Hottinger Baldwin Messtechnik GmbH
Im Tiefen See 45
D-64239 Darmstadt
Tel. +49 6151 803-0
Fax +49 6151 803-9100
info@hbm.com
www.hbm.com
Mat.: 7-2002.4354
DVS: A4354-2.2 HBM: public
11.2016
E Hottinger Baldwin Messtechnik GmbH.
Subject to modifications.
All product descriptions are for general information only.
They are not to be understood as a guarantee of quality or
durability.
The PMX amplifier system is to be used exclusively for
measurement tasks and directly related control tasks.
Use for any purpose other than the above is deemed to
be non-designated use. In the interests of safety, the
device should only be operated as described in the
Operating Manuals. It is also essential to comply with the
legal and safety requirements for the application
concerned during use. The same applies to the use of
accessories.
Each time, before starting up the equipment, you must
first run a project planning and risk analysis that takes
into account all the safety aspects of automation
technology. This particularly concerns personal and
machine protection.
Additional safety precautions must be taken in plants
where malfunctions could cause major damage, loss of
data or even personal injury. In the event of a fault, these
precautions establish safe operating conditions.
Safety instructions
This can be done, for example, by mechanical
interlocking, error signaling, limit value switches, etc.
Notice
The device must not be directly connected to the DC
voltage supply system. Supply voltage 10 V … 30 V
(DC).
Before commissioning, ensure that a suitable supply
voltage is used and that the circuit used is sufficiently
protected.
PMXA4354-2.2HBM: public13
Page 14
Safety instructions
The device may only be supplied with separated extralow voltage (safety isolating transformer according to DIN
VDE 0551 / EN60742). Only operate built-in devices once
they are installed in the housing provided. The device
development is based on DIN EN 61010 Part 1 (VDE
0411 Part 1).
General dangers of failing to follow the safety
instructions
The PMX system is a state of the art unit and as such is
reliable. The module may give rise to dangers if it is
inappropriately installed and operated by untrained
personnel.
Any person instructed to carry out installation, starting up,
maintenance or repair of the module must have read and
understood the Operating Manual and in particular the
technical safety instructions.
Conditions at the place of installation
SProtect the device from direct contact with water.
SProtect the PMX system from moisture and humidity
or weather conditions such as rain, snow, etc.
SDo not expose the device to direct sunlight.
SPlease observe the permissible maximum ambient
temperatures stated in the specifications.
SThe permissible relative humidity at 31 °C is 95 %
(non condensing); linear reduction to 50 % at 40 °C.
SInstall the device so that it can be disconnected from
the supply voltage at any time without difficulty.
SIt is safe to operate the PMX system up to a height of
2000 m.
14A4354-2.2HBM: publicPMX
Page 15
Safety instructions
Maintenance and cleaning
The PMX system is maintenance-free.
SBefore cleaning, disconnect all connections.
SClean the housing with a soft, slightly damp (not wet!)
cloth. You should never use solvents, since these
could damage the labeling.
SWhen cleaning, ensure that no liquid gets into the
module or connections.
Residual dangers
The scope of supply and performance of the PMX system
only covers a small area of measurement technology. In
addition, equipment planners, installers and operators
should plan, implement and respond to the safety
engineering considerations of measurement technology
in such a way as to minimize remaining dangers. On-site
regulations must be complied with at all times. There
must be reference to the remaining dangers connected
with measurement technology.
Product liability
In the following cases, the protection provided for the
device may be adversely affected. Liability for device
functionality then passes to the operator:
SThe device is not used in accordance with the
operating manual.
SThe device is used outside the field of application
described in this Chapter.
SThe operator makes unauthorized changes to the
device.
PMXA4354-2.2HBM: public15
Page 16
Safety instructions
Warning signs and danger symbols
Important instructions for your safety are specifically
identified. It is essential to follow these instructions in
order to prevent accidents and damage to property.
Safety instructions are structured as follows:
WARNING
Type of danger
Consequences of non-compliance
Averting the danger
SWarning sign:
draws your attention to the danger
SSignal word:
indicates the severity of the danger (see table below)
SType of danger:
mentions the type or source of the danger
SConsequences:
describes the consequences of non-compliance
SDefense:
indicates how the danger can be avoided/bypassed
16A4354-2.2HBM: publicPMX
Page 17
Danger class according to ANSI
SymbolSignificance
WARNING
CAUTION
Notice
This marking warns of a potentially dangerous
situation in which failure to comply with safety
requirements can result in death or serious physical
injury.
This marking warns of a potentially dangerous
situation in which failure to comply with safety
requirements can result in slight or moderate physical
injury.
This marking draws your attention to a situation in
which failure to comply with safety requirements can
lead to damage to property.
Working safely
Notice
The device must not be directly connected to the DC
voltage supply system. Supply voltage 10 V … 30 V
(DC).
Safety instructions
The supply connection, as well as the signal and sense
leads, must be installed in such a way that
electromagnetic interference does not adversely affect
module functionality (HBM recommendation: "Greenline
shielding design", downloadable from the Internet at
http://www.hbm.com/Greenline
).
Automation equipment and devices must be installed in
such a way that adequate protection or locking against
unintentional actuation is provided (e.g. access checks,
password protection, etc.).
When devices are working in a network, these networks
must be designed in such a way that malfunctions in
individual nodes can be detected and shut down.
PMXA4354-2.2HBM: public17
Page 18
Safety instructions
Safety precautions must be taken both in terms of
hardware and software, so that a line break or other
interruptions to signal transmission, such as via the bus
interfaces, do not cause undefined states or loss of data
in the automation device.
Conversions and modifications
The device must not be modified from the design or
safety engineering point of view except with our express
agreement. Any modification shall exclude all liability on
our part for any damage resulting therefrom.
In particular, any repair or soldering work on
motherboards is prohibited. When exchanging complete
modules, use only original parts from HBM. The product
is delivered from the factory with a fixed hardware and
software configuration. Changes can only be made within
the possibilities documented in the operating manual.
Qualified personnel
Qualified personnel means persons entrusted with siting,
mounting, starting up and operating the product, who
possess the appropriate qualifications for their function
(qualified electrician, or by someone with electrical
training under the supervision of a qualified electrician).
This device is only to be installed and used by qualified
personnel strictly in accordance with the specifications
and with the safety rules and regulations which follow.
This includes people who meet at least one of the three
following requirements:
SKnowledge of the safety concepts of automation
technology is a requirement and, as project personnel,
you must be familiar with these concepts.
18A4354-2.2HBM: publicPMX
Page 19
Safety instructions
SAs automation plant operating personnel, you have
been instructed how to handle the machinery. You are
familiar with the operation of the equipment and
technologies described in this documentation.
SAs commissioning engineers or service engineers,
you have successfully completed the training to qualify
you to repair the automation systems. You are also
authorized to activate, ground and label circuits and
equipment in accordance with safety engineering
standards.
It is also essential to comply with the legal and safety
requirements for the application concerned during use.
The same applies to the use of accessories.
The PMX system must only be installed by qualified
personnel, strictly in accordance with the specifications
and with the safety requirements and regulations listed
below.
Maintenance and repair work on an open device with the
power on may only be carried out by trained personnel
who are aware of the dangers involved.
Important
The safety instructions are also included in paper format
with the product ("Documentation and Safety instructions
PMX" A3260-2.0).
PMXA4354-2.2HBM: public19
Page 20
Symbols on the device
2Symbols on the device
SymbolMeaning
Statutory waste disposal mark
See chapter 28, „Waste disposal and environmental protec
tionpage 539
Statutory mark of compliance with emission limits in electronic
equipment supplied to China
See chapter
tion
“, page 539
CE mark
The CE mark is used by the manufacturer to declare that the
product complies with the requirements of the relevant EC
directives (the Declaration of Conformity can be found at
Take details in the operating manual into account.
CODESYS is a software platform for programmable logic
controllers. The license for CODESYS is already
implemented in WG001 basic housings.
28, „Waste disposal and environmental protec
20A4354-2.2HBM: publicPMX
Page 21
3User information
Important
Obsolete documentation !
If you use an obsolete version of this document or an
obsolete version of any of the documentation it mentions,
this may result in the product being mounted and
operated incorrectly.
Make sure that all the documents you possess and
use are always the current version. The current
documentation version for your HBM products can be
found at http://www.hbm.com/en/menu/products/industrial-
amplifiers/pmx/
3.1Using this manual
► Read this operating manual thoroughly and in full
before operating the device for the first time.
User information
► Look upon this operating manual as part of the
product and keep it so that it is accessible to all users,
all of the time.
► If you pass the device on to a third party, always pass
it on together with the requisite documentation.
Should you lose this manual, the current version can be
found on our website, at http://www.hbm.com/en/menu/
products/industrial-amplifiers/pmx/
Failure to comply with this manual can result in personal
injury or damage to equipment.
We can assume no liability for any damage arising
from non-compliance with this operating manual.
PMXA4354-2.2HBM: public21
Page 22
User information
To help you quickly find the information you require, there
is a full list of contents right at the front of this operating
manual.
There is also a comprehensive index at the end of the
manual, where you can look for individual keywords.
3.2About the PMX documentation
The documentation of the PMX measuring amplifier
system comprises
SThis operating manual in PDF format
SA printed quick start guide for initial start-up
SA printed summary of the safety instructions
SThe specifications (data sheet) in PDF format
SA PDF version of the web server Online help.
The complete description of the functionalities and
operation can be found in the Online help of the PMX
web server.
Important
These documents can be found
on the PMX system CD supplied with the device
always up-to-date, on our website, at http://www.hbm.
com/en/menu/products/industrial-amplifiers/pmx/
You can find additional information at http://www.hbm.
com/en/menu/support/software-firmware-downloads/indus
trial-amplifiers/ regarding e.g. device description files for
the realtime Ethernet cards (PROFINET / EtherCat) and
configuration examples.
22A4354-2.2HBM: publicPMX
Page 23
You can find additional information at http://www.hbm.
com/en/menu/products/industrial-amplifiers/pmx/ ,as well
as a PMX video tutorial.
3.3Symbols used in this manual
So that you can start working quickly and safely with your
product, the symbols and terms used in this manual are
standardized and are explained below (see Chapter 28Waste disposal and environmental protection).
SymbolSignificance
-
List
List (second level)
Cross-reference to another point in this
document or to other documents
This prompts you to take action (a single,
independent action)
User information
1.
2.
3.
Important
Tip
PMXA4354-2.2HBM: public23
Carry out this sequence of actions in the
given order.
Important information
This draws your attention to important information
about the product or about handling the product.
Information/Application instructions
Practical tips or other useful information for you.
Page 24
PMX product description
4PMX product description
By buying the PMX measuring amplifier system, you
have chosen a high-quality HBM measurement system
that is compact, powerful and variable. The data rate for
all the measurement and calculation channels is 19,200 /
38,400 measurements per second. The device thus
achieves an overall processing rate of approx. 400,000
measured values per second.
Numerous different measurement, control and
automation tasks can be resolved with this measurement
system.
Connection to a PC (HOST)
The PMX measuring amplifier system is connected to a
PC via the standard ETHERNET interface, and can be
parameterized and operated via the internal web server.
The link to an automation system can be connected via
digital and analog inputs/outputs, as well as over the
fieldbus interfaces of the PMX, to a control (PLC) or a
primary automation system.
Internal calculation channels
The PMX has 32 internal calculation channels as
standard, which are freely available for evaluations and
mathematical measurement signal calculations.
Automation tasks from peak values to PID controllers can
be implemented simply and elegantly.
24A4354-2.2HBM: publicPMX
Page 25
PMX product description
The following types of plug-in card are available:
PX401
SThe PX401 measurement card offers four individually
configurable current or voltage inputs with TEDS
sensor detection.
SExtreme accuracy is guaranteed, as all the channels
have their own A/D converter with 24-bit resolution.
This also allows the acquisition of all the channels to
be totally synchronized.
PX455
SThe PX455 measurement card is available for
measurement with strain gages (SG) and also has
four channels with 24-bit resolution and TEDS sensor
detection.
SThe measurement card is suitable for strain gages in
both half bridge and full bridge circuits, as well as for
inductive transducers in half bridge and full bridge
circuits, LVDTs, potentiometric sensors and PT100
resistance thermometers.
PX460
SThe frequency measurement card PX460 can be used
to operate torque flanges (torque, speed, angle of
rotation), angle/incremental encoders, SSI/PWM
sensors and frequency measurements up to 2 MHz.
Channels 1 and 3: Frequency measurement (fixed)
Channels 2 and 4: Frequency (digital, inductive),
counter, encoder, SSI, PWM (adjustable)
PMXA4354-2.2HBM: public25
Page 26
PMX product description
The following measurement modes are available:
- Up to four torque flanges (T10, T12, T40) for torque
or speed measurement (without direction of rotation
detection)
- Or two measuring channels for simultaneous
measurement of speed and angle (with direction of
rotation detection)
- Or one measuring channel for simultaneous speed,
angle and direction of rotation or reference pulse
detection
- Or two angle/incremental encoders each, SSI/PWM
sensors, magnetic transducers or pulse counters
- Or four measuring channels for frequency
measurement up to 2 MHz, incl. two shunt
calibrations and two 1-Wire TEDS (sensor
detection)
PX878
SThe PX878 input/output card has a total of eight digital
inputs, eight digital outputs and five analog voltage
outputs. The PMX can be controlled by this or be
operated with a downstream controller (PLC). All real
or calculated measurement signals can be freely
assigned to the outputs.
PX01EC, PX01PN and PX01EP
SThese interface cards can be optionally included in
the PMX, enabling operation of the PMX in an
automation system via the interface formats
EtherCat
, PROFINET or EtherNet/IP. Only one
variant can be used in each case.
26A4354-2.2HBM: publicPMX
Page 27
PMX product description
Connection technique
Transducers are connected to the amplifiers via plug
terminals.
Plug terminals using push-in technology are available as
standard, with a screw-on system also available as an
option. If required, both types can be coded with the
enclosed coding pins to prevent mix-ups.
TEDS (Plug&Measure)
PMX amplifiers support TEDS (Transducer Electronic
Data Sheet, IEEE1451.4). When transducers are
connected, they are automatically detected. This
efficiently minimizes setup times and user error.
PMX web server
An easy to operate web server that is specifically tailored
to the PMX and is suitable for the measurement cards, is
integrated into the device for configuration, data
acquisition and visualization.
This gets you quickly to your measurement result, where
you can visualize the measured data and view them at a
later date.
PC software catman®EASY/AP
The HBM catman® software can be used to acquire,
condition and analyze the PMX measurement data, as an
option. This allows vast quantities of measurement data
to be displayed (stripchart function), analyzed and
exported in the most commonly used formats.
PMXA4354-2.2HBM: public27
Page 28
PMX product description
Software driver
The PMX command set and a dotNET-API are available
to create customer-specific interfaces. This allows your
own operating concepts and the inclusion in existing
software solutions to be realized.
With the PMX-LabView driver, PMX can be integrated via
Virtual Instruments (VI) in the software of National
Instruments.
LabVIEW is a registered trademark of National
Instruments.
Device implementation
The multi-client capability of the PMX makes it possible to
access the device simultaneously and without loss of
speed via all the interfaces - including web server, field
bus and analog outputs.
Calibration certificates
Documented quality: at the time of delivery, PDF docu
ments of the HBM calibration certificates for the fitted
measurement cards, and a certificate of compliance as
per EN 10204 2.1 are already stored in the PMX device
memory. The PMX browser can be used to download
them.
28A4354-2.2HBM: publicPMX
Page 29
Model overview, scope of supply, accessories
5Model overview, scope of supply, accessories
5.1The PMX system
The PMX is a modular and universally applicable
measuring amplifier system.
6
7
5
8
4
3
2
1
12
NrBezeichnung
1Communication card: EtherCAT or PROFINET; EtherNet/IP
2Fieldbus status LEDs
3Measurement card status LED
4RJ45‐Ethernet socket to PC/Network
5USB Host
6CANbus (WGX001 only)
72x RJ45 socket for synchronization of up to 20 modules
810 … 30V voltage supply
9System status LED
10Ground
11Positioning for the support rail
12Max. 4 measurement card or input/output cards, e.g. : PX455, PX460, PX878, PX401
9
10
11
PMXA4354-2.2HBM: public29
Page 30
Model overview, scope of supply, accessories
The PMX comprises
- Basic device
- Measurement cards
- I/O cards and
- Communication cards.
The measurement cards, input/output cards and commu
nication cards can be individually combined and intelli
gently configured, in accordance with the measurement
task.
The PMX system is a modular and universally applicable
measuring amplifier system. The measurement cards,
input/output cards and communication cards can be
individually combined and intelligently configured in
accordance with the measurement task.
Basic device
ConnectionsDescription
ETHERNETConnection to an Ethernet network or a PC, 100 MBit/s; half
and full duplex
USBDevice backup, data storage and special device functions
CANLocal connection to CANBus nodes (WGX001 only)
SYNCSynchronization of up to 20 PMX devices
POWERVoltage supply (10 … 30 V DC)
30A4354-2.2HBM: publicPMX
Page 31
Model overview, scope of supply, accessories
Measurement cards
Measurement cardDescriptionTransducers that can be
connected
PX401
Current/voltage
amplifier
4 Current/voltage sources, always
individually user-selectable
between current and voltage input,
TEDS (1-wire)
PX455
Strain gage amplifier
4 SG full or half bridges (CF). The
bridge excitation voltage is 2.5 V
Inductive full or half-bridges
LVDT, potentiometric sensors,
piezoresistive sensors, TEDS
(0-wire)
PX460Frequency/counter
measuring amplifier
Up to four torque flanges (T10,
T12, T40) for torque or speed
measurement (without direction of
rotation detection)
Or two measuring channels for
simultaneous measurement of
speed and angle (with direction of
rotation detection)
Or one measuring channel for
simultaneous speed, angle and
direction of rotation or reference
pulse detection
Or two angle/incremental encoders
each, SSI/PWM sensors, magnetic
transducers or pulse counters
Or four measuring channels for
frequency measurement up to 2
MHz, incl. two shunt calibrations
and two 1-Wire TEDS (sensor
detection)
PMXA4354-2.2HBM: public31
Page 32
Model overview, scope of supply, accessories
Input/output cards (I/O)
Basic device, modelInterfacesTransducers that can be
connected
PX878I/O card8 digital inputs, 8 digital outputs,
5 analog voltage outputs, all
individually configurable
EtherCAT® is a registered brand and patented technology, licensed by Beckhoff Automation
GmbH, Germany
32A4354-2.2HBM: publicPMX
Page 33
Model overview, scope of supply, accessories
Overview of measurement cards and input/output card
PMXA4354-2.2HBM: public33
Page 34
Model overview, scope of supply, accessories
5.2Scope of delivery
Order No.
1 PMX basic device, with a wall mounting kit (1 wall bracket, 4
screws, 4 washers) and a set for DIN rail mounting
with CAN connection and CODESYS‐V3 Soft‐PLC
without CAN connection, without CODESYS
For each measurement card : A mating connector for each
channel, all mating connectors with Push-In technology
(4 connectors are enclosed for each measurement card,
complete with coding pins)
DIN rail mounting
(2 units, packed in film cushion packaging with the mounting
material in an Etimex bag)
(4 M5x10 close tolerance screws, 4 spring washers)
PMX system CD with operating manual and data sheet, safety
instructions and quick start guide
For WGX001: Delivery with CODESYS‐CD (CODESYS‐V3
software, PMX package quick start guide and program
examples)
Mating connector M12x1 for CAN interface for WGX0011-CON-S1002
Mating connector for PMX voltage supply (WGX001 / WGX002)1-CON-S1010
1-WGX001
1-WGX002
1-CON-S1008
1-CON-S1012 for
PX460
1-RAILCLIP
5.3Accessories
AccessoriesOrder No.
Ethernet crossover cable, for direction operation of devices on a
PC or notebook, length 2 m, type CAT5+
AC/DC power supply unit;
Input: 90 V … 264 VAC, 1.5 m cable, output: 24 V DC, max.
1.25 A, 2 m cable with ODU plug
34A4354-2.2HBM: publicPMX
1-KAB239-2
1-NTX001
Page 35
Model overview, scope of supply, accessories
Spare partsOrder No.
PX01, PMX blue blanking plate for plug‐in card slot 01-PX01
PX02, PMX gray blanking plate for plug‐in card slot 1-41-PX02
RAILCLIP, PMX DIN rail mounting set (2 pieces), incl. screws1-RAILCLIP
Phoenix plug terminals
Set of plug terminals (push‐in) for PMX plug‐in cards (4 x
7‐pin, incl. coding plug and labeling sheets)
Set of screw terminals for PMX plug‐in cards
(4 x 7‐pin, incl. coding plug and labeling sheets)
Set of screw terminal for PMX voltage supply
(1 x 2‐pin, incl. coding plug and labeling sheets)
Set of plug terminals (push‐in) for PMX plug‐in cards (2 x
13‐pin and 2 x 2‐pin, incl. coding plug and labeling sheets)
Mating connector M12x1 for CAN interface for WGX001
1-CON-S1008
1-CON-S1009
1-CON-S1010
1-CON-S1012
1-CON-S1002
In general, the mating connectors are always included for
all plug-in cards (PX401, PX455, PX460 and PX878).
When ordering a PMX basic device, the delivery always
includes DIN rail mounting and wall mounting elements.
Important
You have the option to retrofit or subsequently remove
measurement cards, I/O cards and communication cards.
PMXA4354-2.2HBM: public35
Page 36
Model overview, scope of supply, accessories
5.4PMX web server/software
A PMX web server, including Help, is integrated in the
device. The web server also has a firmware loader
function, which can transfer new PMX firmware and web
server versions to the PMX.
An online Help is integrated into the web server to supp
ort PMX operation and handling (click on the Help icon,
top right in the overview menu).
36A4354-2.2HBM: publicPMX
Page 37
Model overview, scope of supply, accessories
PC software catman®EASY/AP
The HBM catman® software can be used to acquire,
condition and analyze the PMX measurement data, as an
option. This allows vast quantities of measurement data
to be displayed (stripchart function) and analyzed (see
following ).
All real and calculated measurement channels, as well as
digital inputs/outputs can be measured. Digital inputs/
outputs are displayed as binary coded values.
PMXA4354-2.2HBM: public37
Page 38
Model overview, scope of supply, accessories
PMX supports up to three sampling rates that can be
independently set. These sampling rates can then be
assigned individual measurement signals.
Times of day, digital inputs/outputs of the PMX or
triggering via limit values can be used to start/stop
(trigger) a measurement.
The PMX can also be partially parameterized with
catman®. This includes:
SSetting the sensor type via the sensor database or
TEDS
SWriting the TEDS sensors via the integrated TEDS
EDITOR in catman®
SZero setting the measurement signal and setting the
filter frequency for each individual channel
You can use the catman
®
Script programming language
to program complete measurement sequences including
automated storage of measured data and creation of
logs.
For more information, see the online Help in
catman®EASY/AP
Important
The PMX device settings are permanently saved in the
active PMX parameter set by catman®. The sensor set
tings (sensor type, scaling and filter) are automatically
changed by catman
® in PMX. This cannot be locked
either in catman ® or in PMX.
®
Before starting catman
filter setting in the catman
activate retaining of the PMX
®
option dialog (under DAQ
channels: "Allow manual filter settings").
For firmware version 2.00 and later, catman
®
version 4.0
or higher is mandatory.
38A4354-2.2HBM: publicPMX
Page 39
Model overview, scope of supply, accessories
Software driver
The PMX command set and a dotNET-API and a
LabView driver are available to create customer-specific
interfaces. This allows your own operating concepts and
the inclusion in existing software solutions to be realized.
With the PMX-LabView driver, PMX can be integrated via
Virtual Instruments (VI) in the software of National
Instruments.
The following functions are supported by the drivers in
PMX:
FunctionDescription
Decice ScanScan Ethernet network
Measurement configurationSet sample rate, filter, zero
Sensor configurationSet scaling (2‐point) or via TEDS
Analog In DAQ and calculated channels
(streaming)
Status information (diagnostic)Read each channel‐ and device status
Peak valuesRead or clear peak values
Limit switchesRead or set limit switches
Analog Out (direct setting)Read or set analog outputs (10 V)
Analog Out (configuration)Set source, scaling
Digital In DAQRead digital outputs and set (high/low)
Digital Out DAQ (direct setting)Read digital outputs and set (high/low)
CAN DAQ (via CODESYS / calculated
channels)
Parameter setsRead and select parameter sets
Read all measuring values and time
stamps from sensors, channels
Read calculated channels with CAN
signals
PMXA4354-2.2HBM: public39
Page 40
Model overview, scope of supply, accessories
Notice
For firmware version 2.00 and later, catman®version 4.0
or higher is mandatory.
Tip
All commands of the PMX command set can be used as
lowlevel commands (see chapter 21) .
You can find extensive support and programming ex
amples in the program help for the individual drivers:
All drivers and the catman
downloaded from the HBM website as a free 30day ver
http://www.hbm.com/en/menu/support/software-
sion:
firmware-downloads/daq-software/catman/
®
software as well can be
40A4354-2.2HBM: publicPMX
Page 41
Degree of protection / housing / shielding design
6Degree of protection / housing / shielding
design
The degree of protection given in the specifications
indicates the suitability of the device for various ambient
conditions and also the protection it gives people against
potential risks when they are using it. The letters IP
(International Protection), which are always present in the
designation, are followed by two digits. These indicate
which degree of protection a housing offers against
contact or foreign bodies (first digit) and moisture (second
digit).
All PMX modules and the basic device are designed with
degree of protection IP20 (as per EN 60529).
IP20
Code indexDegree of
protection
against contact
and foreign
bodies
2Protection against
contact with fingers,
protection against
foreign bodies with
> 12 mm
Code indexDegree of
protection
against water
0No water protection
PMXA4354-2.2HBM: public41
Page 42
Degree of protection / housing / shielding design
New Greenline shielding design
To improve electromagnetic interference protection, HBM
has developed an effective measure, the Greenline
shielding design. The shield is connected to the
connector housing. Appropriate routing of the cable
shield means that the entire measuring chain is
completely enclosed by a Faraday cage.
wh
bk
rd
bu
gn
gy
ye
Fig. 6.1Greenline shielding design
Important
Use standard HBM cables for connecting the
transducers. When using other shielded, low-capacitance
measurement cables, attach the shield of the transducer
cable to the ground connection provided on the multipoint
connector, in accordance with HBM Greenline
information http://www.hbm.com/Greenline
.
This results in greater EMC protection.
42A4354-2.2HBM: publicPMX
Page 43
Mounting/Dismounting/Replacing
7Mounting/Dismounting/Replacing
7.1Mounting tools and tightening torques
MountingRequired toolTightening torque
Fasten Rail-Clip to support rail
M 2.5 hexagon socket screw
Mounting DIN rail on housing
M5 hexagon socket screw
Mounting front panel
M2.5 Torx screws
Mounting wall bracket
M4 hexagon socket screw
Mounting side parts
M3 Torx screws
Grounding screw on the PMX
M4 Torx screws
Hexagon socket
screwdriver
2.5 a.f.
Hexagon socket
screwdriver
3 a.f.
Torx screwdriver
TX8
Hexagon socket
screwdriver
3 a.f.
Torx screwdriver
TX10
Torx screwdriver
TX20
1.0 - 1.2 Nm
3 Nm
0.5 - 0.6 Nm
1.5 … 2 Nm
0,8 … 1 Nm
1,5 … 2 Nm
PMXA4354-2.2HBM: public43
Page 44
Mounting/Dismounting/Replacing
7.2Support rail mounting
75 mm
2
Allen screw
3 mm a.f.
1
3
A
B
3
4
D
Fig. 7.1Mounting on a support rail
► Undo the four back-panel screws (Torx T10) (1)
► Slide the side parts forward
► Screw on the support rail mounting (Rail-Clip) (3)
(approx. 5 Nm) in a choice of four positions (A to D)
(two positions for the 7.5 mm support rail)
► Screw the side panels (2) back on
► Hook the PMX device into the support rail (4)
(2)
C
44A4354-2.2HBM: publicPMX
Page 45
Mounting/Dismounting/Replacing
Important
Device damage caused by the PMX falling due to resis
tance in hooking/unhooking the PMX.
HBM recommends using a DIN support rail (DIN EN
60715), 15 mm in height. When using a smaller support
rail (7.5 mm high), it should be shimmed to make it easy
to hook/unhook the PMX device.
The 7.5 mm support rail can only be used in the top two
positions (A and B).
Attach the support rail mounting (Rail-Clip) to the
DIN rail
Allen screw
2.5 a.f.
At the time of delivery, the self-locking (2.5 mm a.f.) Allen
screws are unscrewed as far as the stop.
► Clip on the support rail mounting (Rail-Clip)
PMXA4354-2.2HBM: public45
Page 46
Mounting/Dismounting/Replacing
► Hand-tighten the self-locking Allen screw (1 - 1.2 Nm)
Important
Device damage caused by electromagnetic irradiation of
external devices.
Faulty measurements due to electromagnetic irradiation
from other devices.
To ensure adequate grounding for the PMX, the support
rail must be connected to a functional earth
Both the DIN rail and the PMX must be free of paint,
varnish and dirt at the point of installation.
► Connect the PMX housing to ground via the grounding
screw.
.
46A4354-2.2HBM: publicPMX
Page 47
Mounting/Dismounting/Replacing
Dimensions and mounting instructions
200
141
*) Height of support rail 15 mm
**) Height of support rail 7.5 mm
***) Min. dimension: Plug plus sensor cable
NOTE:
To ensure sufficient ventilation/cooling, a 2 cm gap must
be maintained above and below neighboring devices.
200
141*)
variable
133.5**)
122
min. 25***)
PMXA4354-2.2HBM: public47
Page 48
Mounting/Dismounting/Replacing
7.3Wall-mounting
4. 3
1
225
168.5
215
Fig. 7.2Mounting on a wall
152
► Attach the wall bracket at the back of the PMX with
the enclosed M4 screws (1) (3 Nm)
48A4354-2.2HBM: publicPMX
Page 49
Mounting/Dismounting/Replacing
Screw
M4
Screw
M4
Screw
M4
Grounding
screw
(Torx TX20)
Screw
M4
► Screw the complete unit to a wall; hole Ø 4 mm
Notice
Device damage caused by electromagnetic irradiation of
external devices.
Faulty measurements due to electromagnetic irradiation
from other devices.
The housing must also be connected to a functional earth
in a wall mounting.
► Connect the PMX housing to ground via the grounding
screw.
PMXA4354-2.2HBM: public49
Page 50
Mounting/Dismounting/Replacing
Dimensions and mounting instructions
215
152
138
200
min. 25*)
122
*) Min. dimension: Plug plus sensor cable
NOTE:
To ensure sufficient ventilation/cooling, a 2 cm
gap must be maintained above and below
neighboring devices.
7.4Replacing measurement and
communication cards
Measurement and communication cards can be
retrofitted or removed at a later date. Please note the
combination options (see chapter 8.2.1 on page 55).
After modification and switching on the supply voltage,
the PMX automatically detects and initializes the
hardware configuration. The factory settings are loaded.
All parameters must be reentered, also for the existing
cards.
50A4354-2.2HBM: publicPMX
Page 51
Mounting/Dismounting/Replacing
Notice
If the measurement or communication cards are
incorrectly removed/replaced, they may be damaged
or destroyed.
These cards must be de-energized before they are
removed or replaced
► Always disconnect the PMX from the power supply
before removing a card. Note that the device settings
have to be re-parameterized if new cards are added.
The instructions below must also be followed:
Removal
1
1. Undo the three M2.5x8 Torx (Tx8) screws (1) of the
card/blanking plate
2. Use a screwdriver to slightly lever up the card at the
lug provided.
3. Carefully take out the board
PMXA4354-2.2HBM: public51
Page 52
Mounting/Dismounting/Replacing
Installation
1. Carefully insert the board into the PMX slot (ribs
prevent tilting)
2. The board centers itself in the VG strip at the back
3. Re-tighten the three M2.5 screws
Notice
Device damage caused by electromagnetic irradiation of
external devices.
Faulty measurements due to electromagnetic irradiation
from other devices.
► Close the open module slots with blanking plates
(Accessories).
52A4354-2.2HBM: publicPMX
Page 53
8Electrical connections PMX
8.1Plug technology and clamping areas
All PMX plug-in cards (PX401, PX455, PX460, PX878)
are delivered as standard with easy to fit plug terminals
using Push-In technology. Plug terminals can also be
delivered with a screw-on system as an option.
Push-In technology
The clamping area is 0.2mm2 (AWG24) to 1.5mm2
(AWG16). If several conductors are to be connected to a
terminal, the line cross-section must be adapted
accordingly. End sleeves (without plastic collars, 10mm)
should be used on the strands to connect the wires to the
terminals.
Electrical connections PMX
The shield of the transducer cable must be connected to
the ground connection provided on the PMX multipoint
connector, in accordance with HBM Greenline informa
tion http://www.hbm.com/Greenline
Important
The grounding terminal on the PMX is not a protective
ground (optional connection).
The measurement system is fitted with an automatic
current limiter for each device card, as well as for the
PMX basic device.
PMXA4354-2.2HBM: public53
.
Page 54
Electrical connections PMX
8.2Overview of PMX functions
1
11
10
9
8
Ethernet plug for PC/network connection
1
2
USB host, e.g. for a memory stick
CAN for CAN driver, M12, option
3
(WGX001)
2 x RJ45 for synchronization
4
Excitation 10 . . . 30 V DC
5
6
Device status - LED
4
2
3
5
6
7
7
Measurement cards (PX401, PX455, PX460)
7
and/or PX878 I/O card or blanking plate
8
LED Measurement card status
Communication cards:
9
PX01EC (EtherCAT®, PX01PN
(PROFINET-IO, PC01EP EtherNet/IP) or
blanking plate
8.2.2Relevance of the basic device connection
sockets
PC or network connection.
Cable: Ethernet cable CAT5, SFTP
USB connection version 2.0 e.g. for mass storage device,
scanner, USB stick
Cable: standard USB cable
PMXA4354-2.2HBM: public55
Page 56
Electrical connections PMX
Synchronization of several (max. 20) PMX devices via
two RJ45 sockets.
See chapter 8.1
Supplying voltage to the PMX by connecting a separate
DC voltage supply.
-
+
(view of device front)
PMX basic device
3
4
5
1
2
Cable
Connector
Female connector
CAN connection (type WGX001 only)xx
PinSignalDescription
1SHLDCAN shielding
2not connected
3GNDGround
4CAN_HCAN_H data lead (high)
5CAN_LCAN_L data lead (low)
56A4354-2.2HBM: publicPMX
Page 57
Electrical connections PMX
8.2.3LEDs for system monitoring (device LED)
Basic device (WGX001/002)
12
34567
ETHERNET LED (1, 2)
LEDLEDStatusSignificance
Ethernet Link (1)
Ethernet RX / TX (2)
green
yellow
SteadyConnection is present
FlashingData are being transmitted
SYNC IN / OUT (3, 4 and 5, 6)
LEDLEDStatusSignificance
IN (3)
IN (4)
IN (3 + 4)OffMaster
OUT (5)
OUT (6)
green
yellow
green
yellow
OnSlave
OnError
Onalways on
Error (always identical to the
On
right-hand LED of the IN
socket)
PMXA4354-2.2HBM: public57
Page 58
Electrical connections PMX
SYS LED (7)
LEDStatusSignificance
green
On
Off
Voltage supply available
Voltage supply missing
yellow
red
OnDevice is booting
Flashing
On
8.2.4Fieldbus LED
PX01EC
Serious internal error or
Firmware update
2
1
58A4354-2.2HBM: publicPMX
Page 59
Electrical connections PMX
EtherCAT®
LEDLEDStatusSignificance
ERR
red
red
red
red
red
LEDLEDStatusSignificance
RUN
green
green
green
OffNo error
FlashingConfiguration error
Single flashSynchronization error
Double flashApplication timeout error
OnPDI timeout error
OffINIT status
FlashingPRE OPERATIONAL status
Single flashSAFE OPERATIONAL status
green
OnOPERATIONAL
PMXA4354-2.2HBM: public59
Page 60
Electrical connections PMX
LEDLEDStatusSignificance
1
2--No function
green
Permanently
on
Flashing
Off
Connection established
Send / Receive
No connection
60A4354-2.2HBM: publicPMX
Page 61
Electrical connections PMX
PX01EP
EtherNET/IP
LEDLEDStatusSignificance
Connected: If the device has at least
LV
green
On
Flashing
green
On
red
one existing connection (even to the
message router), the network status
display will light up and remain green.
No connections: If the device has no
existing connections, but has received
an IP address, the network status
display flashes green.
Double IP*: If the device detects that
the IP address has already been used,
the network status display lights up and
remains red.
2
1
PMXA4354-2.2HBM: public61
Page 62
Electrical connections PMX
Connection time-out: If one or more of
the connections to this device are in
time-out, the network status display
red
green
red
-
LEDLEDStatusSignificance
MS
green
green
red
red
Flashing
Flashing
Off
On
Flashing
Flashing
Flashing
flashes red. This status is only
terminated when all connections in time-
out have been restored or when the
Self-test: The network status display
flashes green/red when the device is
carrying out a self-test.
Not switched on, no IP address: If the
device has no IP address (or is switched
off), the network status display does not
Device operational: If in operation and
running correctly, the network status
display will light up and remain green.
Standby: If the device has not been
configured, the module status display
Serious error: If the device detects a
serious error that cannot be rectified, the
module status display lights up and
Simple error*: If the device detects a
serious error that cannot be rectified, the
module status display lights up and
NOTE: A faulty or inconsistent
configuration is classified, e.g. as a simple
SignificanceStatusLEDLED
device is reset.
light up.
flashes green.
remains red.
remains red.
error.
62A4354-2.2HBM: publicPMX
Page 63
Electrical connections PMX
SignificanceStatusLEDLED
Self-test: The module status display
green
red
-
LEDLEDStatusSignificance
Flashing
Off
flashes green/red when the device is
carrying out a self-test.
Not switched on: If the device is not
switched on, the module status display
does not light up.
1
2
green
-
yellow
OnConnection to Ethernet established
Off
Flashing
The device has no connection to the
Ethernet
The device is sending/receiving Ethernet
frames
PMXA4354-2.2HBM: public63
Page 64
Electrical connections PMX
PX01PN
2
P2
Port 2
Port 1
P1
PROFINET
LEDLEDStatusSignificance
On
SF
BF
LEDLEDStatusSignificance
1
2--No function
red
red
green
Flashing
On
Flashing
Permanently on
Flashing
Off
System error, incorrect configuration
Flashing for device detection is controlled
by the IO controller
No connection or no configuration
Bus error, incorrect configuration, not all
IO devices are connected
Connection established
Send / Receive
No connection
1
64A4354-2.2HBM: publicPMX
Page 65
Electrical connections PMX
8.2.5Measurement card LEDs
PX401, channel status
Status LED
LEDStatusSignificance
green
yellow
red
Onno errors
FlashingFirmware update
Parameter not OK,
On
overloaded
PMXA4354-2.2HBM: public65
Page 66
Electrical connections PMX
PX455, channel status
Status LED
LEDStatusSignificance
green
yellow
red
Onno errors
On
Flashing
On
No transducer connected or wire break (self-
calibration ongoing)
Firmware update
Parameter not OK, transducer error,
overloaded
66A4354-2.2HBM: publicPMX
Page 67
Status LED
Electrical connections PMX
PX460, channel status
LEDStatusSignificance
green
yellow
red
Onno errors
On
Flashing
On
No transducer connected or wire break or
firmware update
Parameter not OK, transducer error,
overloaded
PMXA4354-2.2HBM: public67
Page 68
Electrical connections PMX
PX878
One status LED per channel
LEDStatusSignificance
Digital
green
green
Analog
green
red
On
Off
On
Off
On
Off
On
Digital output: High
Digital output: Low
Digital input: High
Digital input: Low
Analog output configured
Analog output not configured
Analog output overloaded,
signal invalid
68A4354-2.2HBM: publicPMX
Page 69
Electrical connections PMX
8.3Supply voltage
Notice
Device damage caused by high voltages.
If you are using power supply 1NTX001 listed in the ac
cessories, note the enclosed safety instructions.
With a separate DC voltage power supply (10 to 30 V
DC, nom. 24 V, power output at least 20 W), the PMX
device is supplied with voltage via the POWER socket (1)(see Chapter 11 Startup).
To operate transducers (load cells, force transducers,
etc.) in potentially explosive atmospheres, intrinsically
safe measurement circuits (Ex II (1) GD, [EEx ia]IIC)
must be set up on the PX455 by connecting safety
barriers (Zener barriers) type SD01A. The safety barriers
must also be mounted on the DIN rail like the PMX. An
ATEX test certificate must be available for the
transducers used. Transducers with 350 ohm bridge
resistance can be used. One transducer per
measurement channel of the PC455 can be operated
(parallel connection is not possible).
The PX455 offers 4 measurement channels with 4.8kHz
carrier frequency measurement. If the measurement
signals need to be added/subtracted or mean values
formed, this can be set using the internal PMX calculation
channels.
connection
Operation with Zener barriers
Important
In addition to the SD01A, the negative operating voltage
of the PMX must be grounded!
Cable lengths of up to max. 100 m are permissible. TEDS
cannot be used.
The measurement accuracy of the PX455 with operation
of the SD01A is max. 0.5%.
70A4354-2.2HBM: publicPMX
Page 71
Electrical connections PMX
8.4.2PX455
Four individually configurable 4 SG full or half
bridges (4.8kHz CF). Inductive full or half bridges,
LVDT, potentiometric sensors, piezoresistive
sensors, TEDS (0-wire), sensor detection
0‐wire TEDS
see Chapter 8.7.3
The bridge excitation voltage is 2.5 V.
PMXA4354-2.2HBM: public71
Page 72
Electrical connections PMX
8.4.2.1Strain gage and inductive full and half
bridges (6-wire configuration)
SG and inductive full bridges
wh
Measurement
signal (+)
bk
Bridge excitation
voltage (-)
rd
Measurement signal (-)
bu
Bridge excitation
voltage (+)
gn
Sense lead (+)
gy
Sense lead (-)
ye
Cable shield
1
2
4
3
'
3
2'
SG and inductive half bridges
Plug terminal
Cable color code for HBM transducer cables:
wh= white; bk= black; bu= blue; rd= red; ye = yellow; gn= green; gy= gray
Transducer connection
Fig. 8.1PX455 plug terminal in a six-wire circuit
wh
Measurement
signal (+)
bk
Bridge excitation
voltage (-)
bu
Bridge excitation
voltage (+)
gn
gy
Sense lead (-)
ye
Cable shield
Sense lead (+)
1
2
3
3'
2'
72A4354-2.2HBM: publicPMX
Page 73
Electrical connections PMX
8.4.2.2Strain gage and inductive full and half
bridges (4-wire configuration)
Four-wire connection:
full bridge
wh
bk
rd
bu
ye
Feedback bridges for a four-wire configuration
Cable color code: wh= white; bk= black; bu= blue; rd= red; ye = yellow; gn= green; gy= gray
1
2
4
3
3'
2'
Connection: Half bridge
wh
bk
bu
gn
gy
ye
Fig. 8.2PX455 plug terminal in four/three-wire configuration
1
2
3
3'
2'
PMXA4354-2.2HBM: public73
Page 74
Electrical connections PMX
8.4.2.3LVDT
LVDT transducer
Plug terminal
Transducer connection
Measurement signal (+)
Bridge excitation voltage (-)
Bridge excitation voltage (+)
Measurement signal (-)
Cable shield
Sense lead (+)
Sense lead (-)
Cable color code for HBM transducer cables:
wh= white; bk= black; bu= blue; rd= red; ye = yellow; gn= green; gy= gray
Transducer connection in a four/three wire
configuration:
When connecting a transducer in a four/three wire
configuration, the sense leads must be connected to the
corresponding bridge excitation circuit (PIN 2'-2 and
Pin 3'-3) by wire bridges, as otherwise a sensor error will
be detected.
TEDS functionality is not available when connecting in a
fourwire configuration.
PMXA4354-2.2HBM: public75
Page 76
Electrical connections PMX
8.4.2.5PX455 with Pt100 temperature measurement
With the PX455 measurement module, temperatures can
be measured with an uncertainty of +/-1°C without an
external preamplifier. A 100 ohm precision resistor
(R_comp) with a maximum tolerance of 0.1% is added to
a Pt100 resistor to make a half bridge circuit. The PMX
calculated channel "Pt100 on PX455" converts the meas
ured bridge unbalance to degrees Celsius and performs a
corrective calculation appropriate to the sensor cable be
ing used (R_wire).
Good heat dissipation is essential to keep the measure
ment error caused by the self heating of the Pt100 as low
as possible! This can be done, for example, by mounting
it on a metallic body.
Important
76A4354-2.2HBM: publicPMX
Page 77
Resistance thermometer
Electrical connections PMX
PX455
Measurement
signal +
Bridge excitation
voltage -
Bridge excitation
voltage +
Sense lead +
Cable shield
Sense lead -
Fig. 8.5PX455 with PT100 element for temperature
measurement
1
2
3
3'
2
'
PMXA4354-2.2HBM: public77
Page 78
Electrical connections PMX
8.4.3PX401
Four individually configurable current or voltage
inputs with 4 TEDS (1-wire) sensor detection.
1‐wire TEDS
see Chapter 8.7.3
IEPE sensors can also be operated in combination with
the Smart module (1-EICP-B-2).
78A4354-2.2HBM: publicPMX
Page 79
Electrical connections PMX
8.4.3.1Voltage source ± 10 V
TEDS
module
(optional)
Excitation for
external
transducers
U
1
2
+ U
TEDS
OUT +
OUT -
Cable shield
Fig. 8.6PX401 plug terminal Voltage source ±10 V
PMXA4354-2.2HBM: public79
Page 80
Electrical connections PMX
8.4.3.2Current source ± 20 mA
I
+ I
TEDS
module
(optional)
Excitation for
external
transducers
1
2
TEDS
OUT +
OUT -
Cable shield
Fig. 8.7PX401 plug terminal Current source ±20 mA (4-wire
circuit)
80A4354-2.2HBM: publicPMX
Page 81
TEDS-
module
(optional)
Electrical connections PMX
8.4.3.3Current source ± 20 mA
+ I
I
1
2
TEDS
OUT +
OUT -
Sensor supply
comes from the
PX401
Sensor supply and measure
ment signal are sent through
a connection cable
Fig. 8.8PX401 plug terminal Current source ±20 mA (2-wire
Cable shield
circuit)
PMXA4354-2.2HBM: public81
Page 82
Electrical connections PMX
Current-fed piezoelectric transducers, IEPE or IPC
transducers, are supplied with a constant current, e.g. 4
mA, and output a voltage signal that can be controlled
with the PX401 via an external module.
8.4.3.4IEPE transducers with external measuring
amplifier
PX401
IEPE
BNC
PIN:
Smart modules
(1-EICP-B-2)
2
6
3
5
4
1
2 Data
3 No function
Signal +
(+)
Signal -
(-)
Cable shield
24 V
0 V
1-wire EEPROM (optional) view from below
123
wh
gn
bk
1
2
7
rd
5
6
PX401
2
4
Fig. 8.9PX401 plug terminal lEPE transducers
82A4354-2.2HBM: publicPMX
Page 83
Electrical connections PMX
8.4.3.5PX401 with charge amplifier
Piezoelectric sensors can be operated with the PX401 via
external CMA or CMD charge amplifiers. The charge
amplifiers convert the sensor signal into a +/-10 V voltage
signal. The reset/operate charge amplifier signal can take
place in the PMX from an external controller or via a
PX878 digital output. The CMA charge amplifier can be
supplied directly by the PX401 measurement module.
Important
Due to the inrush current of the CMD charge amplifier,
the current feed of the CMD must be separate and not via
the PX401 measurement module.
PMXA4354-2.2HBM: public83
Page 84
Electrical connections PMX
Piezoelectric sensors with external charge amplifier
PX401
Charge cable
Piezoelectric sensor
PIN:
5
6
4
3
8
1
CMA
charge amplifier
Signal +
(+)
Signal -
(-)
TEDS transducer
identification
Cable shield
24 V
0 V
Measurement signal (reset)
gy
pk
ye
rd
wh
gn
1
2
4
7
5
6
PX878
1
5
6
PX401
Terminal 2
(digital
input/output)
Fig. 8.10PX401 pin assignment with external charge amplifier
84A4354-2.2HBM: publicPMX
Page 85
Electrical connections PMX
External transducers are supplied with power via the
PX401 measurement card (OUT + and OUT -). The
supply voltage is identical to the device supply voltage.
The maximum current is 400 mA per measurement card
and is distributed across the used transducers.
Important
The individual measurement channels on the PX401
measurement card are not electrically isolated from each
other. The PX401 measurement card has common elec
trical isolation from the basic device.
PMXA4354-2.2HBM: public85
Page 86
Electrical connections PMX
Channel
1
WGX basic device
PX401 housing
All four channels atone potential
Channel 4
Potential isolation
Fig. 8.11Potential isolation PX401
86A4354-2.2HBM: publicPMX
Page 87
Electrical connections PMX
8.4.4PX460
Torque flanges (torque, speed, angle of rotation),
angle/incremental encoders, SSI/PWM sensors and
frequency measurements up to 2 MHz
Channels 1 and 3: Frequency measurement (fixed)
Channels 2 and 4: Frequency (digital/inductive), counter,
encoder, SSI, PWM (adjustable)
The following measurement modes are available:
- Up to four torque flanges (T10, T12, T40) for torque
or speed measurement (without direction of rotation
detection)
- Or two torque flanges for simultaneous torque and
rotational speed measurement (without direction of
rotation/angle of rotation measurement)
- Or one torque flange for simultaneous
measurement of torque, speed, angle and direction
of rotation or reference pulse detection
- Or two angle/incremental encoders each, SSI/PWM
sensors, magnetic transducers or pulse counters
- Or four torque flanges for frequency measurement
up to 2 MHz, incl. two shunt calibrations and two
1-Wire TEDS (sensor detection)
PMXA4354-2.2HBM: public87
Page 88
Electrical connections PMX
SSI
1‐wire TEDS
see Chapter 8.7.3
Notice
The sensors for the PX460 are supplied externally with
voltage with the contacts (IN + -). The PX460 card then
provides the supply for 24 V (OUT + - ) and 5 V (5 V
OUT).
The input signals fed from the sensor into the PX460
must not exceed max. ±12 V. Otherwise the measure
ment inputs of the PX460 may be destroyed.
A shunt can be operated via Pin7. It can be activated via
the PMX web browser or the PMX command /dot
NET‐API/Catman.
88A4354-2.2HBM: publicPMX
Page 89
Electrical connections PMX
8.4.4.1Voltage supply for signal transmitter and
transducer up to 24 VDC nominal (rated)
voltage.
NOTE
The input signals fed from the
sensor into the PX460 must not
max.. ± 12 V
max. ± 12 V
max. ± 12 V
exceed max. ±12 V. Otherwise
the measurement inputs of the
PX460 may be destroyed.
Sensor
input signals
looped through, protected, 2 A max.
Sensor supply
Nominal (rated) voltage
24 VDC
Feed sensor supply
+
max. 24 VDC
-
Fig. 8.12PX460 Voltage supply options up to 24 VDC
nominal (rated) voltage
PMXA4354-2.2HBM: public89
Page 90
Electrical connections PMX
8.4.4.2Voltage supply for signal transmitter and
5 VDC, 200 mA
transducer up to 5 VDC nominal (rated)
voltage
Sensor
supply
Nominal
(rated)
voltage
5 VDC
Feed sensor supply
+
10 … 24 VDC
-
Fig. 8.13PX460 Voltage supply options up to 5 VDC nominal
3Supply voltage 18 V 30 Vbu
4Torque measurement signal
5Measurement signal 0V, symmetricgy
6Shunt signal trigger 5 V 30 Vgn
Top view
6
1
2
7
5
3
4
7
1)
RS-422 complementary signals; with cable lengths exceeding 10 m, we
recommend using a termination resistor R = 120 ohms between the (wh) and (rd) wires.
2)
RS 422: Pin 1 corresponds to A, Pin 4 corresponds to B.