HELIX 9600 Installation, Operation And Maintenance Instructions

Page 1
9600 (High Voltage) Compressor Installation,
Operation, and Maintenance Instructions
©2007 Helix Technology Corporation Pub. No. 8040437, Rev. 104, 06/12/2007 ECO No.18068 Printed in USA
HELIX TECHNOLOGY CORPORATION
http://www.helixtechnology.com
Page 2
Page 3
The information in this document is believed to be accurate and reliable. However, Helix Technology Corporation, cannot accept any financial or other responsibilities that may
Helix Technology Corporation reserves the right to change any or all information contained herein without prior written notice. Revisions may be issued at the time of such changes and/or deletions.
Any duplication of this manual or any of its parts without expressed written permission from Helix Technology Corporation is strictly prohibited.
Any correspondence regarding this document should be forwarded to:
Helix Technology Corporation Mansfield Corporate Center Nine Hampshire Street Mansfield, Massachusetts 02048-9171 U.S.A.
Telephone: (508) 337-5000 FAX: (508) 337-5464
The following Helix Technology Corporation trademarks and service marks may appear in this document:
Conductron
Cryogenerator
GOLDLink
®
®
®
Helix Technology.. Your Vacuum Connection
On-Board
Stabil-Ion
Vacuum Assurance
®
®
SM
Convectron
Cryo-Torr
Granville-Phillips
Micro-Ion
SM
RetroEase
ThinLine™
®
®
®
®
®
Cryodyne
CTI-Cryogenics
GUTS
Mini-Convectron
RetroFast
TurboPlus
®
®
®
®
®
®
Cryogem
FastRegen™
Helix
Mini-Ion™
Stabil-1
TrueBlue
®
®
®
SM
All other trademarks or registered trademarks are the property of their respective holders.
©2007 Helix Technology Corporation Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 iii
Page 4
Page 5

Table of Contents

Table of Contents

Figures

Tables

Safety Conventions

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . S-1
Signal Word Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . S-1
Safety Shape Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . S-2
References . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . S-2

Section 1 - 9600 Compressor Description

General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Compressor Configurations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
System Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
CTI-CRYOGENICS Helium Refrigeration System . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Electrical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Cooling Water . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Component Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
System Circuit Breaker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9
Elapsed Time Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9
Control Circuit Breaker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9
Power ON Indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9
Gas Charge Flared Fitting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9
Helium Pressure Gauge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10
Power Inlet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10
Return Gas Coupling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10
Cooling Water IN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10
Cooling Water OUT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10
Cryopump Electrical Outlet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10
Compressor Remote Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-11
Supply Gas Coupling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-11
©2007 Helix Technology Corporation Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 v
Page 6
Multiple On-Board Cryopump Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-12
Multiple Cryo-Torr Cryopump Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-13

Section 2 - Unpacking and Inspection

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Shipping Carton Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Removal from Shipping Carton . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Compressor Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Compressor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Helium "OFF" Condition Pressure Verification . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Shipping Carton Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2

Section 3 - Installation

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Supply and Return Water Line Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Hard Water Lines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Flexible Water Lines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Electrical Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Phase Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Connecting/Disconnecting Helium Flex Lines . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Connecting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Disconnecting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Single On-Board Cryopump Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Multiple On-Board Cryopump Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Helium Line Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Power Cable Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
Multiple Cryo-Torr Cryopump Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
Helium Line Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
Power Cable Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14

Section 4 - Operation

Adjusting System Helium Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
"OFF" Condition Helium System Pressure Verification . . . . . . . . . . . . . . . . . . . . . 4-1
Compressor Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Replacement of Helium Circuit Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3

Section 5 - Maintenance

Scheduled Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Suggested Maintenance Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Adsorber Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Adjusting System Helium Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
vi Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 ©2007 Helix Technology Corporation
Page 7
Reducing Helium Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
Increasing Helium Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Adding Helium . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6

Appendix A - Customer Support Information

Customer Support Center Locations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
Guaranteed Up-Time Support (GUTS®) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
Product Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
E-mail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1

Appendix B - Flow Diagram

Appendix C - Troubleshooting Procedures

Troubleshooting the Compressor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-1
Technical Inquiries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-1

Appendix D - Schematic

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-1

Index

©2007 Helix Technology Corporation Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 vii
Page 8
viii Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 ©2007 Helix Technology Corporation
Page 9

Figures

Figure 1-1: 9600 Compressor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Figure 1-2: 9600 Compressor Dimensions. . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Figure 1-3: Water Flow Rate versus Pressure Drop . . . . . . . . . . . . . . . . . . . . . 1-6
Figure 1-4: 9600 Compressor Rear View Component Locations. . . . . . . . . . . 1-8
Figure 1-5: 9600 Compressor Connected to Multiple On-Board Cryopumps 1-12
Figure 1-6: 9600 Compressor Connected to Multiple Cryo-Torr Cryopumps 1-13
Figure 3-1:9600 Compressor Installation Flowchart . . . . . . . . . . . . . . . . . . . . 3-2
Figure 3-2: 9600 Compressor Circuit Breaker Terminals (Cover Removed) . 3-5
Figure 3-3:Connecting/Disconnecting Helium Flex Line Couplings. . . . . . . . 3-8
Figure 3-4: Single On-Board Cryopump Connections . . . . . . . . . . . . . . . . . . . 3-9
Figure 3-5: Single Cryo-Torr Cryopump Installation. . . . . . . . . . . . . . . . . . . 3-10
Figure 3-6: Recommended Multiple On-Board Cryopump or Waterpump Installation
(Splitter Box located at Process Tool) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
Figure 3-7: Alternative Multiple On-Board Cryopump or Waterpump Installation
(Splitter Box located at Compressor) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13
Figure 3-8: Multiple Cryo-Torr Cryopump Installation . . . . . . . . . . . . . . . . . 3-15
Figure 3-9: Multiple Cryo-Torr Cryopump Installation . . . . . . . . . . . . . . . . . 3-16
Figure 5-1: Disconnecting Self Sealing Couplings. . . . . . . . . . . . . . . . . . . . . . 5-2
Figure 5-2: Adsorber Location in 9600 Compressor (Rear Panel Removed). . 5-4
Figure 5-3: Helium Pressure Control Components. . . . . . . . . . . . . . . . . . . . . . 5-5
Figure B-1: 9600 Compressor Flow Diagram P/N 8135001 Rev. A . . . . . . . . B-2
Figure D-1: 9600 High Voltage Compressor . . . . . . . . . . . . . . . . . . . . . . . . . . D-4
©2007 Helix Technology Corporation Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 ix
Page 10
x Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 ©2007 Helix Technology Corporation
Page 11

Tables

Table 1-1:Compressor Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Table 1-2:Electrical Input Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Table 1-3:Cooling Water Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Table 1-4:General Compressor Operating Specifications . . . . . . . . . . . . . 1-7
Table 1-5:Cryopump Electrical Outlet Pin Assignments . . . . . . . . . . . . . 1-11
Table 1-6:Compressor Remote Connector Pin Assignments . . . . . . . . . . 1-11
Table 4-1:9600 Compressor Helium ("OFF" Condition) Charge. . . . . . . . 4-1
Table 5-1:Suggested Maintenance Equipment . . . . . . . . . . . . . . . . . . . . . . 5-1
Table C-1:Compressor Troubleshooting Procedures . . . . . . . . . . . . . . . . . C-2
Table D-1:Basic Control Assembly Legend. . . . . . . . . . . . . . . . . . . . . . . . D-1
Table D-2:On-Board/Cryo Electrical Module P/N 8135127G002 Legend D-2
Table D-3:On-Board Output Module P/N 8135148G001 Legend . . . . . . . D-3
©2007 Helix Technology Corporation Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 xi
Page 12
xii Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 ©2007 Helix Technology Corporation
Page 13

Safety Conventions

Introduction

Follow all safety precautions during installation, normal operation, and when servicing CTI-Cryogenics products.
This chapter explains the safety conventions used throughout this manual. CTI-Cryogenics uses a specific format for cautions and warnings, which includes standard signal words and safety shapes.
See also the Customer Support appendix or call your local Customer Support Center for assistance.

Signal Word Descriptions

All cautions and warnings contain signal words, which call attention to safety messages and designate the degree of hazard seriousness. The following table shows the signal words and their meanings that may be used in this document.
Term Example Definition
CAUTION
CAUTION
WARNING
A signal word accompanied by a safety shape that indicates a potentially hazardous situation or unsafe practice.
If not avoided, the action may result in minor
or moderate personal injury or equipment damage. A CAUTION is highlighted in
yellow. A signal word that indicates a situation or
unsafe practice, which if not avoided may result in equipment damage. A CAUTION is highlighted in yellow.
A signal word accompanied by a safety shape that indicates indicates a potentially hazardous situation.
If not avoided, the action may result in serious injury or death. A WARNING is highlighted in orange.
©2007 Helix Technology Corporation Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 S-1
Page 14

Safety Shape Descriptions

Safety Shape Descriptions
Example Term Shape Definition
All cautions and warnings contain safety shapes, which have specific safety meanings. The following table shows some of the safety shapes used in this document and their meanings.

References

General Warning
High Voltage Indicates a high voltage hazard.
Hot Surface
For more information about safety standards, see the following documents:
Indicates a general hazard. Details about this hazard appear in the safety notice explanation.
Indicates a surface is hot enough to cause discomfort or a burn.
ISO 7010: 2003(E), Graphic symbols - Safety colours and safety signs - Safety signs used in workplaces and public areas
ISO 3864-1: 2002(E), Graphic symbols - Safety colours and safety signs - Part 1: Design principles for safety signs in workplaces and public areas
S-2 Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 ©2007 Helix Technology Corporation
Page 15

Section 1 - 9600 Compressor Description

General

This manual provides the information required to install, operate, and maintain the CTI-CRYOGENICS 9600 Compressor .
NOTE: All personnel with installation, operation, and maintenance responsibilities should become familiar with the contents of both the 9600 Compressor Installation, Operation, Maintenance, and appropriate cryopump manuals to ensure safe, high quality, and reliable system performance.
CAUTION
Refer to “Appendix A - Customer Support Information” to contact the local Customer Support Center for information on connecting 9600 Compressors to a manifold with other CTI-CRYOGENICS compressors.

Compressor Configurations

The 9600 Compressor supports either On-Board or Cryo-Torr Cryopumps. For multiple cryopump installations, an On-Board Splitter Box or Cryo­Torr Interface can be used for cold head power distribution that reduces total cable requirements as shown in Figure 1-5 and Figure 1-6.

System Documentation

The manuals for a system cover two basic components: the cryopump and the Compressor. A manual is shipped with each system component to provide information for installation and operation of that component. A loose-leaf binder with index tab separators is also provided so you can compile a complete set of tabulated manuals.

CTI-CRYOGENICS Helium Refrigeration System

The operation of CTI-CRYOGENICS’ cryopumps is based upon a closed loop helium expansion cycle. The system is made up of two major
©2007 Helix Technology Corporation Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 1-1
Page 16
CTI-CRYOGENICS Helium Refrigeration System
components: the cryopump, which contains the cold head, and the helium Compressor which compresses the helium gas.
Refrigeration is produced in the cryopump cold head through periodic expansion of high pressure helium in a regenerative process. The high pressure helium is provided by the Compressor. Low pressure helium returning from the cold head is compressed into the necessary high pressure to be returned to the cold head. The energy required to compress the helium is rejected as heat through the cooling water.
High pressure room temperature helium is transferred to the cold head through the supply lines. After expansion, low pressure helium is returned to the Compressor (at or near room temperature) to repeat the cycle in a closed loop fashion. Large separation distances can be accommodated between the Compressor and the cryopump.
In the Compressor, helium is compressed using a highly reliable oil lubricated commercial Compressor. Helium purification takes place via several stages of oil removal. The final stage of purification is performed with a replaceable adsorber cartridge. In order to maintain peak efficiency, the adsorber must be replaced every three years. The 9600 Compressor is shown in Figure 1-1.
1-2 Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 ©2007 Helix Technology Corporation
Page 17
0
0
R
6
9
O
S
0
R
ES
0
6
R
O
9
S
P
S
M
E
R
CO
P
M
O
C
Figure 1-1: 9600 Compressor
©2007 Helix Technology Corporation Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 1-3
Page 18

Specifications

Specifications
Dimensions
The dimensions of the Compressor are shown in Figure 1-2.
19.52 in. (496 mm)
14.4 in. (366 mm)
21.50 in. (546 mm)
23.36 in. (593 mm)
3.18 in. (81 mm)
Figure 1-2: 9600 Compressor Dimensions
1.28 (32mm)
CAUTION
Do not place a weight greater than 75 lbs. (34Kg) on top of the Compressor.
1-4 Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 ©2007 Helix Technology Corporation
Page 19
Weight
Electrical
The weight of the Compressor is listed in Table 1-1.
Table 1-1: Compressor Weight
Weight lbs./kg
267/121kg
The electrical specifications of the Compressor are listed in Table 1-2.
Table 1-2: Electrical Input Specifications
Parameter Value
Operating Voltage Range/Frequency 342-457V/50 Hz
342-506V/60 Hz
Phase 3
Nominal Input Power 5.5 KW/60 Hz
5KW/50 Hz Nominal Power Factor 0.85 Rated FL/LR* Current 8/40
Minimum Electrical Service 15 Amps
*FL/LR = Full Load/Locked Rotor
Cooling Water
The water used to cool the Compressor must meet the specifications shown in Table 1-3 for proper system operation.
Table 1-3: Cooling Water Specifications
Parameter Value
Maximum Inlet Temperature 90°F (32°C)
Minimum Inlet Temperature 50°F (10°C)
Flow Rate 2.75 ±1.25 gpm (10.4 ± 4.7 lpm)
NOTE: Water conditioning may be required for ap plications not meetin g these requirements.
©2007 Helix Technology Corporation Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 1-5
Page 20
Specifications
Table 1-3: Cooling Water Specifications (Continued)
Parameter Value
Pressure Drop (inlet-to-outlet)
See Figure 1-3
Maximum Inlet Pressure 100 psi (6.9 bars)
Alkalinity 6.0 - 8.0 pH
Calcium Carbonate < 75 ppm
NOTE: Water conditioning may be required for ap plications not meetin g these requirements.
WATER PRESSURE DROP (Bars)
.07 .49
.21
.35
.63
4.00
3.50
E
N
I
L
3.00
2.50
2.00
1.50
WATER FLOW RATE (GPM)
AC
C
EPT
E
P
O
E
BL
A
G
N
I
T
A
R
.77
15.2
13.3
11.40
9.50
7.60
5.70
WATER FLOW RATE (LPM)
1.00
0.50 1
NOTE: Figure 1-3 defines the water flow rate through the Compressor as a function of the pressure drop from
water inlet to water outlet. You must provide the correct pressure drop in your water supply system to ensure that
3
5
WATER PRESSURE DROP (PSID)
7
9
3.80
1.90
11
-
Figure 1-3: Water Flow Rate versus Pressure Drop
General
The information in Table 1-4 provides general Compressor operating specifications.
1-6 Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 ©2007 Helix Technology Corporation
Page 21
Table 1-4: General Compressor Operating Specifications
Specification Values Part Number 8135901G001
Input Power Cable
(Customer Supplied)
Nominal Helium Pressure
Ambient Operating Tem-
perature Range
Interface
Gas Supply Connector
Gas Return Connector
Remote Control Receptacle
Adsorber Service Schedule 3 Years
600 VAC 12 Gauge, 3 conductor wire with ground Must conform to local electrical codes
Refer to 50 - 100º F (10 - 38º C)
Cryopump Power Receptacles: mates with the CTI-CRYOGENICS supplied cryopump power cable for single pump use.
Mates with remote junction box power cable for multiple cry­opump use.
1/2 in. Aeroquip self-sealing coupling
1/2 in. Aeroquip self-sealing coupling
24VAC, 2.7A inductive mates with P5 connector Type MS3106A*
Table 4-1
* Supplied by CTI-CRYOGENICS
NOTE: The 9600 Compressor is designed for continuous operation and should remain ON when the cryopumps are in a regeneration cycle.
©2007 Helix Technology Corporation Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 1-7
Page 22

Component Description

Component Description
The components of the 9600 Compressor that are accessible from the rear panel are shown in Figure 1-4 and described in the following paragraphs.
Elapsed Time Meter
Control Circuit Breaker
Gas Charge Flared Fitting
Helium Pressure Gauge
Gas Charge Control Valve
Return Gas Coupling
Power Inlet
Power ON Indicator
System Circuit Breaker
Cryopump Electrical Outlet
Compressor Remote Connector
Supply Gas Coupling
Rear Panel
Cooling Water IN
Cooling Water OUT
Figure 1-4: 9600 Compressor Rear View Component Locations
1-8 Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 ©2007 Helix Technology Corporation
Page 23
System Circuit Breaker
The System Circuit Breaker protects main input power to the Compressor pump and module. The circuit breaker positions are labeled ON (1), which is in the UP position, and OFF (0), which is in the DOWN position.
NOTE: The phase monitor in the Compr essor will cause the system cir cuit breaker to open when input power phases are incorrect.
Elapsed Time Meter
The Elapsed Time Meter records the number of Compressor operating hours. Since the meter is digital, it is not illuminated unless the system circuit breaker is in the ON position and power is connected to the Compressor. The Elapsed Time Meter maintains the correct accumulated operating hours while system power is turned OFF.
NOTE: The meter cannot be reset.
Control Circuit Breaker
The Control Circuit Breaker provides current overload protection for all internal components of the Compressor except the Compressor motor. The Compressor motor is protected by a separate overload protector. The Control Circuit Breaker opens automatically and must be reset manually.
Power ON Indicator
The Power On Indicator illuminates when the system circuit breaker is placed in the ON position. The Compressor pump is energized when the power indicator is illuminated and the elapsed time meter records system operation time.
Gas Charge Flared Fitting
The Gas Charge Flared Fitting is used to connect a 99.999% pure helium supply to the Compressor when helium charging is required. The fitting has a 45º flare and 7/16 in. x 20 threads/inch. Refer to “Section 5 -
Maintenance” for information on adding helium to the Compressor.
©2007 Helix Technology Corporation Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 1-9
Page 24
Component Description
Helium Pressure Gauge
Power Inlet
Return Gas Coupling
The Helium Pressure Gauge indicates system ("OFF" Condition) helium charge pressure when the Compressor and cryopumps are OFF and Compressor suction or inlet pressure when the Compressor is ON. Refer to
Table 4-1 for the appropriate ("OFF" Condition) helium charge pressure.
The Power Inlet is used to connect your power cable to the Compressor. Refer to “Section 3 - Installation” for information on power cable installation.
The Return Gas Coupling returns the helium, which has been cycled through the cryopump, back to the Compressor. Refer to “CTI-
CRYOGENICS Helium Refrigeration System” in this section for more
information.
Cooling Water IN
The Cooling Water IN connector provides water to the Compressor from your facility to cool the Compressor during operation. The connector thread size is a 1/2 in. female pipe thread. The water must meet the specifications outlined in Table 1-3. Refer to “Section 3 - Installation” for more information on cooling water connections.
Cooling Water OUT
The Cooling Water OUT connector returns the water that has been used to cool the Compressor to your facility. The connector thread size is a 1/2 in. female pipe thread. Refer to “Section 3 - Installation” for more information on cooling water connections.
Cryopump Electrical Outlet
The Cryopump Electrical Outlet provides power to a single On-Board or Cryo-Torr Cryopump, an On-Board Splitter Box, or a Cryo-Torr Interface. The Compressor requires the use of an On-Board Splitter Box or Cryo-T orr Interface for multiple cryopump system connections. Refer to T able 1-5 for connector pin identification. Refer to “Multiple On-Board Cryopump
Connections” or “Multiple Cryo-Torr Cryopump Connections” in this
1-10 Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 ©2007 Helix Technology Corporation
Page 25
section for more information.
Table 1-5: Cryopump Electrical Outlet Pin Assignments
Identifier Function
A and B Heater Power - 208 VAC nominal
C Center tap for D and E
D and E 24 VCT @ 4.6 Amps
F-G and G-H Cold Head Voltage Output 130-160 VAC @ 4.5 Amps
J Chassis Ground
K Not Used
Compressor Remote Connector
The Compressor Remote Connector is a 14 pin connector that can be used in conjunction with the On-Board setpoint relays, relays in the Cryo-Torr Interface, or a signal from the vacuum system to turn the Compressor ON or OFF. Refer to Table 1-6 for connector pin identification. Switching contacts must be rated at 24VAC, 2.7A inductive.
NOTE: The Compr essor is shipped with a mating plug which must remain installed in the Compressor Remote Connector to ensure Compressor operation when the Compressor remote feature is not being used.
Table 1-6: Compressor Remote Connector Pin Assignments
Identifier Function
D and E Compressor Remote Control
F, G, I, J Answer Back
A,B,C,H,K,L,M,N Open
Supply Gas Coupling
The Supply Gas Coupling provides a connection for high pressure compressed helium to the cryopump cold head. Refer to “CTI-
CRYOGENICS Helium Refrigeration System” in this section for more
information.
Make = ON, Break = OFF
©2007 Helix Technology Corporation Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 1-11
Page 26

Multiple On-Board Cryopump Connections

Multiple On-Board Cryopump Connections
The On-Board Splitter Box permits the connection of multiple On-Board Cryopumps or Waterpumps to one 9600 Compressor as shown in Figure 1-
5. Refer to “Section 3 - Installation” for more information on connecting
single or multiple On-Board Cryopumps or Waterpumps to the Compressor.
WATER
SUPPLY
REMOTE
POWER
RETURN
On-Board Splitter Box
On-Board Cryopump Power Cable
Figure 1-5: 9600 Compressor Connected to Multiple On-Board Cryopumps
NOTE: Your installation (number of pumps per compressor) will vary based upon the On-Board Cryopumps or Waterpumps used. Refer to
“Appendix A - Customer Support Information” to consult your local CTI-
CRYOGENICS Customer Support Center for information on specific compressor/pump applications.
1-12 Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 ©2007 Helix Technology Corporation
Page 27

Multiple Cryo-Torr Cryopump Connections

The Cryo-Torr Interface permits the connection of multiple Cryo-Torr Cryopumps to one Compressor as shown in Figure 1-6.
Refer to “Section 3 - Installation” for more information on connecting single or multiple Cryo-Torr Cryopumps to the Compressor.
SUPPLY
RETURN
Cryo-Torr Interface
Remote
WATER
POWER
Cryopump Electrical Input
User Remote (from vacuum system)
Figure 1-6: 9600 Compressor Connected to Multiple Cryo-Torr Cryopumps
NOTE: Your installation (number of pumps per compressor) will vary based upon the Cryo-Torr Cryopump models used. Refer to “Appendix A -
Customer Support Information” to consult your local CTI-CRYOGENICS
Customer Support Center for information on specific compressor/ cryopump applications.
©2007 Helix Technology Corporation Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 1-13
Page 28
Multiple Cryo-Torr Cryopump Connections
1-14 Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 ©2007 Helix Technology Corporation
Page 29

Section 2 - Unpacking and Inspection

Introduction

The 9600 Compressor is shipped in a shipping carton incorporating a ramp system which makes removing the Compressor from the carton safe and easy.

Shipping Carton Inspection

Inspect the exterior of the shipping carton for visible signs of damage before opening the shipping carton. Report any damage to the shipping company at once.

Removal from Shipping Carton

1. Cut the two straps on the exterior of the shipping pallet.
2. Lift the cardboard carton straight up and remove it from the pallet.
3. Cut the tape which holds the ramp in the vertical position.
4. Swing the ramp down until the end touches the floor.
5. Remove any excess shipping material from around the Compressor .
Maintain control over the movement of the Compressor as it rolls down the ramp. Injury to personnel may result if the Compressor is allowed to roll freely down the ramp.
CAUTION
Maintain control over the movement of the Compressor as it rolls down the ramp. Damage to the Compressor may result if the Compressor is allowed to roll freely down the ramp.
©2007 Helix Technology Corporation Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 2-1
Page 30

Compressor Inspection

Compressor Inspection
Compressor
Helium "OFF" Condition Pressure Verification
6. Carefully roll the Compressor down the ramp and onto the floor.
Inspect the Compressor for visible signs of damage as indicated in the following paragraphs.
Inspect the exterior of the Compressor for visible signs of damage, evidence of an oil leak, and check the Helium Pressure Gauge for proper helium pressure. Report any damage to the shipping company at once.
Refer to “Section 4 - Operation” for more information on the ("OFF" Condition) helium charge pressure of the 9600 Compressor.
Shipping Carton Contents
The shipping carton should contain the following items:
Compressor
Two barbed fittings for flexible water lines
Installation, Operation, and Maintenance manual
Compressor remote start connector and strain relief
2-2 Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 ©2007 Helix Technology Corporation
Page 31

Introduction

Section 3 - Installation

Section 3 provides you with the information required to install the 9600 Compressor and connect it to single or multiple On-Board or Cryo-Torr Cryopump configurations.
Figure 3-1 highlights the major tasks for Compressor installation and refers
to the appropriate installation procedures in this section.
©2007 Helix Technology Corporation Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 3-1
Page 32
Introduction
START
Supply and Return Water Line Connections
(page 3-3)
Electrical Connections (page 3-4)
Phase Check (page 3-6)
Connect/Disconnect Helium Flexlines (page 3-7)
Single Cryopump Connections (page 3-8)
Single or Multiple Cryopump Installation?
Multiple Cryopump Connections (page 3-11)
See
“Section 4 - Operation”
END
Figure 3-1: 9600 Compressor Installation Flowchart
3-2 Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 ©2007 Helix Technology Corporation
Page 33

Supply and Return Water Line Connections

NOTE: The water used for cooling the Compressor must meet the specifications outlined in “Section 1 - 9600 Compressor Description”.
Hard Water Lines
1. Apply a light coating of standard plumbing thread sealant to the hard line pipe threads.
2. Install the Supply hard line into the INPUT connection on the rear panel of the Compressor. Tighten the fitting by hand.
3. Install the Return hard line into the OUTPUT connection on the rear panel of the Compressor. Tighten the fitting by hand.
CAUTION
Do not overtighten the ferrules. Damage to the input and output connector threads may occur.
4. Using a wrench, tighten the fittings.
5. Allow water to flow and check for leaks at the rear of the Compressor.
Flexible Water Lines
1. Apply a light coating of standard plumbing thread sealant to the barbed fitting threads.
2. Install the barbed fittings into the INPUT and OUTPUT connections on the rear panel of the Compressor.
Do not overtighten the barbed fittings. Damage to the INPUT and OUTPUT connector threads may occur.
3. Using a wrench, tighten the barbed fittings.
4. Connect the Supply flexible water line to the INPUT barbed fitting and secure with a hose clamp.
5. Connect the Return flexible water line to the OUTPUT barbed fitting and secure with a hose clamp.
CAUTION
6. Allow water to flow and check for leaks at the rear of the Compressor.
©2007 Helix Technology Corporation Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 3-3
Page 34

Electrical Connections

Electrical Connections
Follow all local high voltage safety precautions when performing this procedure to reduce the possibility of electrical shock. Make sure all electrical power is OFF before proceeding with this procedure.
The following procedures provide information for making all three phase (180 - 250 VAC) electrical connections to the Compressor.
Power Cable Preparation
WARNING
CAUTION
The cable used for making the Compressor power cable must be 10 gauge, 3 conductor cable with ground rated at 600 VAC.
1. Cut a 10 AWG (6.00 mm2), 3 conductor cable with ground to an appropriate length.
2. Strip the cable jacket back 4 in. (101.6 mm).
3. Strip the insulation back 3/8 in. (9.3 mm) on each individual conductor.
4. Install a #10 ring tongue terminal on the end of each conductor using the appropriate size double crimping tool.
5. Remove the rear panel as shown in Figure 3-2.
6. Remove the circuit breaker terminal cover as shown in Figure 3-2.
7. Install the cable into the Compressor through the cable strain relief.
8. Remove the 10-32 nut and install the grounding wire on the ground stud. Install the nut and tighten to 18 in.-lbs (0.21m-kg).
NOTE: Use a slotted screwdriver which is capable of holding a screw when performing steps 9 and 10.
9. Remove the screws from the Compressor circuit breaker terminals X, Y, and Z as shown in Figure 3-2.
3-4 Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 ©2007 Helix Technology Corporation
Page 35
See Detail A
Circuit Breaker
Terminal Cover
Y
Z
Rear Panel
X
Three Phase Power
Detail A
Figure 3-2: 9600 Compressor Circuit Breaker Terminals (Cover Removed)
NOTE: The phase order in which the conductor terminal lugs are connected to circuit breaker terminals X, Y, and Z will be determined during the Phase Check Procedure.
NOTE: For installation where one of the three phase legs is at or near ground potential, connect that leg to terminal Y on the Compressor as shown in Figure 3-2.
10. Install the conductor terminal lugs to the circuit breaker terminals X, Y, and Z as shown in Figure 3-2.
11. Torque the screws to 12 in.-lbs (0.14m-kg).
12. Allow enough cable to stay in the electrical enclosure to prevent strain on the electrical connections and tighten the screws on the cable strain relief.
13. Install the power source end of the power cable according to the local electrical codes.
14. Install the circuit breaker terminal cover.
15. Proceed with Phase Check.
©2007 Helix Technology Corporation Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 3-5
Page 36
Phase Check
WARNING
Follow all high voltage safety precautions when performing this procedure to prevent the possibility of electrical shock.
1. Make sure power is applied to compressor circuit as described in Table 1-2.
NOTE: The cir cuit breaker will trip immediately during step 2 if the power phase connections are not correct.
2. Turn the Compressor circuit breaker to the ON position. If the circuit breaker trips, refer to step 3. If the circuit breaker does not trip, refer to step 4.
3. If the circuit breaker trips, perform the following steps: a. Turn the Compressor circuit breaker to the OFF position. b. Disconnect the power cord from the power source. c. Remove the circuit breaker terminal cover. d. Reverse the wiring order of Compressor circuit breaker
terminals X and Y. e. Torque the circuit breaker terminal screws to 12 in.-lbs. f. Install the circuit breaker terminal cover. g. Repeat steps 1- 2 of this procedure.
4. Install the rear panel.
5. Proceed with appropriate cryopump connections.
3-6 Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 ©2007 Helix Technology Corporation
Page 37

Connecting/Disconnecting Helium Flex Lines

CAUTION
Make sure the helium flex lines are connected and disconnected from the 9600 Compressor using the following procedure and as shown in
Figure 3-3. Failure to follow this procedure could damage connector
O-ring seals or cause a helium circuit leak.
Connecting
1. Remove all dust plugs and caps from the Gas Supply and Return lines, and the Compressor and cryopump Supply and Return connectors. Check for the presence of a flat gasket in the male connector, and no gasket in the female connector.
2. Connect the Gas Return line to the GAS RETURN connector on the rear of the Compressor and then to the GAS RETURN connector on the cryopump. Using two wrenches as shown in Figure 3-3, tighten the connector.
3. Connect the Gas Supply line to the GAS SUPPLY connector on the rear of the Compressor and then to the GAS SUPPLY connector on the cryopump. Using two wrenches as shown in Figure 3-3, tighten the connector.
Disconnecting
4. Attach the Supply and Return line identification labels to each end of the appropriate lines.
1. Using two wrenches as shown in Figure 3-3, disconnect the two self sealing coupling connectors quickly to minimize helium leakage.
©2007 Helix Technology Corporation Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 3-7
Page 38

Single On-Board Cryopump Connections

This 1 1/8 (or 1) in. wrench holds the coupling in a sta­tionary position.
HOLD WITH 1 1/8 INCH WRENCH
TURN WITH 1 3/16 INCH WRENCH
TO LOOSEN
Figure 3-3: Connecting/Disconnecting Helium Flex Line Couplings
This 1 3/16 in. wrench is used to turn the self sealing coupling con­nector.
TO TIGHTEN
TURN WITH 1 3/16 INCH WRENCH
HOLD WITH 1 INCH WRENCH
Single On-Board Cryopump Connections
CAUTION
Make sure the Compressor power is OFF before making any connections to the rear panel.
1. Connect the Supply and Return lines to the 9600 Compressor as described in “Connecting/Disconnecting Helium Flex Lines” in this section.
2. Connect one end of the Cryopump cold head cable to the CRYOPUMP ELECTRICAL OUTLET on the rear panel of the Compressor as shown in Figure 3-4.
3. Connect the opposite end of the cold head cable to the cryopump
3-8 Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 ©2007 Helix Technology Corporation
Page 39
cold head cable connector.
SUPPLY
RETURN
REMOTE
On-Board Cryopump Power Cable
WATER
POWER
Figure 3-4: Single On-Board Cryopump Connections
CAUTION
Make sure the Compressor power is OFF before making any connections to the rear panel.
1. Connect the Supply and Return lines to the 9600 Compressor as described in “Connecting/Disconnecting Helium Flex Lines” in this section.
2. Connect one end of the Cryopump cold head cable to the CRYOPUMP ELECTRICAL OUTLET on the rear panel of the Compressor as shown in Figure 3-5. Connect the opposite end of the cold head cable to the cryopump cold head cable connector as shown in Figure 3-5.
©2007 Helix Technology Corporation Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 3-9
Page 40
Single On-Board Cryopump Connections
SUPPLY
RETURN
REMOTE
Cryo-Torr Power Cable
WATER
POWER
Figure 3-5: Single Cryo-Torr Cryopump Installation
3-10 Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 ©2007 Helix Technology Corporation
Page 41

Multiple On-Board Cryopump Connections

CAUTION
Make sure the Compressor power is OFF before making any connections to the rear panel.
Helium Line Connections
CAUTION
The use of several compressors on a single manifold feeding a common supply header and a common return header requires special precautions. Contact CTI-CRYOGENICS for a review of the intended installation and for specific technical instructions.
The use of a 9600 compressor on a manifold with other CTI-CRYOGENICS compressor models requires a reduction of the helium charge pressure to 200 - 210 psig charge pressure to avoid helium safety valves from inadvertently venting. Refer to “Section 4 -
Operation” and “Section 5 - Maintenance” for more information.
1. Connect the Supply and Return lines to the 9600 Compressor as described in “Connecting/Disconnecting Helium Flex Lines” in this section.
2. Connect the Gas Return Line to the customer supplied helium manifold and then to the GAS RETURN connector on the On-Board Cryopump.
3. Connect the Gas Supply Line to the customer supplied helium manifold and then to the GAS SUPPLY connector on the On-Board Cryopump.
©2007 Helix Technology Corporation Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 3-11
Page 42
Multiple On-Board Cryopump Connections
Power Cable Connections
1. Connect the On-Board Splitter Box power cable between the CRYOPUMP ELECTRICAL OUTLET on the rear panel of the Compressor and the On-Board Splitter Box power connector as shown in Figure 3-6.
2. Connect the On-Board Cryopump or Waterpump power cables to the CRYOPUMP 1, 2, or 3 connectors on the On-Board Splitter Box and the respective cryopumps as shown in Figure 3-6.
NOTE: Your installation (number of pumps per compressor) will vary based upon the On-Board Cryopump models used. Refer to “Appendix A -
Customer Support Information” to consult your local CTI-CRYOGENICS
Customer Support Center for information on specific compressor/pump applications.
NOTE: The On-Board Splitter Box can be installed at the process tool containing the pumps as shown in Figure 3-6, or on the rear of the Compressor as shown in Figure 3-7. CTI-CRYOGENICS recommends that the On-Board Splitter box be installed near the pr ocess tool to r educe cable requirements.
WATER
SUPPLY
POWER
NOTE:
RETURN
REMOTE
On-Board Power Cable
On-Board Splitter Box
Figure 3-6: Recommended Multiple On-Board Cryopump or Waterpump Installation
3-12 Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 ©2007 Helix Technology Corporation
Page 43
SUPPLY
RETURN
REMOTE
On-Board Power Cables
On-Board
WATER
POWER
Splitter Box
(Splitter Box located at Process Tool)
Figure 3-7: Alternative Multiple On-Board Cryopump or W aterpump Installation
(Splitter Box located at Compressor)
©2007 Helix Technology Corporation Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 3-13
Page 44

Multiple Cryo-Torr Cryopump Connections

Multiple Cryo-Torr Cryopump Connections
CAUTION
Make sure the Compressor power is OFF before making any connections to the rear panel.
Helium Line Connections
CAUTION
The use of several compressors on a single manifold feeding a common supply header and a common return header requires special precautions. Contact CTI-CRYOGENICS for a review of the intended installation and for specific technical instructions.
The use of a 9600 compressor on a manifold with other CTI-CRYOGENICS compressor models requires a reduction of the helium charge pressure to 200 - 210 psig charge pressure to avoid helium safety valves from inadvertently venting. Refer to “Section 4 -
Operation” and “Section 5 - Maintenance” for more information.
1. Connect the Supply and Return lines to the 9600 Compressor as described in “Connecting/Disconnecting Helium Flex Lines” in this section.
2. Connect the Gas Return Line to the customer supplied helium manifold and then to the Gas Return connector on the Cryo-Torr Cryopump.
3. Connect the Gas Supply Line to the customer supplied helium manifold and then to the Gas Supply connector on the Cryo-Torr Cryopump.
Power Cable Connections
1. Connect the Cryo-Torr power cable between the CRYOPUMP ELECTRICAL OUTLET on the rear panel of the Compressor and the CRYOPUMP ELECTRICAL INPUT on the Cryo-Torr Interface as shown in Figure 3-8 or Figure 3-9.
2. Connect the Cryo-Torr Power Cables between the CRYOPUMP 1,
3-14 Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 ©2007 Helix Technology Corporation
Page 45
SUPPLY
2, or 3 connectors on the Cryo-Torr Interface and the respective Cryo-Torr Cryopumps as shown in Figure 3-8 or Figure 3-9.
3. Connect the User Remote cable to the Cryo-T orr Interface as shown in Figure 3-8 or Figure 3-9.
4. Connect the Remote cable between the Cryo-Torr Interface and the Compressor as shown in Figure 3-8 or Figure 3-9.
NOTE: Your installation may vary based upon the Cryo-Torr Cryopump models used. Refer to “Appendix A - Customer Support Information” to consult your local CTI-CRYOGENICS Customer Support Center for information on specific compressor/cryopump applications.
NOTE: The Cryo-Torr Interface can be installed at the process tool containing the cryopumps as shown in Figure 3-8 or near the Compressor as shown in Figure 3-9. CTI-CRYOGENICS recommends that the Cryo­Torr Interface be installed at the process tool to reduce cable requirements.
RETURN
Cryo-Torr Interface
User Remote (from vacuum system)
WATER
POWER
Remote
Cryopump Electrical Input
Figure 3-8: Multiple Cryo-Torr Cryopump Installation
©2007 Helix Technology Corporation Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 3-15
Page 46
Multiple Cryo-Torr Cryopump Connections
SUPPLY
RETURN
Cryopump Power
Remote
Cryo-Torr Interface
WATER
POWER
User Remote (from vacuum system)
Figure 3-9: Multiple Cryo-Torr Cryopump Installation
3-16 Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 ©2007 Helix Technology Corporation
Page 47

Section 4 - Operation

Adjusting System Helium Pressure

Your CTI-CRYOGENICS high vacuum pump system is comprised of several pressurized components i.e. compressor, flex lines, and cryopumps. Each component is charged with helium before shipment. After all cryopumps, helium lines, and manifolds are attached to the compressor, the system ("OFF" Condition) helium charge pressure must be verified before system operation. Once the ("OFF" Condition) helium system pressure has been verified, the system is ready for operation. After cooldown, the normal system operating pressure is recorded.
NOTE: The 9600 Compressor is designed for continuous operation and should remain ON even when the cryopumps are in a regeneration cycle.

"OFF" Condition Helium System Pressure Verification

The proper system ("OFF" Condition) helium charge pressure is necessary so that the cryopumps operate at maximum performance as well as to assure that the compressor will operate below the maximum design motor winding temperature which will maximize the life of the compressor motor.
1. Make sure the Compressor and Cryopump(s) are OFF.
2. Make sure all system components are connected together as described in “Section 3 - Installation”.
3. Allow all system components to acclimate to a temperature between 60º F and 80º F (15.5º C - 26.6º C).
4. Read the compressor helium pressure gauge located on the compressor rear panel as shown in Figure 1-4. Compare the gauge reading to the appropriate 50/60 Hz line frequency value (depending upon your system installation) indicated in Table 4-1.
Table 4-1: 9600 Compressor Helium ("OFF" Condition) Charge
Line Frequency Helium ("OFF" Condition) Charge Pressure
60 Hz 240 - 250 psig (16.5 - 17.2 bars ) 50 Hz 255 - 265 psig ( 17.6 -18.3 bars )
©2007 Helix Technology Corporation Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 4-1
Page 48

Compressor Operation

Exceeding the recommended system ("OFF" Condition) helium charge pressure will result in the compressor and cryopump safety relief valves opening and releasing excess helium gas.
Compressor Operation
NOTE: The use of a higher helium charge pressure for 50 HZ operation is necessary in order to compensate for the slower speed at which the compressor operates at 50 HZ. The ("OFF" Condition) charge level for 60 Hz remains at 240-250 PSIG.
CAUTION
5. If the ("OFF" Condition) helium charge pressure is not in the ranges as indicated in Table 4-1, then adjust the charge pressure as described in “Section 5 - Maintenance”.
CAUTION
While the 9600 Compressor is operating, the helium pressure gauge indicator should never be in the red zone. If the gauge indicator is below 65 psig, then the system must be checked for insufficient helium or helium leaks. If the gauge indicator is above 160 psig, then the system has been over pressurized. Refer to “Section 5 -
Maintenance” and either add or remove helium before operating the
9600 Compressor.
The system may be operated once the helium charge pressure is correct. Perform the following steps to start the compressor:
1. Set the System Circuit Breaker to the ON (UP) position.
2. Set the Control Circuit Breaker to the ON (UP) position.
3. Close all Cryopump gate valves.
4. Refer to the On-Board Module Programming and Operation manual or Cryo-Torr Cryopump Installation and Service manual (that came with your cryopump) and perform the cryopump start-up procedure.
5. Once the second stage temperature for all cryopumps is below 17K, record the compressor pressure gauge reading as the normal system operating pressure.
NOTE: During compressor operation, the compressor gauge reads the pressure of the gas entering the compressor prior to it being compressed.
4-2 Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 ©2007 Helix Technology Corporation
Page 49
6. Affix a copy of the data next to the compressor gauge on each compressor . This data is to be verified for each tool installa tion and whenever a configuration change is made affecting the amount of system helium gas and line volume.
The compressor pressure reading will decrease from the normal system operating pressure during cryopump regeneration or if fewer cryopumps are being operated. These are normal variations in the compressor pressure reading and should not be cause for concern.
If you have concerns about system performance changing, then check the normal system operating pressure which was determined in “Compressor
Operation” in this section. If the normal system operating pressure is not
correct, check the system for leaks.
Once the leaks have been repaired, helium must be added to return the system to normal operating system pressure as described in “Section 5 -
Maintenance”.

Replacement of Helium Circuit Components

On occasion, it may be necessary to replace components such as cryopumps, helium gas lines or compressors, or change the configuration of the system. Whenever any of these conditions occur, "OFF" Condition Helium System Pressure Verification should be performed to ensure that ("OFF" Condition) helium pressure has not changed.
CAUTION
The use of several compressors on a single manifold feeding a common supply header and a common return header requires special precautions. Contact CTI-CR YOGENICS for a review of the intended installation and for specific technical instructions. The use of a 9600 compressor on a manifold with other CTI-CRYOGENICS compressor models requires a reduction of the helium charge pressure to 200--210 psig charge pressure to avoid helium safety valves from inadvertently venting.
©2007 Helix Technology Corporation Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 4-3
Page 50
Replacement of Helium Circuit Components
4-4 Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 ©2007 Helix Technology Corporation
Page 51

Section 5 - Maintenance

Scheduled Maintenance

Suggested Maintenance Equipment
It is recommended to have the following equipment and disposable supplies available as listed in Table 5-1.
Table 5-1: Suggested Maintenance Equipment
Supply CTI-CRYOGENICS P/N
Helium, 99.999% pure -
Pressure regulator (0-3000/0-400 psi) Assy. 8031403
Helium charging line terminating in a 1/4-inch
female flare fitting
Lint-free gloves and cloth -
Oakite or equivalent detergent soap -
Denatured alcohol -
Refer to
“Appendix A - Customer Support Information” and contact the local Customer
Support Center to obtain the
CTI-CRYOGENICS parts listed in this table.
7021002P001
©2007 Helix Technology Corporation Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 5-1
Page 52
Scheduled Maintenance
Adsorber Replacement
Use the following procedure to change the adsorber every three years.
1. Set the System Circuit Breaker, on the rear of the 9600 Compressor, to the OFF position.
2. Remove the 4 screws which secure the rear panel to the Compressor and remove the rear panel.
NOTE: Use two wrenches in Step 3 to prevent loosening the body of the coupling.
3. Using a 1-3/16 in. wrench, and a 1-1/8 in. wrench, as shown in
Figure 5-1, disconnect the two self sealing coupling connectors
quickly to minimize helium leakage.
This 1 1/8 (or 1) in. wrench holds the coupling in a sta­tionary position.
HOLD WITH 1 1/8 INCH WRENCH
TURN WITH 1 3/16 INCH WRENCH
TO LOOSEN
Figure 5-1: Disconnecting Self Sealing Couplings
This 1 3/16 in. wrench is used to turn the self sealing coupling con­nector.
TO TIGHTEN
TURN WITH 1 3/16 INCH WRENCH
HOLD WITH 1 INCH WRENCH
4. Using a 7/16 in. (11mm) wrench, remove the adsorber mounting bolt as shown in Figure 5-2.
5-2 Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 ©2007 Helix Technology Corporation
Page 53
5. Move the adsorber from under the mounting tabs in the base as shown in Figure 5-2 and remove the adsorber from the Compressor .
6. Install the replacement adsorber under the mounting tabs and secure it into place with the bolt removed during Step 4.
7. Using two wrenches as shown in Figure 5-1, connect the two self sealing couplings quickly to minimize helium leakage.
8. Install the Compressor rear panel.
9. Ensure that the pressure gauge reads the proper value as shown in
Table 4-1. If additional gas pressure is required, refer to “Adding Helium” in this section. If gas pressure needs to be reduced, refer to “Reducing Helium Pressure” in this section.
10. Record the adsorber replacement date on the label as shown in
Figure 5-2, and also note that the next adsorber replacement should
be performed every three years.
©2007 Helix Technology Corporation Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 5-3
Page 54
Scheduled Maintenance
DATE
Adsorber
Adsorber Mounting Bolt
Figure 5-2: Adsorber Location in 9600 Compressor (Rear Panel Removed)
5-4 Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 ©2007 Helix Technology Corporation
Page 55

Adjusting System Helium Pressure

NOTE: These pr ocedur es can be performed on a compressor that is turned ON or OFF. However, the helium pressure gauge should be set to the ("OFF" Condition) helium charge pressure value if the compressor is turned OFF or set to the normal system operating pressure if the compressor is turned ON. Refer to “Section 4 - Operation” for more information.
Reducing Helium Pressure
NOTE: You must obtain the normal system operating pressure from the
“Compressor Operation” procedure in “Section 4 - Operation” in order to
perform this procedure. If the normal system operating pressure is unknown, then shut the compressor OFF and perform the “"OFF"
Condition Helium System Pressure Verification” procedure in “Section 4 ­Operation” instead.
1. Remove the flare cap from the gas charge fitting as shown in
Figure 5-3.
Gas Charge Flared Fitting
Helium Pressure Gauge
Gas Charge Control Valve
Figure 5-3: Helium Pressure Control Components
2. Open the gas charge control valve very slowly to allow a slight amount of helium to escape. Leave the valve open until the helium pressure gauge indicates one of the following:
To the appropriate value in Table 4-1 if the compressor is OFF and
acclimated to a temperature between 60º F and 80º F
Rear Panel
©2007 Helix Technology Corporation Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 5-5
Page 56
Adjusting System Helium Pressure
Increasing Helium Pressure
If helium is being added more than once every several months, check for leaks caused by improperly connected self-sealing connections or any mechanical joint in the Compressor.
(15.5º C - 26.6º C).
To the value previously recorded in the “Compressor Operation”
procedure in “Section 4 - Operation” if the compressor is ON.
3. Close the gas charge control valve and install the flare cap.
Use the following procedure to increase the helium pressure if the indicated pressure is below the appropriate value as shown in Table 4-1.
CAUTION
Adding Helium
NOTE: You must obtain the normal system operating pressure from the
“Compressor Operation” procedure in “Section 4 - Operation” in order to
perform this procedure. If the normal system operating pressure is unknown, then shut the compressor OFF and perform the “"OFF"
Condition Helium System Pressure Verification” procedure in “Section 4 ­Operation” instead.
This procedure ensures that both the regulator and the charging line will be purged of air and that the air trapped in the regulator will not diffuse back into the helium bottle. For best results, CTI-CRYOGENICS suggests a dedicated helium bottle, regulator, and line, which are never separated, for adding helium.
NOTE: You are required to supply the helium charging line terminating in a 1/4-inch female flare fitting, and a two-stage pressure regulator rated at 0-3000/0-400 psig for this operation.
CAUTION
Use only 99.999% pure helium gas. Helium circuit contamination may result if a lower quality of helium is used.
5-6 Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 ©2007 Helix Technology Corporation
Page 57
1. Attach a regulator (0-3000/0-400 psig) and charging line to a helium bottle (99.999% pure).
NOTE: Do not open the bottle at this time.
2. Purge the regulator and charging lines as follows: a. Open the regulator a small amount by turning the adjusting
knob clockwise until it contacts the diaphragm, then turn approximately 1/8 to 1/4 turn more, so that the regulator is barely open.
b. Loosely connect the charge line to the helium pressure
regulator.
c. Slowly open the bottle valve, and purge the regulator and
line for 10 to 15 seconds. Turn the regulator knob counterclockwise until the helium stops flowing.
3. Remove the flare cap of the gas charge flared fitting on the rear of the Compressor.
4. Loosely connect the charging line from the helium pressure regulator to the 1/4-inch male flare fitting installed on the helium charge valve. Purge the charge line again, as in step a, for 30 seconds, and tighten the charge line flare fitting onto the gas charge fitting while the helium is flowing.
5. Set the helium pressure regulator to 300 psig (20.7 bars). If the compressor is ON, proceed with step a. If the compressor is OFF, proceed with step b.
a. Obtain the previously recorded normal system operating
pressure from the “Compressor Operation” procedure in
“Section 4 - Operation”. Open the gas charge control valve
very slowly and allow helium to flow until the compressor gauge reading is the same as the value obtained from
“Section 4 - Operation”. Quickly close the gas charge
control valve.
b. Obtain the appropriate (50 or 60 Hz) ("OFF" Condition)
system operating pressure from Table 4-1. Open the gas charge control valve very slowly and allow helium to flow until the compressor gauge reading is the same as the appropriate value in T able 4-1. Quickly close the gas char ge control valve.
6. Ensure that the helium charge valve on the Compressor is tightly closed. Shut off the helium pressure regulator on the helium bottle and remove the charging line from the male flare fitting. Reinstall the flare cap.
©2007 Helix Technology Corporation Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 5-7
Page 58
Adjusting System Helium Pressure
5-8 Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 ©2007 Helix Technology Corporation
Page 59

Appendix A - Customer Support Information

Customer Support Center Locations

To locate a Customer Support Center near you, please visit our website www.helixtechnology.com on the world wide web and select CONTACT on the home page.

Guaranteed Up-Time Support (GUTS®)

For 24-hour, 7-day per week Guaranteed Up-Time Support (GUTS) dial:
1 800-367-4887 - Inside the United States of America
+1 508-337-5599 - Outside the United States of America
Product Information
Please have the following information available when calling so that we may assist you:

E-mail

Product Part Number
Product Serial Number
Product Application
Specific Problem Area
Hours of Operation
Equipment Type
Vacuum System Brand/Model/Date of Manufacture
For your convenience, you may also e-mail us at:
techsupport@helixtechnology.com
©2007 Helix Technology Corporation Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 A-1
Page 60
E-mail
A-2 Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 ©2007 Helix Technology Corporation
Page 61

Appendix B - Flow Diagram

©2007 Helix Technology Corporation Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 B-1
Page 62
A
.
v
e R 1
0
0
5
3
1
8 N
/ P
m
a
r
g
a
i D
w
o
l F
r
o
s
s
e
r p
m
o C 0
0
6
9 :
1
­B e
r u
g
i F
B-2 Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 ©2007 Helix Technology Corporation
Page 63

Appendix C - Troubleshooting Procedures

Troubleshooting the Compressor

The compressor troubleshooting procedures are summarized in Table C-1.

Technical Inquiries

Please refer to “Appendix A - Customer Support Information” of this manual for a complete list of the
CTI-CRYOGENICS’ world wide customer support centers.
CAUTION
Disconnect the compressor before performing any troubleshooting procedures. The compressor pump is hot after operating. Wait for the pump to cool down before working on the inside of the compressor
Do not change or modify any compressor internal wiring circuits, this may cause failure of the compressor and cold head due to improper phasing.
©2007 Helix Technology Corporation Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 C-1
Page 64
Technical Inquiries
Table C-1: Compressor Troubleshooting Procedures
Problem Possible Cause Corrective Action
1) System circuit breaker (CB1) trips immediately to the OFF (0) position when switched to the ON (1) position.
2) System (CB1) and Control Circuit (CB2) circuit break­ers remain in the ON (1) position when switched ON but the compressor will not run.
1) Incorrect phasing of input power.
1) No power coming from source.
2) Insufficient power
3) Remote control jumper plug not in place. This will apply only if remote circuit is not being used.
1) Check phasing of input power. Refer to
“Phase Check” in “Section 3 ­Installation”.
1) Check source fuses, circuit breakers, and wiring asso­ciated with the power source. Repair as needed.
2) Verify adequate phase-to­phase input voltage. Refer to
Table 1-2.
3) Check to insure that remote
jumper plug is fully seated. See
Figure 1-4 for loca-
tion. Refer to
“Compres-
sor Remote Connector”
in
“Section 1 - 9600 Compressor Descrip­tion” for more information.
4) Improperly wired external remote control circuit. NOTE: Only applies if remote control feature is being used.
C-2 Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 ©2007 Helix Technology Corporation
4) Verify correct installation of remote control feature. Refer to
Table 1-6.
Page 65
Table C-1: Compressor Troubleshooting Procedures (Continued)
Problem Possible Cause Corrective Action
3) System circuit breaker (CB1) will not remain in the ON (1) position when switched ON. The Control Circuit circuit breaker (CB2) trips when exces­sive current is being drawn by the cold head or 24 volt compressor control cir­cuits.
4) System circuit breaker (CB1) remains in the ON (1) position and the com­pressor stops after several minutes of operation and remains OFF (0).
1) Damaged On-Board power cable, connectors, or drive motor.
2) Damaged component in the
compressor power or control circuit.
1) Thermal protective switches
are open.
2) Very cold water has caused a
restriction of oil flow through the oil injection orifice dur­ing start-up.
1) Check for compressor oper­ation with cryopump cable disconnected from compres­sor. Refer to
“Appendix A
- Customer Support Infor­mation” to contact the Cus-
tomer Support Center if the compressor operates improperly.
2) Refer to
“Appendix A ­Customer Support Infor­mation” to contact the Cus-
tomer Support Center.
1) Check for inadequate water cooling. Refer to
2) Recheck for proper cooling water temperature. Refer to
Table 1-3.
Table 1-3. Restart compres-
sor repeatedly until continu­ous operation is achieved.
5) System circuit breaker (CB1) trips after a period of running.
1) Loss or degradation of power from the source.
2) Defective motor windings.
1) Check that line voltage is correct on all phases.
2) Check running current on all phases.
3) Refer to
“Appendix A ­Customer Support Infor­mation”
Customer Support Center.
to contact the local
©2007 Helix Technology Corporation Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 C-3
Page 66
Technical Inquiries
C-4 Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 ©2007 Helix Technology Corporation
Page 67

Introduction

Appendix D - Schematic

The schematic in Appendix D supports the 9600 (High Voltage) Compressor CTI-CRYOGENICS P/N 8135164P002.
Table D-1: Basic Control Assembly Legend
Identifier Description
1M Compressor Motor CB1 Main Circuit Breaker (15A) CB2 Control Circuits Circuit Breaker (4A)
ETM Elapsed Time Meter
ETM1 Elapsed Time Meter
J15 Module Power Receptacle
J1/P1 Autoset Power Connector J2/P2 Unload Solenoid Connector J3/P3 Oil Solenoid Connector J4/P4 Helium Thermal Switch J5/P5 Water Thermal Switch J6/P6 Compressor Contactor Coil J7/P7 Autoset Input Connector
J8 ETM Connector
M1 Contactor 7.5 HP IEC
J9/P9 LT1 Connector
J10/P10 Control Transformer
J11 Open
J12 Module Signal Connector
J13/P13 Phase Monitor
J14 Open
©2007 Helix Technology Corporation Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 D-1
Page 68
Introduction
Table D-1: Basic Control Assembly Legend (Continued)
Identifier Description
K2 Over Temperature Lockout Relay
LT1 Lamp, 24-28V LED Green
M1OL Relay, Overload (10-16A - set at 10A)
PM1 Phase Monitor OMRON RDR-TFY-M
PWB1 PWB Autoset
T3 Transformer Assembly Control
Table D-2: On-Board/Cryo Electrical Module P/N 8135127G002 Legend
Identifier Description
JT1 Open JT2 Open
JT3/PT3 T2 Cold Head Supply, 2 phase JT4/PT4 T1 Cold Head Supply, 2 phase JT5/PT5 T2 Cold Head Transformer Output JT6/PT6 T1 Cold Head Transformer Output JT7/PT7 T4 Supply JT8/PT8 T4 Low Voltage Output (23/26 VCT)
J18 Power Output
J19 Signal Output P12 Signal Input Connector P15 Power Input Connector
K1 Cold Head Voltage Relay K2 Signal Voltage Relay K3 Cryo Power Relay
T1-T2 Cold Head Motor Transformers
T4 On-Board Logic Power Transformer T5 On-Board Heater Power Transformer
D-2 Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 ©2007 Helix Technology Corporation
Page 69
Table D-3: On-Board Output Module P/N 8135148G001 Legend
Identifier Description
J30 On-Board Output Receptacle
J31 Remote Control Receptacle P18 Power Connector P19 Signal Connector P31 Remote Jumper
©2007 Helix Technology Corporation Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 D-3
Page 70
Introduction
D
8
1
7
REVISIONS
80828486889094969892100
6
E REVISED PER DEO 11883 UM 4/18/97 VG
REV DATE APPROVEDDESCRIPTION
2
5
RED VIO
GRA
3ph
BLK
JT2
3ph
JT1
14 1234
JT7
T5 RED ORN YEL
VIO
1234
1
JT4
T1
RED ORN YEL
234 234 1234
1234
VIO
1
JT3
T2
YEL
RED
VIO
W/YEL WHT W/ORN
GRA
BLK
ORN
T4
BLK
P15
C
J30
A
C
DDEEF
RED
WHT
WHT/ORN
8135148G001
ON-BOARD
OUTPUT MODULE
P18
1234567812345
102
104
106
2
JT9
1
JT6
1
JT5
12345 234 234 1
JT8
124
1234567812345
JP3
JI8 BLK
JP2
K2 K1
BRN
225
BLU
BLK
ABCN
W/BK W/GY
GRA
BLK
ABCN
W/BK W/GY
GRA
YEL
RED
G/YEL
6
A
BBC
ORN
GRA
YEL
W/RED REMOTE
GRN/YEL
P19
110
112
114
118
VIO VIO
GRA
YEL
K3
JP1
GRN/Y J19
W/RED
W/ORN
W/BLU
W/BRN
WHT
135246P12
P31
FGGHHJJKKILMN
BLU
WHT/BLU
6
120
122
6
B
J31
124
A
8
11
SHEET OF
1
8135164P002 E
W/ON-BOARD
DWG NO
SCHEMATIC, 9600
CTI-CRYOGENICS
HELIX TECHNOLOGY CORPORATION
COMPRESSOR, HIGH VOLTAGE
CAGE CODE
SIZE
2/96
2/96
DATEBY
VG
GF
ENG APPVL
DRAWN
CHECKED
ENG APPVL
PRE-PROD REV
ENG APPVP
7
31949
B
6
N/A
QA APPVL
PROD
23
5
4
X Y Z GND
D
OPENS CONTACTOR
OPENS CONTACTOR
POWER DOWN TO RESET
TS;
REMOTE OR SMOD;
He
Comp
TS2
Water
TS1
TS3
OIL
SOL1
SOL2
J12
40
13524
123 4
P10
UNLOAD
VIO
WHT
VIO
CRYO
MODULE
SIGNAL
WHT
P3
1
2P21
2
1
2J31
2
6
J2
K2B
24AC
line HI
23 W/YEL
23 1
1 J7
P7
AUTOSET
464842
54
52
50
B
C C
P14
P5
P4
J4
1
2
1
2
1
2
1
2
1
2
1
2
J5
J14
K2A
K2
COM
J9
1
RED
LT1
58
62
56
65
60
M1
V
COMPRESSOR
GRN/YEL
RED #12
M1OL
M1
5
RED #12
.02A
P4
BLK
97 98
M1OL
W/RED
P2 P1
15A
CB1
P3
9
5
CRYO
POWER
MODULE
U
W
124
J15
YEL #12
BLK #12
T3
T2
T1
6
4
123
BLK
YEL
RED
PM1
RST
YEL #12
BLK #12
BLU W/ORN W/RED
T
W
S
V
4A
CB2
R
U
11
16
20
13
6
YEL
450
T3
BLK
22
26
SMOD
123
4
J11
JP1
POWER AUTOSET BD
P13
1234
0
WHT
23
ORN
26
P1
225
BRN
1234
1234
J1
BLU
GRN/Y
28
32
36
C
B B
34
3
2
1
TRIPS BREAKER
PHASE OR O/L;
COMPRESSOR BASIC HI VOLTAGE CONTROL ASSY ON-BOARD/CRYO-TORR ELECTRICAL MODULE
1
A A
P16
J16
2J81 2
WHT
STATEMENT
ENVIRONMENTAL
SELLER'S GOODS SHALL
NOT BE MANUFACTURED
WITH OR CONTAIN CLASS
I ODS'S AS DEFINED
UNDER SECTION 602 OF
THE FEDERAL CLEAN AIR
ACT (42 USC SECTION
7671A) OR ANY NEW
APPLICABLE
REGULATIONS.
BLU
1
2
1
2
M1
1314
BRN
M1
WHT
P6
1
2
1
2
J6
K2A
8135132G002
2
1
2P91
P8
VIO W/YEL
WHT
ETM
69
73
CHANGE TO SCHEMATIC 8135230P005.
1. ANY CHANGE TO THIS SCHEMATIC MAY RES ULT IN A
NOTES:
A
D D
4
3
ORCAD V7.11
CAD FILENAME: 8135164P002.DSN
2
4
1
Figure D-1: 9600 High Voltage Compressor
D-4 Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068 ©2007 Helix Technology Corporation
Page 71

Index

ELIXH
A
Adsorber
replacing, 5-2
C
Components
compressor remote connector, 1-11 control circuit breaker, 1-9 cooling water IN, 1-10 cooling water OUT, 1-10 Cryopump electrical outlet, 1-10 description, 1-8 elapsed time meter, 1-9 gas charge flared fitting, 1-9 helium pressure gauge, 1-10 power inlet, 1-10 power ON indicator, 1-9 return gas coupling, 1-10 supply gas coupling, 1-11 system circuit breaker, 1-9
Compressor
description, 1-1 installing, 3-1 operation, 4-2
schematic, D-4 Compressor remote connector, 1-11 Control circuit breaker, 1-9 Cooling water IN, 1-10 Cooling water OUT, 1-10 Cryopump electrical outlet, 1-10
E
Elapsed time meter, 1-9 Electrical
connecting, 3-4
G
Gas charged
flared fitting, 1-9
H
Helium
adding, 5-6 refrigeration system, 1-1
Helium circuit components
replacing, 4-3
Helium flex lines
connecting/disconnecting, 3-7
Helium pressure
increasing, 5-6 reducing, 5-5
system, adjusting, 4-1 Helium pressure gauge, 1-10 Helium pressure, system
adjusting, 5-5
M
Maintenance, 5-1–5-7 Multiple Cryo-Torr connections, 1-13 Multiple Cryo-Torr pumps
connecting, 3-14 Multiple On-Board Cryopump connections, 1-12 Multiple On-Board Cryopumps
connecting, 3-11
O
OFF condition helium system pressure
verifying, 4-1
P
F
Flow diagram, B-1
©2007 Helix Technology Corporation Pub. No. 8040437, Rev. 104, 06/12/2007 E CO No. 18068 Index-1
Phase checks
performing, 3-6 Power inlet, 1-10 Power ON indicator, 1-9
Page 72
R
Return gas coupling, 1-10
S
Safety precautions, S-1 Schematic, D-1, D-4
basic control assembly legend, D-1 On-Board output module legend, D-3 On-Board/Cryo electrical module
legend, D-2
Single On-Board Cryopumps
connecting, 3-8 Specifications, 1-4 Supply and return water lines
connecting, 3-3 Supply gas coupling, 1-11 System circuit breaker, 1-9
ELIXH
T
Troubleshooting, C-1–C-3
U
Unpacking, 2-1–2-2
©2007 Helix Technology Corporation Pub. No. 8040437, Rev. 104, 06/12/2007 E CO No. 18068 Index-2