Helmer Scientific HLR105-GX, iPR105-GX, HPR105-GX, iBR105-GX, HBR105-GX Operation Manual

...
Laboratory
i.Series
iLR105-GX
Horizon Series
HLR105-GX
Pharmacy
i.Series
iPR105-GX
Horizon Series
HPR105-GX
i.Series
iBR105-GX
Horizon Series
HBR105-GX
Refrigerator Operation Manual
i.Series® and Horizon Series™ - Undercounter
360397/A
Document History
Revision Date CO Supersession Revision Description
A 15 JAN 2019 14198 n/a Initial release
* Date submitted for Change Order review. Actual release date may vary.
Document Updates
The document is furnished for informational use only, is subject to change without notice and should not be construed as a
commitment by Helmer Scientic. Helmer Scientic assumes no responsibility or liability for any errors or inaccuracies that may appear in the informational content contained in this material. For the purpose of clarity, Helmer Scientic considers only the most recent revision of this document to be valid.
Notices and Disclaimers
Condential / Proprietary Notices
Use of any portion(s) of this document to copy, translate, disassemble or decompile, or create or attempt to create by reverse engineering or otherwise replicate the information from Helmer Scientic products is expressly prohibited.
Copyright and Trademark
Copyright © 2019 Helmer, Inc. Helmer®, i.Series®, i.C³®, Horizon Series™, and Rel.i™ are registered trademarks or trademarks of Helmer, Inc. in the United States of America. All other trademarks and registered trademarks are the property of their respective owners. Helmer, Inc., doing business as (DBA) Helmer Scientic and Helmer.
Disclaimer
This manual is intended as a guide to provide the operator with necessary instructions on the proper use and maintenance of certain Helmer Scientic products.
Any failure to follow the instructions as described could result in impaired product function, injury to the operator or others, or void applicable product warranties. Helmer Scientic accepts no responsibility for liability resulting from improper use or maintenance of its products.
The screenshots and component images appearing in this guide are provided for illustrative purposes only, and may vary slightly from the actual software screens and/or product components.
Helmer Scientic 14400 Bergen Boulevard Noblesville, IN 46060 USA
www.helmerinc.com Part No. 360397/A
Helmer Scientic i.Series® and Horizon Series™ Refrigerator - Undercounter Operation Manual
Contents
1 About this Manual ................................................................................................ 4
1.1 Intended Audience .......................................................................................... 4
1.2 Model Reference ............................................................................................4
1.3 Intended Use ...............................................................................................4
1.4 Safety Symbols and Precautions ...............................................................................4
1.5 Avoiding Injury ..............................................................................................6
1.6 General Recommendations ................................................................................... 6
2 Installation ...................................................................................................... 7
2.1 Location ...................................................................................................7
2.2 Placement and Leveling ...................................................................................... 7
2.3 Stacked Undercounter Units ...................................................................................7
2.4 AC Power Cord .............................................................................................7
2.5 Temperature Probes .........................................................................................8
2.6 Chart Recorder (if included) ...................................................................................8
3 i.Series® Operation .............................................................................................. 10
3.1 Initial Power-up ............................................................................................10
3.2 Operation .................................................................................................11
3.3 Change Temperature Setpoint .................................................................................11
3.4 Set Alarm Parameters ....................................................................................... 12
3.5 Active Alarms ............................................................................................. 13
3.6 Mute Active Alarms ......................................................................................... 13
3.7 Light Operation (optional) ....................................................................................13
4 Min/Max Temperature Monitoring ..................................................................................15
5 i.Series® Access Control (Optional) .................................................................................16
5.1 Setup ....................................................................................................16
5.2 Open Refrigerator with Access Control .......................................................................... 16
360397/A ii
Helmer Scientic i.Series® and Horizon Series™ Refrigerator - Undercounter Operation Manual
6 Horizon Series™ Operation ....................................................................................... 17
6.1 Initial Power Up ..............................................................................................17
6.2 Display Minimum and Maximum Monitor Temperature Recordings ....................................................... 17
6.3 Set Parameter Values .........................................................................................19
6.4 Set Temperature Units ......................................................................................... 19
6.5 Set Alarm Setpoints (Parameters) ................................................................................ 19
6.6 Temperature Calibration O󰀨sets ..................................................................................19
6.7 Active Alarms ................................................................................................ 20
6.8 Mute and Disable Audible Alarms ................................................................................. 20
6.9 Light Operation (optional) .......................................................................................20
7 Horizon Series™ Access Control (Optional) ......................................................................... 21
7.1 Setup
8 ProductSpecications ........................................................................................... 22
8.1 Operating Standards ........................................................................................22
9 Compliance .................................................................................................... 24
9.1 Safety Compliance .........................................................................................24
9.2 Environmental Compliance ...................................................................................24
Appendix A: Parts .................................................................................................. 25
i.Series Parts
......................................................................................................21
....................................................................................................25
Appendix B: Parts
Horizon Series Parts ..............................................................................................26
.................................................................................................. 26
360397/A iii
Helmer Scientic i.Series® and Horizon Series™ Refrigerator - Undercounter Operation Manual
1 About this Manual
1.1 Intended Audience
This manual provides information on how to use i.Series® and Horizon Series™ undercounter laboratory, blood bank, and pharmacy refrigerators. It is intended for use by end users of the refrigerator and authorized service technicians.
1.2 Model Reference
Models are indicated by a distinguishing model number that corresponds to the series, type, and capacity of the refrigerator. For example, “iLR105-GX” refers to an i.Series Laboratory Refrigerator with 1 door and a capacity of 5 cu ft.
1.3 Intended Use
Note
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
Helmer refrigerators are intended for the storage of blood products and other medical and scientic products.
1.4 Safety Symbols and Precautions
Symbols found in this document
The following symbols are used in this manual to emphasize certain details for the user:
Task Indicates procedures which need to be followed.
Note Provides useful information regarding a procedure or operating technique when using Helmer
Scientic products.
NOTICE Advises the user against initiating an action or creating a situation which could result in damage to
equipment; person injury is unlikely.
CAUTION Advises the user against initiating an action or creating a situation which could result in damage
to equipment or impair the quality of the products or cause minor injury.
WARNING Advises the user against initiating an action or creating a situation which could result in damage
to equipment and serious personal injury to a patient or the user.
360397/A 4
Helmer Scientic i.Series® and Horizon Series™ Refrigerator - Undercounter Operation Manual
DD.1.5 370198-1/A
CAUTION - RISK OF FIRE OR EXPLOSION. DISPOSE OF
DD.1.4A 370196-1/A
DANGER - RISK OF FIRE OR EXPLOSION.
DD.1.6 370199-1/A
ENTRAÎNANT UNRISQUE D’INCENDIE OU D’EXPLOSION. MANIPULATION.UTILISATION DE FLUIDE FRIGORIGÈNE
INFLAMMABLE.
CAUTION - RISK OF FIRE OR EXPLOSION DUE TO
DANGER - RISK OF FIRE OR EXPLOSION. FLAMMABLE
DD.1.3 370195-1/A
UTILISATION DE FLUIDE FRIGORIGÈNE INFLAMMABLE. NE PAS UTILISER DE DISPOSITIFS MÉCANIQUES POUR
CAUTION - RISK OF FIRE OR EXPLOSION. FLAMMABLE OWNER’S GUIDE BEFORE ATTEMPTING TO INSTALL OR
DD.1.4B 370197-1/A
CAUTION - RISK OF FIRE OR EXPLOSION DUE TO
FLAMMABLE REFRIGERANT USED. FOLLOW HANDLING INSTRUCTIONS CAREFULLY IN COMPLIANCE WITH U.S.
DD.2.2 370203-1/B
AUX RÈGLEMENTS DU GOUVERNMENT DES ÉTATS-UNIS.
THIS UNIT IS INTENDED FOR USE IN COMMERCIAL,
DD.1.10 370202-1/A
Symbols and Labels found on the units
The following symbols may be found on the refrigerator or refrigerator packaging:
PUNCTURE OF REFRIGERANT TUBING; FOLLOW HANDLING INSTRUCTIONS CAREFULLY. FLAMMABLE REFRIGERANT USED. ATTENTION - TUBES FRIGORIFIQUES PERCÉS
SUIVRE ATTENTIVEMENT LES INSTRUCTIONS DE
REFRIGERANT USED. CONSULT REPAIR MANUAL / SERVICE THIS PRODUCT. ALL SAFETY PRECAUTIONS
MUST BE FOLLOWED.
ATTENTION - RISQUE D’INCENDIE OU D’EXPLOSION. UTILISATION DE FLUIDE FRIGORIGÈNE INFLAMMABLE. CONSULTER LE MANUEL DE RÉPARATION / LE GUIDE D’ENTRETIEN AVANT DE TENTER D’INSTALLER OU D’ENTRETENIR CET APPAREIL. TOUTES LES PRÉCAUTIONS DE SÉCURITÉ DOIVENT ÊTRE SUIVIES.
Caution: Risk of damage to equipment or
Earth / ground terminal
danger to operator
Caution: Hot surface Protective earth / ground terminal
Caution: Shock / electrical hazard Refer to documentation
Warning: Flammable material Caution: Unlock all casters
PROPERLY IN ACCORDANCE WITH FEDERAL OR
Warning: Crushing of hands / ngers
Warning: Cold surface below -30°C
Caution: Follow handling instructions carefully.
Caution: Consult instruction manual prior to installation or service.
LOCAL REGULATIONS. FLAMMABLE REFRIGERANT USED. ATTENTION - RISQUE D’INCENDIE OU D’EXPLOSION. ÉLIMINER DE FAÇON ADAPTÉE SELON LES RÈGLEMENTS FÉDÉRAUX OU LOCAUX. UTILISATION DE FLUIDE FRIGORIGÈNE INFLAMMABLE.
FLAMMABLE REFRIGERANT USED. TO BE REPAIRED ONLY BY TRAINED SERVICE PERSONNEL. DO NOT PUNCTURE REFRIGERANT TUBING.
DANGER - RISQUE D’INCENDIE OU D’EXPLOSION. UTILISATION DE FLUIDE FRIGORIGÈNE INFLAMMABLE. A FAIRE RÉPARER UNIQUEMENT PAR LE PERSONNEL D’ENTRETIEN QUALIFIÉ. NE PAS PERCER LES TUBES FRIGORIFIQUES.
REFRIGERANT USED. DO NOT USE MECHANICAL DEVICES TO DEFROST REFRIGERATING EQUIPMENT. DO NOT PUNCTURE REFRIGERANT TUBING. DANGER - RISQUE D’INCENDIE OU D’EXPLOSION.
DÉGIVRER LES APPAREILS DE RÉFRIGÉRATION. NE PAS PERCER LES TUBES FRIGORIFIQUES.
GOVERNMENT REGULATIONS.
ATTENTION - RISQUE D’INCENDIE OU D’EXPLOSION EN RAISON DE L’UTILISATION DE FLUIDE FRIGORIGÈNE INFLAMMABLE. SUIVRE ATTENTIVEMENT LES INSTRUCTIONS DE MANIPULATION CONFORMÉMENT
Caution: Dispose of properly
Danger: Repair only by trained service personnel.
Danger: Do not use mechanical devices to defrost.
Caution: Follow handling instructions carefully in compliance with U.S. Government regulations
INDUSTRIAL, OR INSTITUTIONAL OCCUPANCIES AS
Pantone 185 (Red) Warning: Puncturing or opening refrigerant
DEFINED IN THE SAFETY STANDARD FOR REFRIGERATION SYSTEMS, ANSI/ASHRAE 15.
CETTE UNITÉ EST DESTINÉE À UNE UTILISATION COMMERCIALE, INDUSTRIELLE OU INSTITUTIONNELLE, COMME DÉFINI DANS LA NORME DE SÉCURITÉ DES SYSTÈMES FRIGORIFIQUES, ANSI/ASHRAE 15.
Intended use per ANSI/ASHRAE
circuit might be expected.
360397/A 5
Helmer Scientic i.Series® and Horizon Series™ Refrigerator - Undercounter Operation Manual
1.5 Avoiding Injury
WARNINGS
• Do not use mechanical devices or other means to accelerate the defrosting process, other than those recommended by the manufacturer.
• Do not damage the refrigerant circuit.
Review safety instructions before installing, using, or maintaining the equipment.
Before moving unit, ensure door is closed and casters (if installed) are unlocked and free of debris. Before moving unit, disconnect the AC power cord and secure the cord. Never physically restrict any moving component. Avoid removing electrical service panels and access panels unless so instructed. Keep hands away from pinch points when closing the door. Avoid sharp edges when working inside the electrical compartment and refrigeration compartment. Ensure products are stored at recommended temperatures determined by standards, literature, or good
laboratory practices.
Proceed with caution when adding and removing product from the refrigerator. Do not open multiple loaded drawers at the same time. Use manufacturer supplied power cord only. Avoid risk of ignition by using only manufacturer supplied components and authorized personnel when servicing the unit. ♦ Using the equipment in a manner not specied by the manufacturer may impair the protection provided by the equipment. Ensure product is stored safely, in accordance with all applicable organizational, regulatory, and legal requirements. ♦ The refrigerator is not considered to be a storage cabinet for ammable or hazardous materials. ♦ Use caution when moving a stacked conguration.
CAUTION
Decontaminate parts prior to sending for service or repair. Contact Helmer Scientic or your distributor for
decontamination instructions and a Return Authorization Number.
1.6 General Recommendations
General Use
Allow refrigerator to come to room temperature before switching power on. During initial startup, high temperature alarm may sound while refrigerator reaches operating temperature.
Initial Loading
Allow chamber temperature to stabilize at the setpoint before storing product.
Product Loading Guidelines
When loading your refrigerator, take care to observe the following guidelines:
♦ Never load refrigerators beyond capacity. ♦ Always store items within shelves or trays. ♦ Temperature uniformity is maintained by air circulation, which could be impeded if unit is overlled, particularly at the top or
back. Ensure a minimum of 2" (50 mm) clearance is provided below the fan.
Note
Products stacked against back wall may obstruct air ow and a󰀨ect performance of unit.
360397/A 6
Helmer Scientic i.Series® and Horizon Series™ Refrigerator - Undercounter Operation Manual
2 Installation
2.1 Location
WARNING
Keep all ventilation openings in the enclosure or, in the structure for building-in, clear of obstruction.
♦ Has a grounded outlet meeting the electrical requirements listed on the product specication label. ♦ Is clear of direct sunlight, high temperature sources, heating vents, and air conditioning vents. ♦ Has a minimum of 3” (76 mm) of space behind the refrigerator for clearance and feature access. ♦ Meets the limits specied for ambient temperature (15 °C to 32 °C) and relative humidity.
2.2 Placement and Leveling
CAUTION
• To prevent tipping, ensure the casters (if installed) are unlocked and the door is closed before moving the refrigerator.
• Use of leveling feet or casters is required.
• Do not sit, lean, push or place heavy objects on top surface.
• Do not lean on or push down on an open door or extended drawers.
• To avoid damaging refrigerant tubing or risking refrigerant leak, use caution when moving or operating the unit.
1. Move refrigerator into place. Lock casters if installed.
2. Ensure refrigerator is level.
Note
Helmer recommends the use of leveling feet and wall and oor brackets (PN 400472-2) for stabilization. Contact Helmer
Technical Service for parts and instruction.
2.3 Stacked Undercounter Units
CAUTION
For stacked conguration, both units must have leveling feet installed.
• Back brace bars and front stabilizing brackets must be installed (Blue - PN 400821-1; Stainless Steel - PN 400821-2).
• When stacking units, place the heavier unit on the bottom.
• Do not open multiple loaded drawers at the same time.
• Do not lean on or push down on an open door or extended drawers.
Contact Helmer or your distributor for more information regarding the stacking kit and methods to secure both units to the wall
and/or oor.
2.4 AC Power Cord
CAUTION
Use manufacturer supplied power cord only.
Install power cord
If packaged with modular cord, insert plug securely into the refrigerator power receptacle prior to connecting to grounded outlet.
360397/A 7
Helmer Scientic i.Series® and Horizon Series™ Refrigerator - Undercounter Operation Manual
2.5 Temperature Probes
A solid ballast or probe bottle and container of glycerin have been provided with this unit. The glycerin is used to create a solution which, when placed in the probe bottle, simulates the product stored in the refrigerator. The product simulation solution
temperature reects the product’s temperature during normal operation.
Notes
• Temperature probes are fragile; handle with care.
• Solid ballast (if installed) should be placed in the bracket in a horizontal position.
CAUTION
Failure to ll probe bottles (if installed) or keep probe bottles lled to the appropriate level may cause the chamber
temperature to display higher or lower than the actual temperature.
Primary Monitor Probe
The primary monitor probe bottle is located at the top left side of the refrigerator.
Primary monitor probe
Fill Temperature Probe Bottle
Note
Use approximately 4 oz. (120 mL) of product simulation solution (10:1 ratio of water to glycerin). Glycerin packet included in refrigerator box.
1. Remove probe(s) from bottle and remove bottle from bracket.
2. Remove cap and ll with approximately 4 oz. (120 mL) of product simulation solution.
3. Secure cap on bottle and place in bracket.
4. Replace probe(s), immersing at least 2” (50 mm) in solution.
2.6 Chart Recorder (if included)
Note
• For complete information, refer to the Temperature Chart Recorder Operation and Service Manual.
• If chart recorder is operated on battery power, the battery should be replaced to ensure the back-up source has proper charge.
The chart recorder has a back-up battery system enabling a period of continuous operation if power is lost. Battery life varies by manufacturer as well as voltage level remaining. If full battery power is available, back-up power for the temperature chart recorder is available for up to 14 hours.
360397/A 8
Helmer Scientic i.Series® and Horizon Series™ Refrigerator - Undercounter Operation Manual
Prior to use:
Route the chart recorder probe through the rear access port and place in bottle with primary monitor probe, or install in ballast and secure with thumb screw.
Set up and Operation
Access the chart recorder by pulling the door open.
Install battery
Connect the leads to the battery to provide back-up power to the chart recorder.
Install / Replace Chart Paper
Note
For accurate temperature reading, ensure the current time is aligned with the time line groove when the chart knob is fully tightened.
1. Press and hold C button. When stylus begins to move left, release button. The LED ashes.
2. When stylus stops moving, remove chart knob then move knob up and away from chart paper.
3. Place new chart paper on chart recorder.
4. Gently lift stylus and rotate paper so current time line corresponds to time line groove.
5. Hold chart paper in place while making sure the chart knob is fully tightened. (Failure to fully tighten the knob can result in paper
slipping and losing time.)
6. Press and hold C button. When stylus begins to move right, release button.
7. Conrm stylus is marking on paper and stops at the correct temperature.
8. Calibrate chart recorder to match primary temperature if needed and close recorder door.
Power Supply
The temperature chart recorder uses AC power when the system is operating. If AC power fails, the recorder continues to record temperature with back-up power provided by the nine volt battery.
♦ The LED indicator glows green continually when main power is functioning and the battery is charged. ♦ The LED indicator glows red continually when main power is functioning and the batteries is either not installed or needs to be
replaced.
♦ The LED indicator ashes red to indicated the recorder is receiving power only from the back-up battery. ♦ The LED indicator ashes during chart paper change mode.
360397/A 9
Helmer Scientic i.Series® and Horizon Series™ Refrigerator - Undercounter Operation Manual
3 i.Series® Operation
3.1 Initial Power-up
1. Plug the power cord into a grounded outlet that meets the electrical requirements on the product specication label.
2. Switch AC ON/OFF switch ON.
3. Switch back-up battery switch ON.
Notes
• For models equipped with optional Access Control, the back-up battery is turned ON with a key switch.
• The Start screen is displayed when the i.C3 is powered on. The i.C3 will take approximately 2-5 minutes to boot up.
Start screen
i.Series
The language screen is displayed when the i.C3 is powered on. Use the Language screen to select the i.C3 display language.
Language screen
Note
English is the default language.
If an alarm sounds, temporarily mute the alarm by touching the Mute icon.
Home screen - alarm muted Mute icon
Note
Active alarms are displayed on the Home screen. If an alarm condition other than High Temperature occurs, refer to the service manual for troubleshooting.
360397/A 10
Helmer Scientic i.Series® and Horizon Series™ Refrigerator - Undercounter Operation Manual
3.2 Operation
Notes
• Refer to the i.C3 User Guide for complete information regarding the i.C3 User Interface.
• The i.C3 Home screen displays temperature and alarm information, and provides icons to gain access to other functions of the i.C3.
• After two minutes of inactivity, the screensaver will be displayed. To return to the Home screen, touch the screensaver.
Home screen Home screensaver (touch to return to Home screen)
3.3 Change Temperature Setpoint
Note
The Temperature Setpoint toggle button can be accessed from either the initial Settings screen or the Device Control Settings screen.
>
> > Enter the Settings password. Select Temperature setpoints. Touch minus (-) or plus (+) on the spin box to change
the value.
Settings screen Device Control Settings screen
i.Series
Notes
• Default Settings password is 1234.
• Default setpoint is 4.0 °C for iLR and iBR models, or 5.0 °C for iPR models.
360397/A 11
Helmer Scientic i.Series® and Horizon Series™ Refrigerator - Undercounter Operation Manual
3.4 Set Alarm Parameters
> > Enter the Settings password. Scroll down to select Alarm Settings. Touch minus (-) or plus (+) on spin box to set
each alarm parameter.
Settings screen
Alarms Settings screens
i.Series
360397/A 12
Helmer Scientic i.Series® and Horizon Series™ Refrigerator - Undercounter Operation Manual
3.5 Active Alarms
Home screen with active alarm
Table 1. Active Alarms
Alarm Description
3
Communication Failure 1 Communication lost between i.C Communication Failure 2 Conguration le is corrupt or i.C Communication Failure 3 Corrupt database Compressor Probe Failure Probe not functioning properly Compressor High T emperature Compressor temperature reading is above high temperature alarm setpoint Control Probe Failure Probe not functioning properly Drive Space Low SD card that holds downloadable data is approaching capacity Drive Space Full SD card that holds downloadable data is full Door Open Door is open beyond user-specied duration Inverter Communication Failure Communication is lost between the i.C Low Battery Monitoring system backup battery voltage is low No Battery Monitoring system backup battery voltage is decient Power Failure Power to the unit has been disrupted Primary Monitor Probe Failure Probe not functioning properly Primary Probe High Temperature Primary monitor probe reading is above high temperature alarm setpoint Primary Probe Low Temperature Primary monitor probe reading is below low temperature alarm setpoint
display board and control board
3
is unable to access the conguration le
3
control board and the VCC inverter
3.6 Mute Active Alarms
i.Series
Audible alarms may be temporarily muted by touching the Mute icon. The delay duration can be set and changed by selecting Sound Settings from the Settings screen. The duration may be set to any value from 1 - 60 minutes. The delay time remaining will be displayed in the bottom right corner of the icon. If the alarm is still active after the mute delay has ended, the audible alarm will sound.
> > Enter the Settings password. Scroll down to select Sound Settings. Touch minus (-) or plus (+) on spin box to set
the mute duration.
3.7 Light Operation (optional)
Press Light Icon to turn LED lights ON or OFF.
Auto ON/OFF feature can be congured in Settings.
360397/A 13
Helmer Scientic i.Series® and Horizon Series™ Refrigerator - Undercounter Operation Manual
Application Icon Table
Image Name Image Name Image Name Image Name
Home Temperature Graph PDF Download Save
Event Log Alarm Test Upload Cancel
Mute Information Logs Access Control Back Arrow
Reset
Zoom Information Display Brightness Alarm Conditions
i.C³ Applications Icon Transfer Cancel Test Battery Power
Settings CSV Download Light On/O󰀨
Contact Information/ Contact Helmer
Access Log Scroll
Temperature Graph Forward/Back
i.Series
360397/A 14
Helmer Scientic i.Series® and Horizon Series™ Refrigerator - Undercounter Operation Manual
4 Min/Max Temperature Monitoring
The Min/Max temperature display provides the highest and lowest Primary Monitor probe temperature reading since the last system reset (power-on event) or manually-initiated reset. Touch the Reset icon to the right of the display to manually reset.
Notes
The Min/Max temperature display can be turned on or o󰀨 through Display Settings.
• Once the time reaches the maximum display of 999 hours and 60 minutes, the message will display “>999:60”, but minimum and maximum temperatures will continue to be tracked.
i.Series
360397/A 15
Helmer Scientic i.Series® and Horizon Series™ Refrigerator - Undercounter Operation Manual
5 i.Series® Access Control (Optional)
Allows user-specic secure access to the refrigerator.
Notes
• During a power failure, the optional Access Control lock will remain locked until battery power is depleted or until the back-up battery key switch is switched OFF.
• Switching the back-up battery key switch OFF will disable the monitoring system during a power failure.
• During a power failure, switch the battery back-up switch OFF and use the mechanical door key to provide secure storage for refrigerator contents.
• Refer to i.C3 User Guide for complete information regarding Access Control.
5.1 Setup
Congure and manage user-specic accounts to allow controlled access to the refrigerator.
> > Access Setup
Access Control Setup password screen Access Control Setup screen
Enter the supervisor PIN to set up Access Control. Select the Add User button and follow the on-screen prompts to set up users.
Notes
• Initial factory supervisor PIN = 5625
• The supervisor PIN cannot be deleted, and should be changed to prevent unauthorized user ID setup. The supervisor PIN does not allow access to the unit. At least one user ID must be set up to gain access to the unit.
5.2 Open Refrigerator with Access Control
Enter a valid PIN using the keypad.
i.Series
360397/A 16
Helmer Scientic i.Series® and Horizon Series™ Refrigerator - Undercounter Operation Manual
6 Horizon Series™ Operation
6.1 Initial Power Up
1. Plug the power cord into a grounded outlet that meets the electrical requirements on the product specication label.
2. Switch AC ON/OFF switch ON.
3. Remove the 9 V battery from the literature box and install it.
4. Press Down Arrow (Mute) if high temperature alarm sounds.
Notes
• For models equipped with the optional Access Control, switch the back-up battery key switch ON.
• During a power failure, the back-up battery continues to provide power to the optional Access Control lock (if equipped).
If the back-up battery is not functioning, the optional Access Control lock will not secure the door.
• If an alarm condition other than High Temperature occurs, refer to the service manual for troubleshooting.
Door Open Lamp
High and Low
Temp Lamps
Temperature
Unit Indicator
LED Display
MODE Button
CONTROL and
MONITOR Lamps
UP and DOWN Arrows
SET Button
Horizon Series™ temperature monitor and control interface
6.2 Display Minimum and Maximum Monitor Temperature Recordings
The minimum and maximum recording feature allows the user to view a minimum temperature occurrence and a maximum temperature occurrence within a given period of time. The timer provides a time reference in which those temperatures occurred.
View minimum temperature recording
1. Press and hold the Down Arrow button for 1 second and listen for a single beep.
2. The display will alternate between LO and a valid temperature value ve (5) times followed by a single beep to indicate exit back to the temperature display.
View maximum temperature reading
1. Press and hold the Up Arrow button for 1 second and listen for a single beep.
2. The display will alternate between HI and a valid temperature value ve (5) times followed by a single beep to indicate exit back to the temperature display.
Horizon Series
360397/A 17
Helmer Scientic i.Series® and Horizon Series™ Refrigerator - Undercounter Operation Manual
View recorded temperature timer
Notes
• The timer denotes the period of time that has elapsed. It does not display the time at which a minimum or maximum
temperature occurred.
• The maximum period of time the timer can record is 99:59 (99 hours and 59 minutes). The timer will reset automatically
when this maximum is reached.
1. Press and hold either the Up or Down Arrow button for 1 second.
2. While the display is ashing the HI or LO value, press and hold the SET button for 1 second.
3. The display will alternate ve (5) times between CLr and a value representing the number of hours and minutes that have elapsed since the last recording (example: 12:47 would represent 12 hours and 47 minutes). A single beep will follow to indicate exit back to temperature display.
Clear minimum and maximum temperature recordings.
1. Press and hold either the Up or Down Arrow button for 1 second.
2. While the display is ashing the HI or LO value, press and hold the SET button for 1 second and listen for a single beep.
3. While the display is ashing the elapsed time since last reset, press and hold the SET button for 2 seconds. CLr will be displayed followed by a series of 3 beeps to indicate exit back to the temperature display.
Notes
The minimum and maximum temperature and timer will reset when:
the unit is powered o󰀨 and battery back-up is not engaged, or
• after 99 hours and 59 minutes have elapsed.
Change Refrigerator Temperature Setpoint
Note
Default setpoint is 4.0 °C for HBR and HLR models; 5.0 °C for HPR models.
1. Press and release MODE to change to Control mode. The CONTROL lamp will illuminate.
2. Press and hold SET to display the current setpoint temperature.
3. Hold SET and press Up or Down Arrow as necessary to set the desired setpoint value.
4. Release all buttons; the setpoint is changed.
5. Press and release MODE to return to Monitor mode. The MONITOR lamp will illuminate.
Horizon Series
360397/A 18
Helmer Scientic i.Series® and Horizon Series™ Refrigerator - Undercounter Operation Manual
6.3 Set Parameter Values
1. Press and hold the Up and Down Arrows simultaneously for 3 seconds to enter program mode.
2. The LED Display will show .C or .F to indicate Celsius or Fahrenheit.
3. Press and release MODE button to scroll through the parameters.
4. Once the desired parameter is selected, press and hold the SET button while pressing the Up or Down Arrow to select the desired value.
5. Release SET button. The new setting is saved.
6. Press and hold the Up and Down Arrows simultaneously for 3 seconds to exit program mode.
Note
Contact Helmer Technical Service to set Rail Limit values.
Table 2. Parameter Values
Parameter Visual Indicator Range Default
Celsius or Fahrenheit °C Lamp or °F Lamp ˚C, ˚F ˚C High Temperature MONITOR Lamp & HIGH Lamp -40.0 to 25.0 (˚C); -40 to 77 (˚F) 5.5 °C
Low Temperature MONITOR Lamp & LOW Lamp -40.0 to 25.0 (˚C); -40 to 77 (˚F) 1.5 °C (HBR models)
Monitor Offset MONITOR Lamp -10.0 to 10.0 (˚C); -18 to 18 (˚F) Varies Control Offset CONTROL Lamp -10.0 to 10.0 (˚C); -18 to 18 (˚F) Varies Upper Rail Limit CONTROL Lamp & HIGH Lamp 0.1 to 10.0 (˚C); 1 to 18 (˚F) 0.7 °C Lower Rail Limit CONTROL Lamp & LOW Lamp 0.1 to 10.0 (˚C); 1 to 18 (˚F) -0.7 °C
(HBR and HLR models)
6.5 °C (HPR models)
2.0 °C (HLR and HPR models)
6.4 Set Temperature Units
Note
If temperature units are changed, the temperature setpoints, o󰀨sets and alarm settings must be recalibrated.
1. Press and hold the Up and Down Arrows simultaneously for 3 seconds to enter program mode.
2. The LED will display .C or .F to indicate Celsius or Fahrenheit.
3. Press and hold the SET button while pressing the Up or Down Arrow to select the desired temperature unit.
4. Release SET button. The new setting is saved.
5. Press and hold the Up and Down Arrows simultaneously for 3 seconds to exit program mode.
6.5 Set Alarm Setpoints (Parameters)
1. Press and hold the Up and Down Arrows simultaneously for 3 seconds to enter program mode.
2. The LED Display will show .C or .F to indicate Celsius or Fahrenheit.
3. Press MODE until HIGH TEMP or LOW TEMP and MONITOR lamps ash.
4. Hold SET, then press Up or Down Arrow to change the setpoint.
5. Release SET button. The new setting is saved.
6. Press and hold Up and Down Arrows simultaneously for 3 seconds to exit program mode.
6.6TemperatureCalibrationO󰀨sets
Temperature calibration o󰀨sets indicate an acceptable margin of error between the actual temperature value and the desired
temperature value.
Horizon Series
Monitor Oset
Value is factory-set to match a calibrated reference thermometer. ♦ Refer to the service manual for instructions regarding changing the Monitor O󰀨set.
360397/A 19
Helmer Scientic i.Series® and Horizon Series™ Refrigerator - Undercounter Operation Manual
Control Sensor Oset and Rail Limits
The control sensor a󰀨ects the reading of the control probe temperature and therefore the actual temperature of the
refrigerator. This should not be adjusted from the original setting unless directed by Helmer Technical Service. The Upper and Lower Rail Limits help control the refrigeration based on the control probe temperature reading and the set
point. These limit values should not be changed from the default setting unless directed by Helmer Technical Service.
NOTICE
Control Sensor O󰀨set and Rail Limits are factory-preset and should not be changed. Contact Helmer T
echnical Service for
instructions regarding changing these values.
6.7 Active Alarms
The controller displays temperature and alarm information.
Table 3. Horizon Series Active Alarms
Alarm Visual Indicator Description
High Temperature HIGH TEMP lamp ashes Chamber temperature reading is above high temperature
Low Temperature LOW TEMP lamp ashes Chamber temperature reading is below low temperature
Display/Control Board Communication Error Er04 Display board fails to communicate with the control board Control Board to Compressor Inverter Error Er05 Communication loss from control board to compressor inverter Power Failure ”PoFF” appears on display Power to unit has been disrupted Primary Monitor Probe Failure (RTD1) Er01 Probe not functioning properly Control Probe Failure (RTD2) Er02 Probe not functioning properly No Battery Er06 Battery voltage is low Conguration Error Er07 Indicates that an EEPROM reading was corrupted or dip switch
Door Open < 3 min. DOOR ALARM lamp lights Door is open (less than three minutes) Door Open > 3 min. DOOR ALARM lamp ashes Door has been open 3 minutes or longer*
*Audible alarm will sound after door is open for 3 minutes.
alarm setpoint
alarm setpoint
settings on the control board have changed since last power-up
6.8 Mute and Disable Audible Alarms
Note
Muting audible alarms does not disable alarm lamps or signals sent through the remote alarm interface.
♦ Press Down Arrow (Mute) to mute audible alarms. ♦ To disable all audible alarms, insert the key in the Alarm Disable switch and turn.
6.9 Light Operation (optional)
The light switch is located on the monitoring and control panel and controls the LED light within the chamber.
360397/A 20
Horizon Series
Helmer Scientic i.Series® and Horizon Series™ Refrigerator - Undercounter Operation Manual
7 Horizon Series™ Access Control (Optional)
Allows user-specic secure access to the refrigerator.
Notes
• During a power failure, the optional Access Control lock will remain locked until battery power is depleted or until the
back-up battery key switch is switched OFF.
• During a power failure, switch the battery back-up switch OFF and use the mechanical door key to provide secure
storage for refrigerator contents.
• Refer to Horizon Series Access Control manual for complete information.
7.1 Setup
The Access Control keypad was programmed at the factory with a master code (0000). The master code is used to program the keypad and enter user codes.
Note
The master code should be changed to prevent unauthorized user code setup.
Enter unique user codes for up to 100 users. Each user code is stored with a specic record location number. Keep a log of the
location numbers and user codes with users’ names.
Add User Code
1. Enter the master code followed by the * (asterisk) key
2. Press 1 to initiate user code programming function
3. Enter the location number (00 - 99)
4. Enter the user code (4 - 8 digit number) followed by the # (pound) key
5. Press * (asterisk) to save changes and return to normal operation
Horizon Series
Delete User Code
1. Enter the master code followed by the * (asterisk) key
2. Press 1 to initiate delete user code programming function
3. Enter the location number (00 - 99) followed by the # (pound) key
4. Press * (asterisk) to save changes and return to normal operation
Open Refrigerator with Access Control
1. Enter the user code
2. Press # (pound) key
360397/A 21
Helmer Scientic i.Series® and Horizon Series™ Refrigerator - Undercounter Operation Manual
8 ProductSpecications
8.1 Operating Standards
These units are designed to operate under the following environmental conditions:
♦ Indoor use only ♦ Altitude (maximum): 2000 m ♦ Ambient temperature range: 15 °C to 32 °C ♦ Relative humidity (maximum for ambient temperature): 80% for temperatures up to 31 °C; 76% at 32°C (90°F) ♦ Temperature control range: 2 °C to 10 °C ♦ Overvoltage Category II ♦ Pollution Degree 2 ♦ Sound level is less than 70 dB(A).
Table4.ElectricalSpecications(Laboratory,BloodBank,andPharmacy)
105
Input Voltage and Frequency 115V 60 Hz 220-240V 50/60 Hz Voltage Tolerance ±10% Circuit Breakers 4.0A Current Draw* 1.4A 0.85A Power Source Grounded outlet, meeting national electric code (NEC) in
the U.S. and local electrical requirements in all locations.
Remote Alarm Capacity 115V or 230V: 1 A at 33 V (AC) RMS or 70 V (DC)
quantity 2
* Amperage values are subject to change. Refer to the product specication label on your unit for current values.
CAUTION
The interface on the remote alarm monitoring system is intended for connection to the end user’s central alarm
system(s) that uses normally-open or normally-closed dry contacts.
If an external power supply exceeding 33V (RMS) or 70V (DC) is connected to the remote alarm monitoring system’s
circuit, the remote alarm will not function properly; may be damaged; or may result in injury to the user.
360397/A 22
Helmer Scientic i.Series® and Horizon Series™ Refrigerator - Undercounter Operation Manual
Table5.RefrigeratorSpecications
Model Voltage Code Amps*
iLR105-GX
iBR105-GX
iPR105-GX
HLR105-GX
HBR105-GX
HPR105-GX
* Exterior dimensions include handle and leveling feet at lowest level.
115V 60 Hz 1.4 220-240V 50/60 Hz 0.85 115V 60 Hz 1.4 220-240V 50/60 Hz 0.85 115V 60 Hz 1.4 220-240V 50/60 Hz 0.85 115V 60 Hz 1.4 220-240V 50/60 Hz 0.85 115V 60 Hz 1.4 220-240V 50/60 Hz 0.85 115V 60 Hz 1.4 220-240V 50/60 Hz 0.85
Cu. Ft/
Liters Cabinet Door
5.3
(150)
5.3
(150)
5.3
(150)
5.3
(150)
5.3
(150)
5.3
(150)
Notes
• The maximum height added with leveling feet or casters installed is 2" (51 mm).
• The maximum weight added per shelf is 6 lbs (3 kg)
• The maximum weight added per ventilated drawer is 10 lbs (5 kg)
• The maximum weight added per drawer with locking lid is 15 lbs (7 kg)
• The maximum weight added per standard drawer is 12 lbs (6 kg) - i.Series; 11 lbs (5 kg) - Horizon Series
• Maximum load per shelf or drawer - 100 lbs (46 kg).
Net weight may vary depending on storage conguration. Weight listed in the table reects standard conguration for
each model.
Undercounter
Undercounter
Undercounter
Undercounter
Undercounter
Undercounter
Single hinged
solid
Single hinged
solid
Single hinged
solid
Single hinged
solid
Single hinged
solid
Single hinged
solid
Dimensions W x H x D in. (mm) Net Wt.
Exterior*
24 x 31.9 x 28.1
(610 x 809 x 714)
24 x 31.9 x 28.1
(610 x 809 x 714)
24 x 31.9 x 28.1
(610 x 809 x 714)
24 x 31.9 x 27.8
(610 x 809 x 705)
24 x 31.9 x 27.8
(610 x 809 x 705)
24 x 31.9 x 27.8
(610 x 809 x 705)
lbs (kg)
174 (79)
186 (85)
182 (83)
174 (79)
184 (84)
182 (83)
360397/A 23
Helmer Scientic i.Series® and Horizon Series™ Refrigerator - Undercounter Operation Manual
9 Compliance
9.1 Safety Compliance
This device complies with the requirements of directive 93/42/EEC concerning Medical Devices, as amended
0086
This product is certied to applicable UL and CSA standards by a NRTL.
9.2 Environmental Compliance
by 2007/47/EC.
This device complies with FCC Radiated and Conducted Emissions Approval to CFR47, Part 15; Class A levels
This device complies with the 2011/65/EU Directive for the Restriction of Hazardous Substances (RoHS).
This device falls under the scope of Directive 2102/19/EU Waste Electrical and Electronic Equipment (WEEE) .
When disposing of this product in countries a󰀨ected by this directive:
♦ Do not dispose of this product as unsorted municipal waste. ♦ Collect this product separately. ♦ Use the collection and return systems available locally.
For more information on the return, recovery, or recycling of this product, contact your local distributor.
Helmer Scientic 14400 Bergen Blvd. Noblesville, Indiana 46060 USA
EC REP
Emergo Europe Prinsessegracht 20 2514 AP The Hague The Netherlands
360397/A 24
Helmer Scientic i.Series® and Horizon Series™ Refrigerator - Undercounter Operation Manual
Appendix A: Parts
i.Series Parts
A
G
H
I
M
R
B C D
E
S
J
K
L
F
U
T
N O
P Q
XWV
Y
Letter Description Letter Description
3
A i.C B Backup battery key switch O RJ45 Ethernet port C AC On/O󰀨 switch P USB Port D Battery switch Q AC output power cord receptacle E Circuit breakers R Rear access port F Casters S iC G Unit cooler with fan guard T iC H Primary monitor probe U iC I Probe bottle V Compressor J Solid ballast K Ventilated shelf W Condenser fan motor L Door switch X Condensing unit M Key lock Y Condenser temperature probe
user interface N Remote alarm contacts
360397/A 25
Not Shown
3
power supply board
3
monitoring system backup battery
3
control oard
Compressor Inverter
Helmer Scientic i.Series® and Horizon Series™ Refrigerator - Undercounter Operation Manual
Appendix B: Parts
Horizon Series Parts
A
I
B
C D
E
F G
H
S T
U
J K
O
L
M
N
P
Q
X
WV
Y
R
Letter Description Letter Description
A Temperature Chart Recorder N Door switch B Display board O Remote alarm contacts C Backup battery key switch P AC output power cord receptacle D AC On/O󰀨 switch Q Casters E Chamber light switch R Rear access port F Circuit breakers S Back-up battery (optional Access Control) G Monitoring system backup battery T Power supply board H Alarm display key switch U Control board I Unit cooler with fan guard V Compressor inverter J Primary probe W Compressor K Probe bottle X Condenser fan motor L Solid ballast Y Condensing unit M Ventilated shelf
END OF MANUAL
360397/A 26
Not Shown
Condenser temperature probe
Helmer Scientic
14400 Bergen Boulevard, Noblesville, IN 46060 USA