Heath Consultants Heath Chart Recorder User Manual

Mechanical Recorder Operation Manual (Press &/or Temp)
1. Place the instrument in the desi red locati on near the pressure & temperature source to monitor.
2. Connect the pressure line to the ¼” female NPT fitting on the back of the instrument. Note: Wall mount fittings will be on the bottom.
3. If Required: Connect the temperature probe into a thermowell or sui table location using the ¼” Male NP T sli p-fitting provided to allow i nstallation i nto different depths of pipe.
4. Remove the chart plate & install the chart drive battery & replace the battery cover.
5. Turn the selector knob (on the chart-drive) to the desired setti ng “Per revolution” (Ex am ple: 1= 24 h ours, 7= 1 week, 30= month ) .
6. Replace and secure the chart plate into the i nstrument us ing t he ( 4 ) char t -plate screws. Note: These screws do not need to be removed- only loosened.
7. Use the pen lifter to raise the pens off the chart.
8. Install a paper chart onto the chart drive hub and make sure the chart is under all (4) chart-plate tabs.
9. Remove the pen tips from both pens.
10. Align the chart to where the correct time &/or date i s at the Ti meline poi nter.
11. Lock down the chart to the hub by pressing “horse-shoe shaped” hub lock towards the center of the chart drive shaft. Note: The old-style hub - tightens clockwise and loosens co unte r-clockwise.
12. Apply pressure to the recorder- if thi s hasn’t been done already.
13. Lower the pens onto the chart. (The pressure pen should be on the zero-line when no pressu r e is applied).The chart should turn and the pens should record the pressure and temperature.
14. After the duration of the chart is over, repeat steps 7-14.
Pressure pen (your pen color may be different)
Temperature pen
(your pen color
Chart hub (your hub
pic)
Arc-line or Timeline
Tim eline pointer
Zero line on chart
Chart plate tab
Chart plate tab
Chart-plate s c rew s
Pen Lifter
may be different)
will be different- this is old style hub in the
Loading...