Hottinger Baldwin Messtechnik GmbH
Im Tiefen See 45
D-64239 Darmstadt
Tel. +49 6151 803-0
Fax +49 6151 803-9100
info@hbm.com
www.hbm.com
Mat.: 7-2002.4354
DVS: A4354-2.2 HBM: public
11.2016
E Hottinger Baldwin Messtechnik GmbH.
Subject to modifications.
All product descriptions are for general information only.
They are not to be understood as a guarantee of quality or
durability.
The PMX amplifier system is to be used exclusively for
measurement tasks and directly related control tasks.
Use for any purpose other than the above is deemed to
be non-designated use. In the interests of safety, the
device should only be operated as described in the
Operating Manuals. It is also essential to comply with the
legal and safety requirements for the application
concerned during use. The same applies to the use of
accessories.
Each time, before starting up the equipment, you must
first run a project planning and risk analysis that takes
into account all the safety aspects of automation
technology. This particularly concerns personal and
machine protection.
Additional safety precautions must be taken in plants
where malfunctions could cause major damage, loss of
data or even personal injury. In the event of a fault, these
precautions establish safe operating conditions.
Safety instructions
This can be done, for example, by mechanical
interlocking, error signaling, limit value switches, etc.
Notice
The device must not be directly connected to the DC
voltage supply system. Supply voltage 10 V … 30 V
(DC).
Before commissioning, ensure that a suitable supply
voltage is used and that the circuit used is sufficiently
protected.
PMXA4354-2.2HBM: public13
Safety instructions
The device may only be supplied with separated extralow voltage (safety isolating transformer according to DIN
VDE 0551 / EN60742). Only operate built-in devices once
they are installed in the housing provided. The device
development is based on DIN EN 61010 Part 1 (VDE
0411 Part 1).
General dangers of failing to follow the safety
instructions
The PMX system is a state of the art unit and as such is
reliable. The module may give rise to dangers if it is
inappropriately installed and operated by untrained
personnel.
Any person instructed to carry out installation, starting up,
maintenance or repair of the module must have read and
understood the Operating Manual and in particular the
technical safety instructions.
Conditions at the place of installation
SProtect the device from direct contact with water.
SProtect the PMX system from moisture and humidity
or weather conditions such as rain, snow, etc.
SDo not expose the device to direct sunlight.
SPlease observe the permissible maximum ambient
temperatures stated in the specifications.
SThe permissible relative humidity at 31 °C is 95 %
(non condensing); linear reduction to 50 % at 40 °C.
SInstall the device so that it can be disconnected from
the supply voltage at any time without difficulty.
SIt is safe to operate the PMX system up to a height of
2000 m.
14A4354-2.2HBM: publicPMX
Safety instructions
Maintenance and cleaning
The PMX system is maintenance-free.
SBefore cleaning, disconnect all connections.
SClean the housing with a soft, slightly damp (not wet!)
cloth. You should never use solvents, since these
could damage the labeling.
SWhen cleaning, ensure that no liquid gets into the
module or connections.
Residual dangers
The scope of supply and performance of the PMX system
only covers a small area of measurement technology. In
addition, equipment planners, installers and operators
should plan, implement and respond to the safety
engineering considerations of measurement technology
in such a way as to minimize remaining dangers. On-site
regulations must be complied with at all times. There
must be reference to the remaining dangers connected
with measurement technology.
Product liability
In the following cases, the protection provided for the
device may be adversely affected. Liability for device
functionality then passes to the operator:
SThe device is not used in accordance with the
operating manual.
SThe device is used outside the field of application
described in this Chapter.
SThe operator makes unauthorized changes to the
device.
PMXA4354-2.2HBM: public15
Safety instructions
Warning signs and danger symbols
Important instructions for your safety are specifically
identified. It is essential to follow these instructions in
order to prevent accidents and damage to property.
Safety instructions are structured as follows:
WARNING
Type of danger
Consequences of non-compliance
Averting the danger
SWarning sign:
draws your attention to the danger
SSignal word:
indicates the severity of the danger (see table below)
SType of danger:
mentions the type or source of the danger
SConsequences:
describes the consequences of non-compliance
SDefense:
indicates how the danger can be avoided/bypassed
16A4354-2.2HBM: publicPMX
Danger class according to ANSI
SymbolSignificance
WARNING
CAUTION
Notice
This marking warns of a potentially dangerous
situation in which failure to comply with safety
requirements can result in death or serious physical
injury.
This marking warns of a potentially dangerous
situation in which failure to comply with safety
requirements can result in slight or moderate physical
injury.
This marking draws your attention to a situation in
which failure to comply with safety requirements can
lead to damage to property.
Working safely
Notice
The device must not be directly connected to the DC
voltage supply system. Supply voltage 10 V … 30 V
(DC).
Safety instructions
The supply connection, as well as the signal and sense
leads, must be installed in such a way that
electromagnetic interference does not adversely affect
module functionality (HBM recommendation: "Greenline
shielding design", downloadable from the Internet at
http://www.hbm.com/Greenline
).
Automation equipment and devices must be installed in
such a way that adequate protection or locking against
unintentional actuation is provided (e.g. access checks,
password protection, etc.).
When devices are working in a network, these networks
must be designed in such a way that malfunctions in
individual nodes can be detected and shut down.
PMXA4354-2.2HBM: public17
Safety instructions
Safety precautions must be taken both in terms of
hardware and software, so that a line break or other
interruptions to signal transmission, such as via the bus
interfaces, do not cause undefined states or loss of data
in the automation device.
Conversions and modifications
The device must not be modified from the design or
safety engineering point of view except with our express
agreement. Any modification shall exclude all liability on
our part for any damage resulting therefrom.
In particular, any repair or soldering work on
motherboards is prohibited. When exchanging complete
modules, use only original parts from HBM. The product
is delivered from the factory with a fixed hardware and
software configuration. Changes can only be made within
the possibilities documented in the operating manual.
Qualified personnel
Qualified personnel means persons entrusted with siting,
mounting, starting up and operating the product, who
possess the appropriate qualifications for their function
(qualified electrician, or by someone with electrical
training under the supervision of a qualified electrician).
This device is only to be installed and used by qualified
personnel strictly in accordance with the specifications
and with the safety rules and regulations which follow.
This includes people who meet at least one of the three
following requirements:
SKnowledge of the safety concepts of automation
technology is a requirement and, as project personnel,
you must be familiar with these concepts.
18A4354-2.2HBM: publicPMX
Safety instructions
SAs automation plant operating personnel, you have
been instructed how to handle the machinery. You are
familiar with the operation of the equipment and
technologies described in this documentation.
SAs commissioning engineers or service engineers,
you have successfully completed the training to qualify
you to repair the automation systems. You are also
authorized to activate, ground and label circuits and
equipment in accordance with safety engineering
standards.
It is also essential to comply with the legal and safety
requirements for the application concerned during use.
The same applies to the use of accessories.
The PMX system must only be installed by qualified
personnel, strictly in accordance with the specifications
and with the safety requirements and regulations listed
below.
Maintenance and repair work on an open device with the
power on may only be carried out by trained personnel
who are aware of the dangers involved.
Important
The safety instructions are also included in paper format
with the product ("Documentation and Safety instructions
PMX" A3260-2.0).
PMXA4354-2.2HBM: public19
Symbols on the device
2Symbols on the device
SymbolMeaning
Statutory waste disposal mark
See chapter 28, „Waste disposal and environmental protec
tionpage 539
Statutory mark of compliance with emission limits in electronic
equipment supplied to China
See chapter
tion
“, page 539
CE mark
The CE mark is used by the manufacturer to declare that the
product complies with the requirements of the relevant EC
directives (the Declaration of Conformity can be found at
Take details in the operating manual into account.
CODESYS is a software platform for programmable logic
controllers. The license for CODESYS is already
implemented in WG001 basic housings.
28, „Waste disposal and environmental protec
20A4354-2.2HBM: publicPMX
3User information
Important
Obsolete documentation !
If you use an obsolete version of this document or an
obsolete version of any of the documentation it mentions,
this may result in the product being mounted and
operated incorrectly.
Make sure that all the documents you possess and
use are always the current version. The current
documentation version for your HBM products can be
found at http://www.hbm.com/en/menu/products/industrial-
amplifiers/pmx/
3.1Using this manual
► Read this operating manual thoroughly and in full
before operating the device for the first time.
User information
► Look upon this operating manual as part of the
product and keep it so that it is accessible to all users,
all of the time.
► If you pass the device on to a third party, always pass
it on together with the requisite documentation.
Should you lose this manual, the current version can be
found on our website, at http://www.hbm.com/en/menu/
products/industrial-amplifiers/pmx/
Failure to comply with this manual can result in personal
injury or damage to equipment.
We can assume no liability for any damage arising
from non-compliance with this operating manual.
PMXA4354-2.2HBM: public21
User information
To help you quickly find the information you require, there
is a full list of contents right at the front of this operating
manual.
There is also a comprehensive index at the end of the
manual, where you can look for individual keywords.
3.2About the PMX documentation
The documentation of the PMX measuring amplifier
system comprises
SThis operating manual in PDF format
SA printed quick start guide for initial start-up
SA printed summary of the safety instructions
SThe specifications (data sheet) in PDF format
SA PDF version of the web server Online help.
The complete description of the functionalities and
operation can be found in the Online help of the PMX
web server.
Important
These documents can be found
on the PMX system CD supplied with the device
always up-to-date, on our website, at http://www.hbm.
com/en/menu/products/industrial-amplifiers/pmx/
You can find additional information at http://www.hbm.
com/en/menu/support/software-firmware-downloads/indus
trial-amplifiers/ regarding e.g. device description files for
the realtime Ethernet cards (PROFINET / EtherCat) and
configuration examples.
22A4354-2.2HBM: publicPMX
You can find additional information at http://www.hbm.
com/en/menu/products/industrial-amplifiers/pmx/ ,as well
as a PMX video tutorial.
3.3Symbols used in this manual
So that you can start working quickly and safely with your
product, the symbols and terms used in this manual are
standardized and are explained below (see Chapter 28Waste disposal and environmental protection).
SymbolSignificance
-
List
List (second level)
Cross-reference to another point in this
document or to other documents
This prompts you to take action (a single,
independent action)
User information
1.
2.
3.
Important
Tip
PMXA4354-2.2HBM: public23
Carry out this sequence of actions in the
given order.
Important information
This draws your attention to important information
about the product or about handling the product.
Information/Application instructions
Practical tips or other useful information for you.
PMX product description
4PMX product description
By buying the PMX measuring amplifier system, you
have chosen a high-quality HBM measurement system
that is compact, powerful and variable. The data rate for
all the measurement and calculation channels is 19,200 /
38,400 measurements per second. The device thus
achieves an overall processing rate of approx. 400,000
measured values per second.
Numerous different measurement, control and
automation tasks can be resolved with this measurement
system.
Connection to a PC (HOST)
The PMX measuring amplifier system is connected to a
PC via the standard ETHERNET interface, and can be
parameterized and operated via the internal web server.
The link to an automation system can be connected via
digital and analog inputs/outputs, as well as over the
fieldbus interfaces of the PMX, to a control (PLC) or a
primary automation system.
Internal calculation channels
The PMX has 32 internal calculation channels as
standard, which are freely available for evaluations and
mathematical measurement signal calculations.
Automation tasks from peak values to PID controllers can
be implemented simply and elegantly.
24A4354-2.2HBM: publicPMX
PMX product description
The following types of plug-in card are available:
PX401
SThe PX401 measurement card offers four individually
configurable current or voltage inputs with TEDS
sensor detection.
SExtreme accuracy is guaranteed, as all the channels
have their own A/D converter with 24-bit resolution.
This also allows the acquisition of all the channels to
be totally synchronized.
PX455
SThe PX455 measurement card is available for
measurement with strain gages (SG) and also has
four channels with 24-bit resolution and TEDS sensor
detection.
SThe measurement card is suitable for strain gages in
both half bridge and full bridge circuits, as well as for
inductive transducers in half bridge and full bridge
circuits, LVDTs, potentiometric sensors and PT100
resistance thermometers.
PX460
SThe frequency measurement card PX460 can be used
to operate torque flanges (torque, speed, angle of
rotation), angle/incremental encoders, SSI/PWM
sensors and frequency measurements up to 2 MHz.
Channels 1 and 3: Frequency measurement (fixed)
Channels 2 and 4: Frequency (digital, inductive),
counter, encoder, SSI, PWM (adjustable)
PMXA4354-2.2HBM: public25
PMX product description
The following measurement modes are available:
- Up to four torque flanges (T10, T12, T40) for torque
or speed measurement (without direction of rotation
detection)
- Or two measuring channels for simultaneous
measurement of speed and angle (with direction of
rotation detection)
- Or one measuring channel for simultaneous speed,
angle and direction of rotation or reference pulse
detection
- Or two angle/incremental encoders each, SSI/PWM
sensors, magnetic transducers or pulse counters
- Or four measuring channels for frequency
measurement up to 2 MHz, incl. two shunt
calibrations and two 1-Wire TEDS (sensor
detection)
PX878
SThe PX878 input/output card has a total of eight digital
inputs, eight digital outputs and five analog voltage
outputs. The PMX can be controlled by this or be
operated with a downstream controller (PLC). All real
or calculated measurement signals can be freely
assigned to the outputs.
PX01EC, PX01PN and PX01EP
SThese interface cards can be optionally included in
the PMX, enabling operation of the PMX in an
automation system via the interface formats
EtherCat
, PROFINET or EtherNet/IP. Only one
variant can be used in each case.
26A4354-2.2HBM: publicPMX
PMX product description
Connection technique
Transducers are connected to the amplifiers via plug
terminals.
Plug terminals using push-in technology are available as
standard, with a screw-on system also available as an
option. If required, both types can be coded with the
enclosed coding pins to prevent mix-ups.
TEDS (Plug&Measure)
PMX amplifiers support TEDS (Transducer Electronic
Data Sheet, IEEE1451.4). When transducers are
connected, they are automatically detected. This
efficiently minimizes setup times and user error.
PMX web server
An easy to operate web server that is specifically tailored
to the PMX and is suitable for the measurement cards, is
integrated into the device for configuration, data
acquisition and visualization.
This gets you quickly to your measurement result, where
you can visualize the measured data and view them at a
later date.
PC software catman®EASY/AP
The HBM catman® software can be used to acquire,
condition and analyze the PMX measurement data, as an
option. This allows vast quantities of measurement data
to be displayed (stripchart function), analyzed and
exported in the most commonly used formats.
PMXA4354-2.2HBM: public27
PMX product description
Software driver
The PMX command set and a dotNET-API are available
to create customer-specific interfaces. This allows your
own operating concepts and the inclusion in existing
software solutions to be realized.
With the PMX-LabView driver, PMX can be integrated via
Virtual Instruments (VI) in the software of National
Instruments.
LabVIEW is a registered trademark of National
Instruments.
Device implementation
The multi-client capability of the PMX makes it possible to
access the device simultaneously and without loss of
speed via all the interfaces - including web server, field
bus and analog outputs.
Calibration certificates
Documented quality: at the time of delivery, PDF docu
ments of the HBM calibration certificates for the fitted
measurement cards, and a certificate of compliance as
per EN 10204 2.1 are already stored in the PMX device
memory. The PMX browser can be used to download
them.
28A4354-2.2HBM: publicPMX
Model overview, scope of supply, accessories
5Model overview, scope of supply, accessories
5.1The PMX system
The PMX is a modular and universally applicable
measuring amplifier system.
6
7
5
8
4
3
2
1
12
NrBezeichnung
1Communication card: EtherCAT or PROFINET; EtherNet/IP
2Fieldbus status LEDs
3Measurement card status LED
4RJ45‐Ethernet socket to PC/Network
5USB Host
6CANbus (WGX001 only)
72x RJ45 socket for synchronization of up to 20 modules
810 … 30V voltage supply
9System status LED
10Ground
11Positioning for the support rail
12Max. 4 measurement card or input/output cards, e.g. : PX455, PX460, PX878, PX401
9
10
11
PMXA4354-2.2HBM: public29
Model overview, scope of supply, accessories
The PMX comprises
- Basic device
- Measurement cards
- I/O cards and
- Communication cards.
The measurement cards, input/output cards and commu
nication cards can be individually combined and intelli
gently configured, in accordance with the measurement
task.
The PMX system is a modular and universally applicable
measuring amplifier system. The measurement cards,
input/output cards and communication cards can be
individually combined and intelligently configured in
accordance with the measurement task.
Basic device
ConnectionsDescription
ETHERNETConnection to an Ethernet network or a PC, 100 MBit/s; half
and full duplex
USBDevice backup, data storage and special device functions
CANLocal connection to CANBus nodes (WGX001 only)
SYNCSynchronization of up to 20 PMX devices
POWERVoltage supply (10 … 30 V DC)