Heiland electronic TRD 2 User Manual

Page 1
Heiland electronic GmbH
B&W-Densitometer
User manual
Version 6.2
Page 2
Table of Contents
1. GENERAL INFORMATION.....................................................................3
2. SAFETY REGULATIONS.......................................................................3
3. AREA OF APPLICATIONS.....................................................................3
4. INSTRUMENT DESCRIPTION
4.1. STANDARDVERSION (FIGURE 1)......................................................4
4.2. OPTIONS (FIGURE 2,3)......................................................................6
5. INSTALLATION
5.1. LOCATION CONDITIONS...................................................................8
5.2. SWITCHING ON..................................................................................9
6. READING
6.1. INSTRUCTIONS TO AVOID INCORRECT READING..........................9
6.3. INSTRUCTIONS FOR RESET TO ZERO...........................................10
6.4. DENSITY MEASUREMENT...............................................................11
7. CARE OF INSTRUMENT......................................................................12
8. CALIBRATION CHECK........................................................................12
9. GUARANTEE.......................................................................................13
10. TECHNICAL DATA.............................................................................14
2
Page 3
1. General information
Dear Customer, thank you for purchasing a Heiland densitometer. In order to become familiar with the operation of your instrument, please read these instructions carefully and completely before your first use.
The packaging of the instrument is shock-absorbent and should be kept for the case of shipping and repairs. Use only the original packaging for transportation and shipment. First lay the cardboard on the surface of the base unit, then place the roll of bubble film between measuring head and base unit. Finally put the instrument inside the plastic bag and then into the shipment-box. Don't forget the calibration strips!
CAUTION: Never use the removable measuring head (Opt. 01) for carrying the instrument. The base unit may separate and drop off! Damages caused by this are not included by guarantee.
2. Safety regulations
§ Assure that the main voltage and the frequency on the power supply unit is the same as the one indicated on the identification label of the power supply unit
§ It is not permitted to exchange the power supply against another one without the permission of the manufacturer.
§ Always keep the instrument and the power supply in a dry place and never immerse it in any kind of liquid.
§ To prevent electric shock do not open, neither the densitometer nor the power supply. There are no user serviceable parts inside.
3. Area of applications
The densitometer is designed for the measurement of density values of B&W and other nearly grey coloured subjects. The reading is possible in units of logarithmic density. Please contact the manufacturer, if you like to use the instrument for other applications. The modular construction of the instrument enables the user to optimise it according to his requirements.
TRD 2 USER MANUAL Version 6.2 3
Page 4
4. Instrument description.
Figure 1, TRD 2 standard version
Measuring
head Mode-LED‘s Display
Probe
Aperture
4
Base unit
Mode and reference buttons
Page 5
4.1 Standard version
§ The id-label is located on the bottom of the base unit.
§ The base unit contains the transmission lamp. The light is emitted
through an opal aperture.
§ The top of the base unit serves as a platform, on which the film or photo rests during measurement.
§ The measuring arm is mounted to the base unit by a precise bearing. The integrated spring pushes the measuring head in the upper position. The head contains the sensing electronic, below of it the probe and the reflective lamp are mounted.
§ One of the lamps and the displays are automatically activated
when the measuring head is lowered.
§ The 3-digit display allows a precise reading down to 0.01 log.D. If the minimum sensitivity of the system is reached or if the measuring head is depressed too fast, the display will flash "-888". If reading a lower density than at reset, a negative sign will appear. If the display flashes E XX (where XX is the number of the error) please consult the manufacturer.
§ The LED's beside the 3-digit display show the selected measurement mode while the measuring arm is lowered. T: Transmission measurement R: Remission measurement D: Density values are displayed
§ The switches on the top of the measuring arm are only active, when the display is on. The Min switch sets the display to zero, with T/R you switch between remission and transmission mode.
TRD 2 USER MANUAL Version 6.2 5
Page 6
Figure 2, Rear side and options
Opt.02
Power Supply
Opt.03
Opt.01
6
Page 7
4.2 Options
Option 01: Removable measuring head
The removed head is used for reflection measurement only! The measuring head, when attached to the base unit or the additional base plate, is kept in correct alignment only, when the screw with the black knob on the rear side is tightened. For measurement of large murals the head is easily removed from the body and then attached to a base plate with an integrated target. The electrical connection is made with a long or a short cable.
Measuring large prints:
§ Turn the instrument so, that the long side is right in front of you.
§ Remove the cable between the base unit and the measuring
head.
§ Loosen the screw on the rear side of the instrument by turning the black knob.
§ Grip the back part of the measuring head in direction "A".
§ Pull the measuring head in direction "B" off the base unit and stick
it onto the additional base plate. Keep attention that the vertical part of the base plate slips into the slot of the measuring head (between case and bear).
§ Tighten the screw with the black knob. Use no tool.
If you attach the head to the base unit later on please connect the cable again, otherwise the transmission lamp will not work.
Figure 3: Moving the measuring head
B
A
TRD 2 USER MANUAL Version 6.2 7
Page 8
Option 02: Changeable apertures
The apertures have 4 different diameters (0.5/1/2/3mm). Like the standard diffuser they are made out of opal material. Smaller diameters reduce the amount of light, thus the maximum density is lowered. The aperture is changed by pressing slightly on the edge of the diffuser, e.g. with the fingernail. After changing the aperture, new reference points must be set for transmission mode.
Option 03: PC port
To connect the densitometer to a PC, use the optional USB connector and cable. See appendix for details.
Option 05: Extended transmission density range
The maximum density in transmission mode is extended to 5,5 log.D.
Option 06:
The resolution of the instrument is increased to 0.001 log.D in the measurement range between –0.999 and +0.999 log.D.
5. Installation
5.1 Location conditions
§ Store and use this instrument only in rooms with a temperature from 17°C to 27°C (degree Celsius) and relative humidity from 0 to 70 percent. Fast temperature variations may cause condensation inside the densitometer or the power supply. After moving the instrument from a cold to warm environment (more than 10°C temperature difference) please wait at least one hour until you connect the power supply again.
§ Keep the densitometer away from these environment:
- Direct Sunlight
- Instruments that emit heat
- Aggressive chemicals or liquids
- Strong magnetic fields, e.g. from transformers, voltage stabilisers or loudspeakers
- Electronic devices that emit strong interference’s, like dimmers
8
Page 9
5.2 Switching on
§ Assure that the main voltage of the power supply unit is the same as the one indicated on the identification label of this unit.
§ Connect the power supply first to the densitometer and then into the wall plug
§ The instrument switches on automatically in a stand by mode. Wait at least 2 minutes before taking the first reading.
.
6. Reading
Caution: Always depress the measuring head softly until it stops on the reading-area. To prevent damage of the precise bearing, always let the measuring head return slowly to the "up" position! Never try to raise it higher than this stop position! On a digital display, the actual value may intermit between two values. The display changing back and forth between two values should not be a cause for concern.
6.1 Instructions to avoid incorrect reading
§ Especially when measuring high densities the subject must be free from dust, lint, finger prints, and other contaminants.
§ The right reading mode is very important: "Transmission" for film, glass plates or filters. "Reflection" for non-transparent subjects (i.e. pictures) The wrong mode causes considerable errors.
§ The densitometer can also be used for process control with B&W images on colour-based film (i.e. silver-free films), if no colour­extracting filter is required.
§ Thick glass plates or filters should be read in low ambient light because of their light conducting tendency.
§ Films should be placed on the reading area with the emulsion
side down.
TRD 2 USER MANUAL Version 6.2 9
Page 10
6.2 Instructions for reset to zero
Reset the display before starting a series of density measurements. If the part of the subject used for zeroing does not have the lowest density, a negative sign will appear while measuring lower density.
Reset of transmission reading
§ In photographic applications you may decide either to measure the density in relation to the absolute light or to base+fog. That is an unexposed, but a developed area of film.
- If you zero to absolute light (no film in the light pass), the base+fog is displayed with a positive value and this value has to be subtracted from all density values.
-If you zero to base+fog, the density values of any area within the negative could be read direct. The absolute value of base+fog could also be measured, while taking a reading to absolute light and ignoring the negative sign in this case
§ If in thick glass plates or filters zeroing to the light source is required, then the subject to be read must be placed on the reading area so that the light source is not covered up, but the measuring head rests on the plate when the measuring head is depressed. The free light flux is equivalent to absolute-zero.
Reset of reflective reading
§ There is no totally reflective and diffusing material i.e. there is no absolute white. Barium sulphate is used for zeroing, because it reflects up to 99% of diffuse incident light, however, it is also sensitive to soiling and damage. Normally, the back of a fibre based paper is sufficient for zeroing.
§ In photography, zeroing is done on the base fog, that is an unexposed but developed part of the subject.
§ If occasionally material thicker than 2 mm is to be measured, then the material and the paper used for zeroing should be placed under the measuring head together to allow for the offset in the reading geometry. This hint can be ignored, if you're working with the removed measuring head (Opt.01).
10
Page 11
6.3 Density measurement
Transmission reading
§ If the instrument is not in mode 'T' select this mode by pushing the "T/R" button.
§ Reset to zero on base fog. Position film so that the lightest part is centred over the light source. Depress measuring head and press "Min" button until display reads 0.00.
§ Measurement is done by slightly lowering the measuring head until light comes on, then centring the area to be read over the light source and reading the value with the fully depressed measuring head.
Reflective reading
§ If the instrument is not in mode 'R' select the mode by pushing the "T/R" button.
§ Place subject in a way that the area to be read is centred under the tip of the measuring head. Press "Min"-button until display reads 0.00.
§ Take series of readings.
TRD 2 USER MANUAL Version 6.2 11
Page 12
7. Care of instrument
§ The housing and especially the reading area should only be cleaned with a lightly dampened soft cloth or chamois, if necessary with the addition of a mild cleaner
§ Do not use solvents or abrasive materials.
§ The measuring head and probe should only be cleaned from dust
with a soft brush or compressed air.
§ Unplug the instrument when not in use for long time periods.
§ For necessary repairs send the densitometer and the power
supply unit to manufacturer or authorised service centre. Any unauthorised opening, will invalidate your guarantee.
§ CAUTION: To prevent an electric shock do not open the instrument! There are no user serviceable parts inside!!
8. Calibration check
The instrument is calibrated at the factory. Certification is shipped with the instrument in the form of a transmission and a reflective light calibration strip on which the density values of the calibration are recorded.
Note: The densitometer and the calibration strips shipped with it belong together. Since measuring geometry may differ, the calibration strips should not be used to check an other densitometer.
The calibration strips are made of photographic material. They should be kept in a dark, cool, dry, and dust-free place to protect density values. Do not touch reading area with bare hands. Soiled reading areas cause incorrect readings, especially on high densities.
12
Page 13
Calibrating should be done under the following conditions:
§ Ambient temperature 20 to 22 ° Celsius.
§ Relative humidity between 0 and 70 %
§ Voltage and frequency must be in tolerance to the one indicated
on identification label.
§ The instrument should be acclimated to ambient conditions and have been connected to the power source for a minimum of 5 minutes before calibrating.
§ For calibrating procedure use the aperture with 3mm diameter (Opt.02 only).
§ Position the calibration strip so, that the left field (0,00) of it is centred over the light source. Reset the display to zero.
§ Reading of densities is done in the centre of each test area. If a slight drifting of values is encountered, the values should be averaged.
§ The tolerance is ± 0.02 log. D. If greater variations are encountered, the TRD and the calibration strip should be sent to the manufacturer, or authorised service centre, since a new calibration is not possible without the necessary testing devices.
9. Guarantee
This instrument is fully guaranteed for the period of one year from date of purchase. Proof of purchase is the invoice or receipt. Damages due to improper handling or unauthorised access are excluded.
This instrument has been carefully manufactured and tested using defect free materials and state-of-the-art technology.
In case of a failure the instrument should be returned free of charge to the manufacturer or authorised service centre accompanied by proof of purchase.
For the warranted period, the manufacturer will pay cost of replacement parts and repair. However, the manufacturer reserves the right to exchange the instrument. Important: Both the instrument and calibration strips must be send back together!
TRD 2 USER MANUAL Version 6.2 13
Page 14
10. Technical data
§ Dimensions length x width x height : 200 x 100 x 100 mm
§ Weight : 1 kg
§ Voltage requirements : 12 V DC by use of a wall
plug power supply
§ Power consumption : 3 VA
§ Aperture diameter : 0,5/1/2/3 mm
according to the order
§ Reading geometry
Transmission : diffuse/directed Reflection : 45/0 degrees Opt.04 : polarised
§ Maximum subject size
Width x Height : 260 mm x 2 mm Opt. 01 : 2m Radius
§ Maximum measurable density Transmission aperture. ∅ 3,0mm : 4,0 D
∅ 2,0mm : 3,5 D ∅ 1,0mm : 3,0 D ∅ 0,5mm : 2,5 D
Transmission Opt. 05 : 5,5 D Reflection : 2,5 D
§ Reading variation : ±(1% + 0,02 log.D.)
§ Repeat accuracy : ± 0,01 D
§ Temperature range : 17 ... 27 °C
§ Relative humidity : 0 ... 70 %
§ Accessories : Power supply unit
Two calibration strips
14
Page 15
Appendix: Communication parameters
§ Communication port type: USB 2, emulating a serial com port with 9600 Baud, 8 Databit, 2 Stopbit.
§ Character which are send to the densitometer are answered with the character NAK (15h).
§ The measurement value is transmitted automatically by the instrument every time there is a stable value for at least 0,5 seconds. The value is transmitted with the following syntax:
Character-No. possible Char. meaning 1 T/R/0 Trans./Refl./StandBy 2 +/- Sign 3 0...9/. Value MSB 4 0...9/. Value 5 0...9/. Value 6 0...9/. Value LSB 7 D/Z/% Density/Zone/Percentage of dot area 8 CR End of string
Examples:
T+1.53D Transmission
Density: +1.53 log D R-0.05D Reflective Density: -0.05 log D T+05.7Z Transmission Zone: 5.7 R+45.0% Reflective 45 Percent of dot area T+.278D Transm. in case of Opt.06 Density: +0,278 log D
0+....D Stand By mode or
measuring head is up Issue date: January 2007 Technical data may change without notice.
TRD 2 USER MANUAL Version 6.2 15
Page 16
Heiland electronic GmbH Schulstraße 8 D-35579 Wetzlar