Long press the “M” will enter into the Setting interface, then press
the up or down key to adjust the interface; if set the data or function,
press “M”, set the flashing parameter, after the set press “OK” to save
the change, if want to cancel, press the “M” again; long press the “M”,
will back to the normal inspection interface.
3.2.1、AL1 setting:
When the gas concentration data exceed the set level, the red
backlight will be lit, meanwhile the K1 relay will act.
Press “M”, the AL1 flashing, the set digit also flashing, press the ▲
could change the position. Press ▼ could adjust the data, from 0~9,
press “OK” will save the setting. If cancel, press the “M”.
Picture 3.3、AL1 setting interface
3.2.2、AL2 setting:
When the gas concentration data exceed the set level, the red
backlight will be lit, meanwhile the K2 relay will act.
Press “M”, the AL2 flashing, the set digit also flashing, press the ▲
10
Page 12
could change the position. Press ▼ could adjust the data, from 0~9,
press “OK” will save the setting. If cancel, press the “M”.
Picture 3.4、AL2 setting interface
3.2.3、Device address setting:
Picture 3.5、setting the device address interface
Users can set the device address through this interface, the range of the
address is 1---247. Among the RS485 bus system, every device is only.
The function is only useful to this type device which has the RS485
function output.
3.2.4、Restore factory setting:
Picture 3.6、Loading the factory setting interface
11
Page 13
If you want to recover the setting parameters as before, press the M
button, then begin to flash, if need to cancel this, you can press M button,
if you want to confirm the recovering factory settings, you can press the
OK button, and if the recovering setting is ok, then it will turn back the
standard interface automatically, if not, it will stay the primary interface.
3.2.5、Zero calibration setting:
Enter into the Zero Calibration interface, you can press the M
button under this interface.
Picture 3.7、Zero calibration interface
The SET ZERO character flickers and the interface display the
relevant number at the same time, after the number is stable, you can
press the OK button, then the zero demarcated can be finished.
3.2.6、Calibrate gas concentration setting:
After entering the setting interface, you can press the ▼ to change
this interface. If you press M button, the character SET CAL and the
relevant setting digits will flicker, meanwhile, you can input the
calibrated gas concentration. The ▲ has the function of removing the
setting digits, press the ▲ to change the digits you want to set, and press
12
Page 14
the ▼ can change the number of the relevant digits. The number can be
cyclic changed from 0~9, after this, you can press the OK button to
finish the demarcated gas concentration setting, the interface of the
standard gas concentration as below:
Picture 3.8、Calibration gas concentration interface
3.2.7、Span calibrations setting:
It refers to that the users can adjust the instrument by using the
calibration gas. After entering this interface, press the M button to
calibrate, then the SET PEAK becomes flickering. After that, pass
through the standard gas(control the flow as 200ml/min), when the data
stable, press the “OK” to save the setting.
Picture 3.9、Span calibration setting
四、Calibration methods
Before the gas detector leave our factory, we have calibrated it. If
the users have the calibration device, they can calibrate it regularly as the
13
Page 15
instruction. If they don’t have the equipment, they can go to the relevant
measure departments for help or send it to our company, then we can
calibrate it with costs.
In order to assure device measure accuracy, the normal calibrated
cycle of the instrument is 3---6 months. If the inspection place long term
with the higher concentration gas, then the calibrated cycle will be
shortened. So in order to ensure the instrument’s measure accuracy,
should timely calibrate it and save the calibration record.
4.1、Zero calibration
Pass the 0ppm concentration standard gas through the gas detector,
usually as the normal air as the zero, some gas(such as the O2, CO2),
take the N2 as the Zero; then enter into the Zero Calibration interface,
press the “M”, then the SET ZERO flashing, then display the data, when
the data stable, press the “ok” to save the setting and finish the Zero
Calibration.
Remark: Zero calibration must be operated in clean air, do not
operate in the normal working environment.
4.2、Span calibration
Firstly, make the standard gas with set concentration through the
flow meter, and control the flow between 160---300ml/min, then link the
conduit into the instrument tube, then choose the menu and enter the
14
Page 16
span calibration interface. Press the “M”, the SET PEAK flashing, after
the value is stable(don’t rise or descend any more), press the OK button
to save and exit.
五、Remote controller
Remote controller mainly for the working place with heavy dust or
other place inconvenient for open the cap. This could avoid the danger.
15
Page 17
Enter the “Menu” enter into the setting interface, then use the “<”
“>”to adjust the parameter interface, press the “Set” to adjust the
parameter. Press the “∧”“∨”to adjust the data, press “ok” to save the
setting, if only press “Menu” will not save, press “ESC” will be back to
normal working mode.
六、Notice and maintenance for sensor
Keep the gas detector away from corrosive gas, lampblack, dust and rain
when it working. Avoid the dropping from height or severe vibrations.
Do not regular contact the device with high concentration gas which has
higher concentration than the detecting range, or the sensor life will
reduce.
For mixed combustible gas or liquid vapor, there may be errors in the
detect results, due to the operating environment is different from the
calibration environment.
Recommend sensor life: Catalytic combustion sensor of 3 years,
Electrochemical sensor of 2 years. PID sensor of 3~5 years.
七、Warranty
OC-F08 gas detector with the warranty of 12 month(from the day leave
16
Page 18
factory). During this period, we will be responsible for providing the
free maintain to quality problems of breakdown under normal usage.
Our company disclaim all responsibility for the damage caused by
misuse, manmade damage or natural disaster.
If the sensor life is end, please contact us for replacement.
For the failures which can not be solved locally, please return the device
to distributor or manufacturer for maintenance. Our company will send
back to you as soon as possible.
Technical Specification
Applicable Gas Oxygen, Combustible gas, toxic gas, etc
Detecting Range According to the detected gas
Resolution According to the detected gas
Sampling Diffusion or pump-suction, circulation selectable
Accuracy
Repeatability ≤±1%
Zero drift
Sensor Type
Sensor Life
Response Time (T90) ≤20s
Alarming mode Audible-visual alarm(Optional)
Temperature: -30℃~60℃(Customized for special
demands)
Humidity: ≤95%RH (non-condensing); if >90%RH, should
add the filter
Pressure: 86kPa~106kPa
(Standard)Three-wire system 4~20mA standard signals
output, available for PLC, DCS, etc;
(Optional) Four-wire system RS485 digital signal output;
17
Page 19
(Optional) Frequency signal 200~1000Hz output;
Two-level passive contact output:220VAC 0.5A, 5VDC
0.5A
Installation Means Wall mounted, Pipe type (Thread:M40*1.5 )
Material of Body Cast aluminum
Explosion-proof Index Ex d IIC T6
Protection Grade IP66
Electrical Interface 3/4″NPT(Female Screw)、1/2″NPT(Female Screw)