Hioki 3193, 3193-10 Instruction Manual

Page 1
3193 3193-10
POWER HiTESTER
Instruction Manual
Feb. 2019 Revised edition 16 3193A981-16 19-02H
EN
Page 2
Page 3

Contents

Introduction i Inspection i Safety Notes iii Notes on Use vi Chapter 1 Overview 1
1.1 Product Overview 1
1.2 Features
Chapter 2 Names and Functions of Parts 5
2.1 Panels and Key Operation 5
2.2 Names and Configuration of Screen
2.2.1 Screen Configuration 8
2.2.2 MEAS Screen (Measurement Screen) 9
2.2.3 STATUS Screen (Setting Screen) 13
2.2.4 FDD Screen 16
2.3 Indicators 17
2.4 Peak Over Indication
18
Chapter 3 Preparation for Measurement 19
3.1 Notes on Use 19
3.2 Basic Operating Procedure
3.3 Powering On
22 23
2
8
3.4 Connecting the Direct Input Unit
3.5 Connecting the Clamp Input Unit
3.6 Measurement Losses
3.7 Error Messages
3.8 System Reset
3.9 Operations During Power Failure
24 26 27 28 29 30
Chapter 4 Setting and Using the Basic Functions 33
4.1 Setting the Wiring Mode (1P2W to 3P4W) 33
4.2 Setting the Coupling Mode (DC/AC+DC/AC)
4.3 Switching the Voltage Range and Current Range
4.4 Effective Value (RMS) or Mean Rectified Value (MEAN) Selection
4.5 Setting the Scaling (PT/CT/SC Ratios)
4.6 Setting the Low-pass Filter (LPF)
3193A981-16
36 37
39 40 42
Page 4
4.7 Setting the Phase Polarity Discrimination Filter 43
4.8 Switching the Waveform Peak Value
4.9 Setting the Response (FAST/MID/SLOW)
4.10 Setting the Averaging
4.11 Setting on the MEAS Screen
4.11.1 Setting the Display Items (for 1 to 6 channels) 50
4.11.2 Setting the SELECT screen 51
44 45 46 50
4.12 Setting on the SYSTEM screen 52
4.12.1 Switching the Interface (GP-IB/RS-232C) 52
4.12.2 Setting the Display Color 53
4.12.3 Setting the Back Light 53
4.12.4 Setting the Equation for Reactive Power (Q) and  Apparent Power (S) 54
4.12.5 Setting the Beep Sound 55
4.12.6 Setting Indications for Out-of-Range Inputs 55
4.12.7 Setting the Display Language (English/Japanese) 56
4.12.8 Setting the Real-time Clock 56
4.13 Degaussing 57
Chapter 5 Frequency Measurement 59
5.1 Setting the Frequency Measurement Source (fa) 60
5.2 Setting the Frequency Range(fa)
61
Chapter 6 Hold/Peak Hold Function 63
6.1 Hold Function 63
6.2 Peak Hold Function
6.2.1 Combination with Control Times 66
65
Chapter 7 Integration Function 67
7.1 Overview 67
7.2 Setting the Control Time
7.2.1 Setting the Interval Time 69
7.2.2 Setting the Timer 70
7.2.3 Setting the Real Time Control 71
7.3 Integration Screen 72
69
7.4 Starting, Stopping, and Resetting the Integration
7.5 Manual Integration
(Controlled by Panel Keys)
74
7.6 Integration Using Time Settings
(Controlled by Panel Keys)
7.6.1 Timer Integration 75
75
73
Page 5
7.6.2 Real-Time Control Integration 76
7.6.3 Interval Integration 76
7.7 Measuring the Load Factor 77
7.8 Zero suppress function
78
Chapter 8 Efficiency Measurement 79
8.1 Overview 79
8.2 Efficiency Screen
8.3 Setting the Calculation Formula
8.4 Example Measurement
8.4.1 Efficiency Measurement of a Switching Power Supply (1φ2W) 82
8.4.2 Efficiency Measurement of a Switching Power Supply (3φ3W) 82
8.4.3 Efficiency Measurement of a Light Fitting (Two-Lamp) 83
8.4.4 Efficiency Measurement of an Inverter (1φ2W) 83
8.4.5 Efficiency Measurement of an Inverter (3φ3W)
and Motor 84
80 81 82
Chapter 9 External Output/ External Control Terminals 85
9.1 Connector Pin Arrangement 86
9.2 Internal Circuit for Analog, Monitor, D/A Outputs
9.3 Internal Circuit for the External Control and Timing
9.3.1 INTEG.EXT.CONT and INTEG.RESET Terminals 88
9.3.2 FDD/PRINTER.START Terminal 89
9.3.3 EXT.A/D START Terminal 89
87 88
Chapter 10 D/A Output 91
10.1 Overview 91
10.2 Selecting Output Item
10.3 Output Rate
92 93
Chapter 11 Using the Floppy Disk Drive (Only 3193) 95
11.1 Overview 95
11.2 Operation Procedure
11.3 Using the Floppy Disk
96 97
11.4 Formatting a Floppy Disk
11.5 Switching the FDD/Printer
11.6 Setting File Names for Saved Measurement Data
98 99
100
Page 6
11.7 Setting the Measurement Items for Saving 101
11.8 Saving the Data on FDD
11.8.1 Automatic Saving Using Time Settings 102
11.8.2 Manual Saving 103
11.8.3 Screen Hard Copy 103
11.8.4 Saving the Settings 103
11.8.5 Saving and Loading Settings 104
102
11.9 Information Which Can Be Saved 105
11.10 Deleting and Confirming Files
11.11 Format for Data Output to Floppy Disk
11.12 Message and Error Displays
106 107 108
Chapter 12 GP-IB and RS-232C Interface 109
12.1 Overview 109
12.2 Specifications
12.2.1 GP-IB Interface 110
12.2.2 RS-232C Interface 111
110
12.3 Interface Outline 114
12.3.1 Messages 114
12.3.2 Command Syntax 115
12.3.3 Headers 115
12.3.4 Message Terminators 116
12.3.5 Separators 116
12.3.6 Data Formats 117
12.3.7 Abbreviation of Compound Commands 118
12.3.8 Output Queue 119
12.3.9 Input Buffer 119
12.3.10 Note on Commands Initiating Events 119
12.3.11 Status Model 120
12.3.12 Status Byte Register 121
12.3.13 Event Registers 122
12.3.14 GP-IB Commands 129
12.4 Command Reference 130
12.4.1 Standard Command 131
12.4.2 Specific Commands 141
12.5 Command Summary 197
12.5.1 Standard Commands 197
12.5.2 Commands Specific to the 3193 198
12.5.3 Valid Command According to Condition (Standard Command) 204
Page 7
12.5.4 Valid Command According to Condition (Specific Command) 205
12.5.5 Execution Time of GP-IB Interface Command 210
12.5.6 Initialization 211
12.5.7 Specific Command Tree 212
12.6 Sample Programs 215
12.6.1 GP-IB 215
12.6.2 RS-232C 216
12.7 Device Compliance Statement 217
12.8 Notes on Interface 219
12.8.1 GPーIB Troubleshooting 219
12.8.2 RS-232C Troubleshooting 220
Chapter 13 Using the Printer (Option) 221
13.1 Overview 221
13.2 Specifications
13.3 Operating Procedure
13.4 Loading Recording Paper
13.5 Switching the FDD/Printer
13.6 Setting the Measurement Items to Print
13.7 Printing Out
13.7.1 Manual Printing 227
13.7.2 Automatic Printing by Time Settings 227
13.7.3 Screen Hard Copy 229
13.7.4 Help Printing Mode 229
13.7.5 External Control Printing 229
222 223 224 225 226 227
13.8 Setting the Printing Direction 230
13.9 Error and Overflow Displays
231
Chapter 14 9600 AC/DC DIRECT INPUT UNIT (Option) 233
14.1 Overview 233
14.2 Notes on Use
14.3 Specifications (using with the 3193)
14.4 Internal Block Diagram
14.4.1 RMS Value (root-mean-square value) 238
14.4.2 MEAN Value (MEAN rectification effective value for display) 239
14.4.3 Active Power 239
14.4.4 Waveform Peak Value Measurement Circuit 240
14.4.5 Crest Factor 240
234 235 238
Page 8
Chapter 15 9601 AC DIRECT INPUT UNIT (Option) 241
15.1 Overview 241
15.2 Notes on Use
15.3 Specifications (using with the 3193)
15.4 Internal Block Diagram
242 243 246
Chapter 16 9602 AC/DC CLAMP INPUT UNIT (Option) 247
16.1 Overview 247
16.2 Notes on Use
16.3 Specifications (using with the 3193)
16.4 Internal Block Diagram
248 249 252
Chapter 17 9603 EXTERNAL SIGNAL INPUT UNIT (Option) 253
17.1 Overview 253
17.2 Display Screen
17.3 Setting Method
17.3.1 Changing the Voltage 255
17.3.2 Setting the Scaling 256
254 255
17.3.3 Setting the Units 256
17.3.4 Setting the Pulse 257
17.3.5 Calculating Motor Power (Pm) 257
17.4 Specifications 258
17.5 Internal Block Diagram
259
Chapter 18 Maintenance and Service 261
18.1 Cautions 261
18.2 Disposing of the Unit
262
Chapter 19 Rack Mounting 263
19.1 Rack Mounting Fittings 263
19.2 Installation Procedures
265
Chapter 20 Specifications (unit only) 267
20.1 General Specifications 267
20.2 Function Specifications
270
20.3 Calculations
20.4 Internal Block Diagram of the 3193
276 281
Appendix APPENDIX 1 Index INDEX 1
Page 9
────────────────────────────────────────────────────

Introduction

Thank you for purchasing this HIOKI "3193, 3193-10 POWER HiTESTER." To get the maximum performance from the unit, please read this manual first, and keep this at hand.
The HIOKI 3193, 3193-10 POWER HiTESTER will be referred to as the
HIOKI 3193in this manual.
The HIOKI 3193 comes with a floppy disk drive (FDD) but the HIOKI
3193-10 does not. The respective specifications will be referred to when a specific product name is mentioned in this manual.
When the FDD is mentioned in the instruction manual, even if there is no
special indication, the corresponding specification will not be supported by the 3193-10.
i

Inspection

When the unit is delivered, check and make sure that it has not been damaged in transit. In particular, check the accessories, panel switches, and connectors. If the unit is damaged, or fails to operate according to the specifications, contact your dealer or HIOKI representative.
■ Standard accessories
Instruction Manual 1 Power cord 1 Connector 1
■Options
9600 AC/DC DIRECT INPUT UNIT 9601 AC DIRECT INPUT UNIT 9602 AC/DC CLAMP INPUT UNIT 9603 EXTERNAL SIGNAL INPUT UNIT 9604 PRINTER UNIT 9605 HARMONIC/FLICKER MEASURMENTS UNIT
────────────────────────────────────────────────────
Introduction
Page 10
ii
────────────────────────────────────────────────────
■ Shipment
If reshipping the unit, preferably use the original packing. Before shipping the 3193, always remove the floppy disk.
ACDC UNIT
AC UNIT
CLAMP UNIT ACDC 20 A
CLAMP UNIT ACDC 200 A
CLAMP UNIT ACDC 500 A
CLAMP UNIT ACDC 50 A
CLAMP UNIT 20 A AC CLAMP
CLAMP UNIT 200 A AC CLAMP
Ex UNITON
Printer ON
9605 ON
9600 is installed. 9601 is installed. 9602 is installed and 9277 is inserted. 9602 is installed and 9278 or CT6863 is inserted. 9602 is installed and 9279 or 9709 or CT6865 is
inserted. 9602 is installed and CT6862 is inserted. 9602 is installed and 9270 or 9272 (20 A) or 9272-10
(20 A) is inserted. 9602 is installed and 9271 or 9272 (200 A) or 9272-10
(200 A) is inserted. 9603 is installed . 9604 is installed. 9605 is installed.
────────────────────────────────────────────────────
Inspection
Page 11
iii
────────────────────────────────────────────────────

Safety Notes

This Instruction Manual provides information and warnings essential for operating this equipment in a safe manner and for maintaining it in safe operating condition. Before using this equipment, be sure to carefully read the following safety notes.
DANGE R
Safety symbols
This instrument is designed to comply with IEC 61010 Safety Standards, and has been thoroughly tested for safety prior to shipment. However, mishandling during use could result in injury or death, as well as damage to the instrument. Be certain that you understand the instructions and precautions in the manual before use. We disclaim any responsibility for accidents or injuries not resulting directly from instrument defects.
・ This symbol is affixed to locations on the equipment where the
operator should consult corresponding topics in this manual (which are also marked with the symbol) before using relevant functions of the equipment.
・ In the manual, this mark indicates explanations which it is
particularly important that the user read before using the equipment.
Indicates AC (Alternating Current).
Indicates both DC (Direct Current) and AC (Alternating Current).
DANGE R
WARNING
CAUTION
NOTE
Indicates a grounding terminal.
Indicates the ON side of the power switch.
Indicates the OFF side of the power switch.
The following symbols are used in this Instruction Manual to indicate the relative importance of cautions and warnings. 
Indicates that incorrect operation presents extreme danger of accident resulting in death or serious injury to the user.
Indicates that incorrect operation presents significant danger of accident resulting in death or serious injury to the user.
Indicates that incorrect operation presents possibility of injury to the user or damage to the equipment.
Denotes items of advice related to performance of the equipment or to its correct operation.
────────────────────────────────────────────────────
Safety Notes
Page 12
iv
────────────────────────────────────────────────────
Accuracy
The specifications in this manual include figures for "measurement accuracy" when referring to digital measuring instruments, and for "measurement tolerance" when referring to analog instruments.
f.s. (maximum display or scale value, or length of scale)
Signifies the maximum display (scale) value or the length of the scale (in cases where the scale consists of unequal increments or where the maximum value cannot be defined).
In general, this is the range value (the value written on the range selector or equivalent) currently in use.
rdg. (displayed or indicated value)
This signifies the value actually being measured, i.e., the value that is currently indicated or di splayed by the measuring instrument.
dgt. (resolution)
Signifies the smallest display unit on a digital measuring instrument, i.e., the value displayed when the last digit on the digital display is "1".
Display items Display items FDD (header) Printer GP-IB/RS-232C
Voltage Voltage peak
Current Current peak
Active power Reactive power Apparent power Power factor Phase angle Frequency Integration
active current
Integration power
|Up|/Pk (enlarged
|Ip|/Pk (enlarged
(+) +Ih PIh Ih (+) PIH (-) -Ih MIh Ih (-) MIH
(total)
(+) +WP PWP WP (+) PWP (-) -WP MWP WP (-) MWP
(total) WP WP WP WP
U U U U
display)
I I I I
display)
P P P P
Q Q Q Q
S S S S
λ PF PF PF
φ DEG DEG DEG
f f f f
Ih Ih Ih IH
PEAK PEAK (Vpeak) Pk
PEAK PEAK (Apeak) Pk
Load factor Maximum averaging
power Efficiency Channel A of 9603 Channel B of 9603 Motor power of 9603
────────────────────────────────────────────────────
Safety Notes
LF LF LF LF
no display Wmax Wmax none
η EFFI EFFI EFF
chA CHA CHA EXTA
chB CHB CHB EXTB
Pm PM PM PM
Page 13
v
────────────────────────────────────────────────────
Measurement categories
9600, 9601 and 9602 instrument comply with CAT III (600 V or less)/ CAT II (600 to 1000 V) safety requirements.
To ensure safe operation of measurement instruments, IEC 61010 establishes safety standards for various electrical environments, categorized as CAT to CAT IV, and called measurement categories.
CATⅡ
CATⅢ
CATⅣ
Primary electrical circuits in equipment connected to an AC electrical outlet by a power cord (portable tools, household appliances, etc.) CAT II covers directly measuring electrical outlet receptacles.
Primary electrical circuits of heavy equipment (fixed installations) connected directly to the distribution panel, and feeders from the distribution panel to outlets.
The circuit from the service drop to the service entrance, and to the power meter and primary overcurrent protection device (distri bution panel).
Using a measurement instrument in an environment designated with a higher-numbered category than that for which the instrument is rated could result in a severe accident, and must be carefully avoided. Use of a measurement instrument that is not CAT-rated in CAT II to CAT IV measurement applications could result in a severe accident, and must be carefully avoided.
────────────────────────────────────────────────────
Safety Notes
Page 14
vi
────────────────────────────────────────────────────

Notes on Use

In order to ensure safe operation and to obtain maximum performance from the unit, observe the cautions listed below.
DANGE R
WARNING
Always connect the powermeter input (including clamp) to the
secondary side of the breaker. On the secondary side of a breaker, even if the lines are shorted the breaker can trip and prevent an accident. On the primary side, however, the current capacity may be large, and in the event of a short-circuit there may be a serious accident.
The maximum input voltage and current for this unit depend on the
input unit being used. Do not apply an input exceeding the maximum input voltage and current specified for the input unit. Exceeding the maximum input voltage or current could damage the unit or cause a serious accident.
Before turning on the power, make sure that the voltage of the power
supply being used matches the supply voltage indicated on the rear panel of the unit. If an attempt is made to use an improper supply voltage, there is danger of damage to this unit and of life-threatening risk to the operator.
The power switch has a microgap construction, and it is therefore
essential to use it close to a power outlet. When the unit is not in use, and while making connections to the circuit being tested, isolate the unit electrically from the power supply, for example by removing the power cord plug from the outlet.
To avoid electrical accidents and to maintain the safety
specifications of this instrument, connect the power cord provided only to a 3-contact (two-conductor + ground) outlet.
Do not remove the case of the unit. There are components inside
carrying high voltages or becoming hot, and this could cause an electric shock accident.
Do not use the unit where it may be exposed to corrosive or
explosive gases. The unit may be damaged, or explosion may occur.
────────────────────────────────────────────────────
Notes on Use
Page 15
vii
────────────────────────────────────────────────────
CAUTION
・ Should the unit emit smoke, or a strange smell or strange sound,
immediately stop testing operations, power the unit off, and remove the power cord from the outlet, shut off the circuit being tested, disconnect the unit, and consult your HIOKI representative. Continued use of the unit could lead to fire or electric shock accidents.
・ Do not insert foreign objects through the ventilation holes in the top
and bottom of the case. Particularly if metallic, liquid, or combustible substances get inside the case, this may lead to fire or electric shock, or to malfunction. When using several devices at the same time, do not stack them.
・ Never allow the ventilation holes in the top and bottom covers to
become blocked while using this unit. Blocking the ventilation will cause internal temperature to rise, possibly resulting in fire or damage to the equipment.
・ To prevent electric shock, do not allow the unit to become wet and
do not use the unit when your hands are wet.
・ This unit is designed for indoor use and can be safely used at
temperatures ranging from 0℃ to 40℃ and should be operated at 80% RH or less.
・ This unit is not constructed to be waterproof or dustproof, so do
not use it in a very dusty environment or in one where it will get wet.
・ Do not store or use the unit where it will be exposed to direct
sunlight, high temperatures, high humidity, or condensation. If exposed to such conditions, the unit may be damaged, the insulation may deteriorate, and the unit may no longer satisfy its specifications.
・ To avoid damage to the unit, do not subject the equipment to
vibrations or shocks during transport or handling. Be especially careful to avoid dropping the equipment.
・ Do not place the unit on an unstable stand, or in an uneven
location. It may fall to the ground, or fall over, and either of these events may lead to malfunction or accident.
・ Do not use the unit near any device which generates strong
electromagnetic radiation or near a static electrical charge, as these may cause errors.
・ Avoid treading on or pinching the cable so as not to damage the
cable sheaths.
・ When unplugging the power cord from the power receptacle or from
the unit, grasp the plug, not the cord, in order to avoid damaging the cable.
・ To avoid damaging the sensor cables or probes, do not bend or pull
them, especially where they connect to the sensor
.
────────────────────────────────────────────────────
Notes on Use
Page 16
viii
────────────────────────────────────────────────────
CAUTION
・ Use caution when taking measurements in circuits where the
power line are hot.
・ Keep the cables well away from heat, to prevent the possibility of
melting the insulation. ・ For long-term storage, remove the power cord. ・ In order to prevent temperature rising, place the device away
from other things so that the vents are not blocked. ・ The instrument should be operated only with the bottom or rear
side downwards. ・ Do not stack up the device. ・ Do not obstruct the ventilation holes. (Do not place anything on
the device as it may block the ventilation holes.)
────────────────────────────────────────────────────
Notes on Use
Page 17
ix
────────────────────────────────────────────────────
NOTE
All options for this unit are factory-fitted, but it is also possible to add
options at a later date after purchase. In this case, however, it is necessary for the unit to be returned to HIOKI headquarters.
With the appropriate combination of direct connection input units, this unit
can function as either an AC power meter or dual AC/DC power meter. When used together with clamp input units, depending on whether the clamp sensor used is for AC or DC, this unit can function as either an AC power meter or dual AC/DC power meter. When used as an AC power meter, it is not possible to measure a DC component superimposed on the AC signal (half-wave rectification, or full-wave rectification upper and lower excluded waveform).
Note that limits are specified for the range in which voltage and current level
accuracies are guaranteed.
In order to assure accurate measurements, allow this unit to warm up for at
least 1 hour before using it.
This power meter uses the calculations indicated in the specifications in
order to determine apparent power (S), power factor (
λ
), and reactive power (Q) on the basis of the measured voltage (U), current (I), and active power (P). The values displayed by this power meter may differ from those produced by other testers that are based on different principles of operation or testers that use different calculations.
Display of a polarity symbol (-) together with reactive power (Q), power
factor (λ) or phase angle (φ) occurs only when TYPE1 is selected as the calculation type, and indicates that current is delayed with respect to voltage. For reasons related to circuit design, the polarity symbol is displayed even when input is "0".
Due to measuring error or a disproportionate load, the effective power may
exceed the apparent power, resulting in a power factor of 1 or more. In such a case, this system is designed to make the apparent power equal to the effective power.
There are two sorts of measurement: using analog calculation by the input
unit or by digital calculation using the harmonic analysis/flicker measurement function, and since these have entirely different principles of measurement, frequency range, and accuracy, and as a result the final measured values may be different.
Accurate measurement may be impossible in locations subject to strong
external magnetic fields, such as transformers and high-current conductors, or in locations subject to strong external electric fields, such as radio transmission equipment.
For the current measurement of the 9600 AC/DC DIRECT INPUT UNIT, the
DC-CT (current transformer) method is used, so after measuring a large current, there may be a very slight residual offset signal. The offset signal produces the largest error effect in the minimum ranges; in this case, shut off the current input, and carry out degaussing (DMAG).
The 9600, 9601, and 9602 active power measurement units operate with an
auto-zero circuit at 2.442 kHz. For this reason, an input signal with a frequency of 2.442 kHz will result in a periodically fluctuating display indication.
────────────────────────────────────────────────────
Notes on Use
Page 18
x
────────────────────────────────────────────────────
NOTE
When the input is less than a certain level of measurement range, depending
on using the input unit, the display value is forced to zero. See the specifications of the input unit to be used.
When measuring a high frequency voltage to earth (for example the
secondary side of an inverter), errors may occur in the measurement values.
To maintain the measurement accuracy of the unit, bear the following
cooling measures in mind: Do not obstruct the ventilation holes Keep away from sources of heat If rack mounted, install a cooling fan Do not stack up the device. Do not place anything on the device.
This unit switches the power supply voltage automatically. Voltage
fluctuations of 10% from the rated supply voltage are taken into account.
This unit has no external fuse. Thus if the unit does not operate when the
power switch is turned on and power is supplied, there is a fault. Disconnect the power cord and measurement lines, and contact your dealer or HIOKI representative.
This instruments may cause interference if used in residential areas.Such use
must be avoided unless the user takes special measures to reduce electromagnetic emissions to prevent interference to the reception of radio and television broadcasts.
────────────────────────────────────────────────────
Notes on Use
Page 19
1
────────────────────────────────────────────────────
Chapter 1
1
2
1.1 Product Overview
The 3193 POWER HiTESTER is a power meter that can test any type of line ranging from single-phase lines to three-phase four-wire lines.
Based on the voltage, current, and active power measurements, this unit calculates and displays reactive power, apparent power, power factor, phase angle, and efficiency. It further has a wide range of measurement functions including frequency measurement function, peak measurement function, current integration, active power integration, analog outputs, monitor outputs, and harmonic analysis/flicker function (option).
Overview
3
4
5
6
7
8
9
10
11
12
13
14
A
────────────────────────────────────────────────────

1.1 Product Overview

Page 20
2
────────────────────────────────────────────────────
1.2 Features
(1) Safe design
The 3193 POWER HiTESTER features a safe design that complies with the IEC61010-1 safety standard.
(2) Capable of measuring power on all types of power lines
This single power meter is capable of measuring power on all types of power lines, ranging from single-phase lines to three-phase four-wire lines by installing the input units.
(3) Simultaneous measurement of multiple systems
Up to six input channels can be installed, so that for example, a single unit can simultaneously measure the input power and output power of a three­phase inverter. By combinations with the optional external input units, the input power and output power of an inverter and the output of a motor can be measured and calculated with a single unit. Further, by using the efficiency calculation function, the efficiency and overall efficiency of up to three points can be measured simultaneously with one unit.
(4) Wide current measurement range
With a direct connection input unit, and no external current transformer, it is possible to measure a maximum of 50 A rms. The internal current transformer design keeps the losses in the current measurement meter extremely low. Using a current sensor input unit, existing current sensors can be used to measure up to 500 A.
(5) High accuracy
The basic accuracy of±0.1% rdg.±0.1%f.s. is high.
(6) Wide frequency range: DC and 0.5 Hz to 1 MHz (using optional
9600
The wide frequency response supports the evaluation of inverter-motor systems, inverter fluorescent lighting systems, ultrasound motors, switching power supplies, and so on.
AC/DC DIRECT INPUT UNIT
)
(7) Built-in low pass filter
The cut-off frequency can be selected from three values. This function allows an inverter fundamental frequency to be extracted, and also supports data exchange with conventional devices.
(8) Three types of calculation expression selectable
Three types of calculation for apparent power and reactive power can be selected, to support compatibility with conventional devices.
────────────────────────────────────────────────────

1.2 Features

Page 21
3
────────────────────────────────────────────────────
(9) Peak measurement function
It is possible to measure peak values of a voltage or current waveform. Also, using the peak hold function, motor surge current peak values, and the peak values of effective values can be measured.
(10) Separate integration values for each polarity
For current and active power, positive, negative, and total integrated values are provided. Each channel can be integrated separately.
(11) Three averaging functions
Time average, sliding average, or exponential average can be selected as the averaging mode.
(12) Three-channel frequency measurement function
The unit has a three-channel frequency measurement function, allowing separate frequency measurement when multiple systems are being tested.
(13) Analog signal input from an external device (using optional 9603
EXTERNAL SIGNAL INPUT UNIT)
It is possible to input a separate analog output (or pulse signal) from the system undergoing power measurement, for easy on-screen conversion. For example, inputting the analog outputs from a torque meter or rotation counter enables the converted torque or rotation values to be shown on the screen. The power can also be computed from the torque or rotation values.
1
2
3
4
5
6
7
8
(14) Harmonic analysis/flicker measurement function (option)
The harmonic analysis function can analyze up to the 50th harmonic of the voltage, current, or active power waveform, for fundamental frequencies from 5 Hz to 440 Hz.
(15) High visibility color LCD
The color LCD screen has a wide viewing angle, and allows simultaneous display of different information without requiring screen switching, giving an at-a-glance grasp of the overall state of the measured system. In combination with the optional harmonic analysis/flicker function, it is possible to use different colors to distinguish harmonic analysis graphs and waveforms.
9
10
11
12
13
14
────────────────────────────────────────────────────
1.2 Features
A
Page 22
4
────────────────────────────────────────────────────
(16) FDD installed (only 3193)
The built-in floppy disk drive facilitates data saving when required, and automatic saving at preset times. It is also possible to save the unit settings and reload them to restore the previous state. Upgrades of the unit are also supported. (No FDD in the HIOKI 3193-10)
(17) Eight-channel D/A output fitted as standard
These output specified items, with an output of ±5 V corresponding to the full scale range.
(18) Efficiency calculation function fitted as standard
This provides three efficiency calculations from measured power values.
(19) Rapid response analog outputs fitted as standard
These outputs provide 5 V full-scale analogs of the voltage, current, and active power ranges. (Excluding 1000 V range) When the response is set to FAST, these have a 100 ms response time.
(20) Waveform outputs fitted as standard
These outputs provide 1 V full-scale waveform outputs corresponding to the voltage and current ranges, allowing wavefor m monitoring with a recorder or oscilloscope.
(21) Built-in printer (option)
This provides a printout of the measurement data and screen displays.
(22) Choice of display language
The display language can be selected as English or Japanese.
(23) GP-IB/RS-232C fitted as standard
────────────────────────────────────────────────────
1.2 Features
Page 23
5
────────────────────────────────────────────────────
Chapter 2
1
2
Names and Functions of Parts
2.1 Panels and Key Operation
(∗)
FDD
PAGE keys FUNCTION keys
3
4
5
6
START/STOP key
OUTPUT keys
7
8
FUNCTION keys
CURSOR keys
RANGE keys
Front Panel
Backlight off LED
SHIFT key
HOLD key
LOCAL key
9
10
11
Power switch
12
13
14
A
────────────────────────────────────────────────────

2.1 Panels and Key Operation

Page 24
6
────────────────────────────────────────────────────
FUNCTION MEAS
STATUS FDD
(∗)
Changes to the measurement value display screen Changes to the settings display screen Used for setting the file name of the floppy disk, and saving
and recalling unit settings.
PAGE
In the MEAS and STATUS screens, used to switch display for the item in the second row from the top.
RANGE
U
+ / U-
Changes the voltage range on the displayed channel. Pressing both keys sets to the auto ranging.
SHIFTU+ Pressing the
SHIFT
key and then pressing the
U
+
key toggles the voltage for the displayed channel between RMS and MEAN.
I
+ /
I
-
Changes the current range on the displayed channel. Pressing both keys sets to the auto ranging.
SHIFTI+ Pressing the
SHIFT
key and then pressing the
I
+
key toggles the current for the displayed channel between RMS and MEAN.
SHIFTI- Pressing the
SHIFT
key and then pressing the
I
-
key executes degaussing. This effects only when using the 9600 input unit, or when using the 9602 in combination with AC/DC clamp.
(∗)
OUTPUT OUTPUT
COPY
SHIFTCOPY
SAVE/PRINT
SHIFT SAVE/PRINT
CURSOR ▲▼
SHIFT
Outputs the display screen to the FDD Sends a copy of the screen to the FDD Prints the current settings of the unit on the FDD
(∗)
Outputs the specified items to the FDD Feeds the printer paper. During printing, pressing this key
(∗)
ends the printing. Used to move the cursor for settings and so on. Changes the connection mode on the measurement screen for
or printer.
(∗)
or printer.
(∗)
or printer.
(∗)
or printer.
each channel.
SHIFT
Changes the response mode on the measurement screen for each channel.
SHIFT
Changes the low-pass filter on the measurement screen for each channel.
HOLD HOLD Stops display updating of all measurement values, then each
subsequent press updates the display.
SHIFT→HOLD
When not in the hold mode, this switches to the peak hold mode. Press this mode, pressing the
SHIFT
and
HOLD
again to release this setting. In
HOLD
key resets and then it is in the
peak value hold mode.
LOCAL LOCAL
SHIFTLOCAL Locks the panel keys. Press
Used to end remote control.
SHIFT
and
LOCAL
again to release
this setting.
START/ STOP
F1 to F5 POWER
START/STOP
SHIFT START/STOP
Used to select setting items. Powers the unit on and off.
Starts and stops each time controls (integration, time averaging, automatic output to FD/printer).
After stopping the integraion, this key combination resets the elapsed time and integration values.
(: There is no FDD or FDD key in the HIOKI 3193-10. SAVE/PRINT key is the PRINT key.)
────────────────────────────────────────────────────
2.1 Panels and Key Operation
Page 25
7
────────────────────────────────────────────────────
AC power inlet
GP-IB connector 9603 Input terminal Voltage input terminal
Output connector
(analog output, monitor output, D/A output, external control terminal)
1
2
3
4
5
RS-232C connector
Grounding terminal
Serial No.
6
Clamp input terminal Current input terminal
7
Rear Panel
8
9
10
11
12
────────────────────────────────────────────────────
2.1 Panels and Key Operation
13
14
A
Page 26
8
────────────────────────────────────────────────────
2.2 Names and Configuration of Screen
2.2.1 Screen Configuration
In 1 P3W, 3P3W mode: channels 1+2, 3+4,
4+5, 5+6 In 3V3A, 3P4W mode: channels 1+2+3, 4+5+6
(in 1P2W mode)
MEAS screen
P
F
MAGNIFY
3ch
DETAILS
4ch
INTEG-
RATED
F
4ITEMS
8ITEMS
16 ITEMS
5ch
6ch
SELECT
EFFI
EXT IN
HARMONIC
/FLICKER
The three basic screens are the MEAS (measurement) screen, the STATUS screen, and the FDD (floppy disk drive) screen
STATUS
,or
FDD
key
(∗)
on the panel switches to the corresponding screen.
(∗)
. Pressing the
MEAS
This configuration is when all options are installed.
Power on
Option installation display screen
Second time or later
Factory settings
Display returns on the channel selected
when the unit was powered off
FUNCTION
1ch
2ch
MEAS FDD
UNIT
P
TIME
FREQ/
OUTPUT
SYSTEM
EFFI
EXT UNIT
HARMONIC/FLICKER
Screen paths when all options are installed.
STATUS
STATUS screen
F
WIRING
Range, PT ratio, CT ratio,
C
SC ratio, LPF, phF, peak
Response, average, interval time, timer time, real-time control time
Output device, output item, saving color, D/A output, frequency source, frequency range
Interface, display color, back­light, calculation, beep sound, language, real time, system reset, out-of-range input display
Calculation, setting
Voltage range, scaling, unit
FDD screen
FDD
C
Data file name Set up file name Screen copy file name File list Disk format Remaining space
Display items P: PAGE keys C: CURSOR keys F: F1 to F5 keys
:ThereisnoFDDor
FDD key,FDD screen in th e H IO KI 3193-10.
,
(∗)
NOTE
When the unit is first powered on after purchase, and after a system reset, the display for channel 1 appears. Thereafter, the display returns on the channel selected when the unit was powered off.
────────────────────────────────────────────────────

2.2 Names and Configuration of Sc ree n

Page 27
9
────────────────────────────────────────────────────
2.2.2 MEAS Screen (Measurement Screen)
This screen displays measurement results. The displays available depend on the options installed. Switch from one display to another using the this case the second row of cursor positions from the top of the screen shows the currently displayed page. Each item in this row is blank if the corresponding option is not installed. The third row on the screen shows the settings for the currently displayed channel.
①②
#201.tif
Use the PAGE keys
Use the Function keys
PAGE
key on the front panel. In
Measurement Screen
① Screen for each channel (channels 1 to 6) [
1chto6ch
]
This is the screen when the cursor position is on 1ch (channel 1) to 6ch
(channel 6). This corresponds to the installation of the 9600, 9601, and 9602 options.
For multi-channel combinations, of single-phase three-wire (1P3W) and
above, the measurement values are displayed combined on a single screen. In this case, the cursor also appears on the corresponding channel numbers together.
#202.tif
────────────────────────────────────────────────────
#203.tif
In 1P2W mode
In 1P3W, 3P3W mode
2.2 Names and Configuration of Sc ree n
Page 28
10
────────────────────────────────────────────────────
3V3A/3P4W mode
#204.tif
NOTE
The subscript numbers on symbols indicate channels. For example, "U1"
indicates that the voltage measured on input unit channel 1 is displayed. The indication "U
" indicates that the SUM value of the voltages measured
123
on input unit channels 1, 2 and 3 is displayed.
When the SUM value of the active power in 3V3A mode is shown as, for
example, "P
", then "P1+P2" is calculated, and "P3" is ignored.
123
In 1P2W mode, when DC mode is selected, the reactive power (Q), power
factor (λ), phase angle (φ) for each channel are displayed, but they are meaningless. In 3P3W or 3V3A mode, when three-phase three-wire is measured, active power (P), apparent power (S), reactive power (Q), power factor (λ), phase angle (φ) for each channel are displayed, but they are also meaningless. Make a setting for display to off not to display them.
See Section 20.3, "Calculations."
Other display screens include enlarged and integration value displays;
function keys
F1
and
switch to these displays.
F3
#205.tif
Enlarged display screen
Integration display screen
1 P2W, DC mode
#206.tif
────────────────────────────────────────────────────
2.2 Names and Configuration of Sc ree n
Page 29
11
────────────────────────────────────────────────────
#207.tif
② Selection screen [
SELECT
]
Required items (except harmonics, flicker, and integration values) can be
selected from all of the measurements being made and displayed.
The screen format can be selected to show 4, 8, or 16 items.
4 items display screen
8 items display screen
#208.tif
#232-9.tif
16 items display screen
────────────────────────────────────────────────────
2.2 Names and Configuration of Sc ree n
Page 30
12
────────────────────────────────────────────────────
③ Efficiency screen [
By combining measurement values (active power, motor power), this calculates and displays the efficiency.
#232-10.tif
④ External input screen [
This is displayed when the optional 9603 EXTERNAL SIGNAL INPUT UNIT is installed. The motor power (Pm) is displayed only when the unit settings for channel A is torque, for channel B is number of rotating (rpm).
EFFI
]
EXT IN
]
#232-11.tif
⑤ Harmonic waveform screen
This is displayed when the optional 9605 HARMONIC/FLICKER
MEASUREMENTS UNIT
is installed. (See 9605 Instruction Manual)
────────────────────────────────────────────────────
2.2 Names and Configuration of Sc ree n
Page 31
13
────────────────────────────────────────────────────
2.2.3 STATUS Screen (Setting Screen)
This screen provides various settings. The screens correspond to the installed options. Switch from one display to another using the panel. In this case the second row of cursor positions from the top of the screen shows the currently displayed page. From this row, you can also check which options are installed.
①② ③ ④
PAGE
key on the front
STATUS screen
① Unit screen [
UNIT
]
This shows a list of the settings for each channel. In this case too, the
settings are collected together according to the channel combinations depending on the connection mode.
Moving the cursor to an item with the
CURSOR
keys allows that item to be
set or changed.
1P2W 6 types
3P4W 2 types
② Time control screen [
TIME
]
This shows the settings for the response, averaging function, the interval time, timer time, and real-time control time.
#CHANGE
233-4.tif
────────────────────────────────────────────────────
2.2 Names and Configuration of Sc ree n
Page 32
14
────────────────────────────────────────────────────
③ Frequency screen [
This shows the settings for the output to FDD/printer, printing direction, saving screen color on FD, D/A output, frequency measurement function source, and frequency range of the unit. With the HIOKI 3193-10, FDD ([FD], [FD&PRINT]) cannot be selected as data output.
#CHANGE
233-5.tif
④ System screen [
FREQ/OUTPUT
SYSTEM
]
]
This shows the settings for the GP-IB/RS-232C, display color, LCD backlight, calculation, beep sound, display of input out-of-range, language, real-time, system reset.
#CHANGE
233-6.tif
────────────────────────────────────────────────────
2.2 Names and Configuration of Sc ree n
Page 33
15
────────────────────────────────────────────────────
⑤ Efficiency screen [
EFFI
]
This sets the items to be substituted in the efficiency calculation expression.
#233-7.tif
(6) External input screen [
This is displayed when the optional 9603 EXTERNAL SIGNAL INPUT UNIT is installed, and some settings are made for the 9603.
EXT UNIT
]
#233-8.tif
(7) Harmonic waveform screen
This is displayed when the optional 9605 HARMONIC/FLICKER
MEASUREMENTS UNIT
is installed. (See 9605 Instruction Manual)
────────────────────────────────────────────────────
2.2 Names and Configuration of Sc ree n
Page 34
16
────────────────────────────────────────────────────
2.2.4 FDD Screen (Only 3193)
This sup ports file name setting of a floppy disk, and saving and loading of the unit settings.
#234.tif
────────────────────────────────────────────────────
2.2 Names and Configuration of Sc ree n
Page 35
17
────────────────────────────────────────────────────
2.3 Indicators
The following indicators are shown by panel key operation.
#235.tif
SHIFT
Indicates when the
SHIFT
key is pressed. Pressing again goes
off.
KL
Indicates key lock state (red), and remote state by GP-IB/RS­232C (yellow).
HOLD
Indicates the displays are held.
PEAK
Indicates the peak hold function is active.
TOTAL
STIME
Indicates during real time control. A blue display indicates
Indicates total value after time averaging.
standby during real-time control, and a yellow display indicates within setting time.
INTEG
Indicates integration or operation by time controls.
A yellow display indicates that operation is in progress, and a blue display indicates during waiting.
FD
Indicates the output method is set to FDD.
PRI
Indicates the output method is set to printer. A yellow display indicates normal and a red display indicates there is no paper or printer lever is head-up.
────────────────────────────────────────────────────

2.3 Indicators

Page 36
18
────────────────────────────────────────────────────
2.4 Peak Over Indication
If the input voltage or current waveform peak exceeds six times the range value, a "PEAK" indication appears. These indications appear on the screen below the channel number, the voltage indication on the left and the current indication on the right, so that even a "PEAK" state can be detected even for channels not currently displayed. For example, the following indications mean that the current on channel 4 and the voltage on channel 6 have peak values exceeding six times the range value.
#236.tif
Indicates when voltage or current
Current on channel 4 has peak values exceeding six times the range value
Voltage on channel 6 has peak values exceeding six times the range value
for any channels has peak values.
NOTE
These indications are only valid within the range of the maximum input voltage and current for each input unit.
────────────────────────────────────────────────────

2.4 Peak Over Indication

Page 37
19
────────────────────────────────────────────────────
Chapter 3
Preparation for Measurement
1
2
3
4
5
3.1 Notes on Use
DANGE R
Always connect the powermeter input (including current sensor) to
the secondary side of the breaker. On the secondary side of a breaker, even if the lines are shorted the breaker can trip and prevent an accident. On the primary side, however, the current capacity may be large, and in the event of a short-circuit there may be a serious accident.
Once the connections are made, do not touch the input terminals,
and the voltage and current transformers. There are exposed live parts, and a danger of electric shock or serious accident.
Check that the terminals are tightened securely. If the connections
should become detached, there is a danger of a short-circuit or electric shock accident. Additionally, if the connections are not properly tightened, the contact resistance increases, which may lead to the generation of heat, or fire.
The maximum input voltage and current for this unit depend on the
input unit being used. Do not apply an input exceeding the maximum input voltage and current specified for the input unit. Exceeding the maximum input voltage or current could damage the unit or cause a serious accident.
6
7
8
9
10
11
12
────────────────────────────────────────────────────

3.1 Notes on Use

13
14
A
Page 38
20
────────────────────────────────────────────────────
WARNING
CAUTION
In order to prevent electric shock and short-circuit accidents, shut
off the power to the line to be measured before connecting the direct connection voltage and current cables to the terminals.
Be sure to connect the voltage input terminals, current input terminals
correctly. Measurement which is attempted with the wiring connected incorrectly may cause damage to the unit or a short-circuit.
To avoid electrical accidents and to maintain the safety
specifications of this instrument, connect the power cord provided only to a 3-contact (two-conductor + ground) outlet.
When using an external voltage transformer, do not leave the
secondary side short-circuited. If a voltage is applied to the primary while the secondary is short-circuited, a high current will flow through the secondary, which could lead to fire or malfunction.
When using an external current transformer, do not leave the
secondary side open-circuit. If a current flows through the primary while the secondary is open-circuit, this can generate a high voltage on the secondary, which is extremely dangerous.
・ To avoid electrical accidents, use wiring with more than adequate
current carrying capacity and voltage insulation properties.
・ When the power is turned off, do not apply voltage or current to the
voltage input terminal, current input terminal, or clamp sensor. Doing so may damage the unit.
NOTE
For 3P3W, 3V3A measurement, the active power values for each channel are
found from the voltages between lines and the currents on each line, and have no individual significance.
If the maximum values of the voltage or current on the lines being measured
exceed the measurement range of this unit, use an external voltage transformer (PT) or current transformer (CT). In this case, by setting the corresponding PT and CT ratios with the scaling function of this unit, you can directly read off the measured values. See Section 4.5, "Setting the Scaling."
For combinations 1P3W and above, there are restrictions on the
combinations of input units. See Section 4.1, "Setting the Wiring Mode."
When using an external voltage transformer (PT) or current transformer
(CT), its precision, phase accuracy, frequency characteristics, and so on, may greatly affect the error in the measured power value. Use transformers with adequate frequency characteristics and small phase error for the frequency band of the line being measured.
When using a voltage transformer (PT) or current transformer (CT), ground
one side of the secondary for safety.
────────────────────────────────────────────────────
3.1 Notes on Use
Page 39
21
────────────────────────────────────────────────────
Pre-Measurement Inspection
Before using the instrument the first time, verify that it operates normally to ensure that the no damage occurred during storage or shipping. If you find any damage, contact your dealer or Hioki representative.
Peripheral Device Inspection
Inspect the voltage measurement cables or connection cables.
Is the insulation of the voltage measurement cables or connection cables to be used dameged, or is bare metal exposed?
Inspect the current sensors
Is a sensor cracked or damaged?
Is the insulation of the any cable to be used dameged, or is bare metal exposed?
No Metal Exposed
No
No Metal Exposed
Inspection complete
Metal Exposed
Yes
Metal Exposed
Do not use if damage is present, as you could receive an electric shock. Replace the damaged  items.
1
2
3
4
5
6
7
8
9
10
11
12
13
14
────────────────────────────────────────────────────
3.1 Notes on Use
A
Page 40
22
────────────────────────────────────────────────────
3.2 Basic Operating Procedure
Check that the line to be measured is shut off, and
測定対象のラインが遮断されていることを確認、
check that this unit is powered off and the power
および本器の電源が OFF され電源コードを抜いて
cord disconnected from the outlet.
あることを確認する
Connect the power cord to the 3P-outlet.
Turn the power on.
Turn the power on.
Turn the power on.
(Warming-up 1 hour or more
When using the 9600 or 9602, carry out degaussing after warming-up.)
Make a connections.
See Section 3.4, 3.5.
Check there is no short circuit
Set the connection mode.
Set the display item.
Set the measurement range.
Set the each items.
Check the connection of line being measured again,
and turn on the power line.
Start the measurement.
────────────────────────────────────────────────────

3.2 Basic Operating Procedure

Page 41
23
────────────────────────────────────────────────────
3.3 Powering On
WARNING
Before turning on the power, make sure that the voltage of the power
supply being used matches the supply voltage indicated on the rear panel of the unit. If an attempt is made to use an improper supply voltage, there is danger of damage to this unit and of life-threatening risk to the operator.
1
2
3
CAUTION
To avoid electrical accidents and to maintain the safety
specifications of this instrument, connect the power cord provided only to a 3-contact (two-conductor + ground) outlet.
The power switch has a microgap construction, and it is therefore
essential to use it close to a power outlet. When the unit is not in use, and while making connections to the circuit being tested, isolate the unit electrically from the power supply, for example by removing the power cord plug from the outlet.
・ Should the unit emit smoke, or a strange smell or strange sound,
immediately stop testing operations, power the unit off, and remove the power cord from the outlet, shut off the circuit being tested, disconnect the unit, and consult your HIOKI representative. Continued use of the unit could lead to fire or electric shock accidents.
・ When the power is turned off, do not apply voltage or current to the
voltage input terminal, current input terminal, or clamp sensor. Doing so may damage the unit.
1. Confirm that the voltage of the power supply being used matches the supply voltage indicated on the rear panel of the unit.
2. Confirm that the power switch on the front panel is off.
3. Connect the supplied power cord to the AC inlet on the rear pane.
4. Connect the power cord to a grounded three-pin outlet. If no grounded outlet is available, use the supplied ground adapter.
5. Turn on the power switch on the front panel.
6. The unit starts the self-test. It is completed after about 10 seconds.
4
5
6
7
8
9
10
11
12
Self-Test
In the self test, the following tests are carried out, then after about 10 seconds the measurement screen automatically appears.
Unit versionInstalled optionsRAM check
NOTE
────────────────────────────────────────────────────
If there is a problem in the settings, this screen remains displayed and the unit stops. If this happens again after powering off and on, the unit has developed a fault. Stop measurement, and shut off the line being measured, then power off the unit. Disconnect the test wiring and the power cord. Contact your
HIOKI
service representative for repair.

3.3 Powering On

13
14
A
Page 42
24
────────────────────────────────────────────────────
3.4 Connecting the Direct Input Unit
The following diagrams show the connections in various modes when using the 9600
AC/DC DIRECT INPUT UNIT
and 9601
■ Single-phase two wires (1P2W) i: 1 to 6
Source
Load
Source
Load
AC DIRECT INPUT UNIT
Source
.
Load
CT
U
±
I
±
Channel (i)
transformers (PT, CT)
U
±
I
±
Channel (i)
Connecting the measured
line directly to the unit
CT
U
±
I
±
Channel (i)
Using the current
transformer (CT)
Using the voltage and current
■ Single-phase three wires (1P3W) i: 1, 3, 4, 5
Source
N
±
±
Load
U
I
U
±
I
±
Source Load
N
N
CT
U
±
I
±
CT
±
±
Source Load
N
N
CT
U
I
PT
U
±
I
±
CT PT
U
±
±
N
I
Channel (i+1)
Connecting the measured
line directly to the unit
────────────────────────────────────────────────────

3.4 Connecting the Direct Input Unit

Channel (i)
Channel (i+1)
Using the current
transformer (CT)
Channel (i)
Channel (i+1)
Channel (i)
Using the voltage and current
transformers (PT, CT)
Page 43
25
────────────────────────────────────────────────────
■ Three-phase three wires (3P3W)
Source R S T
U
±
I
±
Channel (i+1)
Channel (i)
U
±
I
±
Load
Source
R
R
S
S
T
T
Connecting the measured
line directly to the unit
■ Three-phase three wires (3V3A)
Source R
S T
Load
Source
R
R
S
S
T
T
CT
U
±
I
±
Channel (i+1)
Using the current
transformer (CT)
CT
Channel (i)
±
±
U
I
Load
Load
Source
R
R
S
S
T
T
CT
PT
U
±
I
±
Channel (i+1)
CT PT
Channel (i)
Using the voltage and current
transformers (PT, CT)
Source
R
R
S
S
T
T
i: 1, 3, 4, 5
Load
R S T
U
±
I
±
i: 1, 4
Load
R S T
1
2
3
4
5
6
U
±
I
±
Channel
(i+2)
U
±
I
±
Channel
(i+1)
U
±
I
±
Channel
(i)
Connecting the measured
line directly to the unit
■ Three-phase four wires (3P4W)
Source R
S T N
U
±
I
±
Channel
(i+2)
U
±
I
±
Channel
(i+1)
±
±
Channel
Load
U
I
(i)
Source
R
R
S
S
T
T
N
N
Connecting the measured
line directly to the unit
CT
±
±
Channel
(i+2)
CT
U
I
Channel
Using the current
transformer (CT)
CT
±
±
Channel
(i+2)
U
I
CT CT
Channel
Using the current
transformer (CT)
U
±
I
±
(i+1)
U
±
I
±
(i+1)
CT
Channel
Channel
U
±
I
±
(i)
Load
U
±
I
±
(i)
PT
CT CT
U
±
I
±
Channel
(i+2)
PT
U
±
I
±
Channel
(i+1)
CT
Using the voltage and current
transformers (PT, CT)
Source
R
R
S
S
T
T
N
N
PT PT
CT
±
±
Channel
(i+2)
U
I
CT CT
U
±
I
±
Channel
(i+1)
Using the voltage and current
transformers (PT, CT)
PT
U
±
I
±
Channel
(i)
i: 1, 4
Load
PT
U
±
I
±
Channel
(i)
7
8
9
10
R S
11
T N
12
13
14
────────────────────────────────────────────────────
3.4 Connecting the Direct Input Unit
A
Page 44
26
────────────────────────────────────────────────────
3.5 Connecting the Clamp Input Unit
The following diagrams show the connections in various modes when using the 9602
AC/DC CLAMP INPUT UNIT
.
Single-phase two wires (1P2W)
Source Load
U
±
SENSOR
I
Channel (i)
i: 1 to 6
Three-phase three wires (3P3W)
Source
R S T
Single-phase three wires (1P3W)
Source Load
N
U
±
SENSOR
I
Channel (i+1)
U
±
SENSOR
I
Channel (i)
Three-phase three wires (3V3A)
Source
Load
R
R
S
S
T
T
N
i: 1, 3, 4, 5
Load
R S T
U
±
SENSOR
I
Channel (i+1)
U
±
SENSOR
I
Channel (i)
i: 1, 3, 4, 5
Three-phase four wires (3P4W)
Source
R S T N
U
±
SENSOR
I
Channel (i+2)
U
±
SENSOR
I
Channel (i+1)
U
±
SENSOR
I
Channel (i)
Load
i: 1, 4
U
±
SENSOR
I
Channel (i+2)
R S T N
U
±
SENSOR
I
Channel (i+1)
U
±
SENSOR
I
Channel (i)
i: 1, 4
────────────────────────────────────────────────────

3.5 Connecting the Clamp Input Unit

Page 45
27
────────────────────────────────────────────────────
3.6 Measurement Losses
This unit is designed to have low measurement losses, and an extremely small effect on the power measurement values, but the following variant connection methods may be used to further reduce the effect of measurement losses.
(1) When the voltage input is connected to the power supply side, the
measurement includes losses from the input resistance of the current input terminals, but this yields the minimum measurement losses when the measured voltage is high and the measured current is low.
Source
±
±
Load
Source Load
U
I
R
Input resistance on
the current side
U
±
I
Input resistance on the voltage side
±
r
(2) When the current input is connected to the power supply side, the
measurement includes losses from the input resistance of the voltage input terminals, but this yields the minimum measurement losses when the measured voltage and measured current is low.
Source
U
±
I
±
Load
Input resistance on
the current side
I
Source Load
Input resistance on
the voltage side
±
r
U
R
±
NOTE
When using a clamp-on input unit as the current sensor, the measurement losses of the current sensor can be ignored, so method (1) above should be used.
────────────────────────────────────────────────────

3.6 Measurement Losses

Page 46
28
────────────────────────────────────────────────────
3.7 Error Messages
"Integration in progress (press START/STOP key to stop)." "Integration on standby (press SHIFT + START keys to reset)." "Reset not possible while integration in progress."
"Time averaging is on."
Operation
Floppy
"Stop time has passed, so real-time control is turned off." "Output in progress." "Hold function operating." "Peak hold function operating." "Program load failed."
"Disk access error" "File cannot be opened" "Save failed" "Load failed" "Formatting failed" "File names may not include spaces." "Disk is write-protected" "Disk full"
Printer
"Printer: head temperature error." "Printer: motor drive voltage error." "Printer: head is up." "Printer: no paper."
If an error message appears when the instrument is turned ON, the unit has malfunctioned. Please contact your local distributor for further assistance.
────────────────────────────────────────────────────

3.7 Error Messages

Page 47
29
────────────────────────────────────────────────────
3.8 System Reset
To reset settings to the initial factory settings, there are following two methods.
・Whenpoweringoff
Turn the power on pressing the
・ On the STATUS screen
1. Press the
2. Using the press the
3. Pressing
STATUS
CURSOR
RESET
(
F5
YES
(
F1
key to display the
keys, move the cursor to "SYSTEM RESET", and
)key.
) carries out system reset.
All settings are reset to the following their initial factory settings.
SHIFT
key until beep sounds.
SYSTEM
screen.
Connection mode 1P2W (all channels) Coupling mode Voltage range Current range PT/CT/SC ratios
LPF phF Peak Response MID Average OFF, averaging t ime: 8 Interval time Timer Real time control Output type Output item ON (all items) Direction of printing Forward Screen save color Monochrome D/A output Frequency
measurement Frequency range Interface GP-IB, address: 1
AC (all channels) AUTO, RMS (all channels) AUTO, RMS (all channels) OFF (all channels) initial
value: 1 OFF (all channels) OFF (all channels) U (all channels)
OFF, initial value: 0h1m00s OFF, initial value: 0h1m00s OFF OFF
all U1 all U1forfa, fb, fc
all AUTO
Display color Normal Backlight Calculation
expression (S,Q) Beep sound ON Out of range input Language Real time Zero suppress
function (Integration) Efficiency screen on
STATUS External input
screen on STATUS Magnification display
for channel 1 to 6 on MEAS
Details display for channel 1 to 6 on MEAS
4 items display on MEAS
8 items display on MEAS
16 items display on MEAS
OFF TYPE1
OFF JAPANESE Current time
0.5%
P1 both denominator and numerator for
Channels A and B: 10 V range, scaling: 1, unit:V
U / I / P /
U / I / P /|Up|/S / Q /
U1/I1/P1/λ1
U1/I1/P1/S1/Q1/λ1/φ1/Pk1
U1/I1/P1/S1/Q1/λ1/φ1/Pk1
both left and right
η1,η2,η
λ
3
λ
────────────────────────────────────────────────────

3.8 System Reset

Page 48
30
t
────────────────────────────────────────────────────
3.9 Operations During Power Failure
CAUTION
・ In the DC and AC+DC modes, after the power is restored, an offset
due to the circuit design may be output. In some cases the data may be invalid.
・ When the unit is powered off as a result of a power failure, continuing
to input voltage and current may damage the unit.
Screen display The screen display goes blank, and after power restoring redisplays the
screen. However, the
STATUSorFDD
screen is displayed before power failure,
the MEAS screen for channel 1 is redisplayed.
Measuremen
data
If the display data was being held when power was lost, all of the data that was being held is not retained.
Integration data ・ Manual integration
A power failure is treated as a zero input and zero elapsed time; after the power is restored integration restarts.
・ Timer integration
A power failure is treated as a zero input and zero elapsed time; after the power is restored integration restarts, and stops when the timer time has elapsed.
・ Real-time control integration
If the power failure starts and ends while the unit remains on standby, there is no effect.
If a power failure starts while the unit is on standby, and ends after the set start time, integration starts from the time when the power is restored. In this case the interval from the set start time until the power is restored is treated as a zero input. The elapsed time is shorter than time from start to stop.
A power failure during integration operation is treated in the same way as for timer integration.
────────────────────────────────────────────────────

3.9 Operations During Power Failure

Page 49
31
────────────────────────────────────────────────────
Floppy disk drive ・ When automatic output is selected (only 3193) After the power is restored, a character string indicating that there was a
power failure is output. (time of power failure and restoring)
・ Power failure during saving
The data being saved is invalid. In the worst case there is a possibility of the file itself being corrupted.
・ Power failure during loading
The setting is invalid. The system reset should be carried out. Turn on the power pressing the
SHIFT
key.
Printer ・ During manual printing
After power is restored, the printing is not started. Restart the printing.
・ During automatic output and before stop time the power is restored
After power is restored, the time when the power failure occurs and the power is restored are printed and then printing is restarted.
・ During automatic output and after stop time the power is restored
After power is restored, the time when the power failure occurs and the power is restored are printed and then operation stops.
────────────────────────────────────────────────────
3.9 Operations During Power Failure
Page 50
32
────────────────────────────────────────────────────
────────────────────────────────────────────────────
3.9 Operations During Power Failure
Page 51
33
────────────────────────────────────────────────────
Chapter 4
1
2
Setting and Using the Basic
Functions
4.1 Setting the Wiring Mode (1P2W to 3P4W)
This unit can have up to six input unit channels, allowing a single unit to measure anything from six 1P2W lines to two 3P4W systems. The connection mode of each channel also appears on the screen as shown below.
41-1.tif
Wiring mode 1P3W 3P3W 3V3A 3P4W
3
4
5
6
7
8
9
Set the line to be measured under " channel combinations set here determine the screen configuration.
UNIT
" on the STATUS screen. The
10
11
12
13
14
A
────────────────────────────────────────────────────

4.1 Setting the Wiring Mode (1P2W to 3P4W)

Page 52
34
────────────────────────────────────────────────────
UNIT
WIRING
UNIT Page
1. Press the
STATUS
keys to display the "
2. Using the
WIRING
"
3. Press
CURSOR
" item.
(
F5
SELECT
key, then use the
UNIT
"page.
keys, move the cursor to the
) to switch to the connection
PAGE
( )
setting screen.
4. In the Wiring screen, a list of the installed input units appears. Move the cursor to the desired combination to be selected.
Wiring Screen
UNIT Page
##41-2.tif
-3
-4
5. Press
F1
SET
(
) to confirm. This automatically
returns to the previous screen.
6. When a number of channels are combined, move the cursor to the wiring item. When using two channels, select from
F2
3V3A
(
3P3W
(
), and for three channels select from
)and
MEAS
key to return to the measurement
F2
3P4W
(
).
and
F1
7. Press the
F1
1P3W
(
screen, where measurement is now possible.
NOTE
On the Wiring screen, if the combination is not changed and then may be initialized. Pressing the
is pressed, the setting items
F1
PAGE
key to exit
from the screen does not initialize the items.
)
────────────────────────────────────────────────────
4.1 Setting the Wiring Mode (1P2W to 3P4W)
Page 53
35
────────────────────────────────────────────────────
NOTE
Measurement line Mode Display item
Single-phase two-wire (1φ2W)
It is only possible to select from the combinations shown in the connection
setting screen. For combinations 1P3W and above, adjacent units must be of the same type.
Channels
1 2 3 4 5 6
1P2W 1P2W 1P2W 1P2W 1P2W 1P2W
1P3W / 3P3W 1P2W 1P2W 1P2W 1P2W
1P3W / 3P3W 1P3W / 3P3W 1P2W 1P2W
1P3W / 3P3W 1P3W / 3P3W 1P3W / 3P3W
3V3A / 3P4W 1P2W 1P2W 1P2W
3V3A / 3P4W 1P3W / 3P3W 1P2W
3V3A / 3P4W 3V3A / 3P4W
When using the 9602
same current sensor type can be selected. In other cases, all are set to 1P2W.
When using the 9602
configuration is changed (including with the sensor not connected), and the unit is then powered on, "Resetting due to configuration change." is displayed. Pressing reset and settings remain unchanged, but the display value for current may be changed. Return the configuration settings and power on again. The message is not displayed.
When using clamp-on units, if the sensor rating is changed and the unit is
then powered on, in the 1P2W mode the CT ratio within the unit is automatically set accordingly. For combinations 1P3W and above, if other channels in the combination have different ratings the combination is disabled for measurement.
The display items in the various modes are as follows.
1P2W U, I, P, Q, S, λ/φ, |Up|/|Ip|
AC/DC CLAMP INPUT UNIT
AC/DC CLAMP INPUT UNIT
(YES)
F1
resets settings. Pressing
, only a combination of the
, if the sensor
(NO)
F2
1
2
3
4
5
6
does not
7
8
9
10
Single-phase three-wire (1φ3W)
Three-phase three-wire (3φ3W)
Three-phase four-wire (3φ4W)
────────────────────────────────────────────────────
1P3W (channels 1+2)
3P3W (2 voltages, 2 currents, 2 power meters method) (channels 1+2)
3V3A (3 voltages, 3 currents, 2 power meters method) (channels 1+2+3)
3P4W (channels 1+2+3)
Although the display will show for each channel 3P3W or 3V3A the active
power (P1, P2 and P3), reactive power (Q1, Q2 and Q3), apparent power (S1, S2 and S3), power factor (
and
φ
3), please be aware that these figures have no meaning.
Power factor (λ) and phase angle (φ) are not be displayed simultaneously. ・ The measurement values for three-phase three-wire in 3P3W and 3V3A mode
are same because of same measurement method.
U1, U2, U12, I1, I2, I12, P1, P2, P12, Q1, Q2, Q12, S1, S2, S12, λ1/φ1, λ2/φ2, λ12/φ12, |U1p|/|I1p|, |U2p|/|I2p|
U1, U2, U12, I1, I2, I12, P12, Q12, S12, λ12/φ12, |U1p|/|I1p|, |U2p|/|I2p|
U1, U2, U3, U123, I1, I2, I3, I123, P123, Q123, S123, λ123/φ123, |U1p|/|I1p|, |U2p|/|I2p|, |U3p|/|I3p|
U1, U2, U3, U123, I1, I2, I3, I123, P1, P2, P3, P123, Q1, Q2, Q3, Q123, S1, S2, S3, S123, λ1/φ1, λ2/φ2, λ3/φ3, λ123/φ123, |U1p|/|I1p|, |U2p|/|I2p|, |U3p|/|I3p|
λ1,λ
2 and
4.1 Setting the Wiring Mode (1P2W to 3P4W)
λ
3), and phase angle (
φ1,φ
2
11
12
13
14
A
Page 54
36
────────────────────────────────────────────────────
4.2 Setting the Coupling Mode (DC/AC+DC/AC)
UNIT
SHIFT
COUPLING
Coupling mode DC AC+DC AC
42-1.tif
##42-2.tif -3
The coupling mode can be selected according to the measurement being performed.
Switching using the panel keys
CURSOR
DC → AC+DC → AC
1. Switch to display the channel for which you wish to change the setting.
2. Press the
SHIFT
key, then use the
CURSOR
change. Pressing the
CURSOR
key switches connection
mode.
Switching on the STATUS screen
1. Press the
STATUS
keys to display the "
2. Select mode from
key, then use the
UNIT
"page.
to
F1
F3
PAGE
( )
.
key to
F1 F2 F3
NOTE
The voltage and current for a single channel and for combinations of 1P3W
and above are forced to the same settings. In this case the setting for the lowest-numbered channel is used.
When DC mode is selected, the polarity is displayed for the voltage and
current.
In DC mode, reactive power (Q), power factor (λ), and phase angle (φ) are
displayed but they are meaningless.
For DC mode, the active power (P) is displayed as a calculated AC+DC
value. For this reason, if there is a superimposed AC waveform the value may not agree with the U×I calculation.
When AC+DC or AC mode is selected, the display values of voltage and
current are always positive values.
When using the 9601 AC DIRECT INPUT UNIT or AC current sensor for the
9602 AC/DC CLAMP INPUT UNIT, DC or AC+DC mode cannot be selected.
────────────────────────────────────────────────────

4.2 Setting the Coupling Mode (DC/AC+DC/AC)

Page 55
37
────────────────────────────────────────────────────
4.3 Switching the Voltage Range and Current Range
When the voltage range and current range is displayed on the screen for each channel, it is also possible to change the ranges directly with the panel keys. This is also possible from the STATUS screen in the "
43-1.tif
Voltage range
43-2.tif
-
Effective input range
The effective input range is 5% to 110% of range. (for the 9600 and 9601, 5% to 100% for 1 kV range only, for the 9602, 5% to 100% for 600 V range only)
UNIT
" display.
Current range
Display range
The value which can be displayed is up to 130% of range. If the value exceeds this range, "o.r" is displayed.
────────────────────────────────────────────────────

4.3 Switching the Voltage Range and Current Range

Page 56
38
────────────────────────────────────────────────────
Switching the range on each channel screen.
RANGE
1. Switch to display the channel for which you wish to change the setting.
U
I
2. Hold down the panel
RANGE
(+,-) key until the
desired setting range is displayed.
6V⇔ 15 V ⇔ 30 V ⇔ 60 V ⇔ 150 V ⇔ 300 V ⇔ 600 V ⇔ 1kV
UNIT
URANGE
F1 F2 F5
NOTE
For channel combinations of 1P3W and above, the channels are forced to the
same range. In this case the range for the lowest-numbered channel is used for all of the channels.
The auto ranging function switches up a range when a measurement value
exceeds 110% of measurement range (out-of-range) or when a waveform peak exceeds six times the range value (peak over), and switches down a range when the value is less than 30% of the nominal range. When measuring a distorted waveform the range selection may not be stable. In this case use manual range setting.
When the integration function time average has started, auto ranging is
disabled, and the range remains fixed from that point.
3. To set auto ranging, hold down the panel
RANGE
(+) key or press both (+,-) keys simultaneously.
4. To cancel auto ranging, press either of the+and keys.
Switching on the STATUS screen
1. Press the keys to display the "
2. Using the
STATUS
CURSOR
key, then use the
UNIT
keys, move the cursor to the
channel to be changed of the "
RANGE
3. Select range from
4. Press the
" item.
(; range down).
F2
MEAS
(AUTO)
F5
key to return to the measurement
"page.
PAGE
U RANGE
,
F1
( )
I
"or"
(; range up),
screen.
-
────────────────────────────────────────────────────
4.3 Switching the Voltage Range and Current Range
Page 57
39
────────────────────────────────────────────────────
4.4 Effective Value (RMS) or Mean Rectified Value
1
(MEAN) Selection
For voltage and current measurement, this unit has two different rectification circuits, which can be selected according to the signal being measured.
SHIFT
44-2.tif
UNIT
RMS/MEAN
RANGE
U
44-1.tif
I
Switching for each channels on MEAS screen
To switch the voltage, press the
then press the+key on the U side.
To switch the current, press the
then press the+key on the I side.
Switching on the STATUS screen
1. Press the keys to display the "
2. Using the channel to be changed of the "U"or"I" item.
3. Select
4. Press the screen.
STATUS
CURSOR
F1
MEAS
(RMS)
key, then use the
"page.
UNIT
keys, move the cursor to the
or
F2
(MEAN)
key to return to the measurement
SHIFT
SHIFT
key, and
key, and
PAGE
.
2
3
4
5
6
( )
7
8
F1 F2
NOTE
Display of "RMS" or "MEAN" following the unit RMS and MEAN values
are distinguished by displaying "RMS" or "MEAN" following the unit, as appropirate. However, these labels do not appear in the DC mode.
The formulas for RMS and MEAN calculation depend on the option.For channel combinations of 1P3W and above, the channels are forced to the
same rectification method, but distinct settings can be made for voltage and current.
The RMS and MEAN values agree when the input is a perfect sine wave,
but do not agree for a distorted waveform.
Whichever of RMS and MEAN is selected, this has no effect on the active
power (P), but does affect the internally derived apparent power (S), reactive power (Q), power factor (λ), and phase angle (φ).
In DC mode the RMS/MEAN selection is not available.
9
10
11
12
13
14
A
────────────────────────────────────────────────────

4.4 Effective Value (RMS) or Mean Rectified Value (MEAN) Selection

Page 58
40
────────────────────────────────────────────────────
4.5 Setting the Scaling (PT/CT/SC Ratios)
This is used for setting the ratio (PT ratio or CT ratio) when using an external voltage transformer (PT) or current transformer (CT), and the scaling factor (SC ratio) for conversion of the active power to other physical units. When a PT ratio, CT ratio, or SC ratio is set for a particular channel, on the measurement screen this appears as "SC". The ranges which can be set are as shown below.
Scaling operation
PT CT SC
46-1.tif
Scaling constant
PT ratio (Kp) CT ratio (Kc)
SC (
Ksc
)
UNIT
Display Setting range
PT 0.0001 to 10000.
CT 0.0001 to 10000.
SC 0.0001 to 10000.
PT
Function Equation
Voltage Current
Active power Apparent power
Reactive power Integrated current
Integrated voltage
1. Press the
STATUS
keys to display the "
2. Using the
CURSOR
U
I P S
Q
Ih
WP
key, then use the
UNIT
"page.
keys, move the cursor to the
U×Kp
I×Kc
P×Kp×Kc×Ksc
S×Kp×Kc×Ksc
Q×Kp×Kc×Ksc
Ih×Kc
WP×Kp×Kc×Ksc
PAGE
desired channel of the PT, CT or SC item.
3. Set the numerical value 0 to 9 by using up), by using use the
(; down), and move the decimal point
F2
F3
CURSOR
(.),
keys.
(.). To move the digit,
F4
( )
F1
(;
4. Press the
MEAS
key to return to the measurement
screen.
46-2.tif
F1 F2 F3 F4 F5
────────────────────────────────────────────────────

4.5 Setting the Scaling (PT/CT/SC Ratios)

Page 59
41
────────────────────────────────────────────────────
NOTE
For channel combinations of 1P3W and above, the PT ratio, CT ratio, or SC
ratio must be the same for all channels in the combination.
When the channel identifies a CT6862, CT ratio of the channel is set to 2.5,
as default. In this state, the current ranges are configured and the measured values can be read off directly.
Display ranges:「1.25(A) / 2.5A / 5A / 12.5(A) / 25A / 50A」
The default CT ratio is changeable. If you want to change this, please set the value calculated by the formula: (your desired CT ratio x 2.5). For example, when you want to set 40 as CT ratio, please set 100(=40 x 2.5). When CT ratio other than 2.5 (including OFF) is being set, the current ranges to be displayed are below.
Display ranges:「0.5A / 1A / 2A / 5A / 10A / 20A」
The actual current range configuration with the CT ratio other than 2.5 can be calculated by multiplying the set CT ratio by each of 0.5A, 1A, 2A, 5A, 10A, 20A.
When a CT6865 is used, please set it to CT ratio 2.
The display ranges are [10A/20A/50A/100A/200A/500A], but the range structure is actually configured to be double the CT ratio after internal computation.
────────────────────────────────────────────────────
4.5 Setting the Scaling (PT/CT/SC Ratios)
Page 60
42
────────────────────────────────────────────────────
4.6 Setting the Low-pass Filter (LPF)
The input units of the 3193 are provided with a low-pass filter function for restricting the frequency characteristics. By using an appropriate filter selection it is possible to eliminate harmonics.
Low-pass filter operation
47-1.tif CHANGE
UNIT
NOTE
Depending on using the input units, the low-pass filter (LPF) may not be selected. The frequency range of accuracy assured varies. For details, see the specifications of the input units.
CURSOR
SHIFT
OFF → 500 Hz→ 5 kHz → 300 kHz
LPF
Switching using the panel keys
1. Switch to display the channel for which you wish to change the setting.
2. Press the
SHIFT
key, then use the
CURSOR
key to change. Pressing the
CURSOR
key switches low-pass
filter.
Switching on the STATUS screen
1. Press the keys to display the "
2. Using the channel to be changed of "
STATUS
CURSOR
key, then use the
UNIT
"page.
keys, move the cursor to the
LPF
3. Select desired low-pass filter from
500 Hz
(
),
F3
5 kHz
(
F4
300 kHz
(
),
" item.
).
PAGE
F1
( )
(OFF)
,
F2
4. Press the
MEAS
key to return to the measurement
screen.
F1 F2 F3 F4
47-3.tif 47-2.tif
────────────────────────────────────────────────────

4.6 Setting the Low-pass Filter (LPF)

Page 61
43
────────────────────────────────────────────────────
4.7 Setting the Phase Polarity Discrimination Filter
For distorted waveforms such as inverter waveforms, the reactive power (Q), power factor (λ), and phase angle (φ) phase angle polarity may not be stable. In this case, by setting the phase polarity discrimination filter to "ON" stable polarity measurements can be taken.
Phase polarity discrimination
48-1.tif
filter operation
UNIT
F1 F2
NOTE
phF
1. Press the
STATUS
key, then use the
keys to display the "
2. Using the
CURSOR
keys, move the cursor to the
channel to be changed of "
3. Select
4. Press the
F1
(OFF)
MEAS
or
key to return to the measurement
UNIT
F2
"page.
phF
(ON).
PAGE
" item.
screen.
48-2.tif
This is valid when the calculation formula for reactive power (Q)and
apparent power (S) is set to "
TYPE2
"
"or"
TYPE3
" the ON/OFF setting of phF has no effect.
This has no effect on voltage (U), current (I), active power (P), or apparent
power (S).
For channel combinations of 1P3W and above, the channels are forced to the
same setting.
The filter cut-off frequency is 200 Hz, and therefore depending on the
frequency components of distorted waveforms, in some cases stabilization may not be possible.
When the frequency of measurement waveform is 200 Hz or above, set to
OFF.
When using averaging function, the polarity is not displayed.When the input signal level is 50% or less of the range, the polarity may not
be discriminated appropriately.
TYPE1
." When the calculation formula is set to
( )
────────────────────────────────────────────────────

4.7 Setting the Phase Polarity Discrimination Filter

Page 62
44
────────────────────────────────────────────────────
4.8 Switching the Waveform Peak Value
The waveform peak value measurement can be set to voltage waveform (|Up(i)|) or current waveform (|Ip(i)|)
UNIT
F1 F2
NOTE
1. Press the keys to display the "
2. Using the
STATUS
CURSOR
key, then use the
UNIT
keys, move the "
the desired input unit and press
"page.
PAGE
Peak
F1
" item of
(I)or
(U).
Peak
420-1.tif
For any one input unit, it is not possible to measure the voltage peak value
and current peak value simultaneously.
The peak value is given as an absolute value, and it is not possible to
determine the sign.
The averaging function has no effect on peak value measurement.When measuring a waveform with a superimposed DC component in AC
mode, the peak value of the AC waveform is found after the DC component has been eliminated.
When a low-pass filter is activated, the peak value is found after the
waveform has passed through the filter.
For the operation principle of peak value measurement and specifications,
see specifications of input units.
( )
F2
────────────────────────────────────────────────────

4.8 Switching the Waveform Peak Value

Page 63
45
────────────────────────────────────────────────────
4.9 Setting the Response (FAST/MID/SLOW)
There are three settings for the response time of analog outputs from this unit: FAST, MID, and SLOW. For measurement of a normal commercial power supply, the FAST setting is adequate, but is the frequency is low or there are sudden fluctuations, setting the response to MID or SLOW makes the display more stable.
49-1.tif
Response display FAST MID SLOW
Switching using the panel keys
SHIFT
TIME
RESPONSE
F1 F2 F3
CURSOR
1. Switch to display the channel for which you wish to change the setting.
2. Press the
SHIFT
key, then use the
CURSOR
key to
change. Pressing the
CURSOR
key switches response
mode.
Switching on the STATUS screen
1. Press the
STATUS
keys to display the "
2. Using the
RESPONSE
"
3. Select
4. Press the
CURSOR
" item.
F1
MEAS
(FAST)
key, then use the
TIME
"page.
keys, move the cursor to the
F2
(MID)
,
key to return to the measurement
PAGE
( )
F3
(SLOW)
.
,
screen.
49-2.tif
CHANGE
NOTE
The response setting applies to all channels together. It is not possible to
make separate settings for each channel.
The display refresh rate does not depend on the response setting.The response setting does not affect a channel for which the DC mode is
selected.
If the display is unstable even with the SLOW setting, use the averaging
function as well.
────────────────────────────────────────────────────

4.9 Setting the Response (FAST/MID/SLOW)

Page 64
46
────────────────────────────────────────────────────
4.10 Setting the Averaging
(Time averaging/Moving averaging/Exponential averaging)
This unit provides three averaging functions. The time average outputs the average over a fixed time interval, and the moving average and exponential average provide values which reflect the previous values.
Averaging display
NOTE
AV-T
AV-M
AV-E
Time averaging
Moving averaging
Exponential averaging
This setting applies to all channels together. It is not possible to make
separate settings for each channel.
This function does not affect the harmonic analysis/flicker function.The D/A outputs give the values obtained by averaging.Waveform peak measurement values are not averaged.If [TYPE1] is selected as the calculation type while using averaging (time
averaging, moving averaging, or indexed averaging), the "si" and "su" polarities for each channel are calculated as +1.
Calculation is performed 16 times per second.
(1) Setting the Time Average (AV-T)
The time average function sequentially sums the data values obtained during the time interval (interval control time, timer control time, or real-time control time), and divides by the number of samples. This can be used, for example, while integrating at particular intervals, to output the average of some other measurement value during those intervals.
ΣZn Display value = Nn
────────────────────────────────────────────────────

4.10 Setting the Averaging

―――
Zn: nth measured data Nn: number of storing during setting time
Page 65
47
────────────────────────────────────────────────────
TIME
AVE TIME
F1 F2 F3 F4
10-3.TIF #CHANGE
NOTE
The time average requires a setting of an interval control time, timer control
time, or real-time control time.
Operation with the START/STOP key also affects integration. Therefore,
after completion of averaging or following forced termination, press the SHIFT key, then press the START/STOP key to reset integration.
When used in conjunction with a timer control time or real-time control
time, when the set time elapses the overall average value is displayed and held, which the unit stops. When this hold state is ended, the unit returns to normal measurement.
For the time average function, floppy disk and printer settings are also
synchronized in the same way.
During the time average function operation, it is not possible to change
settings.
If an interval control time is set and the
updated each time the interval elapses.
During the averaging operation, if an out-of-range data value is included, the
unit indication is red.
No polarity is shown for time averaging when [TYPE1] is selected as the
calculation type for reactive power (Q).
1. Press the
STATUS
keys to display the "
2. Using the "
AVERAGING
3. Press
F2
CURSOR
" item.
(
ave time
key, then use the
" (time control) page.
TIME
PAGE
( )
keys, move the cursor to the
).
4. Select desired time setting from interval time, timer time, real control time. See Section 7.2.
5. Press the
MEAS
screen. Pressing the
key to return to the measurement
START/STOP
key starts
averaging.
6. To return to the normal measurement, release the hold state, because the operation stops in hold state ("HOLD" indicates) when the time averaging is completed. When used together with the interval time, the hold value is the average within the final interval time. To switch display to the total average, press the
HOLD
key. The TOTAL mark
appears during display of the total average.
7. To complete forcibly, press the
START/STOP
key
again.
HOLD
key is pressed, the display is
────────────────────────────────────────────────────
4.10 Setting the Averaging
Page 66
48
────────────────────────────────────────────────────
(2) Setting the Moving Average (AV-M)
The moving average function displays a simple average calculated by summing the measurement values from the beginning of averaging, and dividing by the number of samples, until the specified number of samples. From that point on it discards the oldest data value as each new value is added, thus yielding a simple average over the most recent specified number of samples.
TIME
TIME
Z Display value =
――――――――――――――
(n-N-1)
+Z
N
1. Press the keys to display the "
2. Using the
AVERAGING
MOVING AVE
"
3. Press
(n-N-2)
F3
+...+ Zn
STATUS
CURSOR
" item.
Moving ave
(
Zn: nth measured data N: number of setting samples
key, then use the
TIME
" (time control) page.
keys, move the cursor to the
).
4. Move the cursor to the specified number of sampling item on the right, and set the number of sampling using the function key.
5. Press the
MEAS
key to return to the measurement
screen, and averaging starts.
PAGE
( )
F1 F2 F3 F4
NOTE
This has no connection with any time settings.
────────────────────────────────────────────────────
4.10 Setting the Averaging
Page 67
49
────────────────────────────────────────────────────
(3) Setting the Exponential Average (AV-E)
The exponential average provides a average of the previous values, but weighted toward the latest value. The effect of previous values thus diminishes exponentially.
(N-1) A Display value = ――――――― N
TIME
EXPONENTIAL
F1 F2 F3 F4
NOTE
This has no connection with any time settings.
+Zn
n-1
1. Press the keys to display the "
2. Using the
AVERAGING
"
3. Press
Zn: nth measured data
: n-1 th dis play value
A
n-1
N: constant setting
STATUS
CURSOR
key, then use the
TIME
keys, move the cursor to the
" item.
Exponential
F3
(
).
PAGE
" (time control) page.
4. Move the cursor to the specified number of sampling item on the right, and set the constant value using the function key.
5. Press the
MEAS
key to return to the measurement
screen, and averaging starts.
( )
────────────────────────────────────────────────────
4.10 Setting the Averaging
Page 68
50
────────────────────────────────────────────────────
4.11 Setting on the MEAS Screen
4.11.1 Setting the Display Items (for 1 to 6 channels)
For items on "
DETAILS
which measurement items to display. It is also possible to select whether to display the power factor (λ)orphase angle (φ). On the "
419-1.tif -2 -3
F1 F2 F5
" for each channel screens, it is possible to select
MAGNIFY
" screen, up to 4 items can be set.
Display item setting for "DETAILS" display
1. Select the screen of the display item to be cleared and press
2. Using the
F5
(SELECT).
CURSOR
keys, move the cursor to the desired item. To clear the display, press
3. Press
F5
(
RETURN
) to return to the measurement
F1
(OFF)
.
screen.
Switching power factor/phase angle
1. Select the screen to switched between power factor or phase angle, and press
F5
(SELECT).
F1 F2 F3 F5
2. Using the
CURSOR
keys, move the cursor to the
desired item and select from
(OFF)
F5
.
(
RETURN
) to return to the measurement
F1
3. Press screen.
F2
(λ),
F3
(φ),
────────────────────────────────────────────────────

4.11 Setting on the MEAS Screen

Page 69
51
────────────────────────────────────────────────────
Setting items on "MAGNIFY" display
1. Select the magnification display to be set on channel screen, and press
F5
(SELECT).
The item list which can be selected is displayed.
2. Using the desired display and select item from
F2
To clear the setting display press
3. Press
F1
F2
NOTE
On the integration display screen, the integration item cannot be set.
F5
On the magnification display, the integration value cannot be selected.
4.11.2 Setting the SELECT screen
This function allows a desired subset of the measurement values to be selected, and displayed together on the screen. There are three modes: with four, eight, or 16 items displayed. The modes are selected with function keys The following procedure describes how to select a four-item display, and the procedure for eight or 16 items is similar. The method of setting, see Section 4.11.1, "Setting the Display Item."
CURSOR
keys, move the cursor to the
() to decide setting.
F5
F1
(RETURN)
to
to complete settings.
.
F3
F2
F1
(OFF).
()or
F1 F2 F3 F5
421-1.tif
NOTE
────────────────────────────────────────────────────
In the SELECT screen, the panel key operations for voltage range, current
range, RMS/MEAN selection, and degaussing (DMAG) are disabled. If changes are required, switch to the STATUS "
" display for the relevant
UNIT
channel.
On the SELECT screen, only response setting can be made. Set for others on
the STATUS screen or on the display screen for each channels.
The integration value can not be selected.
4.11 Setting on the MEAS Screen
Page 70
52
────────────────────────────────────────────────────
4.12 Setting on the SYSTEM screen
4.12.1 Switching the Interface (GP-IB/RS-232C)
This unit has GP-IB and RS-232C interfaces fitted as standard, and either one can be used as required.
F1 F2
SYSTEM
INTERFACE
1. Press the
STATUS
keys to display the "
2. Using the "
INTERFACE
CURSOR
" item.
key, then use the
SYSTEM
"page.
PAGE
( )
keys, move the cursor to the
413-1.tif CHANGE
3. Select from
4. When "RS-232C" is selected, set the baud rate (2400/9600 bps), data length (7/8 bits), stop bit (STOP1/STOP2), parity (none; PN/ odd:PODD/ even; PEVEN). When "GP-IB" is selected, set address (0 to 30).
5. Press the
MEAS
screen.
F1
(
RS-232C
)or
F2
(
GP-IB
).
key to return to the measurement
F1 F2
413-2.tif CHANGE
NOTE
────────────────────────────────────────────────────

4.12 Setting on the SYSTEM screen

GP-IB and RS-232C interfaces cannot be set simultaneously.For setting the GP-IB and RS-232C, see Chapter 12, "GP-IB and RS-232C
Interface."
Page 71
53
────────────────────────────────────────────────────
4.12.2 Setting the Display Color
You can select from four patterns for the screen display colors.
SYSTEM
DISP COLOR
F1 F2 F3 F4
423-1.tif CHANGE
4.12.3 Setting the Back Light
The backlighting time of the color LCD panel on the unit can be set. In the absence of any key presses, the backlighting goes off automatically after the specified time has elapsed.
1. Press the
STATUS
keys to display the "
2. Using the
CURSOR
item.
3. Select color from
key, then use the
SYSTEM
"page.
keys, move the "
F1
to
F4
.
( )
PAGE
DISP COLOR
"
F1 F2
NOTE
SYSTEM
BACKLIGHT
The time can be set to "OFF" (00 minutes), or from 1 to 99 minutes.The backlighting lifetime is about 50,000 hours.Pressing any key while the backlighting is off turns it on again. This first
1. Press the
416-1.tif 416-2.tif
keys to display the "
2. Using the "
BACKLIGHT
3. Select from
4. Press the screen.
F5
416-1.tif CHANGE
key press is otherwise ignored.
STATUS
CURSOR
" item.
F1
MEAS
key, then use the
SYSTEM
"page.
PAGE
( )
keys, move the cursor to the
(↑)
or
F2
(↓).
key to return to the measurement
────────────────────────────────────────────────────
4.12 Setting on the SYSTEM screen
Page 72
54
────────────────────────────────────────────────────
4.12.4 Setting the Equation for Reactive Power (Q)and Apparent Power (S)
This unit provides three different internal ways of computing the reactive power and apparent power. Select whichever is appropriate. See Section 20.3, "Calculations."
SYSTEM
CALCULATION
F1 F2 F3
NOTE
For calculation, see Section 20.3, "Calculations"In general, use "
The values yielded by the different methods of calculation coincide when the
When "
The power factor and phase angle are also derived from the selected
If [TYPE1] is selected as the calculation tpe while using averaging (time
1. Press the
STATUS
keys to display the "
2. Using the "
CALCULATION
CURSOR
3. Select calculation type from (
4. Press the
TYPE2
),
F3
MEAS
key, then use the
SYSTEM
"page.
PAGE
( )
keys, move the cursor to the
" item.
,
F2
(
TYPE3
F1
(TYPE1)
).
key to return to the measurement
screen.
414-1.tif CHANGE
". Select "
TYPE1
compatibility with previous models.
input is a sine wave on a balanced line, but may be different when there is distortion present or unbalanced line.
" is selected, the reactive power calculation for 1P3W and
TYPE1
above includes the lead/lag of the current with respect to the voltage. The detection of the sign of this lead or lag is carried out by zero-crossing detection on the voltage and current waveforms by the input units. For this reason, stable measurement may not be possible when the waveforms are distorted. In such cases, use a phase polarity discrimination filter (phF) in addition. See Section 4.7.
calculation result.
averaging, moving averaging, or indexed averaging), the "si" and "su" polarities for each channel are calculated as +1.
TYPE2
"or"
" when required for
TYPE3
────────────────────────────────────────────────────
4.12 Setting on the SYSTEM screen
Page 73
55
────────────────────────────────────────────────────
4.12.5 Setting the Beep Sound
This unit sounds a "beep" each time a key is pressed.
BEEP
SYSTEM
1. Press the keys to display the "
2. Using the
BEEP
"
3. Select from
4. Press the
STATUS
CURSOR
" item.
MEAS
key, then use the
SYSTEM
keys, move the cursor to the
(OFF)
F1
or
F2
key to return to the measurement
screen.
F1 F2
415-1.tif
NOTE
When an error occurs, the beep sounds, irrespective of this setting.
4.12.6 Setting Indications for Out-of-Range Inputs
This functions selects whether or not the numerical values appear in a different color when outside the set range.
"page.
(ON).
PAGE
( )
F1 F2
SYSTEM
DEFINED
422-1.tifCHANGE
Only indication of numerical value is faint.
1. Press the
STATUS
keys to display the "
2. Using the
3. Press
NOTE
CURSOR
F1
OFF
(
For the out of range input, see the specifications of
each input units.
An active power value is treated as out of range
when either of the voltage and current values is out of range.
A reactive power (Q), apparent power (S), power
factor (λ), or phase angle (φ) value is treated as out of range when either of the voltage (U), current (I), power (P) values is out of range.
key, then use the
SYSTEM
"page.
keys, move the "
)or
F2
(ON).
PAGE
( )
DEFINED
" item.
────────────────────────────────────────────────────
4.12 Setting on the SYSTEM screen
Page 74
56
────────────────────────────────────────────────────
4.12.7 Setting the Display Language (English/Japanese)
Display messages can be selected to appear in either Japanese or English.
SYSTEM
1. Press the keys to display the "
2. Using the
LANGUAGE
"
3. Select from
4. Press the
LANGUAGE
screen.
Instead of above procedure, hold down the STATUS
F1 F2
key and turn the power on, the display in English appears.
4.12.8 Setting the Real-time Clock
This sets the internal real-time clock.
SYSTEM
1. Press the keys to display the "
STATUS
CURSOR
key, then use the
SYSTEM
"page.
keys, move the cursor to the
" item.
(JAPANESE)
F1
MEAS
key to return to the measurement
STATUS
key, then use the
SYSTEM
or
F2
"page.
PAGE
ENGLISH
(
PAGE
( )
).
( )
REAL TIME CLOCK
F1 F2 F5
418-1.tif CHANGE
NOTE
The clock uses 24-hour time.Support until 2078.
F5
(SET)
is not pressed after time is set, the setting time after powering
If
off is returned to the previous setting.
2. Using the digit to be set on the right of the "
CLOCK
3. Set the time using
CURSOR
" item.
keys, move the cursor to the
REALTIME
(↑)
F1
or
F2
(↓).
The settings are completed after the time is set.
4. Press
F5
(SET)
to end setting.
────────────────────────────────────────────────────
4.12 Setting on the SYSTEM screen
Page 75
57
────────────────────────────────────────────────────
4.13 Degaussing
When a large DC current or large transient current is measured with the
AC/DC DIRECT INPUT UNIT
9600 the 9602
AC/DC CLAMP INPUT UNIT
magnetized, thus outputting an offset even for a zero input. Use the degaussing function if this occurs, and also before measurement after warming-up.
45-1.tif
or an AC/DC type of current sensor for
, the internal DC-CT may become
NOTE
1. Display the channel screen for which degaussing is desired.
2. Press the
SHIFT
key once, then press the current
range-key.
SHIFT
RANGE
U
I
3. A degaussing message appears on the screen. "Will now degauss" Pressing
F1
YES
(
) starts degaussing, and
(NO) does not degauss and the message goes off.
4. The degaussing operation is completed in about 10 seconds.
Carry out degaussing when the input current is zero.This function is not available for AC direct input units and AC current
sensors.
The specification accuracy applies only after degaussing.The degaussing function operates once after powering on.For channel combinations of 1P3W and above, degaussing takes place
simultaneously for all channels in the combination.
In extreme cases, for example, when there is an input of the maximum input
current or above, complete degaussing may not be possible. In this case, power the unit off and on again.
F2
────────────────────────────────────────────────────

4.13 Degaussing

Page 76
58
────────────────────────────────────────────────────
────────────────────────────────────────────────────
4.13 Degaussing
Page 77
59
────────────────────────────────────────────────────
Chapter 5
1
2
NOTE
Range
Frequency Measurement
This unit has internal circuits for three frequency measurement channels (fa, fb, fc), and can thus measure a number of systems simultaneously. The frequency ranges can be combined with high-pass filters (HPF) and low­pass filters (LPF)
Depending on the frequency range and the frequency and distortion of the
waveform, stable measurement may not be possible. In that case, set the range manually.
The frequency effective measurement range is within the range of frequency
characteristics of combination input unit.
When using the 9603
measurement, this is unconditionally assigned to the 9603 uni t. For details see Chapter 17, "9603
(9603 has no HPF and LPF)
500.00 mHz
to 50.000 Hz
20.00 Hz to
EXTERNAL SIGNAL INPUT UNIT
EXTERNAL SIGNAL INPUT UNIT
200.00 Hz to
500.00 Hz
5.0000 kHz
you select pulse
."
2.0000 kHz to
50.000 kHz
20.000 kHz to
2.0000 MHz
3
4
5
6
7
8
9
HPF 0.5 Hz 100 Hz 10 kHz
LPF 360 Hz 50 kHz 1.2 MHz
The frequency measurement data is displayed on the lower right of the "DETAILS" page on each channel screen
51.tif
Frequency measurement data
10
11
12
13
14
A
────────────────────────────────────────────────────
Page 78
60
────────────────────────────────────────────────────
5.1 Setting the Frequency Measurement Source (fa)
FREQ/OUTPUT
FREQUENCY
F1 F2 F3 F4 F5
NOTE
After settings are completed, always pressBy setting fa to "U1" and fb to "I1", for example, it is not possible to
simultaneously measure the voltage and current waveforms from a single input unit.
1. Press the display the "
2. Using the
STATUS
CURSOR
key, then use the
FREQ/OUTPUT
keys, move the cursor to the
PAGE
"page.
source item of "fa", and the window of settable source opens.
3. Press desired source item, and press
F1
()and
() to move the cursor to
F2
(
SET
).
F5
For fb, fc, same method
52.TIF CHANGE
(
SET
).
F5
key to
────────────────────────────────────────────────────

5.1 Setting the Frequency Measurement Source (fa)

Page 79
61
────────────────────────────────────────────────────
5.2 Setting the Frequency Range(fa)
FREQ/OUTPUT
FREQ RANGE
F1 F2 F3 F4 F5
53.tif CHANGE
NOTE
〜ー>ー
If the measurement is not possible, the following error is displayed. When the input is out of setting frequency range: "o.r." When it is in AUTO range: "-----"
1. Using the
CURSOR
keys, move the cursor to the frequency range item of "fa", and the window of frequency range opens.
2. Press
F1
()and
() to display the desired
F2
range.
0.5 Hz to 50 Hz ⇔ 20 Hz to 500 Hz ⇔ 200 H to 5 kHz ⇔ 2kHzto50kHz⇔ 20 kHz to 2 MHz
Sets fb and fc in the same way.
────────────────────────────────────────────────────

5.2 Setting the Frequency Range(fa)

Page 80
62
────────────────────────────────────────────────────
────────────────────────────────────────────────────
5.2 Setting the Frequency Range(fa)
Page 81
63
v
────────────────────────────────────────────────────
Chapter 6
1
2
6.1 Hold Function
Pressing the panel switching from one screen to another it is possible to compare different simultaneously captured values. Since internally the measurement continues, each time you press the key the values at that time are displayed. To end the hold function, hold down the
SHIFT
Hold/Peak Hold Function
HOLD
key freezes the display values of all items. By
HOLD
key and press the
HOLD
key.
Hold state
61.tif
3
4
5
6
7
8
Display value
Start holding Update Release holding
Hold
Display
alue
Internal measured values
HOLD
Hold
HOLD
UpdateUpdate
SHIFT
HOLD
9
10
11
12
13
Time
14
A
────────────────────────────────────────────────────

6.1 Hold Function

Page 82
64
v
────────────────────────────────────────────────────
NOTE
In the hold state, it is not possible to change settings.In the auto-ranging, the range when the
HOLD
key is pressed is fixed.
In the hold state, external output values (for floppy disk or printer, through
the GP-IB or RS-232C interface, or D/A output) are the values displayed on the screen. When combined with the interval timer, the display is updated at the specified interval. In this case the previous value is held until the next interval time.
For timer and real-time control, the display is updated at the stop time, and
then held.
The HOLD key operates at any time, including prior to and during timer
operation.
Combination with external control signal
When the unit is in the held state, the display can be held or updated using an external control signal to the EXT. A/D START terminal of the OUT (ANALOG WAVE D/A), EXT. CONT connector on the rear panel.
Display value
Start holding Update
Update
Hold
Update
Hold
Release holding
Hold
Update
HOLD
Display
alue
Lo Lo
Time
SHIFT
HOLD
────────────────────────────────────────────────────
6.1 Hold Function
Page 83
65
────────────────────────────────────────────────────
6.2 Peak Hold Function
1
When the peak hold function is activated, only items exceeding the previous maximum value are updated continuously. For example, this can be used for measuring transient currents in an electric motor.
62.tif
Display value
Start peak holding
Update
Peak hold state
Release peak holding
Peak hold state
Display value
2
3
4
5
6
7
8
Internal measured values
Time
SHIFT
HOLD
To activate or deactivate this function, press the
HOLD
the key resets the peak value, and starts a new peak hold operation from that point.
key. When the peak hold function is activated, pressing the
Display value
Start peak holding
Update
Display value
Internal measured values
Peak hold state
Release peak holding
Reset
SHIFT
HOLD
SHIFT
key and then press
HOLD
Time
9
10
11
12
13
14
SHIFT
HOLD
────────────────────────────────────────────────────
HOLD
SHIFT
HOLD

6.2 Peak Hold Function

A
Page 84
66
g
v
v
────────────────────────────────────────────────────
NOTE
If the display value is out of range, the indication "o.r." appears. In this case,
first stop the peak hold function, then change the range.
The maximum value refers to the maximum absolute value. For example,
after an input of "+50 W", an input of "-60 W" causes the display to be updated, because the absolute value of "-60 W" is greater.
In the peak hold state, it is not possible to change settings.In the peak hold state, external output values (for floppy disk or printer,
through the GP-IB or RS-232C interface, or D/A output) are the values displayed on the screen.
6.2.1 Combination with Control Times
When an interval time is set, the maximum value within each interval can be measured. When a timer time or real-time control time is set, the maximum value from the start time to the stop time is found, and then the unit stops.
Interval time
Display value
Start peak holding
Display value
Start peak holding
SHIFT
HOLD
Peak hold
SHIFT
HOLD
Internal measured
alues
START /STOP
Start interval
Maximum value measured
Internal measured
alues
START /STOP
Start timer
Start timer
Display value
Release interval
Display value
End timer
START /STOP
Hold
End timer
Time
Release peak holdin
SHIFT
HOLD
Time
SHIFT
HOLD
Release peak holding
NOTE
The peak hold function can be activated either before the time setting is
made, or after the time has started. However, if after the time has started, the maximum value is only found from the time after the peak hold function is activated.
The time of occurrence of the maximum value is not shown.
────────────────────────────────────────────────────
6.2 Peak Hold Function
Page 85
67
────────────────────────────────────────────────────
Chapter 7
1
2
7.1 Overview
Integration Function
For a 1P2W system in DC mode, the integration function in this unit can simultaneously integrate positive, negative, and total values for current (I) and active power (P) for all channels. There are six ways of controlling integration by the various time settings, as listed below. All of these ways can be combined with the floppy disk drive and printer. Load factor (LF) calculation is also possible using the interval timer.
1. Manual integration
2. Interval integration
3. Timer control integration
4. Timer + interval integration
5. Real time integration
6. Real time + interval integration
3
4
5
6
7
8
9
NOTE
Because of the internal circuit design, the maximum integration time is
10,000 hours, and the integration automatically stops if this time elapses.
Start, stop, and reset control of integration by either a panel operation or the
external control connector applies simultaneously to all items being integrated. However, the GP-IB or RS-232C interface provides separate control for each channel. For details see Chapter 12, "GP-IB and RS-232C Interface."
The items which can be integrated depend on the connection mode and
coupling mode as shown in the following table.
Mode Display items
1P2W, DC +Ih, -Ih, Ih, +WP, -WP, WP, LF
1P2W Ih, +WP, -WP, WP, LF
(When using channels 1, 2)
1P3W, 3P3W
3V3A, 3P4W
(When using channels 1, 2, 3)
Ih1, Ih2, +WP12, -WP12, WP12, LF12
Ih1, Ih2, Ih3, +WP123, -WP123, WP123, LF123
10
11
12
13
14
A
────────────────────────────────────────────────────

7.1 Overview

Page 86
68
────────────────────────────────────────────────────
NOTE
Data for each interval of the interval time setting is displayed on the screen
in hold state. When the value is not held, it must be combined with the floppy disk drive or printer function to display.
Calculation results (DC voltage) from the various input units are integrated
at the rate of 64 samples per second. Therefore, integration results may vary with instruments that use different response speeds, sampling rates or calculation methods.
During integration operation, if the integration item is "o.r" (out-of-range),
the value of 130% of range is integrated. In this case, the integration value indication is red.
In the DC mode, current is integrated from instantaneous current waveforms.
In the AC+DC/AC mode, it is integrated as RMS or MEAN values.
In the DC mode, power is integrated from instantaneous power waveforms.
In the AC+DC/AC mode, it is integrated as effective power.
Once integration starts, items for which auto ranging is set have their ranges
fixed from the start time. Set the range so that it will not be exceeded before beginning integration.
During integration (even "on standby" under real-time control), all settings
except changing screens and the hold and peak hold functions are disabled.
While the hold function is activated, the display is frozen, but internally the
integration continues normally. When the floppy disk drive or D/A outputs are used, however, the held values continue to be output.
While the peak hold function is activated, the integration operation is not
affected.
After integration ends, until the integration values are reset, it is not possible
to change settings.
The calculation results (for DC voltage) from each input units are integrated
64 times per second.
Operation Procedure
1. Set the required time settings (interval, timer, or real-time control). For the manual integration, make the time settings to OFF, however, the timer time operates as 10000 hours.
2. Set the external output for FDD, printer, D/A if necessary. See Chapter 10, Chapter 11, Chapter 13
3. Start integration. Press the
START/STOP
key.
────────────────────────────────────────────────────
7.1 Overview
Page 87
69
────────────────────────────────────────────────────
7.2 Setting the Control Time
1
Using the three time control functions provided by this unit, it is possible to control time averaging, the floppy disk drive, printer, and integration functions.
NOTE
It is not possible to make separate settings for time averaging, floppy disk
drive, printer, and integration functions.
The integration function always operates. Therefore, even when the
integration value is not being measured, during the operation period of time control, the " (
SHIFT→START/STOP
When using time averaging integration, some settings are disabled. If auto
ranging is set, the range is fixed at the start time. The HOLD function, however, can still be used.
Even when time settings are made, the unit does not operate until the panel
START/STOP
The operation starts by real-time control and it is completed by timer
control. In this case, the stop time by real-time control is ignored.
" indication appears. After end of time control, reset
INTEG
) to turn off the "
key is pressed.
7.2.1 Setting the Interval Time
INTEG
2
3
4
" display.
5
6
7
TIME
F1 F2
TIME
INTERVAL
Interval control operates the unit repeatedly at the specified interval.
1. Press the keys to display the "
2. Using the "
INTERVAL
3. Move the cursor to desired time digit, and set the time u sing the
4. After setting completed, press the return to the measurement screen.
5. Press the
STATUS
CURSOR
" and press
START/STOP
key, then use the
" (time control) page.
TIME
keys, move the cursor to the
to set to ON.
F2
F1
and
(↑)
key to start operation.
F2
PAGE
(↓).
MEAS
key to
411-1.tif 411-2.tif CHANGE
8
( )
9
10
11
12
13
14
F1 F2
────────────────────────────────────────────────────

7.2 Setting the Control Time

A
Page 88
70
────────────────────────────────────────────────────
NOTE
The interval setting is in steps of 10 seconds, to a maximum of 100 hours 00
minutes 00 seconds.
Even when operated without the timer or real-time control time set, the timer
operates at 10,000 hours. For this reason, once 10,000 hours have elapsed, pressing the In this case, press the
If the interval timer setting is longer than the time set by the timer or real-
time control function start/stop setting, then the interval setting has no effect.
If the timing at which timer or real-time control ends does not coincide with
the timing of the interval timer, the timing of timer or real-time control takes precedence and the unit stops.
When operating in conjunction with the floppy disk drive, or printer, the
minimum setting of the interval timer changes, according to the amount of data being written. See Chapter 11, "Using the Floppy Disk Drive.",Chapter 13, "Using the Printer (Option)."
7.2.2 Setting the Timer
Timer control provides a single time interval. This can also be used in combination with the interval time, to subdivide the timer time.
START/STOP
SHIFT
key does not operate the unit.
and
START/STOP
keys to reset.
TIME
F1 F2
NOTE
1. Set the timer time in the same way of the interval time setting.
2. Press the
key to return to the measurement
MEAS
screen.
TIMER
3. Press the
START/STOP
stop automatically the control after timer time. During operation, "
4. To stop operation, press the
key to start operation and
" is displayed.
INTEG
START/STOP
again.
411-3.TIF CHANGE
The timer time setting is in steps of 1 minute, to a maximum of 10000
hours.
Even if the real-time control function is enabled, The operation starts by
start time, and it is completed by timer control. In this case, the stop time is ignored.
key
────────────────────────────────────────────────────
7.2 Setting the Control Time
Page 89
71
────────────────────────────────────────────────────
7.2.3 Setting the Real Time Control
Using the real-time control function, the internal real-time clock in the unit can be used to start and stop operation at specified times. This can also be used in combination with the interval time, to subdivide the time specified by real-time control.
1
2
TIME
F1 F2
NOTE
1. Set the start and stop time of the real time control time in the same way of the interval time setting.
REAL TIME CONTROL
2. Press the screen.
3. Press the are displayed. Once the set stop time is reached, operation automatically stops.
4. To stop operation during controlling, press the
START/STOP
MEAS
key to return to the measurement
START/STOP
key again.
key, "
INTEG
"and"
411-4.TIF CHANGE
The time setting is in steps of 1 minute.The time is specified with a four-digit year, and 24-hour times.If a time point has already passed, the real-time control is ignored.If it stops operation during control, the real-time control is forcibly ignored.Even if the real-time control function is enabled, The operation starts by
start time, and it is completed by timer control. In this case, the stop time is ignored.
STIME
3
4
"
5
6
7
8
9
10
11
12
13
14
A
────────────────────────────────────────────────────
7.2 Setting the Control Time
Page 90
72
────────────────────────────────────────────────────
7.3 Integration Screen
On each channel screen, pressing the to integration screen.
Real-time control indication
STIME INTEG
73.tif
F1 F2 F3
INTEG
During integration, this display is shown in yellow.
When integration end or during waiting integration, it is shown in blue.
INTEGRATED
(
F3
Integration operation
indication
) function key moves
Start time Stop time Integration
elapsed time
Integration item
Load factor
STIME
Operates by real time control.
During operation, it is shown in yellow.
────────────────────────────────────────────────────

7.3 Integration Screen

Page 91
73
────────────────────────────────────────────────────
7.4 Starting, Stopping, and Resetting the Integration
1
There are three ways of starting, stopping, resetting integration, as shown below. These controls operate whether or not the integration screen is displayed.
Panel key control
Start
Starts integration by pressing the
Stop
Stops integration by pressing the
integration.
Reset
Resets integration value by pressing the
integration completed.
External control by the terminal
The operation by the panel key can be also controlled by the external control terminal. For details, see Chapter 9, "External Output/ External Control Terminals."
GP-IB/RS-232C control
START/STOP
START/STOP
START/STOP
key.
key during
key after
2
3
4
5
6
7
8
Control is possible in the same way as with the panel keys. When measuring a number of systems, separate control is also possible. Refer to Chapter 12, "GP-IB and RS-232C Interface."
9
10
11
12
13
14
────────────────────────────────────────────────────

7.4 Starting, Stopping, and Resetting the Integration

A
Page 92
74
────────────────────────────────────────────────────

7.5 Manual Integration

Manual integration continues from the time that integration is started until any later point when it is stopped.
Integration value
Start
START /STOP
Integration value
Start
Integration display value
Integration display value
Stop
START
/STOP
Stop
Hold
Hold
(Controlled by Panel Keys)
1. Select the channel and item to be integrated.
Reset
2. If an interval, timer, or real-time control time is set, switch it off. See section 7.2, "Setting the Control Time."
START/STOP
Time
3. Press the integration. "INTEG" (yellow) is indicated.
SHIFT
4. After the required time has elapsed,
START /STOP
press the
START/STOP
integration. "INTEG" (blue) is indicated.
5. To continue with cumulative integration, press the
Start
again. "INTEG" (yellow) is indicated.
6. To reset the integration value, press the
Add
SHIFT
key and then
"INTEG" goes off.
key to start
key to end the
START/STOP
START/STOP
key
key
START /STOP
NOTE
Time
START /STOP
START /STOP
If an interval, timer, or real-time control time is set, manual integration is
not possible.
Because of the internal circuit design, the maximum integration time is
10,000 hours.
If the FDD output or printer output is set, these operate when the
START/STOP
key is pressed. If not necessary, set to
"OFF"
.
While the hold function is activated, the display is frozen, but internally the
integration continues normally. When the floppy disk drive, printer, or D/A outputs are used, however, the held values continue to be output.
────────────────────────────────────────────────────
7.5 Manual Integration (Controlled by Panel Keys)
Page 93
75
────────────────────────────────────────────────────

7.6 Integration Using Time Settings

By first setting the interval, timer, or real-time control time, and then pressing the time.
NOTE
When the interval time is set, data for each interval of the interval time
setting is not displayed on the screen. This must be combined with the floppy disk drive (only 3193) or (optional) printer function.
If the timer (or real-time control) end timing does not coincide with the
interval timer timing, then the integration ends with the timer (or real-time control) end timing, and the last interval timing is ignored.
While the hold function is activated by pressing the
interval time is set, the display is updated every interval time. When the timer time or real control time is set, the final data is displayed at the end of setting time.
7.6.1 Timer Integration
After integrating for the time specified the unit automatically stops. In this case the integration result is held. When the interval time is set, for each interval specified the total integration value at that point is written to the floppy disk or printer.
START/STOP
key, integration can be carried out for the specified
(Controlled by Panel Keys)
HOLD
key, and when the
Integration value
Start
START /STOP
Integration value
Start
Interval time
Timer time
Integration display value
Timer time
Integration display value
Stop
Auto stop
Hold
Stop
Reset
SHIFT
START /STOP
Reset
Hold
Time
Time
START /STOP
────────────────────────────────────────────────────
7.6 Integration Using Time Settings (Controlled by Panel Keys)
Auto stop
Auto stop
SHIFT
START /STOP
Page 94
76
────────────────────────────────────────────────────
7.6.2 Real-Time Control Integration
Integration starts automatically at the start time of the real-time control time, and stops at the stop time. When the interval time is set, for each interval specified the total integration value at that point is written to the floppy disk or printer.
Integration value
Waiting
START /STOP
7.6.3 Interval Integration
The integration calculation is that in the case that only the interval time is set, and is the same as in the case that the timer time is set to 10,000 hours.
Start time
Auto start
Stop time
Integration display value
Auto stop
Hold
Reset
SHIFT
START /STOP
Time
────────────────────────────────────────────────────
7.6 Integration Using Time Settings (Controlled by Panel Keys)
Page 95
77
────────────────────────────────────────────────────
7.7 Measuring the Load Factor
When the timer or real-time control time setting is combined with the interval time, the load factor (LF) can be measured. The load factor result appears in the integration display.
Active power
Start
Interval time
START /STOP
Wav Load factor (LF)=───×100% Wmax Total integration value (+)
Wav = ─────────────── Timer time or real-time control time
Maximum value of integration value for interval Wmax = ────────────────── Interval time
Timer time
Stop time
Hold
Display value
Auto stop
Auto stop
1. Set the interval time, and timer time or
Reset
real time control time. See Section 7.2, "Setting the Control Time."
2. Press the
START/STOP
key to start. During operating, the load factor up to that point is shown.
3. To reset the measured value, press the
Time
SHIFT
START /STOP
Wav : Total averaging Wmax : Ma ximum value of averaging for interval
SHIFT START/STOP
key and then press the
key. ("-----" is displayed.)
NOTE
The load factor is calculated from the positive value integration only.
Therefore, the load factor does not match the normal average. The negative integration value is also taken as zero.
On the screen, the load factor only is displayed.If no interval time is set, the load factor is not displayed.When combined with the floppy disk drive or printer, the maximum average
power (Wmax) and load factor (LF) recorded during operation for each interval are output.
This function can operate simultaneously with integration.Once load factor measurement starts, items for which auto ranging is set
have their ranges fixed from the start time. Set the range so that it will not be exceeded before beginning load factor measurement.
During load factor measurement (even "on standby" under real-time control),
all settings except changing screens are disabled.
After load factor measurement ends, until reset, it is not possible to change
settings.
When combined with a timer setting, after the timer period has elapsed, if
the process is started again the load factor is calculated including the data for the previous timer period.
────────────────────────────────────────────────────

7.7 Measuring the Load Factor

Page 96
78
────────────────────────────────────────────────────
7.8 Zero suppress function
You can set the minimum value of data to be integrated. You can change the value if the input level is too low for the preset range.
F1
F2
NOTE
INTEG-ZERO
1. Press the
STATUS
keys to display the "
2. Using the "
INTEG-ZERO
CURSOR
3. Select from
4. Press the
MEAS
key, then use the
SYSTEM
"page.
PAGE
keys, move the cursor to the
" item.
F1
(0.5%)
,
F2
(0.1%)
,
F3
(0.0%)
key to return to the measurement
( )
.
screen.
F3
In zero suppress, figures less than the percentage set for the full range are
regarded as "0".
The setting is 0.5% by default.When the setting is at 0.0% or 0.1%, the values not displayed as
instantaneous values will also be counted. Counts may be displayed even when there is no input due to internal off-set.
This change function is not transferred when data is saved or loaded via PC
communications or floppy disk. Therefore, settings must be made manually before measurement.
The zero suppress in the display of instantaneous values is not changed.
────────────────────────────────────────────────────

7.8 Zero suppress function

Page 97
79
────────────────────────────────────────────────────
Chapter 8
1
2
8.1 Overview
NOTE
Measuring the motor power (Pm) requires the optional 9603 EXTERNAL
If the load is subject to violent fluctuations or transients, stable measurement
The coupling mode on the channels for which the efficiency is to be
If the calculated efficiency exceeds 100%, it is treated as 100%.
Efficiency Measurement
This unit can calculate the efficiency from the measured values (active power, motor power). For example, the input/output efficiency of an inverter, input/output efficiency of a motor, and overall efficiency can be calculated simultaneously with a single unit.
SIGNAL INPUT UNIT.
may not be possible.
measured should be the same. (DC/AC + DC/AC)
3
4
5
6
7
8
9
10
11
12
13
14
15
────────────────────────────────────────────────────

8.1 Overview

Page 98
80
────────────────────────────────────────────────────
8.2 Efficiency Screen
In the MEAS screen, use the
key to move the cursor to "EFFI"
PAGE
(efficiency) to display the efficiency screen. The calculation formula can be set in the "EFFI" display of the STATUS screen.
82-1.tif
────────────────────────────────────────────────────

8.2 Efficiency Screen

Page 99
81
────────────────────────────────────────────────────
8.3 Setting the Calculation Formula
1
F1 F2
NOTE
A maximum of three formulas can be set.
1. Press the display the "EFFI" page.
2. Next, use the the denominator or numerator. The items which can be substituted in the formula appear in the lower part of the screen. Use function keys ()and
The same setting is made forη2andη3.
83-1.tif
If the measurement value is not inserted in either the numerator or denominator, this is shown as
η
STATUS
F2
= "--".
key, then use the
CURSOR
() to select the item.
keys to move the cursor to
PAGE
key to
F1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
────────────────────────────────────────────────────

8.3 Setting the Calculation Formula

Page 100
82
────────────────────────────────────────────────────
8.4 Example Measurement
The following is an example of measuring the efficiency.
NOTE
In either the 3P3W or 3V3A connection mode of the 3193, the active power (P)ofa3φ3W system is found by the two-power calculation method, and the efficiency calculation result is also the same.
8.4.1 Efficiency Measurement of a Switching Power Supply (1φ2W)
Input side (1φ2W) DC/50/60 Hz
P
1
Switching
power supply
When channel 1 is the input and channel 2 the output Input: 1P2W, AC mode / output: 1P2W, DC or AC+DC mode
η
P
1= ─── ×100
2
P
1
Measuring the efficiency of a single-phase power converter (switching power supply etc.)
Output side (DC)
P
2
8.4.2 Efficiency Measurement of a Switching Power Supply (3φ3W)
Input side (3φ3W) 50/60 Hz
P
12
Switching
power supply
When channel 1, 2 are the input and channel 3 the output Input: 3P3W, AC mode / output: 1P2W, DC or AC+DC mode
η
P
3
1= ── ×100
P
12
Measuring the efficiency of a three-phase power converter (switching power supply etc.)
Output side (DC)
P
3
────────────────────────────────────────────────────

8.4 Example Measurement

Loading...