HEALTHCARE
Imaging Services
Installing the CR 10-X at NX 2.0.8700 / 3.0.8700
Quick Installation Guide
A604993.1 DD+DIS171.11E
Safety
WARNING:
Improper operation or service activities
may cause damage or injuries.
INSTRUCTIONS:
(1) Read the "Generic Safety Directions"
document (See digitizer service manual)
prior to attempting any operation, repair
or maintenance task on the equipment.
(2) Strictly observe all safety directions within
the "Generic Safety Directions" and on the
product.
(3) Strictly observe all safety directions in
safety.cdr
chapter 3.1 of the digitizer service manual.
Workstation Setup
The following steps describe:
Loading the license file at NX
Running the CR 10-X config wizard at NX
Synchronizing the configuration database at NX
Connections
The following steps describe:
Connections at NX workstation
(1) Perform connections at workstation.
License
Dongle
(2) Perform connections at digitizer.
connection
Direct
network
Connections at digitizer
Mains
(3) Check NX SW version and switch on devices.
A
Check label on PC. It must show:
Installed software: NX_X.0.8700
(4) Load license.
Copy the application license file (ALF) to the NX directory:
A
C:\AGFA\Healthcare\NX\Configuration\Licensing
For details refer to CR 10-X
Service Manual chapter 1.
B
Switch on both devices.
B
C
Starts
red
C
blinking.
Enter User Name / Password
obtained in Web
based Training .
install_poster_04.cdr
User name
Password
D
E
HDMI to
DVI converter
Mains
(optional)
install_poster_02a.cd
Double-click the Licensing Manager
icon on the desktop.
Select and browse to the folder
<Add License>
where the ALF file is stored. Click: .
License
Manager
<open>
Wait a few seconds until the message appears:
The licenses have been updated. <OK>.
Close the License Manager.
Press
install_poster_05.cdr
(5) Copy CR 10-X config wizard to D:\drive.
A
Insert digitizer USB flash drive at NX.
Mains
install_poster_003.cdr
Copy the file in folder
B
CR10–X Configuration Wizard
from USB flash drive to drive on the NX.
Copy the file in folder
C
ErrorViewer
from USB flash drive to drive on the NX.
.exe
.msi
D:\
D:\
install_poster_06.cdr
(6) Re-insert USB flash drive at digitizer. (7) Run the CR 10-X config wizard. (8) Continue with the CR 10-X config wizard. (9) Continue with the CR 10-X config wizard.
install_poster_07a.cdr
Region and Language
Double-click file:
Keyboards and Languages
D:\ActivateNX4CR10.NX8700.XX.exe
Warning
Choose Language
Ready to be activated
Date and Time
OK
Next >
Date and Time
Date and Time
Change date and time...
Time zone
Change time zone...
Notify me when the clock changes
OK Cancel Apply
Next >
Activate
Change time zone...
Change date and time...
install_poster_08.cdr
Text Services and Input Languages
Default input Language
English
Installed Services
DE
German (Germany)
Keyboard
German
OK
OK
ab: Keyboards and Languages
Change keyboards...
OK Cancel Apply
Add
Cancel
Change keyboards...
Add keyboard (if required)
Change Input Language (if required)
Apply
Select tab "Administrative".
install_poster_09.cdr
Region and Language
Administrative
Change system locale
OK Cancel Apply
OK
Tab: Administrative
Copy settings...
Copy settings...
Activating (takes approx. 5 minutes)
Next
Reboot now
Welcome screen
New user accounts
Finish
Page 1 of 2
Agfa Company Confidential
Copyright © 2012 Agfa HealthCare N.V.
Type 5151/100
Document ID 34000041
Welcome screen and new user accounts settings
Current user
Welcome screen
New user accounts
Welcome screen and system acc ounts
New user accounts
install_poster_10.cdr
OK Cancel
OK
Quick Installation Guide
Installing the CR 10-X at NX 2.0.8700 / 3.0.8700
DD+DIS171.11E
(10) Login and stop NX. (11) Synchronize NX database.
A
Login
B
Login
C
Double-click "Stop NX"
Follow Instructions
D
Stop NX
C:\Agfa\Healthcare\N\Bin\Service>echo off
*************************************************
ARE YOU SURE?
*************************************************
Press any key to continue ...
User name
Password
User name
Password
(13) Expose flatfield.
1,3 m
Automatic
reboot
NX
Stop NX
Insert CU Filter
A
B
Expose Cassette:
Large Focus
12 mAs
75 kVp
C
Rotate cassette by 180°
Expose again
D
Remove CU Filter
E
install_poster_22.cdr
A
Double-click: "HostName Synchronizer"
B
Click: "Synchronize"
C
Wait a few seconds for message:
<<<Synchronization of hostname succeeded ! >>>
D
Close the HostName Synchronizer.
E
Double-click: "Restart NX"
and press any key to continue.
After restart of NX, digitizer status LED turns green.
(14) Insert and identify cassette.
Insert cassette with
A
black side facing up.
NX
Restart NX
B
Identify cassette.
HostName
Synchronizer
in
System
Diagnosis
Genrad
Flat
Field
200
r
Collimated
Area
Black side
facing up
install_poster_013.cdr
Wait approx. 3 minutes until image transfer is complete.
C
installposter 25.cdr
Installation Verification
The following steps describe:
Erasing a cassette
Creating a flatfield image
Evaluating the image quality
(15) Evaluate flatfield image quality.
If image quality is not OK:
See digitizer service manual
chapter "Troubleshooting".
install_poster_15.cdr
(12) Erase cassette.
Press the "Erase" button.
A
Insert cassette with
B
black side facing up.
Wait approx. 3 minutes until
C
status LED turns green.
D
Remove cassette.
Installation Finalization
The following steps describe:
Performing the digitizer installation wizard
Creating a backup
Finalizing the installation
install_poster_24.cdr
(16) Connect to digitizer service GUI. (17) Perform digitizer installation wizard. (18) Create a backup. (19) Finalize Installation.
192.192.192.192
Start
A
Double-click the Internet Explorer icon.
B
Enter "192.192.192.192" in the address window.
C
Enter user name and password
obtained in the Web based Training.
install_poster_16.cdr
Authentication Required
User Name:
Password:
OK
Authentication Required
User Name:
Password:
A
Digitizer starts up automatically:
Cancel
Digitizer Settings
Destination Settings
Start
Site Specific Data
Installation Date Confirmation
Cancel
OK
Installation Wizards
Adapt local settings
Do not change
Adapt
Next
Next
Next
Confirm installation
install_poster_17.cdr
Select Service Menu "Backup".
B
Click: "Save to USB"
C
Click:
Logout
Backup
Complete Backup
Backup single files only
Select singe files (active only if "Backup single files
only" selected):
Infocounter Logfiles
Calibration Files
Service Activity Log
install_poster_18.cdr
The mandatory connection and configuration steps for a
standalone system are finished.
Refer to the CR 10-X service manual chapter 1 for:
Info about current system configuration status and for
Following additional steps:
o Mandatory:
- Perform tests according to national regulations
- Train customer
o Optional, depending on the local situation:
- Configure NX
- Add additional components
- Put NX and digitizer in the local network
Page 2 of 2
Agfa Company Confidential
Copyright © 2012 Agfa HealthCare N.V.
Type 5151/100
Document ID 34000041