Heidelberg HRT3 Installion manual

Rostock Cornea Module
Heidelberg Retina Tomograph 3
Installation Instructions
Revision 1.3-1E, June 2006 © Heidelberg Engineering GmbH 2006 Printed in Germany Art. No. 11223 / 97 085-002
The manufacturer hereby declares that this product conforms to the requirements of Directive 93/42/EEC of the Council of the European Community dated 14 June 1993 regarding medical products (MDD 93/42/EEC).
Caution! Do not use the Heidelberg Retina Tomograph 3 / Rostock Cornea Module without reading the Operating Instructions. User errors may result in a false diagnosis!
Table of Contents
1 HRT3 / RCM System Setup and Installation..........................................................5
1.1 Computer and System Requirements...................................................................5
1.2 System Components ............................................................................................5
1.2.1 Detachable Components and Accessory......................................................5
1.2.2 Camera Mount .............................................................................................5
1.2.3 Additional Components...............................................................................6
1.2.4 Cables...............................................................................................................6
1.3 Installation ...........................................................................................................7
1.3.1 Replacing the HRT 3 Front Panel....................................................................7
1.3.2 Mounting the Headrest and the Chinrest .....................................................9
1.3.3 Mounting the RCM Objective and the CCD Camera................................11
1.3.4 Attachment of the cables............................................................................12
2 Software Installation................................................................................................13
2.1 RCM Operating Software Installation...............................................................13
2.1.1 Software Upgrade......................................................................................13
2.1.2 Software Update ........................................................................................14
2.2 CCD Camera Driver Installation.......................................................................15
2.2.1 Very first Camera Driver Installation........................................................15
2.2.2 Update of the Camera Driver.....................................................................21
2.3 HRTC.INI Settings ............................................................................................26
2.3.1 Section [BoardType]..................................................................................26
3 Troubleshooting.......................................................................................................27
3.1 No CCD Image is Displayed During Image Acquisition. .................................27
3.1.1 Windows 2000...........................................................................................27
3.1.2 Windows XP..............................................................................................28
3.2 Wrong or Missing Display of Focus Position....................................................28
3
4

1 HRT3 / RCM System Setup and Installation

1.1 Computer and System Requirements

The Heidelberg Retina Tomagraph 3 / Rostock Cornea Module (HRT 3 / RCM) is both a hardware and a software add-on to the HRT 3. The operation of the RCM requires full hardware installation of the HRT 3. No additional PC or laptop requirements for the HRT 3 are required.

1.2 System Components

The HRT 3 / RCM comprises all components listed below. The headrest must be perma­nently modified to image with the RCM. All other components are detachable and are stored together with the accessory tools in the wooden case.

1.2.1 Detachable Components and Accessor y

When the HRT 3 is operated in glaucoma or retina mode, the components of the HRT 3 / RCM should be kept in the wooden case provided:
1 2
- (1) Rostock Cornea Module objective
- (2) IEEE 1394 CCD camera and camera mount
- (3) Front cap wrench
- (4) Tool for demounting the field lens
- (5) Alignment pin
- (6) Field lens, FOV 300 µm
- (7) Standard cornea gel (Comfort Gel™ , Dr. Mann Pharma, Europe or GenTeal Gel™, Novartis Ophthalmics, US).

1.2.2 Camera Mount

The camera mount of the HRT 3 / RCM dif­fers from the standard HRT 3 camera mount in:
3 4 5 6
8
1 2
9
10
- (8) an adjustable forehead rest
5
- (9) a telescopic chin rest
- (10) a front panel, internally equipped with
the RCM board
These three components replace the original components of the standard HRT 3 camera mount.
If these components are not already installed, please see sections 1.3.1 and 1.3.2 for installa­tion instructions.

1.2.3 Additional Components

Additional components included in delivery:
- Sterile TomoCaps (50 pcs.)
- HRT / RCM Software Installation CD-ROM
- Operation Instructions Manual
10
- Installation Instructions Manual
1.2.4 Cables
Delivery of the HRT 3 / RCM encloses two interconnection cables:
- (11) IEEE 1394 cable: connects the CCD­Camera to the front panel of headrest
- (12) RCM cable: connects the RCM ob­jective to the front panel of headrest
11 12
6

1.3 Installation

Generally, the HRT 3 / RCM will be delivered with the appropriate camera mount. This means the headrest will already be adapted to the HRT 3 / RCM including the RCM chin rest, the RCM forehead rest and the front panel (see 1.2.2). In this case, please proceed to section 1.3.3.
Note: The HRT 3 / RCM headrest can be used for all exam modes: glaucoma, retina and
cornea.
In the case that an existing HRT 3 needs to be upgraded to HRT 3 / RCM, the compo­nents mentioned above have to replace the corresponding components of the HRT 3. The mount is described in 1.3.1 and 1.3.2.
1.3.1 Replacing the HRT 3 Front Panel
Detachment of the HRT 3 front panel:
Unscrew the front panel at the bottom of the HRT 3 (see arrows). For this purpose, the HRT 3 can be positioned to slightly exceed the edge of a table.
Open the front panel and detach the protec­tive earth, then remove the front panel.
7
Installation of the RCM front panel:
Connect the protective earth exactly as it is shown in the figures. The cable must pass be­low the RCM board.
Connect the ribbon cable of the HRT 3 / RCM board to the appropriate port on the carrier board of the HRT 3.
Close the front panel - the ribbon cable
e should become S-shaped - and tighten th
screws.
8
1.3.2 Mounting the Headrest and the
Chinrest
Remove the two small black caps (1) on the top of the HRT 3 mount (using a small screw driver, if required). Then remove the screws.
Detach the headrest, then detach the chin rest (2). Mount the RCM chin rest (3), then mount the RCM headrest (4).
(3)
(1)
(4)
(2)
Make sure that you do NOT invert the orien­tation of the headrest! The black screw (see arrow) will be located on the right side seen from the patient’s point of view.
Tighten the headrest with the scre fore. Insert the two black caps.
o switch between Cornea Module and
T Glaucoma / Retina Module image acqui tion:
Unbloc
k the black screw, flip the headrest and tighten it again.
ws as be-
si-
9