Heyer Narkomat+ User manual

N a r k o m a t
+
A n e s t h e s i a S y s t e m
Operator’s Manual
Rev. 3.1 – 05/09
OD ER N SOL UT IO NS F OR ANE ST HE TI CS A ND INH AL AT IO N
H E Y E R N a r k o m a t+
Table of Contents
Page
Table of Contents ................................................................................................................................3
Table of Figures...................................................................................................................................6
1 Details of the Manufacturer / Apparatus.......................................................................................8
2 Description and Utilization of the Apparatus................................................................................9
2.1 General.....................................................................................................................................9
2.1.1 Product improvements.................................................................................................10
2.1.2 Responsibilities of operators........................................................................................11
2.1.3 Liability of the manufacturer.........................................................................................11
2.2 General precautions ...............................................................................................................12
2.2.1 Warning notes..............................................................................................................12
2.2.2 Caution notes...............................................................................................................13
3 Functional Description................................................................................................................14
3.1 Anesthesia ventilator ..............................................................................................................14
3.1.1 Fresh gas decoupling ..................................................................................................14
3.1.2 Constant volume provided by controlled ventilation....................................................14
3.1.3 Compliance compensation ..........................................................................................15
3.1.4 Bag-in-bottle system....................................................................................................15
3.2 Fresh gas dosing ....................................................................................................................15
3.3 Vaporizer mounting device and Vaporizers............................................................................16
3.4 Patient module........................................................................................................................17
3.4.1 Circuit absorber system...............................................................................................17
3.4.2 CO2 absorber...............................................................................................................17
3.4.3 Reservoir and manual ventilation bag .........................................................................17
3.4.4 Volume measurement..................................................................................................17
3.4.5 Oxygen measurement*................................................................................................17
3.4.6 Patient module heating................................................................................................18
3.5 Gas measurement*.................................................................................................................18
4 Operating Elements / Connections for Appliances ....................................................................19
4.1 Views of apparatus .................................................................................................................19
4.1.1 Front of apparatus........................................................................................................19
4.1.2 Rear of apparatus........................................................................................................20
4.1.3 Left side of apparatus ..................................................................................................21
4.1.4 Right side of apparatus................................................................................................22
4.2 Ventilator unit..........................................................................................................................23
4.2.1 Operating panel, ventilator keyboard...........................................................................24
4.2.1.1
"Ventilation Modes"- selection switch .................................................................24
4.2.1.2
"Push to Enter" - encoder button to change and confirm settings ......................25
4.2.1.3
Operating panel keys for insp. pause and exp. pause........................................25
4.2.1.4
"Settings" - panel keys for basic set-up ..............................................................25
4.2.1.5
"Alarms" - panel keys and displays for alarms....................................................26
4.2.1.6
Vent. Settings" - panel keys for direct selection display windows ......................26
4.3 Flow meter tube block including fresh gas dosing..................................................................27
4.4 Vaporizer mounting device .....................................................................................................28
4.5 Patient module (circuit system) ..............................................................................................29
4.6 Sample gas recirculation ........................................................................................................30
4.7 Vacuum source for bronchial suction (optional) .....................................................................31
4.8 Symbols on the Unit................................................................................................................32
5 Alarm Messages and Safety Devices ........................................................................................33
5.1 Alarm messages direct after the System start........................................................................34
5.2 Alarm Messages in Standby...................................................................................................35
5.3 Compliance test alarm messages ..........................................................................................36
5.3.1 Compliance test alarm messages................................................................................36
5.4 System test alarm messages .................................................................................................38
5.5 Alarm messages during normal operation..............................................................................40
5.5.1 Alarm messages during normal operation with the integrated Gas Module * .............45
5.5.2 Alarm messages during normal operation with the optional ventilation mode PCV....47
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 3 / 150
H E Y E R N a r k o m a t+
5.6 Technical failure alarm message............................................................................................48
6 Start-Up and Functional Test.....................................................................................................48
6.1 Preparation of apparatus ........................................................................................................48
6.2 Pre-Operation Tests ...............................................................................................................49
6.3 Sample Tube Connection (optional integrated Gas Module) .................................................50
6.4 System Start ...........................................................................................................................51
6.5 Battery Control........................................................................................................................52
6.6 Sensor Test ............................................................................................................................52
6.6.1 Compliance Test..........................................................................................................54
6.7 System Test............................................................................................................................56
6.7.1 O2 sensor calibration* .................................................................................................58
6.7.2 Leak Test Fresh Gas Delivery System........................................................................62
6.8 Gas module verification*.........................................................................................................64
7 Operation in the Individual Functions.........................................................................................71
7.1 Standby mode.........................................................................................................................72
7.1.1 Standby mode menu window.......................................................................................73
7.1.2 Standby mode graphics window..................................................................................77
7.1.3 Standby mode options window....................................................................................78
7.1.4 Standby mode screen saver function..........................................................................79
7.2 Control Panel Keys in the Ventilation Modes CMV / (S)CMV / PCV * Adult / Child...............80
7.2.1 Menu-Window in the Ventilation Modes CMV / (S)CMV / PCV * Adult / Child............80
7.2.2 Graphic Window in the Ventilation Modes CMV / (S)CMV / PCV * Adult / Child ........81
7.2.3 Options Window in the Ventilation Modes CMV / (S)CMV / PCV * Adult / Child.........82
7.2.4 Automatic Anesthetic Agent and Agent Mixture Identification * ..................................83
7.2.5 Panel Key VT/V in the Ventilation Modes CMV / (S)CMV Adult / Child ......................85
7.2.6 Panel Key VT/V in the Ventilation Mode PCV Adult / Child.........................................86
7.2.7 Panel Key f in the Ventilation Modes CMV / (S)CMV / PCV Adult / Child...................87
7.2.8 Panel Key I:E in the Ventilation Modes CMV / (S)CMV / PCV Adult / Child................88
7.2.9 Panel Key PEEP in the Ventilation Modes CMV / (S)CMV / PCV Adult / Child ..........89
7.3 “Manual/spontaneous” ventilation mode.................................................................................90
7.3.1 “Manual/spontaneous” ventilation mode......................................................................90
7.3.1.1
Alarm limit setting in “manual/spontaneous” mode.............................................94
7.4 Controlled Ventilation Mode....................................................................................................98
7.4.1 CMV adult ventilation mode.........................................................................................99
7.4.1.1
Parameter Settings in CMV Adult Mode...........................................................104
7.4.1.2
Alarm limits setting in CMV adult mode............................................................105
7.4.1.3
Alarm limits setting in CMV child mode.............................................................108
7.5 Ventilation mode (S)CMV .....................................................................................................109
7.5.1 Ventilation mode (S)CMV ..........................................................................................110
7.5.1.1
Changing parameter settings in (S)CMV adult mode .......................................115
7.5.1.2
Alarm limits setting in (S)CMV adult mode .......................................................116
7.5.1.3
Alarm limits setting in (S)CMV child mode........................................................119
7.6 Ventilation mode PCV...........................................................................................................120
7.6.1 PCV adult ventilation mode .......................................................................................121
7.6.1.1
Changing graphic settings in PCV Adult mode.................................................126
7.6.1.2
Alarm limits setting in PCV adult mode.............................................................127
7.6.1.3
Alarm limits setting in PCV child mode.............................................................130
8 Dismantling and Reassembling................................................................................................131
8.1 Patient module......................................................................................................................131
8.1.1 CO2 absorber canister ...............................................................................................131
8.1.2 Bag-in-bottle system..................................................................................................131
8.1.3 Replacing the expiratory flow sensor.........................................................................133
8.1.4 Dismantling the ventilation pressure valve (APL)......................................................133
8.1.5 Replacing the diaphragm valves................................................................................135
8.1.6 Valves (emergency air, inspiration, and expiration valve).........................................137
8.2 Connecting and disconnecting of the vaporizers..................................................................137
9 Cleaning ...................................................................................................................................138
9.1 Cleaning and sterilizing the apparatus .................................................................................138
9.1.1 Cleaning the housing.................................................................................................138
9.2 Patient module......................................................................................................................138
9.2.1 Sterilizing the patient module.....................................................................................138
4 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
10 Preventive Maintenance and Servicing................................................................................139
10.1
General...........................................................................................................................139
10.1.1 Preventive Maintenance by a qualified technician.....................................................139
10.1.1.1 6 Month Service Interval ...................................................................................139
10.1.1.2 12 Month Service Interval .................................................................................140
10.1.1.3 36 Month Service Interval .................................................................................141
10.2
Servicing the Vaporizer..................................................................................................141
10.3
Other Servicing...............................................................................................................141
11 Specifications........................................................................................................................142
12 Warranty ...............................................................................................................................146
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 5 / 150
H E Y E R N a r k o m a t+
V
Table of Figures
Fig. 1 Front view of the apparatus......................................................................................................................... 19
Fig. 2 Rear of the apparatus.................................................................................................................................. 20
Fig. 3 Left side view on the apparatus................................................................................................................... 21
Fig. 4 Right side view on the apparatus................................................................................................................. 22
Fig. 5 Ventilator display and airway pressure gauge ............................................................................................. 23
Fig. 6 Operating panel, ventilator keyboard........................................................................................................... 24
Fig. 7 Flow meter block including fresh gas dosing............................................................................................... 27
Fig. 8 Vaporizer mounting device.......................................................................................................................... 28
Fig. 9 Patient module (circuit system).................................................................................................................... 29
Fig. 10 Sample gas recirculation and .................................................................................................................... 30
Fig. 11 Vacuum source.......................................................................................................................................... 31
Fig. 12 Example of an alarm message during the booting procedure.................................................................... 34
Fig. 13 Example of an alarm message during the booting procedure.................................................................... 35
Fig. 14 Example of a sensor test alarm message.................................................................................................. 36
Fig. 15 Example of a system test alarm message................................................................................................. 38
Fig. 16 Example of an alarm message during normal operation............................................................................ 40
Fig. 17 “Major Technical Failure” alarm message.................................................................................................. 48
Fig. 18 First display message during system start................................................................................................. 51
Fig. 19 Second display message during system start............................................................................................ 51
Fig. 20 Display before sensor test......................................................................................................................... 52
Fig. 21 Display after the sensor test....................................................................................................................... 53
Fig. 22 Display prior to the compliance test........................................................................................................... 54
Fig. 23 Display after the compliance test as extended information........................................................................ 55
Fig. 24 Display after the compliance test as pass/fail information ......................................................................... 55
Fig. 25 System test selection display..................................................................................................................... 56
Fig. 26 Display of system test selection sub-menu................................................................................................ 57
Fig. 27 First display message of O2 sensor calibration.......................................................................................... 58
Fig. 28 Second display message of O2 sensor calibration..................................................................................... 59
Fig. 29 Third display message of O2 sensor calibration......................................................................................... 60
Fig. 30 Fourth display message of O2 sensor calibration....................................................................................... 61
Fig. 31 Display prior to the fresh gas system leak test .......................................................................................... 62
Fig. 32 Display after the Leak Test Fresh Gas Delivery System............................................................................ 63
Fig. 33 Display of the gas module verification ....................................................................................................... 64
Fig. 34 Connection of the tube system to the verification gas bottle...................................................................... 65
Fig. 35 Display No. 1 for gas measurement verification ........................................................................................ 65
Fig. 36 Display No. 2 for gas measurement verification ........................................................................................ 66
Fig. 37 Display No. 3 for gas measurement verification ........................................................................................ 67
Fig. 38 Display No. 4 for gas measurement verification ........................................................................................ 68
Fig. 39 Display No. 5 for gas measurement verification ........................................................................................ 69
Fig. 40 Display No. 6 for gas measurement verification ........................................................................................ 70
Fig. 41 Control panel ............................................................................................................................................. 71
Fig. 42 Standby mode display ............................................................................................................................... 72
Fig. 43 Standby mode menu window..................................................................................................................... 73
Fig. 44 "Set time" within the menu window............................................................................................................ 74
Fig. 45 "Load/save personal alarm limits" within the menu window....................................................................... 75
Fig. 46 "System test options“ within the menu window.......................................................................................... 76
Fig. 47 Standby mode graphic settings window..................................................................................................... 77
Fig. 48 Stand-by mode options display.................................................................................................................. 78
Fig. 49 Screen saver ............................................................................................................................................. 79
Fig. 50 Menu-Window in the ventilation mode CMV-Adult..................................................................................... 80
Fig. 51 Graphic-Window in the ventilation mode CMV-Adult................................................................................. 81
Fig. 52 Options-Window in the ventilation mode CMV-Adult ................................................................................. 82
Fig. 53 Vt adjustment in the ventilation mode CMV Adult...................................................................................... 85
Fig. 54
Fig. 55 Rate adjustment in the ventilation mode CMV Adult.................................................................................. 87
Fig. 56 I:E adjustment in the ventilation mode CMV Adult..................................................................................... 88
Fig. 57 PEEP adjustment in the ventilation mode CMV Adult................................................................................ 89
Fig. 58 "Manual/spontaneous settings" display 1 .................................................................................................. 90
Fig. 59 "Manual/spontaneous settings" display 2 .................................................................................................. 91
Fig. 60 "Manual/spontaneous settings" display 3 .................................................................................................. 92
Fig. 61 Ventilation pressure valve (APL) in CMV/SP setting.................................................................................. 93
Fig. 62 Ventilation pressure valve (APL) in 20 cmH2O setting.............................................................................. 93
Fig. 63 Alarm limits 1 in the Manual/spontaneous ventilation mode...................................................................... 94
Fig. 64 Display “Alarm limits 2 in the Manual/spontaneous ventilation mode”....................................................... 95
Fig. 65 Alarm limits in manual/spontaneous mode................................................................................................ 96
&
adjustment in the ventilation mode PCV Adult....................................................................................... 86
6 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
Fig. 66 Graphics window in manual/spontaneous ventilation mode ...................................................................... 97
Fig. 67 Display “ventilation mode adult”................................................................................................................. 99
Fig. 68 "CMV-Settings / Adult" display................................................................................................................. 100
Fig. 69"Start in CMV adult" ventilation mode display........................................................................................... 102
Fig. 70" CMV adult" ventilation mode display...................................................................................................... 103
Fig. 71 "Parameter setting during the CMV adult" ventilation mode display........................................................ 104
Fig. 72 Display “Alarm limits 1 in the CMV Adult ventilation mode ...................................................................... 105
Fig. 73 Display “Alarm limits 2 in the CMV Adult ventilation mode” ..................................................................... 106
Fig. 74 Alarm limits in the CMV Adult ventilation mode ....................................................................................... 107
Fig. 75 Alarm limits in the CMV Child ventilation mode ....................................................................................... 108
Fig. 76 Display “ventilation mode adult”............................................................................................................... 110
Fig. 77 "(S)CMV-Settings / Adult" display............................................................................................................ 111
Fig. 78 "(S)CMV adult" ventilation mode display 1 .............................................................................................. 113
Fig. 79 "(S)CMV adult" ventilation mode display 2 .............................................................................................. 114
Fig. 80"Parameter setting during the (S)CMV adult" ventilation mode display .................................................... 115
Fig. 81 "Alarm Settings" ventilation mode (S)CMV adult ..................................................................................... 116
Fig. 82 Display “Alarm limits 2 in the (S)CMV Adult ventilation mode” ................................................................ 117
Fig. 83 Alarm limits in the (S)CMV Child ventilation mode................................................................................... 118
Fig. 84 Alarm limits in the (S)CMV Child ventilation mode................................................................................... 119
Fig. 85 Display “ventilation mode adult”............................................................................................................... 121
Fig. 86 "PCV -Settings / Adult" display ................................................................................................................ 122
Fig. 87 "Start in PCV adult" ventilation mode display .......................................................................................... 124
Fig. 88 "PCV adult" ventilation mode display....................................................................................................... 125
Fig. 89 "Parameter setting during the PCV adult" ventilation mode display......................................................... 126
Fig. 90 Display “Alarm limits 1 in the PCV adult ventilation mode”...................................................................... 127
Fig. 91 Display “Alarm limits 2 in the PCV adult ventilation mode”...................................................................... 128
Fig. 92 Alarm limits in the PCV adult ventilation mode ........................................................................................ 129
Fig. 93 Alarm limits in the PCV Child ventilation mode........................................................................................ 130
Fig. 94 Dismantling of the patient module, front view .......................................................................................... 132
Fig. 95 Dismantling of the patient module, top view ............................................................................................ 134
Fig. 96 Dismantling of the valve bodies............................................................................................................... 136
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 7 / 150
H E Y E R N a r k o m a t+
1 Details of the Manufacturer / Apparatus
Apparatus: Anesthesia system HEYER NARKOMAT
Manufacturer: HEYER Medical AG Carl-Heyer-Strasse 1-3
56130 Bad Ems Germany
Tel: ++49 2603 / 791-3 Fax: ++49 2603 / 70424
E-Mail: info@heyermedical.de
Software version: Display driver software: 3.0
Ventilator software: 2.5.n
+
Certificate-No.: 04 207-1743/98
8 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
2 Description and Utilization of the Apparatus
2.1 General
Introduction
The HEYER NARKOMAT+ anesthesia system represents a flexible anesthesia workplace for implementing and monitoring inhalation anesthesia in the half-closed system and the nearly closed system for low-flow techniques with minimum gas- and anesthetic agent utilization.
During the development of the system special emphasis was placed on the ergonomic design and consequently, a secure and easy operation. Furthermore the excellent air-tightness of the system ensures the economical daily high and low pressure utilization.
The standard model contains the following system components:
A. Electronic ventilator
The process-controlled ventilator allows constant-volume ventilation for all patient groups with a body weight of 3 kg upwards. Due to the system compliance compensation, even small tidal volumes can be precisely administered. The ventilation type CMV as the standard, the (S)CMV and PCV as optional ventilation modes, as well as various forms of ventilation and considerable variation options of the artificial ventilation cycle, facilitate a secure ventilation even for complicated lung conditions. A comprehensive test and alarm management ensures the required safety for patients and prevents out-of-control operating conditions. The clear design of the user interface and the display allow the secure operation and a quick detection of the selected ventilation parameters.
B. The patient module
The circular patient absorber system is highly integrated and compacted in an aluminum block. The block is tempered to prevent the formation of condensation. The block also contains an emergency air valve, a fresh gas reservoir in form of a hand-held anesthesia bag and an expiratory flow sensor. A motor drive connects the module to the basic apparatus. All sensors are continuously monitored during operation. The sensors are automatically calibrated during the start-up of the apparatus.
C. Integrated fresh-gas dosing system including vaporizer unit
The flow meter tube block contains all mandatory safety equipment as well as a pneumatic control system to maintain a minimum oxygen concentration of 25% in the fresh-gas flow (ratio system).
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 9 / 150
H E Y E R N a r k o m a t+
D. Integrated gas measurement CO2
The continuous CO2 measurement is graphically shown as a capnogramm. The end expiratory and inspiratory CO2 measurement will be shown as a numerical value. The CO2 values can be displayed in Vol% or mmHg.
N2O
The nitrous oxide measurement is similar to the CO2 measurement and will be shown in Vol% in the form of numerical values.
O2
The display of the oxygen measurement is based on the same principles as the measurement of CO2 and N2O and will be shown in Vol% in the form of numerical values.
Anesthesia gas
The gas module is equipped with an automatic anesthetic agent and anesthetic agent mixture identification. The apparatus measures the agents Halothane, Isoflurane, Enflurane, Desflurane and Sevoflurane. The display of the Anesthesia gas measurement is based on the same principles as the measurement of CO2, N2O and O2 .
MAC
The displayed Minimal Alveolar Concentration is calculated according to the following formula:
EtAA[%] EtN20[%] MAC (AA) = ------------ + ------------­ xAA 100
AA = Anesthetic agent Et = End expiratory concentration x = Alveolar concentration of the Anesthetic agent for MAC =1 at 100% O2 and patient age of 40 years.
2.1.1 Product improvements
HEYER MEDICAL AG retains the right to carry out modifications or to update the apparatus and/or operating instructions without prior notification. These Operating Instructions explain all features of the HEYER NARKOMAT+ anesthesia system and are correct at going to print. Instructions and models produced at a later stage may already contain improvements or modifications that were not included in previous models.
10 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
2.1.2 Responsibilities of operators
The correct functioning of the HEYER NARKOMAT+ anesthesia system can only be guaranteed if the apparatus is operated and serviced in accordance with the information provided by the manufacturer. The non-compliance with this information voids all guarantee claims against HEYER MEDICAL AG.
NOTE: Before using the apparatus please study the Operating Instructions as well as the section "General precautions", observing all information contained in these Operating Instructions that are highlighted with CAUTION or WARNING. These Operating Instructions only describe the operation of the apparatus. Information about service and repair by qualified trained personnel are contained in the HEYER NARKOMAT+ SERVICE INSTRUCTIONS.
The apparatus may only be operated by qualified and trained skilled personnel. All operators must fully observe these Operating Instructions or relevant additional documentation and information provided by the manufacturer. They must also comply with the general precautions detailed below and must be trained by authorized medical product consultants.
The apparatus may only be operated with an additional gas monitoring if a gas monitoring is not included. The following conditions must be fulfilled (DIN 13 252):
At least the:
- anesthesia-gas concentration and
- the carbon dioxide concentration must be monitored. Upper and lower alarm limits must be set for these monitoring parameters. Upon reaching
one of these upper or lower alarm limits, an optical and acoustic alarm must be triggered. The measuring adapter to be installed in the circular system or patient hose system must
contain ISO cones (DIN 13 252). These must be applied to the inspiration hose connection or ideally to the Y piece. Measuring close to the tube is, however, recommended as this allows the recording of the inspiratory and expiratory gas values. Monitors using the side-stream procedure should be clearly preferred as a supplement to the additionally required gas monitoring.
Any apparatus not functioning as described in these Operating Instructions must not be used until the fault has been removed. The operators are responsible for any damage or injuries caused by the incorrect operation or repair/servicing of the apparatus by unauthorized personnel.
2.1.3 Liability of the manufacturer
HEYER MEDICAL AG shall only be liable for the safety, reliability and functionality of the apparatus, if:
- the apparatus was operated in accordance with the information issued by the manufacturer;
- extensions, new adjustments, modifications or repairs have been carried out by an expert
authorized by the manufacturer;
- the apparatus was only operated in buildings containing grounding facilities according to the
IEC regulations.
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 11 / 150
H E Y E R N a r k o m a t+
2.2 General precautions
2.2.1 Warning notes
NOTE: A warning note points out potentially dangerous situations which may cause injuries to the patient or operating personnel.
Carry out the daily checks specified on the checklist and do not operate the system in case of a fault until the fault has been repaired.
If possible, always connect the output of the ventilation pressure valve (APL) to the anesthetic gas removal line, usually installed in the operating theatre.
The patient should furthermore be visually monitored by qualified personnel. In certain situations, life-threatening circumstances may occur which may not necessarily trigger an alarm.
Always set the alarm limits so that the alarm is triggered before hazardous situations occur. Incorrectly set alarm limits may result in operating personnel not being aware of drastic changes in the patient’s condition.
Always check that the displayed ventilation pressure value lies within an acceptable range before activating the "Set Pmin/Pmax automatically" function, automatically setting new alarm limits.
In order to prevent an electric shock, the apparatus (protection class I) may only be connected to a correctly grounded mains connection (socket outlet with grounding contact).
Explosion hazard! The apparatus may not be operated near flammable anesthetic agents or other flammable substances. No flammable anesthetic agents (i.e. ether, cyclopropane) may be used.
As the apparatus may not be used with inflammable anesthetic agents (i.e. ether, cyclopropane), no antistatic ventilation hoses or face masks are required (DIN VDE750, part
214). Electric shock and fire hazard! Always switch off the apparatus and disconnect from mains
before cleaning. Fire hazard! Fuses (i.e. additional sockets) may only be replaced by fuses of the same type
and with the same fuse value. Electric shock hazard! The apparatus may only be opened by qualified or authorized experts. The connection of apparatus via the additional socket may, in case of a failure of the
protective conductor, lead to a discharge current exceeding the permissible values. Electromagnetic radiation disturbances exceeding the values of EN 60601-1-2, can affect the
function of the machine.
12 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
2.2.2 Caution notes
NOTE: A CAUTION note refers to a situation which may cause damage or the incorrect function of the apparatus.
This apparatus may only be operated by trained, skilled medical staff. Before starting the apparatus, the operating personnel must be familiar with the notes and
information contained in these Operating Instructions and must have been trained by a medical product consultant.
If the apparatus does not function as described, the apparatus must be examined and possibly repaired by qualified service personnel, before being used again.
Handle the apparatus with care to prevent damage or functional faults. Ensure that the gas supply of the apparatus always complies with the technical specification. Before the operation, the apparatus must be correctly calibrated and/or the respective
apparatus tests, as described in the Operating Instructions, must be carried out. If the apparatus should show faults during the initial calibration or testing, the apparatus may
not be operated until the fault has been repaired by a qualified expert. After servicing, a functional test and a sensor and system test must be carried out before
clinical use. Only bacteriological filters with a low flow resistance must be connected to the patient module
and/or the patient connection. * After the unit has been switched on, a warming period of 4 minutes is necessary to ensure
an exact measurement of the anesthesia gas values. The other breathing gas values are within the ISO specifications after 1 min
* Only use the recommended pressure controller when doing the gas verification. A to high or to low calibration gas flow can lead to faulty result.
* Use only the original tube system as measuring tube. * The gas measurement tube system at the monitor must be connected before putting the
apparatus into use or changing the flow rate. Otherwise the apparatus might falsely detect a tube blockage.
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 13 / 150
H E Y E R N a r k o m a t+
3 Functional Description
3.1 Anesthesia ventilator
Ventilators are described according to the principle of controlling the change over from inspiration to expiration. The NARKOMAT+ apparatus offers the following characteristics in
the controlled ventilation mode or so-called CMV mode (Controlled Mandatory Ventilation).
time-controlled:
The timely sequence of inspiration and expiration has been specified by the ventilation frequency settings. The ratio of the inspiration to the expiration time of the individual ventilation cycle is determined by the adjustable I/E ratio.
pressure limitation:
The tidal volume during controlled ventilation is supplied during the entire period of the inspiratory flow and can be set as ventilation parameter. The inspiration is, however, terminated before the tidal volume has been administered, once the measured airway pressure reaches the set Peak pressure alarm limit.
constant volume:
The inspiratory flow to the patient required for ventilation with the set parameters, frequency f, tidal volume T. Vol. and ventilation time ratio I/E, is automatically calculated by the ventilator. This inspiratory flow is generated via the drive gas for the patient module. In standard anesthesia ventilators, deviations to the tidal volume actually supplied to the patient could occur due to the respective fresh gas setting and system compliance of the ventilation
system. The ventilator of the NARKOMAT+ apparatus supplies in each setting a constant volume. The patient module containing the bellows is decoupled from the fresh gas, and the system compliance of the patient module is automatically taken into consideration by the ventilator when generating the tidal volume.
3.1.1 Fresh gas decoupling
During the fresh gas decoupling in the CMV mode, the fresh gas flow is directed into the manual ventilation bag. The manual ventilation bag serves as a fresh gas reservoir. This principle offers the following advantages for controlled ventilation:
1. The tidal volume is completely independent from the set fresh gas flow. The ventilation is therefore also referred to as constant volume ventilation.
2. The fresh gas flow can be maintained at a very low level, i.e. below 500 ml/min, depending on the patient. The manual ventilation bag serves as a reservoir for the fresh gas administered during the inspiration. The entire fresh gas volume is available during the next inspiration, i.e. not only the fresh gas stored in the reservoir but also the fresh gas supplied during the expiration period.
3.1.2 Constant volume provided by controlled ventilation
During controlled CMV ventilation, the set tidal volume is administered irrespective of the pulmonary circumstances. The ventilator drive represents, in principle, a constant flow generator. The inspiratory flow of the ventilation gas is automatically adapted to the respective settings of the tidal volume VT, the ventilation frequency RATE and the ventilation time ratio I/E.
14 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
3.1.3 Compliance compensation
The administered tidal volume is corrected to the set value, set by the user, with the aid of the compliance compensation. The system compliance of each ventilation system, i.e. the compliance of patient hoses and the patient module, always leads to losses in the
administered tidal volume. The NARKOMAT+´s automatic compensation of the system compliance can correct this volume loss with the aid of a control loop. For this purpose several ventilation cycles are required with the drive gas flow being slightly increased above the normal values, i.e. the values for achieving the set ventilation volume. The correct tidal volume is, however, administered to the patient, whilst the slightly higher volume is absorbed by the system compliance due to the effective compliance compensation.
3.1.4 Bag-in-bottle system
The so-called bag-In-bottle system is part of the patient section or circuit system. The gas­conducting sections are divided from the ventilator into a primary (ventilator) and a secondary circuit (patient). The gas volume provided by the drive is not directly administered to the patient but instead compresses a bellows inside a pressure dome. As a result, the ventilation gas contained in the bellows is administered to the patient. An increased drive volume, flowing into the pressure dome, also increases the tidal volume. Once the drive-gas flow has finished, also the pressure compensation between the primary and secondary circuit is ended. A distinctive plateau in the ventilation pressure curve is formed, if the system does not switch over to the expiration directly after the end of the inspiratory gas flow. For this purpose, the drive volume contained in the pressure dome is maintained at a steady level for some time. The Bellow is suitable for adults and children. An exchange of the bellows for different patient groups is not necessary.
3.2 Fresh gas dosing
The set-up of the gas quantities administered to the patient is carried out on a measuring tube block. This block contains measuring tubes that are also referred to as rotameters. These measuring tubes consist of a vertical glass tube, containing a suspended body. As the glass tubes expand towards the top, a certain gas flow lifts the suspended body to a respective height. The gas flow is adjusted via valve spindles, located below the individual measuring tube. The setting of the O2/AIR or O2/N2O is selected with the change-over switch which opens the respective gas line to the measuring tube block. The fresh gas cannot be set with a mixture of AIR and N2O as, in this case, a decrease of the oxygen content to below 21% could not be avoided. The decrease of the oxygen content to below 21% is in theory also possible with O2 and N2O gas dosing. Such unfavorable settings are, however, prevented by a pneumatic safety system. This system, which is also referred to as RATIO system ensures the continuous existence of a minimum O2 content of 25% in the dosed gas mixed with N2O. In case of an increase of the N2O flow, the required O2 flow is also automatically increased. The thus set fresh gas is then fed over a vaporizer, where it is mixed with anesthetic agent.
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 15 / 150
H E Y E R N a r k o m a t+
3.3 Vaporizer mounting device and Vaporizers
The apparatus contains a Selectatec compatible vaporizer accommodating two vaporizers. The vaporizer contains two chambers, with the bottom section containing the liquid anesthetic agent. Via a woven metal wick, the top section of the chamber is enriched with saturated anesthetic agent steam. The concentration of the saturated steam at ambient temperatures is considerably higher than that acceptable for clinical purposes. The desired concentration is achieved by a suitable mixing ratio of the gas and anesthetic agent with a gas flow that is passed around this chamber. This is achieved with the aid of the setting wheel. For this purpose, the ratio of the flows of the carrier gas is adjusted by a bypass channel and the vaporizer channel in such a way, that the chosen concentration is achieved at the vaporizer outlet. In the zero position of the vaporizer, this bypass channel remains open whilst the vaporizer chamber is completely closed for the gas flow. Although the anesthetic agent steam concentration of the vaporizer chamber is saturated, the absolute anesthetic agent content depends, however, on the temperature. Consequently, the bypass channel contains a temperature compensation valve which in case of changes in the steam pressure caused by temperature fluctuations, changes the set dilution ratio in such a way that the anesthetic agent concentration issued is no longer dependant on the temperature. For further information see: Instructions of the respective anesthetic agent vaporizer.
mounting device
16 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
3.4 Patient module
3.4.1 Circuit absorber system
The circuit absorber system consists of a ventilation system with CO2 absorber. This system allows anesthetics to be carried out at extreme low fresh gas settings. The ventilation gas contains various parts of re-breathing gas, i.e. expiratory gas freed from CO2 parts. This is achieved with a circuit ventilation system, facilitating a re-breathing of the expiratory CO2­containing gas. A circuit system with high re-breathing contents causes a reduction of the consumption of anesthetic agents. This system also offers an improved breathing gas conditioning. The patient module is designed as a circuit absorber system in the form of a compact aluminum block. The hose connections normally required between the ventilator and the circuit system are thus no longer needed.
3.4.2 CO2 absorber
The absorber serves to absorb the soda lime and aims to remove the CO2 from the expiration air. The absorption process is a chemical reaction in which carbon dioxide is bound, most of the reaction water evaporates and the calcium is removed. Consequently the spent soda lime is dry and hard. The soda lime must be stored hermetically sealed, cool and dry in order not to become malabsorbant.
3.4.3 Reservoir and manual ventilation bag
The reservoir consisting of a manual ventilation bag provides an inspiratory interim store for the fresh gas. The reservoir pressure during machine and spontaneous breathing is limited to 1-2 cmH2O. During manual breathing, this valve also allows the manual adjusting of the desired ventilation pressure.
3.4.4 Volume measurement
Volume measurement is carried out by using a flow meter, operating according to the hot-wire anemometer principle, to measure the flow in the expiration branch. The ventilator processors integrate this measured value with the displayed tidal and ventilation minute volume. The tidal volume shown in the display is a measured value. The tidal volume displayed during controlled ventilation is measured by an internal flow sensor and is not dependent on the expiratory volume measurement.
3.4.5 Oxygen measurement*
The oxygen is metered by a measuring cell installed on the inspiration valve. This single­cathode measuring cell, also referred to as fuel cell, offers a longer life compared to other oxygen cells and is less sensitive to existing anesthetic gases.
*not valid with the apparatus option with integrated gas module
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 17 / 150
H E Y E R N a r k o m a t+
3.4.6 Patient module heating
The heating prevents the formation of condensation in the patient module and on the valve caps of the inspiration and expiration valve. The heating positively contributes to a ventilation gas conditioning. The heating mat also functions as a sealing mat and is installed between the top and bottom section of the patient module. An electronic control integrated in the ventilator, keeps the temperature of the patient module constant at an approx. 36°C. An over­temperature protection protects the apparatus against overheating.
3.5 Gas measurement*
The measurement of the single gases in the breathing gas is based on the principle that different gases absorb a different wavelength of infrared light. A pump inside the apparatus continuously sucks a breathing gas probe out of the respiratory circulation into the measurement chamber via a sampling line which is included with the gas module. This gas probe flows through the measurement chamber where the absorption of the different wave lengths of infrared light is measured. Based on these measurements, a microprocessor calculates the concentration of CO2, N2O and anesthesia gas.
The measurement of oxygen is done with a paramagnetic oxygen sensor. This principle of measurement is based on the paramagnetic characteristics of the oxygen molecule, through which the oxygen concentration can be measured magnetically.
* only valid with the apparatus option with integrated gas module
18 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
4 Operating Elements / Connections for Appliances
4.1 Views of apparatus
4.1.1 Front of apparatus
Fig. 1 Front view of the apparatus
1 Vaporizer mount
Selectatec* compatible vap. mount for connection of 2 vaporizers.
2 Flow meter block
6-fold flow meter block with integrated Ratio­system, O2-Bypass and N20/AIR change over switch.
3 Ventilator unit
Microprocessor controlled ventilator with TFT-display.
4 Control panel
Control panel with keys, selector switch and encoder (rotary switch).
5 Patient module
Absorber circuit system with integrated bag-in­bottle system, active and passive valves, such as APL-valve.
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 19 / 150
H E Y E R N a r k o m a t+
4.1.2 Rear of apparatus
Fig. 2 Rear of the apparatus
1 Mains cable
Voltage supply cable for connecting the apparatus to a socket with grounding contact.
2 Cable hooks 3 Sockets
Additional devices can be connected.
4 Circuit breakers
One 5A circuit breakers for the ventilator unit , one 5A circuit breaker for the battery and one 10A (or two 5A) circuit breaker for the convenient receptacles
5 Cylinder jokes
Pin-index or DGAI cylinder jokes, one for N2O, one for O2.
6 Serial number 7 Fan
A fan provides ventilation of the housing and cooling of the integral components.
8 Pipeline supply connections
Connections for O2, AIR and N2O from central gas supply. Three pipeline pressure gauges inform about supply pressure.
20 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
4.1.3 Left side of apparatus
Fig. 3 Left side view on the apparatus
1 Holder for patient arm
2 Oxygen-sensor with cable*
3 Patient module
4 Suction stand with glasses (option)
5 Water trap with gas measurement inlet**
* Only valid without the option gas module
**Only valid with the option gas module
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 21 / 150
H E Y E R N a r k o m a t+
4.1.4 Right side of apparatus
1 Machine operation checklist
The checklist informs about several tests which have to be carried out in the start-up procedure.
2 Mains switch
With this switch, the apparatus is switched on/off and the mains connection is activated. During the operation, the switch remains in the ON position.
3 Rail mount 25 x 10
Fig. 4 Right side view on the apparatus
22 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
4.2 Ventilator unit
1
2
3
Fig. 5 Ventilator display and airway pressure gauge
1 Ventilator display
The ventilator uses a color TFT display. This display is high in contrast, clearly visible also from the side view and provides all measured values and settings of the ventilator.
2 Airway pressure gauge
Shows the airway pressure on a pressure gauge within a range of
-20 to 80 cmH2O.
3 Cylinder pressure gauges
Shows the cylinder pressure for each cylinder on a pressure gauge within a range of 0 to 315 kPa x 100 (315 bar).
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 23 / 150
H E Y E R N a r k o m a t+
4.2.1 Operating panel, ventilator keyboard
Fig. 6 Operating panel, ventilator keyboard
4.2.1.1 "Ventilation Modes"- selection switch
This switch for the selection of ventilation modes has four positions: Standby: Position for starting and implementing sensor and system tests. Manual/spont.: In this position the ventilator is switched to the manual or
spontaneous ventilation mode.
Child (20-400 ml): In this position, the ventilator is switched to the mode for the
controlled ventilation of children.
Adult (300-1400 ml): In this position the ventilator is switched to the mode for the controlled
ventilation of adults.
24 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
4.2.1.2 "Push to Enter" - encoder button to change and confirm settings
With this encoder button the different parameter windows can be selected in the display. By turning the button anti-clockwise or clockwise, the respective windows or selection buttons in the display are selected. By pushing the direct selection button, functions or further windows may be selected or opened. Rotary ventilator set-up buttons shown on the display, are also operated by the direct selection button. The respective parameters are increased in clockwise direction and decreased in anti-clockwise direction.
In case of changes, the changed settings or selections are confirmed and stored once the button is pressed.
4.2.1.3 Operating panel keys for insp. pause and exp. pause
Insp. Pause: Upon pressing this key in the CMV mode, the green LED flashes and an
inspiratory pause of up to 5 sec. is inserted at the end of the next inspiration. In order to terminate the pause early, the key may be pressed a second time, after which the inspiratory pause is terminated.
Exp. Pause: Upon pressing this key in the CMV mode, the green LED flashes and an
expiratory pause of up to 30 sec. is inserted at the end of the next expiratory. In order to terminate the pause early, the key may be pressed a second time, after which the expiratory pause is terminated.
4.2.1.4 "Settings" - panel keys for basic set-up
Menu: This key opens a ventilator display window menu in all ventilation modes and
the standby position. The time and temperature of the patient module are displayed. The alarm level of the acoustic alarms can be set and the alarm limits for three personal settings may be stored or loaded. The test option system display key opens another window in which comprehensive or brief information may be selected for the result of the compliance test.
Graphics: This key opens a ventilator display graphic window in all ventilation modes
and the standby position. The time axis of the real-time ventilation pressure and expiratory flow graphics to be displayed can be set to 16, 24 or 32 seconds.
Options This key opens a ventilator display graphic window in all ventilation modes
and the standby position. Here the anesthesia gas, the sample flow and the displayed CO2 unit can be set. * In the standby mode the patient module can be unlocked and the gas measurement can be verified. *
*Only valid with option gas module
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 25 / 150
H E Y E R N a r k o m a t+
V
V
4.2.1.5 "Alarms" - panel keys and displays for alarms
This key opens a display window in the manual/spontaneous, CMV,
(S)CMV or PCV ventilation modes for setting all available alarm limits. The respective alarm limits can then be changed to the new settings. By pressing the limits key a second time, an additional alarm window is opened with additional adjustable alarm settings. By pressing the limits key a third time the window is closed and the changed alarm limits are stored.
This key switches off the sound of the acoustic alarms for a maximum
of 2 minutes. By pressing the key again, the sound is activated again before the end of the 2 minute period.
4.2.1.6 Vent. Settings" - panel keys for direct selection display windows
For the respective parameters a smaller window is opened for a new setting or the alteration of existing settings. With the aid of the encoder button, the current parameter setting can be changed with a rotary display button. The changed setting is confirmed and stored by pressing the encoder. Or the display window is closed by depressing the key again or automatically after about 5 sec.
VT/
: In the CMV ventilation mode this key opens a display window for
setting the tidal volume VT.
In the PCV ventilation mode this key opens a display window
setting the maximum inspiratory flow.
f: This key opens a Rate display window in the CMV ventilation mode for
setting the ventilation frequency Rate.
I:E: This key opens an I:E display window in the CMV ventilation mode for
setting the ventilation time ratio I:E from inspiration to expiration time.
PEEP: This key opens a PEEP display window in the CMV ventilation mode
allowing a PEEP setting.
for
26 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
4.3 Flow meter tube block including fresh gas dosing
Fig. 7 Flow meter block including fresh gas dosing
1 Flow meter tube O 2 Flow meter tube O 3 Flow meter tube AIR
with a low measuring range for settings between 0 to 1000 ml/min
2
with a high measuring range for settings between 1,5 and 10 l/min
2
with a low measuring range for settings between 0 to 1000
ml/min
4 Flow meter tube AIR
with a high measuring range for settings between 1,5 and 10
l/min
5 Flow meter tube N2O
with a low measuring range for settings between 0 to 1000
ml/min
6 Flow meter tube N2O
with a high measuring range for settings between 1,5 is 10 l/min
7 Valve spindle for O2 gas dosing 8 Valve spindle for AIR gas dosing 9 Valve spindle for N2O gas dosing 10 O2 bypass
The O2 bypass supplies a high O2 flow (approx. 50 l/min) directly to the fresh gas outlet or
into the patient module. Upon releasing the bypass key, the key returns to its original position and the O2 bypass is automatically interrupted.
11 N2O/AIR change-over switch
This switch allows the pre-selection of N2O or AIR, which can then be dosed with the
respective valve spindles. The previously set volume flow is retained after switching back to the same gas type.
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 27 / 150
H E Y E R N a r k o m a t+
4.4 Vaporizer mounting device
Fig. 8 Vaporizer mounting device
1 Fixing knob for patient arm 2 Valve cartridge of vaporizer mount 3 Locking device 4 Stopping face
28 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
4.5 Patient module (circuit system)
Fig. 9 Patient module (circuit system)
1 Emergency air valve 2 Inspiration valve 3 Oxygen sensor*
(fuel cell)
4 Ventilation pressure valve (APL valve): Including rotary regulator for setting the
pressure control during manual ventilation or for setting it as a 2 cmH2O release valve during CMV or spontaneous ventilation.
5 Expiration valve 6 Hose connection branch 7 Outlet of ventilation pressure valve
(APL-valve):
The anesthetics gas scavenging system is connected to this point.
8 Hose connection for inspiration branch 9 Hose connection for reservoir/manual ventilation bag 10 Bellows including bellows dome 11 CO
absorber canister
2
*only the apparatus option without the gas module
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 29 / 150
H E Y E R N a r k o m a t+
4.6 Sample gas recirculation
Fig. 10 Sample gas recirculation
1 Sample gas recirculation connector*
The gas return line from external monitors operating according to the side stream procedure are connected to the "Sample gas recirculation" adapter at the rear side. The sample gas is putted back to the reservoir bag by this adapter.
* only the apparatus option without the gas module
30 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
4.7 Vacuum source for bronchial suction (optional)
Fig. 11 Vacuum source
1 Tube clamp for suction tube/finger tip 2 Quick coupling AIR for vacuum source 3
Vacuum source (injector) with vacuum gauge and control knob
4 Connection for tube to suction glass "vacuum" connector
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 31 / 150
H E Y E R N a r k o m a t+
4.8 Symbols on the Unit
Attention, check the accompanying documents
I/0
On/Off (connection to the power supply)
32 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
5 Alarm Messages and Safety Devices
The NARKOMAT+ anesthesia system displays alarm messages on the ventilator display during operation. The alarm message is displayed until the fault condition that triggered the alarm is resolved. High priority alarms are displayed against a light background in the alarm window. Low priority alarms are displayed against a dark background. The switching of the alarm muting key does not influence the alarm shown on the display.
In the following, all alarm situations that may arise during the different operating conditions are explained.
There are three priority levels of the alarms:
No. Alarm LEDs Acoustic
1 Continues red Continues tone 2 Red blinking Intermitting tone 3 Yellow blinking
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 33 / 150
H E Y E R N a r k o m a t+
5.1 Alarm messages direct after the System start
Direct after the switching on and booting of the system, alarm messages are displayed in the bottom section of the screen. The alarm has to be acknowledged by pushing the encoder button before the booting procedure continues.
Direct after the switching on and booting of the system, alarm messages are displayed in the bottom section of the screen. The operating conditions can be checked and, if necessary, errors can be corrected.
Fig. 12 Example of an alarm message during the booting procedure
34 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
5.2 Alarm Messages in Standby
In standby, the alarm messages are display in the middle section of the screen.
In standby, the alarm messages are display in the middle section of the screen. The operating conditions can be checked and, if necessary, errors can be corrected.
Fig. 13 Example of an alarm message during the booting procedure
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 35 / 150
H E Y E R N a r k o m a t+
5.3 Compliance test alarm messages
During or after the compliance test, alarm messages are displayed in the middle or top section of the screen
During or after the compliance test, alarm messages are displayed in the middle or top section of the screen The operating conditions can be checked and, if necessary, errors can be corrected.
Fig. 14 Example of a sensor test alarm message
If the system was locked because of a sensor failure, only use the apparatus in case of an emergency for manual ventilation and contact a service technician if the compliance test fails repeatedly.
5.3.1 Compliance test alarm messages
Alarm message Significance/ cause Corrective action
System Resistance too high Resistance of patient hoses or
bacteria filter is too high.
Compliance test Passed. Leak rate is higher than 600 ml/min. Tighten valve rings
The leakage of the circle system and the patient hoses is higher than
600 ml/min at 40 cm H2O. Check breathing circuit Press OK to continue
Replace bacteria filter and breathing circuit. Use the Compliance test in Standby / OPTIONS to retest. If all else fails call Service.
Ventilator may be used safely with adequate fresh gas flows. If necessary use the Compliance test in Standby / OPTIONS to retest after tightening the breathing circuit and valve rings.
36 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
Compliance test alarm messages (continued)
Alarm message Significance/ cause
Compliance out of range The compliance of the connected
patient hoses lies outside the range
of 3.0 to 9.9 ml/cm H2O.
System Error Cal Required – Call Service
Data for parameter or alarm limit
settings was not saved correctly,
the data exchange between the
ventilator modules and on-screen
display is faulty or the startup test
for Internal circuit EEPROM has
failed. Check Diaphragm Valves The system has detected a
pressure increase. The diaphragm
valves may be leaking drive gas
into the circle system.
Pressure Reading out of tolerance Perform Compliance Test
The system has detected a fault on
one or both pressure sensors. when convenient Flow/Volume Readings not
available Replace Flow Sensor - Call Service
During the test, the hotwire sensor
did not pass its test. Sensor may be
faulty.
Flow/Volume Readings not available Replace Flow Sensor - Call Service
During the test, the hotwire sensor
did not pass its test. Sensor may be
faulty.
Valve Error: Use Manual Ventilation Call Service
Vent Error: Use Manual Ventilation
The proportional valve for
generating the ventilation volume
does not function correctly.
The automatic test routine has
detected a processor control fault. Call Service
Calibrate Breathing System Perform Compliance Test when convenient
One or more solenoid valves
actuating the bellows valve have
failed.
Patient module unlocked The Patient module is not adapted
correctly in the docking station.
Corrective action
Replace bacteria filter and breathing circuit. Use the Compliance test in Standby / OPTIONS to retest. If all else fails call Service.
Retry function. Reboot machine. If all else fails call Service.
Remove the Breathing circuit from the Docking Station and check the Decoupling and Expiratory valves for intact membranes. Replace as required or call Service.
Perform the Compliance test in Standby / OPTIONS. If all else fails call Service.
Retry function. Perform the Compliance test in Standby / OPTIONS Reboot machine. If all else fails call Service. Retry function. Reboot machine. If all else fails call Service.
Perform the Compliance test in Standby / OPTIONS. If all else fails call Service.
Call Service.
Remove the Breathing circuit from the Docking Station. Check the Bellows valve actuation. Perform the Compliance test in Standby / OPTIONS. If all else fails call Service.
Check if the Patient module is adapted correctly. If else fails, call service
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 37 / 150
H E Y E R N a r k o m a t+
5.4 System test alarm messages
During or after the sensor test, alarm messages are displayed in the lower section of the ventilator display:
If the system cannot create a sufficient pressure increase in the breathing circuit, the following fault message is for instance displayed in the lower section of the system test window.
Fig. 15 Example of a system test alarm message
38 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
System test alarm messages (continued)
Alarm message
System Resistance too high Resistance of patient hoses or
Leakage greater than 200 ml/min The fresh gas system leakage test
Compliance out of range The compliance of the connected
System Error Cal Required – Call Service
Calibrate Breathing System Perform Compliance Test when convenient
CPU Error: Use Manual Ventilation Call Service
O2 Cell is out of range Replace O2 Sensor Press OK to Start
O2 concentration too high. Expose Sensor to room air. Press OK to start.
Significance/ cause Corrective action
bacteria filter is too high.
has detected a leakage of over 200 ml/min.
patient hoses lies outside the range of 3.0 to 9.9 ml/cm H2O.
Data for parameter or alarm limit settings was not saved correctly, the data exchange between the ventilator modules and on-screen display is faulty or the startup test for Internal circuit EEPROM has failed.
One or more solenoid valves actuating the bellows valve have failed.
The watchdog control system has detected a failure.
O2 cell is either in a concentration of O2 that is lower than 21%, is disconnected from the cable or is faulty.
O2 cell is either in a concentration of oxygen that is higher than 21% or cell is faulty.
Replace bacteria filter and breathing circuit. Use the Compliance test in Standby / OPTIONS to retest. If all else fails call Service.
Ventilator may be used safely with adequate fresh gas flows. Retest after checking the breathing circuit, tighten valve rings, o-rings, seals on vaporizer mounts, bellows and absorber domes
Replace bacteria filter and breathing circuit. Use the Compliance test in Standby / OPTIONS to retest. If all else fails call Service.
Retry function. Reboot machine. If all else fails call Service.
Remove the Breathing circuit from the Docking Station. Check the Bellows valve actuation. Perform the Compliance test in Standby / OPTIONS. If all else fails call Service.
Call Service.
Expose cell to 21% oxygen, with its cable plugged in, or replace the cell.
Expose cell to 21% oxygen or replace the cell.
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 39 / 150
H E Y E R N a r k o m a t+
5.5 Alarm messages during normal operation
During normal operation, alarm messages are displayed in the lower section of the ventilation mode window:
Fig. 16 Example of an alarm message during normal operation
After each activation of the CMV mode, the following alarm message is, for instance, displayed because the selected minimum airway pressure has not been reached and the anesthesia gas concentration is not sufficient for identification.
Alarm messages during normal operation (continued)
40 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
Alarm message
Significance/ cause Corrective action
CPU Error: Use Manual Ventilation Call Service
The watchdog control system has
detected a failure.
Vent Error: Use Manual Ventilation Call
The automatic test routine has
detected a processor control fault. Service
High system pressure Set APL valve to OPEN
The ventilation pressure does not
decrease during expiration.
Valve Error: Use Manual Ventilation Call Service
Flow Error: Use Manual Ventilation Call Service
The proportional valve for
generating the ventilation volume
does not function correctly.
The internal flow sensor,
measuring the generated
ventilation volume, does not
functioning correctly. Calibrate Breathing System
Perform Compliance Test when convenient
One or more solenoid valves
actuating the bellows valve have
failed.
Pressure Reading out of tolerance Perform Compliance Test
The system has detected a fault
on one or both pressure sensors. when convenient No Drive Gas; Please Check The compressed gas (AIR or O2)
supply with which the drive gas of
the ventilator is generated, has
failed. Resume Ventilation Drive gas now available after the
ventilator had stopped, because
of a loss of drive gas. Flow/Volume Readings not
available Replace Flow Sensor - Call Service
During the test, the hotwire sensor
did not pass its test. Sensor may
be faulty.
Call Service.
Call Service.
Remove the Breathing circuit from the Docking Station. Check Expiratory valve actuation.
Perform the Compliance test in Standby / OPTIONS. If all else fails call Service.
Perform the Compliance test in Standby / OPTIONS. If all else fails call Service.
Remove the Breathing circuit from the Docking Station. Check the Bellows valve actuation. Perform the Compliance test in Standby / OPTIONS. If all else fails call Service.
Perform the Compliance test in Standby / OPTIONS. If all else fails call Service.
Check AIR or O2 pressure supply.
Select ventilation mode. Restart the ventilation.
Retry function. Reboot machine. If all else fails call Service.
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 41 / 150
H E Y E R N a r k o m a t+
Alarm messages during normal operation (continued)
Alarm message
A P N E A No breathing by the patient is
System vented or drive gas missing
PEEP greater than Pmin The measured PEEP exceeds the
FiO2 lower than FiO2 min The measured FiO2 value is less
Tidal Volume lower than Vt min
Peak pressure greater than alarm limit
Peak pressure below alarm limit
Minute Volume below alarm limit
PEEP greater than PEEP­Setting
FiO2 greater than FiO2 alarm limit
Check Vent Dial position The Vent dial is in an invalid
Temp Sensor out of tolerance Check Heating System – Call Service
Temp Sensor readings not available Call Service
Significance/ cause Corrective action
Check connections and patient detected in manual mode when measuring the expiratory volume.
breathing. Initial Apnea indication
(beeps) is adjustable to 15, 30, or
45 sec. A steady tone indicates
Apnea after 2 minutes. The circle system is vented during
the pressure or leak test.
Check the O2 cell is present,
APL valve is set to max (leak
test) and breathing circuit is
connected properly.
Check that APL valve setting is set Pmin pressure limit.
appropriate. Check Pmin. Alarm
setting.
Check that supplied FiO2 is than the set minimum oxygen concentration FiO2 min.
appropriate. Check O2 Alarm
setting. The measured tidal volume VTE
value is less than the set VTE min. limit.
Check that Tidal Volume setting
is appropriate. Check Vt min
Alarm setting. The peak pressure Ppeak
exceeds the set Pmax pressure limit.
Check that Tidal Volume setting
is appropriate. Lungs deflate
when Pmax is reached. Check
Pmin. Alarm setting. The peak pressure Ppeak is lower
than the set Pmin pressure limit.
Check that Tidal Volume or
Plateau Pressure setting is
appropriate. The ventilation minute volume M
vol. is lower than the set M vol. min. limit.
Check that Minute Volume
setting is appropriate. Check
M.Vol. Alarm setting. The PEEP is greater than the
PEEP setting. The measured FiO2 exceeds the
set FiO2 max. oxygen concentration limit.
Check that PEEP setting is
appropriate.
Check that supplied FiO2 is
appropriate. Check O2 Alarm
setting.
Reposition the vent mode vent position longer than 3 sec.
dial to a valid position. Reselect
ventilation mode. Restart
ventilator. The control of the breathing circuit
Call Service at next opportunity. heating, or the heating mat, is faulty.
The temperature sensor of the breathing circuit has failed.
Perform the Compliance test in
Standby / OPTIONS. If all else
fails call Service.
42 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
Alarm messages during normal operation (continued)
Alarm message
Fan Error Check Fan – Call Service
AC Power lost, using Battery The mains voltage supply has
AC Power lost, using Battery 30 min. remaining
AC Power lost, using Battery 25 min. remaining
AC Power lost, using Battery 20 min. remaining
AC Power lost, using Battery 15 min. remaining
AC Power lost, using Battery 10 min. remaining
AC Power lost, using Battery 5 min. remaining
AC Power lost, using Battery Battery running low Use Manual Ventilation
Check Battery – Call Service The battery is empty or defective Check the fuses. Call Service if
Significance/ cause Corrective action
The fan that ventilates the
Call Service at next opportunity. housing and cools the integral modules has stopped.
Check AC line cord is plugged failed, the machine is being powered by the battery.
into AC outlet. Call Service if
required. Less than 30 minutes of battery
Plug in AC Line cord. power available.
Less than 25 minutes of battery
Plug in AC Line cord. power available
Less than 20 minutes of battery
Plug in AC Line cord. power available
Less than 15 minutes of battery
Plug in AC Line cord. power available
Less than 10 minutes of battery
Plug in AC Line cord. power available
Less than 5 minutes of battery power available
Mechanical ventilation will
discontinue in approximately 5
minutes. Display and numeric
values will continue until battery
power is exhausted. Patient can
be manually ventilated as long as
there is compressed gas
available. The power supplied by the battery
is too low for automatic mechanical ventilation. The
Check AC line cord is plugged
into AC outlet. Call Service if
required. display will fade within minutes.
required.
Continuous Pressure A continuous ventilation pressure
without significant change of the
Check that APL valve is set to
the CMV/SP position. pressure value is detected.
Ambient Air Intake; Check Fresh Gas setting
The emergency air valve in the patient section has opened; check the fresh-gas setting and increase the fresh gas flow, if necessary.
Check if the fresh gas flow is is
sufficient. Check for Leaks.
Check if patient is breathing
Spontaneously. call Service at
next opportunity.
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 43 / 150
H E Y E R N a r k o m a t+
Alarm messages during normal operation (continued)
Alarm message
Significance/ cause Corrective action
Set APL Valve to CMV/SP position
Message to remind operators to set the ventilation pressure valve (APL) to the CMV/SP position after activating the CMV mode.
Set the airway pressure limiting
valve (APL) to the CMV/SP
position after activating either
mechanical ventilation mode. This message also appears if there is a peep that is higher then the chosen peep.
Rate lower than f min
The measured ventilation frequency Rate value lies below the set f min. limit.
Only in the manual mode.
Increase the ventilation
frequency
Rate greater than f max
The measured ventilation frequency Rate value exceeds the set f max. limit.
Only in the manual mode
Decrease the ventilation
frequency
Check Settings The set parameters for CMV
mode ventilation are for an
Check Tidal Volume setting is
appropriate. inspiratory flow that cannot be achieved by the machine. The max. limit is a minute volume of < 20 l/min, or an inspiratory flow < 75 l/min.
Compliance Test bypassed The Startup Compliance test has
been bypassed
Perform the Compliance test in
Standby / OPTIONS as soon as
time allows to ensure the safe
function of the system
44 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
Insp. Agent
Insp. Agent
5.5.1 Alarm messages during normal operation with the integrated Gas Module *
Alarm message
Insp. Agent H I G H
FiO2 L O W ! The measured FiO2 value is less
FiO2 H I G H ! The measured FiO2 exceeds the
CO2 Re-breathing; please check Absorber
Insp. N2O H I G H
ET CO2 H I G H
ET CO2 L O W
Insp. Agent L O W
Insp. N2O L O W
Gas Monitor Technical Failure Sampling Line Occlusion
No Patient detected; no Breathing
Water trap is full, will be full soon or connection faulty
Significance/ cause Corrective action
The measured Anesthesia agent concentration is higher than the set maximum value for the end
Check that supplied Insp. Agent is appropriate. Check Alarm setting.
tidal concentration
Check that supplied FiO2 is than the set minimum oxygen concentration FiO2 min.
appropriate. Check O2 Alarm
setting.
Check that supplied FiO2 is set FiO2 max. oxygen concentration limit.
appropriate. Check O2 Alarm
setting. The inspiratory CO2 value is too
high The measured N2O concentration
is higher than the set value N2O max
Exchange the soda lime. If all
else fails call Service.
Check that supplied N2O is
appropriate. Check N2O Alarm
setting.
The measured CO2 concentration is higher than the set value CO2 max
Check the soda lime, the
conditions of the patient and the
alarm settings If all else fails call
Service.
The measured CO2 concentration is lower than the set value CO2 min
Check the conditions of the
patient and the alarm settings If
all else fails call Service.
The measured Anesthesia agent concentration is lower than the set minimum value for the inspiratory
Check that supplied Insp. Agent
is appropriate. Check
Alarm setting. concentration
The measured N2O concentration is lower than the set value N2O min
Check that supplied N2O is
appropriate. Check N2O Alarm
setting.
Internal fault in the gas module Call Service Due to an improper tube
connection, the exact value can not be measured
Check if the sample tube is
kinked or obstructed. If else fails
exchange the sample tube
The CO2 measurement cannot detect any breathing or respiration activities
Check if the sample tube is
connected and check the
conditions of the patient
The water trap is full, nearly full or the water trap is not connected correctly
Check if the water trap is
connected correctly or, if full,
exchange it
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 45 / 150
H E Y E R N a r k o m a t+
Alarm messages during normal operation with the integrated Gas Module* (continued)
Alarm message
Gas Monitor NOT ready
Anesthetic Agent Mixture
Anesthetic Agent not identified or in Startup Mode
Anesthetic Agent Measurement in Startup Mode
*Only valid with option gas module
Significance/ cause Corrective action
Power supply or data
Call Service communication to the gas module is defective
The presence of a second anesthetic agent has been detected
The concentration of the anesthetic agent is too low for an auto identification or the sensor is in the startup phase and cannot
Wait up to 4 min for the sensor to
warm up. If all else fails call
Service. deliver the measurements within
the ISO specifications
The sensor is in the startup phase and cannot deliver the measurements within the ISO specifications
Wait up to 4 min for the sensor to
warm up. If all else fails call
Service.
46 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
5.5.2 Alarm messages during normal operation with the optional ventilation mode PCV
Alarm message
Unable to attain target pressure; Adjust f, flow or I:E Ratio
PCV Setting not valid
Expiratory time too short
Significance/ cause Corrective action
The inspiration time and/or the flow is too low for the selected plateau pressure to be attained
Check that I:E ratio and breathing frequency setting are appropriate.
The combination of inspiratory
Change the PCV settings flow, frequency and I:E ratio cannot be realized by the ventilator.
The set expiration time does not allow the below to refill completely
Check that I:E ratio and
breathing frequency setting are
appropriate.
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 47 / 150
H E Y E R N a r k o m a t+
5.6 Technical failure alarm message
Alarm message
Major Technical Failure Switch to Manual Ventilation! Call Service - Technician
This alarm is not shown in the usual alarm windows of the various displays. Due to its extremely high priority, this fault is displayed in its own window.
Significance/ cause Corrective action
The ventilator has a serious technical fault and stops the controlled ventilation in the CMV mode.
Call Service
Fig. 17 “Major Technical Failure” alarm message
6 Start-Up and Functional Test
6.1 Preparation of apparatus
1. Install the apparatus in such as way that the touch screen operating panel at the front of the apparatus is within easy reach and the flow measuring tubes can be easily read.
2. Once the apparatus has been installed in the correct location activate the breaks on the front rollers to prevent accidental movement.
3. Connect the gas connection lines to the gas connection at the rear of the unit. Check that the pressure of the gas supply complies with the specifications of the apparatus. Connect the gas supply by plugging in the connectors in the gas supply sockets. Ensure that the connectors are not in the park position.
4. Plug the mains cable into a grounded socket. Switch on the power supply using the mains switch at the right side of the apparatus.
48 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
5. Wait until the ventilator display requests the compliance test. If the breathing circuit should not be connected to the apparatus remove the transportation cover, if necessary, and slide the breathing circuit into position. After sliding in the breathing circuit nearly up to the stop, you will need to turn the lever below the docking station clockwise to adapt the breathing circuit.
6. Next equip the breathing circuit with the bag-in-bottle system, consisting of a bellows and the bellows dome. Install the CO2 absorber canister. Connect the set of hoses for the anesthesia gas scavenging to the outlet of the ventilation pressure valve (APL­valve). Connect the hose set of anesthesia gas scavenging to the gas scavenger installed in the operation theatre.
7. Connect the manual ventilation bag and the respective hose to the breathing circuit. Connect the set of patient hoses to the insp. and exp. connections.
6.2 Pre-Operation Tests
The tests and examinations described below must be carried out before each system start. They are also specified on the check list kept at the apparatus.
1. Install and lock the vaporizers to be used on the device.
2. Start the compliance test of the device. Follow the instructions on the ventilator display.
3. Carry out the leak test. The O2 calibration should by carried at least every 30 running hours Respective instructions are shown on the ventilator display.
4. Ensure that a suitable independent device for manual ventilation (i.e. ambulatory bag) is available at or near the apparatus.
5. If required start the additional monitoring for CO2 and anesthesia gases ( Not necessary with the optional integrated gas module) and if available ECG and check function according to the respective operating instructions. The gas return line from monitors operating according to the side stream procedure should be connected to the connection at the rear side labeled as "Sample gas recirculation".
6. Ensure via a manual test that the APL valve does not allow a pressure greater than 100 cmH2O
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 49 / 150
H E Y E R N a r k o m a t+
6.3 Sample Tube Connection (optional integrated Gas Module)
WARNING
The gas measurement tube system at the monitor must be connected before putting the apparatus into use or changing the flow rate. Otherwise the apparatus might falsely detect a tube blockage.
WARNING
Use only the original tube system as measuring tube. If a different tube system is used, faulty measurements can occur due to condense water and anesthesia gases. Condensed water can also damage the internal sensors.
WARNING
Ensure the sample tube is going direct upward from the y-piece to avoid condense water going into the system.
50 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
6.4 System Start
After switching on the apparatus the message “Please wait...” appears for approx. 20 seconds while the software is loading. During the booting process the apparatus shows two displays for the system start and the automatic apparatus test. The system then asks the operator to carry out the sensor test
Fig. 18 First display message during system start
Fig. 19 Second display message during system start
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 51 / 150
H E Y E R N a r k o m a t+
6.5 Battery Control
After the apparatus has been switched on, the charging status of the battery will be checked. Depending on the charging status of the battery, the apparatus may start charging the battery which can take up to 7 hours.
6.6 Sensor Test
After the automatic system start has been completed, the following display for implementing the sensor test appears.
Fig. 20 Display before sensor test
In the case of an emergency, the sensor test can be bypassed via turning the selector switch. Otherwise follow the instructions of the ventilator display and activate by pressing the “START” button.
The system now checks and calibrates all sensors, i.e. the expiratory flow sensor and the pressure sensors. Also the internal sensors of the ventilator drive and the active components such as pneumatics control valves are checked during the compliance test.
The breathing circuit, patient connection hoses, the Y-piece etc. are also tested for leaks. Furthermore the system compliance to which in particular also the patient hoses contribute is detected during the compliance test.
52 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
Compliance test (continued)
During the compliance test a display message confirms that the test is running. After the successful sensor test the text message "Sensor Test ok" with information about
leakage rate and system compliance appears.
Fig. 21 Display after the sensor test
After the successful completion of the sensor test, a similar test, the compliance test can be carried out in the standby mode. Apart from detecting the system compliance this compliance test also checks for leaks in the patient module, patient connecting hoses, Y-piece, etc. The compliance test should be carried out, if the patient tubes were exchanged against others, with a relevant difference in the tube compliance.
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 53 / 150
H E Y E R N a r k o m a t+
6.6.1 Compliance Test
Fig. 22 Display prior to the compliance test
Follow the instructions of the ventilator display and activate by pressing the “START” button.
54 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
Compliance test (continued)
Fig. 23 Display after the compliance test as extended information
As extended information to the test result of a compliance test, the leakage of the patient module at 40 cmH2O and the detected system compliance is displayed.
Fig. 24 Display after the compliance test as pass/fail information
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 55 / 150
H E Y E R N a r k o m a t+
Compliance test (continued)
The menu window allows the operator to select whether the result should be displayed as a brief pass/fail message or if extended information should be displayed.
By activating the “menu” key, the menu window is opened. Under the sub-menu “System test options” and “Sensor test options” “Pass/Fail Info” or “Extended Info” can be selected.
6.7 System Test
Once the successful sensor test is confirmed with the “Quit” display key, the system test selection screen is displayed.
Fig. 25 System test selection display
Only in emergencies, where the system is required immediately, should the system test be skipped. The system test involves a calibration of the oxygen sensor and a fresh-gas system leak test and should be carried out during a correct system start.
* Only valid with option gas module
56 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
System test (continued
)*
Fig. 26 Display of system test selection sub-menu
Both tests should be carried out during the start of the system; select the respective system test sub-menu by turning the encoder button. Confirm the selection by pressing the encoder on the “Run” display key.
* The oxygen calibration is not valid with the apparatus option with integrated gas module
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 57 / 150
H E Y E R N a r k o m a t+
6.7.1 O2 sensor calibration*
When starting the O2 sensor calibration the following display appears.
Fig. 27 First display message of O2 sensor calibration
Pull the O2 sensor out of the inspiration valve cap. For calibration, place the sensor into the calibration adapter in ambient air left near the port of the O2 sensor cable.
* The oxygen calibration is not valid with the apparatus option with integrated gas module
58 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
O2 sensor calibration (continued)*
Fig. 28 Second display message of O2 sensor calibration
In order to allow sufficient time for aerating the O2 sensor in ambient air, a 15 second countdown is displayed in a calibration start window.
* The oxygen calibration is not valid with the apparatus option with integrated gas module
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 59 / 150
H E Y E R N a r k o m a t+
O2 sensor calibration (continued)*
During the calibration of the O2 sensors a window providing information about the progress of the calibration process appears.
Fig. 29 Third display message of O2 sensor calibration
* The oxygen calibration is not valid with the apparatus option with integrated gas module
60 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
O2 sensor calibration (continued)*
The result of the O2 sensor calibration is then displayed as follows:
Fig. 30 Fourth display message of O2 sensor calibration
Remove the O2 sensor from the calibration adapter and reinsert it in the inspiration valve cap. This completes the calibration of the O2 sensor.
* The oxygen calibration is not valid with the apparatus option with integrated gas module
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 61 / 150
H E Y E R N a r k o m a t+
6.7.2 Leak Test Fresh Gas Delivery System
To carry out a leak test for the entire fresh gas system, select menu item “Leak Test Fresh Gas Delivery System” on the display by turning and pressing the encoder button.
Fig. 31 Display prior to the fresh gas system leak test
Follow the instructions on the ventilator display. Connect the patient hoses and the hose of the reservoir/manual ventilation bag to the Y piece. Set the ventilation pressure valve (APL) to max. and start the leak test by activating the display key “Push to enter” by pushing the encoder button.
Apart from the patient module, patient connection hoses, the Y piece, etc., the leak test of the fresh gas system checks also the flow meter tube block, the vaporizer mount and the vaporizers if activated , absorber canister, manual ventilation hose and ventilation pressure valve (APL) for any leaks.
62 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
Leak Test Fresh Gas Delivery System (continued)
During the fresh gas system leak test, a message confirming that the test is carried out, is displayed. After a successful fresh gas system leak test, the following window is displayed.
Fig. 32 Display after the Leak Test Fresh Gas Delivery System
Return the hoses of the patient module to their normal condition and reconnect the reservoir/manual ventilation bag to its hose. Return the ventilation pressure valve to the CMV/SP position. The fresh gas system leak test has now been completed.
Leave the system test window by selecting the display key “Quit” by pressing the encoder button.
The ventilator is now operational after the completion of the sensor and system test and observation of the checklist test points. The Standby window of the normal operating mode is displayed.
In the following section 7 “Operation in the individual functions” this window is described under section 7.1 “Standby mode”.
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 63 / 150
H E Y E R N a r k o m a t+
6.8 Gas module verification*
The verification mode can be called up by pressing the button “options” on the display. For Verification, the selection switch must be in the “standby” position. Following display will appear:
Fig. 33 Display of the gas module verification
A calibration of the gas module is not necessary because it is maintenance free. Should a value of the gas measurement be outside of the tolerance, call an authorized service technician.
*only for apparatus option with integrated gas module
64 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
Gas module verification (continued)*
The verification procedure is described step by step on the screen.
WARNING:
Use only the recommended verification gas and pressure regulator. If using a different pressure regulator, reinsure that the sample gas flow rate is about 300 ml/min. A to high or to low flow rate can lead to faulty readings.
Fig. 34 Connection of the tube system to the verification gas bottle
Fig. 35 Display No. 1 for gas measurement verification
*only for apparatus option with integrated gas module
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 65 / 150
H E Y E R N a r k o m a t+
Gas module verification (continued)*
Fig. 36 Display No. 2 for gas measurement verification
1. Connect the gas measurement tube to the verification gas bottle
2. Open the Valve of the verification gas bottle
3. Confirm “start” through turning and pressing the encoder button.
*only for apparatus option with integrated gas module
66 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
Gas module verification (continued)*
Fig. 37 Display No. 3 for gas measurement verification
To ensure the correct display of the values, a zeroing of the gas module is performed first. The maximum time for a zeroing is 1 min. Do not press the „Quit“ button during this time.
*only for apparatus option with integrated gas module
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 67 / 150
H E Y E R N a r k o m a t+
Gas module verification (continued)*
After the zeroing is completed, open the gas cylinder valve.
Fig. 38 Display No. 4 for gas measurement verification
*only for apparatus option with integrated gas module
68 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
Gas module verification (continued)*
If continues values are being displayed, activate the button quit and close the valve of the verification gas bottle.
Fig. 39 Display No. 5 for gas measurement verification
*only for apparatus option with integrated gas module
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 69 / 150
H E Y E R N a r k o m a t+
Gas module verification (continued)*
Compare the display results with the values printed on the verification bottle.
Fig. 40 Display No. 6 for gas measurement verification
Activate the button “finalize verification“ to return to the options menu.
*only for apparatus option with integrated gas module
70 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
7 Operation in the Individual Functions
The preparation and operation of the system up to the standby mode is described in section 6 “Start-up and functional test” and its sub-sections. The following sections describe how to operate the system for the various ventilation modes in the standby mode and the operating modes.
Fig. 41 Control panel
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 71 / 150
H E Y E R N a r k o m a t+
7.1 Standby mode
This is the normal operating mode prior to switching on a ventilation mode or in between ventilation operations. The system returns to this mode after the sensor test and the fresh gas system leak test have been carried out. The display shows the following window:
Fig. 42 Standby mode display
The ventilator display shows the current time, the total number of system operating hours and the status of the system tests.
72 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
7.1.1 Standby mode menu window
The standby mode menu window is opened by pressing the "Menu" key on the control panel. The screen displays the following:
Fig. 43 Standby mode menu window
The menu window shows the measured temperature of the heated patient module. The system time is shown in hours, minutes and seconds. Using the “Sound level” display regulator the volume of the acoustic alarms can be preset
from 1 to 10. This setting will remain stored until the next system start. In the Manual/spontaneous mode the alarm acoustic is automatically set to “0”. When leaving the Manual/spontaneous mode, the alarm acoustic is automatically set to the previous set value.
The display keys "Clock Set", "Load/save" and "System Test Options" open further sub­menus determining further basic settings.
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 73 / 150
H E Y E R N a r k o m a t+
Standby mode menu window (continued)
By activating the display key "Clock Set", the window "Set time" is opened.
Fig. 44 "Set time" within the menu window
The options of this window can be selected by turning the encoder button and can be activated for setting different values by pressing the encoder button.
Turning the encoder anti-clockwise reduces, turning it clockwise increases the value. To quit the respective option, press the encoder button again. The new date or time settings are confirmed with “OK” and will then be stored in the system. To quit the "Set date and time" window, the display key "Cancel" is pressed.
74 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
Standby mode menu window (continued)
By activating the "Load/save" display key, the window "Load and Save private Alarm Limits” is opened. Here you can save the save the alarm limits which were set in the ventilation mode, or recall alarm limits which were previously saved.
Fig. 45 "Load/save personal alarm limits" within the menu window
Using following scheme, you can load or save your personal alarm limits:
1. Choose one of the four available memory slots through turning and pressing the encoder button in the panel “Memory Slot”. The saved alarm limits will be shown in the menu.
2. If you wish to save the currently set alarm limits, confirm it through turning and pressing the encoder button in the panel “Memory Slot”. Please notice that you will overwrite settings which have saved in this slot previously
3. If you want to use previously saved alarm settings, confirm this slots through turning and pressing the encoder button in the panel “Load”.
4. You can leave the menu "Load and Save private Alarm Limits” slots through turning and pressing the encoder button in the panel “Quit”.
Note: Due to safety reasons, the system will activate the default settings after every new start
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 75 / 150
H E Y E R N a r k o m a t+
Standby mode menu window (continued)
By confirming the “System test options" key, a window bearing the same name is opened.
Fig. 46 "System test options“ within the menu window
Through selecting and confirming with the encoder button:
1. you can switch the “sensor test” between the options “Extended Info” and “short Info”.
2. the window “System Tests” can be called up.
3. the “System Compliance” can be checked. This test, which determines the system compliance, should be done when e.g. the ventilation tubes are changed between two patients and their compliance have a significant difference.
This window can be closed and the settings saved through the button “Quit”.
76 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
7.1.2 Standby mode graphics window
The standby mode graphic window is opened by pressing the "Graphics" key on the operating panel. The following window is displayed.
Fig. 47 Standby mode graphic settings window
The graphic window allows the setting of the time axis for the display of the ventilation pressure and expiratory flow or CO2 capnogram real-time curves.
16, 24 or 32 seconds can be set as values. 16 seconds is the default setting. The desired time axis representation for the real time curves can be chosen by selecting the
respective screen key. By pressing the "Graphics" key for the second time, the window is closed again and the
selected setting is stored.
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 77 / 150
H E Y E R N a r k o m a t+
7.1.3 Standby mode options window
The "Options" window in the standby mode is opened by pressing the “Options” key on the operating panel. The following window is displayed:
Fig. 48 Stand-by mode options display
The patient module can be unlocked by pressing the "unlock" screen key and is then unlocked from the apparatus by the integral motor drive. If the patient module has been removed, the display key shows "unlocked". The patient module is also reconnected by the motor drive. The motor starts automatically, once the patient section is sufficiently pushed into the apparatus. Upon pressing the "Options" key again, the window is closed.
In the field “Gas Measurement“ the date of the last verification is displayed. Activating the button “verify“ opens the menu fort the Gas verification.
78 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
7.1.4 Standby mode screen saver function
If the unit is switched to standby mode for longer than 10 minutes without any operation of a key or selection switch, the display is switched dark. The text line "Press ENTER for Standby screen" is displayed on different positions of the display. To deactivate this screen saver function just press or turn the encoder button.
Fig. 49 Screen saver
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 79 / 150
H E Y E R N a r k o m a t+
7.2 Control Panel Keys in the Ventilation Modes CMV / (S)CMV / PCV * Adult / Child
Described is only the ventilation mode CMV-Adult. Differences in the ventilation CMV-Child will be explained specifically.
(The screen titles change according to the ventilation modes CMV / (S)CMV / PCV * Adult / Child)
7.2.1 Menu-Window in the Ventilation Modes CMV / (S)CMV / PCV *
Adult / Child
Fig. 50 Menu-Window in the ventilation mode CMV-Adult
The menu window shows the measured temperature of the heated patient module. The system time is shown in hours, minutes and seconds. Using the “Sound level” display regulator the volume of the acoustic alarms can be preset
from 1 to 10. This setting will remain stored until the next system start. The display keys "Clock Set", "Load/save" and "System Test Options" open further sub-
menus determining further basic settings.
* The ventilation modes (S)CMV / PCV are optional.
80 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
7.2.2 Graphic Window in the Ventilation Modes CMV / (S)CMV / PCV *
Adult / Child
Fig. 51 Graphic-Window in the ventilation mode CMV-Adult
The graphic window allows the setting of the time axis for the display of the ventilation pressure and expiratory flow or CO2 capnogram real-time curves.
16, 24 or 32 seconds can be set as values. 16 seconds is the default setting. Expiratory flow or CO2 capnogram real-time curves can be chosen for the lower graphic By pressing the "Graphics" key for the second time, the window is closed again and the
selected setting is stored. * The ventilation modes (S)CMV / PCV are optional.
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 81 / 150
H E Y E R N a r k o m a t+
7.2.3 Options Window in the Ventilation Modes CMV / (S)CMV / PCV *
Adult / Child
Fig. 52 Options-Window in the ventilation mode CMV-Adult
Through selecting and confirming the encoder button the used anesthesia gas can be defined and a sample flow can be set.
The display of the CO2 unit can be changed to mmHg or %.
* The ventilation modes (S)CMV / PCV are optional.
82 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
7.2.4 Automatic Anesthetic Agent and Agent Mixture Identification *
The gas module incorporates automatic anesthetic agent and agent mixture identification. The data of the agent measurement is displayed as follows:
Anesthetic Agent Identification
Automatic
Automatic
Automatic Agent identified
Automatic
Specific agent elected
Specific agent elected
** In Startup mode
*** The accuracy of all gas measurements corresponds to the requests of the ISO 11196 (Accuracy of the indicated reading for Anesthetic agents) and ISO 9918 (Capnometers for use with humans, requirements) the latest 4 min after switching on the unit.
Accuracy (ISO): CO2 : +/- 0,5 % O2 : +/- 3% Agent : +/-0,15% N2O : +/- 2%
**** The minimum concentration for the identification of anesthetic agents is approx. 0,3 Vol%.
* only for apparatus option with integrated gas module
Status of the measurement
Startup Mode **or insufficient accuracy ***
concentration of the anesthetic agent is too low for an auto identification ****
The concentration of an 2nd agent has become so high that the concentration of the 1st agent cannot be measured accurately
Startup Mode **or insufficient accuracy ***
Specific agent elected
Name of the Agent
Auto
Auto
Hal./Enfl./Isofl./ Sev./Des.
Hal./Enfl./Isofl./ Sev./Des.
Hal./Enfl./Isofl./ Sev./Des.
Concentration Color Error Message
--.-
Concentration value
Concentration value
White
White
Color coded
Concentration value of the 1st anesthetic agent
--.-
Concentration value
Color coded
Color coded
Anesthetic Agent not identified or in Startup Mode Anesthetic Agent not identified or in Startup Mode
Anesthetic Agent Mixture
Anesthetic Agent Measurement in Startup Mode
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 83 / 150
H E Y E R N a r k o m a t+
Automatic Anesthetic Agent and Agent Mixture Identification (continued)
Anesthetic Agent and MAC values are color coded as follows:
Gas Abbreviation Color Coding
Halothane Hal. Red Enflurane Enfl. Orange Isoflurane Isofl. Purple Sevoflurane Sev. Yellow Desflurane Des. Blue No gas identified Auto White
After the unit has been switched on, the gas module is in the startup mode for approx. 4 min. During this period following rules apply:
Cold Start:
Time Relative
Accuracy
Event
Time
(Min) (Min) Agent 1/2 N2O CO2
00:00 00:12 - - - Communication
not possible
00:12 00:12 Not ready Not ready Not ready 00:20 00:20 Start phase Start phase Start phase 00:40 00:20 Start phase Unknown Reduced
accuracy
Start phase Start phase Start phase 02:50 02:20 Zero in Progress Zero in Progress Zero in Progress
03:00 00:10 Start phase Reduced
accuracy
04:00 01:00 Start phase Reduced
accuracy
Reduced accuracy Reduced accuracy
Zeroing required
04:00 00:00 Zeroing in
progress
04:30 00:30 ISO accuracy ISO accuracy ISO accuracy
Zeroing completed
124:00 120:00 ISO accuracy ISO accuracy ISO accuracy Zeroing required
84 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
V
7.2.5 Panel Key VT/
Child
in the Ventilation Modes CMV / (S)CMV Adult /
Fig. 53 Vt adjustment in the ventilation mode CMV Adult
The Menu window shows the screen adjustment Vt for the setting of the tidal volume in ml during the controlled ventilation.
This adjustment is stored until the unit is switched off. The window closes after pressing the encoder button or automatically after 10 s.
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 85 / 150
H E Y E R N a r k o m a t+
V
V
V
7.2.6 Panel Key VT/
in the Ventilation Mode PCV Adult / Child
Fig. 54
&
adjustment in the ventilation mode PCV Adult
The Menu window shows the screen adjustment
&
for the setting of the maximum inspiratory
flow in l/min during the controlled ventilation. This adjustment is stored until the unit is switched off. The window closes after pressing the encoder button or automatically after 10 s.
86 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
7.2.7 Panel Key f in the Ventilation Modes CMV / (S)CMV / PCV Adult /
Child
Fig. 55 Rate adjustment in the ventilation mode CMV Adult
The Menu window shows the screen adjustment f for the setting of the ventilation rate in 1/min.
This adjustment is stored until the unit is switched off. The window closes after pressing the encoder button or automatically after 10 s.
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 87 / 150
H E Y E R N a r k o m a t+
7.2.8 Panel Key I:E in the Ventilation Modes CMV / (S)CMV / PCV Adult /
Child
Fig. 56 I:E adjustment in the ventilation mode CMV Adult
The Menu window shows the screen adjustment I:E for the setting of the inspiration­exspiration time ratio
This adjustment is stored until the unit is switched off. The window closes after pressing the encoder button or automatically after 10 s.
88 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
7.2.9 Panel Key PEEP in the Ventilation Modes CMV / (S)CMV / PCV
Adult / Child
Fig. 57 PEEP adjustment in the ventilation mode CMV Adult
The Menu window shows the screen adjustment PEEP for the setting of the PEEP in cmH2O. This adjustment is stored until the unit is switched off. The window closes after pressing the encoder button or automatically after 10 s.
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 89 / 150
H E Y E R N a r k o m a t+
7.3 “Manual/spontaneous” ventilation mode
For a manual ventilation or the application of the system for spontaneous ventilation, set the selection switch of the operating panel to "Manual/spontaneous".
7.3.1 “Manual/spontaneous” ventilation mode
In the ventilation mode “Manual/spontaneous” the display informs you about the measured values for pressure, flow or volume, ventilation frequency and gas concentration. The displayed alarms will appear below the real time curves.
Fig. 58 "Manual/spontaneous settings" display 1
In the panels for the measurement values the following will be displayed: Mean Average pressure in cmH2O
PEEP Positive end expiratory pressure in
cmH2O
T. Vol. Tidal volume in ml
M. Vol. Respiratory minute volume in l (1/min)
Peak Airway peak pressure in cmH2O
Rate Ventilation or spontaneous breath rate
in 1/min
MAC minimal alveolar Concentration
90 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
“Manual/spontaneous” ventilation mode (Continued)
Fig. 59 "Manual/spontaneous settings" display 2
In the panels for the gas measurement values the following will be displayed:
CO2 Et End tidal CO2 concentration in
mmHg or %
O2 In Inspiratory O2 concentration in %
Iso. In Inspiratory Anesthetic agent
concentration in %. Here for example Isoflurane.
N2O In Inspiratory N2O concentration in
%
CO2 In Inspiratory CO2 concentration in
mmHg or %
O2 Et Expiratory O2 concentration in %
Iso. Et Exspiratory Anesthetic agent
concentration in %. Here for example Isoflurane.
N2O Et Expiratory N2O concentration in
%
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 91 / 150
H E Y E R N a r k o m a t+
“Manual/spontaneous” ventilation mode (Continued)
Fig. 60 "Manual/spontaneous settings" display 3
In the panels for the alarm configuration the following will be displayed:
The displayed symbol
P: ON The airway pressure shows that the patient alarms are suppressed acoustically
CO2: ON The CO2 alarms CO2
Ax: Off The anesthetic gas alarms
Vol: ON The Volume alarms T.Vol
min and M.Vol are switch on
N2O: Off The N2O alarms N2O
Rate: Off The frequency alarms f
min and f max are switched off
alarms Pmin and Pmax are switched on
high, CO2 low and RA CO2 are switched on
e.g. ISO high and ISO low are switched off
high, N2O low are switched off
92 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
Fig. 61 Ventilation pressure valve (APL) in CMV/SP setting
Set the ventilation pressure valve (APL) to the CMV/SP setting for spontaneous ventilation.
Fig. 62 Ventilation pressure valve (APL) in 20 cmH2O setting
For a manual ventilation, the ventilation pressure valve (APL) must be set to a respective pressure limiting value, for example 20 cmH2O. The patient will be ventilated with the manual ventilation bag.
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 93 / 150
H E Y E R N a r k o m a t+
7.3.1.1 Alarm limit setting in “manual/spontaneous” mode
The alarm limits in the “Manual/spontaneous” ventilation mode are set by pressing the “Limits” key on the operating panel.
Above the graphics representation the "Alarm limits page 1 of 2" window is opened, showing the respective alarm parameters in pairs.
Fig. 63 Alarm limits 1 in the Manual/spontaneous ventilation mode
With the buttons “On” and “Off” the alarms, with the exception of In O2 (O2 max, O2 min), can be activated and deactivated.
Explanation of individual alarms:
T.Vol. min Limit for minimum tidal volume
in ml.
Rate max Limit for maximum breath or
ventilation rate in 1/min
P max Maximum pressure limit (high
pressure limit) in cmH2O
High FiO2 Limit for maximum oxygen
concentration in %
MVol. min Limit for minimum minute
volume in l.
Rate min Limit for minimum breath or
ventilation rate in 1/min
P min Minimum pressure limit (low
pressure limit) in cmH2O
Low FiO2 Limit for minimum oxygen
concentration in %
94 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
Alarm limit setting in “manual/spontaneous” mode (continued)
Through pressing the button “limits” a second time, the alarm limits for the gas module can be set.
Above the graphics representation the "Alarm limits page 2 of 2" window is opened, showing the respective alarm parameters in pairs.
Fig. 64 Display “Alarm limits 2 in the Manual/spontaneous ventilation mode”
Explanation of individual alarms:
High CO2 Limit for the maximum end
CO2 concentration
Low CO2 Limit for the minimum end
CO2 concentration
RB CO2 CO2 re-breathing alarm.
The maximum inspiratory CO2 concentration
High Iso. Limit for the maximum
inspiratory anesthesia gas concentration
Low Iso. Limit for the minimum
inspiratory anesthesia gas concentration
High N2O Limit for the maximum
inspiratory N2O concentration
Low N2O Limit for the maximum
inspiratory N2O concentration
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 95 / 150
H E Y E R N a r k o m a t+
Alarm limit setting in “manual/spontaneous” mode (continued)
The alarm limits are selected and changed with the aid of the encoder or are switched “On” or “Off” in pairs.
To change an individual alarm parameter, it is selected with the encoder. The rotary display button is activated by pressing the encoder. The frame around the rotary
button is highlighted. The alarm limit value is decreased by turning the encoder anti-clockwise and increased by a
clockwise rotation. The alarm limit settings of the parameters offer different value graduations. The changed setting is stored by pressing once again on the encoder and the frame around
the rotary button returns to normal. The alarm limit window is closed by pressing once again on the “Limits” key, or closes
automatically after 10 seconds.
Table of adjustable alarm limits in manual/spontaneous mode
Parameter default for ON/OFF default setting value graduation adjustable values TVol. min. On 350 [ml] 50 [ml] 50 – 1800 [ml] MVol. min 4 [l] 0.5 [l] 0.5 – 15 [l] Rate max. Off 25 [1/min] 5 [1/min] 10 – 96 [1/min] Rate min. 2 [1/min] 2 [1/min] 2 – 30 [1/min] P max On 50 [cmH2O] 5 [cmH2O] 10 – 80 [cmH2O] P min 12 [cmH2O] 2 [cmH2O] 2 – 70 [cmH2O] High FiO2 On (always) 100 [%] 5 [%] 30 – 100 [%] Low FiO2 18 [%] 2 [%] 18 – 90 [%] High CO2 On 7,9 [%] 0,2 [%] 0-10 [%] Low CO2 3,2 [%] 0,2 [%] 0-10 [%] RB CO2 --- 1,2 [%] 0,2 [%] 0-10 [%] High Iso. Off 3,0 [%]* 0,2 [%] 0-19 [%] Low Iso. 0,0 [%] 0,2 [%] 0-19 [%] High N2O Off 82 [%] 2 [%] 0-82 [%] Low N2O 0 [%] 2 [%] 0-82 [%]
Fig. 65 Alarm limits in manual/spontaneous mode
For manual ventilation mode the alarm parameters should be switched on and adjusted to values, referred to the requirements for a save monitoring of the patient.
For a spontaneous ventilation mode the airway pressure alarms P max and P min have not to be switched on.
The default setting for the alarm volume is set to “0” in the manual/spontaneous mode, which is displayed through a crossed loudspeaker. The acoustic alarms for the configurable alarms will be suppressed. Technical alarms and the O2 min alarm are not affected by this feature. The acoustic alarm suppression can be switched off by changing the alarm volume in the menu window.
* The default setting for the Desflurane high alarm is 8 %
96 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
“Manual/spontaneous” ventilation mode (continued)
Fig. 66 Graphics window in manual/spontaneous ventilation mode
The graphics menu in the manual/spontaneous ventilation mode allows the setting of the time axis for the display of the ventilation curves.
16, 24 or 32 seconds can be set as values. 16 seconds is the default setting. The desired time axis representation for the real time curves can be chosen by selecting the
respective screen key. In the real time display 2nd window you can choose which value shall be displayed. CO2
capnogram* or expiratory flow curve The Y scaling of the pressure graph can be set to 20, 40, 60 or 80 cmH2O. By pressing the "Graphics" key for the second time, the window is closed again and the
selected setting is stored.
*only for apparatus option with integrated gas module
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 97 / 150
H E Y E R N a r k o m a t+
7.4 Controlled Ventilation Mode
The NARKOMAT+ anesthesia system offers two modes for controlled ventilation, the CMV child and the CMV adult settings on the selection switch of the operating panel. The differences between the two modes are the various ventilation parameters and alarm limits for these patient groups.
Select the ventilation mode, depending on the ventilation requirements of the patient. Different bellows are not required for the ventilation of adults and children. The used tube and filter systems should be fitted to the patient.
Described is only the ventilation mode CMV-Adult. Differences in the ventilation CMV-Child will be explained specifically.
98 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
H E Y E R N a r k o m a t+
7.4.1 CMV adult ventilation mode
(The screen titles change according to the ventilation modes Adult / Child) Set the operating panel selection switch to “CMV adult” for the controlled ventilation of adults. In the appearing display the requested ventilation mode “CMV-adult” can be selected through
turning and pressing the encoder button. The ventilation modes (S)CMV and PCV are options. The appearance of this window
depends on the apparatus configuration.
Fig. 67 Display “ventilation mode adult”
Rev. 3.1 – 05/09 HEYER Narkomat+, Operator’s manual 99 / 150
H E Y E R N a r k o m a t+
CMV adult ventilation mode (continued)
The parameters for the machine ventilation of adults can be determined and controlled on the display. The figure shows the standard setting for this patient group.
To implement a controlled ventilation, these parameters must be confirmed with the encoder button by pressing the "Push ENTER to Activate" display key.
Fig. 68 "CMV-Settings / Adult" display
The parameter setting fields determine the following:
T. Vol. Tidal volume of the controlled
ventilation in ml
Rate Ventilation rate in 1/min
PEEP Setting of a PEEP in cmH2O
Plateau A plateau for the inspiratory
pressure curve expressed in % to the overall inspiration period.
I:E Ventilation time ratio of inspiration
and expiration
Sigh Setting for a sigh ventilation. Every
100 breathing cycles an inspiration with 1.5 times the volume and 1.5 times the duration is carried out, the display key shows "ON".
100 / 150 HEYER Narkomat+, Operator’s manual Rev. 3.1 - 05/09
Loading...