OWNER’S MANUAL
MicroTPI & MicroTPW
Field Portable Turbidimeters
HF scientific, inc.
3170 Metro Parkway
Ft. Myers, FL 33916
Phone: 239-337-2116
Fax: 239-332-7643
Email: Info@hfscientific.com
Website: www.hfscientific.com
Manual Part No. 24378 (2/07)
Rev. 1.6
DECLARATION OF CONFORMITY
Application of Council Directive
Standard to which Conformity is Declared:
Product Safety - Tested and passed CE EN61010-1:1990 + A1:1992 (73/32 EEC)
Immunity and Emissions – Tested and passed EN61326:1997 + A1:1998
Manufacturer’s Name: HF scientific, inc.
Manufacturer’s Address: 3170 Metro Parkway, Fort Myers, Florida 33916-7597
Type of Equipment: Field Portable Turbidimeter
Model No: MicroTPI & MicroTPW
I, the undersigned, hereby declare that the equipment specified above conforms to the above
Directive and Standard
Place: Fort Myers, Florida USA _________________________
(Signature)
Rowan T. Connelly, General Manager
Micro TPI & MicroTPW (2/07) i
Rev. 1.6
Table of Contents
DECLARATION OF CONFORMITY............................................................................................i
Specifications................................................................................................................................. 1
1.0 Overview............................................................................................................................ 2
1.1 Unpacking and Inspection of the Instrument and Accessories ....................................... 2
1.2 The Display..................................................................................................................... 3
1.3 The Instrument and Touch Pad....................................................................................... 4
2.0 Safety.................................................................................................................................. 4
3.0 Normal Operation.............................................................................................................5
3.1 Routine Measurement ..................................................................................................... 6
4.0 Instrument Calibration .................................................................................................... 7
4.1 Calibration Standards...................................................................................................... 7
4.2 Indexing the Calibration Standards................................................................................. 7
4.3 Calibration Procedures.................................................................................................... 8
4.4 Calibration Error ............................................................................................................. 9
4.5 Restore Factory Calibration ............................................................................................ 9
5.0 Routine Maintenance...................................................................................................... 10
5.1 Cuvette Cleaning and Care ........................................................................................... 10
5.2 Battery Replacement..................................................................................................... 11
6.0 Troubleshooting.............................................................................................................. 12
6.1 System Warning Messages ........................................................................................... 12
6.2 Battery Low Indicator................................................................................................... 12
6.3 System Error Messages................................................................................................. 13
6.4 Technical and Customer Assistance ............................................................................. 13
7.0 Accessories and Replacement Parts List....................................................................... 14
8.0 Warranty ......................................................................................................................... 15
8.1 Waterproof Seal
9.0 Glossary ........................................................................................................................... 16
……………………………………………………………………………………………... 15
Micro TPI & MicroTPW (2/07) ii
Rev. 1.6
Specifications
Measurement Range
Accuracy
Resolution
0.01 –1100 NTU/FNU
±2% of reading or 0.01 NTU (0-500 NTU)
±3% of reading (500-1100 NTU)
0.01 NTU below 100.0 NTU,
0.1 NTU for 100.0 NTU – 999.9
1.0 NTU for 1000 to 1100 NTU
Regulatory Compliance MicroTPI IR (860 nm) LED Compliant to ISO 7027
MicroTPW Tungsten Lamp Compliant to U.S. EPA 180.1
Principle of Operation Nephelometry
Response Time Less than 14 Seconds
Display 4 digit 7 Segment Liquid Crystal Display with Special Icons
Operating Temperature Range
0°C – 50°C (32°F – 122°F)
Sample Temperature Range 0°C – 50°C (32°F – 122°F)
Sample Size 10-15 mL
Relative Humidity 0-90% Non Condensing
Power Supply 4 – AAA Batteries
Battery life is approx.5000 readings with standard alkaline batteries
Certification CE
Insulation Rating Pollution Degree 2
Approximate Dimensions Instrument: 7 cm x 16.5 cm x 4.8 cm (2 ¾” x 6 ½” x 1 7/8”)
Instrument in Minilab Case: 25 cm x 21.6 cm x 6.3 cm
(10” x 8 1/2” x 2 1/2”)
Enclosure Rating:
NEMA 4X
Designed to meet the specifications of IP67
Shipping Weight: 1.22 kg (2.70 lbs)
Shipping Dimensions: 28 cm x 30.5 cm x 7.6 cm (11” x 12” x 3”)
Warranty 1 Year from date of shipment
MicroTPI & MicroTPW (2/07)
Rev. 1.6
1