Heinen & Löwenstein Leoni Service Manual

Heinen & Löwenstein GmbH
Servicecenter
LEONI baby ventilator service manual
Changing the sintered filters Page 3
Sensor adjustment Page 5 Flow sensor
O2 sensor
Flow adjustment AIR 10 Lpm. Page 6
O
10 Lpm. Page 7
2
Pressure sensor adjustment 0 cm H2O Page 8
60 cm H2O
Expiration valve adjustment Page 9
Desktop\Eigene Dateien\Eigene Dateien\Dokumente\Service an Leoni vers.1,0 Christoph H. Schmitt 21.05.2003
Test report : 1LED’s, Display, Signal Generator Page 10 2
Patient alarms Page 11 3System alarms Page 12 4Flow sensor alarm Page 12 5Pressure measurement Page 13
6Volume measurement Page 14
7Oxygen measurement Page 15
8Safety check Page 15 Intervals
Service and Safety check Page 16
Spare parts Page 17
Desktop\Eigene Dateien\Eigene Dateien\Dokumente\Service an Leoni vers.1,0 Christoph H. Schmitt 21.05.2003
2
Changing the sintered filters
. Hex key : 4mm Sintered filter : 0207157 .
The NIST connectors are at the rear of the device. Using the hex key, the fixing screws can be removed in order to take out the filters (or, in newer versions, the filters can be removed directly with the hex key).
Software version check:
In the onscreen menu „settings“, press the keys “menu” and “alarm limits” in parallel and keep pressed for 5 seconds. The onscreen menu “system diagnosis” is opened and here you take the readings of the following parameters in the left column:
CHK PO DAT PO CHK P1 Ver P1 CHK P2 Ver P2
Check whether the latest version is installed.
Desktop\Eigene Dateien\Eigene Dateien\Dokumente\Service an Leoni vers.1,0 Christoph H. Schmitt 21.05.2003
3
Sensor adjustment:
Flow sensor:
Before each adjustment, the device must warm up for 15 min.
Assemble LEONI with its sensors, tubing system and test lung. Use the calibration key to choose the onscreen menu “sensor check”. With the tubing system connected, press soft key “flow” and keep pressed until “sensor check” and then “ADJUSTMENT” shows up. With the message „sensor adjusted“, adjustment of the flow sensor has been effected successfully.
O2 Sensor:
Oxygen calibration must always be done almost pressureless, so
first disconnect the expiration tube in order to reach a minimum
constant positive pressure.
In the same onscreen menu now press the „21% O2“ key and keep pressed until the hourglass starts on the monitor. 21% adjustment has been successful when the hourglass disappears without error message and you see the message „sensor adjusted“.
Finally, press the “100%“ key and keep pressed until the hourglass starts. 100% adjustment has been successful when the hourglass disappears without error message and you see the message „sensor adjusted“.
Desktop\Eigene Dateien\Eigene Dateien\Dokumente\Service an Leoni vers.1,0 Christoph H. Schmitt 21.05.2003
4
Flow adjustment : AIR
Screwdriver : slot-head 3,5 x 75
For flow adjustment, disconnect the expiration tube and kink it. As soon as the system alarm starts, acknowledge and mute the system alarms by pressing the keyswitch and the alarm suppression key in parallel. In the lower left corner of the monitor a message shows up. Enter the setup screen via keyswitch and adjust the following parameters:
Insp.-Flow
20
Basic flow
10
PEEP
10
P.-Insp.
50
Frequency
30
Re-connect the expiration tube but remove the test lung. Now switch the screen to the graphical mode by means of the menu key and then set to 20 litres compressed air. Check by displaying the flow graph in the 20 l setting. (Switch from pressure to flow by repeatedly pressing the „graphs“ key).
Insp.-Time
1
Oxygen
21%
Do adjust the flow sensor first !
Pressure reducing valves for adjustment are at the rear of the device. Pressure adjustment is achieved by turning the pressure reducer set screw. Compressed air is controlled by the pressure­reducer next to the NIST-AIR screwing. Turn until the pressure shown in the screen display corresponds to 20 or 10 litres respectively.
Desktop\Eigene Dateien\Eigene Dateien\Dokumente\Service an Leoni vers.1,0 Christoph H. Schmitt 21.05.2003
5