The MTOL+™ Online Turbidimeter has been specifically designed
for measuring the NTU of filtered water, re-use water, raw water
and many industrial applications. The MTOL+ features built-in
Ultrasonic auto-cleaning and data logging as well as easy set-up,
maintenance, and calibration. Calibration is completed using
ProCal™ EPA approved* NTU calibration standards.
The MTOL+ is compatible with the M100+ laboratory turbidimeter which offers a color touchscreen display and sample
data logging for multiple sample ID's. The M100+ also stores
calibration NTU values and the date/time of calibration.
The two units share calibration standards eliminating the need
for different calibration standards between laboratory and
process NTU.
• Same Standard set for both versions for
most applications
MTOL+
Online Process Turbidimeter
Factory default 0 - 100 NTU
M100+
Laboratory Turbidimeter
* ProCal NTU Standards are EPA approved for
use as primary NTU standards
HFscientific.com
Page 2
Model MTOL+ Turbidimeter System
The continuous monitoring NTU system shall include the online
model with power supply, display, user interface and sensor as
one single SKU. It will not be necessary to purchase a separate
module for display or power supply. The turbidimeter shall meet
all requirements specified by the USEPA Method 180.1 (White
Light Model), ISO 7027 (Infrared Model) and Standard Methods
2130B. The turbidimeter shall be Modbus compatible and have
a 4-20 mA output. The turbidimeter shall have the same optical
design as the M100+ laboratory turbidimeter.
The online turbidimeter shall have an ultrasonic auto-clean
system. Accuracy shall be 2% of reading or plus or minus 0.02,
whichever is greater, from 0-40 NTU, and 5% of reading or
plus or minus 0.02, whichever is greater, from 40-1000 NTU.
Resolution will be 0.001 NTU (user selectable).
The sensor shall utilize a rotational flow through assembly with
a 30ml measuring cuvette. The specially designed flow head
bubble rejection system will reduce the need for a bubble trap
and provide a faster response time. The sample stream will
remain under pressure, ensuring it will not be exposed to open
atmosphere, which can cause precipitants in some circumstances.
Precipitation of oxidized sample constituents can result in false
NTU readings. The sensor shall allow grab samples to be
measured without the need to stop sample flow to the flow head.
Calibration and standardization will be accomplished using small
volumes (30ml) of reusable primary standards. Primary Standards
shall be reusable for multiple online turbidimeters and also with
the M100+ laboratory turbidimeter. Calibration procedures can
be completed without disrupting the sample flow. Details of
calibration (date and time of last successful calibration) will
be stored on the analyzer. An integral part of the turbidimeter
system is the inclusion of a laboratory instrument specifically
designed to use the same calibration standards. The required
supporting laboratory instrument must be an HF scientific M100+
touchscreen model with USB data logging and ability to store the
date/time of the last calibration.
The process turbidimeter and the laboratory turbidimeter lamp
source and detector shall not come in contact with the sample,
eliminating false low readings. The turbidimeters shall use menu
driven software. The online turbidimeter enclosure shall be ABS.
The online turbidimeter will have data logging features, will store
the latest date/time of calibration and be user selectable for
ranges of 0 – 10 NTU, 0 – 100 NTU (factory default) or 0 – 1000
NTU. The online turbidimeter shall be HF scientific MTOL+
Online Turbidimeter.
MTOL+ Features & Benefits:
Optical design
The unique, shared optical design allows consistent readings
between the M100+ laboratory turbidimeter and the MTOL+. The
unique optical design ensures that the same calibration
standards can be used with the M100+ laboratory unit and
the MTOL+.
Bubble rejection system
The optical chamber of the MTOL+ has been designed to eliminate air in the sample while simultaneously creating a vortex
cleaning action throughout the optical chamber.
Calibration
(Online & Lab NTU—Use the same Standards)
Calibration with primary standards is completed using sealed
cuvettes, similar to laboratory procedures. This method of
calibration is fast, clean and reusable. On-screen menu prompts
guide you through the calibration procedure quickly and easily.
The layouts for the MTOL+ and M100+ have a similar "feel" making the UI transition from one to the other easy.
Modular design
One-piece mounted design allows for simple mounting and
minimal use of space. Wide range of measurement (0-1000 NTU)
allows for measurement of both low NTU filtered water or raw
water with one analyzer, eliminating the need to learn multiple
models. Optical design increases accuracy and provides more
consistent readings with online, laboratory and portable
turbidimeters. Calibrate the new M100+ laboratory turbidimeter
with the same standards as the MTOL+ for greatly improved comparison functionality between lab and online.
Light source
White light is recommended for use in turbidimeters reporting
results under US EPA 180.1 (US standard) jurisdiction. The
white light is Tungsten filament, krypton gas filled, and has a life
expectancy of 5 to 7 years. Infrared light is recommended for
use in turbidimeters reporting results under ISO 7027 (European
standard) jurisdiction. Infrared light is also recommended for
some waste water final effluent and industrial applications where
color is present in the sample stream.
Sensor Base Installation allows
up to 1/4" hardware
Power Supply/Junction Box installation
allows up to 3/16" hardware
2% of reading ±.02 NTU (below 40 NTU) 5% of reading
(above 40 NTU)
Resolution 0.001 (selectable) 0.0001 (selectable)
Response Time Adjustable (5-500 seconds) Selectable-Manual or Auto Mode
Flow Rate 0.026-.26 gpm (100 ml/min-1000 ml/min) NA
Standard Outputs 4-20 mA Galvanic Isolated or RS-485 Modbus NA
User Alarms 2 User Settable Alarms NA
Security Code User Selectable-Prevents Unauthorized Access NA
Storage Temperature -4°F to 140°F (-20°C to 60°C) -4°F to 140°F (-20°C to 60°C)
Data Logging Yes Yes
Store Calibration Data Yes Yes
Operating Temperature 34°F to 122°F (1°C to 50°C) 34°F to 104°F (1°C to 40°C)
Input Pressure 1-101 psi (built in regulator set at 15 psi) NA
Enclosure ABS ABS
White Light Version compliant to U.S. EPA 180.1
Regulatory Compliance
and Certifications
Infrared Version compliant to ISO 7027 and DIN 27027
CE Approved, LC listed to UL 61010-1 3rd Edition: 2012
LC: Certified to CSA 22.2 No.61010-1-12: 2012
Warranty 2 Years from ship date 1 year from ship date
2% of reading ±.02 NTU (below 40 NTU) 5% of
reading (above 40 NTU)
White Light Version compliant to U.S. EPA 180.1
Infrared Version compliant to ISO 7027
LC mark tested to UL and CSA.
Conforms to CE
Online Process Turbidimeter
Backpressure Valve to prevent dissolved
gases from releasing in the optics.
Shut off Clamp
Inlet Pressure Regulator
Max input 101psi
Factory set to 5psi
Laboratory Turbidimeter
Dimensions:
10.75" W x 10" L x 4.75" H
50.80
[2,00]
88.77
[3,50]
133.71
[5,26]
350.19
[13,79]
Sample Inlet and Outlet
1/4" OD push to
connect fittings
87.12
[3,43]
147.65
61.91
[2,44]
198.20
[7,80]
Provide at least 8 inches of free space above the sensor to allow removal
for the flow head for maintenance and calibration.
Page 4
Ordering Information Turbidimeter System:
When ordering a complete turbidimeter system please order as many online units (MTOL+) as are required and at least
one M100+ laboratory turbidimeter (M100+ includes a set of ProCal Calibration Standards, 0-4000 NTU).
MTOL+M100+
Ordering Information
Cat. No.Model
28053
28052
White Light0-100
Infrared0-100
NTU
Range
Calibration Kits/Standards
Part #Description
39953
39950
39957
39825
39824
50040
ProCal Calibration Kit 0.02, 10, 100 NTU
ProCal Calibration Kit 0.02, 1, & 10 NTU
ProCal Calibration Kit 0.02, 10, & 1000 NTU
ProCal Calibration Standard 10 NTU, 125 ml
ProCal Calibration Standard 100 NTU, 125 ml
Formazin Stock Solution Kit
Spare Parts
Part # Description
21555R
20779S
24166S
24082S
21369S
50036
24306S
21062
02852
02853
Dessicant Refill
Power Cord 120 VAC / 240 VAC
Replacement Ultrasonic Cuvette Assembly
Replacement Lamp Assembly White Light
Replacement Lamp Assembly IR Light
Replacement Cuvettes, non-ultrasonic (3 Pack)
Pressure Regulator
Replacement Tubing Kit
Electronic Service Module for MTOL+ IR Light
Electronic Service Module for MTOL+ White Light
USEPA
Method
180.1
x x
ISO
7027
Ordering Information
Cat. No.Model
28060
28061
White Light0-4000
Infrared0-4000
NTU
Range
Calibration Kits/Standards
Part #Description
39940
39941
39957
39825
50040
ProCal Calibration Kit
(0-4000 NTU for M100+ White Light)
ProCal Calibration Kit
(0-4000 NTU for M100+ Infrared)