Heidelberg SL-OCT User manual

SL-OCT
Slitlamp- OCT Installation Instructions
Version 001, February 2009 © Heidelberg Engineering GmbH 2009 Art. No 18107 QM-Nr. 97 201-001
SL-OCT Installation Manual
Art. No.: 18107 Page 2 of 42
Corporate Headquarters
Heidelberg Engineering GmbH • Tiergartenstr. 15 • 69121 Heidelberg • Germany
Phone +49 6221 6463-0 • Fax +49 6221 646362 • www.HeidelbergEngineering.de
US Main Office
Heidelberg Engineering, Inc. • 1499 Poinsettia Avenue, Suite 160 • Vista, CA 92081
Phone 760 598-3770 • Fax 760 598-3060 • www.HeidelbergEngineering.com
US Service Center
Heidelberg Engineering, Inc. • 410 Harris Road • Smithfield, RI 02917
Phone 401 349-0500 • Fax 401 349-0504 • www.HeidelbergEngineering.com
SL-OCT Installation Manual
Art. No.: 18107 Page 3 of 42
Contents
1 INTRODUCTION 4 2 SCOPE 6 3 TOOLS 7 4 UNPACKING THE SYSTEM 8
4.1 Lift table 8
4.2 OCT-box 8
4.3 Slit lamp 8
5 INSTALLATION 9
5.1 Lift table 9
5.2 Optional printer table 16
5.3 OCT-box 19
5.4 Slit lamp 25
5.5 Electrical and fiber connections to the slit lamp 26
5.6 Electrical connections to the OCT-box 28
5.7 AC connections 29
5.8 Cleaning 30
6 TEST MEASUREMENT 31 7 DISASSEMBLY OF A SL-OCT UNIT 33 8 APPENDIX A: TROUBLESHOOTING 35 9 APPENDIX B: LIST OF SCREWS 37 10 APPENDIX C: PACKING LISTS 38
10.1 Lift table 38
10.2 OCT-box 39
10.3 Slit lamp 40
10.4 Optional printer table 42
SL-OCT Installation Manual
Art. No.: 18107 Page 4 of 42
1 Introduction
The SL-OCT is a tomographic device for the viewing and axial, cross sectional imaging of anterior ocular structures. It is used for the in vivo imaging and measurement of the anterior segment structures of the eye. This document describes the installation of the SL-OCT device. It also contains important safety information.
NOTE: To ensure the overall safety of the system, the SL-OCT must be installed by certified staff
of your local Heidelberg Engineering distributor.
Scanner-box
Slit lamp
Printer table (optional)
Lift table
OCT-box
Pic. 1 SL-OCT with optional printer table
SL-OCT Installation Manual
Art. No.: 18107 Page 5 of 42
1.4 Cautions, Warnings, and Contraindications WARNING Carefully read the operation manual before operating the device. Misuse of the device
may lead to incorrect diagnostic results.
WARNING Do not open the device component housings. Doing so may result in electrical shock and
laser radiation.
WARNING Do not use the device outside the scope of its “Intended use”. Doing so may lead to
malfunctions or damage of the device.
WARNING Do not use PCs, components or accessories that have not been approved by Heidelberg
Engineering. Do not install other software as this may interfere with the functionality of the Heidelberg Engineering software or equipment. This could include damage to the system as well as incorrect measurement results.
WARNING Do not use a network connection without network isolation in accordance with
IEC 60601-1. In the event of a failure in the network, user and patient could be at risk of electrical shock.
WARNING Make sure that the environmental requirements are met when the system is operated.
Exceeding environmental conditions may damage the system or lead to incorrect measuring results.
WARNING Make sure the patient is correctly positioned in front of the device before starting the
examination. Wrong positioning may lead to poor images and incorrect diagnostic results.
WARNING Artifacts on the images could falsify the measured results. Do not use the measured
results for further treatment if there are artefacts in the images.
WARNING Do not make a diagnostic decision on the basis of one single examination. The device is
not clinically evaluated for the diagnosis of specific pathologies. So always use alternative information; history data etc. to assist in a final diagnostic determination.
WARNING Do not use the analysis results from patients with pathological eyes without an inspection
of the segmentation. In case of pathological eyes, the segmentation might be incorrect, and lead to wrong calculations of the analysis results.
WARNING Prepare safeguards to ensure that only authorized personnel can access the patient
data. Data loss impedes follow-up analyses and may result in inappropriate diagnostic decisions.
WARNING Be sure to perform periodic data backup procedures. Check the success of the backup to
avoid data loss caused by backup errors.
WARNING To avoid the risk of electric shock, this equipment must only be connected to a grounded
power supply.
WARNING To avoid the risk of electric shock, do not touch conductive parts of connectors and the
patient simultaneously.
WARNING Do not operate the system directly after large temperature changes. Let the device
acclimatize itself for a minimum of 2 hours to avoid device damage or incorrect measurement results.
CAUTION Never leave the patient alone with the instrument during the examination! CAUTION The instrument must not be used under any circumstances if mechanical, optical or
electrical faults occur. Any change or addition to the system must comply with the relevant legal guidelines. Repairs, particularly to the electronic and optical components, must be carried out only by Heidelberg engineering authorized, trained personnel.
CAUTION Unusual noises and/or vibrations can indicate a fault. Should this happen, please turn the
instrument off immediately and contact the technical support center responsible for your area. Do not attempt to repair the instrument yourself in the event of a fault.
CAUTION This instrument contains a diode laser and emits invisible laser light through the slit lamp.
The Heidelberg SL-OCT is a Class I laser system. The laser does not pose any safety hazard.
CAUTION This equipment was tested in accordance to IEC 60601-1-2, Electromagnetic
Compatibility (EMC). Nevertheless, it might be affected by strong electromagnetic fields. Portable high frequency communication devices may affect the device.
CAUTION The operator must be sure that the device settings and adjustments are correct before
starting an examination and making any diagnostic decision. Wrong settings and adjustments may lead to poor image quality or incorrect examination information.
CAUTION The physician must be sure to have the correct patient data before making a diagnostic
decision. Mismatched patient data may lead to inappropriate diagnostic decisions.
SL-OCT Installation Manual
Art. No.: 18107 Page 6 of 42
CAUTION Read subsection “Imaging Process” of the Operation Instructions carefully before starting
the examination. Incorrect preparation of the patient may lead to poor image quality and incorrect diagnoses.
CAUTION Do not start an examination without informing the patient about the examination
procedure. Inappropriate patient behavior during the examination may lead to poor image quality and incorrect diagnoses.
CAUTION Read subsection “Service Maintenance and Cleaning” of the Operation Instructions
carefully. A failure to carry out maintenance or incorrect adjustment of the device may lead to poor image quality and incorrect diagnoses.
CAUTION Before starting the system check the regional power supply specifications to verify that
they comply with the required tolerances (100V < U < 240V; 50Hz < f < 60Hz). Wrong power supply conditions may lead to malfunctions of the system.
CAUTION A computer failure during image acquisition or analysis could lead to incorrect results. CAUTION United States of America:
Federal law restricts this device to sale by or on the order of a Physician or Practitioner.
CAUTION While printing using the optinal printer table, it is not possible to take measurements due
to the vibrations of the printer
CONTRAINDICATIONS No contraindications are known.
For the United States of America only:
The laser class label “Laser Class 1” is located on the rear of the OCT housing. The Laser safety class is approved and defined in accordance to IEC 60825 part 1 and part 2 and in accordance to the 21 CFR Part 1040 “Performance Standard for Light emitting Products”.
2 Scope
This installation instruction is referring to model SL-OCT Ver. B01.
SL-OCT Installation Manual
Art. No.: 18107 Page 7 of 42
3 Tools
The following tools are required:
 Allen wrenches
o 5 mm o 3 mm o 2.5 mm
 Screwdriver for slotted head screws 3 mm  Wire cutter  Cutter.
For the assembly of the optional printer table, following the tools are required:
 Allen wrench with ball head
5 mm
 Screwdriver for recessed head screws  Wire cutter  Long nose pliers  Cutter.
Pic. 2 Tools for installation of SL-OCT
Pic. 3 Tools for the assembly of the printer table
SL-OCT Installation Manual
Art. No.: 18107 Page 8 of 42
4 Unpacking the system
The SL-OCT is delivered in three cardboard boxes – lift table, OCT-box, and slit lamp. The packing lists for each box are listed in section 10 Appendix C.
4.1 Lift table
The lift table package contains three layers:
1. Cable conduit, column, keyboard fixation arm, and monitor fixation arm.
2. Cables, wheels, and table top.
3. Foot.
Pic. 4 Cardboard box containing the lift table
4.2 OCT-box
The OCT-box package contains two layers:
1. Folder with operating instructions etc., and mouse pad.
2. Cables, mouse, dust cover, protective caps, replacement fuses, OCT-box.
4.3 Slit lamp
The slit lamp package contains two layers:
1. Head rest in a separate cardboard box.
2. Monitor, slit lamp with permanently mounted scanner-box, slit lamp accessories in a separate cardboard box, eyepieces for the microscope, and keyboard.
Pic. 5 Cardboard box containing OCT-box
Pic. 6 Cardboard box containing slit lamp
SL-OCT Installation Manual
Art. No.: 18107 Page 9 of 42
5 Installation
The installation of the SL-OCT unit is started with the build up of the lift table. The next part is the assembly of the OCT-box onto the lift table. Afterwards, the slit lamp including the permanently mounted scanner-box is installed onto the lift table. At the end of the installation procedure, test measurements are carried out to ensure correct operation of the device.
5.1 Lift table
The first step is the assembly of foot and column. Take the column out of its package and place it on top onto e.g. a cellular plastic backing (see Pic. 7).
cellular plastic backing
Pic. 7 Column prepared for assembly of foot
Remove the four Allen screws (M6 x 40) and spring washers needed for the assembly of the foot.
All screws and washers necessary for assembly are provided, slightly screwed into their respective screw holes!
Control the cable laying. The mains cable of the column must lay in the slit (see Pic. 9).
Pic. 8 Premounted screws for
assembly of the foot
Pic. 9 Cable laying of mains cable
SL-OCT Installation Manual
Art. No.: 18107 Page 10 of 42
Mount the foot onto the column (see Pic. 10). The longer part of the foot must be oriented towards the the cable outlet of the column. The computer screen will later be mounted onto this side of the table. Therefore, the correct orientation of the foot is essential for the stability of the table.
Pic. 10 Foot onto column mounted
Now clip in the four wheels onto the foot.
Pic. 11a Wheels
Pic. 11 Clip in the wheel
Turn around column and mounted foot (see Pic. 12a). Then, connect mains and lift the column until the stop at end (see Pic. 12).
Attention: Be very careful not to touch the cables in the open control console as they are alive! As soon as the column is lifted, disconnect the mains!
Pic. 12a Column not lifted
Pic. 12 Column lifted till stop at end
SL-OCT Installation Manual
Art. No.: 18107 Page 11 of 42
After disconnecting mains, the tabletop is prepared. Remove the screws and washers from the screw holes for the column (M6 x 25 and spring washers) and control console (M6 x 16 and spring washers) (see Pic. 13 red and green).
Multible power socket
Assembly of head rest
Assembly of control console
Assembly of column
Pic. 13 Table top (bottom) with premounted screws and washers
Mount the tabletop onto the column. Use the short Allen screws for the console and the long ones for the column.
Assembly of monitor fixation
Pic. 14 Mounting of table top onto the column
If there is an optional printer arm delivered with the SL-OCT, it can be assembled at this state of the installation or at the end of the assembly procedure. Please refer to section 5.2.
SL-OCT Installation Manual
Art. No.: 18107 Page 12 of 42
Pic. 15a rear side of mounted monitor unit
Pic. 15 Mounted monitor including keyboard
fixation arm
The computer screen is mounted together with the keyboard fixa tion arm on its fixation arm. The other end of the monitor fixation arm is mounted beneath the tabletop (see Pic. 15a and Pic. 15).
There are two possible ways for mounting the monitor unit onto the table:
1. Premounting of monitor arm, keyboard arm, and monitor before assembly of the whole unit onto the lift table. This is recommended if several persons are installing the device together.
2. Premounting of monitor arm and lift table before assembly of keyboard arm and monitor. This might be more comfortable if there is a single person installing the device. It is easy to adjust the monitor fixation arm with one hand while looking beneath the tabletop and mounting the screws. Handling the whole premounted unit with one hand while adjusting the screws beneath the table needs more strength.
Avoid touching the surface of the screen.
Pic. 16a Monitor arm and keyboard arm
Pic. 16 Rear side computer screen with its screws
SL-OCT Installation Manual
Art. No.: 18107 Page 13 of 42
1. Premounting of monitor unit
Put the computer screen with the screen facing downwards onto a table parallel to the table’s edge (see Pic. 16). Remove the screws (Allen screws M4 x 20). Place the keyboard arm onto the monitor with the angled part hanging down. Now place the monitor arm on top and screw all parts together (see Pic. 17).
Pic. 17 Mounted monitor unit
Remove the screws from the tabletop (see magenta arrow in Pic. 13 and Pic. 18).
Pic. 18 rear side of tabletop in detail
Place the monitor unit so that the keyboard arm is on top of the lift table and the angled part of the monitor arm is beneath the table (see Pic. 15a). Assembly the 4 Allen screws (M6 x 20). Adjust the computer screen into the right position by turning it slightly around the middle screw before fastening the screws.
2. Premounting of the monitor arm
Remove the screws from the tabletop (see Pic. 13 magenta and Pic. 18). Assemble the monitor arm (see Pic. 16a) at the lift table (see Pic. 20). Do not fully fasten the screws (M6 x 20) until the computer screen is mounted and adjusted. Plug in mains and lower the lift table to a convenient height.
Pic. 19 Mounted monitor arm
Pic. 20 Mounted monitor arm
Loading...
+ 29 hidden pages