Heidelberg HRA2 ERG Upgrade User manual

HRA2 ERG Upgrade
Art. Nr.: 97–078–001 Page 1 / 4 2006-04-03
Heidelberg Retina Angiograph 2
Electro-Retinography (ERG) Upgrade
Installation and Use
This document describes the Heidelberg Retina Angiograph 2 (HRA2) upgrade for electro­retinography. This upgrade is needed for use of the HRA2 together with third party ERG de­vices.
1. Connect the PC to the HRA2 as described in the HRA2 manual.
2. Switch on the HRA2.
3. Insert the CD labeled “Heidelberg Retina Angiograph 2 – ERG upgrade” into the CD-ROM drive. The installation program will start automatically after a few seconds. The following window will be shown:
HRA2 ERG Upgrade
Art. Nr.: 97–078–001 Page 2 / 4 2006-04-03
4. Compare the power supply serial number to the serial number on your upgrade CD. The serial numbers must correspond otherwise the upgrade will fail. If the serial num­bers differ or if an error message appears that complains about the current firmware version, please contact our support.
5. Press the “Yes“ button and then wait for the upgrade to complete. Do not switch off the system during upgrade. The HRA2 system will perform a reset after a successful up­grade.
6. Switch off the HRA2 system.
Loading...