Hioki 3118 Instruction Manual

INSTRUCTION MANUAL
3118
November 2002 Revised edition 1
3118A981-01 02-11H Printed in Japan
Introduction
Thank you for purchasing the Hioki 3118 M HiTESTER. To obtain maximum performance from the product, please read this manual first, and keep it handy for future reference.
Specifications
Operating temperature and humidities
0to40 (32 t o 10 4 ), 90% RH max (with no condensations)
Storage temperature and humidities
-10 t o 50 (-4 t o 12 2 ), 90% RH max (with no condensations)
Operating environment
Altitude up to 2000 m (6562 feet)
Power source
Rated power voltage: 1.5 VDC X 6, R6P manganese battery X 6 External power (DC8 .5 V/600mA, MAX. 10 V)
Maximum rated power
4VA
Continuous operating time
Approx. 8 ho urs a t 500 V, 1 M (using mannganese batteries)
Dielectric strength
3 kVAC 50/60 Hz for one minute Between electric circuit and case
Dimensions (excluding protrusions)
106W X 145H X 52D mm approx.
4.17"W X 5.71"H X 2.05"D approx.
Mass
500 g, (17.6 oz.) approx. (including batteries)
Accessories
9294 TEST PROBE 9363 CARRYING CASE (3118-11) 9364 CARRYING CASE (3118-12) Instruction Manual, Instruction label R6P manganese battery X 6, Spare fuse
Option
9288 BREAKER PIN, 9289 TEST PROBE, 9257 CONNECTION CORD
Model 3118-11 3118-12
Rated voltage 250 V 500 V 500V 1000 V
Maximum effective
reading
50 M 100 M 200 M 2000 M
1st effective
measurement
range
0.05 to 20 M0.1 to 50 M0.1 to 50 M2to500M
5% of scale reading
2nd effective
measurement
range
20 to 50M50 to 100M50 to 200M500 to 2000M
10% of scale reading
Other accuracy
0.7% of scale length
(including 0 and reading)
Measurement
terminal voltage
accuracy
10% of rated voltage at infinity
90% min. of rated voltage at 1 M (3118-12: 1000 V range at 20 MΩ )
Effective of
temperature
(0 to 40 )
5% of scale reading plus basic allowance at 1 M
(3118-12: 1000 V range at 20 MΩ )
0.7% of scale length plus basic allowance at 0 and
Measurement current (when
short-circuited)
3.4 mA 0.32 mA
Response time
3 s or less at 1 MΩ or 0
(3118-12: 1000 V range at 20 MΩ )
Overload
protection
(measurement
terminal)
280 V rms 600 V rms 600 V rms 1100V rms
Discharge function Yes No Yes No
Resistance range
and accuracy
0to100Ω/ 3% of scale length
Effective of
temperature
(0 to 40 )
3% of scale length plus basic allowance
Open terminal
voltage
310 mV max.
Overload
protection
250 V rms
(Fuse protection: 1 A/250 V with non-arcing type
AC voltage range
and accuracy
0 to 600 V
7% of maximum scale value (50/60 Hz)
Effective of
temperature
(0 to 40 )
7% of maximum scale value
Input resistance Approx. 160 k Approx. 3.3 M
Overload
protection
700 V rms
Safet
y
The
symbol printed on the product indicates that the user should refer to a corresponding topic in the manual (marked with the
symbol) before using the
relevant function.
In the manual, the
symbol indicates particularly important information that the user should read before using the product.
Indicates that dangerous voltage may be present at this terminal.
Indicates AC (Alternating Current).
Indicates DC (Direct Current).
DANGER
Indicates that incorrect operation presents an extreme hazard that could result in serious injury or death to the user.
WARNING
Indicates that incorrect operation presents a significant hazard that could result in serious injury or death to the user.
CAUTION
Indicates that incorrect operation presents a possibility of injury to the user or damage to the product.
NOTE
Advisory items related to performance or correct operation of the product.
Notes on Use
DANGER
Before connecting probes to the product, check that the probes are d isconnected from the object to be measured.
WARNING
To avoid electric shock, do not al low the product to get wet, and do not use it when your hands are wet. Before using the product, make sure that the insulation on the probes is undamaged and that no bare conductors are improperly exposed. Using the product in such conditions could cause an electric shock. Replace the test probes with the specified Hioki Model 9294. Do not use the product where it may be exposed to corrosive or c ombustible gases. The product may be damaged or cause an explosion.
CAUTION
Do not store or use the product where it could be exposed to direct sunlight, high temperature or humidity, or condensation. Under such conditions, the product may be damaged and insulation may deteriorate so that it no longer meets specifications. To avoid damage to the product, protect it from vibration or shock during transport and handli ng, and be especially careful to avoid dropping.
Fuse and BatteryReplacement
WARNING To avoid electric shock when replacing the batteries, first disconnect the test probes from the object to be measured. Before using the product after replaci ng the batteries, replace the cover and screw. Do not mix old and new batteries, or different types of batteries. Also, be careful to obser ve battery polarity during installation. Otherwise, poor performance or damage from battery leakage could result. To avoid the possibility of explosion, do not short circuit, disassemble or incinerate batteries. Keep batteries away from children to prevent accidental swallow ing. Replace the fuse only with one of the specified characteristics and voltage and current ratings. Using a non-specified fuse or shorting the fuse holder may cause a life-threatening hazard. Fuse type: MF51NR1-S 1 A, 250 V (non-arc ing) 20 mm X5mmdia.
Battery cover
Spare fuse
Inspection
When you receive the product, inspect it carefully to ensure that no damage occurred during shipping. If damage is evident, or if it fails to operate according to the specifications, contact your dealer or Hioki representative.
Maintenance and Service
To clean the product, wipe it gently with a soft cloth moistened with water or mild detergent. Never use solvents such as benzene, alcohol, acetone, ether, ketones, thinners or gasoline, as they can deform and discolor the case. If the product seems to be malfunctioning, confirm that the batteries are not discharged, and that the test probes are not open circuited before contacting your dealer or Hioki representative. Pack the product carefully so that it will not be damaged during shipment, and include a detailed written description of the problem. Hioki cannot be responsible for damage that occurs during shipment.
General Specifications
Specifications
Guaranteed at 23 5 (73 9 ), 90% RH or less, for 1 year.
This manual contains information and warnings essential for safe operation of the product and for maintaining it in safe operating condition.
Before using the product, be sure to carefully read the following safety
notes.
Safety Symbols
The following symbols in this manual indicate the relative importance of cautions and warnings.
NOTE
Before use check that batteries are charged. Replace the old batteries with fresh one if they are low. When not in use for a long time , to prevent possible corrosion caused by battery leakage, remove the batt eries before storage. Always release the MEASURE button after use. The meter display is glass, so avoid subjecting it to strong impacts. This product is provided with a label containing measurement instruction and precautions. After purchase, be sure to fix this label to the battery cover on the back of the product. Before use, check whether the product operates properly. You can do this by performing Insulation Resistance Measurement, then pressing the MEASURE button in the Msetting (the indicator should read zero). If the needle remains in the position, the fuse has blown. In this case, replace the fuse with a new one of the same rating (refer to "Fuse and Battery Replacement").
1. Rotate the dec orative sc rew with a coin or similar object.
2. Remove the battery cover.
3. Insert six new batteries. If the fuse has blown, replace it with a new one.
Names and Functi ons of Parts
1
2
3
7
5
10
8
9
6
11
12
13
4
Measurement Procedure
WARNING Observe the follow ing to avoid electric shock and short circuits.
Do not attempt to measure insulation resistance on a live conductor. D oing so could damage the product or cause an accident that might result in injury or death. Always turn off power to the conductor being measured before sta rting. The maximum pe rmissible input is indicated in the specifications. Do not measure voltage in excess of these limitations, as doing so may damage the product or cause an accident that might result in injury or death . When measuring i nsulation resistance, dangerous voltage is a pplied to the measurement termi nals. To avoid electric shock, do not touch the probe. To avoid electrical shock, never touch the external power input terminal and BATT.CHECK terminal while measuring insulation resistance.
CAUTION
When discharging, do not set the function selector to the position. It may result in damage to the product.
1
2
3
4
Source (primary side)
Load (secondary side)
Always turn off the breaker or the measurement LINE.
EARTH terminal
LINE terminal
Circuit to be measured
-
+
3118
ON
OFF
MEASURE button
Low voltage M
ACV
Function selecto
r
Discharge resistance
High voltage M
Capacitance
3118
Core
Mete
r
Power suppl
y
LINE
EARTH
GUARD
Insulato
r
Sheathings
DANGER Test probes should only be connected to the secondary side of a breaker, so the breaker can pre vent an accident if a short circuit occurs. Connections should never be made to the pri mary side of a breaker, because unrestricted current flow could cause a serious accid ent if a short circui t occurs. The maximum permissible input is 700 Vrms. Attempting to measure voltage in excess of the maximum rating cou ld destroy the product and result in personal injure or death. To avoid electrical shock, never touch the external power input terminal and BATT.CHECK terminal while measuring voltage. To avoid electrical shock, be careful to a vo id shorting live lines with the test probes.
WARNING Never press the MEASURE button while measuring voltage. Doing so could damage the circuitry or cause a life-threatening accident.
1
3
Do not press the MEASURE button.
Always connect the probe to the secondary side of the breaker.
Source (primar
y
side
)
Load
(
secondaryside
)
2
DANGER Never apply voltage to test probes when the Resistance function is selected. Doing so may damage the product and result i n personal injury. To avoid electrical accidents, remove power from the circuit before m easuring.
BatteryCheck Method
1
2
Do not press the MEASURE button.
Usingthe CarryingCase
Hook
9288 BREAKER PIN
Option
DANGER
To avoid electrical shock, be careful to a vo id shorting live lines with the breaker pin.
1. MEASURE button
Press to measure insulation resistance.
2. Zero-ajustment 3. Di splay and scale 4. Indicator needle
5. Function selector (Ω/ACV/MΩ)
6. Earth measurement terminal
Connect the EARTH probe (black).
7. Guard terminal
8. Battery check terminal
Use with LINE probe.
9. DC IN: External power supply jack (DC8.5 V/600mA, MAX. 10 V, has φ5 mm-dia.-ter minals and has a center pin with negative polarity)
10. Line measurement terminal
Connect the LINE probe (red).
11. Battery cover 12. decorative screw
13. Instruction label (fix the label here)
Insulation Resistance Measurement
1. Set the f unction selector to either Mposition. (3118-11: 250/500 V, 3118-12: 500/1000 V)
2. Connect the black EARTH probe to the ground side of the object being measured.
3. Connect the red LINE probe to the LINE to be measured.
4. Press the MEASURE button. (To make continuous measurements, pull the button up.)
5. Read th e valu e aft er th e needle has stabil ized.
NOTE
Insulation resistance is the ratio of leakage current to input voltage, and is therefore unstable. Depending on the circuit being measured the needle may stabilize, but this in not a meter malfunction. When measuring insulation between c ommunication cables or other systems with a capacitance element, the needle will move first quickly to the left, then gradually return to the right, and it may take some t ime to become stabilized. This is not a sign of malfunction. Read the value after the needle has stabilized.
Discharge Function
When measuring an insulation resistance that contains a capacitance element, that element will be charged with an electric load proportional to the measurement voltage. After measuring a circuit with capacitance, release the MEASURE button and set the f unction selector to 250 V (3118-11) or 500 V (3118-12) before detaching the probes. This will activate the 3118 built-in discharge circuit.
The time required for discharge depends on the capacitance value. As a reference, observe the indicator needle. During discharge, it will go beyond, then return gradually. discharge is practically completed when needle reaches the
symbol.
Make it a practice t o discharge the subject conductor after measurement.
Guard terminal use
The guard terminal eliminates the effects of test object surface leakage resistance to measure only volumetric resistance.
For measurement of cables and similar objects, as indicated in the diagram, wrap a bare wire (about 0.5 mm in diameter) around the insulator to be measured and connect this to the GUARD terminal. The leakage current flowing through the surface of the insulator will not be displayed, making it possible to accurately measure volumetric resistance.
AC Voltage Measurement
1. Set the function selector to ACV.
2. Connect the black EARTH probe to the ground.
3. Connect the red LINE probe to the line to be measured.
4. Read the voltage value on the AC scale after the needle stabilizes.
Resistance Measurement
1. Set the function selector to Ω.
2. Press the MEASURE button and attach the LINE and EARTH probes.
3. Read th e res istance value on the scale.
1. Set the function selector to B CH.
2. Connect the LINE probe tip to the BATT CHECK terminal.
3. If the needle is within the B range, the batteries can be used. If it is to the outside of the B range, replace the batteries with new ones.
For use, fold the cover back as shown. Attach the belt to the hook as indicated when carried from your shoulder.
Remove the hook if the product is to be used while hung aroud your neck. Carry probes here. Carry the 9288 BREAKER PIN (option) here.
9288 BREAKER PIN
Attach this to the tip of the LINE probe or pin type EARTH probes. It allows the tip to be extended, when other wise it would be too short to ma ke a measurement.
9289 TEST PROBE
The black EARTH probe is a pin type probe like the LINE probe. This can not be attach to the 9288 BREAKER PIN.
9257 CONNECTION CORD
The black EARTH probe is a clip type probe like the LINE probe.