Heyer Narkomat Service manual

+
HEYER Narkomat

Anesthesia System
Rev.2.1.0
Software version Service software: 1.8.3
HEYER MEDICAL AG
Carl-HEYER-Strasse 1-3
56130 Bad Ems
Germany
Tel. +49 - 2603 – 791 - 3
Fax. +49 - 2603 – 791 - 209
E-mail: info@HEYERmedical.de
2
Table of Contents
0 Table of Contents
0 Table of Contents ............................................................................................... 3
0.1 Table of Figures ........................................................................................................................ 6
1 General Information............................................................................................ 7
1.1 Guidelines ................................................................................................................................. 7
1.2 Product improvements .............................................................................................................. 7
1.3 Manufacturer’s Liability .............................................................................................................7
1.4 Manufacturer’s specification ..................................................................................................... 8
1.5 Updating status .........................................................................................................................8
1.6 Warning, Precautions and Notes ..............................................................................................9
1.6.1 WARNINGS .......................................................................................................................9
1.6.2 PRECAUTIONS............................................................................................................... 10
1.6.3 NOTES.............................................................................................................................11
2 Theory of Operation ......................................................................................... 13
2.1 Microprocessor-Controlled Ventilator......................................................................................13
2.2 Patient Module ........................................................................................................................13
2.3 Gas Conditioning Unit .............................................................................................................13
2.4 The Ventilator Unit ..................................................................................................................13
2.5 Adjustable Alarms ...................................................................................................................13
2.5.1 P
2.6 Fresh Gas Decoupling ............................................................................................................14
2.7 Compliance Compensation..................................................................................................... 14
2.8 Electrical supply ......................................................................................................................15
2.8.1 Electrical components......................................................................................................16
2.9 Power supply module..............................................................................................................17
2.9.1 Connectors on power supply board AVM 2-1.................................................................. 18
2.9.2 Fuses on power supply board AVM 2-1...........................................................................19
2.9.3 Charging / discharging control for the battery:................................................................. 20
2.9.4 Status Indicators Of Battery Control:...............................................................................20
2.10
2.10.1
2.11
2.11.1
2.11.2
2.11.2.1 Plug connectors on board AVM 3-1 ........................................................................ 24
2.11.2.2 Plug connectors on board AVM3-2 ......................................................................... 25
2.12
2.13
2.14
2.14.1
2.14.2
2.14.3
2.14.4
2.14.5
2.14.6
2.14.7
2.15
2.15.1
2.15.2
2.15.3
2.15.3.1 CMV mode, inspiration............................................................................................32
2.15.3.2 CMV mode, expiration............................................................................................. 33
2.15.3.3 Manual mode, inspiration ........................................................................................34
2.15.3.4 Manual mode, expiration.........................................................................................35
2.15.3.5 Spontaneous mode, inspiration............................................................................... 36
2.15.3.6 Spontaneous mode, expiration................................................................................37
2.15.4
2.15.4.1 Ventilation bellows system ......................................................................................38
2.15.4.2 Manual Respiration Bag / Reservoir........................................................................38
limiting on alarm violation.........................................................................................14
max
Module 1..............................................................................................................................21
Connectors on module 1..............................................................................................22
Ventilator Module 2, AVM 3-1 and AVM 3-2 ....................................................................... 23
Parts list module 2........................................................................................................24
Connectors on module 2..............................................................................................24
Display.................................................................................................................................25
Battery .................................................................................................................................25
Ventilator pneumatic............................................................................................................26
Ventilator pneumatic drive............................................................................................27
HP pressure reducer....................................................................................................27
LP- double stage pressure reducer.............................................................................. 27
Solenoid valves MV1 to MV4 .......................................................................................27
Pneumatic driving module............................................................................................ 28
Flow metering module.................................................................................................. 28
Tube color coding.........................................................................................................28
The patient module (circle system) ..................................................................................... 29
Top and back side view on the patient module............................................................ 29
Bottom and back side view on the patient module....................................................... 30
Functional representations of the patient module........................................................31
Components of the patient module..............................................................................38
3
Table of Contents
2.15.4.3 CO2 absorber...........................................................................................................38
2.15.4.4 Inspiratory and expiratory valves.............................................................................38
2.15.4.5 Airway pressure relief valve .................................................................................... 38
2.15.4.6 Room Air valve........................................................................................................38
2.15.4.7 Diaphragm valves....................................................................................................38
2.15.2.7.1 Status of the diaphragm valves:....................................................................... 39
2.15.2.7.2 Machine OFF or Ventilator in Standby............................................................. 39
2.15.2.7.3 Manual / Spontaneous Mode, INSPIRATION And EXPIRATION.................... 39
2.15.2.7.4 CMV Child / Adult Mode, INSPIRATION..........................................................39
2.15.2.7.5 CMV Child / Adult, EXPIRATION..................................................................... 39
2.15.2.7.6 CMV Adult, EXPIRATION And PEEP..............................................................39
2.15.2.7.7 Compliance Test Patient Module .....................................................................40
2.15.2.7.8 Leak Test Patient Module and Fresh Gas Module...........................................40
3 Repair Information............................................................................................ 41
3.1 Introduction .............................................................................................................................41
3.2 Warnings and precautions ......................................................................................................41
3.2.1 Precautions...................................................................................................................... 41
3.2.2 Warnings..........................................................................................................................41
3.3 Troubleshooting Guidelines ....................................................................................................42
3.4 Troubleshooting Charts...........................................................................................................42
3.4.1 Error Messages during ventilation...................................................................................42
3.4.2 Alarm messages during the compliance test, leak test and O2 calibration..................... 46
3.4.2.1
3.4.2.2
3.5 Required Tools........................................................................................................................50
3.6 Disassembly instructions.........................................................................................................50
3.6.1 Connecting and disconnecting vaporizers....................................................................... 50
3.6.2 Removing the compressed gas tanks (PIN-Index)..........................................................50
3.6.3 Removing the Patient Module..........................................................................................51
3.6.4 Removing the CO2 Absorber Canister............................................................................. 51
3.6.5 Removing the Bellows and Dome System ...................................................................... 51
3.6.6 Removing the Airway Pressure Limiting Valve (APL)...................................................... 51
3.6.7 Inspecting/replacing the Decoupling, Bellow and Expiratory Valves...............................52
3.6.8 Disassembling the Room Air, Inspiration and expiration valves...................................... 52
3.6.9 Removing the Gas Block Module .................................................................................... 52
3.6.10
3.6.11
3.6.12
3.6.13
3.6.14
3.6.15
Messages during the system tests........................................................................... 47
Symptoms of Fuse Failures...................................................................................... 49
Removing the flow tubes.............................................................................................. 52
Removing the pressure gauges ...................................................................................53
Removing the Module 2 Circuit Board Set................................................................... 53
Removing Module1 Circuit Board Set..........................................................................53
Removing the Power Supply Module........................................................................... 53
Removing Internal Regulators, Proportional Valve and Flow Divider..........................54
4 Maintenance and Calibration........................................................................... 55
4.1 Introduction .............................................................................................................................55
4.2 Calibration Warnings and Precautions....................................................................................55
4.3 Test Procedure........................................................................................................................56
4.3.1 General ............................................................................................................................56
4.3.2 Inspecting/replacing consumable parts ........................................................................... 56
4.3.3 Power supply checks.......................................................................................................61
4.3.4 Functional Tests...............................................................................................................62
4.3.4.1
4.3.4.2
4.3.4.3
Pneumatic tests........................................................................................................62
Alarm Tests...............................................................................................................63
Electrical Tests ......................................................................................................... 65
5 Service Software............................................................................................... 66
5.1 Identity Check .........................................................................................................................67
5.2 Main Menu ..............................................................................................................................68
5.3 Set Date and Time ..................................................................................................................69
5.4 Choose Service.......................................................................................................................70
5.5 Calibration Routines................................................................................................................71
5.5.1 O2 Sensor ADC Calibration* ........................................................................................... 72
5.5.2 Pressure Sensor #1 ADC Calibration..............................................................................74
5.5.3 Pressure Sensor #2 ADC Calibration..............................................................................76
4
Table of Contents
5.5.4 Temperature Sensor Calibration......................................................................................78
5.5.5 Calibration Of Proportional Valve ADC............................................................................ 80
5.5.6 Calibration Of Internal Flow Sensor................................................................................. 82
5.5.7 Fixing Of Calibration Characteristic Proportional Valve................................................... 84
5.5.8 Calibration Of External Flow Sensor................................................................................86
5.5.9 Test Menu Routines.........................................................................................................89
5.5.10
Test Characteristics Proportional Valve....................................................................... 90
5.5.11
Pressure, Oxygen and Temperature-Sensors ............................................................. 91
5.5.12
Keyboard Test.............................................................................................................. 92
5.5.13
Miscellaneous Functions.............................................................................................. 93
5.5.14
Solenoid Valves MV1 - MV4 ........................................................................................ 94
6 Preventive Maintenance................................................................................... 95
6.1 6 Month Service Interval .........................................................................................................95
6.2 12 Month Service Interval .......................................................................................................96
6.3 36 Month Service Interval .......................................................................................................96
6.4 Cleaning.................................................................................................................................. 97
6.4.1 Cleaning and disinfecting the apparatus .........................................................................97
6.4.2 Cleaning and sterilizing the Patient Module ....................................................................97
6.5 Battery Replacement and Maintenance..................................................................................97
6.5.1 Battery replacement.........................................................................................................97
6.5.2 Battery Maintenance........................................................................................................ 97
7 Order information............................................................................................. 98
5
Table of Contents
0.1 Table of Figures
Fig. 1 Line Terminal Block, 120/230VAC Supply .................................................................................15
Fig. 2 Overview electrical components.................................................................................................16
Fig. 3 Power supply module .................................................................................................................17
Fig. 4 Module 1 File-PC........................................................................................................................21
Fig. 5 View on module 2....................................................................................................................... 23
Fig. 6 Ventilator pneumatic...................................................................................................................26
Fig. 7 Top and backside view on the patient module...........................................................................29
Fig. 8 Bottom and backside view on the patient module...................................................................... 30
Fig. 9 Survey of the patient module......................................................................................................31
Fig. 10 Survey of the patient module, CMV mode, inspiration.............................................................32
Fig. 11 Survey of the patient module, CMV mode, expiration.............................................................. 33
Fig. 12 Survey of the patient module, Manual mode, inspiration ......................................................... 34
Fig. 13 Survey of the patient module, Manual mode, expiration..........................................................35
Fig. 14 Survey of the patient module, Spontaneous mode, inspiration................................................ 36
Fig. 15 Survey of the patient module, Spontaneous mode, expiration ................................................ 37
Fig. 16 Patient module, front view........................................................................................................ 57
Fig. 17 Patient module, top view ..........................................................................................................58
Fig. 18 Supply inlet filters and water traps ...........................................................................................59
Fig. 19 Alarm screen ............................................................................................................................63
Fig. 20 O2 Whistle adjustment (appearance may differ due to different configuration options) .......... 64
6
General Information
1 General Information
1.1 Guidelines
The information in this service manual exclusively refers to servicing and maintenance work of the anesthesia system HEYER Narkomat+. Please, read from the operator’s manual HEYER Narkomat+ the service, setting, maintenance and care of the apparatus, normally carried out by the user. These service instructions are to be used only by a skilled, trained and authorized service staff. The servicemen must be provided with the specified special tools and accessories. This service manual is utilized for business affairs and general customized information. HEYER gives no guaranty when using the information. The service technicians should have read and fully understood the service instructions prior to beginning with their service duties. The functional principles of the apparatuses have also been described in the user’s manual. The user’s manual contains general precautions, which are also of importance to the service technician.
1.2 Product improvements
HEYER reserves the right to improve their products or revise the instructions without prior notice. This manual deals with the status of the anesthesia system HEYER Narkomat+ at the time of issue. HEYER is not obliged to retrofit former models subject to improvements and modifications. An
exception to be examined will be made when improvements or modifications due to design and production deviations are influencing the patient’s safety or would entail malfunctioning of the apparatus.
1.3 Manufacturer’s Liability
HEYER can only be held liable for safety, reliability and fail-safe operation of the system, provided
that:
the system was being operated in conformity with the instructions given by the manufacturer,
certifications, readjustments, changes, or repairs have been carried out by authorized
personal,
service and maintenance was being made in conformity with the instructions given by the manufacturer.
the system was operated in a building with grounding equipment in compliance with the regulations issued by IEC, NFPA, and UL.
In no event will HEYER be liable for any special, incidental, or consequential damages, including loss of profits, whether or not foreseeable and even HEYER has been advised of the possibility of such loss or damage. HEYER disclaims any liability arising from a combination of its product with products from other manufacturers if the combination has not been endorsed by HEYER. Buyer understands that the remedies noted in HEYER's limited warranty are its sole and exclusive remedies.
7
General Information
1.4 Manufacturer’s specification
Product: HEYER NARKOMAT
Manufacturer:
+
HEYER Medical AG
Carl-HEYER-Strasse 1-3 56130 Bad Ems
Germany Tel. + 49 2603 / 791-3 Fax. +49 2603 / 791-209
E-mail: info@HEYERmedical.de
1.5 Updating status
State of this service manual: Rev. 2.1.0 of March 2004
8
General Information
1.6 Warning, Precautions and Notes
Warnings alert the user to potential serious outcomes (death, injury or serious adverse events) to the
patient or user. Precautions alert the reader to exercise special care necessary for the safe and effective use of the
device Notes are a general information statement concerning the Narkomat+ Please read adhere to all warning, precautions and notes listed here and in the appropriate areas
THROUGHOUT THIS MANUAL.
1.6.1 WARNINGS
WARNING: The Narkomat+ anesthesia machine works on line voltage and at high pressure.
Therefore, an electric shock hazard may exist when the instrument covers are removed. Repair and calibration procedures should only be performed by qualified personnel who follow proper servicing techniques. Warnings are given in appropriate locations
WARNING: In order to prevent an electrical shock, the machine (protection class I) may only be
connected to a correctly grounded mains connection (socket outlet with grounding contact).
WARNING: Possible explosion hazard. Do not operate near flammable substances. WARNING: The use of anti-static or electrically conductive breathing tubes, when utilizing high
frequency electric surgery equipment, may cause burns and is therefore not recommended in any application of this machine.
WARNING: Possible fire hazard. Fuses (i.e., additional sockets) must only be replaced by fuses of
the same type and with the same rating.
WARNING: Possible shock hazard. The machine may only be opened by qualified and authorized
service personnel.
WARNING: Compressed gasses are considered dangerous goods/hazardous materials per
I.A.T.A. Regulations. It is a violation of international law to offer any package or over pack of dangerous goods for transportation without the package being appropriately identified, packed, marked, classified, labeled and documented according to I.A.T.A. Regulations. Please refer to the applicable I.A.T.A dangerous goods regulations for further information.
WARNING: Never block airflow at the drive gas outlet. Blocking the drive gas outlet raises internal
pressures above specified limits and will result in permanent damage to internal sensors.
9
General Information
1.6.2 PRECAUTIONS
CAUTION: Refer to the maintenance intervals in the preventive maintenance section for guidance
on which steps are preformed when.
CAUTION: Use surgical gloves whenever touching or disassembling valves or other internal
components of the patient module.
CAUTION: If possible, always connect the output of the APL valve to the anesthetic removal line,
usually installed in the operation theater.
CAUTION: Carry out the daily checks specified on the checklist and do not operate the system in
case of a fault until the fault has been repaired.
CAUTION: The patient should be visually monitored by qualified personnel. In certain situations
circumstances may occur which may not necessarily trigger an alarm.
CAUTION: Always set the alarm limits so that the alarm is triggered before a hazardous situation
occurs. Incorrectly set alarm limits may result in operation personnel not being aware of changes in the patient’s condition.
CAUTION: This machine must only be operated by trained, skilled medical staff. CAUTION: Before starting the machine, the operating personnel must be familiar with operating
instructions and must have been instructed by a qualified instructor.
CAUTION: If the machine does not function as described, the machine must be examined and
possibly repaired by qualified service personnel, before being returned to use.
CAUTION: Handle the machine with care to prevent damage or functional faults. CAUTION: Ensure that the gas supply of the machine always complies with the technical
specification.
CAUTION: Before clinical use, the machine must be correctly calibrated and/or the respective
machine tests performed, as described in the operation instructions.
CAUTION: If the machine should show faults during the initial calibration or testing, the machine
should not be operated until the fault has been repaired by a qualified service technician.
CAUTION: After servicing, functional, sensor and system tests must be carried out before clinical
use.
CAUTION: Only bacteria filters with a low flow resistance must be connected to the patient
module and/or patient connector.
CAUTION: Failure to connect device to a grounded mains outlet may elevate leakage current in
excess of permissible values.
CAUTION: During transportation of the patient module, transportation should be applied to the
rear to protect the valve connectors.
CAUTION: After changing the co2 absorbent, carry out a system leak test. CAUTION: The spring in the top of the APL valve may not be stressed. After removal, place to
one side, taking care that the spring is not unduly loaded.
CAUTION: Only selectatec™ compatible vaporizers with interlock system may be used with the
Narkomat
+
unit.
10
General Information
CAUTION: After each exchange of a vaporizer, carry out a system leak test. CAUTION: Use cleaning agent sparingly. Excess fluid could enter the machine causing damage. CAUTION: The patient dome of the bellows system cannot be autoclaved. It is not in contact with
the ventilation gas. If soiled, the patient dome should be cleaned with water and liquid cleaning agent. The unit can be disinfected with a standard surface-disinfecting agent. Do not use alcohol.
CAUTION: Do not clean the machine while it is on and/or plugged in. CAUTION: Pressing quit at any time during the calibration procedure will cancel the session’s
settings and reload the previously stored calibration coefficients.
1.6.3 NOTES
NOTE: Unauthorized servicing may void the remainder of the warranty. Check with the factory
or with a local authorized HEYER dealer to determine the warranty status of a particular instrument.
11
General Information
THIS PAGE INTENTIONALLY LEFT BLANK
12
Theory of Operation
2 Theory of Operation
The anesthesia system HEYER NARKOMAT+ represents a flexible employable workstation to apply and monitor anesthesia inhalation in semi-closed and almost closed circuits in low flow techniques for minimized consumption of gas and anesthetics. The basic model Narkomat+ includes the following components:
2.1 Microprocessor-Controlled Ventilator
The Microprocessor-controlled ventilator allows with its dedicated patient module time-controlled, pressure limited, and a constant volume controlled ventilation for all patient groups of a body weight of 3 kg upwards. The system compliance of the patient module and respiratory tubes are automatically compensated by the ventilator. Low tidal volumes, therefore, can be dosed very accurately. Thanks to the different ventilation modes, a ventilation is feasible, even in case of complicated lung conditions. A comprehensive test and alarm management system prevents uncontrollable operating conditions and therefore provides the patient a high safety standard. The ergonomically designed control surface allows an easy operation of the ventilator. The display informs about the selected ventilation parameters and the measured values for volume, pressure, FiO2 and shows real time curves of the expiratory flow and airway pressure.
2.2 Patient Module
The patient circuit is integrated in a compact aluminum block. This block is brought to a moderate temperature to avoid the formation of water vapor. In additional, it includes a controlled emergency valve, a reservoir in form of a manual respiratory bag and an expiratory flow sensor. The patient module is connected to the basic unit through a locking handle. Activating the locking handle pulls the block toward the basic unit, and all connections to the docking block are checked in the compliance- and leak tests.
2.3 Gas Conditioning Unit
The flow meter block features all the necessary safety facilities, such as O2 pressure loss alarm and N2O shut-off. An integrated pneumatically controlled system provides a minimum of 25% oxygen concentration in the fresh gas flow at all flow settings (ratio system).
2.4 The Ventilator Unit
The microprocessor-controlled ventilator allows time-controlled, volume-constant and pressure-limited (CMV) or pressure targeted (PCV) artificial respiration without assisted or synchronized control functions. The system also allows manual respiration as well as spontaneous respiration of the patient. Adjustable PEEP, breathing time ratios (I:E) and pressure sustaining plateau functions are available. Volume-constant respiration is effected by the time control of the respirator and the fresh gas decoupling of the patient module. The ventilator computes the inspiratory flow required for the settings tidal volume (Vt), respiratory rate and the ventilation time ratio (I:E). The resultant tidal volume (Vt) is delivered to the patient at high accuracy
2.5 Adjustable Alarms
Minimum and maximum alarm limit settings are available for Peak Pressure, Mean Pressure and FiO2 minimum alarm settings are available for tidal volume and minute volume.
13
Theory of Operation
2.5.1 P
limiting on alarm violation
max
Exceeding the P from exceeding the high alarm setting. in the CMV mode, the setting of the p
alarm limit automatically halts the inspiratory phase preventing airway pressure
max
peak pressure
max
provides pressure limitation. when reaching this pressure limit, a warning (alarm) “peak greater than p
” is displayed, and the inspiration is discontinued. The next inspiration is at a regular time interval,
max
the time control of the respiratory unit does not increase the respiratory rate. the result is a decrease of the tidal volume “Vt” and minute volume “m vol.”. The respirator responses to pressure limitation through displaying the p
alarm constantly.
max
2.6 Fresh Gas Decoupling
The tidal volume (Vt) is adjusted independently of the adjusted fresh gas flow. This is achieved by fresh gas decoupling. In inspiratory procedure the fresh gas flow is uncoupled by the decoupling and expiratory diaphragm valves from the internal subsystem of the patient module. Fresh gas either is filling the manual respiration bag (reservoir) or is led into the scavenging through the airway pressure relief valve.
2.7 Compliance Compensation
The tidal volume (Vt) is automatically corrected in response to the compliance of the patient module and patient tubes. Volume reduction due to system compliance is compensated as a result of this.
14
Theory of Operation
receptacles
2.8 Electrical supply
The line cord of the Narkomat+ is connected to an internal terminal block. Directly connected to this terminal are convenience receptacles and the unit itself. Line supply to the receptacles is available with the line cord connected. Line supply for the unit is switched by the main switch at the right side of the unit. The two pole switched power supply is protected by each one 5A circuit breaker. The switched line supply is connected to the AC IN connector on power supply board AVM 2-1. Another third pole of the main switch is connected via the terminal to the board AVM 2-1 (X5:5,6), enabling to detect a main switch OFF position or missing line supply.
Line cord
N 120/230 V
L 120/230 V
X5 pin 6
PE
X5 pin 5
X5 pin 4
X5 pin 3
power supply board
int. ground
wiring
Circuit breakers ventilator
Main switch
Power supply board
5 A
1
8
1 1
1
1
2
2
2
2
2
3 4
7
3
3 4
4
5
6
6
5
Circuit breaker 5 A
battery
Convenience
Fig. 1 Line Terminal Block, 120/230VAC Supply
The 120 V configuration has
additional circuit breakers
15
2.8.1 Electrical components
floppy drive
flow meter backlight
Theory of Operation
hard disc
module 1 PC
RS 232 interface
ventilator control panel
transformer
power supply
battery
motor
fan
proportional valve
oxygen sensor
main switch
I²C interface
module 2 AVM 3-1
module 2 AVM 3-2
PT 100
heat blanket
P1P2
ext. flow sensor
int. flow sensor
solenoid valve block
Fig. 2 Overview electrical components
16
Theory of Operation
2.9 Power supply module
The power supply module is located in the frame behind the drawers. This module serves for the voltage supply of the ventilator modules 1 and 2, the flow meter block illumination, the patient circuit heating blanket and the charging / discharging control for the battery. The power supply board allows the supply with 230 VAC or 115 VAC, the supply voltage can be selected with a switch on the power supply board.
Fig. 3 Power supply module
17
Theory of Operation
2.9.1 Connectors on power supply board AVM 2-1
Connector connected to Pin. def. Func. / Signal Color / No.
X1 Module 1 X1-1 +5V yellow X1-2 GND red X1-3 GND green X1-4 +12V blue
Connector connected to Pin. def. Func. / Signal Color / No.
X2 X2-1 +5V X2-2 GND X2-3 GND X2-4 +12V
Connector connected to Pin. def. Func. / Signal Color / No.
X3 X2-1 - X2-2 - X2-3 GND X2-4 +12V
Connector connected to Pin. def. Func. / Signal Color / No.
X4 X4-1 +5V X4-2 GND X4-3 GND X4-4 +12V
Connector connected to Pin. def. Func. / Signal Color / No.
X5 Battery, X5-1 BATT+ brown Fuse for Battery X5-2 BATT- black Main switch X5-3 Cir.break. bat. black 6 X5-4 Cir.break. bat. black 5 X5-5 Main switch black 4 X5-6 Main switch. black 3
Connector connected to Pin. def. Func. / Signal Color wire isolat.
X6 Transformer X6-1 sec.IIa 15VAC blue RT110/12 secondary X6-2 sec IIa green
X6-3 sec. IIb 15VAC blue X6-4 sec IIb green
Connector connected to Pin. def. Func. / Signal Color wire isolat.
X7 Transformer X7-1 prim.I, L120/230V RT110/12 primary X7-2 prim.I, N120/230V X7-3 prim.II,L120/230V X7-4 prim. II, . violet
Connector connected to Pin. def. Func. / Signal Col. wire isolat.
X11 Transformer X11-1 Sec.Ia 22VAC red RT110/22 secondary X11-2 Sec Ia yellow
X11-3 Sec. Ib 22VAC red X11-4 Sec Ib yellow
AC
AC
AC
black/ orange white
18
Theory of Operation
Connector connected to Pin. def. Func. / Signal Color / No.
X12 int. connection X12-1 AC, N 110V/230V blue AC IN terminal X12-2 AC , L 110V/230V black
Connector connected to Pin. def. Func. / Signal Color / No.
X13 Module 2 X13-1 +12V red Board X13-2 GND blue AVM 3-2 X13-3 +5V(1) red X13-4 GND blue
Connector connected to Pin. def. Func. / Signal Color / No.
X14 Module 2 X14-1 +5V(2) red Board X14-2 GND blue AVM 3-2 X14-3 -12V red X14-4 GND blue
Connector connected to Pin. def. Func. / Signal Color / No.
X15 Module 2 X15-1 Control Sign. white Board X15-2 Control Sign. white AVM 3-2 X15-3 Control Sign. white X15-4 Control Sign. white
Connector connected to Pin. def. Func. / Signal Color / No.
X16 Module 2 X16-1 GND blue AVM 3-2 X16-2 +31V red
Connector connected to Pin. def. Func. / Signal Color / No.
X17 solid state relay X17-1 22VAC to heat. bl. black heating blanket X17-2 22VAC to relay black X17-3 + Control voltage black X17-4 GND black
Connector connected to Pin. def. Func. / Signal Color / No.
X22 DC-AC convert. X22-1 +supply voltage red for flow meter X22-2 - illumination X22-3 - X22-4 GND black
2.9.2 Fuses on power supply board AVM 2-1
Fuse No. Fuse value Fuse protects F1 4 A MT, medium slow-blow DC / DC-converter +5V F7 8 A M, medium -blow DC / DC-converter +12V F8 8 A M, medium -blow Transformer, Sec. IIa F9 8 A M, medium -blow Transformer, Sec. IIb F10 10 A T, slow-blow Transformer, Sec. Ia F11 10 A T, slow-blow Transformer, Sec. Ib
19
Theory of Operation
2.9.3 Charging / discharging control for the battery:
The charging / discharging control for the battery is also located on the power supply board AVM 2-1. Status indicators as red, green and yellow LEDs are located on the board to show functions like check or charging of the battery.
2.9.4 Status Indicators Of Battery Control:
LED LED light Battery status
LED yellow (V28) Continuous charging LED yellow (V28) Flashing check LED green (V37) Continuous fully charged LED red (V38) flashing or continuous. defect
20
Theory of Operation
2.10 Module 1
Ventilator module 1 consists of a File-PC with an I/O card, memory card and 3,5” floppy drive. This module serves for the operation of the user surface with different menus displayed on a 10.4” color TFT display in combination with the settings from the control panel. All safety functions, e.g. the complete ventilator control, are located on ventilator module 2. Module 1 is connected to module 2 by means of an RS 232C- interface. Module 1 has no safety function for the ventilator control. It serves to give the functions for parameter and alarm settings to module 2. Module 2 sends the measured parameters or generated alarm signals to module 1 to display.
Fig. 4 Module 1 File-PC
21
Theory of Operation
2.10.1 Connectors on module 1
Connector connected to Plug conf. Func. / Signal X4 Module 2 10 pole Communication with module 2
Connector connected to Plug conf. Func. / Signal X5 EL Display 20 pole Connected to the
display
Connector connected to Plug conf. Func. / Signal X6 Power supply 4 pole Power suppy
Connector connected to Plug conf. Func. / Signal X7 Touch screen 32 pole communication with touch screen
Connector connected to Plug conf. Func. / Signal X8 Module 2 14 pole communication with module 2
Connector connected to Plug conf. Func. / Signal X9 Module 2 28 pole communication with module 2
Connector connected to Plug conf. Func. / Signal X10 Selector switch 14 pole communication With the selector
switch
Connector connected to Plug conf. Func. / Signal X19 LCD Display Flat cable
connector
Connector connected to Plug conf. Func. / Signal X20 (on the backside)
Display illumination
4 pole Power supply
Connected to the display
display illumination
22
Theory of Operation
2.11 Ventilator Module 2, AVM 3-1 and AVM 3-2
Ventilator module 2 serves for the active ventilator control. Module 2 consists of two CPU boards, the CPU board AVM3-1 with µP1 serves for a continuous validation of control actions generated by CPU board AVM3-2 and for the communication between module 2 and module 1. The CPU board AVM3-2 with µP2 generates all active control signals, after a validation of these signals by µP1, AVM3-2 is enabled to operate the different active elements like the solenoid vales and proportional valve.
Keyboard Alarm Panel
RS232
X4
P1
P2
X3X2
CPU-board
IF
EF
1
O2
1
P9
PV
1
X8
X9
X10
X11
AVM 3-2
P3
P5
1
1
1
1
P6
P1
P8
P7
P1
P1
P11 P10
R33
Pressure Sensor #2
Pressure Sensor #1
P2P1
X5
1
X7
GND
X1
X8
P2
P2
P2
X3
X1
1
7
23
456
8
MD
1
ON
CPU-board AVM 3-1
X13
X7
1
GND
Q1
V29
V30
V38
V19
AVM 3-1 Software
+12 GND
GND
+5(1)
1
X15
GND
GND
+5(2)
Power Supply
-12
Fig. 5 View on module 2
DIS/AC
HTON
M
BATT
+31
1
swrtrs wsswge
Motor
(-) (+) -> lösen
X16
Endsch.
Start
X11
1
gn
23
Theory of Operation
2.11.1 Parts list module 2
Pos. No. Description Order No. 1 Ventilator module 2, complete 460-1150
2.11.2 Connectors on module 2
2.11.2.1 Plug connectors on board AVM 3-1
Connector connected to Plug conf. Func. / Signal
X2 Module 1 28 pole communication with module 1
Connector connected to Plug conf. Func. / Signal
X3 Module 1 14 pole communication with module 1
Connector connected to Plug conf. Func. / Signal
X4 Module 1 10 pole communication with module 1
Connector connected to Pin. def. Func. / Signal Color / No.
X7 PT-100 sensor X7-1 temperature dep. blue X7-2 resistance red
Connector connected to Pin. def. Func. / Signal Color / No.
X9 Fan ventilator X9-1 +12V red FAN pneumatic X9-2 fan control signal yellow module X9-3 n.c. X9-4 n.c. X9-5 GND black
Connector connected to Pin. def. Func. / Signal Color / No.
X10 Driving gas X11-1 +12V white Driving gas switch switch to detect X11-2 GND brown a failure of the gas X11-3 n.c. supply X11-4 n.c. X11-5 n.c.
Connector connected to Pin. def. Func. / Signal Color / No.
X11 over temperature X11-1 22VAC when red HTSEC switch (NC) X11-2 switch is open blue
patient X11-3 n.c. block heating X11-4 n.c. X11-5 n.c.
24
Theory of Operation
2.11.2.2 Plug connectors on board AVM3-2
Connector connected to Pin. def. Func. / Signal Color / No. X5 internal X5-1 signal (+1-5 V) orange IF flow sensor X5-2 +10V mint X5-3 GND. violet X5-4 n. c. X5-5 n. c.
Connector connected to Pin. def. Func. / Signal Color / No. X7 external X7-1 flow dep. signal. red EF flow sensor X7-2 n.c. X7-3 n.c. X7-4 n.c.
X7-5 flow dep. signal. blue
Connector connected to Pin. def. Func. / Signal Color / No. X8 Oxygen cell X8-1 n. c. O2 X8-2 + voltage fuel cell red X8-3 - voltage fuel cell blue X8-4 n.c. X8-5 n.c.
Connector connected to Pin. def. Func. / Signal Color / No. X3 proportional X3-1 + 12V blue PV valve X3-2 n. c. X3-3 control volt. 0-5V gray X3-4 n.c. X3-5 GND green
Connector connected to Pin. def. Func. / Signal Color / No. X11 solenoid X11-1 GND ( MV4) brown / 1 valve X11-2 + 10V (MV4) red / 2 block X11-3 GND (MV3) orange / 3 X11-4 + 10V (MV3) yellow / 4 X11-5 GND (MV1) green /5 X11-6 + 10V (MV1) blue / 6 X11-7 GND (MV2) violet / 7 X11-8 + 10V (MV2) gray / 8
2.12 Display
The display of the ventilator is a 10.4” color TFT. It is connected to the module 1.
2.13 Battery
The battery is a maintenance free seal lead acid type. The recharging time is a maximum of 7 hours with a fully depleted battery. The backup time is about 30 minutes with a fully charged battery. To prevent unintended loss of battery operation, it is recommended to replace it with a new HEYER battery every 3 years.
25
2.14 Ventilator pneumatic
AIR or O2
HP-pressure regulator
Theory of Operation
AIR P = 3.5 - 6.5 bar
O2 p = 3.5 - 6.5 bar
1
B
LP-pressure regulator assembly
C
2
Proportional valve
Flowsensor (internal)
3
A
Flowdivider
4
Solenoid valve bank
6
MV1
5 2
MV4
1 8
MV2
7
4
MV3
3
Freshgas
Connecting block patient module
Fig. 6 Ventilator pneumatic
5
P
57 6
3
124
P1
P2
CPU-BOARD AVM 3-2
26
Theory of Operation
2.14.1 Ventilator pneumatic drive
Air or Oxygen serves as driving gas for the ventilator. In addition to the flow meter block, a pressure reducer, reducing the supply pressure to 200kPa (2 bar; 29 PSI), is supplied by the Air or oxygen connection. This pressure represents the high system pressure of the ventilator. The pressure reducer is named HP pressure reducer. The HP pressure reducer is placed ahead of the proportional valve that generates the driving gas flow during the inspiratory phase. This flow is led through the flow divider and fills the bellows dome that surrounds the silicone bellow. In response to the inspiratory and expiratory phases of a respiration cycle, the pneumatic valves (diaphragm valves) of the patient module are activated by solenoid valves. These solenoid valves, MV1, MV3, and MV2 are supplied by a second pressure reducer assembly (LP- double stage pressure reducer) at the low system pressure, together with MV4, with about 20kPa (200 mbar; 2.9 PSI)// 25kPa (250 mbar; 3,63 PSI). Cyclic activation of the proportional valve and solenoid valves is steered by the processors of module 2 according to the parameters set on the control panel of the ventilator. A description of the individual components will be found in the following chapters:
2.14.2 HP pressure reducer
The HP pressure reducer serves for a stabilization of the supply pressure for the proportional valve. The flow generated by the proportional valve thus becomes independent of pressure variations. The setting of the HP pressure reducer (200 kPa (2 bar; 29 PSI)) at the same time determines the maximum inspiratory flow of 75 +0/-4 L/min with respect to the respirator.
2.14.3 LP- double stage pressure reducer
This pressure reducer assembly is supplied with oxygen in parallel to the measuring tube block or ratio system, respectively. It reduces the input pressure to 20 kPa (200 mbar; 2.9 PSI)// diaphragm valve, To 25 kPa (250 mbar; 3,63PSI)// membrane valve and is connected to the solenoid valve bank.
2.14.4 Solenoid valves MV1 to MV4
The solenoid valves MV1, MV2, MV3 and MV4 are mounted on a bank behind the patient module docking station to which the tubes to the docking station of the patient module and the LP- double stage pressure reducer are connected.
MV1: This solenoid valve activates, in the CMV mode, the valve for closing the outlet of the bellows dome (bellows control valve) through the line 5, when driving gas flows in for inspiration. It is supplied by the LP- double stage pressure reducer.
MV2: This solenoid valve activates, in the CMV mode, the valve closing the expiratory channel of the patient module (expiratory valve), via MV2 through the line 1. It is supplied by the LP- double stage pressure reducer via MV4.
MV3: This solenoid valve activates, in the CMV mode, the valve for the fresh gas decoupling (decoupling valve) through the line 6. It is supplied by the LP- double stage pressure reducer.
MV4: This solenoid valve is used for internal supply of MV2 with pressure of the LP- reducer.
27
Theory of Operation
2.14.5 Pneumatic driving module
The driving module consists of the proportional valve and flow divider with an internal flow sensor. The proportional valve supplied by the HP pressure reducer generates a driving gas flow of 0-75 l/min in relation to the control voltage of the proportional valve of 0 - 5VDC.
The control voltage of the PV, required for the pre-selected parameter settings, is generated by the CPU board AVM 3-2. The driving gas flow Q
Q
drive gas Vt = generated tidal volume Ti = Inspiratory time
= Vt/Ti with Q
drive gas
is in the following in relationship with the tidal volume:
drive gas
= driving gas flow
2.14.6 Flow metering module
The flow metering module attached to the proportional valve consists of a flow sensor with a measuring range of 0 - 1 l/min and a flow divider. The flow divider splits the total flow in a ratio of 79: 1, and the bypass flow is lead to the flow sensor. A closed loop that allows high accuracy and back up in the generation of the tidal volume results from this configuration via the module 2 CPU-boards.
2.14.7 Tube color coding
Despite the local color coding system, all the pneumatic tubes inside the apparatus will be according to the coding label inside the unit, if it is not according to the local color coding system, next to the flow meter block. Typically it will be according to one of the follow standards.
Gas ISO 32 Standard US standard
O2 White Green N2O Blue Blue AIR Black/White Yellow
On the outside of the apparatus, the coding will be as ordered.
28
Theory of Operation
2.15 The patient module (circle system)
2.15.1 Top and back side view on the patient module
1 2 3 4
6 5 12 15 8 9 10 11 13 14 7 7
Fig. 7 Top and backside view on the patient module Pos. Description
1. Expiratory valve (passive)
2. Airway pressure limiting valve (APL)
3. Inspiratory valve (passive)
4. Room air valve, emergency air valve
5. Connector of expiratory flow sensor
6. Connectors of power supply heating blanket
7. Bores for guiding pins of connecting block
8. Expiratory valve (active)
9. Port for pressure measurement
10. Port for expiratory valve activation in the manual mode
11. Port for driving gas
12. Locking bolt
13. Bellows diaphragm valve
14. Decoupling diaphragm valve
15. Port for fresh gas
29
Theory of Operation
1 2
3
4
5
2.15.2 Bottom and back side view on the patient module
6
7
8
Fig. 8 Bottom and backside view on the patient module Pos. Description
1 2 3 4 5 6 7 8
Fixing screws for locking bolt Exhaust for driving gas Driving gas in- and outlet to bellows dome Thread for CO2 absorber canister fixing, outlet from patient module to absorber Inlet for re-breathing gas from absorber to patient module Connector for bellows Inlet for ambient air valve / emergency air valve Port for anesthetic gas scavenging tube (30 mm cone)
30
Theory of Operation
2.15.3 Functional representations of the patient module
gas­monitoring
2
V
5
3
41
6
11
7
8
9
10
Fig. 9 Survey of the patient module
Pos. Description
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19
Patient's y-piece Side stream gas monitor Inspiratory bacterial filter Expiratory bacterial filter Airway pressure monitor connection Inspiratory valve (passive) Expiratory valve (passive) Spirometry sensor Airway pressure limiting valve (APL) Waste gas outlet Expiratory valve Room air valve, emergency air valve Ventilator control pneumatic CO2 absorber Bellows Decoupling valve Bellows dome Bellows valve Reservoir / manual ventilation bag
P
12
13
14
16
15
17
18
V
19
31
2.15.3.1 CMV mode, inspiration
gas monitoring
Theory of Operation
P
expiratory valve
V
V
is > 2 mbar
MV3 ="1" MV2 ="1"
decoupling valve
bellows valvewhen reservoir pressure
Fig. 10 Survey of the patient module, CMV mode, inspiration
MV4 ="1" MV1 ="1"
32
2.15.3.2 CMV mode, expiration
gas monitoring
Theory of Operation
expiratory valve
V
when reservoir pressure is > 2 mbar
P
V
MV3 ="0" MV2 ="0"
MV4 ="0"
decoupling valve
bellows valve
Fig. 11 Survey of the patient module, CMV mode, expiration
MV1 ="0"
33
2.15.3.3 Manual mode, inspiration
gas monitoring
Theory of Operation
P
expiratory valve
V
when the set respiratory pressure is attained
V
MV3 ="0" MV2 ="0"
MV4 ="0" MV1 ="0"
Fig. 12 Survey of the patient module, Manual mode, inspiration
decoupling valve
bellows valve
34
2.15.3.4 Manual mode, expiration
gas monitoring
expiratory valve
Theory of Operation
decoupling valve
V
MV3 ="0" MV2 ="0" MV4 ="0"
MV1 ="0"
Fig. 13 Survey of the patient module, Manual mode, expiration
bellows valve
35
Theory of Operation
2.15.3.5 Spontaneous mode, inspiration
gas monitoring
P
emergency air valve opens at p < - 3 mbar
expiratory valve
V
decoupling valve
when reservoir pressure is > 2 mbar
V
MV3 ="0" MV2 ="0"
MV4 ="0" MV1 ="0"
Fig. 14 Survey of the patient module, Spontaneous mode, inspiration
bellows valve
36
Theory of Operation
2.15.3.6 Spontaneous mode, expiration
gas monitoring
P
expiratory valve
V
decoupling valve
waste gas
V
MV3 ="0" MV2 ="0" MV4 ="0" MV1 ="0"
Fig. 15 Survey of the patient module, Spontaneous mode, expiration
bellows valve
37
Theory of Operation
2.15.4 Components of the patient module
2.15.4.1 Ventilation bellows system
The ventilator’s driving system can be characterized as a constant flow generator. The driving gas of this generator fills the bellows dome to compress the bellows. The breathing gas is pressed out of the bellows into the circuit. Due to the weight of the descending bellows, it is self-filling when the driving gas flow is stopped. The bellows control valve enables the escape of the driving gas.
2.15.4.2 Manual Respiration Bag / Reservoir
In the manual mode this device acts as a normal breathing bag, enabling the user to ventilate the patient manually. In CMV mode this bag acts as a reservoir for fresh gas in the inspiration phase. In the expiratory phase the re-breathing gas and the fresh gas are accumulated in this reservoir. The bag’s in- and deflation corresponds to the volume balance of the breathing circuit and the patient. When a volume loss occurs, the bag deflates breath by breath and finally collapses. Such a collapsing breathing bag is an indicator for a leakage. The breathing bag must collapse if the fresh gas flow, reduced by the patient’s uptake, is not able to compensate the leakage.
2.15.4.3 CO2 absorber
The soda lime inside the absorber retains the carbon dioxide from the re-breathing gas. The flow orientation is upstream.
2.15.4.4 Inspiratory and expiratory valves
To ensure the gas flow direction the two one-way-valves in the inspiratory limb and expiratory limb are integrated inside the patient module. These valves are seen activating form the top of the module.
2.15.4.5 Airway pressure relief valve
In CMV position the APL valve closes the re-breathing system towards the scavenging line and limits the reservoir pressure to a constant value of 2 cmH2O. In manual mode the APL valve acts as a normal spring loaded pressure relief valve, limiting the maximum pressure in the re-breathing system.
2.15.4.6 Room Air valve
Due to the tendency of the descending bellows to refill itself by creating a negative pressure inside the bellows this valve prevents the patient from becoming any negative pressure by opening the diaphragm and allowing ambient air to refill the bellows.
2.15.4.7 Diaphragm / Membrane valves
Inside the patient module diaphragm/membrane valves operate the expiratory line, the bellows dome outlet and the fresh gas decoupling line during the CMV-mode. These valves are controlled by pressured gas to open or close the valve’s path. All diaphragm valves are open when not inflated. The power gas for these valves is controlled by the ventilator’s valve bank using a supply pressure of about 20 kPa (200 mbar; 2.9 PSI)/ 25 kPa (250 mbar; 3,63PSI) from the ventilator pneumatic.
38
Theory of Operation
2.15.2.7.1 Status of the diaphragm/membrane valves:
The diaphragm valves are pneumatically operated by solenoid valves. The following list shows the correspondence between diaphragm valve, solenoid valve and pressure port.
Valve Bellows valve Exp. valve Decoupling valve
controlled by MV 1 MV 2 MV 4 MV 3 connected to port # 5 1 3 6 wire # 5, 6 7, 8 1, 2 3, 4
In the following the status of the diaphragm valves are shown corresponding to the selected ventilation mode of the ventilator.
2.15.2.7.2 Machine OFF or Ventilator in Standby
Valve MV 1 MV 2 MV 3 MV 4
Activity OFF OFF OFF OFF Pressure on port 5, No 1, No 6, No no sup. of MV2
2.15.2.7.3 Manual / Spontaneous Mode, INSPIRATION And EXPIRATION
Valve MV 1 MV 2 MV 3 MV 4
Activity OFF OFF OFF OFF Pressure on port 5, No 1, No 6, No no sup. of MV 2
2.15.2.7.4 CMV Child / Adult Mode, INSPIRATION
Valve MV 1 MV 2 MV 3 MV 4
Activity ON ON ON ON Pressure on port 5, Yes 1, Yes 6, Yes supply of MV 2
2.15.2.7.5 CMV Child / Adult, EXPIRATION
Valve MV 1 MV 2 MV 3 MV 4
Activity OFF OFF OFF OFF Pressure on port 5, No 1, No 6, No no sup. of MV 2
2.15.2.7.6 CMV Adult, EXPIRATION And PEEP
Valve MV 1 MV2 MV3 MV4
Activity OFF ON OFF ON Pressure on port 5, No 1, Yes 6, No supply of MV2
39
Theory of Operation
2.15.2.7.7 Compliance Test Patient Module
The compliance test patient module is divided in 3 phases. Phase 1. Flow into the re-breathing system and test of flow sensors.
Valve MV 1 MV2 MV3 MV4
Activity ON OFF ON ON Pressure on port 5, Yes 1, No 6, Yes no sup. of MV2
Phase 2. Pressure increase up to test pressure 40 cmH2O, calculation of system compliance.
Valve MV 1 MV2 MV3 MV4
Activity ON ON ON ON Pressure on port 5, Yes 1, Yes 6, Yes supply of MV2
Phase 3. Test of pressure loss.
Valve MV 1 MV2 MV3 MV4
Activity ON ON ON ON Pressure on port 5, Yes 1, Yes 6, Yes supply of MV2
2.15.2.7.8 Leak Test Patient Module and Fresh Gas Module
Valve MV 1 MV2 MV3 MV4
Activity ON OFF OFF OFF Pressure on port 5, Yes 1, No 6, No no sup. of MV2
40
Repair Information
3 Repair Information
3.1 Introduction
This chapter of the service manual provides the necessary technical information to perform repairs to the instrument. The most important perquisites for effective troubleshooting are a trough understanding of the instruments functions, as well as understanding the principals of operation.
3.2 Warnings and precautions
In the event the instrument covers are removed, observe the following warnings and guidelines
3.2.1 Precautions
1. Do not short components leads together
2. The instrument covers must not be removed by anyone other than qualified technical personnel
who have received supplementary instructions regarding maintenance of medical equipment and/or have equivalent experience in this area.
3.2.2 Warnings
1. This device operates using compressed gas at high pressures. When attaching emergency gas
tanks, always open tank valves slowly, watching the cylinder gauge indicate the tank pressure. When disconnecting the tanks, always close the valves slowly. Use the Narkomat+ flow meters to bleed down the pressure, watching the cylinder gauge indicating the depleting tank pressure, before disconnecting the tank from the yoke. Always open and close tank valves fully.
2. This device operates using compressed gas at high pressures from the hospital central supply.
When connecting gas supply lines, attach the hose connection to the machine before connecting the quick disconnect fitting to the hospital source. Disconnect the supply hose from the hospital source connection prior to disconnecting it from the Narkomat+ gas connection fittings.
3. Whenever flowing anesthetic gases, nitrous oxide, oxygen, or any hospital gas always use the
appropriate agent evacuation system.
4. Never oil or grease any oxygen equipment unless the lubricant used is made and approved for
this kind of service. In general, oils and greases oxidize readily, and in the presence of oxygen burn violently.
41
Repair Information
3.3 Troubleshooting Guidelines
1. Identify the problem – Due to the wide variety of potential symptoms, certain problems may be
more subtle than others. Following the guidelines of the tests will help determine the problem if one exists.
2. Avoid shorting component leads together – During repair procedures, it can be tempting to
make a quick series of measurements. Always turn the power off before connecting and disconnecting the test leads and probes. The accidental shorting of leads can easily stress the components and cause a second failure (aside from the safety risk).
3. Use the proper equipment – The equipment listed below is suggested to fulfill a wide range of
troubleshooting requirements. It is imperative to use the designated equipment in order to ensure proper results of any and all test procedures.
4. Clean up the repair area – After any repair, clean off the repair area.
3.4 Troubleshooting Charts
3.4.1 Error Messages during ventilation
Display message Cause Corrective action User
Vent Error: Use Manual Ventilation Call Service
CPU Error: Use Manual Ventilation Call Service
Set APL valve to CMV/SP position
Flow Error: Use Manual Ventilation Call Service
Valve Error: Use Manual Ventilation Call Service
No communication between AVM 3-1 and AVM 3-2 (ventilator microprocessor boards) for more than 8 sec. No communication between AVM 3-1 and module 1 (user interface) for more than 20 sec.
System Failure (Module 1) Watchdog Error (Module
2) AVM3-2
PPEAK is greater than 10 mbar and lowest expiration pressure after the expiration still greater than PPEAK–2mbar No pressure relief in the expiration CMV, PCV After the first five breaths in CHILD resp. after the first three breaths ADULT: Last inspiratory flow is lower than 0,05 L/min, even though more than 2.00 L/min were set at the prop. valve Sensor does not deliver any values AVM3-2 CMV The internal actual flow value is measured during the inspiration and may not exceed the following desired value x factor
1.5: AVM3-2 PCV
The same as CMV, but other tolerance
Prop. Valve faulty
Finish the case with manual ventilation; Take the machine out of use; Call Service Finish the case with manual ventilation; Take the machine out of use; Call Service
Set APL to the CMV/SP position Reset automatically after pressure relief, Restart of the ventilation after request.
Finish the case with manual ventilation; Take the machine out of use; Call Service
The intermitting tone is reset via the mute button (acknowledgement) Reset the error message via the compliance test
Corrective action
Service technician
Exchange module 2 (ventilator microprocessor boards)
Exchange either the module 2, the module 1 or the communication cable connecting the two modules
If resetting the APL valve does not solve the problem, check the correct functions of the APL valve, solenoid valves and the diaphragm valves
Recalibrate the proportional valve, internal flow sensor and the characteristic of the proportional valve. If that does not solve the problem, exchange the
1. proportional valve
2. internal flow sensor
Recalibrate the proportional valve, internal flow sensor and the characteristic of the proportional valve. If that does not solve the problem, exchange the
1. proportional valve
2. internal flow sensor
42
Repair Information
Display message Cause Corrective action User
Valve Error: Use Manual Ventilation Call Service
Calibrate Breathing System Perform Compliance Test when convenient Pressure Reading out of tolerance Perform Compliance Test when convenient No drive gas; please check
Flow/Volume Readings not available Replace Flow Sensor – Call Service A P N E A
Breathing Circuit Disconnect
PEEP greater than Pmin
FiO2 lower than FiO2 min
Tidal Volume lower than Vt min
Peak pressure greater than alarm limit
Peak pressure below alarm limit
Minute Volume below alarm limit
PEEP greater than PEEP-Setting
FiO2 greater than FiO2 alarm limit
Check Vent Dial position
AVM3-2 CMV/ PCV The actual internal flow value
should be 0 during expiration
AVM3-2 CMV, PCV Difference between the set and the detected state of each solenoid valve for more than 200 ms AVM3-1, AVM3-2 A difference between the two pressure sensors of more than 10 mbar
AVM3-1 Monitoring of the pressure switch for the driving gas
AVM3-1 during ventilation External flow sensor is broken
or short-circuited AVM3-1. Manual/ spontaneous
Within a period of 15 to 60 seconds a expiration cannot be detected AVM3-1 CMV, (S)CMV, PCV The peak pressure PPEAK is lower than 0.5 mbar and lower then the set pressure limit Pmin AVM3-1 CMV, (S)CMV, PCV The end expiratory pressure is greater than the set pressure limit Pmin AVM3-1 The measured O2 value is lower than the set O2min limit AVM3-1
The measured lower than the set
AVM3-1 The measured Ppeak value is greater than or equal to the set Pmax limit AVM3-1 The measured Ppeak value is lower than the set Pmin limit AVM3-1 The calculated M.Vol value is lower than the set M.Vol min limit AVM3-1 The measured PEEP value is 5 mbar higher than the set PEEP AVM3-1 The measured FiO2 value is greater than the set FiO2max limit AVM3-1 The selector switch position is monitored An invalid position must be eliminated within two seconds. After this the alarm message appears.
Vt
Vt min
value is
Finish the case with manual ventilation; Take the machine out of use; Call Service
The intermitting tone is reset via the mute button (acknowledgement) Reset the error message via the compliance test The intermitting tone is reset via the mute button (acknowledgement) Reset the error message via the compliance test Ensure the driving gas supply, restart the ventilation on request.
The intermitting tone is reset via the mute button (acknowledgement) Reset the error message via the compliance test Create expiration Ensure the correct function of
Create a pressure increase in the inspiration phase
Lower the end expiratory pressure or raise the pressure limit Pmin
Increase the O2 value or raise the limit O2min
Increase the Vt value or raise the limit
Vt min
limit
Decrease the Ppeak value or raise the limit Pmax
Increase the Ppeak value or lower the limit Pmin
Increase the ventilation rate or the Vt value or lower the limit Pmin
Decrease the PEEP value or raise the set PEEP
Decrease the FiO2 value or raise the set FiO2max limit
The last detected valid position is maintained until a new valid position is detected, then the message disappears
Corrective action
Service technician
Recalibrate the proportional valve, internal flow sensor and the characteristic of the proportional valve. If that does not solve the problem, exchange the
1. proportional valve
2. internal flow sensor Check the correct voltage supply for the solenoid vales. If that does not solve the problem exchange the according solenoid valve Recalibrate the pressure sensors and check the according tubing. If that does not solve the problem exchange the module 2 Check the gas supply, the correct tubing and the correct connection and function of the pressure switch Exchange the external flow sensor
the circle system
Ensure the correct function of the circle system and the correct electrical and pneumatic function of the solenoid valves Check the correct position and function of the APL valve. Ensure the correct function of the circle system Recalibrate and check the correct function of the O2 cell. Exchange if necessary
Check the correct position and function of the APL valve
Re-calibrate the O2 sensor
1. Check the correct function of the Vent Dial
43
Repair Information
Display message Cause Corrective action User
System Error Cal Required Call Service
Temp Sensor readings not available Call Service
Temp Sensor out of tolerance Check Heating System ­Call Service
Fan Error Check Fan - Call Service
AC Power lost, using Battery
Continuous Pressure
Battery running low Use Manual Ventilation
Ambient Air Intake: Check Fresh Gas setting
Set APL to CMV/SP position
Check Settings
30 min. remaining
25 min. remaining
20 min. remaining
15 min. remaining
EEPROM is checked during the start up and during the sensor test for CRC errors
AVM3-1 The measured temperature value is out of tolerance
AVM3-1 The over-temperature switch activated or the temperature of the patient module has not increased for 20 min even though heater is switched on AVM3-1 The Fan does not deliver any impulses AVM3-1 The line power fail signal reports use of battery
AVM3-1 Manual/ spontaneous Over a time of 10 seconds the minimum and the maximum pressure is measured. If the actual pressure is greater than 10 mbar and the difference between the minimum and the maximum pressure is smaller than 5 mbar, the error message is issued
AVM3-1 Battery mode The valid discharge time is over
AVM3-1 CMV, PCV In the expiration phase there is a sub-atmospheric of more than 2 mbar, an error message is issued
AVM3-1 Controlled ventilation
The measured PEEP value is 5 mbar higher than the set PEEP over a period of 5 ventilation cycles
The controlled ventilation is being started AVM3-2 CMV The maximum position of the Proportional valve has been reached AVM3-1 Battery mode 30 minutes of battery use left
AVM3-1 Battery mode 25 minutes of battery use left
AVM3-1 Battery mode 20 minutes of battery use left
AVM3-1 Battery mode 15 minutes of battery use left
Take the machine out of use; Call Service
Restore the function of the temperature sensor
Restore the function of the heating blanket
Restore the function of the Fan 1. Check for physical
Restore the line supply 1. Ensure the correct line
Increase the pressure difference
Restore the line supply 1. Charge the battery
Correct the fresh gas flow. When the smallest Pressure zero higher is, the error message disappears
Set APL to VENT (CMV) position
Decrease ventilation frequency or Vt
Restore the line supply 1. Charge the battery
Restore the line supply 1. Charge the battery
Restore the line supply 1. Charge the battery
Restore the line supply 1. Charge the battery
Corrective action
Service technician
Recalibrate the complete system. If that does not solve the problem exchange the module 2 Recalibrate the temperature sensor in the service mode. If that does not solve the problem exchange the module 2 Recalibrate the temperature sensor in the service mode. Check the correct function of the solid state relay and the according power supply
obstructions
2. Exchange the fan voltage
2. Ensure the correct function of the power supply module
2. Check the capacity of the battery, if necessary, exchange it. Check the correct calibration of the pressure sensors
1. Ensure the correct function of the APL valve.
2. check the breathing circuit for obstructions
2. Check the capacity of the battery, if necessary, exchange it.
2. Check the capacity of the battery, if necessary, exchange it.
2. Check the capacity of the battery, if necessary, exchange it.
2. Check the capacity of the battery, if necessary, exchange it.
44
Repair Information
Display message Cause Corrective action User
10 min. remaining
5 min. remaining
Unable to attain target pressure, Adjust flow or I:E ratio PCV Setting not valid
Expiratory time too short
Resume Ventilation
Compliance Test bypassed
CO2 Re-breathing; please check Absorber
Gas Monitor Technical Failure
AVM3-1 Battery mode 10 minutes of battery use left
AVM3-1 Battery mode 5 minutes of battery use left
AVM3-2 PCV The peak pressure PPEAK is lower than the set volume
AVM3-2 PCV Bellow monitor active. additional volume delivery is not possible AVM3-2 PCV The calculated expiratory time is greater than 1 second (Adult) resp. 0.5 second (Child) AVM3-1
Message to restart the ventilation The system is started without having performed the compliance test
The inspiratory CO2 value is too high
Internal fault in the gas module
Restore the line supply 1. Charge the battery
Restore the line supply 1. Charge the battery
Increase driving gas flow or lower the plateau
Reduce ventilation frequency or increase “I”
Reduce ventilation frequency or increase “E”
Restart the ventilation
Perform the compliance test
Exchange the soda lime. If all else fails call Service. Call Service
Sampling Line Occlusion
Due to an improper tube connection, the exact value can not be measured
Check if the sample tube is kinked or obstructed. If else fails exchange the sample tube
No Patient detected; no Breathing
The CO2 measurement cannot detect any breathing or respiration activities
Check if the sample tube is connected and check the conditions of the patient
Water trap is full, will be full soon or connection faulty
Gas Monitor NOT ready
The water trap is full, nearly full or the water trap is not connected correctly Power supply or data communication to the
Check if the water trap is connected correctly or, if full, exchange it
Call Service
gas module is defective
Corrective action
Service technician
2. Check the capacity of the battery, if necessary, exchange it.
2. Check the capacity of the battery, if necessary, exchange it. Check the system for leakages
Verify the gas module calibration. If else fails, exchange the gas module
Check the internal cable connections. If else fails, exchange the gas module
Calibrate the sample flow rate in the service mode
Check and if necessary recalibrate the sample flow rate in the service mode
Check and if necessary recalibrate the sample flow rate in the service mode
Check for correct power supply and ensure correct cable connections. If else fails, exchange the gas module
45
Repair Information
3.4.2 Alarm messages during the compliance test, leak test and O2 calibration
Display message Cause Corrective action User
System Resistance too high
Leak rate too high
Compliance out of range
System Error Cal Required Call Service
Check Diaphragm Valves
Pressure Reading out of tolerance Perform Compliance Test when convenient
Flow Error: Use Manual Ventilation Call Service
Flow/Volume Readings not available Replace Flow Sensor ­Call Service
System Error Cal Required - Call Service
Calibrate Breathing System Perform Compliance Test
AVM 3-2
If with in 1 sec. after a flow of 12 l/min has been generated, a pressure increase of 10 mbar is measured AVM 3-2 If after 10 mbar have been reached, 40 mbar are not reached within 2s AVM 3-2 The measured compliance has to be between 2.0 and 9.9
AVM 3-2 Data for parameter or alarm limit settings was not saved correctly, the data exchange between the ventilator modules and on-screen display is faulty or the startup test for Internal circuit EEPROM has failed.
AVM 3-2 The PV closes after 40mbar has been reached. If after this a pressure increase of 4 mbar takes place within 4 s
If the pressure sensors measure a difference greater than 3 mbar during the offset comparison, even though the system is pressure relieved.
Compliance Test
When during the offset calibration a flow of greater than 3 L/min is detected
During the compliance test a flow of 12 L/min is delivered through the prop. valve. If the delivered flow is out of tolerance
Flow Sensor Error AVM3-2 Pressure test
Cannot be set to zero via the analogue output
Set DAC value greater than 1,83V
The amplifier circuit is not working correctly
EEPROM is checked during the start up and during the sensor test for CRC errors
The checking of the solenoid valves failed
Repeat the compliance test Check the breathing circuit for
Troubleshoot the breathing circuit for leakages
Replace bacteria filter and breathing circuit. Use the Compliance test in Standby / OPTIONS to retest. If all else fails call Service. Retry function. Reboot machine. If all else fails call Service.
Remove the Breathing circuit from the Docking Station and check the Decoupling and Expiratory valves for intact membranes. Replace as required or call Service. Take the machine out of use; Call Service
Take the machine out of use; Call Service
Retry function. Reboot machine. If else fails call service
Take the machine out of use; Call Service
Retry function. Reboot machine. If else fails call service
when convenient
Patient module unlocked
The system does not detect a signal from the stop switch in the docking station
Insert Breathing system, else call service
Corrective action
Service technician
obstructions
Troubleshoot the breathing circuit for leakages
Replace bacteria filter and breathing circuit. Check the complete breathing system for obstructions.
Recalibrate the system. If this fails, exchange the module 2 processor board
Remove the Breathing circuit from the Docking Station and check the Decoupling and Expiratory valves for intact membranes. Replace as required Recalibrate the pressure sensors and check the according tubing. If that does not solve the problem exchange the module 2
Recalibrate the proportional valve, internal flow sensor and the characteristic of the proportional valve. If that does not solve the problem, exchange the
1. proportional valve
2. internal flow sensor
Exchange the external flow sensor
Recalibrate the complete system. If that does not solve the problem exchange the module 2 Check the correct voltage supply for the solenoid vales. If that does not solve the problem exchange the according solenoid valve Test the correct function of the stop switch in the service mode. If the function ids OK, readjust it’s positioning. If the function is faulty, exchange the stop switch
46
Repair Information
Display message Cause Corrective action User
System not pressure less
System vented or drive gas missing
AVM 3-2 A pressure larger than 10 mbar is measured before the start of the compliance test AVM 3-2 During the compliance test or Leak test a pressure higher than 10 mbar is not measured
Vent the system by opening the Breathing circuit. Monitor
system pressure. Check the O2 cell is present,
APL valve is set to max (leak test) and breathing circuit is connected properly.
3.4.2.1 Messages during the system tests
Result Compliance Test
Start Up
OK Compliance Test
Passed Leak rate < 300 ml/min Press Enter to
continue OK with a leak of 300ml to 600ml
Compliance Test
Complete
Leak rate is
between 300-600
ml/min
Press Enter to
continue OK with a leak of
Leak Test 500ml to 1000ml
OK with a leak of >600ml/min
Compliance Test
Complete
Leak rate is higher
than 600 ml/min
Tighten valve rings
Check breathing
circuit
Press Enter to
continue OK with a leak of
Leak Test >1000ml/min
Compliance Test Leak Test O2 calibration
Test Passed Leak Test Passed
Set APL Valve to VENT (CMV) position
Test Complete Leak rate is xxx ml/min Compliance is x.x ml/cmH2O Please check breathing circuit
Complete Leak rate is xxx ml/min Verify that APL Valve is set to Max CO2 absorber,
vaporizers locked Test Complete Leak rate is higher than 600ml/min Please check for leaks
Complete
Leak rate is higher
than 1000 ml/min
Verify that APL
Valve
is set to Max CO2
absorber,
vaporizers locked
Corrective action
Service technician
Relieve the system pressure, recalibrate the pressure sensors
Troubleshoot the breathing circuit for leakages
47
Repair Information
Result Compliance Test
Start Up
OK O2Sensor
O2 too high O2 concentration
O2 Sensor too old O2 Sensor is out of
Compliance Test Leak Test O2 calibration
calibration successful
too high Expose Sensor to room air Press Enter to Start
rage Replace O2 Sensor Press Enter to Start
48
Repair Information
3.4.2.2 Symptoms of Fuse Failures
The following table shows the reaction to a failed fuse on the power supply board. The fuses F2, F3, F4, F5 and F6 have been removed from the power supply module and are therefore subject of the production date.
FUSE Failure during working Failure during start
F1
F7
F8/F9
F10/F11
Display is black; module 1 and 2 do not work
alarm „Solenoid Valves faulty“, „Fan faulty“ and „Ambient Air Intake“; real time curve shows a zero line
alarm „AC power lost, using battery“ after the machine has booted alarm
During the ventilation there is no failure; after you have turned the selector switch to Standby the message „Insert patient module“ is shown; during using battery the message disappeared
Display is black; module 1 and 2 do not boot
after booting the message „Selected mode is not Standby“ is shown; module 2 does not boot; fan does not work
„AC power lost, using battery“ is shown after the machine has booted the message„Insert patient module“ is shown; during using battery the message disappeared
49
Repair Information
3.5 Required Tools
Item Number
1 - Operation manual 2 - Service manual 3 900-4726 Calibration pump, Pressure sensors 0 – 80 cmH2O 4 - Safety analyzer For electrical tests according to IEC-
5 - Digital Volt Meter 6 - Calculator Standard four function 7 900-4715 Temperature simulator 8 - Mass flow meter 0 – 80 l/min ± 2% of reading 9 - Surface cleaner Standard hospital grade 10 340-3000 Gasket set 11 460-0610 Driving gas test adapter with 1 meter
12 603-4000 Test lung Adult size 13 - Test lung Neonate size 14 - High pressure gauge 0 – 2.5 kPa x 100(0 – 2.5 bar;
15 - Low pressure gauge 0 – 250 Pa x 100(0 – 250 mbar; 16 - Hand tools, calibration tool, flat screw 17 800-6009 Pressure regulator for verification gas 18 800-6004 Verification gas 50% O2, 43% N2O, 5.5% CO2, 1.5% Hal
19
Part
Number
Description Specification
HEYER specification HEYER specification
601-1,B
HEYER specification
HEYER specification HEYER specification
tubing
0 – 38 PSI) 0 – 3.8 PSI)
Metric standard
driver, socket and Allen wrench set cylinder
3.6 Disassembly instructions
3.6.1 Connecting and disconnecting vaporizers
1. To connect the vaporizer, place it carefully onto the vaporizer manifold. The manifold is Selectatec type.
2. Lock the vaporizer with the locking lever ¼ turn to the fully locked position.
3. Only vaporizers with interlock-system may be used with this system.
4. After each exchange of a vaporizer, perform a leak test.
3.6.2 Removing the compressed gas tanks (PIN-Index)
1. Ensure that the tank is closed
2. Rotate the hand tight knob that retains the tank against the port seat.
3. Flip the swiveling yoke collar 180 degrees.
4. Grip the tank firmly and remove it from the yoke.
5. Connect a new tank in reverse order, ensuring a single tank washer is present on the yoke port seat.
50
Repair Information
3.6.3 Removing the Patient Module
1. Remove the CO2 absorber dome.
2. Remove the bag in bottle system.
3. Remove the O2 cell cable from the O2 cell, if present.
4. Remove the scavenger hose.
5. Remove the breathing circuit.
6. Enter the OPTIONS window in the operating STANDBY mode. AC power and drive gas are required.
7. Select UNLOCK and remove the Patient Module, or
8. Enter the Misc screen in the TEST menu of the service Program Main menu. AC power is required.
9. Select UNLOCK and remove the patient module.
3.6.4 Removing the CO2 Absorber Canister
1. To release the absorber canister from the patient module, turn the unit counter clockwise. Secure the Canister that it does not fall to the ground when the thread has ended.
Note: The discoloration of the spent soda lime disappears again after some time if it has not
been used. If used again, the color discoloration returns. Dry soda lime becomes malsorbant.
3.6.5 Removing the Bellows and Dome System
1. Turn the unit counter clockwise and unscrew the unit ¼ turn.
2. Once released, remove the silicon bellows by pulling it down form the Patient Module fitting.
3.6.6 Removing the Airway Pressure Limiting Valve (APL)
1. Unscrew the union nut. The top section can now be removed.
2. The membrane can be removed from the bottom section and replaced, if necessary.
3. The membrane is placed back into the bottom section with the metal facing upwards.
4. Replace the top part of the Airway Pressure Limiting valve by lining up the index line in its correct orientation.
5. Secure with union nut.
Caution: The spring in the top of the APL valve may not be stressed. After removal from
the bottom section place the top section to one side, taking care that the spring is not unduly loaded.
51
Repair Information
3.6.7 Inspecting/replacing the Decoupling, Bellow and Expiratory Valves
1. To replace the Patient modules rear side valves’ membrane, undo the coupling nut.
2. After removing the valve assembly, undo the black nut securing the valve.
3. Inspect and /or replace the valve, ensuring no hole or deformity that would cause a leak.
4. Assemble in reverse order
CAUTION: During transportation of the patient module, transportation protection should be
applied at the rear to protect the valves.
3.6.8 Disassembling the Room Air, Inspiration and expiration valves
1. To dismantle the valves, the coupling ring must be unscrewed from the valve body. The valve cover can now be removed.
2. The O-ring and the metal baskets in the expiration or inspiration valve can now be removed. The valve plate can be removed. Assemble in reverse order.
3. The membrane of the room air valve can be carefully removed from the valve body after removing the coupling ring and the valve cover. Assemble in reverse order.
4. The O-rings of the valve bodies have only to be exchanged if the valves were pulled out of the Patient Module’s top.
3.6.9 Removing the Gas Block Module
1. Disconnect all sources of pressurized gas.
2. Remove the screws holding the rear access panel.
3. Remove the internal access panel containing the Gas Block.
4. Remove the screws and nuts holding the gas block to the enclosure.
5. Disconnect the gas lines running to the various connections.
6. Remove the Gas Block as a single module.
7. Reinstall the Gas Block in reverse order. Apply gas pressure and test prior to final button-up.
WARNING: An error in the tubing can lead to a serious health damage or death of the
patient. Reinsure the correct connection of the gas lines before the unit is put back to use.
3.6.10 Removing the flow tubes
1. Remove the Gas Block Module
2. Remove the spring-loaded flow tubes by pushing it up and removing from the bottom.
3. Reinstall in reverse order.
52
Repair Information
3.6.11 Removing the pressure gauges
1. Remove the Gas Block Module
2. Remove the metal compression fittings from the gauge.
3. Remove the gauges from the front panel enclosure.
4. Reinstall in reverse order
5. Connect gas supplies at each gauge. Observe and test compression fittings for leaks.
3.6.12 Removing the Module 2 Circuit Board Set
1. Disconnect AC power.
2. Remove the cables connecting the Module 2 circuit boards to the rest of the Narkomat+.
3. Remove the two nuts holding the Module to the housing.
4. Remove the Module 2 circuit as a module.
5. Reinstall in the reverse order.
3.6.13 Removing Module1 Circuit Board Set
1. Remove the display cover and open module 1 housing by unscrewing the screws on the top.
2. Remove the cables connecting the Module 1 circuit board to the rest of the Narkomat+.
3. Remove the screws holding the Module 1 to the button of the housing.
4. Remove the Module 1 circuit boards as a module
5. Reinstall in the reverse order.
3.6.14 Removing the Power Supply Module
1. Disconnect AC power
2. Remove the housing cover of the power supply module.
3. Remove cables connecting the power supply module to the rest of the Narkomat+.
4. Remove the screws holding the board/module to the access panel.
5. Remove the power supply as a module.
6. Reinstall in reverse order.
53
Repair Information
3.6.15 Removing Internal Regulators, Proportional Valve and Flow Divider
1. Disconnect all sources of pressurized gas.
2. Remove vaporizers.
3. Remove the screws of pneumatic drawer and pull it out.
4. Disconnect each gas hose and electric cable running through each section.
5. Remove the drawer exposing the internal regulators and proportional valve. Replace the targeted component. Replace each hose and cable.
6. Reconnect hoses and cables to the new component.
7. Rebuild in reverse order.
8. Apply gas pressure and power to functional test prior to final button-up
54
Maintenance and Calibration
4 Maintenance and Calibration
4.1 Introduction
The Narkomat+ maintenance and calibration is comprised of four distinct activities:
1. Inspecting and/or replacing consumable parts
2. Adjusting individual potentiometers
3. Functional testing
4. Programming coefficients into stored memory
Consult the Preventive Maintenance chapter of this manual to determine the appropriate interval for inspection and replacement of consumables, mechanical and electronic adjustments, testing and programming. Ensure that all testing materials, including gas drive, breathing circuits, test fixtures, tools and documents are available and in good working order prior to beginning test procedures. Testing and programming requires utilizing internal service software. A password is provided by
HEYER.
4.2 Calibration Warnings and Precautions
WARNING: The Narkomat+ Anesthesia Machine operates on line voltage and at high
pressure. Therefore, an electric shock hazard may exist when the backside panel covering the power supply boards is removed. Repair and calibration procedures should only be preformed by qualified personnel who follow proper servicing techniques. Warnings are given in appropriate location.
WARNING: Possible fire hazard. Fuses (i.e., additional sockets) must only be replaced by
fuses of the same type with the same rating.
WARNING: Possible electrical shock hazard. The machine may only be opened by HEYER
authorized service personnel.
CAUTION: Refer to the maintenance intervals in the preventive maintenance section for
guidance on which steps are preformed when.
CAUTION: Use surgical gloves whenever touching or disassembling valves or other
components of the patient module.
CAUTION: Ensure that the gas supply of the machine always complies with the technical
specification.
CAUTION: If the machine should show faults during the initial calibration or testing, the
machine should not be operated until the fault has been repaired by a qualified service technician.
CAUTION: After servicing, functional, compliance and leak tests must be carried out before
clinical use
CAUTION: Failure to connect device to a grounded mains out may elevate leakage current
in excess of permissible values.
CAUTION: During transportation of the patient module, transportation protection should be
applied to the rear to protect the diaphragm valves
55
Maintenance and Calibration
CAUTION: The spring in the top of the APL valve may not be stressed. After removal,
ensure that the APL valve is in position “CMV” and place to one side, taking care that the spring is not unduly loaded.
CAUTION: Use cleaning agent sparingly. Excess fluid could enter the machine, causing
damage.
4.3 Test Procedure
4.3.1 General
1 Ensure that the Service Manual and Operation Manual are present. 2 Ensure there is no external damage to machine and accessories by visual inspection. 3 Ensure all accessories are complete and functional by visual inspection.
4.3.2 Inspecting/replacing consumable parts
1 Remove the patient module from the Narkomat+
a Turn the lever below the docking station counter clockwise to unlock the patient module.. b Pull out the patient module slightly and verify that the patient module stops. c Remove the patient module. d Verify that the diaphragm valve connectors are not damaged..
56
Maintenance and Calibration
1
valve inlet 40° turned drawn
reservoir /
bag port
insp. port
7
2
3 4
5
6
exp. port
8
9
Fig. 16 Patient module, front view
57
expiratory diaphragm valve
16
Maintenance and Calibration
bellows control diaphragm valve
decoupling diaphragm valve
15
14
13
12
11
10
Fig. 17 Patient module, top view (diaphragm version)
Pos. No. Description Order No. 1 Airway pressure relief valve, cpl. 323-0095 2 Emergency air valve, cpl. 323-0098 3 Inspiratory valve, cpl. 323-0096 4 Expiratory valve, cpl. 323-0097 5 Heat blanket, cpl 323-0314 6 Flow sensor, cpl. 323-0220 7 Ventilation Bellow 323-0158 8 Absorber vessel 323-0046 9 Bellow dome 323-0045 10 Fixing nut diaphragm valve 323-0294 11 Bessel diaphragm valve 323-0110 12 Ring plate diaphragm valve 323-0042 13 O-ring 28.3x1.78 323-0292 14 Gasket 13x6.5x2 980-1170 15 Diaphragm valve body 323-0100
58
Maintenance and Calibration
2 Ensure that the O-ring (323-0147) and the packing ring (980-1170) of the CO2 absorber are intact. 3 Ensure that the O-ring (323-0147) on the ventilator bellow is intact. 4 Ensure that the O-ring (049-3182) on each vaporizer mount is intact. 5 Check O-rings (049-3052) of the docking station ports (located between the patient module and its
docking station). a Replace O-rings at recommended interval
6 Inspect O-ring washer (980-1174 for O2 and N2O) between tank and yoke connection. Replace if
necessary.
7 Ensure the inspiratory valve (610-3156) is intact, replace if necessary. 8 Ensure the expiratory valve (610-3156) is intact, replace if necessary. 9 Ensure the APL valve (323-0310) is intact, replace if necessary. 10 Ensure the Room air valve (323-0099) is intact, replace if necessary. 11 Ensure that the inspiratory and expiratory valve O-rings (610-3157) are intact, replace if
necessary.
12 Ensure the Fresh gas decoupling, Driving gas and expiratory diaphragm valves (323-0100) are
intact, replace if necessary.
13 Check gas supply inlet filters (370-0017) and water traps (340-0344).
a Verify that minimal to no condensation or dirt are present in the inlet filter and bottles. Drain
any liquid from bottles.
b Clean and/or replace the contaminated inlet filters and O-rings, if necessary.
Fig. 18 Supply inlet filters and water traps
14 Reassemble the patient module and install back into the docking station.
59
Maintenance and Calibration
15 Turn off the Narkomat+ power switch. 16 Verify O2 and N2O tank check valves. *
a Connect and open a full gas tank on each yoke b Withdraw the gas connecting coupling from the supply outlet. c Verify the check valves prevent gas from escaping through the open supply outlets. d Verify there is no external damage to the gas connecting coupling
17 Verify O2, N2O and AIR supply check valves.
a Remove the tanks and reconnect the gas connecting couplings to the supply outlets. b Verify the check valves prevent gas from escaping through the open tanks yokes. c Verify there is no external damage to the gas tank or yoke pins.
* Subject of configuration
60
Maintenance and Calibration
4.3.3 Power supply checks
1 Remove the screws from the lower panel and keep the panel at it’s position. Carefully remove the
panel.
Warning: AC voltage is present 2 Adjust P1 on the power supply board for +5.2 VDC. Use the ground pin located to the right as
meter ground.
3 Adjust P2 on the power supply board for +12.5 VDC. Use the ground pin located to the right as
meter ground.
4 Verify that the V54 LED is illuminated, indicating -12 VDC. 5 Reinstall the panel.
61
Maintenance and Calibration
4.3.4 Functional Tests
4.3.4.1 Pneumatic tests
1 Perform automatic compliance test.
a Turn on the Narkomat+ and ensure adequate drive gas is present b Attach a breathing circuit the patient module. c Seal the Y-piece by attaching the open port to the parking place on the side of the patient
module. d Turn off the fresh gas flow. e Select START to start the test. The test calibrates the ventilator for tubing compliance at 40
Pa x 100 (40 mbar; 40 cmH2O) and performs a leak test on the patient circuit pneumatics
between the decoupling valve and the expiratory valve. f Verify successful completion and that no leaks are indicated.
2 Perform automatic leak test.
a Turn on the Narkomat+ and ensure adequate drive gas is present. Enter the Leak Test via the
Options menu. b Attach a breathing circuit the patient module. c Connect the y-piece to the hose end where the reservoir bag normally is connected. d Turn off the fresh gas flow. e Set the APL valve to max (fully closed) f Select START to start the test. The test pressurizes the system the pneumatic system all the
way back to the spindle valve of the flow meters. g Verify successful completion and that a minimal leak (less than 300 ml/min at 40 cmH2O) is
indicated.
3 Perform the manual pressure test.
a With the selector switch of the Narkomat+ turned to the manual position, attach a breathing
circuit to the patient module. b Connect the y-piece to the hose end where the reservoir bag normally is connected. c Set the APL valve to max (fully closed). d Rotate the O2 or AIR spindle until 60 Pa x 100 (60 mbar; 60 cmH2O) is observed on the
screen. e Reduce the gas flow by counter rotating the spindle until the pressure stabilizes at 60 Pa x
100 (60 mbar; 60 cmH2O) f The gas flow indicates the Narkomat+’s leak rate. If the automatic Compliance and Leak test
have passed, this test will indicate a leak in the drive gas system, or may be used to
troubleshoot system leaks at a higher-pressure level.
62
Maintenance and Calibration
4 Verify APL Valve accuracy.
a Attach a breathing circuit the patient module. b Connect the y-piece to the hose end where the reservoir bag normally is connected. c Set a fresh gas flow of 5 l/min. d Set the APL valve to each pressure graduation (10, 20, 30, 40 and 50 Pa x 100 (mbar;
cmH2O)) e Compare the valves on the pressure gauge with the valve settings. Deviations from pressure
indication and setting must not exceed ±20% of the set value or 10 Pa x 100 (10 mbar; 10
cmH2O), whichever is greater.
4.3.4.2 Alarm Tests
1 Turn on the Narkomat+ and ensure adequate driving gas is present. 2 Perform the Compliance test. 3 Verify the Vt min (low alarm), Peak pressure (high and low alarms), M.Vol (low alarm) and O2 (high
and low alarms). a Ventilate a test lung in the CMV mode. b Set each parameter’s alarm limit one at a time, to violate it’s high and low alarm setting. c Verify that alarm indications function.
Fig. 19 Alarm screen
63
Maintenance and Calibration
4 Verify O2 pressure loss alarm (whistle) and N2O cutoff.
a Ventilate a test lung in the CMV mode. b Set the AIR/N2O switch to N2O c Flow 1 l/min O2 and 1 l/min N2O using the flow meter spindle valves. d Interrupt the O2 supply to the Narkomat+. e Verify the O2 pressure loss alarm whistle for approximately 7 seconds, and the flow of nitrous
oxide lowers to 0 l/min. If O2 is the ventilator drive gas, an electrical alarm and message will
also activate. f Verify at the same time, Air is available by adjusting the AIR spindle valve. AIR flow will be
available regardless of the position of the AIR/N2O switch position. Air flow can be increased
by rotating the Air flow spindle. g If necessary, adjust the time and loudness level of the O2 deficiency whistle by setting the
screw adjustment next to the Whistle reservoir.
Fig. 20 O2 Whistle adjustment (appearance may differ due to different configuration options) 5 Verify line voltage alarm.
a Interrupt AC line voltage while the respirator unit is in manual/spontaneous, or automatic
ventilation mode. b Verify that the line power alarm sounds and that the AC connector icon appears on the
display. The alarm is reset when reconnecting the line power supply. c Verify the battery icon appears on the screen after AC reconnection. This indicates battery
charging.
64
Maintenance and Calibration
6 Verify N2O:O2 ratio system.
a Set the O2 and N2O valves to minimum. b Rotate the N2O valve throughout its range from 0.5 to 10 l/min. c Using the readings from the flow tube, verify that no less than 25% ± 5% O2 can be achieved
at any N2O flow rate.
4.3.4.3 Electrical Tests
1 Check convenient AC outlets.
a Verify AC voltage is present at each AC outlet with the Narkomat+ Mains switch in the ON or
OFF position.
CAUTION: Perform the following electrical safety inspection as the last step after completing a
repair or after routine maintenance. 2 Conduct electrical safety inspection. Test of electric safety according to IEC 601-1: Test the protective conductor resistance. For the measurement, use a test set according to IEC 601-1. The maximum protective conductor resistance must not exceed 0.1 Ohms. Enter the result into the test protocol. Test the leakage current. For the measurement use a test set according to IEC 601-1. Before the measurement of the leakage current, withdraw the power supply cable of all units form the
convince receptacles at the rear side of the apparatus. The maximum leakage current must not exceed 500µA (0.5 mA). Enter the result into the test protocol. Test the insulation resistance. For the measurement, use a test set according to IEC 601-1. The insulation resistance must be higher than 70 MOhms.
a Enter the result into the test protocol.
65
Maintenance and Calibration
5 Service Software
How To Start The Service Software
Turn off the main switch. The ”INSP. PAUSE” and ”EXP. PAUSE” control panel keys must be pressed while turning on the main switch, the keys have to keep pressed 3 seconds longer than the LED’s are lit. The PC now starts the service software.
Fig. 1 Start Window
66
5.1 Identity Check
Maintenance and Calibration
Fig. 2 Identity Check
Enter a valid key-code to execute the service software.
Leave this window with ”Enter”.
67
5.2 Main Menu
Maintenance and Calibration
Fig. 3 Main Menu
Choose the next application by operating one of the buttons.
68
5.3 Set Date and Time
Maintenance and Calibration
Fig. 4 Set Date and Time
First check the date and time, if it is necessary, set the clock.
Change date or time by moving the display pointer to the corresponding button, change the value
by turning the encoder knob to the right or to the left. Press the encoder knob to leave the display button.
When clock is set, save the information and turn back to the main menu by validating the “Cancel”
button. If the change should not be saved, quit the application to turn back to main menu.
69
5.4 Choose Service
Maintenance and Calibration
Fig. 5 Choose Service
Press the ”Start” button at the main menu to open the calibration routines window.
70
Maintenance and Calibration
5.5 Calibration Routines
Fig. 6 Calibration Routines
Choose from the service menu the sensor to calibrate. In case of a complete calibration it’s
recommended to perform the calibration step by step in order of the menu.
To exit the service software operate the “Quit” button of this window. Two acoustic signals are
indicating the correct end of the service software and a new boot procedure of module 2.
Caution: Shut down the complete unit each time you have finished the service software. Warning: Perform a Sensor Test and the System Tests after you have finished the service software.
71
Maintenance and Calibration
5.5.1 O2 Sensor ADC Calibration*
Fig. 7 O2 Sensor ADC Calibration 1
Place oxygen sensor into the test adapter in ambient air.
Turn offset potentiometer screw O2 OFFS (P5, AVM3-2) until the pointer at the screen shows 20
±
Operate the ”MIN” button.
*Not valid with the gas measuring unit
1
72
O2 Sensor ADC Calibration , cont.*
Maintenance and Calibration
Fig. 8 O2 Sensor ADC Calibration 2
Put the oxygen sensor into 100% O2.
Open the Y-piece and apply an oxygen flow of 5 l/min for some minutes.
Turn gain potentiometer screw O2 GAIN (P6, AVM3-2) until the pointer at the screen shows 80
1
±
.
Operate the ”MAX” button.
Check both of the values alternately for correctness and readjust if necessary.
When both of the calibration points are correct within a tolerance range +/-1.0, operate the ”Save” button.
By operating ”Quit” turn back to calibration menu.
*Not valid with the gas measuring unit
73
Maintenance and Calibration
5.5.2 Pressure Sensor #1 ADC Calibration
Fig. 9 Pressure Sensor #1 ADC Calibration 1
Connect a pressure gauge and a pressure applicator to the pressure gauging line.
Apply ambient pressure.
Turn offset potentiometer screw P1 OFFS (P11, AVM3-2) until screen pointer shows 20.
Operate the ”MIN” button.
74
Maintenance and Calibration
Pressure Sensor #1 ADC Calibration , cont.
Fig. 10 Pressure Sensor #1 ADC Calibration 2
Apply a pressure of 80 mbar to the gauging tube.
Turn gain potentiometer screw P1 GAIN (P10, AVM3-2) until screen pointer shows 80 and operate
”MAX” button.
Check both of the values alternately for correctness and readjust if necessary.
When both of the calibration points are correct within a tolerance range +/-0.2, operate the ”Save” button.
By operating ”Quit” turn back to calibration menu.
75
Maintenance and Calibration
5.5.3 Pressure Sensor #2 ADC Calibration
Fig. 11 Pressure Sensor #2 ADC Calibration 1
Perform the same calibration procedure like in case of pressure sensor P1.
Apply ambient pressure.
Turn offset potentiometer screw P2 OFFS (P8, AVM3-2) until screen pointer shows 20 and
operate the ”MIN” button.
76
Maintenance and Calibration
Pressure Sensor #2 ADC Calibration , cont.
Fig. 12 Pressure Sensor #2 ADC Calibration 2
Apply a pressure of 80 mbar to the gauging tube.
Turn gain potentiometer screw P2 GAIN (P7, AVM3-2) for gain calibration until screen pointer
shows 80 and operate ”MAX” button.
Check both of the values alternately for correctness and readjust if necessary.
When both of the calibration points are correct within a tolerance range +/-0.2, operate the ”Save” button.
By operating ”Quit” turn back to calibration menu.
Caution: Avoid differences in adjustment between pressure sensors P1 and P2.
77
Maintenance and Calibration
5.5.4 Temperature Sensor Calibration
Fig. 13 Temperature Sensor Calibration 1
Connect a Pt-100 simulator to the connector of the docking station, or directly to the board
AVM3-1 of ventilation module 2 (X7).
Apply a simulated temperature of 0 °C.
Turn the offset potentiometer screw (P1, AVM3-1) until the screen pointer show 20 and operate
the “MIN” button.
78
Maintenance and Calibration
Temperature Sensor Calibration , cont.
Fig. 14 Temperature Sensor Calibration 2
Apply a simulated temperature of 60 °C.
Turn the gain offset potentiometer screw (P2, AVM3-1) until the screen pointer show 80 and
operate the “MAX” button.
Check both of the values alternately for correctness and readjust if necessary.
When both of the calibration points are correct within a tolerance range +/- 0.2, operate the ”Save” button.
By operating ”Quit” turn back to calibration menu.
79
Maintenance and Calibration
5.5.5 Calibration Of Proportional Valve ADC
Caution:
“Internal Flow Sensor Calibration” and “Characteristic of The Proportional Valve” must always be performed together in the order as in the manual, never individually.
The three procedures “Proportional Valve Calibration”,
Fig. 15 Calibration Of Proportional Valve ADC 1
Operate the ”MIN” button to close the proportional valve.
Turn the offset potentiometer screw PV OFFS (P1, AVM3-2) until screen pointer shows 3
Operate the ”MIN” button a second time to store the value.
80
1
±
.
Maintenance and Calibration
Calibration Of Proportional Valve ADC , cont.
Fig. 16 Calibration Of Proportional Valve ADC 2
Operate the ”MAX” button to open the proportional valve.
Turn the gain potentiometer screw PV GAIN (P2, AVM3-2) until the screen pointer shows 97
2
±
.
Operate the ”MAX” button a second time.
Check at this point the driving gas flow by measurement at the driving gas outlet of the docking
station. The flow should be in a range of 80 +0/-5 L/min. If necessary check the adjustment of the pressure regulator of the proportional valve. The supply pressure of the proportional valve should not exceed 2.1 bar. Check the signal of the internal flow sensor by measurement on connector X5 (IF) on board AVM 3-2. The flow signal for max. flow measured at X5-3 and X5-1 should be in a range of 4,6 +/- 0,1 Volts. If necessary adjust by turning the restrictor at the flow splitter.
Check all of the values alternately for correctness and readjust if necessary.
When the calibration points are correct within a tolerance range of +/- 2, operate the ”Save” button.
By operating ”Quit” turn back to calibration menu.
81
Maintenance and Calibration
5.5.6 Calibration Of Internal Flow Sensor
Caution:
“Internal Flow Sensor Calibration” and “Characteristic of The Proportional Valve” must always be performed together in the order as in the manual, never individually.
The three procedures “Proportional Valve Calibration”,
Fig. 17 Calibration Of Internal Flow Sensor 1
Operate the ”OFFSET” button to apply a zero flow.
Change the offset value by rotary mouse until pointer shows 2.5
is about 127. To change the moving range of the DAC setting values use the ”MENU” and ”GRAPHICS” control panel keys.
Operate the “ OFFSET “ button again to store the value.
82
1
±
. A typically value for the offset
Maintenance and Calibration
Calibration Of Internal Flow Sensor , cont.
Fig. 18 Calibration Of Internal Flow Sensor 2
Operate the “GAIN” button to apply a maximum flow.
Change the gain value until pointer shows 80, a typically value is about 108 for the gain setting.
Operate the “ OFFSET “ button to close the valve and check the offset value. If necessary correct
the offset adjustment.
Operate the “ OFFSET “ button again to store the adjustment.
Operate the “ GAIN “ button to open the valve.
Operate the “ GAIN “ button again to store the adjustment.
Operate the “Save” button to memorize the stored values for gain and offset adjustments.
By operating ”Quit” turn back to calibration menu.
83
Maintenance and Calibration
5.5.7 Fixing Of Calibration Characteristic Proportional Valve
Caution:
“Internal Flow Sensor Calibration” and “Characteristic of The Proportional Valve” must always be performed together in the order as in the manual, never individually.
The three procedures “Proportional Valve Calibration”,
Fig. 19 Fixing Of Calibration Characteristic Proportional Valve
Connect a flow meter to the driving gas outlet.
Operate the 1 l/min button. Change the DAC value by rotary mouse until a flow of 1 l/min is
applied and operate the rotary mouse again to store the value. To change the moving range of the DAC setting values use the ”MENU” and ”GRAPHICS” keys. The ”OPTION” key reduces the DAC value of 256.
Increase the DAC value until a flow of 2 l/min is applied and operate the rotary mouse again.
Continue this procedure until 80 l/min are reached.
Operate the ”Save” button with a pause of about 30 seconds to store the calibration curve.
Turn back to calibration menu by operating the ”Quit” button.
Note: If a flow of 80 l/min cannot be reached with the DAC value of 4095, interpolate the last three
values to ensure a steady increase of the values and to avoid a calculation fault.
84
Tolerance range for the flow values: 1 l/min: +0 / - 0.1
2 l/min: +0 / - 0.1 4 l/min: +0 / - 0.1 6 l/min: +0 / - 0.1 8 l/min: +0 / - 0.1 10 l/min: +0 / - 0.2 12 l/min: +0 / - 0.2 15 l/min: +0 / - 0.2 20 l/min: +0 / - 0.4 25 l/min: +0 / - 0.4 30 l/min: +0 / - 0.4
Maintenance and Calibration
35 l/min: +0 / - 0.4 40 l/min: +0 / - 0.4 45 l/min: +0 / - 1 50 l/min: +0 / - 1 55 l/min: +0 / - 1 60 l/min: +0 / - 1 65l/min: +0 / - 1 70 l/min: +0 / - 1 75 l/min: +0 / - 5 80 l/min: +0 / - 5
85
Maintenance and Calibration
5.5.8 Calibration Of External Flow Sensor
Calibration routine for ventilator software AVM3nVer. 1.3.n:
Fig. 20 Calibration Of External Flow Sensor 1
Connect the patient block to unit, if necessary. Disconnect the breathing tube from breathing gas
inlet at the patient block to ensure that there is no flow via external flow sensor.
Calibration with warmed-up patient module, for preparation disconnect the plug of ext. flow sensor
on board AVM 3-2, connector X7, EF.
Set DAC to 1024 and turn gain potentiometer (P3) on AVM 3-2 until the pointer shows 90.
Connect the plug of ext. flow sensor to connector EF on AVM 3-2 and press the “MIN” button.
Press “Save” button.
Leave this window with “Quit”.
Start the unit in CMV-Adult mode with default settings to ventilate a test lung.
Check the measured values for tidal volume and adjust the gain potentiometer (P3) to the values
externally measured, respectively the settings of tidal volume on the ventilator (500 ml). Switch during adjustment of P3 sometimes to Standby mode, that the automatic offset calibration is performed. Check the measured values of the external flow sensor again in CMV-Adult mode. Start the sensor test., if a message “ext. flow sensor out of range appears”, turn the gain potentiometer (P3) smoothly until the test will be passed.
86
Maintenance and Calibration
Calibration Of External Flow Sensor, cont. Calibration routine for ventilator software AVM3nVer. 1.4.1:
0
Fig. 21 Calibration Of External Flow Sensor 2
Calibration with warmed-up patient module. Disconnect the breathing tube from breathing gas inlet
at the patient block to ensure that there is no flow at external flow sensor.
Increase the DAC value until the pointer for AVM 3-2 shows 50.0 +/- 0.2
Don’t change the DAC value setting after this.
Measure the voltages U1 and U2 (see page 2 for measurement points of this voltage) and make a
note of the values.
U1= -(Pin 8 of V13) to +(GND PIN), U2 = -(GND PIN) to + (between R35-R19)
The voltages U2 must be 1.4 times of the voltage U1.
Example: U1 = 1.448V , U2 = 1,901V U2/U1 = 1.313, This value is to low. The value has to be in a range of 1.39 to 1.41.
Connect the voltmeter to measure U2 and increase the value by turning on potentiometer P3.
Check U1 and U2 again for correct relation of U1 * 1.4 = U2. Repeat adjustment of P3 if necessary. Finally seal the potentiometer and leave the calibration routine be pressing QUIT.
87
Maintenance and Calibration
Calibration Of External Flow Sensor, cont.
Fig. 22 How to measure the voltages U1 and U2
88
Maintenance and Calibration
5.5.9 Test Menu Routines
For testing the sensor calibrations or check different functional tests open test menu and choose the wanted application.
Fig. 23 Test Menu Routines
89
Maintenance and Calibration
5.5.10 Test Characteristics Proportional Valve
Fig. 24 Test Characteristics Proportional Valve
By operating a flow button the chosen flow is set by the proportional valve. The measured flow
values of proportional valve at the internal flow sensor are shown and can be checked with an external flowmeter.
By operating the ”0” button the flow is switched off.
“Quit” returns to the test menu.
90
Maintenance and Calibration
5.5.11 Pressure, Oxygen and Temperature-Sensors
Fig. 25 Pressure, Oxygen and Temperature-Sensors
Measured values of the oxygen sensor are shown in %O2.
Pressure sensor values of the transducers P1 and P2 are shown in mbar.
The measured temperature sensor value is shown in C.
The adjustment of the sensors can be checked in the window.
91
5.5.12 Keyboard Test
Maintenance and Calibration
Fig. 26 Keyboard Test
Control panel keys and the rotary mouse may be tested.
To close the application operate the ”confirm” function of encoder, rotary mouse.
92
Maintenance and Calibration
5.5.13 Miscellaneous Functions
Fig. 27 Miscellaneous Functions
The ”unlock”, ”lock” and ”STOP” buttons may be operated to test the patient module draw-in
facility(if available).
By operating the Heat ”Off” button heating is switched on and off when operating button again.
Operating the 1 SEC and 2 SEC ”Off” buttons cause a test of the ”shut down” signals of µP1 and
µP2.
The Motor and Power windows show the status of the appropriately elements.
The S-Start and S-Stop button are showing the status of the micro switches for motor drive
control.
The power button AC-Fail indicates AC power supply fail. The Batt. button is showing the charging
of the battery.
93
Maintenance and Calibration
5.5.14 Solenoid Valves MV1 - MV4
Fig. 28 Solenoid Valves MV1 - MV4
This application enables the test of the solenoid valves.
Result of this test are the messages ”ON” , ”OFF” or “Faulty” for each of the four solenoid valves.
94
Preventive Maintenance
6 Preventive Maintenance
The following is a list of activities required for periodic maintenance of the HEYER Narkomat+ anesthesia system. Physical inspection, replacement of consumables and performance checks should be periodically performed per the schedule listed below. Certain calibration adjustments are only required only after replacing one or both of the active devices. HEYER is not responsible for component failure or loss resulting from the use of stated consumables beyond their recommended replacement interval. These are noted in the Preventive Maintenance Checklist on the following pages.
6.1 6 Month Service Interval
                                
Ensure operators manual is present Ensure preoperative Checkout list is attached Check unidirectional valves operate by visual inspection Check low O2 pressure alarm whistle Check “No driving gas” alarm Check the function of the N2O cutoff incase of O2 pressure loss Check the function of the N2O / AIR switch Check the differentiation of the flow meters knobs Check the function and correct flow of the O2 flush. Check that all flow meters work throughout their range Check the function of the Hypoxic Guard Check the function of the back light illumination Perform the compliance test at power up Perform Leak Test (Options Menu) without vaporizers Perform Leak Test (Options Menu) with vaporizers Perform O2 Cell calibration (Options Menu) Perform Manual tightness test Check the function and the accuracy of the APL valve in the manual mode Check the O2 display, 21% in room air, >95 in 100% O2 Check the correct function of the of the individual ventilation alarms Check the correct function of the Mute button Check the correct function of the Alarm LEDs Check that all cylinder and pipeline supply pressure gauges operate Inspect the line power cord Inspect/Clean water traps and filters Verify the gas tank check valves* Verify the pipeline supply check valves* Verify line voltage interruption alarm message Verify battery operation Verify the 115/230 volts AC at the convenient outlets Perform ventilation performance check Replace the O-rings on the vaporizer manifold (4 x 049-3182) Replace the APL valve membrane (323- 0310)
95
Preventive Maintenance
         
*Subject of the configuration
Replace O-rings on the docking station (4 x 049-3052) Replace O-rings between tanks and yokes (N2O 800-5947; O2: 800-5946)* Replace the O-ring of the CO2 absorber (323-0147) Replace the flat seal of the CO2 absorber (980-1170) Replace the O-ring of the Patient dome (323-0147) Replace the O-rings of the O2 Cell (609-3021 + 049-3074) Replace the fan filter (800-5580) Check the threaded insert for the CO2 absorber Check the correct function of the brakes Test of electric safety according to IEC 601-1
6.2 12 Month Service Interval
                   
Perform all the points of the 6 Month Service Interval Inspect and if necessary, replace the spindle nut of the patient module Examination of the internal hose connection Replace the o rings gas pipeline inlet block (1 x 980-1170; 2 x 980-1174) Replace the silicon valve plates (2 x 610-3156) Replace the O-rings of the in-and expiratory valve seat (2 x 610-3157) Replace the diaphragm valves (3 x 323-0100) Replace the diaphragm valve O-rings (3x 323-0292) Replace the flat rings at the bellows valve screw fitting (3x 980-1170) Replace the O-rings of the APL valve (3 x 323-0272) Replace the O-rings of the valve seatings (3 x 2 x 323-0272) Replace the O-rings for the valve seating screws (4 x 960-1036) Replace the O-rings of the test adapter (2 x 049-3052) Enter the service software Check and if necessary calibrate the O2 sensor Check and if necessary calibrate the Pressure sensors P1 and P2 Check and if necessary calibrate the temperature sensor Check the 2 kPa x 100 (2 bar; 30 PSI) pressure reducer for the driving gas Check the 200 Pa x 100 (200 mbar; 3 PSI) pressure reducers for the solenoid valve block Check the electrical and pneumatic function of the solenoid valves MV1 to MV4
6.3 36 Month Service Interval
  
Perform all the points of the 12 Month Service Interval Replace the internal battery (340-2020) Replace the room air valve membrane (323-0099)
96
Preventive Maintenance
6.4 Cleaning
6.4.1 Cleaning and disinfecting the apparatus
Before cleaning, switch off apparatus and disconnect from mains The apparatus housing can be cleaned with a cloth moistened with a liquid cleaning agent. To avoid abrasion, the cloth must be moist. A surface-disinfecting agent may be used. Use cleaning agent sparingly. Excess fluid could enter the apparatus causing damage.
6.4.2 Cleaning and sterilizing the Patient Module
The Patient Module can be autoclaved at a temperature of up to 134°C. Always autoclave the patient module with the protection cover connected to avoid a damage of the diaphragm valves.
Before sterilizing the Patient Module, remove the bellows dome, bellows, CO2 absorber and APL valve. The patient module requires no further dismantling. Autoclave the APL upper part and the APL membrane, as separate parts together with the patient module.
The bellows dome cannot be autoclaved! It does not come in contact with patient gas. If soiled, they can be cleaned with water and liquid cleaning agent. Do NOT use alcohol! The silicon bellows and the CO2 absorber canister can also be autoclaved at 134°C. To enhance the life of sealing materials, a temperature of 121°C is recommended.
6.5 Battery Replacement and Maintenance
6.5.1 Battery replacement
1. Open the access door (located at the rear of the Narkomat+)
2. Remove the old battery
3. Ensure the new battery is electrically connected and secured in the same manner as the original
one.
4. Close battery access door
5. Use only HEYER Battery (340-2020)
6.5.2 Battery Maintenance
Due to the self-discharging characteristics of this battery type. It is imperative that it is charged after three months of storage (or when not in use). If not charged, a permanent loss of capacity may occur as a result of sulfation. The batteries used in the Narkomat+ are of sealed lead acid construction. This battery type may be subject to local regulations regarding disposal. At the end of the battery life, dispose of batteries in accordance with the local regulations.
97
Preventive Maintenance
7 Order information
Ord. No. Description
606-2400
016-2410
022-3645
034-2950
556-2955
323-0094
603-3240 Breathing bag 2.3 l Latex
603-3241 Breathing bag 2.3 l Silicone
Suction set universal (includes bacteria filter and finger tip)
Finger tip
Plastic adapter, angled, for secretion collecting jar
Silicone tube φ 6 x 2 for suction
Bacteria filter for suction set
Bacteria filter
98
Preventive Maintenance
Order information, cont.
Ord. No. Description
800-5191
Connector for breathing bag
800-4853
M-Connector
800-4854
Y-piece, right angled
323-0092 323-0091 323-0084 323-0093 323-0089
Respiratory tube, silicone, 1.5m Respiratory tube, silicone, 1.2 m Respiratory tube, silicone, 1.1 m Respiratory tube, silicone, 0.9 m Respiratory tube, silicone, 0.6 m
323-0096
Inspiration valve, cpl.
323-0097
Expiration valve, cpl.
323-0098
Emergency air valve, cpl.
99
Preventive Maintenance
Order information, cont.
Ord. No. Description
323-0095
Ventilation pressure valve (APL valve), cpl.
323-0310
Ventilation pressure valve membrane
323-0099
Emergency air valve membrane
323-0313
Sealing set valve body, cpl.
323-0159
Coupling ring valve body
610-4020
Valve cover
610-4000
610-3156
Metal basket
Valve plate (Silicone)
100
Loading...