8.1.2 Manual setting of calibration value
The calibration value can be set manually into the RallyComp.
The procedure is as follows:
1. Start the calibration by entering the setup menu, see chapter 8.
2. Go with until 1.Calibrate Distance is showing in the display.
3. Press OK.
4. Press SET.
5. Set the new calibration value, see Setting of values chapter 5.4.
6. Press OK.
7. The new calibration value is now shown in the display.
8. Press OK.
9. Exit the setup menu by pressing BACK.
The new calibration value is now saved in the memory and will be saved
when the power is cut.
8.2 Set clock
To set the clock follow these steps:
1. Start the installation by entering the setup menu, see chapter 8.
2. Go with until 2.Set Clock is showing in the display.
3. The installation of the clock is done by a regular setting, see setting
of values chapter 5.4.
4. Press OK.
5. Exit the setup menu by pressing BACK.
The clock has now been set and will be remembered in the installation
even if the power is cut.
8.3 Adjusting the contrast level of the display
1. Start the installation by entering the setup menu, see chapter 8.
2. Go with until 6.Set Screen Contrast is showing in the display.
3. Press OK.
4. Increase the contrast by repeated pressing the + button.
Decreaset he contrast by repeated pressing the – button.
21