HEALTHCARE CR 10-X User Manual

HEALTHCARE
Imaging Services
Document No: DD+DIS205.12E
This number is needed for production line only: A604725.1
Enclosure
CR 10-X
Type 5151 / 100
Replacement Instructions for the
"PMT with Light Collector"
Purpose of this Document
This enclosure is part of the spare part kit "PMT with Light Collector", spare part number*: CM+6040420
It describes all necessary steps for the replacement.
* The last digit in the spare part number indicates the spare part revision at release of this document. When ordering, the actual revision of the spare part is delivered.
Referenced Documents
Document Title
None
NOTE:
Revision 0 09-2012 printed in Germany
Agfa Company Confidential
Use, dissemination, distribution or reproduction of this document by unauthorized personnel is not permitted and may be unlawful.
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
CONFIDENTIALITY NOTE:
DOCUMENT CONTROL NOTE:
Document Node ID: 33997185
Enclosure_e_template_v07
Copyright © 2012 Agfa HealthCare N.V.
DD+DIS205
.12E
Manufacturer
Agfa HealthCare N.V.
Publisher
Agfa-Gevaert HealthCare GmbH Tegernseer Landstraße 161 D - 81539 München Germany
Copyright 2012 Agfa HealthCare N.V.
All rights reserved. Technical modifications reserved.
Agfa and the Agfa rhombus are trademarks of Agfa-Gevaert N.V., Belgium, or its affiliates.
Enclosure
WARNING:
Improper operation or service activities may cause damage or injuries.
INSTRUCTION:
(1) Read the "Generic Safety Directions" document
(see Agfa HealthCare Library => General Info => Agfa HealthCare => Publications => Service Manual) prior to attempting any operation, repair or maintenance task on the equipment.
(2) Strictly observe all safety directions within the "Generic Safety Directions" and on
the product.
CAUTION:
Static discharge! Electrical components may be destroyed:
For the repair on electrical components, wear a grounding strap (Order number: CM+9 9999 0830 0) around the wrist and connect the other end of this strap on a grounded conducting metal piece.
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0 CR 10-X Page 2 of 16 09-2012 Type 5151 / 100 Agfa Company Confidential
DOCUMENT CONTROL NOTE:
DD+DIS205
.12E
Enclosure
LIST OF CONTENTS
REPLACING THE PMT WITH LIGHT COLLECTOR ...............................................................6
1
2 PERFORMING A SOFTWARE UPDATE IF CURRENT SOFTWARE < ARC_1302 .............10
3 PERFORMING A SHADING CALIBRATION..........................................................................11
3.1 Checking and Erasing the Image Plate ..................................................................................11
3.2 Exposing the Cassette ............................................................................................................12
3.3 Scanning the Image Plate.......................................................................................................13
3.4 Verifying a successful Shading Calibration.............................................................................15
3.5 Appendix: Possible Reasons for a failed Shading Calibration................................................16
4 VERIFYING THE REPLACEMENT ........................................................................................16
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0 CR 10-X Page 3 of 16 09-2012 Type 5151 / 100 Agfa Company Confidential
DOCUMENT CONTROL NOTE:
*
DD+DIS205
Subject
.12E
Replacement Instructions "PMT with Light Collector"
SCOPE OF DELIVERY:
REQUIRED TOOLS:
Enclosure
Figure 1
TX20
1,5 mm Copper Filter** Spare part number*: CM+9 5155 1015 2
Polynit wipes***
Spare Part Number*:
10+9 9999 1273 0
CR Phosphor Plate Cleaner***
Spare part number*:
10+9 9999 1197 0
Lint free cloth**
* The last digit in the spare part number indicates the spare part revision at release of this document. When ordering, the actual revision of the spare part is delivered.
** The Cu filter is part of digitizer delivery.
*** For cleaning of an image plate either use Polynit wipes or CR Phosphor plate cleaner and a lint free cloth.
REQUIRED SOFTWARE:
The PMT with Light Collector which is part of delivery requires software ARC_1302.
REQUIRED TIME:
Approximately 55 minutes
20 minutes exchange
15 minutes software update (if required)
20 minutes shading calibration including verification
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0 CR 10-X Page 4 of 16 09-2012 Type 5151 / 100 Agfa Company Confidential
DOCUMENT CONTROL NOTE:
DD+DIS205
.12E
Enclosure
CAUTION:
The PMT light sensitivity decreases for a certain time (hours to days) if exposed to bright light.
Expose the light sensitive side of the PMT as little as possible to daylight.
CAUTION:
The Light Collector can break when mechanically stressed.
Handle the Light Collector with great care.
CAUTION:
Finger prints on the mirror before the Light Collector may cause reduced image quality.
Be careful not to touch the mirror.
Mirror before Light Collector
515131ak.cdr
Figure 2
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0 CR 10-X Page 5 of 16 09-2012 Type 5151 / 100 Agfa Company Confidential
DOCUMENT CONTROL NOTE:
Loading...