HF Scientific M100+ 28060, M100+ 28061 Owner's And User's Manual

M100+ (April 2019) Rev. 1.76
OWNER’S AND USER’S MANUAL
M100+TM Laboratory Turbidimeter
Models 28060 & 28061
HF scientific 3170 Old Metro Parkway Ft. Myers, FL 33916 Phone: 239-337-2116 Fax: 239-332-7643 Toll-free: 888-203-7248 E-Mail: HF.Info@WattsWater.com Website: HFscientific.com
Read Manual and all product labels BEFORE using the equipment. Do not use unless you know the safe and proper operation of this equipment. Keep this Manual available for easy access by all users. Replacement Manuals are available at HFscientific.com
M100+ (April 2019) Rev. 1.76
M100+ (April 2019) Rev. 1.76
M100+ (April 2019) Rev. 1.76
Attention Owners and Users
Thank you for purchasing the M100+ turbidimeter. This equipment will provide safe and productive operation as long as it is installed, used, maintained, and serviced in accordance with the instructions in this manual and is properly maintained. Importantly, unless the user is adequately trained and supervised, there is a possibility of personal injury, property damage or damage to the equipment.
Owners and users of this equipment bear the responsibility to make certain that this equipment is used properly and safely. READ THIS MANUAL carefully, learn how to use and service this equipment correctly, and strictly follow all of the instructions contained in this manual and the requirements of local, state and federal law. Failure to do so could result in personal injury, property damage or damage to the equipment. This manual should be considered a permanent part of your machine and should be kept available for easy reference by any user.
Owners should not permit anyone to touch this equipment unless they are over 18 years of age, are adequately trained and supervised, and have read and understood this manual. Owners should also ensure that no unauthorized personnel comes in contact with this equipment.
If this equipment, or any of its parts, becomes damaged or needs repair, stop using the equipment and contact an experienced service individual immediately. If the warning labels or this manual are misplaced, damaged or illegible, or if you require additional copies, please contact customer service at +1 (239) 337-2116 or 888-203-7248 for these items at no charge.
Please remember that this manual and the warning labels do not replace the need to be alert, to properly train and supervise users, and to use common sense when using this equipment.
If you are ever uncertain about a particular task or the proper method of operating this equipment, ask your supervisor, consult this manual, access www.hfscientific.com, or contact us at 888-203-7248.
PRODUCT IDENTIFICATION Please record your product’s identification and purchase information which will help in the
event you have questions or need any service.
Model: Serial #:
Date of purchase: Seller name/address:
M100+ (April 2019) Rev. 1.76
Table of Contents
Section Page
Understanding Safety Information .............................................................................................................. 1
Specifications ............................................................................................................................................. 2
1.0 Overview ............................................................................................................................................ 3
1.1 The M100+ Product Description .................................................................................................... 3
2.0 Unpacking and Inspection ................................................................................................................. 4
3.0 Site Selection ..................................................................................................................................... 5
4.0 Setup, Configuration and Calibration ................................................................................................. 5
4.1 Operating Screens ........................................................................................................................ 6
4.2 Configuring the M100+ .................................................................................................................. 7
4.2.1 Screen Brightness ................................................................................................................ 8
4.2.2 Auto/Manual Readings ......................................................................................................... 8
4.2.3 Auto Power Down ................................................................................................................. 9
4.2.4 Touch Screen Calibration ..................................................................................................... 9
4.2.5 Set Date/Time ..................................................................................................................... 10
4.2.6 Delete Samples .................................................................................................................. 10
4.2.7 Data Resolution .................................................................................................................. 11
4.2.8 Offset .................................................................................................................................. 11
4.2.9 Device ID ............................................................................................................................ 11
4.2.10 Calibration Reminder ...................................................................................................... 12
4.2.11 Calibration Download ..................................................................................................... 12
4.2.12 Firmware Update ............................................................................................................ 13
4.2.13 Reset Defaults ................................................................................................................ 13
4.2.14 About Device .................................................................................................................. 14
4.3 Calibrating the M100+ ................................................................................................................. 14
4.3.1 Calibration Standards ......................................................................................................... 15
4.3.2 Care of ProCal Standards .................................................................................................. 16
4.3.3 Indexing Calibration Cuvettes ............................................................................................. 16
4.3.4 Calibration Procedures ....................................................................................................... 17
4.3.5 Alternate Calibration Ranges .............................................................................................. 19
4.3.6 Failed Calibration ................................................................................................................ 19
5.0 Operation ......................................................................................................................................... 20
5.1 Home Screen .............................................................................................................................. 20
5.2 Trend Screen ............................................................................................................................... 21
5.3 Data Retrieval .............................................................................................................................. 21
5.3.1 USB Connection ................................................................................................................. 22
6.0 Troubleshooting ............................................................................................................................... 23
6.1 M100+ Fault Detection ................................................................................................................ 23
6.2 Diagnostic Table .......................................................................................................................... 23
6.3 Technical and Customer Assistance ........................................................................................... 23
7.0 Routine Maintenance ....................................................................................................................... 24
7.1 Cleaning the Cuvette ................................................................................................................... 24
7.2 Replacing the Source Lamp ........................................................................................................ 24
7.3 Battery ......................................................................................................................................... 24
8.0 Accessories and Replacement Parts List ........................................................................................ 25
9.0 Limited Warranty .............................................................................................................................. 26
1

Understanding Safety Information

This manual contains safety and use instructions that must be followed during the installation, commissioning, operation, care and maintenance, and service of the M100+. All responsible personnel must read this manual prior to working with this instrument and should familiarize themselves with the following safety symbols, signals, and pictorials.
This is a safety-alert symbol.
The safety alert symbol is shown alone or used with a signal word (DANGER,
WARNING or CAUTION), a pictorial and/or a safety message to alert you to hazards.
When you see this symbol alone or with a signal word on this instrument or in this
Manual, be alert to the potential for death or serious personal injury.
Safety signal words have the following meaning:
Identifies hazards which, if not avoided, will result in death or serious injury.
Identifies hazards which, if not avoided, could result in death or serious injury.
Identifies hazards which, if not avoided, will result in minor or moderate injury.
Identifies practices, actions or failure to act which will result in property damage or damage to the equipment.
Pictorials used on the equipment and in this Manual have the following meanings:
This pictorial alerts you to the need to read the Manual.
This pictorial alerts you to electricity, electrocution and shock hazards.
2

Specifications

M100+ White Light Model 28060
Conforms to specifications set forth in EPA method 180.1 (Nephelometric Method)
§ This Model uses a tungsten filament
lamp.
M100+ Infrared Model 28061
Conforms to specifications set forth in ISO 7027: Water Quality - Determination of Turbidity. This model uses an IR LED
lamp.
Measurement Range
0-4000 NTU
Accuracy
†, ††
±2% of reading or ±0.02 NTU below 40 NTU whichever is greater
±5% of reading above 40 NTU
Repeatability††
< ±1% of reading or ± 0.02 NTU whichever is greater
Resolution
Menu settable up to 0.0000 over the entire range
USB
Printer or download logged data & calibration data & upload new firmware via thumb drive (USB flash stick).
Power Supply
UL, CSA & CE approved 12V DC, Wall Mount
Miscellaneous Specifications
1. USB port for logged data download and software updates.
2. Built-in Diagnostics
3. Three Year Battery Backup With no External Power
Operating Temperature Range
1°C – 40°C (34°F – 104°F)
Sample Temperature Range
0°C – 40°C (32°F – 104°F)
Dimensions
237mm W x 254mm L x 121mm H (10.75” W x 10” L x 4.75” H)
Shipping Weight
2.5 kg (5.5 lbs.)
Certifications
LC mark tested to UL and CSA. Conforms to CE
Warranty
1 Year from date of shipment
§
The specifications found in EPA method 180.1 are essentially the same as the specifications set out in method 2130B of the Standard
Methods for the Examination of Water and Wastewater 22nd edition and the specifications set out in ASTM Standard Method D1889-94. The M100 PLUS meets or exceeds, the specifications set forth in these methods.
Instrumental accuracy measured under controlled laboratory conditions at 25°C (77°F)
††
Both the accuracy and repeatability specifications for the M100+ are valid only for measurement of static (non-flowing) samples.
3

1.0 Overview

1.1 The M100+ Product Description

The M100+ is a process turbidimeter that allows for the laboratory measurement of the turbidity of sample water. The white light M100+ has been designed to meet the design criteria specified by the US EPA 180.1 on turbidity measurement. The infrared M100+ was designed to meet the design criteria specified in ISO 7027 and DIN 27027 for the measurement of the turbidity of a sample. Both models have long life lamps.
The M100+ instruments are equipped with a USB slot in order to download logged data and calibration information and to update operating software.
The M100+ is equipped with a color touch screen for easy and efficient operation.
Power Supply
Touch screen
Standard Inserted in Optics
USB Connector
4

2.0 Unpacking and Inspection

The table below indicates the items in the turbidimeter shipment.
Item
Quantity
M100+ Laboratory Turbidimeter
1
Accessory Kit for M100+ (2 empty sample cuvettes with caps)
1
Full Range Calibration Kit High Range A, High Range B, 1000 NTU, 10 NTU & 0.02 NTU
(shipped in a separate box to ensure freshness)
1
Quick Start Guide
1
Flash Drive Containing Full Owner’s and User’s Manual
1
Wall Mount Power Supply with Ferrite
1
A ferrite is included with the power supply. This should be looped twice through the cord near the side that connects to the instrument to eliminate EMI interference.
Remove the instrument from the packing carton. Carefully inspect all items to ensure that no visible damage has occurred during shipment. If the items received do not match the order, please immediately contact the local distributor or the HF scientific Customer Service department.
5

3.0 Site Selection

The M100+ must be used in a suitable dry location with adequate ventilation. It should not be located where chemicals such as chlorine are located. It must be operated and stored between 1 and 40°C. The instrument should be used on a stable table top and less than 90 cm (3 feet) from a power source.
Power Requirements:
The M100+ requires 12VDC for proper operation. The provided supply is rated at
1A.
The M100+ is supplied with a wall mount power supply intended for operation with
100-240 VAC, 47-63 Hz power source. BEFORE installing, verify that the line
voltage falls within these specifications.
Connect the wall mount power supply to the rear panel of the M100+. An optional Euro wall mount power supply is available.

4.0 Setup, Configuration and Calibration

Before using your M100+, you will need to familiarize yourself with the instrument’s operating modes and functions and perform some setup, configuration, and calibration.
The M100+ allows for the measurement of the turbidity of water. The
turbidity of the water is usually reported in Nephelometric Turbidity Units (NTU) but may be reported in Formazin Nephelometric Units (FNU). This is user selectable.
Readings up to 4400 are possible but readings above 4000 are outside of
the rated specifications. Readings above 4400 NTU will cause the display to flash indicating an over-range condition.
During normal operation, the instrument will be at the HOME screen. This is
this screen where measurements can be viewed.
Please note, some features depicted in this manual may change with future
firmware updates.