HF Scientific Micro TOL2 WL, Micro TOL3 WL, Micro TOL3 IR, Micro TOL4 WL, Micro TOL2 IR User Manual

...
OWNER’S MANUAL
MicroTOL Series Turbidimeter
3170 Metro Parkway Ft. Myers, FL 33916 Phone: 239-337-2116 Fax: 239-332-7643
Catalog No. 24034 (8/07) E-Mail: hfinfo@watts.com
Rev. 4.1 Website: www.hfscientific.com
HF scientific
DECLARATION OF CONFORMITY
Application of Council Directive
Standards to Which Conformity is Declared:
Product Safety - Tested and passed ETL (tested to UL 61010B-1), 1
24, 2003
- Tested and passed ETLc (tested to CSA C22.2#1010.1-92)
Emissions & Immunity – Tested and passed EN61326: 1997 + A1:1998 + A2:2001 + A3:2003
Manufacturer’s Name: HF scientific
Manufacturer’s Address: 3170 Metro Parkway, Fort Myers, Florida 33916-7597
Importer’s Name:
Importer’s Address:
Type of Equipment: Process Turbidimeter
Model No: Micro TOL
I, the undersigned, hereby declare that the equipment specified above conforms to the above Directive and Standard
Place: Fort Myers, Florida USA (Signature)
Rowan T. Connelly, General Manager
st
Edition, Dated January
Table of Contents
Section Page
Specifications ....................................................................................................1
1.0 Overview ........................................................................................................... 2
1.1 The Micro TOL Series .........................................................................2
1.2 Unpacking and Inspection of the Instrument and Accessories ............. 3
1.3 The Display ..........................................................................................3
1.4 The Touch Pad ...................................................................................... 4
1.5 Vapor Purge........................................................................................... 4
2.0 Safety.................................................................................................................4
3.0 Installation and Commissioning ....................................................................... 5
3.1 Mounting and Site Selection .................................................................5
3.2 Plumbing ............................................................................................. 6
3.2.1 Drain Vent .................................................................................7
3.2.2 Wetted Materials ....................................................................... 7
3.3 Electrical Connections .......................................................................... 7
3.3.1 Power ........................................................................................ 8
3.3.2 RS-485 ..................................................................................... 8
3.3.3 Relays ........................................................................................8
3.3.4 4-20 mA .................................................................................... 8
4.0 Operation ..........................................................................................................9
4.1 Routine Measurement ........................................................................... 9
4.2 Security Access Feature ........................................................................9
5.0 Instrument Calibration .................................................................................... 11
5.1 Calibration Standards ......................................................................... 11
5.2 Calibration Procedures ....................................................................... 12
5.3 Calibration Error ................................................................................. 13
6.0 Instrument Offset ........................................................................................... 14
6.1 Indexing Calibration Cuvettes ............................................................ 15
6.2 Restoring Factory Settings ..................................................................15
7.0 Instrument Configuration (CONFIG mode)....................................................16
7.1 Selecting the Output (O/P) .................................................................16
7.2 Setting the 4-20 mA ............................................................................ 16
7.3 Configuring the RS-485 Port ..............................................................17
7.4 Configuring the Alarms....................................................................... 18
7.4.1 Alarm 1....................................................................................18
7.4.2 Alarm 2 ................................................................................... 19
7.5 Offset Calibration ................................................................................ 19
MICRO TOL (6/08) Rev. 4.1
Table of Contents (continued)
Section Page
7.6 Enabling the Security Access ..............................................................20
7.7 Extended Settings ................................................................................ 20
7.8 Speed of Response............................................................................... 20
7.9 Displayed Resolution ......................................................................... 21
7.10 LCD Backlight Brightness ..................................................................21
7.11 Setting the Units ..................................................................................22
7.12 Ultrasonic Cleaning............................................................................. 22
7.13 RS- 485 Parameters ............................................................................. 22
7.14 Desiccant Alarm .................................................................................. 23
7.15 Saving Configuration Settings............................................................. 23
8.0 Additional Features and Options .................................................................... 24
8.1 Backlit LCD ........................................................................................ 24
8.2 Ultrasonic Cleaning............................................................................. 24
8.3 RS-485 Output.....................................................................................25
8.3.1 HF Online ................................................................................ 25
8.3.2 Simple Communication........................................................... 25
8.3.3 Modbus Communication .........................................................26
8.4 Flow Alarm.......................................................................................... 26
8.5 Flow Controller ................................................................................... 26
8.6 Remote Panel Meter ............................................................................ 26
9.0 Troubleshooting & Maintenance .................................................................... 27
9.1 Micro TOL Fault Detection.................................................................27
9.2 System Fail Message .......................................................................... 27
9.3 Diagnostic Chart ................................................................................. 28
9.4 Technical and Customer Assistance ................................................... 28
10.0 Routine Maintenance....................................................................................... 29
10.1 Cleaning the Flow Through Cuvette .................................................. 29
10.2 Replacing or Installing the Desiccant Pouch ...................................... 29
10.3 Replacing the Source Lamp ...............................................................29
11.0 Accessories and Replacement Parts List ........................................................ 30
12.0 Warranty ......................................................................................................... 31
MICRO TOL (6/08) Rev. 4.1
Specifications
Measurement Range
Accuracy
Resolution Response Time
Display Alarms
Analog Output Communications Port
Maximum Water Pressure
Flow Rate Operating Temperature
Wetted Materials
0 – 1000.0 NTU 0 – 100 NTU (Model 20055 & 20056) ±2% of reading or ±0.02 NTU below 40 NTU whichever is greater ±5% of reading above 40 NTU
0.0001 NTU (below 10 NTU) Adjustable
Multi-Line Liquid Crystal Backlit Display
Two Programmable, 120-240VAC 2A Form C Relay
Powered 4-20 mA, 600 Ω drive
Bi-directional RS-485, Modbus
Integral pressure regulator rated 1380kPa (200 PSI.) Also refer to
Flow Rate
100 ml/min. – 1 liter/min. (.026-.26 Gal/min)
1°C – 50°C (34°F – 122°F) Nylon, Borosilicate Glass, Silicon, Polypropylene, Stainless Steel
Sample Temperature Range Power Supply
Insulation Rating Environmental Conditions
Enclosure Rating Regulatory Compliance And Certifications
Shipping Weight Warranty
1°C – 50°C (34°F – 122°F)
100 – 240 VAC, 47 – 63 Hz, 80VA
Double Insulated, Pollution Degree 2, Overvoltage Category II
Not recommended for outdoor use.
Altitude up to 2000 meters
Up to 95 % RH (non-condensing)
Designed to meet IP 66 /NEMA 4X
White Light Version compliant to U.S. EPA 180.1
Infrared Version compliant to ISO 7027
CE Approved, ETL listed to UL 61010B-1 &
ETL Certified to CSA 22.2 No. 1010-1-92
2.5 kg (5.5 lbs.) 1 Year from date of shipment
MICRO TOL (6/08) 1 Rev. 4.1
1.0 Overview
The MICRO TOL process turbidimeter allows for the measurement of the turbidity of process water on-line. The White Light MICRO TOL has been designed to meet the design criteria specified by the US EPA 180.1 on turbidity measurement. The infrared MICRO TOL was designed to meet the design criteria specified in ISO 7027 and DIN 27027 for the measurement of the turbidity of a sample. Both models have long life lamps.
Some models have ultrasonic cleaning. Refer to section 8.2 for more information.
A pressure regulator on the incoming line is a standard on all Micro TOL instruments and
will reduce pressures up to 1380kPa (200 PSI) down to (104kPa) 15 PSI.
1.1 The Micro TOL Series
The Micro TOL series instruments have a wide variety of options available. Refer to the table below to determine which factory installed options are available.
Catalog
No.
20053 Micro TOL2 WL Standard Standard Standard N/A 0-1000 Option 20054 Micro TOL2 IR Standard Standard Standard N/A 0-1000 Option
20055 Micro TOL3 WL Standard Standard Standard Standard 0-100 Option 20056 Micro TOL3 IR Standard Standard Standard Standard 0-100 Option
20063 Micro TOL4 WL Standard Standard Standard Standard 0-1000 Option 20064 Micro TOL4 IR Standard Standard Standard Standard 0-1000 Option
Description RS-485 Modbus Backlight Ultrasonic
Cleaning
Range
NTU
Flow
Alarm
MICRO TOL (6/08) 2 Rev. 4.1
1.2 Unpacking and Inspection of the Instrument and Accessories
The table below indicates the items in the turbidimeter shipment.
Item Quantity
MICRO TOL Turbidimeter c/w Field Terminal Box & Flow Through
1
Assembly
Instruction Manual
Desiccant Pack
Cuvette (Single Pack)
Tubing Kit: 1-shutoff clamp
1 1 1 1
1-backpressure valve
2-connecting tubing with fittings for flow through assembly 1-drain vent screw (used in pressurized systems)
Remove the instrument from the packing carton. Carefully inspect all items to ensure that no visible damage has occurred during shipment. If the items received do not match the order, please immediately contact the local distributor or the HF scientific, inc. Customer Service department.
Note: The spare cuvette part# 50033 is not included for models 20055, 20056 20063 & 20064.
1.3 The Display
Figure 1 illustrates all the items that can appear on the display. The upper row of the display (1) is used for reporting the turbidity levels and to provide user guidance in the customer setting routine. The lower row of the display (2) is used to communicate error messages and provide user guidance. The display has two icons (3) that are used to indicate the use of access code and offset mode. In addition, mode arrows (4) are used to indicate the current instrument operating mode; AUTO (normal operation), CAL (calibration) and CONFIG (configuration).
Figure 1 – Display used in the instrument.
All items used on the display are shown in this figure
MICRO TOL (6/08) 3 Rev. 4.1
1.4 The Touch Pad
Figure 2 illustrates the touch pad. The touch pad has four buttons: MODE/EXIT, ↵, t, and u. The MODE/EXIT button is used to cycle between the three operational modes of the instrument: CAL, CONFIG, and AUTO (Measurement) mode. The button enters
the option (or mode that is highlighted or chosen. The tand u buttons are used to change settings.
Figure 2: Touch Pad
1.5 Vapor Purge
The Micro TOL is equipped with a continuous vapor purge system. A replaceable desiccant pouch in the lower portion of the instrument dries the air. System heat is used to warm the air. A fan inside the instrument continuously circulates heated dry air around the optical well and the flow through cuvette. This feature eliminates the need for a dry purge line.
The Micro TOL monitors the replaceable desiccant pouch condition continuously. The
LCD display will show DESC on the lower line in the event that the desiccant pouch
needs replacement. Replacement desiccant pouches are available from HF scientific, inc. or the local representative (Part # 21555R). Refer to section 10.2 Replacing or installing
the Desiccant Pouch.
The desiccant can activate an alarm to notify the operator of a saturated desiccant. See section 7.14 Desiccant Alarm.
Note: Prior to installing the desiccant pouch for the first time the shipping support must be
removed. This tube can be discarded.
2.0 Safety
This manual contains basic instructions that must be followed during the commissioning,
operation, care and maintenance of the instrument. The safety protection provided by this
equipment may be impaired if it is commissioned and/or used in a manner not described in this manual. Consequently, all responsible personnel must read this manual prior to
working with this instrument.
In certain instances Notes, or helpful hints, have been highlighted to give further
clarification to the instructions. Refer to the Table of Contents to easily find specific topics and to learn about unfamiliar terms.
MICRO TOL (6/08) 4 Rev. 4.1
3.0 Installation and Commissioning
Prior to use for the first time, the supplied desiccant pouch will need to be installed. Refer to section 10.2 Replacing or Installing the Desiccant Pouch.
3.1 Mounting & Site Selection
The instrument is designed for wall mounting. If wall mounting is not practical, the instrument can be mounted on any suitable level surface. For ease of service there should be about 20 cm (8”) free area above the instrument; this will ensure enough room for calibration and cuvette maintenance. Choose a location that is easily accessible for operation and service and ensure that the front display rests at eye level. The overall mounting dimensions of the instrument are shown in Figure 3. The recommended mounting screws are M6 (¼”) for the instrument enclosure and M4 (#8) for the field terminal box. The Micro TOL is designed to have the field terminal box cradled under the sensor portion of the instrument. It is recommended that the field terminal box be mounted first, and then the rest of the instrument be mounted on top. The template on the last page of this manual may be used to establish mounting hole locations.
Figure 3: Overall Mounting Dimensions of the Instrument
It is critical that the instrument be mounted as close as possible to the sampling point to ensure a quick response time (within 2-3 meters (6-10 ft) of the sampling point).
MICRO TOL (6/08) 5 Rev. 4.1
3.2 Plumbing
The recommended plumbing for the instrument is shown in Figure 4. The instrument is designed to require very little head pressure to operate; around 6.9kPa (1 PSI). The flow through cuvette is rated for a flow of 100ml/min. – 1 liter/min. (0.026-0.26Gal/min). The integral pressure regulator is rated for a maximum pressure of 1380 kPa (200 PSI.). The maximum allowable fluid temperature is 50°C (122°F).
Figure 4: Recommended Plumbing for the Instrument
The instrument is equipped to be plumbed using 4.75 mm (3/16”) ID, 8 mm (5/16”) OD flexible tubing. Opaque tubing should be used if the tubing will be exposed to sunlight, to prevent algae growth.
In figure 4, there are two flow devices shown. The one on the input side is a shutoff clamp used during cuvette maintenance. The other device is a backpressure valve. Backpressure may be required to prevent air from coming out of solution, which may be observed as tiny air bubbles.
MICRO TOL (6/08) 6 Rev. 4.1
3.2.1 Drain Vent: The Micro TOL has been fitted with a drain vent in the “OUT”
bulkhead fitting. This fitting allows for atmospheric equalization, thus helping to alleviate bubble formation in the cuvette. Refer to Figure 4.
Upon initial flow minor leakage may occur through the drain vent. This will subside once
normal flow is established.
For some high pressure systems, where the vent hole continuously leaks, a 6:32 seal screw
is provided which should be inserted into the vent hole and tightened.
The sensor drain tubing should be routed to a suitable drain. It is not recommended to reintroduce the drain sample to the process stream.
3.2.2 Wetted Materials: HF scientific inc. accepts no responsibility for damage caused
by the introduction of vapors, fluids or other materials into the instrument process stream which is not compatible with the instrument’s wetted materials. A list of the wetted materials can be found in the specifications on page 1 of this manual.
3.3 Electrical Connections
All of the electrical connections to the instrument are made through the field terminal box, which should be located directly under the sensor portion of the instrument. The connections are labeled within the terminal box and are self-descriptive (see Figure 5). Please follow all local and government recommendations and methods for installation of electrical connections to and between the instrument and other peripheral devices.
Plugs are inserted into the alarm and 4-20mA/RS-485 cable bulkheads when shipped, to ensure a watertight seal. These plugs should be removed and discarded when cabling to either of these connections.
The power cable bulkhead will accept cable diameters from 5.8mm (.230 in.) up to 10 mm (.395 in.). All terminals are designed to accept wires in the range of 14-28 AWG. All wires should be stripped to a length of 6 mm (¼”). A strain relief strap is provided to reduce tension on the power terminals.
It is the user’s responsibility to assure that the watertight seal is maintained after the terminal box has been wired for operation. If any of the bulkheads are not tightened properly around a cable or plug, the ratings of the instrument will be jeopardized and there is a possibility of creating a shock hazard.
Note: Only qualified electricians should be allowed to perform the installation of the
instrument as it involves a line voltage that could endanger life.
MICRO TOL (6/08) 7 Rev. 4.1
Loading...
+ 29 hidden pages