Thank you for purchasing the HIOKI 3173 PORTABLE WITHSTANDING
VOLTAGE HiTESTER. To obtain maximum performance from the product,
please read this manual first, and keep it handy for future reference.
Warranty
HIOKI cannot be responsible for losses caused either directly or indirectly by
the use of the 3173 with other equipment, or if ownership is transferred to a
third party.
Inspection
WARNING
When you receive the product, inspect it carefully to ensure that no damage
occurred during shipping. In particular, check the connectors and
accessories. If damage is evident, or if it fails to operate according to the
specifications, contact your dealer or Hioki representative.
Accessories
9615 H.V.TEST LEAD (HIGH/LOW) 1 each
Power cord 1
Spare fuse 1
Instruction Manual 1
Preliminary Checks
Before using the product the first time, verify that it operates normally to
ensure that the no damage occurred during storage or shipping. If you find
any damage, contact your dealer or Hioki representative.
Use the original packing materials when reshipping the product, if possible.
Before using the product, make sure that the insulation on the test
leads is undamaged and that no bare conductors are improperly
exposed. Using the product under such conditions could result in
electrocution. Replace the test leads specified by Hioki .
This product is designed to conform to IEC 61010 Safety Standards, and
WARNING
Safety Symbols
has been thoroughly tested for safety prior to shipment. However,
mishandling during use could result in injury or death, as well as
damage to the product. Be certain that you understand the instructions
and precautions in the manual before use. We disclaim any
responsibility for accidents or injuries not resulting directly from
product defects.
This manual contains information and warnings essential for safe operation
of the product and for maintaining it in safe operating condition. Before
using the product, be sure to carefully read the following safety notes.
Thesymbol printed on the product indicates that the user should
refer to a corresponding topic in the manual (marked with the
symbol) before using the relevant function.
In the manual, thesymbol indicates particularly important
information that the user should read before using the product.
Indicates that dangerous voltage may be present at this terminal.
Indicates AC (Alternating Current).
Indicates a fuse.
Indicates a protective conductor terminal.
Indicates the ON side of the power switch.
Indicates the OFF side of the power switch.
The following symbols in this manual indicate the relative importance of
cautions and warnings.
Indicates that incorrect operation presents an extreme hazard that
DANGE R
WARNING
CAUTION
could result in serious injury or death to the user.
Indicates that incorrect operation presents a significant hazard that
could result in serious injury or death to the user.
Indicates that incorrect operation presents a possibility of injury to
the user or damage to the product.
To ensure safe operation of measuring instruments, IEC 61010 establishes
safety standards for various electrical environments, categorized as CAT II to
CAT IV, and called measurement categories.
Using a measuring instrument in an environment designated with a
WARNING
higher-numbered category than that for which the instrumen is rated
could result in a severe accident, and must be carefully avoided.
Never use a measuring instrument that lacks category labeling in a
CAT II to CAT IV measurement environment. Doing so could result in
a serious accident.
CAT II
CAT III
CAT IV
When directly measuring the electrical outlet receptacles of the
primary electrical circuits in equipment connected to an AC
electrical outlet by a power cord (portable tools, household
appliances, etc.)
When measuring the primary electrical circuits of heavy equipment (fixed
installations) connected directly to the distribution panel, and feeders
from the distribution panel to outlets
When measuring the circuit from the service drop to the service entrance,
and to the power meter and primary overcurrent protection device
(distribution panel)
Accuracy
We define measurement tolerances in terms of f.s. (full scale) and rdg.
(reading) values, with the following meanings:
(maximum display value or scale length)
The maximum displayable value or the full length of the scale.
This is usually the maximum value of the currently selected range.
(reading or displayed value)
The value currently being measured and indicated on the measuring
product.
Follow these precautions to ensure safe operation and to obtain the full
benefits of the various functions.
Observe the following precautions to avoid electric shock.
DANGE R
Do not remove the product's case. The internal components of the
product carry high voltages and may become very hot during
operation.
Do not touch the alligator clip while high voltage is being generated.
The vinyl sheath on the alligator clip of the 9615 H.V.TEST LEAD does
not have an insulation withstand voltage.
WARNING
CAUTION
To avoid electric shock, do not allow the product to get wet, and do
not use it when your hands are wet.
To avoid electric shock, connect the protective ground terminal to a
grounded conductor before making any other connections.
Before turning the product on, make sure the source voltage matches
that indicated on the product's power connector. Connection to an
improper supply voltage may damage the product and present an
electrical hazard.
Replace the fuse only with one of the specified characteristics and
voltage and current ratings. Using a non-specified fuse or shorting the
fuse holder may cause a life-threatening hazard.
Fuse type: 250VT1AL (
3173, 3173-01),
250VT0.5A L (
3173-02, 3173-03, 3173-04)
To avoid electrocution, turn off the power to all devices before
pluggingor unplugging any of the interface connectors.
To avoid damaging the power cord, grasp the plug, not the cord, when
unplugging the cord from the power outlet.
To avoid damaging the test leads, do not bend or pull the 9615 H.V. TEST
LEADs.
To avoid damage to the product, protect it from vibration or shock during
transport and handling, and be especially careful to avoid dropping.
Do not use the product near a device that generates a strong
electromagnetic field or electrostatic charge, as these may cause
erroneous measurements.
To avoid electric shock, do not exceed the lower of the ratings shown on
the instrument and test leads.
This product may cause interference if used in residential areas. Such use must be
avoided unless the user takes special measures to reduce electromagnetic emissions
revent interference to the reception of radio and television broadcasts.
Test timer setting dial
Test timer switch
DANGER lamp
HIGH terminal
LOW terminal
Output-voltage adjustment
knob
Test result LED
START key
STOP key
Indicates output-voltage when testing withstand voltage.
The LED lights up during testing when the timer is set to
ON.
Sets the current threshold.
(0.1 to 9.9 mA, resolution of 0.1 mA)
Sets the timer (test time). (1 to 99 s, resolution of 1 s)
Turns ON/OFF the test timer.
This lamp lights to warn that voltage is present between the
terminals during testing.
The HIGH terminal is a high-voltage terminal for voltage
outputs.
The LOW terminal is a low-voltage terminal for voltage
outputs. It has the same electric potential as the unit body.
Sets the output-voltage.
One of the two LEDs will light, depending on the test result.
Used to start a test.
Normally used to terminate a test.
Power switch
Powers the 3173 on or off.
The LED lights up when power is turned ON.
To avoid electric shock, never touch the output-voltage terminals (HIGH
WARNING
and LOW), high-voltage test leads or the device being tested while high
voltage is being applied.
External I/O terminal
Power inlet
Fuse holder
Fuse holder
External I/O terminal
Power inlet
Protective ground
terminal
Protective ground
terminal
Contains a power fuse.
Each signal is output by ON/OFF of the relay depending on the test
result. Terminals for start and stop input signals are also available.
Connect the grounded three-core power cord supplied here.
Used to earth a protective ground wire. Be sure to make grounding
Temperature: 0 to 40(32 to 104)
Humidity: 80% RH or less (no condensation)
Do not place the tester upside down. Be sure to place the tester with
its bottom side down.
Avoid the following locations:
・Subject to direct sunlight.
・Having poor air circulation.
・Subject to high levels of dust, steam, or corrosive gases (Avoid
using the equipment in an environment containing corrosive gases
(e.g., H2S, SO2,NI2,andCI2) or substances that generate harmful
gasses (e.g., organic silicones, cyanides, and formalins)).
・Subject to vibrations.
・In the vicinity of equipment generating strong electromagnetic
To avoid electric shock, be sure to connect the protective ground
WARNING
terminal to a grounded conductor.
To avoid electric shock, connect the protective ground terminal to a
grounded conductor before making any other connections.
1
. Using a Phillips-head screwdriver, remove the screw of the protective
ground terminal from the rear of the unit.
2
. Connect an electric wire with a sufficient current capacity to the
protective ground terminal, and secure the wire using a Phillips-head
screwdriver.
Connect the external I/O terminal before turning on the power. If the external
I/O terminal is installed or removed following startup, malfunction may
result.
1
. Press and hold the button on the terminal block using a flathead
screwdriver.
2
. While holding the button, insert a signal cable with insulation removed
from the end into the hole.
3
. Release the button to secure the cable to the terminal.
NOTE
For the specifications of the external I/O terminal, see Section 4.1.
Observe the following precautions to avoid electric shock.
WARNING
Turn off the power unit and tested object, make sure that there is no
high voltage being applied to the output, confirm the following items,
and connect the 9615 H.V. TEST LEAD.
(1) The analog voltmeter reads 0 kV.
(2) The
Make sure the 9615 H.V. TEST LEAD is securely connected before
starting a test, as a loose test lead can cause a hazard when a voltage
is output.
Before using the product, make sure that the insulation on the test
leads is undamaged and that no bare conductors are improperly
exposed. Using the product under such conditions could result in
electrocution. Replace the test leads specified by Hioki .
1
. Remove the LOW terminal by turning it counterclockwise.
DANGER
lamp is OFF.
LOW terminal
2
. As shown in the figure, insert the plug on the H.V. TEST LEAD (low
voltage side).
3
. Secure the LOW terminal by turning it clockwise.
4
. Connect the plug on the H.V. TEST LEAD (high voltage side) to the
Observe the following precautions to avoid electric shock.
WARNING
Turn off the power unit and tested object, make sure that there is no
high voltage being applied to the output, confirm the following items,
and connect the 9615 H.V. TEST LEAD.
(1) The analog voltmeter reads 0 kV.
(2) The
Make sure that no high voltage is being applied between the outputvoltage terminals (HIGH and LOW), before touching the output-voltage
terminals, H.V. TEST LEAD, or tested object. Even following a test,
there may be a residual voltage at the output-voltage terminals.
1
. Make sure the analog voltmeter is at 0 kV and the
2
. Connect the LOW terminal probe to the tested object. Fasten the probe
DANGER
lamp is OFF.
DANGER
lamp is OFF.
securely to prevent it from loosening during a test.
NOTE
3
. Following the procedure described above, connect the HIGH terminal
probe to the tested object.
If the test lead touches to other metallic parts, it may cause an increase in leakage
current. Prevent the test lead from coming into contact with other parts as muc h as
possible.
Do not the test lead and the EXT/IO cable arranged closely to each other.
Doin
so maylead to a malfunction of the external control due to a noise.
Before turning on the power, make sure that the voltage of the power
WARNING
supply being used matches the supply voltage indicated on the rear
panel of the unit. If an attempt is made to use an improper supply
voltage, there is danger of damage to this unit and of life-threatening
risk to the operator. Apply a voltage within the acceptable powervoltage range. Otherwise, damage to the unit or electrical accidents
may result.
NOTE
Allow 10 minutes warming up after powering on.
Output voltage adjustment knob
Power switch
(1) Powering on the unit
1
. Make sure that the voltage adjusting knob is turned counter-clockwise as
far as possible.
2
. Turn the power switch to ON ( ).
(2) Powering off the unit
1
. Following a test, make sure the analog voltmeter is at 0 kV and the
DANGER
2
. Turn the output voltage adjustment knob counterclockwise until the
To ensure safe testing, check the following before starting operation:
(1) Breaking current
1
. Calculate the resistance based on the output-voltage set for withstand
voltage testing and the upper test value, and prepare a resistor
corresponding to the calculation result.
Note the power rating of the resistor.
(Output voltage / test upper limit (interrupting current) = resistance)
(The 2nd power of output voltage / resistance = power consumption <
powerratingofresistor)
A high-voltage resistor is recommended.
Example high-voltage resistor:
KOA Corp.'s GS Series High Voltage High Resistance Thick Film
Resistor
2
. Set an upper-limit value.
3
. Connect the resistor to the H.V. TEST LEAD.
4
. Increase the output voltage beyond the set value, and make sure the
current is cut off (i.e., make sure the unit is in the FAIL state).
(2) Analog voltmeter
1
. Before turning on the power, make sure the analog voltmeter is at 0 kV.
2
. If the voltage reading is not at zero, adjust the value to zero using a
slotted screwdriver.
(3) Key inspection
1
. Turn off the power, and unplug the power cord from the power outlet.
2
. For both the START and STOP keys on the front panel of the unit, press
the center of the key, and make sure you feel it click. The click is less
noticeable when the edges of the keys are pressed.
*Clicking
When a key is pressed slowly, there is a moment of slight resistance and
a feeling that the key cannot be pressed any further. When the key is
pressed further after this point, a clicking sensation can be felt.
3
. If you do not feel a click, the key may be broken.
This chapter describes how to set withstand voltage test conditions and the
proper testing procedure.
Read Chapter 2, and make the necessary preparations for testing.
NOTE
Note that the output waveform may be distorted when conducting an AC
withstand voltage test for a voltage-dependent device or object (e.g., ceramic
capacitor). Excessively large distortion may damage the device or tested object.
When the device or tested object is a capacitive load, resonance may occur (when
conducting an AC withstand voltage test) with the coil inductance of the output
high-voltage transformer, depending on the capacitance value. This condition
may generate several tens of volts at the voltage output terminal before the test is
started. If the START key is disabled, reset the output voltage adjustment knob
to zero
then set it back to the desired voltage after startingthe test.
The flowchart below explains how a test is carried out.
Setting the test parameters
「READY状態」
See Section 3.1.1 "Setting the Current Threshold Value (Upper Limit Value)"
and Section 3.1.2 "Setting the Timer (Test Time)"
Set the following before starting the test:
Current threshold value (upper limit value) setting
・
Timer (test time) setting
・
Setting the output-voltage
「READY状態」
See Section 3.1.3 "Setting the Output-voltage"
Before starting a tes t, make output voltage settings using the output-voltage knob.
Never touch the output-voltage terminals (HIGH and LOW), H.V. TEST LEAD, or tested object. A voltage is
output during output-voltage setting using the output-voltage adjustment knob.
Starting a test
「READY状態」
See Section 3.1.4 "Executing a Test"
When you press the START key, the DANGER lamp lights up and the test begins.
If the test timer is ON, the test timer LED starts blinking.
Determination
「READY状態」
See Section 3.2 "PASS or FAIL Determination"
Each test returns a PASS or FAIL judgment by comparing the currents measured during the set
test time with the set threshold. When a measurement equals or exceeds the threshold, the test
returns a FAIL judgment.
A PASS judgment is not returned if the test timer is OFF.
3.1.1Setting the Current Threshold Value (Upper Limit Value)
NOTE
. Make sure the analog voltmeter is at 0 kV and the
1
. Turn the current threshold setting dial to set the threshold to the desired
2
DANGER
lamp is OFF.
level.
The current thresold value (upper limit value) can be set from 0.1 to 9.9mA
(setting resolution: 0.1mA).
To prevent malfunctions, during testing avoid changing the test current threshold
settin
. Make sure the analog voltmeter is at 0 kV and the
1
. Turn the test timer setting dial to set the timer (test time).
2
. Turn the test timer switch ON.
3
The timer (test t ime) can be set from 1 to 99 s (setting resolution: 1 s).
If the test time doesn't need to be set, turn the test timer switch OFF. This will
disable test timer setting.
To prevent malfunctions, during testing avoid turning the test timer switch
ON/OFF or chan
Before starting a test, make output-voltage settings using the output-voltage
adjustment knob.
Observe the following precautions to avoid electric shock.
DANGE R
Make sure that no high voltage is being applied to the output, confirm
the following items, and connect the H.V. TEST LEAD.
(1) The analog voltmeter reads 0 kV.
(2) The
Never touch the output-voltage terminals (HIGH and LOW), H.V. TEST
LEAD, or tested object. A voltage is output during output-voltage
setting using the output-voltage adjustment knob.
Do not touch the alligator clip while high voltage is being generated.
The vinyl sheath on the alligator clip of the 9615 H.V.TEST LEAD does
not have an insulation withstand voltage.
DANGER
lamp is OFF.
NOTE
. Make sure the analog voltmeter is at 0 kV and the
1
. Connect the high voltage test leads to the device to be tested.
2
DANGER
lamp is OFF.
(See Section 2.6 "Connection to the Measured Equipment")
. Press the
3
START
key.
DANGER
lamp will light up and a voltage will be
output.
. Set an output-voltage using the output-voltage adjustment knob.
4
The output-voltage knob increases the output-voltage when turned clockwise, and
decreases the voltage when turned counterclockwise.
(The maximum value: 3.0 kV)
The output-voltageisdisplayed on the analogvoltmeter.
When adjusting the output voltage, make sure that the test timer is OFF and that the
current threshold is set sufficiently high. Since PASS/FAIL judgments are made
even durin
Observe the following precautions to avoid electric shock.
DANGE R
Make sure that no high voltage is being applied to the output, confirm
the following items, and output-voltage.
(1) The analog voltmeter reads 0 kV.
(2) The
DANGER
lamp is OFF.
Never touch the output-voltage terminals (HIGH and LOW), H.V. TEST
LEAD, or tested object. A voltage is output during output-voltage
setting using the output-voltage adjustment knob.
Avoid testing grounded devices. If you must test a grounded device,
be sure to connect the alligator clip of the LOW (black) 9615 H.V.
TEST LEAD to the grounded side of the device. Connecting the HIGH
(red) test lead to the grounded side may lead to electric shock or
damage to the tester.
CAUTION
NOTE
For output, the unit uses a high-voltage transformer that boosts the power
voltage. If an unstable power voltage is used to operate the unit, the tested
object may be damaged by the distortion of the output-voltage waveform
and by the output of a voltage higher than the preset voltage.
If a capacity load is applied to the tested object, the output-voltage may
exceed the preset voltage, thereby damaging the equipment.
Continuous output of a high voltage may heat the bottom of the unit. Take
special care when handling the unit (e.g. transporting the unit).
The
START
signal and
key and
――――――――
STOP
STOP
key are connected with the
external input signal, respectively, as OR gates. The signal
―――――――――
START
external input
works in the same way as the key. Check the level of the external input signals if
the test fails to start or stop properly.
Allow 10 minutes warmingupafterpoweringon.
Confirm the following items, before touching the output-voltage
WARNING
terminals (HIGH and LOW), H.V. TEST LEAD, or tested object. Even
when a test has been terminated, there may still be voltage in the
output-voltage terminals.
(1) The analog voltmeter reads 0 kV.
(2) The
DANGER
lamp is OFF.
When the preset test time has elapsed, the unit switches to the PASS state
and immediately stops outputting a voltage.
If the test time has not been set, PASS screening is not performed. To the
test, press the
STOP
key, which will forcibly terminate the test.
Flow of PASS determination
Output-voltage
Comparative
voltage value
1. Press the
The output is synchronized with the supply voltage waveform. The tester
will not output a voltage until the supply voltage waveform becomes 0 V
after the
Since the output is not stabilized, the output voltage will fluctuate with
the load current.
Set the output voltage with the tester connected to the device to be tested.
2. The tester continues to output voltage until the set test time has elapsed.
If a current measurement equals or exceeds the set threshold during the
test time, the tester enters the "FAIL state" and halts output.
3. After the test time has elapsed, the tester enters the "PASS state" and halts
voltage output.
In the "PASS state," the green LED and the PASS relay for the external
I/O output are both ON.
4. After the "PASS state (which lasts for about 1 second)", the tester returns
to the ready state.
START
Ready
START
START
The current measurements sta
below the set threshold.
Confirm the following items, before touching the output-voltage
WARNING
Output-voltage
terminals (HIGH and LOW), H.V. TEST LEAD, or tested object. Even
when a test has been terminated, there may still be voltage in the
output-voltage terminals.
(1) The analog voltmeter reads 0 kV.
(2) The
DANGER
lamp is OFF.
If a current measurement equals or exceeds the tests upper limits during the
test, the tester enters the "FAIL state."
The unit switches to the FAIL state and immediately stops outputting a
voltage.
Flow of FAIL determination
The current measurement exceeds the threshold value.
Comparative
voltage value
Testing time
when it set up
Zero-cross waiting time
START
Ready
1. Press the
START
Test
DANGER lamp
key to begin testing.
STOP input
FAIL
The output is synchronized with the supply voltage waveform. The tester
will not output a voltage until the supply voltage waveform becomes 0 V
after the
START
key is pressed.
Since the output is not stabilized, the output voltage will fluctuate with
the load current.
Set the output voltage with the tester connected to the device to be tested.
2. The tester continues to output voltage until the set test time has elapsed.
If a current measurement equals or exceeds the set threshold during the
test time, the tester enters the "FAIL state" and halts output.
3. The buzzer sounds and the red LED lights. The FAIL relay for the
external I/O output is turned ON. The tester remains in the "FAIL state"
until the next STOP input.
Measurement begins when this signal becomes low at the specified timing.
(This is equivalent to pressing the START key.) This terminal is connected
with START key input as a wired OR gate. When the
START
key is pressed,
the input signal becomes low. It is internally pulled up to 5V.
――――――――
(2)
STOP
(Wired "OR" connection with
STOP
key: Low active)
Measurement is halted when this signal becomes low at the specified timing.
(This is equivalent to pressing the
with
STOP
key input as a wired OR gate. When the
STOP
key.) This terminal is connected
STOP
key is pressed, the
input signal becomes low. It is internally pulled up to 5V.
When the
STARTorSTOP
key is pressed, the
―――――――――
START
becomes low respectively. When using a gate IC, connect it via a transistor, or take
other similar measures. If an IC is connected directly to the terminals, the external
IC may be short-circuited with GND, damaging the IC.
This signal is turned ON when the tester returns a PASS judgment.
(2) FAIL (Relay contact output)
This signal is turned ON when the tester returns a FAIL judgment.
――――――――
(3)
TEST
Indicates that the tester is outputting a high voltage. This signal stays low
while the high voltage is output. A pull-up resistor is not connected
internally.
(Open collector output: Low active)
NOTE
When using
―――――――――
START
and
――――――――
STOP
relay contacts, check the power rating of the
relays. Remain alert for chattering caused by relay ON/OFF.
signals using a relay or switch, connect
them as shown below. To prevent switch or relay chattering, we recommend
using the connection shown in Figure b. For this connection, implement the
circuit so that a current of 10 mA flows to the transistor.
START or STOP at SW ON
SW
NOTE
For connection to the input signal, provide a circuit that protects the relay and
switch from chattering to prevent malfunctioning.
(2) To Control with External IC or Computer
When controlling START and STOP signals using an external IC or
computer, connect them as shown below. Design the circuit so that a current
of 10 mA flows to each signal terminal.
(See the note in "4.1.1 Signal Line.")
The maximum current that can be directed to the TEST signal terminal is 20 mA.
For larger currents, use a transistor or similar device for amplification, as shown
in Figure b.
When connecting a relay or other inductive load to the terminals, connect the
diode in parallel with the coil.
When connecting a load to the output relay for PASS or FAIL, check the power
ratin
of the relay.
NOTE
(2) To Generate Signal Levels
To generate signal levels, connect the terminals as shown below. Note the
output current.
―――――――
TEST
Inside of 3173
signal
20 mA or less
MaximumMinimum
V_outHV_extV_ext-1.5
V_outL0.5
The output signal status upon power-on may be undetermined. Care should be
taken in the operation of equipment connected to the external I/O.
DisplayAnalog meter (voltage only)
Operating temperature
and humidity
Storage temperature and
humidity
Operating temperature
and humidity for
guaranteed accuracy
Operating environmentMax. 2000 m height (6562 feet), Pollution Degree 2, Indoors
Power supply3173 : 100 V AC
Power supply frequency50/60 Hz
Dielectric strength1.35 kV AC, 10 mA, 1 minute between power supply and frame
Maximum rated power3173: 100 V AC
Dimensions149W X 200H X 215D mm (5.87"W X 7.87"H X 8.46"D) Approx.
0to40 (32 to 104 ), 80%RH or less (no condensation)
-10 to 50(14 to 122 ), 90%RH or less (no condensation)
235(739 ), 80%RH or less (no condensation)
after 10 minutes minimum warm-up
Accuracy guaranteed for 1 year.
3173-01: 120 V AC
3173-02: 220 V AC
3173-03: 230 V AC
3173-04: 240 V AC
(Voltage fluctuations of 10% from the rated supply voltage are taken into account.)
3173-01: 120 V AC
3173-02: 220 V AC
3173-03: 230 V AC
3173-04: 240 V AC (Common to all voltages: 50 VA)
(excluding projections)
Mass3173 : 6.3 kg (222.2 oz.) Approx.
3173-01, 3173-02, 3173-03, 3173-04 : 7.0 kg (246.9 oz.) Approx.
Fuse250VT1AL (3173, 3173-01), 250VT0.5AL (3173-02, 3173-03, 3173-04)
Accessories9615 H.V.TEST LEAD (HIGH/LOW side)1 each
To ensure the safe operation of this unit, perform maintenance regularly.
・If the unit has been subject to moisture, or if oil and dust have accumulated in
the unit interior, the danger of electrical shock or fires resulting from the
deterioration of insulation increases greatly. If the unit is ever subject to
excessive moisture, oil, or dust, cease use immediately, and return the unit to us
for maintenance.
・Spare and replacement parts for this product are guaranteed to be available
only until 7 years after manufacture of this model is terminated.
・Replaceable Parts
Main parts to be replaced periodically, and their life times: (Useful life depends on
the operating environment and frequency of use. Operation cannot be guaranteed
beyo nd the follo w ing periods.)
・If damage is suspected, check the "Troubleshooting" section before contacting
your dealer or Hioki representative.
When reshipping the product, pack the product carefully so that it will not be
damaged during shipment, and include a detailed written description of the
problem. Hioki cannot be responsible for damage that occurs during shipment.
Troubleshooting
SymptomWhat to check and Solution
The power switch is ON,
but no power is supplied.
The tester does not begin
testing when the START
key is pressed.
The external
is not output.
――――――――
TEST
signal
Is the power cord disconnected?
Connect the power cord.
→
Has the fuse blown?
Replace the fuse.
→
(See "6.3 Fuse Replacement.")
Are the external input terminals (
fixed to low level?
Set the external input terminals so that they
→
become high when there is no input.
Is the pull-up resistor for
connected? Is
appropriate resistance?
Pull up
→
resistance level. (See "4.2.2 Example of
Output Signal Connection.")
If any of the following should occur, stop using the unit, disconnect the
power cord and 9615 H.V.TEST LEAD, and contact your dealer or HIOKI
representative.
・If you are certain that the unit is damaged.
・If the measurement you wish to perform is inoperative.
・If the unit was stored for a long period of time in high temperatures and
humidity, or other undesirable conditions.
・If the unit was damaged in transit.
6.2Cleaning
To clean the product, wipe it gently with a soft cloth moistened with water
or mild detergent. Never use solvents such as benzene, alcohol, acetone,
ether, ketones, thinners or gasoline, as they can deform and discolor the
case.
To avoid electric shock, turn off the power switch and disconnect the
WARNING
H.V. TEST LEAD before replacing the fuse.
Replace the fuse only with one of the specified characteristics and
voltage and current ratings. Using a non-specified fuse or shorting the
fuse holder may cause a life-threatening hazard.
Fuse type: 250VT1AL (
Turn the power OFF, and disconnect the power cord.
1.
Unlock the fastener on the fuse holder on the rear panel using a slotted