNLOCK" indication
disappears, and confirm that the lever is completely
locked.
Accurate measurement is not possible if "UNLOCK" is
visible.
(6) The current waveform can now be recorded or monitored
with the measuring equipment.
CAUTIO
NOTE
When opening the core section of the sensor, be sure to
operate with the lever. If an upper core is forced to open
when the sensor is locked, the open-close mechanism
can be damaged.
o not pres
• To eliminate noise and other high-frequency components,
set the filter switch to "ON". (The cutoff frequency is
approximately 100 kHz.)
• If it is necessary to go to 0.2 Hz or less, outside of the
frequency band, then set the coupling to DC. (The offset
voltage value for this unit is weighted as the DC
component. The offset voltage values changes, depending
on the range.)
• Immediately after the power is turned on and open or
close sensor, depending on the condition of the internal
circuitry, the output waveform will fluctuate up and down
for approximately several ten seconds until it stabilizes.
(However, a small amount of fluctuation will remain.)
• Accurate measurements are not possible when measuring
overlapping DC components of several amperes or more.
• The zero adjustment dial does not function.
Chapter 1 3270 Current Monitor
Page 17
Facing surface
Lever
Core section
Sensor connector
Sensor cable
C
"OPEN" indication
Figure 3 9274 Identification of Indicators
1
• If the demagnetizing switch is pressed during
measurement, the output changes for approximately 1
second, but afterwards the measured value is not affected.
(The demagnetizing function is not used when an AC
sensor is connected.)
.4.2.2 When Using the 9274
urrent direction mark "UNLOCK" indication
13
(1) Select both ranges of the 3270 and the measuring
equipment according to the input current value. (Refer to
Table 1 in Section 1.4.3.3.)
(2) Set the coupling switch to DC.
(3) Adjust the GND level of the measuring equipment
(oscilloscope, etc.).
(4) Set the measuring equipment for DC coupling.
(5) Before clamping the sensor to the conductor to be
measured, press the demagnetizing switch.
(6) Turn the zero adjustment dial and adjust the output
voltage to zero (GND level).
(7) Pull the lever of the sensor towards the "OPEN"
indication to open the core section. (Refer to Figure 3,
9274 Identification of Indicators.)
Chapter 1 3270 Current Monitor
Page 18
14
N
,
D
s
Lock state
(8) Position the sensor so that the current direction mark on
the sensor part points in the direction of the load and so
that the conductor being measured is centered, and then
close the core section with the lever.
NOTE
If the conductor being measured is not centered, the
results may be affected by its position.
(9) Press the lever forward until the "UNLOCK" indication
disappears, and confirm that the lever is completely
locked.
NOTE
Accurate measurement is not possible if "UNLOCK" is
visible.
(10) The current waveform can now be recorded or monitored
with the measuring equipment.
CAUTIO
NOTE
When opening the core section of the sensor, be sure to
operate with the lever. If an upper core is forced to open
when the sensor is locked, the open-close mechanism
can be damaged.
o not pres
• To eliminate noise and other high-frequency components,
set the filter switch to "ON". (The cutoff frequency is
approximately 100 kHz.)
• If the coupling of this unit is set to AC, or the coupling of
the measuring equipment is set to AC, then components
below the respective cutoff frequencies will be attenuated.
• Immediately after powering on, there may in some cases
be a large offset drift caused by self-generated heat from
this unit and the sensor.
Chapter 1 3270 Current Monitor
Page 19
• When performing continuous measurements, it is
necessary to be aware that the offset voltage for this unit
and the sensor drifts, depending on factors such as the
ambient temperature.
• Always perform demagnetizing and the zero adjustment
when taking measurements.
• The offset voltage value for this unit changes according to
the range. If you change the range, be sure to perform
the zero adjustment again.
• Perform the zero adjustment when there is no input and
after performing demagnetizing. (The coupling setting
must also be set to DC coupling.)
• Do not press the demagnetizing switch during
measurement. (If the switch is pressed, the output will
change for approximately 1 second. If you press the
switch inadvertently, perform demagnetizing and zero
adjustment again.)
• Before doing low current measurement, demagnetize the
sensor heads and zero-adjust as follows. After opening
and closing the sensor heads, the zero point may drift by
an amount equivalent to several mA.
1. Press the demagnetizing switch of the 3270.
2. Open and close the sensor head clamps two or three
times.
3. Turn the zero adjustment dial of the 3270 to zero the
output voltage.
15
Chapter 1 3270 Current Monitor
Page 20
16
e
n
r
e
k
Figure 4 9275 Identification of Indicators
1
.4.2.3 When Using the 9275
Current direction mar
Lever
Facing surfac
Core sectio
Sensor connecto
(1) Select both ranges of the 3270 and the measuring
equipment according to the input current value. (Refer to
Table 2 in Section 1.4.3.3.)
(2) Set the coupling switch to AC.
(3) Gently press the lever of the sensor to open the core
section. (Refer to Figure 4, 9275 Identification of
Indicators.)
(4) Position the sensor so that the current direction mark
points in the direction of the load and so that the
conductor being measured is centered, and then close the
core section.
NOTE
If the conductor being measured is not centered, the
results may be affected by its position.
Sensor cabl
Chapter 1 3270 Current Monitor
Page 21
NOTE
NOTE
17
(5) Confirm that the lever is completely locked.
Accurate measurement is not possible if the sensor is not
locked.
(6) The current waveform can now be recorded or monitored
with the measuring equipment.
• To eliminate noise and other high-frequency components,
set the filter switch to "ON". (The cutoff frequency is
approximately 100 kHz.)
• If it is necessary to go to 0.2 Hz or less, outside of the
frequency band, then set the coupling to DC. (The offset
voltage value for this unit is weighted as the DC
component. The offset voltage values changes, depending
on the range.)
• Immediately after the power is turned on and open or
close sensor, depending on the condition of the internal
circuitry, the output waveform will fluctuate up and down
for approximately several ten seconds until it stabilizes.
(However, a small amount of fluctuation will remain.)
• Accurate measurements are not possible when measuring
overlapping DC components of ten and several amperes or more.
• The zero adjustment dial does not function.
• If the demagnetizing switch is pressed during
measurement, the output changes for approximately 1
second, but afterwards the measured value is not affected.
(The demagnetizing function is not used when an AC
sensor is connected.)
Chapter 1 3270 Current Monitor
Page 22
18
e
n
r
e
k
Figure 5 9276 Identification of Indicators
1
.4.2.4 When Using the 9276
Current direction mar
Lever
Facing surfac
Core sectio
Sensor connecto
(1) Select both ranges of the 3270 and the measuring
equipment according to the input current value. (Refer to
Table 2 in Section 1.4.3.3.)
(2) Set the coupling switch to DC.
(3) Adjust the GND level of the measuring equipment
(oscilloscope, etc.).
(4) Set the measuring equipment for DC coupling.
(5) Before clamping the sensor to the conductor to be
measured, press the demagnetizing switch.
(6) Turn the zero adjustment dial and adjust the output
voltage to zero (GND level).
(7) Gently press the lever of the sensor to open the core
section. (Refer to Figure 5, 9276 Identification of
Indicators.)
Sensor cabl
Chapter 1 3270 Current Monitor
Page 23
(8) Position the sensor so that the current direction mark
points in the direction of the load and so that the
conductor being measured is centered, and then close the
core section.
NOTE
If the conductor being measured is not centered, the
results may be affected by its position.
(9) Confirm that the lever is completely locked.
NOTE
Accurate measurement is not possible if the sensor is not
locked.
(10) The current waveform can now be recorded or monitored
with the measuring equipment.
NOTE
• To eliminate noise and other high-frequency components,
set the filter switch to "ON". (The cutoff frequency is
approximately 100 kHz.)
• If the coupling of this unit is set to AC, or the coupling of
the measuring equipment is set to AC, then components
below the respective cutoff frequencies will be attenuated.
• Immediately after powering on, there may in some cases
be a large offset drift caused by self-generated heat from
this unit and the sensor.
• When performing continuous measurements, it is
necessary to be aware that the offset voltage for this unit
and the sensor drifts, depending on factors such as the
ambient temperature.
• Always perform demagnetizing and the zero adjustment
when taking measurements.
• The offset voltage value for this unit changes according to
the range. If you change the range, be sure to perform
the zero adjustment again.
• Perform the zero adjustment when there is no input and
after performing demagnetizing. (The coupling setting
must also be set to DC coupling.)
19
Chapter 1 3270 Current Monitor
Page 24
20
N
,
N
1
1
1
• Do not press the demagnetizing switch during
measurement. (If the switch is pressed, the output will
change for approximately 1 second. If you press the
switch inadvertently, perform demagnetizing and zero
adjustment again.)
.4.3 Explanation of Each Function
.4.3.1 Connector
Connect any one of clamp-on sensors (9273 through 9276)
to the connector.
CAUTIO
• In order to prevent damages of this unit and the sensor
never plug in or unplug the sensor connector when the
power is turned on.
• Because± 12 V is output from the connector on the
equipment to provide power to the sensor, do not plug
anything into this connector except for the sensor
connector.
.4.3.2 Output Terminals
Output terminals are located on both the front and rear
of the unit; signals are output from these terminals
simultaneously.
Front: Johnson terminals (connect a shielded cord)
Rear: BNC terminal
The output resistance is approximately 50Ω..
CAUTIO
In order to prevent malfunctions, do not short the output
terminal or input voltage to them.
Chapter 1 3270 Current Monitor
Page 25
21
N
1
NOTE
• Connect measuring equipment with an input resistance of
at least 1 MΩ.. When using the front and rear panel
output terminals at the same time, make sure that the
combined input resistance of both measuring equipments
is 1 MΩ or more. (Accuracy is not guaranteed if the
input resistance is less than 1 MΩ.)
• Note that the frequency band is not guaranteed if the
load capacitance (the input capacitance of the measuring
equipment and the capacitance of the output cord) is
several hundred pF or more.
• Do not connect the GND terminal (L) of a recorder or an
oscilloscope to the H terminal on this unit. (Accurate
measurement may not be possible.)
.4.3.3 Measurement ranges
The measurement ranges differ according to the sensor
that is connected. (When a sensor is connected, the LED
indicating the range lights.)
When using the 9273 or 9274: 0.1/0.2/0.5/1/2/5/10 A
When using the 9275 or 9276: 1/2/5/10/20/50/100 A
In each measurement scale, output of 1V is generated
when input is at the top of the scale.
Determine the range to be set based on this unit’s
measurement ranges and the range of the measuring
equipment (such as a recorder) being used. (Refer to
Tables 1 and 2.)
CAUTIO
Be sure to press only one range switch; avoid having
none of the LEDs that indicate the range lit, or having two
or more lit at the same time.
Chapter 1 3270 Current Monitor
Page 26
22
Table 1. Current value conversion table when the 9273 or 9274 (5 mm
Table 2. Current value conversion table when the 9275 or 9276 (20 mm
dia. type) is connected
Measuring
equipment
range/DIV
10 A range0.1 A0.2 A0.5 A1A2A5A10 A20 A
5 A range0.05 A 0.1 A0.25 A 0.5 A1A2.5 A5A10 A
2 A range0.02 A 0.04 A 0.1 A0.2 A0.4 A1A2A4A
1 A range0.01 A 0.02 A 0.05 A 0.1 A0.2 A0.5 A1A2A
0.5 A range 5mA 0.01 A 0.025 A 0.05 A 0.1 A0.25 A 0.5 A1A
0.2 A range 2mA 4mA 0.01 A 0.02 A 0.04 A 0.1 A0.2 A0.4 A
0.1 A range 1mA 2mA 5mA 0.01 A 0.02 A 0.05 A 0.1 A0.2 A
10 mV20 mV50 mV 100 mV 200 mV 500 mV1V2V
*The values indicate the current value per DIV of the
measuring equipment, such as a recorder.
dia. type) is connected
Measuring
equipment
range/DIV
100 A range1A2A5A10 A20 A50 A100 A200 A
50 A range0.5 A1A2.5 A5A10 A25 A50 A100 A
20 A range0.2 A0.4 A1A2A4A10 A20 A40 A
10 A range0.1 A0.2 A0.5 A1A2A5A10 A20 A
5 A range0.05 A 0.1 A0.25 A0.5 A1A2.5 A5A10 A
2 A range0.02 A 0.04 A 0.1 A0.2 A0.4 A1A2A4A
1 A range0.01 A 0.02 A 0.05 A0.1 A0.2 A0.5 A1A2A
10 mV20 mV50 mV 100 mV 200 mV 500 mV1V2V
*The values indicate the current value per DIV of the
measuring equipment, such as a recorder.
Chapter 1 3270 Current Monitor
Page 27
1
.4.3.4 Demagnetizing (DEMAG)
1
Because demagnetizing is necessary after range settings
when an AC/DC sensor (9274 or 9276) is connected, be
sure to press the demagnetizing switch. (Press the switch
when there is no input.)
Demagnetization is performed because there is a
possibility the sensor will become magnetized when the
power for the main unit is turned on/off or as the result of
excessive input. (Accurate measurements are not possible
if the sensor is magnetized.)
Demagnetization occurs in less than 1 second. In
addition, because the signal lines are disconnected by
internal circuitry during the demagnetization process, the
output voltage changes.
If you inadvertently press the demagnetizing switch while
taking a measurement, perform the demagnetizing
process again while there is no input.
23
NOTE
• Always demagnetize before performing the zero
adjustment.
• Because demagnetizing is only effective when an AC/DC
sensor is connected, the demagnetizing function does not
work when there is no sensor connected or when an AC
sensor (9273 or 9275) is connected.
.4.3.5 Zero adjustment (ZERO ADJ.)
When an AC/DC sensor (9274 or 9276) is connected,
demagnetizing and zero adjustment are necessary after
setting the range. (Perform zero adjustment while there is
no input.)
Zero adjustment is performed in order to compensate for
the offset voltage values of the unit and the sensors and
for the drift of those values caused by the temperature.
Chapter 1 3270 Current Monitor
Page 28
24
1
1
Even when using an AC/DC sensor and setting the
coupling to AC for measurement, first set the coupling to
DC and perform the demagnetizing and zero adjustment
procedures.
Turning the zero adjustment dial to the left decreases the
output voltage, and turning the dial to the right increases
the output voltage. (The lower the measurement range,
the broader the zero adjustment range becomes. )
NOTE
Because zero adjustment is only effective when an AC/DC
sensor is connected, the zero adjustment function does not
work when there is no sensor connected or when an AC
sensor (9273 or 9275) is connected.
.4.3.6 Coupling (COUPLING)
The coupling method can be either AC or DC.
AC: Eliminates the DC component and outputs only the
AC signal. (The cutoff frequency is 0.2 Hz.)
DC: Outputs all signals.
Pressing the coupling switch sets the coupling to AC.
(The LED lights.)
.4.3.7 Filter (FILTER)
Use the filter to eliminate noise and other high-frequency
components. (The cutoff frequency is approximately 100
kHz.)
Pressing the filter switch turns the filter on. (The LED
lights.)
Chapter 1 3270 Current Monitor
Page 29
N
G
1
.4.3.8 Overload indicator (OVERLOAD)
1
This LED lights when the input (peak value) exceeds the
range by a factor of 5.5.
When the LED is lit, the input exceeds the crest factor
(maximum of 5.5 for each range value), resulting in
waveform distortion. Adjust the range so that the LED
goes off.
25
CAUTIO
NOTE
In order to avoid malfunctions, confirm the maximum
measurable current range of the sensor being connected
before beginning to take measurements. (Refer to
specifications of the sensor.)
The LED may light immediately after an AC sensor (9273
or 9275) is connected or depending on the position of the
zero adjustment dial when an AC/DC sensor (9274 or
9276) is connected.
.5Fuse Replacement
If the fuse blows, replace it as shown in Figure 6.
WARNIN
In order to prevent electric shock, always
disconnect the power cord from the receptacle
before replacing the fuse.
Always replace the fuse with a new fuse of the
specified rating.
Never use a fuse of other than the specified rating,
and never short circuit the fuse holder to prevent an
accident resulting in injury or death.
100 V, 120 V : F1.0 A/250 V 20 mmX 5 mm dia.
220 V, 240 V : F0.5 A/250 V 20 mmX 5 mm dia.
Chapter 1 3270 Current Monitor
Page 30
26
Spare fuse
Fuse holder
Power cord
Figure 6 Fuse Replacement
1. Unplug the power cord.
2. Remove the fuse holder, check the fuse and then
replace.
Chapter 1 3270 Current Monitor
Page 31
2
9273 Clamp-On AC Senso
r
2
.1Product Specifications
(Accuracy is guaranteed at 23°C± 3°C after the power has
been on for 30 minutes)
♦ Rated current 20AAC
♦ Output voltage 2 V AC/20 A AC
♦ Output resistance 50Ω
♦ Input impedance 0.1 mΩ or less at 55 Hz
(see Figure 7)
♦ Continuous maximum 20 A (see Figure 8)
input range
27
Chapter
♦ Maximum peak current 50 A noncontinuous (peak value)
value
♦ Amplitude accuracy ± 0.5 % rdg.± 0.05 % f.s.
(45 to 66 Hz)
♦ Phase accuracy ± 0.2° (45 to 66 Hz)
♦ Frequency band 0.7 Hz to 10 MHz (-3dB)
(see Figure 9, example of
characteristics)
♦ Frequency characteristics 2 Hz to 10 kHz:± 2.0 % or less
(Deviation from accuracy) 10 kHz to 100 kHz:± 3.0 % or less
♦ Temperature coefficient ± 0.1 % f.s./° C or less
for sensitivity (within a range of 0 to 40° C)
♦ Noise 1 mV rms or less (with a
voltmeter for up to 20 MHz)
♦ Power consumption 1.5 VA max. (with rated input)
Chapter 2 9273 Clamp-On AC Sensor
Page 32
28
♦ Supply voltage ± 12 V± 1V
♦ Operating temperature 0to40°C , 80 % RH or less
and humidity (no condensation)
♦ Storage temperature -10 to 50°C , 80 % RH or less
and humidity (no condensation)
♦ Effect of external Equivalent to a maximum of
magnetic fields 20 mA (in a 400 A/m AC current
magnetic field)
♦ Effect of conductor Within± 0.2 %
position
♦ Dielectric strength 2200 V AC for 1 minute
(between electric circuit and case)
♦ Insulation resistance 500 V DC, 100 MΩ or more
(between electric circuit and case)
♦ Maximum permitted 600 V peak; (the core, shield
circuit voltage case, and electrical circuits are
(insulated conductor) not insulated.)