HF Scientific MTOL+ Operating Manual

Page 1
OWNER’S AND USER’S MANUAL
MTOL+TM Turbidimeter
Models 28052 & 28053
Read Manual and all product labels BEFORE using the equipment. Do not use unless you know the safe and proper operation of this equipment. Keep this Manual available for easy access by all users. Replacement Manuals are available at HFscientific.com
HF scientific 3170 Old Metro Parkway Ft. Myers, FL 33916 Phone: 239-337-2116 Fax: 239-332-7643 Toll free: 888-203-7248 E-Mail: HF.Info@WattsWater.com Website: HFscientific.com
Catalog No. 29290
Page 2
Catalog No. 29290
Page 3
DECLARATION OF CONFORMITY
Application of Council Directive
Standards to Which Conformity is Declared:
Product Safety – Tested and passed: LC (Part 1: General Requirements UL61010-1 Issued May 11 2012 Ed 3 and CAN/CSA C22.2 #61010-1 issued May 11 2012)
Manufacturer’s Name: HF scientific, inc.
Manufacturer’s Address: 3170 Old Metro Parkway, Fort Myers, Florida 33916-7597
Importer’s Name:
Importer’s Address:
Type of Equipment: Process Turbidimeter
Model Name MTOL+TM
Model No: Model #28052 (Infrared), Model #28053 (White Light)
I, the undersigned, hereby declare that the equipment specified above conforms to the above Directives and Standards.
Place: Fort Myers, Florida USA (Signature) Date: 26 January, 2018 Renato Trovo,
General Manager,
HF scientific inc.
Catalog No. 29290
Page 4
Attention Owners and Users
Thank you for purchasing the MTOL+™ turbidimeter. This equipment will provide safe and productive operation as long as it is installed, used, maintained, and serviced in accordance with the instructions in this manual and is properly maintained. Importantly, unless the user is adequately trained and supervised, there is a possibility of death, serious personal injury, property damage or damage to the equipment.
Owners and users of this equipment bear the responsibility to make certain that this equipment is used properly and safely. READ THIS MANUAL carefully, learn how to use and service this equipment correctly, and strictly follow all of the instructions contained in this manual and the requirements of local, state and federal law. Failure to do so could result in death, serious personal injury, property damage or damage to the equipment. This manual should be considered a permanent part of your machine and should be kept available for easy reference by any user.
Owners should not permit anyone to touch this equipment unless they are over 18 years of age, are adequately trained and supervised, and have read and understand this manual. Owners should also ensure that no unauthorized personnel come in contact with this equipment.
If this equipment, or any of its parts, becomes damaged or needs repair, stop using the equipment and contact an experienced service individual immediately. If the warning labels or this manual are misplaced, damaged or illegible, or if you require additional copies, please contact customer service at +1 (239) 337-2116 or 888-203-7248 for these items at no charge.
Please remember that this manual and the warning labels do not replace the need to be alert, to properly train and supervise users, and to use common sense when using this equipment.
If you are ever uncertain about a particular task or the proper method of operating this equipment, ask your supervisor, consult this manual, access www.hfscientific.com, or contact us at 888-203-7248.
PRODUCT IDENTIFICATION
Please record your product’s identification and purchase information which will help in the event you have questions or need any service.
Model:
Serial #:
MTOL+ (Mar 5, 2018) Rev. 1.71
Date of purchase:
Seller name/address:
Page 5
Table of Contents
Section Page
Understanding Safety Information ........................................................ 1
Specifications ........................................................................................ 2
1.0 Overview ............................................................................................... 3
1.1 The MTOL+ ......................................................................... 3
1.2 The Display .......................................................................... 4
1.3 The Touch Pad .................................................................... 4
1.4 Vapor Purge ........................................................................ 5
2.0 Unpacking and Inspection .................................................................... 5
3.0 Installation ............................................................................................ 6
3.1 Site Selection ............................................................................. 7
3.2 Mounting .................................................................................... 7
3.3 Plumbing & Assembly ............................................................ 10
3.3.1 Initial Assembly ........................................................... 10
3.3.2 Connection of MTOL+ to System Plumbing ................ 10
3.3.3 Connecting the drain ................................................... 12
3.3.4 Connecting the water intake ........................................ 12
3.4 Electrical Connections ............................................................. 14
3.4.1 Electrical Power Connection ....................................... 15
3.4.2 Terminal Block (P1) Connections ................................. 16
3.4.3 Closing the Junction Box ............................................. 18
4.0 Setup, Configuration and Calibration ............................................... 19
4.1 Modes of Operation .................................................................. 19
4.2 Configuring the MTOL + .......................................................... 22
4.2.1 Configuring the 4-20 mA Output .................................. 22
4.2.2 Baud Rate ..................................................................... 24
4.2.3 Modbus Address ........................................................... 24
4.2.4 Configuring the Alarms .................................................. 25
4.2.4.1 Configuring Alarm 1 ......................................... 26
4.2.4.2 Configuring Alarm 2 ......................................... 27
4.2.5 Offset ............................................................................. 27
4.2.6 Access Code ................................................................. 28
4.2.7 Signal Averaging ........................................................... 29
4.2.8 Display Resolution ........................................................ 30
4.2.9 LCD Brightness ............................................................. 31
4.2.10 Display Units ................................................................. 31
4.2.11 Ultrasonic Cleaning ....................................................... 32
4.2.12 Modbus Parity & Modbus Stop Bits ............................... 32
4.2.13 4mA and 20mA Adjustment .......................................... 33
4.2.14 Instrument ID ................................................................. 34
4.2.15 Logging Interval ............................................................. 34
MTOL+ (Mar 5, 2018) Rev. 1.71
Page 6
Table of Contents (continued)
Section Page
4.2.16 Download Logged Files ................................................ 35
4.2.17 Date and Time ............................................................... 35
4.2.18 Update Software ........................................................... 38
4.3 Calibrating the MTOL+, CAL Mode .......................................... 40
4.3.1 Calibration Standards .................................................... 40
4.3.1.1 Alternate Ranges ............................................... 40
4.3.2 Indexing Calibration Cuvettes ........................................ 41
4.3.3 Calibration Procedures .................................................. 41
4.3.4 Abort Calibration ........................................................... 43
4.3.5 Calibration Failure ......................................................... 43
4.3.6 Instrument Offset ........................................................... 43
5.0 Operation ........................................................................................... 45
5.1 Replacing or Installing the Desiccant Pouch ............................ 45
5.2 Starting the System .................................................................. 46
5.3 Routine Measurement: AUTO mode ........................................ 47
5.4 Data Retrieval .......................................................................... 47
5.4.1 USB Connection ............................................................ 46
5.4.2 File Downloads ............................................................. 47
5.4.3 Modbus Communication ................................................ 48
5.5 Patented Ultrasonic Cleaning ................................................... 49
5.6 Condensate on the Cuvette ..................................................... 49
5.7 Cleaning the Flow through Cuvette .......................................... 51
5.8 Replacing the Source Lamp ..................................................... 52
5.9 Factory Installed 24V Option ................................................... 52
6.0 Troubleshooting .................................................................................. 53
6.1 MTOL+ Fault Detection ............................................................ 53
6.1.1 Warnings ...................................................................... 53
6.1.2 Errors ............................................................................ 53
6.1.3 Failures ......................................................................... 53
6.2 Diagnostic Table ..................................................................... 54
6.3 Bubbles in the Sample ............................................................ 54
6.4 Technical and Customer Assistance ....................................... 55
7.0 Accessories and Replacement Parts List ........................................... 56
8.0 Limited Warranty ................................................................................ 58
MTOL+ (Mar 5, 2018) Rev. 1.71
Page 7
Understanding Safety Information
This manual contains safety and use instructions that must be followed during the installation, commissioning, operation, care and maintenance, and service of the MTOL+. All responsible personnel must read this manual prior to working with this instrument, and should familiarize themselves with the following safety symbols, signals, and pictorials.
This is a safety-alert symbol.
The safety alert symbol is shown alone or used with a signal word (DANGER, WARNING or CAUTION), a pictorial and/or a safety message to alert you to hazards.
When you see this symbol alone or with a signal word on this instrument or in this Manual, be alert to the potential for death or serious personal injury.
Safety signal words have the following meaning:
Identifies hazards which, if not avoided, will result in death or serious injury.
Identifies hazards which, if not avoided, could result in death or serious injury.
Identifies hazards which, if not avoided, will result in minor or moderate injury.
Identifies practices, actions or failure to act which will result in property damage or damage to the equipment.
Pictorials used on the equipment and in this Manual have the following meanings:
This pictorial alerts you to the need to read the Manual.
1
This pictorial alerts you to electricity, electrocution and shock hazards.
Page 8
Specifications
Measurement Range
Accuracy
Resolution
Response Time
Display
Alarms
Analog Output
Communications Port
Data Storage
Software
Maximum Water Pressure
Flow Rate
0 – 100 NTU, factory configured
Model #28052 (Infrared) & Model #28053 (White Light)
0-10 NTU & 0-1000 NTU optional ranges (loaded on USB thumb drive)
±2% of reading or ±0.02 NTU below 40 NTU whichever is greater (0-100 and
0-1000 ranges)
±5% of reading above 40 NTU (0-100 and 0-1000 NTU ranges)
±2% of reading or ±0.02 NTU (0-10 NTU range)
0.0001 NTU (below 10 NTU)
Adjustable
Multi-Line Liquid Crystal Backlit Display
Two Programmable, 120-240VAC 2A Form C Relay
Powered 4-20 mA, 1000 drive
Bi-directional RS-485, Modbus RTU
16 Gigabytes download via USB flash drive
Upgradeable via USB flash drive
Integral pressure regulator rated 7 bar (101.5 PSI)
100 ml/min. – 1.5 liter/min. (.026-.40 Gal/min)
Operating Temperature
Wetted Materials
Sample Temperature
Power Supply
Insulation Rating
Environmental Conditions
Enclosure
Regulatory Compliance
And Certifications
Instrument Weight
Shipping Weight
Limited Warranty
1°C – 50°C (34°F – 122°F)
Nylon, Borosilicate Glass, Silicone, Polypropylene, Stainless Steel, Viton, Acetyl
1°C – 50°C (34°F – 122°F)
100 – 240 VAC, 47 – 63 Hz, 80VA
Double Insulated, Pollution Degree 2, Overvoltage Category II
Not recommended for outdoor use.
Altitude up to 2000 meters
Up to 95 % RH (non-condensing)
ABS
White Light Version compliant to U.S. EPA 180.1
Infrared Version compliant to ISO 7027 and DIN 27027
CE Approved, LC listed to UL 61010-1 3rd Edition: 2012
LC: Certified to CSA 22.2 No.61010-1-12: 2012
1.8 kg (4.0 lbs.)
2.8 kg (6.2 lbs.)
2 Years from date of invoice
2
Page 9
1.0 Overview
Shutoff Clamp
Display
Keypad
Water Inlet
Backpressure valve
USB Slot
Drain
Junction Box
1.1 The MTOL+ - Product Description
The MTOL+ process turbidimeter allows for the measurement of the turbidity of process water on-line. The MTOL+ works by nephelometery; which means it
measures reflected light at 90° incident to the light beam. The white light MTOL+ (Model #28053) has been designed to meet the design criteria specified by the
US EPA 180.1 on turbidity measurement. The infrared MTOL+ (Model #28052) was designed to meet the design criteria specified in ISO 7027 and DIN 27027 for the measurement of the turbidity of a sample. Both models have long life lamps.
All models have ultrasonic cleaning. Refer to section 5.5 for more information.
A pressure regulator on the incoming line is a standard on all MTOL+ instruments
and will reduce pressures up to 7 bar (101.5 PSI) down to 1.03 bar (15 PSI).
The MTOL+ instruments are fully equipped with isolated 4-20mA, Modbus, two alarms and ultrasonic cleaning. The only option is a flow alarm which must be factory installed.
3
Page 10
1.2 The Display
Figure 1 illustrates all the items that can appear on the display.
The upper row of the display is used for reporting the turbidity levels and to provide user guidance in the customer setting routine.
The lower row of the display is used to communicate error messages and provide user guidance.
The display has two icons that are used to indicate the use of access code and offset mode.
In addition, mode arrows are used to indicate the current instrument operating mode: AUTO (normal operation), CAL (calibration) and CONFIG (configuration).
Figure 1 – Display used in the instrument.
All items used on the display are shown in this figure
1.3 The Touch Pad
Figure 2 illustrates the touch pad. The touch pad has five buttons arranged in star pattern with a center button.
Right and left buttons allow scrolling forward and backward through
the menus
Up and down buttons are used to change selections
The center button is used to accept a change.
Figure 2: Touch Pad
4
Page 11
Item
Quantity
1.4 Vapor Purge
The MTOL+ is equipped with a continuous Vapor Purge system.
A replaceable desiccant pouch in the lower portion of the instrument dries the air.
System heat is used to warm the air. A fan inside the instrument continuously circulates heated dry air around the optical well and the flow through cuvette. This feature eliminates the need for a dry purge line.
The MTOL+ monitors the replaceable desiccant pouch condition continuously. The LCD display will show DESC on the lower line in the event that the desiccant pouch needs replacement. Replacement desiccant pouches are available from HF scientific or the local representative (Part # 21555R). Refer to section 5.1
Replacing or installing the Desiccant Pouch.
The MTOL+ creates its own internal environment. In this controlled environment condensation cannot form. The cuvette is essentially bathed in warm, dry air.
2.0 Unpacking and Inspection
The table below indicates the items in the turbidimeter shipment.
MTOL+ Turbidimeter with Flow Through Assembly
Field Terminal Box with Interconnect cable
Owner’s and User’s Manual
Desiccant Pack
Tubing Kit: 1-shutoff clamp
1-backpressure valve
2-connecting tubing with fittings for flow through assembly
Screwdriver to make electrical connections in junction box
USB Flash Drive (Including files for optional ranges and user manual)
Remove the instrument from the packing carton. Carefully inspect all items to ensure that no visible damage has occurred during shipment. If the items received do not match the order, please immediately contact the local distributor or the HF scientific Customer Service department.
1
1
1
1
1
1
1
5
Page 12
3.0 Installation
The MTOL+ must be mounted and installed in a suitable location, plumbed to the sample water source (section 3.34), and hard-wired to electrical power (section 3.4). Installation MUST be performed by qualified technicians, including licensed electricians, following all instructions, complying with all local, state, federal and other governmental requirements, and with all building and construction codes and standards.
If you have any questions or are uncertain about proper site location, installation or connection of the MTOL+, please contact customer service at HF scientific.
• Installation MUST be performed by qualified technicians, including licensed electricians, following all instructions, complying with all local, state, federal and other governmental requirements, and with all building and construction codes and standards.
• MTOL+ is electrically powered. ALWAYS take proper precautions to recognize, evaluate, and control electricity hazards during installation, programming, use and service/maintenance.
• Use ONLY with water free of corrosive chemicals or other hazardous substances.
ALWAYS leak check plumbing connections following installation and on a regular basis.
• MTOL+ MUST be installed in an area isolated from vibration, and not susceptible to freezing or temperature extremes. Mount on any suitable vertical and flat surface that can support the product’s weight.
• Keep installation and work area clean, well-lighted, free of clutter and distractions, and accessible only by authorized personnel and workers.
• MTOL+ must be located in accessible and well-lighted area for use, servicing, repair or replacement by authorized personnel.
6
Page 13
3.1 Site Selection
Choose a site that is accessible and well-lighted for use, servicing, repair or replacement by authorized personnel.
MTOL+ should be located and mounted within 2-3 meters (6 to10 ft) of the sampling point to ensure a quick response time.
Installation/mounting site must be isolated from vibration, and not susceptible to freezing or temperature extremes.
Site must have a suitable vertical and flat surface that can support the product’s weight and service operations 5kg (10lbs).
There should be 20 cm (8 in) free area above the instrument. This will ensure enough room for calibration and cuvette maintenance.
Site must allow for mounting MTOL+ with front display positioned at eye level 1.4 to 2 meters (approximately 4.5 to 6.5 feet from floor).
MTOL+ is designed to have the Junction Box cradled under the upper portion of the instrument – Select a site with enough space for proper configuration when mounted.
3.2 Mounting
Described is for installation to a concrete surface. If needed
contact HF scientific for instruction to install on other surfaces.
Tools required:
Appropriate screwdriver or driver gun.
Mounting Template – last page of this Manual.
Masking tape for attaching Mounting Template to wall.
Any tools required by wall anchor manufacturer if used.
Hardware required:
7
4 Mounting screws for instrument enclosure - M6 (¼”). Either use concrete
screws or use a wall anchors & the supplied screws.
2 Mounting screws for terminal junction box - M4 (#8). Either use concrete
screws or use a wall anchors & the supplied screws.
Page 14
1. Using masking tape, affix the mounting template to the vertical surface of the selected mounting location and mark location for screws.
The MTOL+ is designed for vertical wall mounting. If wall mounting is not practical, the instrument can be mounted on any suitable vertical, flat surface.
Make sure that the mounting position is level.
2. The overall mounting dimensions of the MTOL+ are shown in Figure 3,
below. Measure to confirm that the mounting location has sufficient space and that there is at least 20 cm (8 in) free area above the instrument when mounted. This will ensure enough room for calibration and cuvette maintenance.
3. Carefully remove the mounting template from the wall, keeping it in one piece for future reference and use.
4. Install wall anchors (if needed) - Following manufacturer’s instructions for the wall anchors, and using the mounting template as a guide, insert wall anchors for each mounting position.
5. Mount the Junction Box First. The MTOL+ is designed to have the Junction Box cradled under the upper portion of the instrument containing the key pad and display.
a. Position the Junction Box of the MTOL+ to match the 2 lower wall
anchors.
b. Mount the Junction Box first by inserting the M4 (#8) screws
through the 2 inside the Junction Box. Use wall anchors as needed.
Using the screwdriver, hand-tighten both screws, ensuring that the Junction Box is firmly and securely attached.
6. Mount the upper key pad and display section. After securely mounting the junction box, mount the upper portion of the MTOL+ so that the junction box is cradled under the upper portion.
a. Position the upper portion of the MTOL+ to match the remaining
holes or wall anchors.
b. Insert M6 (1/4”) screws though the 4 tabs on the upper portion of
the MTOL+ and into the wall anchors.
c. Using the screwdriver or driver gun, to tighten the screws, ensuring
that the upper portion is firmly and securely attached.
8
7. Confirm that MTOL+ is securely mounted to wall (or vertical surface) and positioned:
Page 15
Within 2-3 meters (6 to10 ft) of the sampling point to ensure a quick
response time.
Isolated from vibration, and not susceptible to freezing or temperature
extremes.
Suitable vertical and flat surface that can support the product’s weight
5 kg (10lbs).
20 cm (8 in) free area above the instrument, allowing room for
calibration and cuvette maintenance.
Front display should be positioned at eye level or a suitable level to
read and reach controls - approximately 1.4 Meters (4.5 feet) to 2 meters (6.5 feet) from floor.
Confirm that the unit is level.
Turbidity Sensor
Junction Box
9
Figure 3: Mounting Dimensions of the Instrument
Page 16
3.3 Plumbing & Assembly
ONLY qualified personnel should perform plumbing connections following all instructions, complying with all local, state, federal and other governmental requirements, and with all building and construction codes and standards.
MTOL+ requires very little head pressure to operate 6.9kPa (1 PSI).
The flow through cuvette is rated for a flow of 100ml/min. - 1 liter/min. (0.026-
0.26Gal/min).
The integral pressure regulator is rated for a MAXIMUM pressure of 7 bar
(101.5 PSI).
The MAXIMUM allowable fluid temperature is 50°C (122°F).
3.3.1 Initial Assembly
The IN and OUT tubes must be connected to the MTOL+ before the system is connected to the sample water source and drain.
1. Identify the IN tube. This is the tube with the white ratcheting clamp.
2. Install the IN tube between the sensor mounting ear marked IN and the flow head connection marked IN. Secure both ends tightly.
3. Identify the OUT tube. This is the tube with the rotary backpressure valve.
4. Install the OUT tube between the sensor mounting ear marked OUT and the flow head connection marked OUT. Secure both ends tightly.
3.3.2 Connection of MTOL+ to System Plumbing
Requirements:
Use ¼ inch OD flexible tubing (polyethylene or polypropylene) for connecting
the MTOL+.
Use Opaque tubing if tubing may be exposed to sunlight. This will help limit algae growth.
10
MTOL+ has push-to-connect fittings for water inlet and drain connections.
Page 17
WATER INLET
TUBING
DRAIN
TUBING CONNECTION
PRESSURE REGULATOR
SHUTOFF CLAMP
This clamp allows for shutting off the intake flow during cuvette maintenance. Press down to restrict water flow.
Flow Through Cuvette
The flow through cuvette is rated for a flow of 100ml/min ­1liter/min. (0.026-0.26Gal/min).
BACKPRESSURE VALVE
This valve allows for an adjustment to reduce bubble formation. Turn clockwise to increase backpressure.
The integral pressure regulator is rated for a maximum pressure of 7 bar (101.5 PSI). The maximum allowable fluid temperature is 50°C (122°F).
CONNECTION
Use ¼” OD tubing with supplied push-to-connect fitting. Polyethylene or polypropylene are recommended tubing types.
Use ¼” OD tubing with supplied push-to-connect fitting. Do not reintroduce drain sample back into the process stream
Figure 4 Instrument parts identification
11
Page 18
3.3.3 Connecting the drain:
a. MTOL+ drain tubing MUST be routed to a suitable
sanitary sewer drain. Wetted materials are not FDA approved.
DO NOT return the drain sample or any water run through MTOL+ into the process stream or into any potable water supply.
b. Select a length of ¼ inch OD flexible tubing (Polyethylene or
polypropylene) long enough to directly connect the MTOL+ drain connection point to a suitable sanitary sewer drain. Avoid unnecessary bends or sags in tubing. Make sure there are no kinks, twists, breaks, or damage of any kind in tubing.
c. Using a sharp utility knife, cut the end of the tubing square.
Remove any burrs or sharp edges before inserting into fitting.
d. Place a piece of tape ½ inch from end of tube to indicate how far
the tube should be inserted.
e. Push the tube into the Quick-Connect fitting on the right facing side
of the MTOL+, as shown in Figure 4.
f. Check to confirm that tube is pushed fully against the tube stop.
Gently but firmly pull on the tube to check that it is secure.
g. Connect the other end of the tube to a suitable sanitary drain or
trench following proper plumbing techniques applicable to your facility and configuration.
3.3.4 Connecting the water intake:
a. Use MTOL+ ONLY with water. Connect water intake
ONLY to water free of corrosive chemicals or other hazardous
substances.
HF scientific accepts no responsibility for damage caused by the introduction of vapors, fluids or other materials into the instrument process stream which is not compatible with the instrument's wetted materials. A list of the wetted materials can be found in the specifications on page 2 of this manual.
b. Select a length of ¼ inch OD flexible tubing (Polyethylene or
polypropylene) long enough to directly connect the MTOL+ water inlet connection point to water sampling point. Avoid unnecessary bends or sags in tubing. Make sure there are no kinks, twists, breaks, or damage of any kind in tubing.
12
Page 19
c. Using a sharp utility knife, cut the end of the tubing square.
Remove any burrs or sharp edges before inserting into fitting.
d. Place a piece of tape ½ inch from end of tube to indicate how far
the tube should be inserted.
e. Push the tube into the Quick-Connect fitting on the left facing side
of the MTOL+, as shown in Figure 4.
f. Check to confirm that tube is pushed fully against the tube stop.
Gently but firmly pull on the tube to check that it is secure.
g. Connect the other end of the tube to the water sample source
following proper plumbing techniques applicable to your facility and configuration.
After completing water inlet and drain connections, check all tubing and connections for any leaks, weeping, breaks or defects of any kind. Repair/replace as needed.
13
Page 20
3.4 Electrical Connections
MTOL+ is electrically powered. Installation MUST be performed by licensed electrician following all instructions and comply with all local, state, federal and other governmental requirements, and with all building and construction codes and standards for all electrical connections and for installation of electrical connections to and between the instrument and any peripheral devices.
ALWAYS take proper precautions to recognize, evaluate, and control electricity hazards during installation, programming, use and service/maintenance.
Requirements:
The MTOL+ is equipped with a 100-240 VAC, 47-63 Hz switching power supply. BEFORE installing, verify that the line voltage falls within these specifications.
A breaker or method of disconnecting power must be placed within 2 meters (6 feet) of the MTOL+.
While making connections, refer to Figure 5. The MTOL+ is not supplied with a power cord.
All of the electrical connections to the instrument are made through the MTOL+ Junction Box (see Figure 5), which should be located directly under the sensor portion of the instrument.
The power cable glands will accept cable diameters from 5.8mm (.230 in) up to 10 mm (.395 in.).
All terminals are designed to accept wires in the range of 12-30 AWG. All wires should be stripped to a length of 6 mm (¼ in).
A strain relief strap is provided to reduce tension on the power terminals.
14
Page 21
The connections are labeled within the terminal box and are self­descriptive (see Figure 5).
Plugs are inserted into the alarm and 4-20mA/RS-485 cable glands when shipped, to ensure a watertight seal. These plugs should be removed and discarded when cabling to either of these connections.
Maintain watertight seal. It is the owner’s and installer’s responsibility to assure that the watertight seal is maintained after the terminal box has been wired for operation. Failure to properly tighten any of the cable glands around a cable or plug can create an electrical shock/electrocution hazard.
3.4.1 Electrical Power Connection:
Make all connections at Junction Box and securely close BEFORE making connections to power source and energizing MTOL+.
ONLY licensed electrician should perform installation. ALWAYS take proper precautions to recognize, evaluate, and control electricity hazards.
a. First, make sure that MTOL+ is completely de-energized and
disconnected from all power.
b. Perform required LOCK OUT/TAG OUT of electrical power.
c. Confirm that a breaker or method of disconnecting power is located
within 2 meters (6 feet) of the MTOL+.
d. All of the electrical connections to the instrument are made through
the MTOL+ Junction Box (see Figure 5), which should be located directly under the upper portion of the instrument.
e. Carefully open the MTOL+ Junction Box (see Figure 5) by removing
the four cover screws.
f. With the Junction Box cover removed, the connections are visible
as shown in Figure 5.
g. The power cable glands will accept cable diameters from 5.8mm
(.230 in) up to 10 mm (.395 in.). All terminals are designed to accept wires in the range of 12-30 AWG.
15
Page 22
h. Obtain and cut proper wires to make connections for your specific
installation and facility. All wires should be stripped to a length of 6 mm (¼ in).
i. The connections are labeled within the terminal box and are self-
descriptive (see Figure 5).
j. Make connections following all local, state, federal and other
governmental requirements, and all building and construction codes and standards for all electrical connections.
Earth or Ground wire must be connected to terminal on the MTOL+ marked “G”
k. A strain relief strap is provided to reduce tension on the power
terminals. Pull this tight after completing the connections.
l. Perform all checks necessary to confirm proper wiring and safe
electrical connection.
3.4.2 Terminal Block (P1) Connections:
Shut off and disconnect all electrical power BEFORE making or changing any connections. The terminal block (P1) is located close to the power connection.
The terminal block for RS-485 (Modbus), 4-20 mA and Alarms has spring loaded connections.
To open the connection, insert a 3mm (1/8 in) flat blade screwdriver into the slot right above the electrical connection. Pulling up slightly on the screwdriver opens the terminal block connection. Removing the screwdriver tightens the connection.
There are two stacked rows of connections which are labeled on the PCB. The alarm connections are on the upper row. The RS-485 and 4-20mA are on the lower row.
.
16
These connections are rated for a wire size from 12-28 AWG, either solid or stranded.
a. Alarm and I/O Connections.
The Alarm 1 and Alarm 2 relays are mechanical relays rated at 120-240 VAC 2A.
Note that the relays are labeled NO (Normally Open), NC (Normally Closed) and C (Common).
The normal condition is with power applied to the MTOL+ and in a non­alarm condition. Operation of these alarms is covered in section 4.2.4 Configuring the Alarms.
Page 23
Connections are labeled beneath the large terminal block. These connections are on the upper row.
b. RS-485/Modbus Connection
The RS-485 half-duplex (2-wire) digital interface operates with differential levels that are not susceptible to electrical interferences. Cable lengths up to 900 meters (3000 ft.) can be implemented.
The last device on each bus may require terminating with a 120-ohm resistor to eliminate signal reflection on the line.
Do not run RS-485 cables in the same conduit as power.
Ensure that power is disconnected prior to making connections to present
damage to the instrument.
Connections are labeled beneath the large terminal block. These connections are on the left side on the lower row.
c. 4-20 mA Connection:
Make sure the MTOL+ is disconnected and not powered BEFORE
connecting the 4-20 mA.
The 4-20 mA output is driven by a 24 VDC power source and can drive recorder loads up to 1000 ohms.
This 4-20 mA output is isolated from line power and earth ground.
Do not run 4-20 mA cables in the same conduit as power.
Connections are labeled beneath the large terminal block. These
connections are on the right side on the lower row.
Operation of this output is covered in section 4.2.1 Setting the 4-20 mA.
17
Page 24
3.4.3 Closing the Junction Box.
Maintain watertight seal. Always make sure that the Junction Box cover is firmly in place and securely closed. Water infiltration can create electric shock/electrocution hazard and will damage equipment.
Securely close the Junction Box by replacing the cover and all four
cover screws.
Power Connection
Terminal Block (P1) Alarm & I/O Connections
Figure 5: Electrical Connections for the MTOL+
---------------------------------------------------
18
Figure 6: Electrical Connections on Terminal Block (P1)
Page 25
4.0 Setup, Configuration and Calibration
Prior to use for the first time, the supplied desiccant pouch will need to be installed.
Refer to section 5.1 Replacing or Installing the Desiccant Pouch.
Before using your MTOL+, you will need to familiarize yourself with the instrument’s operating modes and functions, and perform some set up, configuration and calibration.
The MTOL+ allows for the measurement of the turbidity of process water on-line. The turbidity of the process water is usually reported in
Nephelometric Turbidity Units (NTU), but may be reported in Formazine Nephelometric Units (FNU).
Readings above 100 NTU are outside the pre-configured range of this instrument. Readings above 110 NTU will cause the display to flash indicating an over range condition.
For a unit calibrated in the 0-10 NTU Range… Readings above 10 NTU are outside range of this instrument. Readings above 11 NTU will cause the display to flash indicating an over range condition.
For a unit calibrated in the 0-1000 NTU Range… Readings above 1000 NTU are outside the range of this instrument. Readings above 1100 NTU will cause the display to flash indicating an over range condition.
During normal operation, the instrument will have the arrow beside AUTO highlighted with the current scale displayed on the lower row of the display and the measured reading on the upper row of the display (see illustration below).
4.1 Modes of Operation
There are four modes of operation, AUTO, CAL, CONFIG & HOLD. To change between AUTO, CAL, HOLD and CONFIG use the and then
the  button to select.
a. AUTO mode is the normal automatic operation. This is the default
mode when power is applied or restored. The other three modes of operation are limited to 15 minutes with no key presses after which they will return to AUTO made operation.
19
Page 26
b. CAL is the calibration mode. All reading outputs and alarms are
held while in this mode of operation. The instrument was calibrated and tested prior to leaving the factory. Therefore, it is possible to use the instrument directly out of the box. Under normal conditions, re-calibration is required once every three months. Quarterly calibration ensures performance within accuracy specifications.
Enter the CAL mode of the instrument by pressing the or
buttons until the arrow beside CAL is illuminated, then
press the
Use the and  buttons to move forward and backward through the calibration points.
To exit the CAL mode start a calibration point after the countdown starts press or . When the screen shows
ABRT press the
Relay contacts and the 4-20mA will be frozen (held at the current state) while the instrument is in the calibration mode.
While in the calibration mode, the instrument has a time-out feature that automatically returns the system operation to the
AUTO mode after a fifteen (15) minute period of inactivity.
c. CONFIG is the configuration mode. In CONFIG mode, you can
customize the instrument according to needs and preferred operation at any time during normal operation. The CONFIG mode has been split into sub-menus to facilitate instrument configuration. This is also where logged files and calibration logs can be downloaded and where new software, when available, can be uploaded.
button.
button.
20
Enter the CONFIG mode of the instrument by pressing the
 or buttons until the arrow beside CONFIG is
illuminated, then press the
To exit the CONFIG mode at any time, press the
The and buttons are used to move forward and
backward through the menus.
All reading outputs and alarms are held while in this mode of operation. Relay contacts and the 4-20mA will be frozen (held at the current state) while the instrument is in the configuration mode. While in the configuration mode, the instrument has a time-out feature that automatically returns the system operation to the AUTO mode after a fifteen (15) minute period.
button.
button.
Page 27
Shown below is a Flow Chart depicting the options under the configuration operation:
d. HOLD mode is intended for servicing the instrument and holds the
outputs and alarms. During this mode the 4-20mA and alarms are frozen. This mode can be used to ensure that no changes accidentally are made to the instrument. This mode will time out after 15 minutes and revert back to AUTO mode.
Enter the HOLD mode of the instrument by pressing the or  buttons until the arrow beside HOLD is illuminated,
then press the
Figure 6 MTOL+ Flow Chart
button.
Once selected the letters above will change from OFF to
21
ON.
Page 28
The HOLD mode can be used during any maintenance functions such as changing the cuvette or desiccant without affecting alarms or the 4-20mA output signals.
CAL, CONFIG & HOLD are limited to 15 minutes after which time they automatically revert to AUTO mode.
4.2 Configuring the MTOL+
The MTOL+ has been designed to provide the ability to customize the instrument according to needs at any time during normal operation. The CONFIG mode has been split into sub-menus to facilitate instrument configuration. This section describes how to use each of the sub-menus to configure the instrument. In the sections below each sub-menu under the CONFIG mode is discussed in the order that they are reached by circling through the CONFIG mode, as identified in the flow chart located previously.
Turn Power “ON:” After confirming that the MTOL+ is safely and properly mounted, connected and installed, turn the instrument ON by connecting or turning on the power.
22
To enter the CONFIG Mode: Enter the CONFIG mode by pressing the or buttons until the arrow beside CONFIG is illuminated, then press the
button.
To exit the CONFIG Mode: To exit the CONFIG mode at any time, press the
button. Use the and buttons to move forward and backward
through the menus.
4.2.1 Configuring the 4-20mA Output
The 4mA can be set higher than the 20mA level to invert the output current if required. This may be required to control a dosing pump.
Page 29
See 4.2.13 to calibrate the 4-20mA output.
Enable 4-20mA – Enable or Disable 4-20mA Output: Enter the CONFIG mode by pressing the and buttons until the arrow beside
CONFIG is displayed, then press the
appear only if the 4-20mA is enabled. If disabled skip down to 4.2.2
This will be the first menu shown unless the option flow switch was
ordered. Use the or buttons to select whether the 4-20mA will be On or OFF. Press the  button to accept the setting and move to the next
menu or the
button to exit and return to AUTO.
button. The next three menus
Setting 4mA - Setting the 4mA:
The first prompt will be the turbidity limit assigned to the 4mA output level:
Select the turbidity level to assign to the 4MA using the and buttons.
The factory setting is 0.02 NTU.
Setting 20mA - Setting the 20mA:
Next, select the turbidity level to assign to the 20mA output level using the
and buttons. The factory setting is 100.00 NTU.
23
20MA is displayed on the lower row of the LCD display. Select the
turbidity level to assign to the 20MA using the  and buttons.
Once the desired level has been set, press the button to accept the
setting and move to the next menu or the
button to exit and return to
Page 30
AUTO.
Error Level - Configuring the Error Level:
In case of an error in the MTOL+, the 4-20 mA reading can be used to indicate a problem by sending the current to either 4.00 mA, 2.00 mA.
0.00 mA, or OFF.
The factory default setting is OFF and the 4-20mA is unaffected by any error condition.
To change the Error Level, enter the CONFIG mode by pressing the and buttons until the arrow beside CONFIG is illuminated, then press
the
button.
Press the  button to move to the next menu until ERLV is displayed.
Select the desired ERLV by using the and buttons then press the 
button to accept the setting and move to the next menu or the exit and return to AUTO.
4.2.2 Baud Rate - Selecting the RS–485 Baud Rate
Select the desired baud rate (1200, 2400, 4800, 9600, or 19200) for operation of the I/O port.
To change the Baud Rate: Enter the CONFIG mode by pressing the and buttons until the arrow beside CONFIG is illuminated, then press
the
button.
Press the  button to move to the next menu until BAUD is displayed.
Select the desired BAUD rate by pressing the or buttons to change
the displayed baud rate.
button to
24
Page 31
Press the  button to accept the setting and move to the next menu or the
button to exit and return to AUTO.
4.2.3 Modbus Address – Address Selection
Select the Modbus address of the MTOL+ from 1 to 255. For more information on Modbus refer to the Modbus Manuals Cat #29291. This manual can also be downloaded for no charge at www.hfscientic.com.
To change the Modbus Address: Enter the CONFIG mode by pressing the and buttons until the arrow beside CONFIG is illuminated, then
press the
Press the  button to move to the next menu until ADDR is displayed.
Select the address using or buttons.
button.
Press the  button to accept the setting and move to the next menu or the
button to exit and return to AUTO.
4.2.4 Configuring the Alarms
Two relays are provided that are designed to operate as two independent programmable alarms. Three types of information must be input to fully program each alarm:
a. The alarm function (HI, LO, OFF or Error) b. The alarm set point (level at which the alarm activates) c. The delay time for the alarm: The time that the set point must be exceeded prior to alarm activation and the time before resetting the alarm (prevents chatter in the relay).
a. Alarm Function: The alarms can either be turned OFF or
programmed to operate in one of three different manners:
25
HI alarm: the relay changes state when the measured turbidity level is higher than the programmed alarm level for a prescribed time.
LO alarm: the relay changes state when the measured turbidity level is lower than the programmed alarm level for a prescribed time.
Error: the relay changes state when a system error occurs. If a
Page 32
system error occurs a message will appear on the lower row of the screen describing the problem.
b. Alarm Set Point: The level at which an alarm activates is called
the alarm set point. On the instrument, the alarm set point is designated as “S/P”. The set point is adjustable to any valid turbidity level over the range of the instrument in steps of 0.01 NTU.
c. Alarm Delay Time: The alarm delay times are used to prevent
tripping the alarm when the turbidity level is close to the set point. The function of the delay times is as follows:
Delay On: The turbidity level must exceed the alarm set point continuously for at least this number of seconds before the alarm activates.
If the “delay on” time is set at 5 seconds and the process
turbidity exceeds the set point continuously for only 4 seconds, the alarm will not be activated. However, if process turbidity exceeds the set point continuously for 5 seconds or more; the instrument will activate the alarm.
Delay Off: The turbidity level must not exceed the alarm set point continuously for at least this number of seconds prior to deactivation of the alarm.
If the “delay off” time is set to 5 seconds and the process has
exited out of the alarm condition, the alarm will be reset only if the process is out of the alarm condition for a continuous 5 seconds. Otherwise, the instrument will still signal an alarm condition.
4.2.4.1 Alarm 1 - Configuring Alarm 1
Alarm 1 Function: The ALM1 is displayed and the display indicates the current function of Alarm 1 (HI, LO, OFF or Error).
Use the or buttons to cycle through and select the desired function.
Press the  button to accept it and move to the next menu.
If the alarm was turned OFF a prompt will appear to set up Alarm 2
(go to section 4.2.10.2). If, on the other hand, one of the other HI or LO was selected, a prompt will appear to set the delay times.
Alarm 1 Set Point: Set Point is indicated by “S/P” shown on the lower row of the display, and is used to select the set point for this alarm.
Select the desired alarm level by using the and buttons.
26
Page 33
Once the desired set point has been set, press the  button to accept it and move to the next menu.
Alarm 1 Delay Times:
1. Delay On: The following display will appear to allow you to
select the number of seconds currently set for the “delay on” time.
The current selected number of seconds will be shown. Select
the desired number of seconds for the “delay on” time for this alarm using the and buttons. Once the desired delay time has been set, press the  button to accept it and move to the next menu.
2. Delay Off: Next, the following display will appear to select the
number of seconds currently set for the “delay off” time.
The current selected number of seconds will be shown. Select
the desired delay off time for this alarm using the and buttons. Once the desired delay time has been set, press the  button to accept it and move to the next menu.
After the settings for Alarm 1 have been completed, prompts will
allow for the set-up of the information on Alarm 2.
4.2.4.2 Alarm 2 - Configuring Alarm 2
Repeat the procedure listed in section 4.2.10.1 to set up the parameters for alarm 2.
27
Page 34
4.2.5 Offset – Offset Reading Adjustment
A reading offset can be made to allow for the MTOL+ to agree with another instrument if desired.
This range is limited to ± 1NTU.
For more information refer to section 4.3.6.
To change the Offset: Enter the CONFIG mode by pressing the and
buttons until the arrow beside CONFIG is illuminated, then press the
button.
Press the  button to move to the next menu until OFST is displayed.
Select the offset using the or buttons to turn the function On.
When the OFST is turned ON it will stay illuminated in AUTO
as a reminder that it is being used.
Select the offset using the or buttons.
Press the  button to accept the setting and move to the next menu or the
button to exit and return to AUTO.
4.2.6 Access Code - Setting Security
The MTOL+ is equipped with a security access feature that can be activated. If this option is turned ON, the user is required to input an access code into the instrument to get to any mode other than AUTO. A three digit code is used.
HF Scientific highly recommends the use of a security access code in order to help prevent unauthorized access to the system and data.
When the CODE is turned On the icon will be illuminated in
AUTO as a reminder that an access code is being used.
Setting Security on/off: Enter the CONFIG mode by pressing the and
buttons until the arrow beside CONFIG is illuminated, then press the
button.
28
Press the  button to move to the next menu until CODE is displayed.
Use the or buttons to select the setting. (On or OFF).
Page 35
Setting Security Code: With the security feature enabled (“ON”), press the  button. The screen shown in the illustration below will appear:
The 3 digit security code is set one digit at a time. Using the or buttons, select each digit. The flashing digit is the number that is currently being adjusted.
Once a digit is selected, press the  button to move on to the next number.
Now enter the second number in the code. Proceed as with the first number followed by .
Repeat the process for the third, and finish with the  button.
When the code is turned On, the code must be used to exit
AUTO
If a valid access code has been selected, the instrument will be directed to the calibration mode. If the wrong access code is selected, the instrument will return to the AUTO mode.
If the access code was set On from AUTO press or . The next screen will allow entering the three digit code into the instrument. The security code is entered one digit at a time using the or to select the digit and
 to advance to the next digit. Accept the code by pressing the
4.2.7 Signal Averaging - Setting Speed of Response
The averaging can be used to smooth out the response to spikes or unstable readings that may be caused by bubbles or small debris.
The response is done by averaging readings. Both displayed and output values of NTU are affected in this menu. The default setting is 1; however,
button.
29
Page 36
up to 60 readings can be averaged.
To avoid reading air and other anomalies, select a higher averaging.
Select the lowest averaging where monitoring of rapid changes is needed.
These are one second intervals so the maximum averaging is 60 seconds.
The factory default is 10.
To change the Averaging: Enter the CONFIG mode by pressing the and buttons until the arrow beside CONFIG is illuminated, then press
the
button.
Press the  button to move to the next menu until RESP is displayed.
Use the or buttons to select the setting for the speed of response
desired.
Press the  button to accept the setting and move to the next menu or the
button to exit and return to AUTO.
4.2.8 Display Resolution - Setting Resolution
The instrument is equipped with the ability to display several levels of resolution. The instrument can display up to four digits to the right of the decimal place for turbidity readings below 10 NTU. The default setting is
000.00 If the last digit or two is not stable, adjust the resolution to hide these insignificant digits.
To change the Resolution: Enter the CONFIG mode by pressing the or buttons until the arrow beside CONFIG is illuminated, then press the
button.
Press the  button to move to the next menu until RES is displayed, then press the
Change the resolution by pressing the or button.
When the desired digit resolution has been selected, press the  button to
accept it and move to the next menu or the AUTO.
button.
button to exit and return to
30
Page 37
4.2.9 LCD Brightness - Setting the LCD backlight brightness
The LCD backlight brightness may need to be adjusted. This is of particular interest if multiple instruments are located in the same area and it is desired for the entire group to have the same appearance. Ten levels are available. The default brightness is 8.
To adjust Backlight Brightness: Enter the CONFIG mode by pressing the and buttons until the arrow beside CONFIG is illuminated, then
press the
Press the  button to move to the next menu until BRT is displayed.
Change the brightness by pressing the or button.
button.
When the desired brightness has been selected, press the  button to accept it and move to the next menu or the
AUTO.
4.2.10 Display Units - Setting Displayed Units
The most common unit is NTU (Nephelometric Turbidity Units) however the instrument can display in FNU (Formazine Nephelometric Units).
All instruments are shipped from the factory set in NTU mode.
Setting/Changing Units: Enter the CONFIG mode by pressing the and buttons until the arrow beside CONFIG is illuminated, then press
the
button.
Press the button to move to the next menu until UNIT is displayed.
Use the or buttons to select the setting NTU or FNU.
Press the  button to accept it and move to the next menu or the
button to exit and return to
31
Page 38
button to exit and return to AUTO.
4.2.11 Ultrasonic Cleaning - Cleaning Function
This allows for a selection menu to turn off the ultrasonic cleaning function if desired. The default mode is On.
Setting the ultrasonic cleaning: Enter the CONFIG mode by pressing the and buttons until the arrow beside CONFIG is illuminated, then
press the
Press the  button to move to the next menu until CLN is displayed.
Select On or OFF using the and buttons.
button.
Press the  button to accept I the setting and move to the next menu or the
button to exit and return to AUTO.
4.2.12 Modbus Parity & Modbus Stop Bits - RS485 Parameters
These two menus will be used to set the balance of the RS-485 parameters.
The default is no Parity (nOnE), 1 Stop Bit.
Modbus Parity – Setting the Modbus Parity: Enter the CONFIG mode by pressing the and buttons until the arrow beside CONFIG is
illuminated, then press the
Press the  button to move to the next menu until PRTY is displayed.
Make selections using the and buttons
Press the  button to accept it and move to the setting stop bits
button.
32
Modbus Stop Bits – Setting the Modbus Stop Bits: Enter the CONFIG mode by pressing the and buttons until the arrow beside CONFIG is
illuminated, then press the
button.
Page 39
Press the  button to move to the next menu until STOP is displayed.
Make selections using the and buttons
Press the  button to accept it and move to the next menu or the
button to exit and return to AUTO.
4.2.13 4mA and 20mA Adjustment
If the 4-20 mA setting is turned On, two menus will appear one for each 4 and 20mA adjustment.
4mA Adjustment:
The first menu outputs a constant 4mA while allowing for a small amount of adjustment. This adjustment will allow the operator to make the MTOL+ agree with a PLC or SCADA system. The adjustment limits are ± 200 counts or about ± 0.2 mA.
This setting will be slightly different on each instrument as
each MTOL+ will be factory set to 4.00mA. Press the  button when adjustments are complete to save this setting and move on to the 20mA adjustment.
Setting the 4mA Levels: Enter the CONFIG mode by pressing the and buttons until the arrow beside CONFIG is illuminated, then press
the
button.
33
Press the  button to move to the next menu until 4MA is displayed.
Adjust setting using the and buttons.
Press the  button when adjustments are complete to save this setting
and move on to the 20mA adjustment.
20mA Adjustment:
This menu operates similar to the previous menu. This menu outputs a
constant 20 mA while allowing for a small amount of adjustment. The adjustment limits are ± 1000 counts or about ± 1 mA.
Page 40
This setting will be slightly different on each instrument as each MTOL+
will be factory set to 20.00mA.
Setting the 20mA Levels: Enter the CONFIG mode by pressing the and buttons until the arrow beside CONFIG is illuminated, then press
the
button.
Press the  button to move to the next menu until 20MA is displayed.
Adjust setting using the and buttons.
Press the  button to accept the setting and move to the next menu or the
button to exit and return to AUTO.
4.2.14 Instrument ID – Instrument Unique Number Identification
This menu provides the instrument with a unique ID up to 4 digits. This ID
can be used when logged data is downloaded. This enables the data to be associated with a particular instrument or location.
Setting the instrument ID: Enter the CONFIG mode by pressing the and buttons until the arrow beside CONFIG is illuminated, then press
the
button.
Press the  button to move to the next menu until ID is displayed.
Use the and buttons to select the value of the first number. Use the
 to select the next number and repeat the process for all numbers. Scroll
through the numbers to the next menu.
4.2.15 Logging Interval – Setting the interval for logged files
This menu determines how often readings are logged. The setting is in
minutes. This can be set from 1 minute to 60 minutes.
34
Setting the logging interval: Enter the CONFIG mode by pressing the
Page 41
and buttons until the arrow beside CONFIG is illuminated, then press
the
button.
Press the  button to move to the next menu until INVL is displayed.
The value can be adjusted by using the and buttons.
Once the value is set, press the button to accept it and move to the next
menu or the
4.2.16 Download Logged Files – Select to Download
This menu is used to download CSV files to a flash drive. The selections
are No download (NO), Calibration file only (CAL), All Files (ALL) or individual month (Jan, Feb…). Once the data stick is inserted in the USB
slot on the right side of the instrument push the is inserted you will get an Error (Err). See section 5.4.2 File Downloads for more information. The upper display will flash while the download is taking place and show dOnE when complete.
Setting the data download: Enter the CONFIG mode by pressing the and buttons until the arrow beside CONFIG is illuminated, then press
the
button.
Press the  button to move to the next menu until DOWN is displayed.
The selection can be adjusted by using the and buttons.
button to exit and return to AUTO.

button. If no data stick
35
Once the selection is made press the  button to accept it and move to
the next menu or the
4.2.17 Date and Time – Setting of Instrument Time & Date
The MTOL+ has a displayed date (month / day / year) and time (hour /
minute / Daylight Savings) that must be set upon initial installation/setup. The next five menus are used for this setup.
Setting Month: Enter the CONFIG mode by pressing the and buttons until the arrow beside CONFIG is illuminated, then press the
button.
Press the  button to move to the next menu until MNTH is displayed,
button to exit and return to AUTO.
Page 42
then press the
Change the month by pressing the or button.
When the desired month has been selected, press the  button to accept
it and move to the next menu or the
Setting Day: Enter the CONFIG mode by pressing the and buttons until the arrow beside CONFIG is illuminated, then press the
Press the  button to move to the next menu until DAY is displayed, then
press the
Change the day by pressing the or button.
button.
button.
button to exit and return to AUTO.
button.
When the desired day has been selected, press the  button to accept it
and move to the next menu or the
Setting Year: Enter the CONFIG mode by pressing the and buttons until the arrow beside CONFIG is illuminated, then press the
Press the  button to move to the next menu until YEAR is displayed,
then press the
Change the year by pressing the or button.
button.
button to exit and return to AUTO.
button.
36
When the desired year has been selected, press the  button to accept it
Page 43
and move to the next menu or the
Setting Hour: Enter the CONFIG mode by pressing the and buttons until the arrow beside CONFIG is illuminated, then press the
Press the  button to move to the next menu until HOUR is displayed,
then press the
Change the hour by pressing the or button.
When the desired hour has been selected, press the  button to accept it
and move to the next menu or the
button.
button to exit and return to AUTO.
button.
button to exit and return to AUTO.
Setting Minute: Enter the CONFIG mode by pressing the and
buttons until the arrow beside CONFIG is illuminated, then press the button.
Press the  button to move to the next menu until MIN is displayed, then
press the
Change the minute by pressing the or button.
When the desired minute has been selected, press the  button to accept
it and move to the next menu or the
Daylight Savings Time: DST is valid for use in the USA, Canada, and other countries or locations that follow the US Energy Savings Act of
2005.
button.
button to exit and return to AUTO.
37
Enter the CONFIG mode by pressing the and buttons until the arrow
beside CONFIG is illuminated, then press the
Press the  button to move to the next menu until DST is displayed, then
button.
Page 44
press the
Change the Daylight Savings Time by pressing the or button until
ON is displayed.
Press the  button to accept it and move to the next menu or the
button to exit and return to AUTO.
4.2.18 Update Software - Install Latest Software or Change Range
Enter the CONFIG mode by pressing the and buttons until the arrow
beside CONFIG is illuminated, then press the
Press the  button to move to the next menu until UPDT is displayed,
then press the
The software version can be determined by pressing and holding down
the and buttons for about 1 second. The first number shown is the version and second number displayed is the model number. This can be performed anytime during AUTO Mode.
Software updates can only be made if a USB flash drive is installed with a
newer version of software than the one currently installed in the MTOL+. If either a flash drive is not installed or a newer version of software is not detected the following screen will appear.
Range changes can be made with either the current or later versions.
These other ranges were included on a flash drive included with the instrument when new.
Once a flash drive with valid files are found, select the desired range to
update using the and buttons.
button.
button.
button.
38
Page 45
While the system is updating, the word “busy” appears as it performs a
self-check of the software installation.
If the MTOL+ finds a flash drive with a previous version of software it will
show the word FILE. You will need an updated file on the flash drive.
Once installed the system will automatically turn off and restart. The entire
process takes about 60 seconds to complete.
Be certain to check the configuration as changes may have taken place either with new software or a new range.
39
Page 46
4.3 Calibrating the MTOL+, CAL Mode
The instrument was calibrated for the 0-100 NTU range and tested prior to
leaving the factory. Therefore, it is possible to use the instrument directly out of the box. Under normal conditions, re-calibration is required once every three months. Quarterly calibration ensures performance within accuracy specifications.
The EPA and ISO recommend that on-line turbidimeters be calibrated with a primary standard at least once every three
months if they are to be used for reporting purposes.
Relay contacts and the 4-20mA will be frozen (held at the current state)
while the instrument is in the calibration mode. While in the calibration mode, the instrument has a time-out feature that automatically returns the system operation to the AUTO mode after a fifteen (15) minute period of inactivity.
4.3.1 Calibration Standards
If the MTOL+ will be used over the entire range of .02 to 100 NTU a
complete calibration as described below will be required, but you must retain the same calibration schedule as described above.
If instrument accuracy is only required below 10 NTU, such as for potable
water, a calibration may be performed using only a 10 NTU and a 0.02 NTU standard.
We recommend that the following materials be used during calibration to
achieve the full-scale accuracy stated in this manual:
100 NTU
ProCal
ProCal
ProCalProCal
Calibration Standard available from HF scientific
40
10.0 NTU
0.02 NTU
4.3.1.1 Alternate Ranges
NTU the instrument must be recalibrated.
The calibration values will be 10 NTU, 1 NTU and .02 NTU for 0-10 NTU
range.
The calibration values will be 1000 NTU, 10 NTU and .02 NTU for 0-1000
NTU range.
Calibration kits are available for the optional ranges.
Make changes as needed to the instructions in section 4.3.3 below.
ProCal
ProCal
ProCalProCal ProCal
ProCal
ProCalProCal
If the range is changed from the factory setting of 0-100
Calibration Standard available from HF scientific
Calibration Standard available from HF scientific
Page 47
It is well known that diluted Formazine is unstable. If Formazine is used to calibrate the instrument, ensure that a fresh stock suspension of Formazine is used to achieve the accuracy quoted for the instrument.
A Formazine Stock Solution Kit is available from HF scientific (Catalog No.
50040).
The HF scientific ProCal, primary calibration standards (refer to section
11.0 Accessories and Replacement Parts List) are more stable than Formazine and have a minimum shelf life of 12 months. Prior to recalibration, review the expiration dates, to ensure that the standards have not expired.
4.3.2 Indexing Calibration Cuvettes
To achieve the greatest accuracy, and account for normal scratches and aberrations in cuvette glass when calibrating, HF Scientific recommends indexing the cuvettes.
Standards and standard kits purchased from HF Scientific are supplied with indexing rings. Complete instructions regarding how to index the cuvettes are included in the calibration kits.
The following steps allow repeatable indexing of calibration standards:
1. With the instrument operating in AUTO mode insert
the standard into the optical well.
2. Slowly rotate the standard, inside the optical well, one
complete revolution (360º). While rotating the standard slowly, observe the measured turbidity and locate the position of the cuvette having the lowest reading.
3. With the calibration standard positioned at the location having the
lowest turbidity reading, install the Indexing Ring over the cap on the standard so that the pointer of the Indexing Ring faces directly forward.
When using the standards in the future, always insert the standard so that the pointer of the indexing ring faces forward. Slowly rotate the standard back and forth about 5° to find the lowest point. The standard is now indexed and ready for use.
41
4.3.3 Calibration Procedures
It is recommended that the measurement chamber be kept covered during the calibration period and that the flow through cuvette be replaced immediately after the calibration to prevent premature saturation
Page 48
of the desiccant.
1. Select the calibration function of the instrument by pressing the once. The arrow beside CAL will be illuminated on the display. The lower display shows alternating 100 (the value of the standard that is requested)
and
. The upper display shows the real-time reading to allow the standard
to be indexed. Refer to section 4.3.2 for information on indexing cuvettes.
2. Remove the flow through unit.
3. Insert the requested 100 NTU standard. Index the standard to the lowest value on the upper display.
4. To calibrate for 10 NTU and below only, press the button to bypass the 100 NTU and proceed to step 7.
5. Press the
6. The lower display will count down from 30.
7. When complete, the lower display will now change to show alternating 10 and
, requesting the 10.0 NTU standard.
8. If the alternating 10 and  is not displayed, push theuntil this display is shown.
9. Insert the requested 10.0 NTU standard. Index the standard to the lowest value on the upper display.
10. Press the
11. The lower display will count down from 60.
12. When complete, the lower display will now change to show alternating 02 and
, requesting the 0.02 NTU standard.
button to initialize the calibration.
button to initialize the calibration.
button
42
Page 49
13. Insert the requested 0.02 NTU standard. Index the standard to the lowest value on the upper display.
14. Press the
15. The lower display will count down from 30.
16. When complete the instrument will show CAL Good. Press the accept. The display will briefly show STRT and then go to AUTO.
4.3.4 Abort Calibration
If at any time during the calibration process, you can abort the calibration
by pushing theuntil the screen displays ABRT then press the The instrument returns to AUTO operation.
The instrument will save and use any points that were calibrated
successfully. It should be noted that the calibration log will report any missed calibration values or failed calibrations.
4.3.5 Calibration Failure
The MTOL+ will test the validity of the calibration and only good
calibrations will be saved for use.
If the calibration fails the message below will show. Press the
accept.
The instrument will revert to the last valid calibration. Check the standards
and recalibrate if desired.
The failed calibration will be discarded, however a record of a failed
calibration with the time and date is stored.
4.3.6 Instrument Offset
Refer to section 4.2.5
It is possible to use an offset factor in the instrument rather than
performing a physical calibration.
button to initialize the calibration.
button to
button.
button to
43
This procedure is not recommended in lieu of regular instrument
calibration but it can be used to make the MTOL+ agree with another instrument.
Page 50
This calibration technique will make the instrument accurate only at
turbidity levels in the immediate vicinity of the grab sample and not in the full range of the instrument.
Note that the OFFSET icon will be illuminated whenever an offset is used.
The maximum offset is ± 1.00 NTU.
If the instrument variation is greater than 1 NTU a full calibration is
recommended.
Below is the offset procedure:
1. Collect a grab sample of the process water that is being monitored by the MTOL+ and record the turbidity reported by the instrument.
2. Take the grab sample and measure its turbidity using a laboratory turbidimeter (contact the HF scientific customer services department for examples of laboratory turbidimeters).
3. Compare the turbidity reported by the MTOL+ to that obtained in the laboratory. If the readings are very close, then no offset adjustment or calibration is required and the procedure may be stopped at this step. However, if the readings are substantially different (but less than 1 NTU), continue on in this procedure to utilize the offset option to improve the turbidity reading of the MTOL+ so that it will agree with the laboratory reading between calibrations.
4. Select the offset function of the MTOL+ by pressing the button until the arrow beside CONFIG is illuminated on the display then
press the
5. Push the  button until OFST is displayed on the lower row.
6. The upper row of the display will indicate the operational status of
the offset function (On or OFF). Change this status to On by using the and buttons. Then press the  button to accept it and move to the adjustment menu
7. This will add or subtract the value of the offset to the measured NTU value. As an example if the MTOL+ measures the process at
0.36 NTU but the laboratory instrument read the sample at 0.12 NTU, adding an offset of –0.24 would result in the MTOL+ displaying 0.12 NTU.
8. Select the desired offset level using the and buttons. Once the desired level has been set, press the
9. This completes the offset configuration.
10. The instrument will return to the AUTO mode with the offset
mathematically added to the reading.
button.
button to accept it.
44
Page 51
5.0 Operation
Prior to use for the first time, the supplied desiccant pouch will need to
be installed.
5.1 Installing or Replacing the Desiccant Pouch
The MTOL+ continuously checks the condition of the desiccant. When the desiccant gets in such a condition that it may cause problems, the instrument will display DESC on the lower portion of the display to indicate the presence of humidity.
Proper use of the supplied desiccant is essential in maintaining the performance of the instrument. The desiccant has been designed to have a long life; however, replacement of the desiccant pouch will be required from time to time.
It is essential that the enclosure seal on the instrument base be maintained to ensure adequate desiccant life. Inspect the seal each time the desiccant pouch is replaced. Replace or reseat the seal if it is found to be defective.
The desiccant should be replaced when the instrument displays DESC. A new sealed desiccant pouch and indicator card are available from HF scientific part #21555R.
1. To initially install or remove the old desiccant, simply unscrew the four corner thumbscrews and remove the electronics half of the instrument.
2. If replacing an old desiccant pouch, remove the used desiccant pouch from the instrument.
3. Open the bag protecting the new desiccant pouch.
4. Insert the new desiccant pouch in the instrument base assembly.
5. To speed up the recognition, of the new desiccant, by the instrument, it may be necessary to reset the instrument by disconnecting the sensor interconnect cable for 2 seconds and then reconnecting it.
45
Page 52
Figure 8: Installing the desiccant
Once the bag is opened, install the desiccant pouch immediately to prevent premature degradation of the desiccant.
46
5.2 Starting the System
The following steps describe how to measure the turbidity of a sample using this instrument:
1. Apply power to the instrument and allow the unit to warm up (typically
45 minutes – 1 hour on initial commissioning).
2. Check the cuvette to make sure it is tight before starting flow. The
cuvette should be handled at the top ½” to prevent fingerprints from interfering with readings.
3. After the initial warm-up period and once a continuous process stream
is flowing through the instrument, the instrument will display the measured turbidity level of the sample by displaying it on the LCD screen. In addition, the equivalent signal is provided on the analog (4­20 mA) output, or the digital output, depending on the options selected.
4. Initial settings like Date and Time (section 4.2.17) must be set now.
Page 53
5.3 Routine Measurement: AUTO Mode
If power is lost at any time, once power is restored, the instrument will
After the instrument has warmed up and the instrument is operating in AUTO mode, readings will be updated (posted) once per second. These updates occur to the screen, Modbus, 4-20mA and can be logged internal to the instrument. These updates also affect the operation of the alarms. Please note that these updates are all affected by the setting of the Signal Averaging (see section 4.2.7) and possibly Offset (see section 4.2.5).
restart and the instrument will continue in AUTO operation with the same settings. There may be some slight instability if the instrument cooled off. If the onboard data logging was used it will record 0.00NTU during the period that instrument is not running.
5.4 Data Retrieval
Data is available in several forms. It can be viewed on the screen, logged in the instrument, read via Modbus or output the 4-20 mA.
5.4.1 USB Connection
The USB connection can be used to upload new software or download logged data and calibration reports via a USB flash drive.
47
Software uploads can be made by loading the software onto a USB flash drive and inserting it into the USB connector. The update will not affect configurations or stored data, however it always advisable to check the configuration as menus may change. Check our website:
www.hfscientific.com for software upgrades. The instructions are in
section 4.2.18 of this manual.
5.4.2 File Downloads
For instructions see 4.2.16
All files download in a CSV (comma separated value) file format. These
are easily imported directly into most spreadsheet and database programs.
Page 54
The main data includes the Date, Time, Reading, Units, Alarm 1 and Alarm 2. The instrument ID is part of the file name.
Some sample data is shown below.
Alarm
Date Time Reading Units 6/9/2017 16:01 0.024 NTU Off Off 6/9/2017 17:02 0.023 NTU Off Off 6/9/2017 17:03 0.022 NTU Off Off 6/9/2017 17:04 0.022 NTU Off Off 6/9/2017 17:05 6/9/2017 17:06 6/9/2017 17:07 0.022 FNU Off Off
The file structure of the logged data will be in Year/Month/Day. At midnight of each day the current day’s file is closed and a new day’s file is opened.
0.02 NTU Off Off
0.02 FNU Off Off
1 Alarm 2
For moisture tightness and humidity control the supplied USB plug cover must be kept installed firmly when not being used. Failure to correctly install the plug will affect desiccant life.
5.4.3 Modbus Communication
Modbus protocol communication is operational on all models. The Modbus information is covered in a separate manual (Catalog # 29291). This manual is also available as a free download from our website at
www.hfscientific.com.
48
Page 55
5.5 Patented Ultrasonic Cleaning 1
This feature is used to continuously clean the flow through cuvette. It is not intended to clean cuvettes that are already dirty, or replace manual cleaning entirely. See section 5.7 for instructions on cuvette cleaning.
The system will increase the time between cleanings dramatically. Please note that the system requires the use of a special cuvette which is included with the instrument. This cuvette must be used for the system to operate correctly.
The system works by sending an ultrasonic frequency through spring connections into a piezo transducer bonded to the bottom of a flow through cuvette (refer to figure 10).
The system can detect that an incorrect cuvette is installed, an error has occurred in the transducer or the transducer is not making contact with the spring connections. This error is indicated by CLN being posted to the lower screen. Since this is an error condition, this may affect the 4-20 mA and alarms depending in the setting of the ERLV (4-20 mA) and if an alarm is set up to Error.
If the correct cuvette is installed, and the error is still posted, try rotating the flow through assembly slightly to improve the connection to the ultrasonic transducer. If this fails to work, the cuvette may have to be replaced (Catalog #24166S). The detection for this cuvette only operates in AUTO mode. If the system is operating correctly AUTO will flash on the display.
The connection can be improved with use of a small film of an anti­oxidant compound such as OX-GARD ™ made by GB Electrical Inc. This product is available in the electrical section at most hardware stores.
5.6 Condensation on Cuvette
The Vapor Purge system prevents moisture, due to condensation, from occurring inside the instrument. To ensure proper performance, the desiccant and seals must be maintained in good condition. If these are distorted or worn they need to be replaced.
Once the flow through is removed from the instrument, it is possible to have condensate form on the cuvette. It is not desirable to bring additional moisture into the instrument as this will lower the desiccant life. For this reason any condensate should be wiped from the cuvette before reinstalling the flow through unit.
If condensate forms on the cuvette, use the following procedure:
1
U.S. Patent No. 7,808,642
49
Page 56
1. Shut off the flow with the provided ratcheting clamp.
2. Empty the water out of the cuvette to lower the cold mass.
3. If clean warm water is available it can be poured into the cuvette.
4. Dry and clean the exterior of the cuvette.
5. Ensure the cuvette is installed into the flow through tightly.
6. Reinstall the flow through unit.
7. Restart the sample water flow.
Once the flow through assembly is inserted into the instrument, the Vapor Purge system will prevent further condensate from forming inside of the instrument no matter the outside environment or the temperature of the sample water.
All seals are important for proper operation and to extend desiccant life. These seals must be in place and maintained. Replace any missing or distorted seals or worn call HF scientific technical support (section 6.4) Three important seals that are often overlooked are the O-ring under the flow head, the orange enclosure seal and the USB plug cover.
Figure 9: Instrument Seals
50
The cuvette must be completely dry before it is inserted into the sensor. If there is any visible moisture present on the cuvette or transducer, there is a great risk of damaging the sensor electronics and the transducer. Be sure to clean and dry the cuvette completely just before inserting it into the sensor.
Page 57
The Vapor Purge system can NOT remove large droplets of water, only
residual moisture.
For the Vapor Purge system to function properly, all instrument seals must be maintained and the desiccant pack must be in good condition (no DESC display).
SENSOR (TOP VIEW)
FLOW THROUGH
CUVETTE
ULTRASONIC
SPRING
CONNECTIONS
ULTRASONIC TRANSDUCER
Figure 10: Operational parts of the Ultrasonic Cleaning System
Certain routine maintenance items, such as calibration and desiccant change, are triggered by internal diagnostic alerts. The time to take other routine maintenance tasks, such as removal of flow through to manually clean cuvette, will vary based on system usage and water quality conditions.
5.7 Cleaning the Flow-Through Cuvette
Measurement cuvettes used for both the grab sample and the flow through should be clean and free of marks or scratches. It is recommended to check, and if required clean, the cuvette every three months when performing the primary calibration. The flow though assembly must be removed to perform the calibration procedure.
Cleaning is accomplished by cleaning the interior and exterior with a detergent solution and then rinsing several times with distilled or de­ionized water.
51
Page 58
The cuvette can be replaced by first shutting off the flow using the provided shutoff clamp; unscrewing the old cuvette and replacing with a fresh clean one.
Figure 11: Flow-Through Assembly
5.8 Replacing the Source Lamp
The source lamps in the MTOL+ instruments are designed for long life. The IR lamp and white light version have long lives. These lamps, however, are not covered by the warranty. If the lamp should need replacement, we recommend calling HF Service Department for assistance.
5.9 Optional Factory Installed 24V Option
This factory installed option allows for the operation of the MTOL+ from 24 volts DC. For this instrument the changes are in the junction box only.
The range of correct operation is from 18-36VDC. Caution must be used to not exceed this voltage.
Connections are not polarity sensitive and are made to the same place as mentioned in section 3.4.1.
Order numbers are 28052-24V for the IR version and 28053-24V for the White Light version.
All other specifications remain unchanged.
52
Page 59
6.0 Troubleshooting
6.1 MTOL+ Fault Detection
The MTOL+ performs continuous diagnostic monitoring. In the MTOL+ there are three levels of fault detection:
1. Warnings
2. Errors
3. Failures
Any faults are displayed in a queue form in the bottom row of the LCD. How these faults are indicated depends on the settings made in sections 4.2.1 Configuring the Error Level and 4.2.4 Configuring the Alarms. If ERLV is set to OFF and Alarms are not set to Error, there will be no remote indication of a problem.
6.1.1 Warnings If the desiccant becomes saturated, only a screen warning of DESC will appear and but alarms will be activated. Other warnings are ALM1 or
ALM2 will be displayed if an alarm is set and the threshold is exceeded.
6.1.2 Errors
An error indicates a failure or a problem that usually can be corrected by the operator. These errors consist of:
LAMP - Lamp out
MA - 4-20 mA loop open
• FLOW - If option equipped, enabled and no flow
CLN - MTOL+ is equipped with ultrasonic cleaning and message
will indicate that the ultrasonic transducer is not making contact or the flow through has been removed
If any of these errors occur DO NOT rely on displayed readings. These errors effect accuracy and reliability of instruments readings.
6.1.3 Failures
A failure is a system fault. This is NOT a problem that the operator can correct, and the unit must be returned to the factory for service. These failures consist of failures in the CPU, A/D, EEPROM or other devices internal to the instrument. The word FAIL will display on the lower row.
53
Page 60
DESC
LAMP
FLOW
FAIL
CLN
FILE
6.2 Diagnostic Table
Symptom Cause Cure
Lower display shows MA 4-20 mA loop open Check wiring. See sections
3.4.2 and 4.2.1
Lower display shows
Lower display shows
Lower display shows
Lower display shows Lower display shows
Upper display shows
Readings are higher than expected
Readings are erratic Bubbles in solution
Desiccant pouch bad Change desiccant pouch.
See section 5.1
Lamp failed Replace lamp. Refer to
section 5.8
Flow option error
Sample flow has stopped
Major system fault Refer to section 6.1.3
Ultrasonic cuvette not
found
Update not possible
because new file not found Bubbles in solution
Condensate or leaky cuvette
Flow through cuvette dirty
Instrument out of calibration
Debris in flow through
Restore flow. Contact HF about factory installed option
Either turn off the CLN in CONFIG or check/replace the ultrasonic cuvette. See Sections 4.2.11 & 5.5 Check USB flash drive. See section 4.2.18 Apply backpressure. See
section 6.3 and figure 4
For severe cases of bubbles
a stilling chamber is available. Call HF scientific.
Part# 20106
Check flow through cuvette
for condensate or leaks. Change desiccant if needed.
Clean cuvette. See section
5.7
Recalibrate. Refer to section
4.3 See above Clean debris from cuvette
Readings are lower than expected Upper display flashes Sample Over-Range Check sample. Sample may
54
Instrument out of calibration
Recalibrate. Refer to section
4.3
be too high to read.
Page 61
6.3 Bubbles in the Sample
If bubbles are present in the sample water the readings may appear higher than expected. These reading may also not be stable.
Remove cuvette if air can be seen coming into flow through unit from the inlet tube, this is air introduced, probably from a small water or air leak. This will need to be resolved in the plumbing.
If air cannot be seen entering the intake tube but appears on the inside of the glass cuvette, it is coming out of solution. This can usually be cured by applying pressure to the rotary backpressure clamp on the OUT side of the cuvette. This helps keep the air in solution so it will not be seen by the instrument. Refer to figure 4.
6.4 Technical and Customer Assistance
If for any reason assistance is needed regarding this instrument please do not hesitate to contact either the HF scientific Technical Service Department or the HF scientific Customer Service Department:
HF scientific
3170 Old Metro Parkway
Fort Myers, Florida 33916-7597
Phone: (239) 337-2116 Fax: (239) 332-7643
Toll free: 888-203-7248
Email: HF.Info@Wattswater.com
www.hfscientific.com
55
Page 62
7.0 Accessories and Replacement Parts List
The items shown below are recommended accessories and replacement parts.
Accessory Cat. # Photo
Electronic Service Module For MTOL+ White Light, factory set to 0-100 NTU
Electronic Service Module For MTOL+ IR Light, factory set to 0-100 NTU
Replacement Lamp Assembly White Light
Replacement Lamp Assembly IR Light
Operating Manual, MTOL+ 29290 N/A
Modbus Manual , MTOL+ 29291 N/A
02853
02852
24082S
21369S
ProCal
ProCal Calibration Kit, .02, 10 & 100 NTU
ProCalProCal
For 0-100 NTU range
Procal
Procal Calibration Kit, .02, 1 & 10 NTU
Procal Procal
For 0-10 NTU range
Procal
Procal Calibration Kit, .02, 10 & 1000 NTU
Procal Procal
For 0-1000 NTU range
56
39953
39950
39957
Page 63
ProCal
ProCal Calibration Standard 10 NTU 125ml
ProCalProCal
Formazine Stock Kit 50040 N/A
39825
Formazine Stock Solution, 4000 NTU, 500 ml
Replacement Desiccant Pouch
Pressure Regulator
Replacement Cuvette – for operation without cleaning (3 pack)
Replacement Cuvette with Ultrasonic Transducer
70914
21555R
24306S
50036
24166S
Tubing Kit Containing:
1-shutoff clamp, 1-backpressure valve,
2-connecting tubing with fittings for flow through assembly.
To order any accessory or replacement part, please contact the HF scientific Customer Service Department. If for any reason technical assistance is needed regarding this instrument please do not hesitate to contact the HF Technical Services Department.
3170 Old Metro Parkway
Fort Myers, Florida 33916-7597
Phone: (239) 337-2116
Fax: (239) 332-7643
Toll free: 888-203-7248
Email: HF.Info@Wattswater.com
www.hfscientific.com
57
HF scientific
21062
Page 64
8.0 Limited Warranty
HF scientific inc., as vendor, warrants to the original purchaser of this instrument that it will be free of defects in material and workmanship, in normal use and service, for a period of two years from date of invoice. HF scientific inc.’s obligation under this warranty is limited to replacing, at its factory, the instrument or any part thereof. Parts, which by their nature are normally required to be replaced periodically, consistent with normal maintenance, specifically reagent, desiccant, sensors, lamps and fuses are excluded. Also excluded are accessories and supply type items.
Original purchaser is responsible for return of the instruments, or parts thereof, to HF scientific’ inc.’s factory. This includes all freight charges incurred in shipping to and from HF scientific inc.’s factory.
HF scientific inc .is not responsible for damage to the instrument, or parts thereof, resulting from misuse, environmental corrosion, negligence or accident, or defects resulting from repairs, alterations or installation made by any person or company not authorized by HF scientific inc.
The remedy described in the first paragraph of this warranty shall constitute the sole and exclusive remedy for breach of warranty. HF scientific inc. assumes no liability for incidental, consequential or special damage of any kind, and the original purchaser, by placement of any order for the instrument, or parts thereof, shall be deemed liable for any and all damages incurred by the use or misuse of the instruments, or parts thereof, by the purchaser, its employees, or others, following receipt thereof.
Carefully inspect this product for shipping damage, if damaged, immediately notify the shipping company and arrange an on-site inspection. HF scientific inc cannot be responsible for damage in shipment and cannot assist with claims without an on-site inspection of the damage.
This warranty is given expressly and in lieu of all other warranties, expressed or implied. Purchaser agrees that there is no warranty on merchantability or fitness for a particular purpose and that there are no other warranties, expressed or implied. HF scientific inc hereby specifically disclaims all other warranties, express or implied. No agent is authorized to assume for HF scientific inc., any liability except as set forth above. Some States do not allow limitations on how long an implied warranty lasts, and some States do not allow the exclusion or
incidental or consequential damages.
Therefore the above limitations may not apply to you. This
limitation of
Limited Warranty gives you specific legal rights, and you may have other rights that vary from State to State. You should consult applicable state laws to determine your rights.
CONSISTENT WITH APPLICABLE STATE LAW, ANY IMPLIED WARRANTIES THAT MAY NOT BE DISCLAIMED, INCLUDING THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, ARE LIMITED IN DURATION TO ONE YEAR FROM THE DATE OF ORIGINAL SHIPMENT.
SO FAR AS IS
HF scientific, inc.
3170 Old Metro Parkway
Fort Myers, Florida 33916-7597
Phone: (239) 337-2116
Fax: (239) 332-7643
Toll free: 888-203-7248
Email: HF.Info@Wattswater.com
Website: www.hfscientific.com
58
Page 65
59
Page 66
Loading...