1.4 Details of the Manufacturer....................................................................................................................................... 6
2.1.1 Range of Use.................................................................................................................................................. 7
3 System Components.........................................................................................................................................................11
3.2 Breathing System ................................................................................................................................................... 17
3.4.1 Front Panel.................................................................................................................................................... 20
3.5 Menu ...................................................................................................................................................................... 25
3.5.2 Menu Diagram............................................................................................................................................... 28
4.1.1 System Self Test............................................................................................................................................ 34
4.1.2 Manual Leak Test .......................................................................................................................................... 36
5.1.2 Main Failure Alarm Test................................................................................................................................. 46
5.2 Test Gas Supply Pipeline........................................................................................................................................ 46
5.3.1 Monitoring without Oxygen............................................................................................................................ 46
5.3.2 Monitoring with Oxygen................................................................................................................................. 48
5.4 Installing and Testing Vaporizer............................................................................................................................... 49
5.6 Breathing System Test............................................................................................................................................ 51
6 Installation and Connection.............................................................................................................................................. 53
6.1 Install the Absorber Canister................................................................................................................................... 54
6.1.1 When to Replace Absorbent.......................................................................................................................... 54
6.2 Connecting Tubes and Lines................................................................................................................................... 55
6.2.1 Connect CO2 Monitor..................................................................................................................................... 56
6.3 Connecting Gas and Electricity............................................................................................................................... 57
6.3.1 AC Inlet......................................................................................................................................................... 58
6.3.3 Communication Port...................................................................................................................................... 59
6.3.4 Gas Inlet Module........................................................................................................................................... 59
7 Cleaning and Disinfecting................................................................................................................................................ 60
7.1 Cleaning and Disinfecting prior to first Use ............................................................................................................. 60
7.2 Cleanable Breathing System Components.............................................................................................................. 61
8 User Maintenance............................................................................................................................................................ 69
8.2 Maintaining Outline and Schedule........................................................................................................................... 69
8.2.1 User Maintenance......................................................................................................................................... 70
8.2.2 Useful Life Estimation.................................................................................................................................... 70
8.2.3 Replacement of Consumable Parts............................................................................................................... 70
8.3 Maintaining the Breathing System........................................................................................................................... 70
8.3.3 Calibrate CO2 Sensor.................................................................................................................................... 74
8.5 Replace the Fuses.................................................................................................................................................. 78
8.7 The IRMA Probe..................................................................................................................................................... 79
9 Alarm and Troubleshooting .............................................................................................................................................. 80
9.1 About Alarm............................................................................................................................................................ 80
9.2 Alarm Message List ................................................................................................................................................ 80
10.3.1 Gas Supply.................................................................................................................................................... 88
10.8.4 Gas Dynamics Performance.......................................................................................................................... 93
HEYER Medical AG (“HEYER”) holds the copyright to this manual, which is not publicly published, and
reserves the right to maintain it as a secure document. Refer to this manual when operating,
maintaining and repairing products only. Releasing information contained within the manual is
prohibited.
Proprietary materials protected by copyright law are included in this manual. No section of it can be
reproduced, copied, or translated into other languages without prior written approval from HEYER.
All instructions in this manual are considered to be correct. HEYER is not legally responsible for
damages caused by incorrect installation or operation. HEYER does not supply privileges endowed by
patent law to any other parties. HEYER is not legally responsible for the consequences of breaking the
patent law or of a third party violation.
Refer to this manual before using the product. The manual includes: Operating procedures which must
be performed with caution; information on actions and situations that may result in damage to the
equipment, and actions and situations that may cause bodily harm. HEYER is not responsible for the
security, reliability, and/or function of the equipment in the event that damage or other irregular actions
occur. Repairs for these malfunctions are not covered by the warranty.
HEYER has the rights to replace any content in this manual without notice.
1.1 Manufacturer Responsibility
HEYER is responsible for the security, reliability, and function of the equipment when the following
conditions are adhered to:
Installation, adjustments, and repairs must be performed by individuals authorized by HEYER;
Necessary electrical equipment and the working environment must be in accordance with national
and professional standards and the requirements listed in this manual;
Equipment must be used as stated in the operating instructions.
CAUTION: This equipment is not for family use.
CAUTION: Malfunctioning equipment may cause damage and/or bodily injury if repair
request are not submitted in a timely manner by the company or organization
using the equipment.
The paid theoretical framework diagram, calibrating method, and other instructions will be supplied to
the customer upon request. With the assistance of qualified technicians and when stipulated by
HEYER, specific equipment parts can be repaired by the customer.
1.2 Security, Reliability and Operating Conditions
HEYER is not responsible for the security, reliability and operating conditions of this product when:
The assemblies are disassembled, extended, or readjusted
The product is not operated correctly in accordance with the manual instructions; the power
supply that is used is incorrect, and/or the product is operated in an environment other than
optimal conditions per this manual.
In the event a product needs to be returned to HEYER, please follow these steps:
1. Obtain the right of return.
Contact our customer service department with the product number and type. The number is
marked on the surface of the product and is required for a return. Enclose a letter containing
the product number, type, and the reason for the return.
2. Transportation charges
Transportation and insurance charges must be prepaid by the user prior to shipping the
product to HEYER for repair.
1.4 Details of the Manufacturer
Apparatus: Anesthesia System HEYER Pasithec
Manufacturer: HEYER Medical AG
The HEYER Pasithec is a compact and integrated anesthesia transmitting system. The anesthetic
ventilator not only provides patients undergoing operations with auto ventilation, but also monitors and
displays the patient’s various parameters.
The anesthetic ventilator used in the system is controlled by a microprocessor, which internally
configures the monitor and the volume mode; other functions are optional.
Not all the optional functions available may be included in the manual. It is also possible to add other
equipment to the top or middle of this system for added functions. For more information with respect to
the existing product, please contact your local representatives.
WARNING: All Pasithec users must be trained.
WARNING: HEYER Pasithec is not suitable for use in an MRI environment.
2.1.1 Range of Use
Pasithec is applicable for patients of over 2 kg with standard configuration. Pasithec is for use in the
Operating Room and/or Emergency Room of a hospital, drug addiction treatment center, or other
medical facilities where anesthesia is used.
2.1.2 Contraindication
Pasithec is not suitable for pneumothorax patients
2.2 Symbols
Warnings and Cautions indicate all the possible dangers in case of violation of the
stipulations in this manual. Refer to and follow them.
WARNING: Indicates potential hazards to operators or patients.
AC135 Anesthetic Breathing System
AGSS Anesthetic Gas Scavenging Transfer & Receiving System
APL Adjustable Pressure Limit
BDU Basic Data Unit
C Compliance
C·G·O Common Gas Outlet
cmH2O Centimeters of Water
IPPV Intermittent Positive Pressure Ventilation
EEPROM Electrically Erasable Programmable Read Only Memory
FiO2 Fraction of Inspired Oxygen
Flow-t Flow-time Waveform
Freq Frequency
Freq
Minimum Frequency in PS Mode
MIN
GUI Graphical User Interface
I:E Inspiratory to Expiratory Ratio
L Liter
L/min Liters Per Minute
Manual Manual ventilation
ml Milliliter
MRI Magnetic Resonance Imaging
MV Minute Volume
Paw Airway Pressure
Pb Plumbum
PEAK Peak Pressure
PLAT Plat Pressure
MEAN Mean Pressure
Paw-t Pressure-time Waveform
PCV Pressure Control Ventilation
PEEP Positive End Expiratory Pressure
PIP Peak Inspiratory Pressure
PS Pressure Support Ventilation
P
Target Pressure
TARGET
SIMV Synchronized Intermittent Mandatory Ventilation
T
Inspiratory Time
INSP
TP Inspiratory Pause Time
Trigger Flow Trigger
T
Inspiratory Slope Time
SLOPE
UI User Interface
VT Tidal Volume
WDT Watch Dog Timer
∆P Differential Pressure
Enf. Enflurane
Hal. Halothane
Iso. Isoflurane
Sev. Sevoflurane
Des. Desflurane
ETCO2 End-Expiratory CO2 Concentration
INSCO2 Inspiratory CO2 Concentration
MAC Minimum Alveolar Concentration
CAUTION: The anesthetic system is intended to be used with the following monitoring
devices, alarm systems, and protection devices:
+ pressure measuring in accordance with 8.1 of ISO 8835-2;
+ system is to be equipped with an ANESTHETIC GAS SCAVENGING
TRANSFER and RECEIVING SYSTEM complying with ISO 8835-3 before
being put into service.
+ pressure limitation device in accordance with 51.101.1 of IEC60601-2-13;
+ exhaled volume monitor in accordance with 51.101.4 of IEC60601-2-13;
+ breathing system integrity alarm system in accordance with 51.101.5 of
IEC60601-2-13;
+ continuing pressure alarm in accordance with 51.101.6 of IEC60601-2-
13;
+ O2 monitor in accordance with ISO 21647.
+ CO2 monitor in accordance with ISO 21647.
+ ANESTHETIC monitor in accordance with ISO 21647.
WARNING: To avoid explosion hazards, flammable anesthetic agents such as Ether and
Cyclopropane shall not be used in the anesthetic workstation. Only use
anesthetic agents that comply with the requirements for non-flammable
anesthetic agents as specified in this manual. Halothane, Desflurane,
Sevoflurane, Enflurane, and Isoflurane have been found to be nonflammable agents.
WARNING: Independent means of ventilation (e.g. a self-inflating manually powered
resuscitator with mask) should be available whenever the anesthetic system
is in use.
WARNING: Do not use antistatic or electrically-conductive breathing tubes and mask.
WARNING: Contact with a liquid, such as anesthetic agent, results in damage within the
device.
WARNING: The incline angle should not exceed 10 degrees whenever the anesthetic
assistant power and controller are started with the main switch.
O can enter into flowmeter.
Descriptions of each control function at the front of Pasithec:
Item Description
3 Aux. O2 Flowmeter Provides O2 with a maximal flow of 15 L/min to Aux. O2.
8 Flowmeter with Flow
Control Knobs
Turn the knob counterclockwise to increase the flow; turn
clockwise to decrease the flow.
Reading point is top of float inside flow tube.
6 Top Light The top light will illuminate if top light is switched on. The top light
switch has two settings: on and off.
10 Pneumatic and
Electronic Switch
The switch can control electricity and gas and has three settings.
Electricity main switch: OFF, ON, ON; gas way main switch: OFF,
O2+N2O, O2+Air.
Function of electricity main switch: When system is shut off, the
The machine performs a system self test, and after the test, the
startup is complete. While system is in startup state, turn off the
main switch and be sure the main unit system is closed
completely.
Function of gas way main switch: when switch is in the OFF
position, O2, N2O and AIR cannot enter the flowmeter. When
switch is in O2+N2O position, O2 and N
2
When the switch is in O2+Air position, O2 and AIR can enter.
14 Writing Board The writing board can hold up to 10 kg and can be used by a
doctor during an anesthesia operation.
WARNING:When performing closed or semi-closed ventilation with breathing system,
the Fresh gas switch should be placed to Circle Absorber. Otherwise, there
will be anesthetic gas leakage and abnormal operation of the machine.
Descriptions of each control function at the side view of Pasithec:
Item Description
21 O2 Flush (O2+) Press the O2 Flush (O2+) button to supply the breathing
system with O2 at a high flow rate.
22 O2 Sensor Socket Socket to connect Oxygen Sensor for monitoring oxygen
concentration of absorber circle. When monitoring patient
oxygen concentration in inspiratory gas at the back of
inspiratory valve, the socket can be selected.
23 / 24 Driving Gas Outlet (Switch) Provide driving gas to other equipment. Pressure: 280
Front panel consists of display screen, keys, indicators, and a knob.
Figure 3-6 Front Panel
3.4.2 Keys
Manual Key Press the key to change original ventilation mode to manual
mode; Press again to change back to the original ventilation
mode.
Alarm Limits Key
MUTE Key
Spirometry Key This key shall toggle the display between the waveform
MENU Key Press the key and a “Menu” window appeared on the
Normal screen Key The key closes the “Spirometry” and other windows and
Press the key to open alarm window on the screen; Press
again to close the alarm window.
Press the key to mute the alarm for 110 seconds. New
alarms shall override the mute.
window and the two loop display configurations of the
Spirometry Loop Window described in section 4.5. The first
key press shall display the “Pressure-Volume” loop display
configuration. The second key press shall display the “FlowVolume” loop display configuration. After both
configurations have been cycled through the display, a third
key press shall return the display to the Normal Screen.
display screen; for more details refer to section 3.5.
The first menu key press after the initial power up will
display the calibrate menu, with “Start Calibration”
highlighted.
returns the screen to pressure and flow waveforms.
If the “Spirometry” and all other windows are already closed
when the NORMAL SCREEN key is pressed again, no
action shall occur.
AC indicator The indicator lights up when AC power is in use;
the indicator is dark when the AC power fails.
3.4.4 Navigator Knob
The user can adjust the rotary knob to select the menu item and modify the setup. It can be rotated
clockwise or counter-clockwise and pressed like other buttons. The knob may be used to select
options on the screen, in the system menu, and in the parameter menu.
The rectangular mark on the screen that moves with the rotation of the knob is the “cursor”. The
cursor can be used to select any menu item on which it lands.
Operating method:
Move the cursor to the item you wish to select
Press the knob
One of the following four situations will happen:
If the background color of the cursor becomes a contrasting color, the content in the
frame can change with the rotation of the knob.
A pull down menu or dialogue box may appear on the screen
The original menu will be replaced by the new menu.
Save setup
3.4.5 Screen Layout
The display of the ventilator is a color TFT, which can display the monitoring and setting parameters,
waveforms, and alarm information on the screen. See Fig. 3-7.
The screen has three areas: information area, monitoring area, and parameter setup area.
The information area is divided into seven areas: Alarm Indication, Alarm Messages, Date, Time,
Power Source, Patient Type and Ventilation Mode.
3.4.5.1.1 Alarm Indication
The bell icon is displayed when an alarm is present. When it appears, the color of alarm bell is white.
Pressing the alarm mute key places an “X” on the alarm bell icon for 110 seconds. Displayed below
the bell icon is a countdown timer that will display remaining silence time.
3.4.5.1.2 Alarm Messages
Technical Messages and Functional Alarm will be displayed in the alarm message area. High priority
alarms will be red. Mid- and low-priority alarms will be yellow. Up to 2 alarm messages can be
displayed on the screen. For more details, refer to Chapter 9.
3.4.5.1.3 Power Source
Located the left of Ventilation Mode tile are two icons: AC and Battery. The display status of the
Battery includes: Full, Charging and Exhausted. The display status of the AC includes: AC power up
and AC power down.
AC power up: (1) The Battery icon is solid and at 100% capacity when fully charged. (2) If in charging,
the Battery icon shows capacity alternately.
AC power down: (1) The Battery icon shows the current capacity. (2) The Battery can supply power for
the machine for about 15 minutes when the low battery alarm sounds.
3.4.5.1.4 Date
The display mode of Date has three types: MM/DD/YY, DD/MM/YY or YY/MM/DD.
3.4.5.1.5 Time
The display mode of Time has 2 types: 12 hour or 24 hour format.
3.4.5.1.6 Patient Type
The patient type shall be displayed as Adult or Child. By highlighting (black text on white background)
the Patient type tile and pressing, the navigator knob shall toggle the patient type between Adult and
Child. Changing patient type is possible in STANDBY mode only. The default patient type when the
machine is powered up is Adult.
3.4.5.1.7 Ventilation Mode
Pressing the navigator knob when the Mode tile is highlighted displays the current ventilation mode in
white text on a black background, and the “MODE” label remains displayed in black text on a green
background. Rotating the navigator knob clockwise allows the user to scroll through the ventilation
mode selections: STANDBY, IPPV, PCV, SIMV, PS and MANUAL. Rotating the navigator knob
counterclockwise allows the user to scroll through the settings in the reverse order. The default mode
when the machine is powered up is standby.
The monitoring area has two parts: Patient waveform and parameters. See Fig. 3-7.
3.4.5.3 Parameters Setup Area
The parameter setup area contains 8 tiles of fixed vertical height and fixed horizontal width; each tile
contains the tile’s parameter value.
Each tile represents a location that may be highlighted by the navigator knob. Current parameter
settings are displayed in reverse video in the parameter window when a parameter tile is highlighted
(black text on white background). Rotating the navigator knob clockwise allows the user to increase
the setting while rotating the knob counterclockwise decreases the setting. When the selection
reaches its maximum or minimum setting, the minimum or maximum setting is displayed.
Pressing the navigator knob selects the parameter setting which is displayed as white text within a
black field. Parameter settings shall be in normal video unless the parameter tile has been selected or
highlighted. Parameter tiles are populated per table 3-1 below.
A timeout shall occur when a parameter tile is selected or changed but not confirmed for 15 seconds.
Upon a timeout, the parameter setting reverts back to the previously confirmed value.
The parameter timeout shall be available for the following items:
• Ventilation Mode
• Patient Type
Table 3-1
In the following table ”Para.” is equated with “Parameter”.
Vent Mode
Para.
Setup 1
Para.
Setup 2
Para.
Setup 3
Para.
Setup 4
Para.
Setup 5
Para.
Setup 6
Para.
Setup 7
Para.
Setup 8
STANDBY Blank Blank Blank Blank Blank Blank Blank Blank
IPPV VT Freq. I:E TP PEEP Blank Blank Blank
PCV P
SIMV VT Freq. T
PS Blank Freq
MANUAL /
COM A Interface Communication extension interface communicates with the IRMA.
COM B Interface Communication extension interface is used for connecting external
communication equipment (RS232 interface).
SIGNAL Interface Signal interface is used for connecting display screen to main unit.
When calibrating or carrying out other functions, an explanation of the process will be displayed on the
screen.
The following diagram is an example.
Step 1
When the MENU key is pressed, a “Menu” window is displayed on the screen.
Note: The “Menu” window opens in the last viewed menu.
The following example illustrates how to operate the settings.
Step 2
Turn the knob to select the Paw-Low option; the selected option will appear as black text on a green
background.
Turn the knob and select “Return” to set other parameters. Follow the above directions to select the
corresponding submenu.
Press the MENU key to exit the screen.
The default alarm setting in the manual mode is “Enabled”. To adjust the setting:
When the Wave Display is in “Flow” mode, flow-time wave is displayed.
When the Wave Display is in “CO2” mode, CO2-time wave is displayed.
The Wave Display default setting is “Flow” when the machine is turned on.
To adjust Wave Display settings, follow these instructions:
Step 1: Turn the knob to select “Wave Display.”
Step 2: Press the knob, and then turn it to change the setting.
Plug the power cord into the AC power outlet and connect the pipeline gas supply and standby gas
supply. Turn on the cylinder valve of the standby gas supply and turn off the flowmeters of O2, N2O,
Air, and Aux. O2.
Step 2: Turn on the Power Switch
Turn the Power Switch to “ON”. The UI will power on and perform a System Self Test.
4.1.1 System Self Test
When the UI is powered on, the startup interface is displayed, followed by the LOGO interface.
Figure 4-1 LOGO Interface
The System Self Test interface will appear after the LOGO interface.
The System Self Test Complete interface will appear when the self test is complete.
Figure 4-3 System Self Test Complete interface
Select the “Continue” option to perform the Leak Test procedure.
Select “Bypass” to enter the STANDBY interface.
If the self test fails, please contact a service representative.
Step 1: Turn the knob and point the cursor to the current ventilation mode.
CAUTION Exit the menu before carrying out this step.
Step 2: Press the knob to ensure the grounding changed.
Step 3: Turn the knob to select the required ventilation mode.
Step 4: Press the knob for the pre-election state; setting parameters can then be changed.
Step 5: Point cursor to the ventilation mode and press the knob to save the setup.
4.1.7 Breathing Parameters Setup
Step 1: Turn the knob and point the cursor to the corresponding parameter.
CAUTION: Exit the menu before carrying out this step.
Step 2: Press the knob to ensure the grounding changed.
Step 3: Turn the knob to select the required ventilation mode.
Step 4: Press the knob to save the setup.
4.2 Start Mechanical Ventilation
WARNING: Before beginning, set the patient circuit installing and controlling correctly.
The following procedures assume that the system is on and the manual
reservoir gas is in ventilating mode.
Step 1: Check that the control settings coincide with the clinical settings.
Step 2: Select “Auto Ventilation”. To adjust settings, refer to section 4.1.6.
Step 3: If necessary, push the O2 flush button to inflate the bellows.
Step 1: Before starting manual ventilation, connect the reservoir bag to the gas operation port and
ensure the APL valve setting is correct. The APL valve is used to adjust the pressure limit of the
breathing system during the manual ventilation period.
Step 2: Set the ventilation mode to “Manual”, or press the “MANUAL/AUTO" key. To use the reservoir
bag to perform manual ventilation if necessary, push the O2 flush button to inflate the reservoir bag.
4.3.1 Stop Manual Ventilation
Set Ventilation Mode to STANDBY mode.
Manual ventilation can also be stopped by not pinching the reservoir bag or by pressing the
“MANUAL/AUTO" key, which will revert the system to mechanical ventilation.
4.4 Shutdown
Turn off gas supply and set the ventilation mode to “Standby” after the gas within the system is
emptied. Turn off the power.
4.5 Waveforms
1. Paw-t Waveform
Y-Axis: airway pressure; X-Axis: time. For more details, refer to section 10.8.6.
Figure 4-15 Paw-t waveform
2. Flow-t Waveform
Flow scale: -90~90 L/min.
Time-Axis: Positive inspiratory direction above 0 L/min level; minus expiratory direction below 0 L/min
level; no gas flow on 0 L/min level.
The “Save Reference” option can be used to preserve a loop for future reference. When a loop is
currently being displayed, or no loop is currently being displayed, Press “Save Reference” when either
a loop is currently being displayed or no loop is being displayed, and the next complete loop
(inspiration and expiration) will be saved as a reference.
After “Save Reference” is pressed down, the Pressure-Volume loop and Flow-Volume loop are
simultaneously preserved.
When “Save Reference” is pressed again, the system immediately deletes the current preserved
Pressure-Volume loop and Flow-Volume loop and preserves the two new reference loops.
When a user exits the loop function, the preserved loops are saved. When the working mode is
STANDBY, the two function loops are deleted.
Test interval: Pre-operative Checkout should be done each day prior to use by the first patient; prior
to the use of each subsequent patient, and after repair or maintenance.
The test schedule appears in the table below:
Prior to use by the first patient each day Prior to use by each patient
System check
Power failure alarm test
Gas pipeline and gas cylinder test
Flow control test
Vaporizer installation and test
Alarm test
Breathing system test
Ventilator test
WARNING: Do not use this system until the operation and maintenance manuals have
been read and understood.
•
Whole system connection
•
All warnings and cautions
•
User guide for each system module
•
Testing method for each system module
Before using this system:
Complete all tests of this section
Test all system modules
If a test fails, do not use this system. Please contact a service representative.
5.1.1 Check System
Breathing System test
Ventilator test
WARNING: Ensure the breathing circuit is connected correctly and in good condition.
Make sure:
1
Equipment is in good condition.
2
All the components are correctly connected.
3
Breathing circuit is correctly connected and in good condition and the breathing system contains
sufficient absorbent.
4
Vaporizer is in lock position and is filled with sufficient anesthetic.
5
The connection and pressure of pipeline gas supply system are correct.
WARNING: Do not leave the cylinder valves open during pipeline gas supply period.
The cylinder gas supply could be emptied and lead to insufficient supply in
case of pipeline malfunction.
6
The required emergency device is ready and in good condition.
7
The device for airway maintenance and organ cannula are ready and in good condition.
8
The applicable anesthetic and emergency medicine are ready.
9
Tighten and lock the truckles to ensure they are free of motion.
10
Connect the power cord to the AC power outlet. The power indicator light will glow when power is
connected. If the indicator does not glow, power is not being supplied to the machine. Plug the
power cord into a different socket, close the breaker, or replace the power cord.
Turn on the power. Screen will display the stand-by interface after running the self-test.
2
After 5 minutes, unplug the power cord.
3
Check the “Power Off Failure Alarm”; it has the following characteristics:
a) Alarm sounds
b) “Mains Failure!” message displays on the screen
c) Mains icon flickers
4
Reconnect power cord.
5
Alarm will stop.
5.2 Test Gas Supply Pipeline
CAUTION: A user must confirm that gas supply is connected correctly. The pressure
should be correct and there should be no leakages or faulty connections in
the gas circuits. Stop using immediately and check gas connections if any
of the above occur.
Disconnect all pipeline gas supplies if the reading of the pipeline pressure gauge is not zero.
•
Switch on O2 supply.
•
Adjust flow control to middle range.
•
Make sure the N2O pressure gauge is reset to zero.
•
Switch off O2 supply.
•
Make sure the O2 pressure gauge is reset to zero. The low O2 supply alarm should
sound when pressure drops.
5.3 Monitoring Flow Control
WARNING: Refer to Step 1 to 14 of Monitoring without Oxygen for monitoring without
oxygen.
Refer to Step 1 to 13 of Monitoring withOxygen for monitoring with oxygen.
5.3.1 Monitoring without Oxygen
WARNING: The monitoring system cannot be replaced by a link system. A fresh supply
of gas that contains oxygen may not sufficiently increase the level of
oxygen in the breathing circuit.
If N2O exists, it will pass through the system during the test, which should
be securely collected and removed.
Patients may be injured by improper gas mixture. The link system should
not be used if a proper ratio of O2 and N2O is not possible.
The following procedures can test whether the link system has serious
malfunction; however, it cannot determine whether the calibration is
correct.
CAUTION: The gas flow control valve should be adjusted slowly. Do not turn it quickly
when the reading of the flowmeter goes beyond the maximum or minimum
flow rate; the control valve could be damaged and/or break.
1. Connect the pipeline gas supply or open the cylinder valves slowly.
2. Turn counter-clockwise O2, N2O, AIR flow control valve.
Make sure no gas flows in the flowmeter.
3. Turn on the Pneumatic and Electronic Switch and choose the “O2+N2O” option.
4. Do not use this system if the battery is not fully charged or other ventilator failure alarm occurs.
•
Step 5 and step 6 are only applicable for the N2O system test.
WARNING: During Step 5 and Step 6, continue to use the link systems.
Only adjust control test (N2O in step 5 and O2 in step 6).
Adjust flow according to order (N2O first, O2 second).
If adjustable range is exceeded, adjust flow control and perform this step
again.
5. To test the flow increase of the link system:
•
Turn the N2O and O2 flow control clockwise to the end.
•
Turn the N2O flow control slowly counterclockwise.
•
Set the N2O flow control to the rate described in the following table. The O2 flow must be
higher than the minimum flow limit.
N2O flow
(liters per minute):
O2 flow (must be higher than the minimum flow)
(liters per minute):
1.5 0.5
3 1
6 2
9 3
6. Test the function of the link system when flow is reduced.
N2O flow
(liters per minute):
O2 flow (must be higher than the minimum flow)
(liters per minute):
6.0 2.0
3.0 1.0
0.6 0.2
7. Adjust full flow of all gases to ensure that the flowmeter float moves smoothly.
8. Shut off the oxygen supply by closing the oxygen cylinder valve or by disconnecting the oxygen
pipeline supply.
9. When using the flow control:
•
As pressure decreases, the oxygen-supply failure alarm must continuously sound.
•
Disconnect the flow of nitrous oxide and oxygen to be sure that the oxygen flow will be the
last to stop.
•
If the oxygen is the driving gas of the ventilator, the oxygen-supply failure alarm must
continuously sound.
10. Turn all flow control valves completely clockwise to close.
11. Change the Pneumatic and Electronic Switch to “O2+AIR”.
12. Adjust full flow of O2 and AIR to ensure that the flowmeter float moves smoothly.
13. Adjust the knob of the N2O flowmeter and ensure that there is no gas in the flowmeter.
WARNING: The monitoring system cannot be replaced by a link system. A fresh supply
of gas that contains oxygen may not sufficiently increase the level of
oxygen in the breathing circuit.
If N2O exists, it will pass through the system during the test, which should
be securely collected and properly disposed of.
Patients may be injured by improper gas mixture. The link system should
not be used if a proper ratio of O2 and N2O is not possible.
CAUTION: Before testing, perform test of the O2 monitoring device according to step 8
in section 5.3.1.
Follow the steps to test the flow control:
1. Connect the pipeline gas supplies, or slowly open the cylinder valve.
2. Turn all flow control valves completely clockwise to the end.
3. Turn on the Pneumatic and Electronic Switch.
4. Do not use this system if the battery is not fully charged or other ventilator failure alarms occur.
5. While testing the flow control, be sure gas is not flowing through any flow tubes. Step 6 and step 7
are only applicable for the N2O system test.
WARNING: During Step 6 and Step 7, continue to use link systems.
Only adjust testing of control (N2O in step 6 and O2 in step 7).
Adjust flow according to order (N2O first, O2 second).
The oxygen sensor must be calibrated correctly.
6 To test the flow increase of the link system:
•
Turn the N2O and O2 flow control clockwise to the end.
•
Turn the N2O flow control slowly counterclockwise.
•
Check that the O2 flow is increasing. The concentration of the oxygen tested must ≥25%
during the complete process.
7 To test the flow increase of the link system:
•
Set the N2O flow to 9 L/min.
•
Set the O2 flow to 3 L/min or higher.
•
Turn the O2 flow control valve slowly clockwise.
•
Check that the O2 flow is being reduced. The FiO2 tested must ≥25% during the complete
process.
8 Adjust the gases to full flow to ensure that the flowmeter floats moves smoothly.
9 Shut off the O2 supply by closing the O2 cylinder valve or by disconnecting the O2 pipeline supply.
10 While testing, be sure:
•
As pressure decreases, the oxygen-supply failure alarm sounds continuously.
•
Disconnect the flow of nitrous oxide and oxygen to be sure that the oxygen flow will be the
last to stop.
•
Air flow remains.
•
If oxygen is the driving gas of the ventilator, the oxygen-supply failure alarm must sound
continuously.
11 Turn all flow control valves completely clockwise to the close.
12 Reconnect O2 pipeline supplies or open the O2 cylinder valve slowly.
13 Turn off the Pneumatic and Electronic Switch.
Confirm the non-return valve in the Breathing Circuit module works normally:
The non-return exhalation valve will ascend during the exhalation period and descend during the
inhalation period.
WARNING: Objects in the breathing system can interrupt or disrupt the delivery of
breathing system gas, resulting in possible patient death or injury.
Do not use a testing plug small enough to slip completely into the breathing
system.
5.6.1 Check O2+ Button
Press the O2+ button (the sound of gas should be heard from the fresh gas outlet) then release. The
button should immediately drop back to its normal position and stop delivering gas.
5.7 Ventilator Test
1 Connect the simulation lung to the patient end.
2 Turn on Power Switch.
3 Set control options:
Ventilation mode: CMV mode
Ventilator: VT: 700 ml
f: 20 bpm
I:E: 1:2
P
: 40 cmH2O
limit
Anesthetic machine: O2 flow: less than 200 ml
All other gas: closed
4 Press the O2+ button to inflate the bellows.
5 Be sure:
•
Auto ventilation starts.
•
No low pressure alarms sound.
•
Ventilator displays the correct data.
•
The bellows ascend and descend during auto ventilation.
6 Set the O2 flow control to 5 L/min.
7 Ensure:
•
Ending expiratory pressure is between 2~3 cmH2O.
•
Ventilator displays the correct data.
•
The bellows inflate and scavenge during auto ventilation.
8 Set the ventilator control and alarm limits to the proper clinical level.
9 Turn off main supply and close all gas cylinder valves if not using the system.
10 Ensure that the following items are complete:
CAUTION: O2 monitoring must be used on this equipment. For the related stipulations,
refer to local standards.
CAUTION: According to the International Standard IEC 60601-2-13 / ISO 8835-1, this
equipment must use expiratory volume monitoring, O2 monitoring (in
accordance with EN 12342 or ISO 7767) and CO2 monitoring (in accordance
with EN 864 or ISO 9918).
CAUTION: Anesthetic monitoring (in accordance with ISO 21647:2004) must be done
while the anesthetic vaporizer is being used according to the European
standard EN 740 and International Standard IEC 60601-2-13 / ISO 8835-1.
WARNING: The operating room environment can be affected by expiratory gas. To
prevent damage and/or injury, the anesthetic must be tested regularly. The
operator must dispose of expiratory gas in a timely fashion and examine all
parts to minimize damage and malfunction.
WARNING: Be sure the gas pipeline supply hoses and the breathing circuit
components are non-poisonous, do not cause an allergic reaction in the
patient, and do not create dangerous by-products by reacting with the
anesthesia gas or the anesthetic.
WARNING: To prevent data reporting errors and malfunction, use cables, hoses, and
tubes supplied by HEYER.
CAUTION: Anesthetic in the absorber is dangerous. Ensure the soda lime in the
absorber does not dry out. Turn off all gas supplies when finished using the
system.
CAUTION: This system can be operated correctly under IEC 60601-1-2 interference.
Higher-level interference may cause alarm and result in auto ventilation
suspension.
CAUTION: To avoid equipment false alarm caused by high strength electric field:
•
Keep the electricity surgical conducting wire away from the breathing
system and the O2 sensor.
•
Do not put the electricity surgical conducting wire on any parts of the
anesthetic system.
CAUTION: To protect the patient while the electrical surgical equipment is being used:
•
Monitor and ensure that all life supporting and monitoring equipment
are working correctly.
•
Ensure that the backup manual ventilator can be used immediately if the
electrical surgical equipment fails.
•
Never use masks or hoses that can conduct electricity.
• Do not use the absorber if the anesthetic is chloroform or
trichloroethylene.
• The materials in the absorber are dangerous; avoid contact with the
skin or eyes. If contact occurs, clean the affected area immediately
and seek medical attention. Do not replace absorber during
ventilation.
• Do not replace the absorber if the Breathing System is in operation.
• Replace the absorbent often to prevent the deposits of non-metabolic
gas when the system is turned off.
• Check the color of the absorbent after each use. The original color of
the absorbent may be restored when not in use. Refer to the labels on
the absorbent for additional details.
• Carbon monoxide is released if completely dried absorbent comes
into contact with the anesthetic. Replace the absorbent for security.
• Perform leakage testing of the breathing system in manual mode after
installing the absorber.
The absorber in this system can be used repeatedly.
The capacity of each absorber canister is 1500 ml.
Only Air, Oxygen, Nitrous Oxide, Halothane, Enflurane, Isoflurane, Sevoflurane and Desflurane are
used for the absorber.
6.1.1 When to Replace Absorbent
Change in color of the soda lime in the absorbent indicates that it has absorbed carbon dioxide;
however, this color is not 100% accurate. To decide whether to replace the absorbent, use the CO2
monitoring machine.
Immediately remove absorbent that has changed color. Soda lime sometimes reverts to its original
color several hours after use.
The absorber is reusable. Follow the disassembling procedures:
1
Turn the absorber clockwise and then disconnect
it according to directions on the absorber. Turn
the absorber counter-clockwise to re-install.
6.1.3 Filling Absorbent
1. Remove the absorbent from the absorber.
2. Clean and sterilize according to section 7.3.1.1.
3. Fill the absorber with fresh absorbent once the canister is dry. Wipe soda lime off the edge of
the absorber, and then re-install it. Ensure the canister is airtight and that there is no leakage
or spillage.
6.2 Connecting Tubes and Lines
CAUTION: CO2 monitor (in accordance with ISO 9918) should be connected at Y-
piece of patient end.
CAUTION: Anesthetic machine should be used with anesthetic agent monitor.
Anesthetic agent monitor (e.g. Datex-ohmeda S/5 Anesthetic gas monitor)
should be used in accordance with ISO 21647:2004 and should be
connected at the Y-piece installed at the inspiratory port.
CAUTION: O2 sensor should be connected at inspiratory port of anesthetic absorber
circle.
Step 1: Connect breathing tube.
Connect the two tubes on the inspiratory and expiratory ports, respectively. Connect the Y-piece to the
patient end.
Step 2: Connect reservoir bag.
Connect the reservoir bag onto the bag arm port of the breathing system.
Step 3: Connect oxygen sensor.
Connect the oxygen sensor between the inspiratory port of the breathing system and patient system.
Turn and insert the oxygen sensor vertically in the connector.
Step 4: Connect the oxygen sensor cable.
Connect one end of the cable (RJ11 connector) to the bottom of the oxygen sensor. Connect the other
end of the cable (round connector) to the oxygen sensor port at the side of the machine.
6.2.1 Connect CO2 Monitor
Step 1
Connect one end of the CO2 monitor to the Ypiece of the patient end and the other end to the
reservoir bag.
Step 2
Plug the data connecter into the COM A port at
the back of ventilator.
Connect one end of the anesthetic agent
monitor to the Y-piece of the patient end and the
other end to reservoir bag.
Step 2
Plug the data connecter into the COM-A port of
the back of ventilator.
6.3 Connecting Gas and Electricity
WARNING: IEC 60601-1-1 applies to combining medical electrical equipment and for
combinations of at least one item of medical electrical equipment with one
or more items of non-medical electrical equipment. Even if there is no
functional connection between the individual pieces of equipment, when
they are connected to an auxiliary main socket outlet, they constitute a
medical electrical system. It is essential that operators are aware of the
risks of increased leakage currents when equipment is connected to an
auxiliary main socket outlet.
WARNING: The equipment connected to the power outlet will increase electric current
leakage. Test electric current leakage regularly.
WARNING: A malfunction of the central gas supply system may cause one or more
devices connected to it to shut down simultaneously.
CAUTION: Disconnect the anesthetic workstation from the gas supply after use to
prevent contamination or pollution of the pipeline system.
CAUTION: Only the medical gas supply should be used. Other types of gas supply
may contain water, oil or other pollutants.
WARNING: Note that all gas supply connectors have different dimensions and
Two RS-232 ports (COM-A and COM-B) on the rear of the display are communicating ports. The
connection of RS-232 permits serial inlet and outlet of command and data.
The interface of COM-A port is a DB9/M needle connector and is used for data communication or to
connect the gas monitor module. The Signal Definition of the port is DTE.
Pin # Signal Name Signal Description
2 RXD RS232 Receive Data, input of the machine.
3 TXD RS232 Transmit Data, output of the machine.
5 GND Signal Ground / Common
9 5V DC5V power, provide output power in gas monitor module, <200mA.
In data communication, this pin and the RI output signal of RS232C
indicates electrical compliance.
Other NC Not connected.
The interface of COM B port is a DB9/M hole connector and is used for data communication. The
Signal Definition of the port is DCE.
Pin # Signal Name Signal Description
2 TXD RS232 Transmit Data, output of the machine.
3 RXD RS232 Receive Data, input of the machine.
5 GND Signal Ground / Common
Other NC Not connected.
6.3.4 Gas Inlet Module
The Gas Inlet Module includes pipeline gas supply and standby gas supply.
The pipeline connector and standby gas supply connector all adopt DISS (Diameter-indexed safety
system), which can prevent wrong connections.
Lock the castors of the machine to prevent movement.
Step 2
Turn the handle (arrow pointed) of absorb circle 90°
counter-clockwise from closed state.
Step 3
Hold the handle (see arrow) of absorb circle with
two hands and then pull it out horizontally from the
supplying board of absorb circle.
The red part of the supplying board is a heater. Do
not touch it!
Refer to “Disassembling the Absorber” in section 6.1.2. Auto cleaning with agent or disinfector
Clean the absorber in the agent or disinfector according to the specified cleaning procedure.
Put the absorber in the heat-up room and set at room temperature or a maximum temperature of
80°C.
Higher-level sterilization is recommended if the agent and disinfector cannot sterilize equipment.
Manual cleaning
Rinse the absorber.
Immerse the absorber completely in the sink filled with water and cleaning agent for three minutes at a
temperature of 40°C.
Rinse the absorber.
Higher-level sterilization must be performed after cleaning by hand.
Advanced Sterilizing
The absorber must be cleaned before advanced sterilizing.
The absorber can be placed in high temperature and high pressure conditions. The maximum
temperature recommended is 80°C (176°F).
Put the soda lime into the absorber after drying; tighten the knob. Clear all soda lime debris.
7.3.1.2 Airway Pressure Gauge
Clean the Airway pressure gauge with a soft cloth soaked with a water soluble sterilizing agent.
7.3.1.3 APL Valve
Clean APL valve with soft cloth soaked with a water soluble sterilizing agent. Dip in soap water or
cidex solution to clean and disinfect.
7.3.1.4 Inspiratory Valve and Expiratory Valve
Dismount the cover of the inspiratory and expiratory valves by rotating them counter-clockwise.
Remove the valve patch, clean the valve seat and put back together.
Clean cover of the inspiratory and expiratory valves and valve patch with gauze soaked in a water
soluble sterilizing agent.
After all parts are cleaned and dry, replace them.
Check for leakage and the movement of the inspiration and expiration valves according to instructions
in the manual.
Handle all parts with care to prevent damage.
7.3.1.5 Absorber Circle Integration
Steam (not more than 50°C) or immersion disinfection can be used. In case of immersion, all
sterilized parts must be dried with high pressure air or oxygen before reuse.
When installing, pull the supply board of the
absorber circle as far as it will go. Hold the handle
(see arrow) of absorb circle with two hands and
insert it in the supply board. Push the supply board
and absorber circle together to ensure a tight seal
with seat module.
Turn the handle (see arrow) of absorber circle 90°
clockwise from open to lock the circle.
7.3.2 Bellows Assembly
This section covers disassembling, assembling, cleaning and sterilizing the bellows assembly. Read
all parts of this section before disassembling, assembling, cleaning and sterilizing to avoid equipment
damage and/or patient injury.
CAUTION: Only the folding gasbag is made of latex.
Cleaning and sterilizing must be performed at the same time. Follow instructions for the common
bellows assembly sterilization methods.
Sterilizing after general patient use:
Clean the inner and outer parts of the bellows assembly in a soap-and-water solution. Rinse
thoroughly in cold water, and dry with soft cloth.
Immerse plastic and latex instruments in 70~80% ethyl alcohol for half an hour. Take them out using
the aseptically transmits pliers, then store in clean containers. Repeat this step before next use.
Components made of metal and glass can be sterilized with high pressure steam. When the steam
pressure is increased by the autoclave, the rising temperature can concrete the bacterium protein
rapidly to kill bacteria. In 1.05 kg/cm2 steam pressure, the temperature rises to 121°C. Al l bacteria and
most sorus can be killed if this temperature is maintained for 15~25 minutes.
Sterilizing after special infection or infectious patient use:
Open pulmonary TB, pulmonary abscess, pseudomonas, tetanus aeruginosa infection, gas gangrene
or infectious hepatitis is included. Used bellows assembly components must be completely sterilized
according to preliminary and final disposal procedures.
1) Preliminary disposal: Perform in accordance with the isolated disposal stipulation. Collect and
leave all the used bellows assembly components during the operation process in the operating
room. Immerse the bellows assembly components in the 1:1000 benzalkonium bromide or 1~5%
cresol for 30 minutes after finishing the operation.
2) Final disposal: Perform the final sterilizing disposal after the bellows assembly components are
processed by the above-mentioned preliminary disposal:
•
Scrub the bellows assembly in a soap-and-water solution. Thoroughly rinse in cold water,
and dry.
•
If conditions permit, suffocating the components directly contacted with patients with formald
or oxirane is preferred, or perform immersing sterilization respectively. For example: the
components used by open pulmonary TB patients must be immersed in 3% cresol for 30
minutes; the components used by tetanus aeruginosa infection patients must be immersed in
0.2% potassium permanganate for 30 minutes; the components used by gas gangrene
patients must be immersed in 0.1% chlorhexidine for 30 minutes; the components used by
pulmonary abscess patients must be immersed in 0.1% benzalkonium bromide for 60
minutes; the components used by pseudomonas patients must be immersed in 0.1%
benzalkonium bromide for 120 minutes.
•
The components being immersed need to be rinsed by water and dried for next use.
•
Scrub and rinse the components that had indirect contact with patients with 1-3% phenol
solution or soap-and-water solution and water. Irradiate them by using the ultraviolet ray for
30 minutes if necessary.
Use the lubricant approved for anesthesia or O2 equipments’ use.
•
Never oil or grease any anesthesia or O2 equipment. In general, oils and
greases oxidize readily, and in the presence of O2 are highly flammable.
•
All the covers or housings for the system use must be made of static-proof
material, as static material may cause fire.
WARNING: Follow sterilizing control and security stipulations because used equipment
may contain blood and body fluids.
WARNING: Movable components and detachable parts can cause injury. Use caution
when system components and parts are being moved or replaced.
WARNING: Shock and strong vibration during transportation can break the fragile glass
cover of the flowmeter.
WARNING: Disposal of waste or invalidated apparatus must be in accordance with local
laws.
8.1 Repair Policy
Do not use malfunctioning equipment. Make all necessary repairs, or have the equipment serviced by
an authorized HEYER Service Representative. After repair, test the equipment to ensure that it is
functioning properly and in accordance with the manufacturer’s published specifications.
To ensure full reliability, have all repairs and service done by an authorized representative. If this is
not possible, replacement and maintenance of parts in this manual should be performed by a
competent, trained individual with experience in Anesthesia Systems repair and appropriate testing
and calibration equipment.
CAUTION: No repair should ever be undertaken or attempted by anyone without proper
qualifications and equipment.
It is recommended that damaged parts be replaced with components manufactured or sold by
HEYER. After any repair work, test the unit to ensure it complies with the manufacturer’s published
specifications.
Contact the nearest HEYER Service Center for service assistance. In all cases, other than where our
warranty is applicable, repairs will be made at current list price for the replacement part(s) plus a
reasonable labor charge.
8.2 Maintaining Outline and Schedule
The following schedule is a recommended minimum standard based upon normal usage and
environmental conditions. Frequency of maintenance for the equipment should be higher if your actual
schedule is more than the minimum standard.
Ventilate the system, open flowmeter, and make sure that the float moves
up and down smoothly. It can prevent blocking and clinging.
Monthly Perform 100% O2 sensor calibration.
Test leakage of bellows assembly.
When cleaning and
installing
Check if any components are broken, and replace or repair them if
necessary.
As required Replace O2 sensor (one year generally).
Open the drain valve and replace absorbent in the absorber.
8.2.2 Useful Life Estimation
CAUTION: The useful life of the following parts should be considered in normal
environment and operating requirements.
Main unit 5 years
Components 5 years
Vaporizer 5 years
Absorb circle 5 years
8.2.3 Replacement of Consumable Parts
Parts are replaced at multiple intervals from the date of installation.
Consumable parts 6 months 12 months 36 months
Check/replace internal battery (3 years) Not needed Check Replace
Check/replace folding bag Not needed Check Replace
Check/replace U-shaped ring (U-ring) Not needed Check Replace
Check/replace O-shaped silicone airtight ring
∅6X1.8 (for micromanometer)
Check/replace O-shaped silicone airtight ring
∅4X1.8 (for micromanometer)
Check/replace O-shaped rubber airtight ring
14×2.65 (on by-pass valve).
Replace O2 sensor Check Replace Replace
8.3 Maintaining the Breathing System
Parts that are broken, crushed, worn or distorted must be replaced immediately when cleaning the
breathing system.
Refer to the sections corresponding to reassembly and testing.
WARNING: Follow local laws when disposing of biohazardous materials. Do not burn.
Replacement steps:
1 Pull out the connector of sampling line from O2 sensor.
2 Replace it with a new one, and connect the sampling line to O2 sensor.
8.3.2 Calibrate O2 Sensor
WARNING: Do not perform the calibration steps when the system is connected to a
patient.
When calibrating O2 sensor, ambient pressure must be equal with
monitoring pressure of delivering O2 in the patient circuit.
If operating pressure is not equal with calibrating pressure, the accuracy of
reading may exceed range stated.
8.3.2.1 Calibrate 21% O2 Sensor
It takes more than 3 minutes to perform 21% O2 sensor calibration.
Before performing 100% O2 sensor calibration, 21% O2 sensor calibration must be finished.
Step 1
Press MENU key and a menu window appears on the screen.
Press the knob and the cursor appears on the O2 Sensor Calibration option.
Step 3
Press the knob and a new window appears right side.
Perform the operation in accordance with the prompt on the screen.
Disassemble O2 sensor from the Y-piece and place it in the air for more than 3 minutes.
Press the knob to perform the calibration.
In the process of calibration, the word “Calibrating” displays on the screen.
“Calibration successful” will be displayed on the window after the O2 Sensor calibration is successfully
completed. Replace the O2 sensor and press “Finish” to continue.
If 21% calibration fails, the word “Calibration failed” is displayed in the window.
If calibration fails:
•
Repeat these steps to calibrate it again.
•
If it fails again, replace the O2 sensor and recalibrate it.
This item must be performed by an authorized service engineer.
8.3.3 Calibrate CO2 Sensor
Step 1
Turn the knob to select “CO2 sensor Calibration” in the “Calibrate” submenu.
Step 2
Press the knob and a new window appears on the right side.
Perform the operation according to the prompts on the screen.
Disconnect the Gas sensor from the Y-piece, and place it in the air for no less than 15 minutes.
Remove the airway adapter (or a new one) until the light turns green.
Wait 1 minute till the State-Area turns green (OK) and select it.
Press NEXT to perform the calibration.
In the process of calibration, the word “Calibrating” displays on the screen.
“Calibration successful” will be displayed on the window after the calibration is completed successfully.
Reinstall the Gas Sensor to the breathing system. Select “Finish” to complete the calibration.
Turn the knob to select “Agents Calibration” in the “Calibrate” submenu.
Step 2
Press the knob and a new window appears on the right side.
Perform the operation according to the prompt on the screen.
Disconnect the Gas Sensor from the Y-piece and place it in the air for no less than 15 minutes.
Remove the airway adapter (or a new one) until the light turns green.
Wait 1 minute till the State-Area turns green (OK) and select NEXT.
Press the NEXT knob to perform the calibration.
During calibration, the word “Calibrating” displays on the screen.
“Calibration successful” will be displayed in the window after the calibration is completed successfully.
Reinstall the Gas Sensor to the breathing system. Select “Finish” to complete the calibration.
8.4 Maintaining O2 Sensor
Perform the calibration periodically; refer to section 8.2.1.
When the machine is not in operation, avoid placing the O2 sensor in a high oxygen concentration to
increase its life.
The O2 sensor’s useful life is 12 months.
SV-03A is the O2 sensor recommended by the manufacturer.
CAUTION: For more detailed information, refer to technical data that is regularly
updated by the manufacturer.
The effect of temperature on the performance of a capillary barrier O2 sensor is relatively small.
Changing the temperature from +20°C to -20°C will t ypically result in 10% loss of the output signal. In
contrast, temperature has a much greater effect on solid membrane O2 sensors. The diffusion of gas
across the membrane is an activated process and as a result has a large temperature coefficient.
Typically a 10°C change of temperature doubles the output signal from the sensor. Solid membrane
sensors require temperature compensation as a result, and many have thermistors in them.
WARNING: To prevent injury or death, disconnect the system from the power supply
before replacing fuses.
WARNING: Replace fuses with only those specified type to prevent damage to the
equipment.
CAUTION: The fuse is fragile, so replacement should be done carefully. Do not use
excessive force.
Replacing steps:
1. Plug the screwdriver into groove on the end of the fuse box.
2. Turn counterclockwise 3 to 5 times and then pull out fuse tubes carefully.
3. Take off fuse tubes.
4. Enclose the new ones.
5. Push fuse tubes gently into place.
6. Turn clockwise 3 to 5 times with screwdriver to tighten.
7. Connect mains supply.
8.6 Maintaining Battery
Specification
24V, lead-acid battery.
Charge: 8 hours typically
Cautions
Charge: Connect mains supply; the system will maintain auto-charging battery. Charging time is more
than 8 hours.
Battery supply will last 120 minutes.
The alarm “Low Battery!” should be displayed on the screen when the battery is nearly drained. The
user/operator should connect mains supply to charge battery and avoid the system shut-off.
Do not disassemble battery device without valid authorization.
Do not short-circuit the battery between the positive plate and the negative plate.
Storage
Batteries should be charged every 3 months if battery power exceeds 3 months.
Store batteries in a dry, cool place.
If battery is damaged, it must be replaced to avoid damage. Contact a service representative.
CAUTION: Only an authorized services representative can replace the battery. If the
battery will not be used, contact a service representative to disconnect the
battery. Dispose used batteries in accordance with local laws.
The IRMA probe is intended for monitoring CO
for a child.
Never sterilize or immerse the IRMA airway adapters in liquid. They can be cleaned using a cloth
moistened with ethanol or isopropyl alcohol. Replace every two weeks.
WARNING: No repair should ever be undertaken or attempted by anyone without
proper qualifications and equipment.
9.1 About Alarm
CAUTION: If alarm occurs, ensure the patient safe before attempting to diagnose
the problem.
Alarm messages display on the top of the screen.
Figure 9-1 Alarm message area
The high priority alarms must be attended to immediately.
Priority Volume Silence Prompt Alarm bell
High 5 tones, 2 hurry;
Every 8 seconds
Medium 3 tones
Every 10 seconds
Low 2 tones
Non-repeat
110 seconds Red background, 3 “!”
Display frequency: 2Hz
110 seconds Yellow background, 2 “!”
Display frequency: 0.5Hz
---- Yellow background, 1 “!”
Display until alarm stops.
Red, flickering
Yellow, flickering
Yellow
When two or more alarms with equal priority are generated at the same time, the signals will be
displayed in turn.
The operator can hear the alarm sound or distinguish the priority of alarm from at least 4 meters. The
alarm message is visible from more than 1 meter away.
CAUTION: There are two alarm display areas, and the array of alarms is accorded
priority from high to low.
CAUTION: When the alarm is silenced, the alarm bell has dashed “X” on it.
CAUTION: A potential hazard can exist if different alarm presets are used for the
same or similar equipment in any single area.
CAUTION: The auditory alarm signal sound pressure range is between 62dB and
75d (measured from a distance of 1 meter).
9.2 Alarm Message List
CAUTION: Patient safety comes first during an operation. Repair the problem once
the operation is finished.
CAUTION: Operation instruction is not included in the alarm message list.
is greater than 10 cmH2O in
Manual mode and lasts more
than 10 seconds, machine
generates continuous high
airway pressure alarm. In
other modes, if current value
of airway pressure (PEEP
value set + 10 cmH2O) lasts
more than 10 seconds,
machine generates
continuous high airway
pressure alarm.
This alarm is decided by
upper machine and displayed
in any mode other than
STANDBY mode.
High Paw greater than upper limit
for 2 consecutive breath
cycles.
Settings of VT higher.
Patient airway blocked.
Exhalation valve blocked.
Machine response:
Immediately enter expiratory
cycle.
No Switch to bag
mode,
manually
ventilate
patient.
Check tubes
and sampling
lines and clear
any blockages.
Yes Reset upper
limit of Paw.
Check
expiratory
cycle, and
clear any
blockages.
Check VT
settings.
Check airway
of patient, and
clear any
blockages.
If the alarm
continues,
please contact
eligible service
representative.
----
Low Airway
Pressure
Mid SW417.2 For Freq >= 4:
Airway pressure below
pressure alarm low limit for >
Yes Reset lower
limit of Paw
Check the
parallel
sampling lines.
15 seconds.
For Freq < 4:
Airway pressure below
pressure alarm low limit for >
30 seconds.
Low FiO2 High FiO2 less than lower limit.
Over-compensation of air or
N2O.
O2 sensor non-calibrated.
O2 sensor failure.
No Reset lower
limit of FiO2.
Reduce
compensation.
Perform the
----
calibration.
Replace O2
sensor.
High FiO2 Mid FiO2 greater than upper limit. No Reset upper
tidal volume is:
<20 ml (for patient Type =
Child) or <150 ml (for patient
Type=Adult)
For a period of:
>60 seconds (Vent Mode =
MANUAL) or >35 seconds
(Vent Mode=(SIMV or PS)
and Breathing Frequency
setting < 6) or >30 seconds
(all other automatic ventilation
modes and Breathing
Frequency settings)
The Minute Volume (MV),
Tidal Volume (VT) and
Breathing Frequency
measurements are blanked
during the APNEA condition
and resume after the
condition is resolved.
Ventilation mode changes
(excluding STANDBY) during
an APNEA condition shall not
restart the apnea timing.
Breathing circuit disconnects
during automatic ventilation
shall cause an apnea alarm.
limit of MV.
Yes Reset lower
limit of MV.
Check patient
end.
N.A. Reset upper
limit of f.
Examine the
patient and
confirm
independent
respiration
exists.
APL valve is on Turn APL valve to off
Canister is not sealed well Reinstall or remove the natrium
calcareousness grains at the joint
Corrugated tubing is broken or the
Replace or reinstall
connector is loose
Valves loose Tighten them
APL valve is adjusted incorrectly Adjust it properly
APL valve failure Please contact eligible service
representative.
Power supply cable is unplugged
Power switch is off
Fuse burned out
Plug in power supply cable
Turn on power switch
Replace with a new one
Power supply is interrupted Use manual ventilation
Maximum pressure
alarm sounds
continuously
Minimum pressure
alarm sounds
continuously
No indication from
the airway pressure
gauge
Tidal volume
readings do not
display normally
The folding gasbag
is inflated
excessively
Breathing System is occluded
Patient’s respiratory tract is occluded
Maximum pressure setting is too low
Ventilator parameters changed
Breathing System leaked
Alarm settings are too high
Patient’s compliance changed
Pressure sampling pipe is disconnected
or broken
Pressure sampling pipe is disconnected
Gas supply exhausted
Flow sensor is disconnected
The sealing ring of bellows base is
broken
Bellows base is broken
Gas scavenging port is occluded
Malfunctioning waste gas scavenging
system created excessive resistance or
vacuum
Check and adjust Breathing System
Check the patient
Readjust the alarm setting
Recalculate the ventilator parameter
Check the pipeline for leak
Reset the alarm settings
Check the patient
Check the pressure sampling pipe
Reconnect the pressure sampling
pipe
Replace the gas supply
Reconnect the flow sensor
Check the bellows and replace the
broken parts, then reassemble it and
carry out testing procedure
according to section 6.2.5
Remove the occlusion
Repair waste gas scavenging
system
All specifications are approximate and may be changed without notice.
CAUTION: Do not use the Pasithec in a volatile environment.
CAUTION: Do not place heavy objects on the top of the machine or in the draws. The
top board and working table can bear 25kg, and the drawer can bear 10kg.
System
Castor 125 mm, with breakers on the front castors.
Drawer 115 mm (H) × 485 mm (W) × 325 mm (D)
Display 10.4’ TFT LCD
Cylinder gauge (O2) Scale: 0 to 25 MPa (0 to 3500 psi). Resolution: 1 MPa.
Cylinder gauge (N2O) Scale: 0 to 10 MPa (0 to 1400 psi). Resolution: 0.4 MPa.
Pipeline gauge (Air & O2 & N2O) Scale: 0 to 1 MPa (0 to 145 psi). Resolution: 0.05 MPa.
Voltage: 100~240 VAC, 50/60 Hz
Maximum input current: 10A
Fuse of AMSO:
250V/10A, ∅5X20 (F)
Socket of AMSO: 4
Fuse above AMSO:
250V/2A,∅5X20 (F)
Maximum output current of AMSO: 1.5A (each); 6A (total)
Earth resistance: <0.2Ω
WARNING: The connection of equipment to the auxiliary mains socket outlets can
increase the patient leakage currents to values exceeding the allowable
limits in the event of a defective protective earth conductor.
10.4.2 Pneumatic
Pipeline supply: O2, Air, N2O
Connect to pipeline: DISS-male, DISS-female, NIST (ISO 5359).
All fittings used to connect O2, Air and N2O pipeline gas
supply are all ready.
Display pressure: Color coded gauges
Connect vaporizer: Two Selectatec®-type interface vaporizers
Input pressure at pipeline inlets: 280~600kPa
Connect to Breathing System:
Insp. Port connector: ∅22 OD /∅15 ID
Exp. Port connector: ∅22 OD /∅15 ID
Breathing System to AGSS: ∅30 OD
10.5 Electromagnetic Compatibility
Changing or reassembling this equipment without authorization may cause electromagnetic
compatibility problems. Contact HEYER for assistance. Designing and testing this equipment is in
accordance with the following stipulations.
WARNING: Using cell phones or other radio equipment near this product may cause
malfunctions. Closely monitor the working condition of this equipment if
radio equipment is in close proximity.
Using other electrical equipment in this system or nearby may cause interference. Check if the
equipment works normally in these conditions before using on a patient.
Be careful of the following when HEYER Pasithec is connected:
Do not put any object which is not in accordance with EN60601-1 within 1.5 meters of patients.
An isolated transformer must be used for alternating current supply (in accordance with IEC60989).
Additional protective ground wires are equipped if all the devices (for medical or non-medical use) are
connected to Pasithec by using signal input/signal output cable.
If a portable all-purpose outlet is used as the alternating current supply, it must be in accordance with
EN60601-1-1 and cannot be put on the floor. Using other portable all-purpose outlets are not
recommended.
Do not connect the non-medical equipment directly to the alternating current outlet on the wall. Only
the alternating current supply of the isolated transformer can be used. Otherwise, the surface leaking
current may exceed the range permitted by EN60601-1 under the normal conditions, and equipment
malfunction may cause injury to patients or operators.
HEYER Pasithec is equipped with all-purpose alternating current outlet for connecting to other medical
equipment. Do not connect non-medical equipment to these outlets, or the surface leaking current
may exceed the range permitted by EN60601-1 under normal conditions and the equipment may
malfunction and cause injury to patients or operators.
A complete system current leaking test (according to EN60601-1) must be performed after any
equipment is connected to these outlets.
WARNING: Medical electrical equipment operators should avoid non-medical electrical
equipment and patients.
10.6 Operation Theory
The HEYER Pasithec is a continuous flow anesthetic system. It is equipped with an airway which can
transfer O2, N2O, AIR and inhalation anesthetic drugs. Airway parts can use O2, N2O and AIR, and the
O2, N2O and AIR which is needed reduce to 280 kPa~600 kPa can enter into the machine through a
high pressure pipeline. Once through the protective pressure reducer in the machine, O2 and Air enter
into Pneumatic and Electronic Switch, while N2O-cut valve can only open when the pressure of O2
more than 50kPa (the FiO2 LOW alarm will be active when the pressure of O2 is less than 0.2MPa in
the state of startup) and the N2O can enter into Pneumatic and Electronic Switch. Otherwise, N2O can
not enter the airway of the machine. The state of gas entering into flowmeter can set to three states
with the Pneumatic and Electronic Switch: OFF, O2+ N2O, and O2+Air. The flow of O2, N2O and AIR
can be adjusted by the knob of the flowmeter. Adjust O2 and N2O proportionally to ensure the O2
concentration is no less than 25%. O2, N2O and AIR are mixed in the flowmeter; the mixed gas in the
anesthetic drug vaporizer can take away part of anesthetic drug. Fresh gas can be transferred to the
patient breathing cycle or C.G.O by the fresh gas selection switch. The O2 from O2 flush is transferred
to the patient breathing cycle or C.G.O directly without going via the flowmeter and vaporizer. The
mixer gas or O2 in the patient breathing cycle can maintain the breathing of a patient by setting
ventilation parameters and ventilation mode via machine or manually. The mixed gas or O2
transferring to C.G.O can be used to maintain patient breathing via Maishi circle.
Flow compensation range: 1 to 10 L/min Compensation of fresh
gas
Absorbent Capacity: 1800 ml
Connection Common Gas Outlet: ISO 5356 connector
Gas components: O2, N2O, Air, Anesthetic Agent
Leakage of breathing
system
Resistance of breathing
system
At pressure of 3 kPa (0.4psi):
Leakage of flow: ≤150 ml/min
At flow of 60 L/min:
Resistance of exhalation: ≤0.6 kPa;
Resistance of inhalation: ≤0.6 kPa.
At flow of 30 L/min:
Resistance of exhalation: 2.2 kPa;
Resistance of inhalation: 2.2 kPa.
Patient cycle of small resistance should be used in accordance with the
relevant standard.
Resistance of APL valve At flow of 3 L/min, resistance of flow: 0.05 to 3 kPa;
At flow of 30 L/min, resistance of flow: 0.1 to 0.5 kPa.
Leakage of connector Resistance of flow: ≤50 ml/min. (APL valve close fully)
Resistance of check valve Dryness: ≤0.15 kPa
The pressure generated by a wet unidirectional valve: <0.14 kPa;
The pressure to open a wet unidirectional valve: <0.1 kPa
Compliance of absorber <50 ml/kPa
Internal volume 2.5 L approximately
O2 flush 25 to 75 L/min
O2 failure alarm and the associate cut-off device:
of airway system:
Noise of whole unit: Not more than 60 dB(A) in normal operation (not including alarm voice)
Warm-up time: More than 5 minutes
Compliance: Not more than 40 ml/kPa
Electrical safety: Meet requirements for Class I, type B equipment specified in
EN60601-1 Medical Electrical equipment: Part one: General
requirement for safety.
Minute volume: Max 20 L/min
Inspiratory flowrate: Max 75 L/min
Pressure transmission
0~9% or
0~74 mmHg
1~10% or
1~75 mmHg
1~10 mmHg 1 mmHg 5 mmHg Unavailable for STANDBY mode
2
1% or
1 mmHg
1% or
1 mmHg
----- Unavailable for STANDBY mode
6%
Unavailable for STANDBY mode
or 50 mmHg
High
Hal. Low 0~8.3% 0.1% ----- Unavailable for STANDBY mode
Hal. High 0.1~8.4% 0.1% 1.5% Unavailable for STANDBY mode
Iso. Low 0~8.3% 0.1% ----- Unavailable for STANDBY mode
Iso. High 0.1~8.4% 0.1% 2.3% Unavailable for STANDBY mode
Enf. Low 0~9.8% 0.1% ----- Unavailable for STANDBY mode
Enf. High 0.1~9.9% 0.1% 3.4% Unavailable for STANDBY mode
Sev. Low 0~9.8% 0.1% ----- Unavailable for STANDBY mode
Sev. High 0.1~9.9% 0.1% 4.2% Unavailable for STANDBY mode
Des. Low 0~21.8% 0.1% ----- Unavailable for STANDBY mode
Des. High 0.1~21.9% 0.1% 12.0% Unavailable for STANDBY mode
CAUTION: All lower limits of parameters in above table may not be set to the upper
limits, nor may the upper limits be set below the lower limits.
The operator should check that the current alarm preset is appropriate prior
to use on each patient.
Do not set alarm limits to extreme values that can render the alarm
system useless.
pressure axis varies with the upper alarm limit of Paw:
Range Gain
0 to 10 cmH2O 5 cmH2O
0 to 30 cmH2O 10 cmH2O
0 to 80 cmH2O 20 cmH2O
Time axis is not only a fixed range but also same scale for Flow-t and
Paw-t waveforms.
Range: 0 to 15 seconds (gas module open).
Range: 0 to 20 seconds (gas module close).
Flow-t waveform: Flow scale: -90 to 90 L/min; gain: 45 L/min. Time scale: 0 to 15 s (gas
module open); 0 to 20 s (gas module close).
CO2-t waveform: CO2 scale: 0 ~ 76 mmHg; gain: 38 mmHg. Time scale: 0 to 15 s (gas
Response time: Not more than 15 seconds
Type of O2 sensor: Chemical fuel cell
Useful life: 12 months (normal operating)
Operational principle: O2 monitoring modules can monitor and display oxygen concentration
of the patient circuit and contain one oxygen sensor. The O2 sensor
can detect the proportionate voltage on its surface, generated with
partial pressure of O2.
The O2 sensor is a chemical fuel cell, and its metal electrode can be
oxidated when oxygen diffuses into it. The current generated from
oxidation proportion O2 partial pressure on the surface of an electrode.
The electrode will be used up gradually during the oxidation process.
The voltage of the sensor will be affected by the temperature of the
gas mixture monitored. Thermistor on the shell of the sensor will autocompensate temperature difference inside the sensor.
Signal processing and circuit analyzing can be used in the O2
monitoring modules. So the signal of O2 sensor could be transformed
to O2 concentration. The concentration is displayed on the screen and
should be comparable with the alarm limit value saved. If the
concentration exceeds the limits, the alarm should be sound.
The effect of disturbance gas (CO2, N2O, anesthetic gas) on the
concentration of the O2 sensor is less than 0.5%VOL.
10.10 IRMA Probe Specification
Model of Sensor: IRMA CO2 or IRMA AX+
Interface : RS232,9600 bps
Useful time: Replace it every two weeks.
Accessory: Airway adapter, adult/infant
Never sterilize or immerse the IRMA airway adapters in liquid.
They can be cleaned using a cloth moistened with ethanol or isopropyl
alcohol.
Principle of IRMA: It is intended to be connected to a patient breathing circuit for
monitoring of inspired/expired gases during anesthesia, recovery and
respiratory care. It may be used in the operating suite, intensive care
unit, patient room and emergency medicine settings for adult, pediatric,
and infant patients.
The IRMA probe adopts infrared measurement technology to measure
the concentration of different gases in the breathing gas mixture, which
is based on the fact that the different gas components absorb infrared
light at specific wavelengths.
Integrated with RS-232 digital interface, it supplies the universal serial
interface communication protocol and is easy to connect.
10.10.1 Intended Use
The IRMA main stream multi-gas probe is intended to be connected to other medical devices for
display of real time and derived monitoring data of CO2, N2O, and the anesthetic agents Halothane,
Enflurane, Isoflurane, Sevoflurane and Desflurane.
It is intended to be connected to a patient breathing circuit for monitoring of inspired/expired gases
during anesthesia, recovery and respiratory care. It may be used in the operating suite, intensive care
unit, patient room and emergency medicine settings for adult, pediatric and infant patients.
Note1: Gas concentration reported in units of volume percent.
Accuracy specifications-during all conditions
Gas Accuracy
CO2 ±(0.3 vol% + 4% of reading)
N2O ±(2 vol% + 5% of reading)
Agents ±(0.2 vol% + 10% of reading)
Note1: The accuracy specification is valid for all specified environment conditions, except for
interference specified in the table “Interfering gas and vapor effect” below.
Note2: The accuracy specification for IRMA AX+ is not valid if more than two agents are present in the
gas mixture.
When connecting IRMA probe to an infant patient circuit, avoid direct contact between the IRMA probe
and the patient’s body.
If the IRMA probe comes into direct contact with any parts of the infant’s body, an insulation material
shall be placed between the IRMA probe and the body.
WARNING: The IRMA probe is not intended to come into long-term skin contact.
10.10.3.3 Pre-use Check
Always verify gas readings and waveforms on the monitor before connecting the airway adapter to the
patient circuit.
Perform the tightness check of the patient circuit with the IRMA probe snapped on the IRMA airway
adapter.
10.10.4 Zeroing Procedure
WARNING: Incorrect probe Zeroing will result in false gas readings.
In order to secure high precision IRMA probe measurements, the following
zeroing recommendations should be followed.
Zeroing is performed by snapping a new IRMA airway adapter onto the IRMA probe without
connecting the airway adapter to the patient circuit. Use the host instrument to transmit a Zero
reference command to the IRMA probe.
Special care should be taken to avoid breathing near the airway adapter before or during the Zeroing
procedure. The presence of ambient air (21% O2 and 0% CO2) in the IRMA airway adapter is of crucial
importance for a successful Zeroing. If a “ZERO-REQ” alarm appears after a Zeroing procedure, the
procedure has to be repeated.
Always perform a pre-use check after Zeroing the probe.
IRMA CO2 probes:
Zeroing needs to be performed ONLY when an offset in gas values is observed, or when an
unspecified accuracy message is displayed.
Allow 10 seconds for the IRMA CO2 probes to warm up after powering on the machine before
proceeding with the Zeroing Procedure.
Allow the IRMA probe to warm up for at least 10 seconds after changing the IRMA airway adapter
before transmitting the Zero reference command.
IRMA AX+ probes:
Zeroing should be performed every time the IRMA airway adapter is replaced, or whenever an offset
in gas values or an unspecified gas accuracy messages is displayed.
Allow 1 minute for warm up of the IRMA AX+ probes after powering on the machine and after
changing the IRMA airway adapter before proceeding with the Zeroing. The green LED on the probe
will blink for approximately 5 seconds while Zeroing is in progress.