Hioki 3173 Instruction Manual

Page 1
3173
Instruction Manual
PORTABLE WITHSTANDING VOLTAGE HiTESTER
Mar. 2019 Revised edition 4 3173A981-04 19-03H
EN
Page 2
Page 3

Contents

Introduction i Inspection Safety Notes Notes on Use
iv
Chapter Summary
Chapter 1 Overview 1
1.1 Product Overview 1
1.2 Names and Functions of Parts
1.2.1 Front panel 2
1.2.2 Rear panel 4
1.2.3 9615 H.V. TEST LEAD 5
1.3 External Dimensions 6
Chapter 2 Testing Arrangements 7
2.1 Installation of the Unit 7
i
ii
v
2
2.2 Connecting the Protective Ground Terminal
2.3 Connecting the External I/O Terminal
2.4 Power Cord Connection
2.5 Connecting the 9615 H.V. TEST LEAD
2.6 Connection to the Measured Equipment
2.7 Powering On and Off the 3173
2.8 Startup Inspection
10 11 12 13 14
Chapter 3 Withstand Voltage Testing Method 15
3.1 Starting a Test 16
3.1.1 Setting the Current Threshold Value (Upper Limit Value) 17
3.1.2 Setting the Timer (Test Time) 18
3.1.3 Setting the Output-voltage 19
3.1.4 Executing a Test 20
3.2 PASS or FAIL Determination 21
8 9
3173A981-
3.2.1 "PASS" State 21
3.2.2 "FAIL" State 22
04
Page 4
Chapter 4 External Interface 23
4.1 External I/O Terminal 23
4.1.1 Signal Line 24
4.1.2 Timing Chart of External I/O Terminal 26
4.2 Example of External I/O Terminal 28
4.2.1 Example of Input Signal Connection 28
4.2.2 Example of Output Signal Connection 30
Chapter 5 Specifications 33
5.1 Basic Specifications 33
5.2 General Specifications
5.3 9615 H .V. TEST LEAD Specifications
35 36
Chapter 6 Maintenance and Servise 37
6.1 Maintenance and Inspection 37
6.2 Cleaning
6.3 Fuse Replacement
38 39
Page 5
_____________________________________________________________________________________________
i

Introduction

Thank you for purchasing the HIOKI 3173 PORTABLE WITHSTANDING VOLTAGE HiTESTER. To obtain maximum performance from the product, please read this manual first, and keep it handy for future reference.
Warranty
HIOKI cannot be responsible for losses caused either directly or indirectly by the use of the 3173 with other equipment, or if ownership is transferred to a third party.

Inspection

WARNING
When you receive the product, inspect it carefully to ensure that no damage occurred during shipping. In particular, check the connectors and accessories. If damage is evident, or if it fails to operate according to the specifications, contact your dealer or Hioki representative.
Accessories
9615 H.V.TEST LEAD (HIGH/LOW) 1 each Power cord 1 Spare fuse 1 Instruction Manual 1
Preliminary Checks
Before using the product the first time, verify that it operates normally to ensure that the no damage occurred during storage or shipping. If you find any damage, contact your dealer or Hioki representative. Use the original packing materials when reshipping the product, if possible.
Before using the product, make sure that the insulation on the test leads is undamaged and that no bare conductors are improperly exposed. Using the product under such conditions could result in electrocution. Replace the test leads specified by Hioki .
______________________________________________________________________________________________
Page 6
ii
_____________________________________________________________________________________________

Safety Notes

This product is designed to conform to IEC 61010 Safety Standards, and
WARNING
Safety Symbols
has been thoroughly tested for safety prior to shipment. However, mishandling during use could result in injury or death, as well as damage to the product. Be certain that you understand the instructions and precautions in the manual before use. We disclaim any responsibility for accidents or injuries not resulting directly from product defects.
This manual contains information and warnings essential for safe operation of the product and for maintaining it in safe operating condition. Before using the product, be sure to carefully read the following safety notes.
The symbol printed on the product indicates that the user should
refer to a corresponding topic in the manual (marked with the symbol) before using the relevant function.
In the manual, the symbol indicates particularly important
information that the user should read before using the product. Indicates that dangerous voltage may be present at this terminal. Indicates AC (Alternating Current). Indicates a fuse. Indicates a protective conductor terminal. Indicates the ON side of the power switch. Indicates the OFF side of the power switch.
The following symbols in this manual indicate the relative importance of cautions and warnings.
Indicates that incorrect operation presents an extreme hazard that
DANGE R
WARNING
CAUTION
could result in serious injury or death to the user.
Indicates that incorrect operation presents a significant hazard that could result in serious injury or death to the user.
Indicates that incorrect operation presents a possibility of injury to the user or damage to the product.
NOTE
______________________________________________________________________________________________
Advisory items related to performance or correct operation of the product.
Page 7
_____________________________________________________________________________________________
iii
Measurement categories
To ensure safe operation of measuring instruments, IEC 61010 establishes safety standards for various electrical environments, categorized as CAT II to CAT IV, and called measurement categories.
Using a measuring instrument in an environment designated with a
WARNING
higher-numbered category than that for which the instrumen is rated could result in a severe accident, and must be carefully avoided. Never use a measuring instrument that lacks category labeling in a CAT II to CAT IV measurement environment. Doing so could result in a serious accident.
CAT II
CAT III
CAT IV
When directly measuring the electrical outlet receptacles of the primary electrical circuits in equipment connected to an AC electrical outlet by a power cord (portable tools, household appliances, etc.)
When measuring the primary electrical circuits of heavy equipment (fixed installations) connected directly to the distribution panel, and feeders from the distribution panel to outlets
When measuring the circuit from the service drop to the service entrance, and to the power meter and primary overcurrent protection device (distribution panel)
Accuracy
We define measurement tolerances in terms of f.s. (full scale) and rdg. (reading) values, with the following meanings:
f.s.
rdg.
______________________________________________________________________________________________
(maximum display value or scale length) The maximum displayable value or the full length of the scale. This is usually the maximum value of the currently selected range.
(reading or displayed value) The value currently being measured and indicated on the measuring product.
Page 8
iv
p
_____________________________________________________________________________________________

Notes on Use

Follow these precautions to ensure safe operation and to obtain the full benefits of the various functions.
Observe the following precautions to avoid electric shock.
DANGE R
Do not remove the product's case. The internal components of the product carry high voltages and may become very hot during operation. Do not touch the alligator clip while high voltage is being generated. The vinyl sheath on the alligator clip of the 9615 H.V.TEST LEAD does not have an insulation withstand voltage.
WARNING
CAUTION
To avoid electric shock, do not allow the product to get wet, and do not use it when your hands are wet. To avoid electric shock, connect the protective ground terminal to a grounded conductor before making any other connections. Before turning the product on, make sure the source voltage matches that indicated on the product's power connector. Connection to an improper supply voltage may damage the product and present an electrical hazard. Replace the fuse only with one of the specified characteristics and voltage and current ratings. Using a non-specified fuse or shorting the fuse holder may cause a life-threatening hazard. Fuse type: 250VT1AL (
3173, 3173-01),
250VT0.5A L (
3173-02, 3173-03, 3173-04)
To avoid electrocution, turn off the power to all devices before pluggingor unplugging any of the interface connectors.
To avoid damaging the power cord, grasp the plug, not the cord, when unplugging the cord from the power outlet. To avoid damaging the test leads, do not bend or pull the 9615 H.V. TEST LEADs. To avoid damage to the product, protect it from vibration or shock during transport and handling, and be especially careful to avoid dropping. Do not use the product near a device that generates a strong electromagnetic field or electrostatic charge, as these may cause erroneous measurements. To avoid electric shock, do not exceed the lower of the ratings shown on the instrument and test leads.
NOTE
______________________________________________________________________________________________
This product may cause interference if used in residential areas. Such use must be avoided unless the user takes special measures to reduce electromagnetic emissions
revent interference to the reception of radio and television broadcasts.
to
Page 9
_____________________________________________________________________________________________
v

Chapter Summary

Chapter 1 Overview
Describes an overview, features, and the names and functions of the parts of the product.
Chapter 2 Testing Arrangements
Describes particulars of testing arrangements.
Chapter 3 Testing Method
Describes procedures for setting, testing, and test results judgment.
Chapter 4 External Interface
Describes use of the external I/O terminal.
Chapter 5 Specifications
Describes the product specifications.
Chapter 6 Maintenance and Servise
Describes the maintenance and inspection, fuse replacement, and ultimate disposal.
______________________________________________________________________________________________
Page 10
vi
_____________________________________________________________________________________________
______________________________________________________________________________________________
Page 11
_____________________________________________________________________________________________
1
Chapter 1
Overview
1.1 Product Overview
(1) Simple design with a streamlined set of core functions.
Simple, clean design with a streamlined set of essential functions - highly portable and perfectly suited to inspections on the production line.
(2) Test timer
The test timer can set test times between 1 and 99 seconds in increments of 1 second.
(3) Analog voltmeter
The test voltage is monitored using the analog voltmeter.
(4) External I/O
The tester has external I/O terminals on its rear panel, which are used to output various status signals and to input test start and stop signals.
______________________________________________________________________________________________

1.1 Product Overview

Page 12
2
_____________________________________________________________________________________________
1.2 Names and Functions of Parts
1.2.1 Front panel
To avoid electric shock, never touch the output-voltage terminals (HIGH
WARNING
and LOW), high-voltage test leads or the device being tested while high voltage is being applied.
Analog voltmeter
Test timer LED Current threshold
setting dial Test timer
setting dial Test timer switch
DANGER lamp
HIGH terminal
LOW terminal
START key
STOP key
Output-voltage adjustment knob
Power LED
Power switch
Test result LED
______________________________________________________________________________________________

1.2 Names and Functions of Parts

Page 13
_____________________________________________________________________________________________
3
Analog voltmeter Test timer LED
Current threshold setting dial
Test timer setting dial Test timer switch DANGER lamp
HIGH terminal
LOW terminal
Output-voltage adjustment knob
Test result LED
START key STOP key
Indicates output-voltage when testing withstand voltage. The LED lights up during testing when the timer is set to
ON. Sets the current threshold.
(0.1 to 9.9 mA, resolution of 0.1 mA) Sets the timer (test time). (1 to 99 s, resolution of 1 s) Turns ON/OFF the test timer. This lamp lights to warn that voltage is present between the
terminals during testing. The HIGH terminal is a high-voltage terminal for voltage
outputs. The LOW terminal is a low-voltage terminal for voltage
outputs. It has the same electric potential as the unit body. Sets the output-voltage.
One of the two LEDs will light, depending on the test result. Used to start a test. Normally used to terminate a test.
Power switch
Powers the 3173 on or off. The LED lights up when power is turned ON.
______________________________________________________________________________________________
1.2 Names and Functions of Parts
Page 14
4
_____________________________________________________________________________________________
1.2.2 Rear panel
To avoid electric shock, never touch the output-voltage terminals (HIGH
WARNING
and LOW), high-voltage test leads or the device being tested while high voltage is being applied.
External I/O terminal
Power inlet
Fuse holder
Fuse holder External I/O terminal Power inlet
Protective ground terminal
Protective ground terminal
Contains a power fuse. Each signal is output by ON/OFF of the relay depending on the test
result. Terminals for start and stop input signals are also available. Connect the grounded three-core power cord supplied here. Used to earth a protective ground wire. Be sure to make grounding
connections before starting a test.
______________________________________________________________________________________________
1.2 Names and Functions of Parts
Page 15
_____________________________________________________________________________________________
5
1.2.3 9615 H.V. TEST LEAD
To avoid electric shock, do not touch the alligator clip while high
DANGE R
voltage is being generated. The vinyl sheath on the alligator clip of the 9615 H.V.TEST LEAD does not have an insulation withstand voltage.
1
High-voltage side (red)
2
1 3
1 Alligator clip 2 High-voltage output
plug
3 Low-voltage output
plug
Low-voltage side (black)
Attach to the test point on the device being measured. Connect to the HIGH terminal on the unit.
Connect to the LOW terminal on the unit.
______________________________________________________________________________________________
1.2 Names and Functions of Parts
Page 16
6
_____________________________________________________________________________________________
1.3 External Dimensions
200 2mm
7.87 0.08"
149 2mm
0.08"
5.87
215 2mm
8.46 0.08"
______________________________________________________________________________________________

1.3 External Dimensions

Page 17
_____________________________________________________________________________________________
7
Chapter 2
Testing Arrangements
2.1 Installation of the Unit
Temperature: 0 to 40 (32 to 104 ) Humidity: 80% RH or less (no condensation) Do not place the tester upside down. Be sure to place the tester with its bottom side down.
Avoid the following locations:
Subject to direct sunlight.Having poor air circulation.
Subject to high levels of dust, steam, or corrosive gases (Avoid
using the equipment in an environment containing corrosive gases (e.g., H2S, SO2,NI2,andCI2) or substances that generate harmful gasses (e.g., organic silicones, cyanides, and formalins)).
Subject to vibrations.In the vicinity of equipment generating strong electromagnetic
fields.
NOTE
______________________________________________________________________________________________
The noise generated by this unit may affect equipment located around the unit.

2.1 Installation of the Unit

Page 18
8
_____________________________________________________________________________________________
2.2 Connecting the Protective Ground Terminal
To avoid electric shock, be sure to connect the protective ground
WARNING
terminal to a grounded conductor. To avoid electric shock, connect the protective ground terminal to a grounded conductor before making any other connections.
1
. Using a Phillips-head screwdriver, remove the screw of the protective
ground terminal from the rear of the unit.
2
. Connect an electric wire with a sufficient current capacity to the
protective ground terminal, and secure the wire using a Phillips-head screwdriver.
3
. Ground the other end of the wire.
______________________________________________________________________________________________

2.2 Connecting the Protective Ground Terminal

Page 19
_____________________________________________________________________________________________
9
2.3 Connecting the External I/O Terminal
Connect the external I/O terminal before turning on the power. If the external I/O terminal is installed or removed following startup, malfunction may result.
1
. Press and hold the button on the terminal block using a flathead
screwdriver.
2
. While holding the button, insert a signal cable with insulation removed
from the end into the hole.
3
. Release the button to secure the cable to the terminal.
NOTE
For the specifications of the external I/O terminal, see Section 4.1.
______________________________________________________________________________________________

2.3 Connecting the External I/O Terminal

Page 20
10
_____________________________________________________________________________________________
2.4 Power Cord Connection
Before turning the product on, make sure the source voltage matches
WARNING
that indicated on the product's power connector. Connection to an improper supply voltage may damage the product and present an electrical hazard.
1
. Be sure that the power switch is turned to OFF.
2
. Connect the grounded three-core power cord provided to the power inlet
on the back of the unit.
3
. Insert the plug into the grounded outlet.
Supply voltage indicated on the rear panel
______________________________________________________________________________________________

2.4 Power Cord Connection

Page 21
_____________________________________________________________________________________________
11
2.5 Connecting the 9615 H.V. TEST LEAD
Observe the following precautions to avoid electric shock.
WARNING
Turn off the power unit and tested object, make sure that there is no high voltage being applied to the output, confirm the following items, and connect the 9615 H.V. TEST LEAD. (1) The analog voltmeter reads 0 kV. (2) The Make sure the 9615 H.V. TEST LEAD is securely connected before starting a test, as a loose test lead can cause a hazard when a voltage is output. Before using the product, make sure that the insulation on the test leads is undamaged and that no bare conductors are improperly exposed. Using the product under such conditions could result in electrocution. Replace the test leads specified by Hioki .
1
. Remove the LOW terminal by turning it counterclockwise.
DANGER
lamp is OFF.
LOW terminal
2
. As shown in the figure, insert the plug on the H.V. TEST LEAD (low
voltage side).
3
. Secure the LOW terminal by turning it clockwise.
4
. Connect the plug on the H.V. TEST LEAD (high voltage side) to the
HIGH terminal.
Plug on the H.V. TEST LEAD (low voltage side)
Plug on the H.V. TEST LEAD (high voltage side)
______________________________________________________________________________________________

2.5 Connecting the 9615 H.V. TEST LEAD

Page 22
12
g
_____________________________________________________________________________________________
2.6 Connection to the Measured Equipment
Observe the following precautions to avoid electric shock.
WARNING
Turn off the power unit and tested object, make sure that there is no high voltage being applied to the output, confirm the following items, and connect the 9615 H.V. TEST LEAD. (1) The analog voltmeter reads 0 kV. (2) The Make sure that no high voltage is being applied between the output­voltage terminals (HIGH and LOW), before touching the output-voltage terminals, H.V. TEST LEAD, or tested object. Even following a test, there may be a residual voltage at the output-voltage terminals.
1
. Make sure the analog voltmeter is at 0 kV and the
2
. Connect the LOW terminal probe to the tested object. Fasten the probe
DANGER
lamp is OFF.
DANGER
lamp is OFF.
securely to prevent it from loosening during a test.
NOTE
3
. Following the procedure described above, connect the HIGH terminal
probe to the tested object.
If the test lead touches to other metallic parts, it may cause an increase in leakage current. Prevent the test lead from coming into contact with other parts as muc h as possible. Do not the test lead and the EXT/IO cable arranged closely to each other. Doin
so maylead to a malfunction of the external control due to a noise.
______________________________________________________________________________________________
2.6 Connection to the Measured Equipment
Page 23
_____________________________________________________________________________________________
13
2.7 Powering On and Off the 3173
Before turning on the power, make sure that the voltage of the power
WARNING
supply being used matches the supply voltage indicated on the rear panel of the unit. If an attempt is made to use an improper supply voltage, there is danger of damage to this unit and of life-threatening risk to the operator. Apply a voltage within the acceptable power­voltage range. Otherwise, damage to the unit or electrical accidents may result.
NOTE
Allow 10 minutes warming up after powering on.
Output voltage adjustment knob
Power switch
(1) Powering on the unit
1
. Make sure that the voltage adjusting knob is turned counter-clockwise as
far as possible.
2
. Turn the power switch to ON ( ).
(2) Powering off the unit
1
. Following a test, make sure the analog voltmeter is at 0 kV and the
DANGER
2
. Turn the output voltage adjustment knob counterclockwise until the
lamp is OFF.
output reaches 0 kV.
3
. Turn the power switch to OFF ( ).
______________________________________________________________________________________________

2.7 Powering On and Off the 3173

Page 24
14
_____________________________________________________________________________________________
2.8 Startup Inspection
To ensure safe testing, check the following before starting operation:
(1) Breaking current
1
. Calculate the resistance based on the output-voltage set for withstand
voltage testing and the upper test value, and prepare a resistor corresponding to the calculation result. Note the power rating of the resistor. (Output voltage / test upper limit (interrupting current) = resistance) (The 2nd power of output voltage / resistance = power consumption < powerratingofresistor) A high-voltage resistor is recommended. Example high-voltage resistor:
KOA Corp.'s GS Series High Voltage High Resistance Thick Film Resistor
2
. Set an upper-limit value.
3
. Connect the resistor to the H.V. TEST LEAD.
4
. Increase the output voltage beyond the set value, and make sure the
current is cut off (i.e., make sure the unit is in the FAIL state).
(2) Analog voltmeter
1
. Before turning on the power, make sure the analog voltmeter is at 0 kV.
2
. If the voltage reading is not at zero, adjust the value to zero using a
slotted screwdriver.
(3) Key inspection
1
. Turn off the power, and unplug the power cord from the power outlet.
2
. For both the START and STOP keys on the front panel of the unit, press
the center of the key, and make sure you feel it click. The click is less noticeable when the edges of the keys are pressed.
*Clicking
When a key is pressed slowly, there is a moment of slight resistance and a feeling that the key cannot be pressed any further. When the key is pressed further after this point, a clicking sensation can be felt.
3
. If you do not feel a click, the key may be broken.
______________________________________________________________________________________________

2.8 Startup Inspection

Page 25
_____________________________________________________________________________________________
,
15
Chapter 3
Withstand Voltage
Testing Method
This chapter describes how to set withstand voltage test conditions and the proper testing procedure. Read Chapter 2, and make the necessary preparations for testing.
NOTE
Note that the output waveform may be distorted when conducting an AC withstand voltage test for a voltage-dependent device or object (e.g., ceramic capacitor). Excessively large distortion may damage the device or tested object. When the device or tested object is a capacitive load, resonance may occur (when conducting an AC withstand voltage test) with the coil inductance of the output high-voltage transformer, depending on the capacitance value. This condition may generate several tens of volts at the voltage output terminal before the test is started. If the START key is disabled, reset the output voltage adjustment knob to zero
then set it back to the desired voltage after startingthe test.
______________________________________________________________________________________________
Page 26
16
_____________________________________________________________________________________________
3.1 Starting a Test
The flowchart below explains how a test is carried out.
Setting the test parameters
READY状態
See Section 3.1.1 "Setting the Current Threshold Value (Upper Limit Value)" and Section 3.1.2 "Setting the Timer (Test Time)"
Set the following before starting the test:
Current threshold value (upper limit value) setting
Timer (test time) setting
Setting the output-voltage
READY状態
See Section 3.1.3 "Setting the Output-voltage" Before starting a tes t, make output voltage settings using the output-voltage knob.
Never touch the output-voltage terminals (HIGH and LOW), H.V. TEST LEAD, or tested object. A voltage is output during output-voltage setting using the output-voltage adjustment knob.
Starting a test
READY状態
See Section 3.1.4 "Executing a Test" When you press the START key, the DANGER lamp lights up and the test begins.
If the test timer is ON, the test timer LED starts blinking.
Determination
READY状態
See Section 3.2 "PASS or FAIL Determination" Each test returns a PASS or FAIL judgment by comparing the currents measured during the set
test time with the set threshold. When a measurement equals or exceeds the threshold, the test returns a FAIL judgment. A PASS judgment is not returned if the test timer is OFF.
(Allow 10 minutes warming up after powering on.)
______________________________________________________________________________________________

3.1 Starting a Test

Page 27
_____________________________________________________________________________________________
g
17
3.1.1 Setting the Current Threshold Value (Upper Limit Value)
NOTE
. Make sure the analog voltmeter is at 0 kV and the
1
. Turn the current threshold setting dial to set the threshold to the desired
2
DANGER
lamp is OFF.
level.
The current thresold value (upper limit value) can be set from 0.1 to 9.9mA (setting resolution: 0.1mA). To prevent malfunctions, during testing avoid changing the test current threshold settin
.
______________________________________________________________________________________________
3.1 Starting a Test
Page 28
18
ging
_____________________________________________________________________________________________
3.1.2 Setting the Timer (Test Time)
NOTE
. Make sure the analog voltmeter is at 0 kV and the
1
. Turn the test timer setting dial to set the timer (test time).
2
. Turn the test timer switch ON.
3
The timer (test t ime) can be set from 1 to 99 s (setting resolution: 1 s). If the test time doesn't need to be set, turn the test timer switch OFF. This will disable test timer setting. To prevent malfunctions, during testing avoid turning the test timer switch ON/OFF or chan
timer settings.
DANGER
lamp is OFF.
______________________________________________________________________________________________
3.1 Starting a Test
Page 29
_____________________________________________________________________________________________
gadj
19
3.1.3 Setting the Output-voltage
Before starting a test, make output-voltage settings using the output-voltage adjustment knob.
Observe the following precautions to avoid electric shock.
DANGE R
Make sure that no high voltage is being applied to the output, confirm the following items, and connect the H.V. TEST LEAD. (1) The analog voltmeter reads 0 kV. (2) The Never touch the output-voltage terminals (HIGH and LOW), H.V. TEST LEAD, or tested object. A voltage is output during output-voltage setting using the output-voltage adjustment knob. Do not touch the alligator clip while high voltage is being generated. The vinyl sheath on the alligator clip of the 9615 H.V.TEST LEAD does not have an insulation withstand voltage.
DANGER
lamp is OFF.
NOTE
. Make sure the analog voltmeter is at 0 kV and the
1
. Connect the high voltage test leads to the device to be tested.
2
DANGER
lamp is OFF.
(See Section 2.6 "Connection to the Measured Equipment")
. Press the
3
START
key.
DANGER
lamp will light up and a voltage will be
output.
. Set an output-voltage using the output-voltage adjustment knob.
4
The output-voltage knob increases the output-voltage when turned clockwise, and
decreases the voltage when turned counterclockwise. (The maximum value: 3.0 kV)
The output-voltageisdisplayed on the analogvoltmeter.
. Press the
5
STOP
key to stop the output.
Output-voltage adjustment knob
NOTE
______________________________________________________________________________________________
When adjusting the output voltage, make sure that the test timer is OFF and that the current threshold is set sufficiently high. Since PASS/FAIL judgments are made even durin
ustment,the output maybe halted byaPASSorFAILjudgment.
3.1 Starting a Test
Page 30
20
_____________________________________________________________________________________________
3.1.4 Executing a Test
Observe the following precautions to avoid electric shock.
DANGE R
Make sure that no high voltage is being applied to the output, confirm the following items, and output-voltage. (1) The analog voltmeter reads 0 kV. (2) The
DANGER
lamp is OFF. Never touch the output-voltage terminals (HIGH and LOW), H.V. TEST LEAD, or tested object. A voltage is output during output-voltage setting using the output-voltage adjustment knob. Avoid testing grounded devices. If you must test a grounded device, be sure to connect the alligator clip of the LOW (black) 9615 H.V. TEST LEAD to the grounded side of the device. Connecting the HIGH (red) test lead to the grounded side may lead to electric shock or damage to the tester.
CAUTION
NOTE
For output, the unit uses a high-voltage transformer that boosts the power voltage. If an unstable power voltage is used to operate the unit, the tested object may be damaged by the distortion of the output-voltage waveform and by the output of a voltage higher than the preset voltage. If a capacity load is applied to the tested object, the output-voltage may exceed the preset voltage, thereby damaging the equipment. Continuous output of a high voltage may heat the bottom of the unit. Take special care when handling the unit (e.g. transporting the unit).
The
START
signal and
key and
――――――――
STOP
STOP
key are connected with the
external input signal, respectively, as OR gates. The signal
―――――――――
START
external input
works in the same way as the key. Check the level of the external input signals if the test fails to start or stop properly. Allow 10 minutes warmingupafterpoweringon.
Pressing the
START
key starts testing. When a voltage is output, the
DANGER
lamp goes on. To interrupt testing, press the
STOP
key. The tester immediately halts voltage
output an d does not return a PASS/FAIL judgment.
______________________________________________________________________________________________
3.1 Starting a Test
Page 31
_____________________________________________________________________________________________
y
21
3.2 PASS or FAIL Determination
3.2.1 "PASS" State
Confirm the following items, before touching the output-voltage
WARNING
terminals (HIGH and LOW), H.V. TEST LEAD, or tested object. Even when a test has been terminated, there may still be voltage in the output-voltage terminals. (1) The analog voltmeter reads 0 kV. (2) The
DANGER
lamp is OFF.
When the preset test time has elapsed, the unit switches to the PASS state and immediately stops outputting a voltage. If the test time has not been set, PASS screening is not performed. To the test, press the
STOP
key, which will forcibly terminate the test.
Flow of PASS determination
Output-voltage
Comparative voltage value
1. Press the The output is synchronized with the supply voltage waveform. The tester will not output a voltage until the supply voltage waveform becomes 0 V after the Since the output is not stabilized, the output voltage will fluctuate with the load current. Set the output voltage with the tester connected to the device to be tested.
2. The tester continues to output voltage until the set test time has elapsed. If a current measurement equals or exceeds the set threshold during the test time, the tester enters the "FAIL state" and halts output.
3. After the test time has elapsed, the tester enters the "PASS state" and halts voltage output. In the "PASS state," the green LED and the PASS relay for the external I/O output are both ON.
4. After the "PASS state (which lasts for about 1 second)", the tester returns to the ready state.
START
Ready
START
START
The current measurements sta below the set threshold.
Testing time when it set up
Zero-cross waiting time
Test time elapsed
Test
DANGER lamp
key to begin testing.
key is pressed.
PASS*
*0.45 seconds.
Time
Ready
______________________________________________________________________________________________

3.2 PASS or FAIL Determination

Page 32
22
_____________________________________________________________________________________________
3.2.2 "FAIL" State
Confirm the following items, before touching the output-voltage
WARNING
Output-voltage
terminals (HIGH and LOW), H.V. TEST LEAD, or tested object. Even when a test has been terminated, there may still be voltage in the output-voltage terminals. (1) The analog voltmeter reads 0 kV. (2) The
DANGER
lamp is OFF.
If a current measurement equals or exceeds the tests upper limits during the test, the tester enters the "FAIL state." The unit switches to the FAIL state and immediately stops outputting a voltage.
Flow of FAIL determination
The current measurement exceeds the threshold value.
Comparative voltage value
Testing time when it set up
Zero-cross waiting time
START
Ready
1. Press the
START
Test
DANGER lamp
key to begin testing.
STOP input
FAIL
The output is synchronized with the supply voltage waveform. The tester will not output a voltage until the supply voltage waveform becomes 0 V after the
START
key is pressed. Since the output is not stabilized, the output voltage will fluctuate with the load current. Set the output voltage with the tester connected to the device to be tested.
2. The tester continues to output voltage until the set test time has elapsed. If a current measurement equals or exceeds the set threshold during the test time, the tester enters the "FAIL state" and halts output.
3. The buzzer sounds and the red LED lights. The FAIL relay for the external I/O output is turned ON. The tester remains in the "FAIL state" until the next STOP input.
Time
Ready
______________________________________________________________________________________________
3.2 PASS or FAIL Determination
Page 33
_____________________________________________________________________________________________
g
23
Chapter 4
External Interface
4.1 External I/O Terminal
NOTE
This tester has five external control signal terminals on its rear panel: PASS and FAIL (relay contact output),
―――――――――
up),
START
The
START
――――――――
and
STOP
to external si
key and
external input signal, respectively, as OR gates. Key input is equivalent
――――――――
and
STOP
STOP
nal input.
(input, 5 V pull-up).
key are connected with the
―――――――――
START
and
―――――――
TEST
――――――――
STOP
(open collector output without pull-
―――――――――
START
signals are internallypulled upto 5V.
external input signal
______________________________________________________________________________________________

4.1 External I/O Terminal

Page 34
24
(
)
_____________________________________________________________________________________________
4.1.1 Signal Line
Input Signals
NOTE
―――――――――
(1)
START
(Wired "OR" connection with
START
key: Low active)
Measurement begins when this signal becomes low at the specified timing. (This is equivalent to pressing the START key.) This terminal is connected with START key input as a wired OR gate. When the
START
key is pressed,
the input signal becomes low. It is internally pulled up to 5V.
――――――――
(2)
STOP
(Wired "OR" connection with
STOP
key: Low active)
Measurement is halted when this signal becomes low at the specified timing. (This is equivalent to pressing the with
STOP
key input as a wired OR gate. When the
STOP
key.) This terminal is connected
STOP
key is pressed, the
input signal becomes low. It is internally pulled up to 5V.
When the
STARTorSTOP
key is pressed, the
―――――――――
START becomes low respectively. When using a gate IC, connect it via a transistor, or take other similar measures. If an IC is connected directly to the terminals, the external IC may be short-circuited with GND, damaging the IC.
See "4.2.1 Example of Input Signal Connection."
or
――――――――
STOP
input signal
______________________________________________________________________________________________
4.1 External I/O Terminal
Page 35
_____________________________________________________________________________________________
25
Output Signals
(1) PASS (Relay contact output)
This signal is turned ON when the tester returns a PASS judgment.
(2) FAIL (Relay contact output)
This signal is turned ON when the tester returns a FAIL judgment.
――――――――
(3)
TEST
Indicates that the tester is outputting a high voltage. This signal stays low while the high voltage is output. A pull-up resistor is not connected internally.
(Open collector output: Low active)
NOTE
When using
―――――――――
START
and
――――――――
STOP
relay contacts, check the power rating of the
relays. Remain alert for chattering caused by relay ON/OFF.
______________________________________________________________________________________________
4.1 External I/O Terminal
Page 36
26
pOp
_____________________________________________________________________________________________
4.1.2 Timing Chart of External I/O Terminal
(1) Timing chart at time of start of testing
―――――――
START
―――――――
STOP
Test timer (Timer is ON)
Voltage output
―――――――
TEST
input
input
output
The figure below shows a timing chart after a test is begun. If the
START
simultaneously, the
―――――――――
(
START
signal).
Sto
―――――――――
key (
START
STOP
100 ms or more
signal) and
――――――――
key (
STOP
20 ms or more
5 ms or less
25 ms or less
20 ms or less
STOP
signal) overrides the
―――――――――
START
key (
eration
signal is input and a
――――――――
STOP
signal) are pressed
START
key
(2) Timing chart at time of termination of testing
―――――――
START
―――――――
STOP
Voltage output
―――――――
TEST
input
input
output
The figure below shows a timing chart after a test is sto. If the
START
simultaneously, the
―――――――――
(
START
signal).
key (
―――――――――
START
STOP
30 ms or more
signal) and
――――――――
key (
STOP
100 ms or more
STOP
signal) overrides the
30 ms or less
25 ms or less
――――――――
STOP
key (
signal is input and a
――――――――
STOP
signal) are pressed
START
key
______________________________________________________________________________________________
4.1 External I/O Terminal
Page 37
_____________________________________________________________________________________________
p
27
(3) PASS Judgment Timing Chart
The figure below shows a timing chart for when the set test time has elapsed and the test returns a PASS judgment.
End of test time
Test timer
PASS relay
Voltage output
―――――――
TEST
output
(4) Test Judgment Timing Chart
Test time
FAIL relay
Operation Sto
5 ms or less
OFF OFFON
20 ms or less
15 ms or less
0.45 s approx.
The figure below shows a timing chart if a current measurement exceeds the threshold during test and the test returns a FAIL judgment.
FAIL
Test is under way.
5 ms or less
OFF
ON (Remains ON until the next
Test is complete.
――――――
STOP
input.)
Voltage output
―――――――
TEST
output
20 ms or less
15 ms or less
______________________________________________________________________________________________
4.1 External I/O Terminal
Page 38
28
_____________________________________________________________________________________________
4.2 Example of External I/O Terminal
4.2.1 Example of Input Signal Connection
The unit can be controlled externally using the external I/O input signal.
EXT I/O Input Signals Specifications
Input signals Active low input (not isolated) Maximum input
voltage HIGH level voltage H LOW level voltage H Signal names
5.5 V DC
: 5.5 V DC, H
max
: 1.0 V DC, H
max
―――――――――
START
――――――――
STOP
: Measurement start : Measurement stop
: 4.5 V DC
min
: -0.5 V DC
min
Each signal is internally pulled up to 5 V with a 560 Ω resistor.
―――――――――
START
――――――――
STOP
input terminal
COM terminal
input terminal
COM terminal
―――――――――
START
5V
560 Ω
56 kΩ
START key
Inside of 3173
input terminal
5V
560 Ω
200 kΩ
STOP key
To the internal circuit
0.1 μF
To the internal circuit
0.1 μF
Inside of 3173
――――――――
STOP
input terminal
______________________________________________________________________________________________

4.2 Example of External I/O Terminal

Page 39
_____________________________________________________________________________________________
29
(1) To Control with External Switch
When controlling
―――――――――
START
and
――――――――
STOP
signals using a relay or switch, connect them as shown below. To prevent switch or relay chattering, we recommend using the connection shown in Figure b. For this connection, implement the circuit so that a current of 10 mA flows to the transistor.
START or STOP at SW ON
SW
NOTE
For connection to the input signal, provide a circuit that protects the relay and switch from chattering to prevent malfunctioning.
(2) To Control with External IC or Computer
When controlling START and STOP signals using an external IC or computer, connect them as shown below. Design the circuit so that a current of 10 mA flows to each signal terminal. (See the note in "4.1.1 Signal Line.")
SW
Figure b
10 mA
Figure a or
10 mA
―――――――――
START
――――――――
STOP
input terminal
input terminal
COM terminal
Inside of 3173
Gate IC (e.g., output of a computer)
START or STOP at H
10 mA
―――――――――
START
――――――――
STOP
input terminal
input terminal
COM terminal
Inside of 3173
______________________________________________________________________________________________
4.2 Example of External I/O Terminal
Page 40
30
_____________________________________________________________________________________________
4.2.2 Example of Output Signal Connection
The external I/O terminals are used to output various tester state signals. For output examples, see "4.1.2 Timing Chart of External I/O Terminal."
External I/O Output Signal Specification
Output signals Relay contact (PASS and FAIL)
Open collector output (TEST)
Maximum load voltage
Maximum output current
Output saturated voltage
Signal name PASS : The test returns a PASS judgment.
250 V AC 30 V DC (PASS and FAIL) 30 V DC (No internal pull-up,
1 A/ 1 signal (PASS, FAIL), 20 mA (
0.5 V DC or less (
FAIL : The test returns a FAIL judgment.
―――――――
: Test is under way.
TEST
―――――――
TEST
―――――――
)
TEST
―――――――
TEST
)
)
(1) To Drive a Load
To interlock with an external device by controlling a load (e.g., relay), connect as shown below.
V_ext 30 V or less
―――――――
Inside of 3173
TEST
―――――――
TEST
signal
20 mA or less
Figure a
signal
20 mA or less
Load
V_ext 30 V or less
Load
Inside of 3173
Figure b
______________________________________________________________________________________________
4.2 Example of External I/O Terminal
Page 41
_____________________________________________________________________________________________
g
31
NOTE
PASS signal FAIL signal
1Aorless
250 V AC, 30 V DC or less
Load
負荷
Figure c
The maximum current that can be directed to the TEST signal terminal is 20 mA. For larger currents, use a transistor or similar device for amplification, as shown in Figure b. When connecting a relay or other inductive load to the terminals, connect the diode in parallel with the coil. When connecting a load to the output relay for PASS or FAIL, check the power ratin
of the relay.
NOTE
(2) To Generate Signal Levels
To generate signal levels, connect the terminals as shown below. Note the output current.
―――――――
TEST
Inside of 3173
signal
20 mA or less
Maximum Minimum
V_outH V_ext V_ext-1.5
V_outL 0.5
The output signal status upon power-on may be undetermined. Care should be taken in the operation of equipment connected to the external I/O.
V_ext
30 V or less
V_out
______________________________________________________________________________________________
4.2 Example of External I/O Terminal
Page 42
32
_____________________________________________________________________________________________
______________________________________________________________________________________________
4.2 Example of External I/O Terminal
Page 43
_____________________________________________________________________________________________
33
Chapter 5
Specifications
5.1 Basic Specifications
Voltage Generator
Voltage 0 to 3.0 kV AC (Single range) Voltage testing
method Transformer
capacity Voltage adjustment
method Voltage
measurement Waveform Mains waveform Frequency Mains synchronous
Zero-toggle switch
30 VA
Manually adjusted
Analog, 0 to 3 kV AC f.s., Accuracy: 5%f.s.
Current Detection Section
Interrupting current setting range
Resolution for setting
Interrupting current accuracy
0.1 to 9.9 mA
0.1 mA, , using a digital switch
(5% + 0.05 mA) of the set level
Decision Function
Decision method Analog comparator method (digital setting) Judgment PASS : The current measurements remain below the set
FAIL : A current measurement exceeds or equals the set
Decision process Output to the display, beeper sound, signals to EXT I/O for
each decision result
______________________________________________________________________________________________
threshold value until the set time has elapsed (only when the test timer is set).
threshold.

5.1 Basic Specifications

Page 44
34
_____________________________________________________________________________________________
Timer Section
Setting range 1to99s Operation ON : Test continues until the set time elapses.
OFF : Test continues until STOP input or F AIL judgment.
Setting resolution
1 s (1 to 99 s), (1% + 50 ms) of the set level
and accuracy
Interface
EXT I/O Input Signals
Input signals Active low input (not isolated)
(Wired "OR" connection with START/STOP key) Pulled up to 5 V with 560 Ω.
Maximum input
5.5 V DC
voltage HIGH limit voltage H LOW limit voltage L Signal names
: 5.5 V DC, H
max
: 1.0 V DC, L
max
―――――――――
START
――――――――
: Measurement stop
STOP
: Measurement start
: 4.5 V DC
min
: -0.5 V DC
min
EXT I/O Output Signals
Output signals Relay contact (PASS and FAIL)
Open collector output ( insulation)
Max. load Relay contact 250 V AC / 30 V DC, 1A
―――――――
signal 30 V DC, 20 mA
TEST
Output saturated voltage 0.5 V or below
Signal names
―――――――
: Test is under way.
TEST
PASS : The relay is turned ON when the test returns a
PASS judgment. (Only when the test timer is used.)
FAIL : The relay is turned ON when the test returns a
FAIL judgment.
―――――――
, without pull-up and
TEST
______________________________________________________________________________________________
5.1 Basic Specifications
Page 45
_____________________________________________________________________________________________
35
5.2 General Specifications
Display Analog meter (voltage only) Operating temperature
and humidity Storage temperature and
humidity Operating temperature
and humidity for guaranteed accuracy
Operating environment Max. 2000 m height (6562 feet), Pollution Degree 2, Indoors Power supply 3173 : 100 V AC
Power supply frequency 50/60 Hz Dielectric strength 1.35 kV AC, 10 mA, 1 minute between power supply and frame Maximum rated power 3173: 100 V AC
Dimensions 149W X 200H X 215D mm (5.87"W X 7.87"H X 8.46"D) Approx.
0to40 (32 to 104 ), 80%RH or less (no condensation)
-10 to 50 (14 to 122 ), 90%RH or less (no condensation)
23 5 (73 9 ), 80%RH or less (no condensation) after 10 minutes minimum warm-up Accuracy guaranteed for 1 year.
3173-01: 120 V AC 3173-02: 220 V AC 3173-03: 230 V AC 3173-04: 240 V AC (Voltage fluctuations of 10% from the rated supply voltage are taken into account.)
3173-01: 120 V AC 3173-02: 220 V AC 3173-03: 230 V AC 3173-04: 240 V AC (Common to all voltages: 50 VA)
(excluding projections)
Mass 3173 : 6.3 kg (222.2 oz.) Approx.
3173-01, 3173-02, 3173-03, 3173-04 : 7.0 kg (246.9 oz.) Approx. Fuse 250VT1AL (3173, 3173-01), 250VT0.5AL (3173-02, 3173-03, 3173-04) Accessories 9615 H.V.TEST LEAD (HIGH/LOW side) 1 each
Standard applying
EMC: Safety:
Product warranty period
Power cord 1
Spare
In
EN61326 CLASS A
EN61010
fuse 1
struction man
3 y
ears
ual 1
______________________________________________________________________________________________

5.2 General Specifications

Page 46
36
_____________________________________________________________________________________________
5.3 9615 H.V. TEST LEAD Specifications
Rated voltage HIGH : 5 kV AC / 5 kV DC
LOW : 600 V AC / 600 V DC
Rated current HIGH : 150 mA AC/ 150 mA DC
LOW :10AAC/10ADC
Dielectric strength HIGH : 6.25 kV AC, sensitivity current 5 mA, 1 minute
LOW : 1.35 kV AC, sensitivity current 5 mA, 1 minute Tested area: (Conductor-cable covering)
Operating temperature and humidity
Storage temperature and humidity
Operating environment
Dimensions 1.5 m (59.06") Approx. Mass 100 g (3.5oz.) Approx. (HIGH and LOW) Standard applying EN61010
0to40 (32 to 104 ), 20 to 80%RH (no condensation)
-10 to 50 (14 to 122 ), 90%RH or less (no condensation)
Max. 2000 m height (6562 feet), Pollution Degree 2, Indoors
______________________________________________________________________________________________

5.3 9615 H.V. TEST LEAD Specifications

Page 47
_____________________________________________________________________________________________
37
Chapter 6
Maintenance and Servise
6.1 Maintenance and Inspection
To ensure the safe operation of this unit, perform maintenance regularly.
If the unit has been subject to moisture, or if oil and dust have accumulated in
the unit interior, the danger of electrical shock or fires resulting from the deterioration of insulation increases greatly. If the unit is ever subject to excessive moisture, oil, or dust, cease use immediately, and return the unit to us for maintenance.
Spare and replacement parts for this product are guaranteed to be available
only until 7 years after manufacture of this model is terminated.
Replaceable Parts
Main parts to be replaced periodically, and their life times: (Useful life depends on the operating environment and frequency of use. Operation cannot be guaranteed beyo nd the follo w ing periods.)
Part Life
Start Switch Approx. 200000 cycles Stop Switch Approx. 200000 cycles
If damage is suspected, check the "Troubleshooting" section before contacting
your dealer or Hioki representative. When reshipping the product, pack the product carefully so that it will not be damaged during shipment, and include a detailed written description of the problem. Hioki cannot be responsible for damage that occurs during shipment.
Troubleshooting
Symptom What to check and Solution
The power switch is ON, but no power is supplied.
The tester does not begin testing when the START key is pressed.
The external is not output.
――――――――
TEST
signal
Is the power cord disconnected?
Connect the power cord.
Has the fuse blown?
Replace the fuse.
(See "6.3 Fuse Replacement.")
Are the external input terminals ( fixed to low level?
Set the external input terminals so that they
become high when there is no input.
Is the pull-up resistor for connected? Is appropriate resistance?
Pull up
resistance level. (See "4.2.2 Example of Output Signal Connection.")
――――――――
TEST
――――――――
TEST
――――――――
TEST
signal pulled up at
signal using an appropriate
―――――――――
START
signal
――――――――
,
STOP
)
______________________________________________________________________________________________

6.1 Maintenance and Inspection

Page 48
38
_____________________________________________________________________________________________
If any of the following should occur, stop using the unit, disconnect the power cord and 9615 H.V.TEST LEAD, and contact your dealer or HIOKI representative.
If you are certain that the unit is damaged.If the measurement you wish to perform is inoperative.If the unit was stored for a long period of time in high temperatures and
humidity, or other undesirable conditions.
If the unit was damaged in transit.
6.2 Cleaning
To clean the product, wipe it gently with a soft cloth moistened with water or mild detergent. Never use solvents such as benzene, alcohol, acetone, ether, ketones, thinners or gasoline, as they can deform and discolor the case.
______________________________________________________________________________________________

6.2 Cleaning

Page 49
_____________________________________________________________________________________________
39
6.3 Fuse Replacement
To avoid electric shock, turn off the power switch and disconnect the
WARNING
H.V. TEST LEAD before replacing the fuse. Replace the fuse only with one of the specified characteristics and voltage and current ratings. Using a non-specified fuse or shorting the fuse holder may cause a life-threatening hazard. Fuse type: 250VT1AL (
Turn the power OFF, and disconnect the power cord.
1.
Unlock the fastener on the fuse holder on the rear panel using a slotted
2.
3173, 3173-01),
250VT0.5A L (
3173-02, 3173-03, 3173-04)
screwdriver, and remove the fuse holder.
Replace the power fuse with a rated fuse.
3.
Reset the fuse holder.
4.
Fuse holder
______________________________________________________________________________________________

6.3 Fuse Replacement

Page 50
40
_____________________________________________________________________________________________
______________________________________________________________________________________________
6.3 Fuse Replacement
Page 51
Page 52
Page 53
Page 54
Page 55
Page 56
Loading...