HBM MP85, MP85DP Operating Manual

Page 1
Operating manual
Two−channel Amplifier
Module MP85
A0897-3.4 en
Page 2
Page 3
PME-MP85
3
Safety instructions 5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1 Introduction 8 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.1 Scope of supply and accessories 8 . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.2 General 8 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.3 Valuable information about PME MP85 documentation 9 . . . . . . . .
2 Commissioning 10 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.1 Philosophy of operation 11 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.1.1 Settings on the device 11 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.1.2 Overview of all groups and parameters 14 . . . . . . . . . . . . . . . . .
2.1.3 Configuring parameters on the device 15 . . . . . . . . . . . . . . . . . .
2.2 Configuring the hardware 16 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.2.1 Voltage supply / Transducer 16 . . . . . . . . . . . . . . . . . . . . . . . . . .
2.2.2 Connecting the CAN adapter (LPT and USB) 17 . . . . . . . . . . . .
2.3 Installing the PME Assistant software 18 . . . . . . . . . . . . . . . . . . . . . . .
2.3.1 System requirements 18 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.4 Starting the software and setting up 19 . . . . . . . . . . . . . . . . . . . . . . . . .
2.4.1 Using the LPT interface 19 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.4.2 Using the USB interface 21 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3 Switch settings / Replacing the battery 22 . . . . . . . . . . . . . . . . . . . . . . . .
4 MP85/MP85DP mounting/disassembly (schematic diagrams) 24 . . .
4.1 Linking several modules 25 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5 Connections 26 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.1 MP85 functional overview 26 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.2 MP85DP functional overview 27 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.3 Supply voltage and control inputs/outputs 28 . . . . . . . . . . . . . . . . . . . .
5.3.1 External supply voltage for control outputs 29 . . . . . . . . . . . . . .
5.3.2 Reference potential for control inputs 29 . . . . . . . . . . . . . . . . . . .
5.4 Transducer 30 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.4.1 Transducers with carrier frequency excitation 30 . . . . . . . . . . . .
5.4.2 Active encoders 33 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.5 CAN interface 35 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.6 Profibus interface (MP85DP only) 36 . . . . . . . . . . . . . . . . . . . . . . . . . . .
A08973.4 en HBM
Page 4
4
PME-MP85
6 Communication with the control system 37 . . . . . . . . . . . . . . . . . . . . . .
6.1 Timing of the test operations 37 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.2 Transducer test 39 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.3 Zero balance 39 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.4 Parameter set switching 40 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7 Error messages/operating status (LED) 41 . . . . . . . . . . . . . . . . . . . . . . .
8 Specifications 43 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
9 Declaration of conformity 49 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
10 Keyword index 51 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
A08973.4 enHBM
Page 5
PME-MP85
5

Safety instructions

Appropriate use
The MP85/MP85DP module and its connected transducers are to be used exclusively for measurement tasks and directly related control tasks. Use for any purpose other than the above shall be deemed to be not in accordance with the regulations.
To ensure safe operation, the device may only be operated in accordance with the information given in the Operating Manual. It is also essential to comply with the legal and safety requirements for the application concerned during use. The same applies to the use of accessories.
The device must not be connected directly to the mains supply. The supply voltage may be 18 to 30 V.
General dangers of failing to follow the safety instructions
The MP85/MP85DP module corresponds to the state of the art and is failsafe. The device may give rise to further dangers if it is inappropriately installed and operated by untrained personnel.
Any person instructed to carry out installation, commissioning, maintenance or repair of the device must have read and understood the Operating Manual and in particular the technical safety instructions.
Conditions on site
Protect the device from direct contact with water (IP20).
Maintenance and cleaning
The MP85/MP85DP module is maintenance-free. Please note the following points when cleaning the housing:
Before cleaning, disconnect the devices from the power supply.
Clean the housing with a soft, slightly damp (not wet!) cloth. You should
never use solvent, since this could damage the labelling on the front panel and the display.
When cleaning, ensure that no liquid gets into the device or connections.
Remaining dangers
The scope of supply and list of components provided with the MP85/MP85DP cover only part of the scope of measurement technology. In addition, equipment planners, installers and operators should plan, implement and respond to the safety engineering considerationsof measurement technology in such a way as to minimize remaining dangers. Prevailing regulations must be complied with at all times. There must be reference to the remaining dangers connected with measurement technology.
A08973.4 en HBM
Page 6
6
Any risk of residual dangers when working with the MP85/MP85DP is pointed out in this introduction by means of the following symbols:
PME-MP85
Symbol:
Meaning: Dangerous situation
Warns of a potentially dangerous situation in which failure to comply with safety requirements can result in death or serious physical injury.
Symbol:
Meaning: Potentially dangerous situation
Warns of a potentially dangerous situation in which failure to comply with safety requirements could result in damage to property or some form of physical injury.
Symbols indicating application notes and useful information:
Symbol:
WARNING
CAUTION
NOTE
Indicates that important information is given about the product or how to handle it.
Symbol:
Meaning: CE mark
With the CE mark, the manufacturer guarantees that his product complies with the requirements of the relevant EC guidelines (see Declaration of conformity at the end of this operating manual).
Working safely
Error messages must only be acknowledged when the cause of the error has been removed and no further danger exists.
The instrument complies with the safety requirements of DIN EN 61010, Part 1 (VDE 0411, Part 1) protection class I.
The device must be mounted on a supporting rail connected to the protective earth potential. Make sure, that at this spot both the supporting rail and the MP85/MP85DP module are free from dirt and other foreign material.
To guarantee adequate immunity from interference, the buses (CAN and in the case of MP85DP, Profibus DP) should be run with shielded and twisted 2-wire lines. The transducer lines should also be run using shielded cables.
A08973.4 enHBM
Page 7
PME-MP85
The shield of each transducer cable should be connected at the PME end by as short a line as possible (t5 cm) and a blade connector (4.8 mm; “Faston”).
The lines to connect the power supply and the digital control inputs and outputs only need to be shielded if the cable length exceeds 30 m or if the lines are installed outside closed buildings.
When connecting the lines (clip on and pull off clamps) and when replacing the MMC, measures must be taken against electrostatic discharge to prevent damage to the electronics.
The MP85/MP85DP module should be operated at a separated extralow voltage (SELV) (supply voltage 18 − 30 V DC) that usually supplies one or more consumers within a control panel.
1)
If the device is operated on a direct voltage network
, additional precautions
must be taken to discharge excess voltages.
7
Conversions and modifications
The MP85/MP85DP module must not be modified from the design or safety engineering point of view except with our express agreement. Any modification shall exclude all liability on our part for any resulting damage.
In particular, any repair or soldering work on motherboards is prohibited. When exchanging complete modules, use only original parts from HBM.
Qualified personnel
This instrument is only to be installed and used by qualified personnel strictly in accordance with the technical data and with the safety rules and regulations which follow. It is also essential to comply with the legal and safety requirements for the application concerned during use. The same applies to the use of accessories.
Qualified personnel means persons entrusted with the installation, assembly, commissioning and operation of the product who possess the appropriate qualifications for their function.
Maintenance and repair work on an open device with the power on must only be carried out by trained personnel who are aware of the danger involved.
1)
Distribution system for electrical power with greater spatial expansion (e.g. via several control panels) that may possibly even supply consumers with high nominal (rated) currents.
A08973.4 en HBM
Page 8
8

1 Introduction

1.1 Scope of supply and accessories

Scope of supply:
D 1 module MP85 or MP85DP
D 4 plugin type screw terminals, coded
1x CAN, 6pin Order no.: 3.33120250 2x transducers, 8pin Order no.: 3.33120299 1x I/O 1, 8pin Order no.: 3.33120301
in addition, for MP85:
1x I/O 2, 8pin Order no.: 3.33120301
D ribbon cable connector 10pin
D 1 Operating Manual module MP85/MP85DP
PME-MP85
Accessories:
D multimedia card (MMC), e.g. Sandisk (www.sandisk.com) D standard ribbon cable, 10pin, grid 1.27 mm

1.2 General

The MP85/MP85DP module from the PME product line is a twochannel measuring amplifier suitable for connecting the transducers of a great variety of technologies. In the case of the MP85DP, there is also a Profibus interface available for CAN-Interface.
Carrier-frequency
excitation voltage 4.8 kHz
Supply for active transducer
5V or 24V
A
D
Intelligent signal conditioning
and powerful algorithms for
A
press-fit monitoring
D
Voltage supply,
electrically isolated
µP
24 V
Control inputs and control outputs
CANopen interface
"10V
SSI
Keyboard &
display
MP85 / MP85DP
MMC memory
card
Fig. 1.1: Block diagram of the MP85/MP85DP module
Profibus DP interface (for MP85DP)
A08973.4 enHBM
Page 9
PME-MP85
The PME Assistant provides a simple user interface under MS-Windows for parameterizing the modules.
The PME Assistant is the sole means of parameterizing and setting up the MP85/MP85DP module (described in the Online Help).
To do this you need an LPT CAN or USB CAN interface converter, that has to be ordered separately (1PMESETUP; 1PMESETUP-USB).
The software allows you to set all the device parameters apart from the one for the CAN-BUS interface used for setup.
The software also allows you to set up the other devices of the PME family.

1.3 Valuable information about PME MP85 documentation

MP85 documentation comprises
9
this Operating Manual,
which describes mainly the configuration of the hardware (transducer, device and PC with the software).
the Online Help of the PME Assistant
which describes how to use the software to set up the device
a separate Operating Manual for the MP85DP with a CAN object index,
CAN interface description and Profibus communication
A08973.4 en HBM
Page 10
10
PME-MP85

2 Commissioning

To prepare the MP85 measurement system for a measurement task, the following conditions must be met or the relevant steps taken:
You need a PME module MP85 or MP85DP and the PME Assistant,
including software
The hardware must be configured (transducer, device, PC with the
software) (from Page 16)
The PME Assistant software must be installed (from Page 18)
To connect the PME module to a PC, you need a CAN adapter (from Page
17)
PCAN parallel interface LPT to CAN or
PCAN USB to CAN or
PCAN card (PCI or ISA) to CAN
The PME module must be connected to the CAN adapter
Once you have started the software, you can set up the PME module with
the PME Assistant (Online Help) (Page 19)
A08973.4 enHBM
Page 11
PME-MP85

2.1 Philosophy of operation

2.1.1 Settings on the device

The display of the MP85/85DP shows you the measured value and status information. The device is actually set up by means of the “PME Assistant” software (described in Online Help).
Display in measuring mode:
11
18.0024
status field
!
kg
x
Function of the keys:
SET
1. Switch from measuring mode to input mode
2. Choose the first parameter within the group.
3. Confirm your Input
4. Back to measuring mode (press for 2 secs.)
measured value unit
+
Select Parameters/Group
+
Flashes in the status field if the parameter value can be edited
Keys :
+
Keep key pressed value runs continuously Short press of the key advance one value at a time
A08973.4 en HBM
Page 12
12
During measurement you can press
1. Measured values
MVx measured value channel x
MVy measured value channel y
2. Process status, “ProcStat”
Alarm the last process was finished via the alarm window
OK the last process was OK
Not OK the last process was not OK
Started a process has been started and not yet finished
+
to view in the display:
PME-MP85
3. Status of the digital inputs and outputs
The MP85 has 5 inputs and 8 outputs. The MP85DP has 1 input and 4 outputs.
Input
set, not set
Output
set, not set
4. Profibus status
BD_SEAR (baud rate search) WT_PARM (waiting for parameterization) WT_CONF (waiting for configuration) DATA_EX (cyclical data traffic) ERROR (bus error)
A08973.4 enHBM
Page 13
PME-MP85
5. MMC-DisplayState (MMCState)
Display Explanation
notinuse No data storage on MMC
with any other display: data storage on MMC no MMC There is no MMC in the device
Init... The MMC is initialized automatically after insertion
SET ³ STOP MMC ready for storing. Press SET to reset to Stop state.
Then the MMC can be removed.
Storing Data is written to the MMC. Upon completion of data sto-
rage, the state changes to “Ready”; SET ³ STOP is dis­played.
Stopped Before removing the MMC from the device,
you must press the SET key to close all files and write the FAT.
13
The message “MMC Disk Closed” or in case of an error “MMC DiskClose Err” is displayed for a short time.
Then the MMC changes to the “Stopped” state and can be removed from the device.
If there is no more free memory available on the MMC, the status automatically changes to “Stopped”.
The only way to exit this state is to remove the MMC from the device. Afterwards, there will be an automatic reinitializa­tion.
5. Error types
During measurement, the character ! in the status field (whilst in measuring mode) indicates a module error.
+
The errors are displayed one after the other (obtainable by using
).
ERROR x (related to screw terminal SENSOR X)
ERROR y (related to screw terminal SENSOR Y)
SENSOR X
SENSOR Y
Possible error messages are summarized in
Chapter 7 ”Error messages”,
Page 41.
A08973.4 en HBM
Page 14

2.1.2 Overview of all groups and parameters

+
SET
+
Up
Down
CAN-BUS PROFIBUS ADDITIONAL FUNCTION
Baud rate Address AmplType
Address
OutpR. ms SrNo
MAINGRP
Overview of parameters
Groups
MAINGRP
Explanation of parameters:
Load MMC: Allows you to load a parameter set that has already been stored on the MMC using the PME Assistant.
PrgVers
HW vers
Load MMC
MAINGRP
We recommend to format the MMC from time to time for optimizing MMC access times.
MAINGRP with
A08973.4 en HBM
SET
back to Group
14
Page 15

2.1.3 Configuring parameters on the device

Measured value
+
2sec
SET
Groups
10kB 20kB 50kB 100kB
+
+/-
+
PROFIBUS
SET
ADDITIONAL FUNCTION
AmplTyp
e
PrgVers
CAN bus
SET SET SET
Baud rate
SET
125kB
Address
OutpR. ms
SET
SET
250kB 500kB 1000kB
1127
000
+/-
+/-
SETSET
Address
back MAINGRP
SET
3 123
+/-
SrNo
+
HW vers
Load MMC
+
SET
Measured value
MEAS MODE
Save ?
0 39
+/-
SET
Yes
No
+/-
SET
back MAINGRP
A08973.4 en HBM
Parameter values
+
Select Parameters
Flashes if the parameter value can be edited
SET
Confirm your Input
+/-
=
+
or
Back to measuring mode:
Print
SET
2sec
SET
back MAINGRP
15
Page 16
16
PME-MP85

2.2 Configuring the hardware

2.2.1 Voltage supply / Transducer

Connect the power supply cable and the transducer to the module
(Chapters 5.3 and 5.4).
CAUTION
Be sure to comply with the safety instructions!
Activate the power supply.
The device will run a self-test (approx. 10 secs) and will then, if working
properly, be in measuring mode. During the self-test, the control outputs stay at 0V.
If ! appears in the display or the status LED shows red, please consult Chapter 7 ”Error messages”.
Connecting the bus system
To find out how to connect several devices to a CAN bus, refer to Chapter 5.5. How to set up the CAN address and the baud rate is described in Chapter
2.1.3.
NOTE
A08973.4 enHBM
Page 17
PME-MP85
17

2.2.2 Connecting the CAN adapter (LPT and USB)

To enable PME devices to communicate with the PC, you first have to link the device to the PC.
1. Connecting LPT to CAN (parallel interface)
Switch off the PC and attach the LPT to CAN adapter to the parallel
interface port of the PC.
Unfortunately it is not possible to operate a printer from this parallel interface at the same time. You need a second parallel interface.
Attach the cable on the adapter between the male connector of the PS2
mouse and the female connector of the PC. This cable is needed to supply power for the adapter. Alternatively you can attach this power supply cable between the keyboard connector and the relevant PC socket.
Attach the CAN connector cable to the MP85 to the adapter.
Activate the PC.
2. Connecting USB to CAN
First install the PME Assistant.
Attach the USB to CAN adapter to a free USB interface on your PC.
The PC will immediately detect the adapter (Plug and Play).
A floppy disk is supplied with the adapter. It includes the driver required for installation.
A08973.4 en HBM
Page 18
18
PME-MP85

2.3 Installing the PME Assistant software

2.3.1 System requirements

To operate the PME Assistant software you need a PC with the following prerequisites:
min. processor requirement, Intel Pentium 400 MHz or equivalent
Windows 95 or higher, Windows NT Version 4.0 or later, Service Pack 3,
Windows 2000
Internet Explorer 5.0 or higher
RAM
32 MByte for Windows 95/98
64 MByte for Windows NT and Windows Millennium
128 MByte for Windows 2000 or later
Graphics card with min. resolution 800 x 600 pixels
min. 10 MBytes memory available on the hard disk
Microsoft or 100 % compatible mouse
configured default printer
LPT or USB interface
The following fonts must be installed:
Arial (TT)
Courier
MS Sans Serif
Times New Roman (TT)
Wingdings
A08973.4 enHBM
Page 19
PME-MP85

2.4 Starting the software and setting up

Interface mode
Opens the Setup window
19
The factory settings
of the interface
CAN bus scan for connected PME modules
Fig. 2.1: Startup window

2.4.1 Using the LPT interface

Start the PME Assistant program and in the startup window, enter the
interface details.
To configure the LPT interface, perform the following steps:
1. In the <Schnittstellentyp> dropdown list box, select the mode PCAN Dongle Std.
2. Enter the port address of the parallel interface in the <LPT Adresse (hex)> text box.
NOTE
Under Windows 2000, open System from the Control Panel and click the Device Manager button on the Hardware tab. Use Connections " Printer connection to display the available hardware. Double-click on an entry to open a dialog box to display the requisite information.
3. Select the interrupt of the parallel interface from the <LPT Interrupt> dropdown list box.
This interrupt must not be used for other devices.
A08973.4 en HBM
Page 20
20
NOTE
Under Windows 2000, make sure that an interrupt is used for the interface: display the properties for the printer connection (double-click). On the Connection settings tab, each Use interrupt assigned to the connection must be activated and the interrupt to be used (interrupt request) must be entered on the Resources tab.
This interrupt must only be used from this interface, double assignment by a second device, for example, an audio card, is not permissible.
4. In the <CAN−Baudrate> dropdown list box, select the baud rate to be set for the connected PME modules.
If you do not know the setting, you can use the keyboard of a PME module to query it (see Page 2.1.1 ).
PME-MP85
The baud rate must be the same for all the PME modules connected to the CAN bus. The factory setting is 1000 kB.
5. Click the <Scan> button.
In a few seconds, the addresses of the connected PME modules will be listed in the <CANID> dropdown list box in the Devices area.
6. Select a module and click on <SET>. This opens the Setup window (see Online Help).
You can now use the PME Assistant software to set up the MP85. This is described in the Online Help.
NOTE
A08973.4 enHBM
Page 21
PME-MP85

2.4.2 Using the USB interface

To configure the USB interface, perform the following steps:
Start the PME Assistant program and in the startup window, enter the
interface details.
1. In the <Interface type> dropdown list box, select the mode PCAN USB
2. In the <CAN-Baudrate> dropdown list box, select the baud rate to be set for the connected PME modules.
If you do not know the setting, you can use the keyboard of a PME module to query it (see Page 2.1.1 ).
NOTE
21
The baud rate must be the same for all the PME modules connected to the CAN bus. The factory setting is 1000 kB.
5. Click the <Scan> button.
In a few seconds, the addresses of the connected PME modules will be listed in the <CAN-ID> dropdown list box in the Devices area.
6. Select a module and click on <SET>. This opens the Setup window (see Online Help).
You can now use the PME Assistant software to set up the MP85. This is described in the Online Help.
A08973.4 en HBM
Page 22
22
PME-MP85

3 Switch settings / Replacing the battery

Modifying the supply voltage for active encoders
Use switch S1 to toggle the supply voltage for active encoders between the internal 5V supply and the external 24V supply. This is the only changeover
that the housing has to be open for.
NOTE
Switch S1 must be set/modified before fitting the PME.
To set switch S1, proceed as shown in Fig. 3.1.
1
Unscrew the
cover
Switch S1 for the supply voltage of active encoders
2
S1
5V
(int)
24V
(ext)
Fig. 3.1: Opening the housing, position of switch S1 (schematic diagram)
Terminating resistance
The CAN terminating resistance is activated/deactivated by switch S2 (see Fig. 3.2).
S2
Toggle switch S2 for the terminating resistance
ON OFF
Fig. 3.2: Switch for the CAN bus terminating resistance (schematic diagram)
A08973.4 enHBM
Page 23
PME-MP85
23
Replacing the battery
The MP85 has a realtime clock which is fed by a type CR2032 lithium battery. This can be removed from the battery holder and replaced at the point shown in Fig. 3.3. The battery should be replaced approx. every 5 years.
NOTE
When replacing the battery, you must remove the MP85/MP85DP from the DIN rail.
1
Battery for realtime clock
Unscrew the
cover
2
Fig. 3.3: Opening the housing, position of the battery (schematic diagram)
CAUTION
Check the polarity of the battery
Battery position
+
Fig. 3.4: Correct battery position
A08973.4 en HBM
Page 24
24
4 MP85/MP85DP mounting/disassembly
(schematic diagrams)
PME-MP85
Fig. 4.1: Mounting on a support rail
Fig. 4.2: Disassembly
CAUTION
The support rail must lie on protective-conductor potential .
A08973.4 enHBM
Page 25
PME-MP85

4.1 Linking several modules

Ribbon cable connector
Subsequent devices will be linked via this connector.
1.
3.
25
Color coding at Pin 1
2.
58 mm
recommended distance for the ribbon cable connectors
Fig. 4.3: Connecting the ribbon cable
Up to four MP85/MP85DP modules can be linked via a ribbon cable. This cable ensures local connection of the supply voltage, the CAN bus and synchronization of the carrier frequency between modules.
A08973.4 en HBM
Page 26
26

5 Connections

WARNING
Before starting the device, read the safety instructions.

5.1 MP85 functional overview

Local connection of CAN bus, supply voltage and synchronization between modules
Screw terminal 1: Voltage supply and CAN bus, synchronization
PME-MP85
LED 1 LED 2
Screw terminal 3:
1 potential-separated control input (24 V level) incl. input ground, 4 potential-separated control outputs (24 V level), injection of external supply for control outputs or incremental or SSI encoders
Screw terminal 2: (same assignment as screw terminal 1) CAN adapter for PC/laptop connection, parameterization via CAN bus
Two line LC display
Control keys
Multi Media Card (MMC) module
Screw terminal 4:
4 potential-separated control inputs (24 V level, related to input ground screw terminal 3) 4 potential-separated control outputs (24 V level, supply via screw terminal 3)
Screw terminal 5:
Transducer connection channel x including transducer excitation
Cable shield connection for transducers
Screw terminal 6: Transducer connection channel y including transducer excitation
A08973.4 enHBM
Page 27
PME-MP85

5.2 MP85DP functional overview

Local connection of CAN bus, supply voltage and synchronization between modules
Screw terminal 1: Voltage supply and CAN bus, synchronization
27
LED 1 LED 2
Screw terminal 3:
1 potential-separated control input (24 V level) incl. input ground, 4 potential-separated control outputs (24 V level), injection of external supply for control outputs or incremental or SSI encoders
Screw terminal 2: (same assignment as screw terminal 1) CAN adapter for PC/laptop connection, parameterization via CAN bus
Two line LC display
Control keys
Multi Media Card (MMC) module
Connection 4:
9-pin sub-D female connector for Profibus DP connection
Screw terminal 5:
Transducer connection channel x including transducer excitation
Cable shield connection for transducers
Screw terminal 6: Transducer connection channel y including transducer excitation
A08973.4 en HBM
Page 28
28
PME-MP85

5.3 Supply voltage and control inputs/outputs

Four (MP85) or three (MP85DP) connectable screw terminals are available for connecting the supply and the control inputs/outputs.
Connecting the voltage supply:
The MP85/MP85DP module must be connected to an external supply voltage of 1830 V (24 V
D The wire ends of the voltage supply must be provided with
D Attach the wire ends to screw terminal 1.
D Insert the screw terminal in the top socket. D Activate the voltage supply.
Screw terminal 2
(CAN adapter; assignment as
screw terminal 1)
WARNING
end sleeves for strands.
Labeling
nom
CAN
0 V
LH
).
24 V
SYNC
Screw terminal 1
Screw terminal 3
(control inputs/outputs)
Out 1 Out 2 Out 3
Out 4
0V
24V
IN 1
IN
IN = digital input OUT = digital output
CAUTION
1 2
OUT
3 4
0V
24V
1 IN
If there is a power failure at the MP85/MP85DP module, all the control outputs will be set to 0 V.
Screw terminal 4 (MP85 only)
(voltage supply CAN bus,
synchronization)
(control inputs/outputs)
Out 5 Out 6 Out 7
Out 8
IN 2 IN 3
IN 4 IN 5
5 6
OUT
7 8
2
3 IN 4
5
Fig. 5.1: Screw terminal assignment
The screw terminals are coded so that there will be no confusion when attaching them to the sockets. Sockets are equipped with coding tabs, screw terminals 1 and 2 with coding pins. In the case of screw terminals 3 and 4, the coding lugs are broken off. Screw terminals 3 and 4, 5 and 6, each have different grid dimensions.
A08973.4 enHBM
Page 29
PME-MP85

5.3.1 External supply voltage for control outputs

Example: PLC connection
29
Module MP85/MP85DP
OUT3
OUT1
24 V*
0 V*
Screw terminal 3
max. 0.5 A
PLC
Relay
max. 0.5 A
24 V
0 V*
Fig. 5.2: Connection to a PLC
Control outputs are available at screw terminals 3 and 4 and are electrically
isolated from the internal supply voltage:
at screw terminal 3: control outputs 1 4
at screw terminal 4: control outputs 5 8 (MP85 only)
)
*
Control outputs must be supplied with an external voltage (ground and 24V) via screw terminal 3.

5.3.2 Reference potential for control inputs

Control inputs are available at screw terminals 3 and 4 and are electrically
isolated from the internal supply voltage and from the control outputs.
at screw terminal 3: control input 1
at screw terminal 4: control inputs 2 5 (MP85 only)
For control inputs, it is necessary to connect an external reference potential (
IN ) to which the control input signals relate.
A08973.4 en HBM
Page 30
30
PME-MP85

5.4 Transducer

2 transducers can be independently connected at screw terminals 5 and 6.

5.4.1 Transducers with carrier frequency excitation

In “carrier frequency amplifier” mode, the following transducer types can be connected:
Screw terminals 5 and 6
for transducer connection
SENSOR X
18
SENSOR Y
18
S.G. and inductive full bridges, piezoresistive transducers
WH
Measurement signal (+)
BK
Excitation voltage (−)
RD
Measurement signal (−)
BU
Excitation voltage (+)
GN
Sensor circuit (+)
GY
Sensor circuit (−)
YE
Cable shield
MP85
1
2
6
4
5
3
Hous.
MP85DP
S.G. and inductive half bridges
WH
Measurement signal (+)
BK
Excitation voltage (−)
BU
Excitation voltage (+)
GN
Sensor circuit (+)
GY
Sensor circuit (−)
YE
Cable shield
1
2
4
5
3
Hous.
Potentiometric transducer *
Measurement signal (+)
Excitation voltage (−)
2
1
3
Excitation voltage (+)
Cable shield
Sensor circuit (+)
Sensor circuit (−)
)
1
2
4
Hous.
5
3
LVDT transducer
Measurement
signal (+) Excitation voltage (−)
Excitation voltage (+)
Measurement
signal (−)
Cable shield
Sensor circuit (+)
Sensor circuit (−)
1
2
4
6
Hous.
5
3
Cable core colors:
WH= white; BK= black; BU= blue; RD= red; YE= yellow; GN= green; GY= gray
Fig. 5.3: Connection of different transducers in “carrier frequency amplifier” mode
*)Function halfbridge
A08973.4 enHBM
Page 31
PME-MP85
31
When connecting a transducer in four-wire technology, the sensor circuits must be linked to the relevant bridge excitation circuit (pin 3 with pin 2 and pin
1)
5 with pin 4)
.
Four-wire connection:
Full bridge
WH
BK
RD
BU
YE
Hous
1
2
6
4
5
3
.
Four-wire connection:
Half bridge
WH
BK
BU
GN
GY
YE
Hous
1
2
6
4
5
3
.
Feedback bridges for four-wire technology
Cable core colors: WH= white; BK= black; BU= blue; RD= red; YE= yellow; GN= green; GY= gray
Fig. 5.4: Transducer connection in four-wire technology
NOTE
To connect the transducers, use standard cable from HBM or another shielded, low-capacitance measurement cable. Connect the shield of each transducer cable via the shortest possible lead (t5 cm) and a blade connector (4.8 mm; “Faston”) to the right of screw terminal 6. This ensures EMC protection.
1)
For cable lengths >50 m, instead of feedback bridges, a resistance with half the value of the bridge resistance (RB/2) must be activated at the transducer. If the transducers are calibrated in a six-wire circuit, the resistances must be activated directly into the sensor circuit.
A08973.4 en HBM
Page 32
32
Synchronization:
Synchronization is advisable for carrier-frequency fed transducers if
D the transducer cables of several devices are installed side by side
D the measuring points are close together and unshielded
Synchronisation prevents carrier frequency differences leading to disturbing superpositions.
PME-MP85
When synchronizing several modules, one device has to be declared the master. The remaining devices must be set to slave (use Setup).
Synchronization between modules should always even if you are working without a CAN bus be effected with the ribbon cable.
NOTE
A08973.4 enHBM
Page 33
PME-MP85
33

5.4.2 Active encoders

In “incremental encoder, SSI transducer or direct voltage encoder” mode, the following transducer types can be connected:
MP85
MP85DP
Pulse counter, incremental transducer (symmetric signals)
Ground
Ground
o
o
f1, 0
f1, 0
o
o
f2, 90
f2, 90
Transducers with SSI interface (symmetric signals)
Measurement signal F1 (+), 0°
Measurement signal F1 (), 0°
Cable shield Measurement signal F2 (+), 90°
Measurement signal F2 (), 90°
Zero index, Ix (+)
Zero index, Ix (−)
Supply voltage +5 V/+24 V
Ground
Clock, Cl (+)
7
6
1
Hous.
5
3
4
2
8
7
6
Clock, Cl (−)
Cable shield
Data, D (+)
Data, D (+)
Supply voltage +5 V/+24 V
DC voltage sources ("10V)
Measurement signal (+)
U
Measurement signal (−)
Cable shield
Fig. 5.5: Connecting active encoders
1
Hous.
5
3
8
5
3
Hous.
A08973.4 en HBM
Page 34
34
PME-MP85
A supply voltage is available at screw terminals 5 and 6, pins 7 and 8 for feeding incremental transducers and transducers with an SSI interface.
Switch S1 is used to toggle between an internal and an external supply. The device has to be opened to do this. (see Chapter 3).
Transducer supplied from the MP85/MP85DP:
Transducer supply voltage 5V "10 %, 150 mA max. (for both channels together). The supply is then not electrically isolated from the measurement system.
Transducer supplied from an external power pack:
At screw terminal 3, pins 5 and 6, a voltage between 10 30 V 24 V
) including ground is applied. A current of max. 300 mA (for both
DC
(nominal
DC
channels together) can then be drawn at transducer terminals 5 and 6.
This external supply voltage is electrically separated from the signal ground and at the same time feeds the digital control outputs.
Screw terminal 3
GND_int 5 V_int
GND_ext
24 V_ext
5
6
S1
78
Screw terminal 5
GND +5 V / 24 V
Screw terminal 6
78
Fig. 5.6: Supplying active encoders (schematic diagram)
A08973.4 enHBM
Page 35
PME-MP85
35

5.5 CAN interface

The CAN bus is connected via screw terminal 1. A maximum of 32 CAN nodes can be connected in one bus segment (under CANopen specification).
The CAN bus needs a terminating resistance of 120 in the first and last bus nodes. The bus line must have no more than two terminating resistances. A terminating resistance is integrated in the MP85DP module and is activated by toggle switch S2.
Low High
1st device
Fig. 5.7: Connecting the CAN interface
CAN low
CAN connection as per Fig. 5.7
First device in
the bus line
Switch in the terminating resistance here (toggle switch)
Do not switch in the
terminating resistance
CAN high
Last device in
the bus line
Switch in the terminating resistance here (toggle switch)
Fig. 5.8: CAN bus mode with several modules (max. 32 according to the standard)
NOTE
If the first or last device in the bus line is not a PME module, a 120 W resistance must be switched in to each of these external devices.
A08973.4 en HBM
Page 36
36
PME-MP85

5.6 Profibus interface (MP85DP only)

On the front panel of the MP30DP is a 9-pin D-sub connection socket for the Profibus connection.
GND
9
RS485-A
5
RS485-RTS
RS485-B
Vcc (5V)
6
1
Profibus
connection socket
Fig. 5.9: Profibus connection in accordance with standard
Installation:
Connect the MP85DP module to the 24 V supply voltage and set the
required Profibus address via the keyboard or the Setup program.
Connect the Profibus cable to the MP85DP module. Ensure that a
terminating resistance is connected to the first and last Profibus unit (the housing of the Profibus connector usually contains a sliding switch for this purpose).
Example:
Profibus connector
First device in
the bus line
sliding switch of Profibus connector
to ”Resistance ON” . . . .
Fig. 5.10: Profibus operation
PLC
Profibus connector
Last device in
the bus line
sliding switch of Profibus connector to ”Resistance ON”
A08973.4 enHBM
Page 37
PME-MP85
37

6 Communication with the control system

All the diagrams relate to a positive logic.

6.1 Timing of the test operations

a.) Storing curves and results without data loss
The below diagrams apply, if you choose the “without data loss” storage me­thod and want to save curves and/or results on the MMC or externally via CAN. The READY signal is not reset to 1 until the device is ready to save data for the next process.
NOTE
Before starting a new measurement, the READY bit must have been re­set to 1.
Start/Stop Input
Ready
Running
Result valid
External stop
The READY signal is not reset to 1 until there has been an external stop.
External stop or end condition occurred
Process in order
State undefined
Once the evaluation time has expired, you can
Initialisization typ. 25 ms
Data acquisition running
Time until stop
Offline evaluation
Data storage
A08973.4 en HBM
read off result 0 or 1 at the output.
Only if “Result valid” = 1, the process result (Pro­cess in order) is valid.
Page 38
38
PME-MP85
b.) Process optimized storage of curves and results
The below diagrams apply, if you choose the “Process optimized storage” sto­rage method and want to save curves and/or results on the MMC or externally via CAN. In contrast to the “Storing without data loss” method, the READY si­gnal can be reset to 1 even if the device is not yet ready to save new data.
Consequently the results and curve files might not be transmittes if you start a new measurement immediately. This method should only be chosen if extre­mely fast processes must be monitored and the results and curves are merely used for random sample checks.
NOTE
Before starting a new measurement, the READY bit must have been re­set to 1.
Start/Stop Input
Ready
Running
Result valid
External stop
The READY signal is not reset to 1 until there has been an external stop.
External stop or end condition occurred
Process in order
Initialisization typ. 25 ms
Data acquisition running
State undefined
Time until stop
Offline evaluation
Once the evaluation time has expired, you can read off result 0 or 1 at the output.
Only if “Result valid” = 1, the process result (Pro­cess in order) is valid.
Data storage
A08973.4 enHBM
Page 39
PME-MP85

6.2 Transducer test

Transducer test timing diagram
39
min. 5 ms
Input Transducer test
Output transducer test result

6.3 Zero balance

Zero balance timing diagram
min. 5 ms
Inputs
Zeroing
1
0
typically 10 ms
1
0
Input START
Settling time e.g. for 100 Hz filter: 20 ms
A08973.4 en HBM
Page 40
40

6.4 Parameter set switching

PME-MP85
Inputs
Switchover occurs when the level changes
Load parameter set bit 0 bit 4
Output Busy
typically 500 ms
Bit 4 Bit 3 Bit 2 Bit 1 Bit 0 Active parameter set
0 0 0 0 0 Factory settings
0 0 0 0 1 1
0 0 0 1 0 2
0 0 0 1 1 3
0 0 1 0 0 4
0 0 1 0 1 5
0 0 1 1 0 6
0 0 1 1 1 7
0 1 0 0 0 8
0 1 0 0 1 9
0 1 0 1 0 10
0 1 0 1 1 11
0 1 1 0 0 12
0 1 1 0 1 13
0 1 1 1 0 14
0 1 1 1 1 15
1 0 0 0 0 16
1 0 0 0 1 17
1 0 0 1 0 18
1 0 0 1 1 19
1 0 1 0 0 20
1 0 1 0 1 21
1 0 1 1 0 22
1 0 1 1 1 23
1 1 0 0 0 24
1 1 0 0 1 25
1 1 0 1 0 26
1 1 0 1 1 27
1 1 1 0 0 28
1 1 1 0 1 29
1 1 1 1 0 30
1 1 1 1 1 31
A08973.4 enHBM
Page 41
PME-MP85

7 Error messages/operating status (LED)

Depending on the display mode, various error messages can appear in the display instead of the measured value.
Error message Cause Remedy
TransdErr. Input signal overflow
Transducer not connected
Transducer wrongly connected
Amplifier not adapted to transducer
type
No sensor circuits connected
Connect transducer See pin assignment
Connect sensor circuits
41
ADC ovfl. Input signal of the A/D converter too
high
Grs ovfl. Gross value overflow Reduce display by one decimal
Scal.Err Input characteristic too steep Modify input characteristic
Flash error An error occurred when reading out
the Flash EOROM.
An error occurred at the CAN bus Check whether there are termi-
CAN bus error
Initial calibr. error No valid initial calibration values Start again, send PME to the
CAN Tx No PDO request on bus Check CAN bus configuration
Adjust hardware measuring
range
place
This can be a one-off read error,
please repeat the operation. If
the error occurs again, please
contact HBM Service.
nating resistances or whether a
channel is faulty. Then switch the
devices back on. If this does not
help, connect one device at a
time to the CAN bus to ascertain
which device is faulty.
manufacturer (HBM)
A08973.4 en HBM
Page 42
42
PME-MP85
Operating status MP85 / MP85DP: The LEDs show the operating status (ready to take measurements, overflow
etc.) of the device. With the MP85DP, however, the Profibus status is displayed instead of the
CAN status (as is the case with the MP85).
LED 1
LED 2
Operating status:
LED1 (OK / NOK)
Operating status LED Red LED Yell LED Green LED Flashes
Initialization during start-up x
Alarm x x
Process starts x x
OK x
Not OK x
LED2 (Status)
MP85 LED Red LED Yell LED Green LED Flashes
Initialization during start-up x
EE_InCalErr x
CANOFF x
No synchronization signal x
Scale error x
MMC error x
Overload x x
LCD Err x x
CAN communitation x x
PreOperational x
Operational x
MP85DP LED Red LED Yell LED Green LED Flashes
Status: Error x
Status: BD_SEAR, WT_PARM, WT_CONF x
Status: DATA_EX x
A08973.4 enHBM
Page 43
PME-MP85

8 Specifications

Type MP85 / MP85DP
Accuracy class 0.1
43
Supply voltage V
DC
24; Potential separation from measuring
system (typically 500 V
Permitted supply voltage range V 18...30
Power consumption
MP85, typically MP85DP, typically MP85 / MP85DP, max.
Backing battery for realtime clock
W W W
10
Years 5
7 9
(CR2032), typical lifetime
Operating modes, independently adjustable for both channels
Carrier-frequency amplifier
Carrier frequency kHz 4.8 "1 %
Excitation voltage V
rms
2.5 "5 %
Attachable process quantity transducer
S.G., half and full bridges Inductive half and full bridges, LVDTs Potentiometric transducer
mH
170 2000
4 160
170 2000
Input sensitivity Measuring range (mV/V)
4 100 1000
mV/V 0.2 4 3.5 100 50 1000
Transducer cable length, max. m 500
DC
.)
Scale range, max. Digits 999999, at 10 % of input measuring
range
Scale range, min. Digits 100, at 100 % of input measuring range
Maximum permitted common-mode voltage
V "5,5
Common-mode rejection
0 60 0 1000 0 4800
dB dB dB
u120
u96 u50
Linearity deviation % t0.03
Noise voltage typically Measuring range (mV/V)
4 100 1000
0 - 1 Hz 0 10 Hz 0 100 Hz 0 1000 Hz
µV/V µV/V µV/V µV/V
pp pp pp pp
0.1
0.25 1 2
2.5 6
25 50
25
60 250 500
Sampling rate, max. 1/s 2400
A08973.4 en HBM
Page 44
44
Measurement frequency range
adjustable
Low pass 4th order with Bessel
characteristic
Nomin
al
(rated)
value
fc (Hz)
1000 980 1400 0.550 0.260 4
,
0.05 0.049 0.081 3750 4280 0
1dB
off 790 1200 0.450 0.290 3
500 440 690 0.860 0.510 1.5 200 190 320 1.6 1.11 1.5 100 100 160 2.9 2.13 1.3
50 51 83 4.6 4.24 1 20 25 41 8.2 8.36 1 10 13 21 15.5 16.8 0
5 6.1 10.3 30.2 33.4 0 2 3.1 5.2 60 67 0 1 1.6 2.6 119 137 0
0.5 0.79 1.30 240 272 0
0.2 0.19 0.32 950 1070 0
0.1 0.09 0.16 2500 2170 0
3dB
(Hz)
(Hz)
Phase
delay
(ms)
PME-MP85
Rise time
(ms)
Over-
shoot
Shunt calibration mV/V 1"3%
Effect of operating voltage
on zero point on sensitivity
Effect of 10 K change in ambient
% f.s. % f.s.
Measuring range (mV/V)
t0.01 t0.01
temperature
4 100 1000
on zero point full bridge on zero point half bridge on sensitivity
Long-term drift over 48 h(meas. range
µV/V µV/V
%
1
10
0.05
20 40
0.05
µV/V 2
200 200
0.05
4 mV/V; 0.5 h after powering up)
DC-voltage transducer
Attachable process quantity transducer DC-voltage transducer, voltage sources
Nominal (rated) measuring range V "10
(%)
Input signal range V "10.5
Scale range, max. Digits 999999, at 10 % of input measuring range
Scale range, min. Digits 100, at 100 % of input measuring range Internal resistance of the signal source k v1
Maximum permitted common-mode
V 2
voltage
Measurement frequency range,
Hz 0.05 1000
adjustable (-1 dB)
Filter characteristics Bessel, 4th order
Linearity deviation % t0.03
Sampling rate max. 1/s 2400
A08973.4 enHBM
Page 45
PME-MP85
Incremental encoder
Attachable process quantity transducer Incremental encoder (up/down counter
with zero index signal)
Voltage supply 5 V, max. 150 mA or 24 V
max. 300 mA
Dual-channel mode Time division multiplex method
Inputs (F1 ("), F2 ("), Ix (")) Differential inputs (RS422)
Input level
LOW level HIGH level Each line to measurement earth, max.
Hysteresis, typically V 0.07
Maximum permitted common-mode voltage
Input impedance, typically k 12
Detection of direction of rotation Via "90o phase-shifted signal F2
Input range pulse counting Imp 0 999999
V V V
V 7 ... +12
t0.8
u2
"14
45
Maximum pulse rate Imp/s 1 000 000 Gap between two consecutive edges
F1("), F2(")
Scale range, max. Digits 20 for 1 pulse
Scale range, min. Digits 1 for 10,000 pulses
Measurement frequency range, adjustable (-1 dB)
Sampling rate max. 1/s 2400
SSI transducer
Attachable process quantity transducer Disp. and angle transducer with SSI inter.
Power supply 5 V, max. 150 mA; 24 V, max. 300 mA
Dual-channel mode Time division multiplex method
Data inputs D (") Differential input (RS422)
Input level, data input D (")
LOW level HIGH level Each line to measurement earth, max.
Hysteresis V 0.07
Maximum permitted common-mode voltage
Clock output Cl (")
Differential output voltage Cl("), without load, max. Differential output voltage Cl("), RL= 50 Ohm, min.
Common-mode output voltage on Cl
("), max.
ns u400
Hz 0.05 1000
V V V
V 7 ... +12
Differential output (RS422)
V
V
V 3
t0,8
u2
"14
5.8
2
Short-circuit current, clock outputs Cl ("), typically
A08973.4 en HBM
mA 100
Page 46
46
Resolution, single turn Bits 12, 13
Resolution, multi turn Bits 24, 25
Scale range, max. Digits 20 for 1 pulse
Scale range, min. Digits 1 for 10,000 pulses
PME-MP85
Measurement frequency range,
Hz 0.05 1000
adjustable (-1 dB)
Sampling rate, max. 1/s 1200
Baud rates kBaud 100, 200, 500, 1000
Coding Gray Code
General specifications
Limit switches
Number 4 per channel
Reference level Gross
Hysteresis % 1 ... 100
Adjustment accuracy Digit 1
Response time, typically (fc=1000 Hz) ms t2
Control outputs
Number 4 (MP85DP) / 8 (MP85)
Nominal voltage, external power supply V
DC
24
Permitted supply voltage range V 10 30
Maximum output current per output A 0.5
Short-circuit current, typically
. = 24 V, RLt 0.1 Ohm)
(U
ext
A 0.8
Short-circuit period Unlimited
Isolation voltage, typically V
DC
500
Control inputs
Number 1 (MP85DP) / 5 (MP85)
Input voltage range LOW V 0 5
Input voltage range HIGH V 10 30
Input current, typically (HIGH level = 24V) mA 12
Isolation voltage, typically V
DC
500
CAN interface
Sampling rate, max. 1/s 1200
Protocol CAN 2.0B, CANopen compatible
Hardware bus link to ISO 11898
Baud rates Maximum cable lengths
Kbit/sm1000 500 250 125 100 50 20 10
25 100 250 500 600 1000 10001000
Terminating resistance Connectable by switch
Connection Terminals
Parameter memory (Flash) 31 plus factory settings
A08973.4 enHBM
Page 47
PME-MP85
Multimedia card (memory card on Flash base)
47
Usable types M-
bytes
Data transmission rate, typically K-byte
/s
File system DOS
Display
Type Two line, 8 character alphanumeric, LCD
Keyboard Touch-sensitive keypad with three keys
Temperature range
Nominal temperature range
Operating temperature range
Storage temperature range
Degree of protection IP20
Dimensions mm 55 x 146 x 156
Weight, approx. g 800
Profibus DP interface (MP85DP only)
o
C 0 50
o
C 20 ... +50
o
C 20 ... +70
8, 16, 32, 64
8
Protocol Profibus DP slave, as per DIN 192453
Baud rate, max. MBaud 12
Node address 3 123, can be set via keyboard
Profibus ID number Hex 699
Configuration data byte 5
Parameter data, max. byte 6 (+7DP standard)
Parameter assignment (asynchronous) to DPV1 standard
Input data, max. byte 142
Output data, max. byte 40
Inputs update rate ms 1 (for 4 measured values)
Outputs update rate ms t10, on zeroing, limit values
Diagnosis data byte 48
Profibus connection 9-pin sub-D (DIN19245-3),
potential-separated from power supply
and signal ground
A08973.4 en HBM
Page 48
48
Specifications for analysis unit
PME-MP85
Max. number of data triples
4000 (auto. data reduction)
(channel x); (channel y); (time) Sampling rate, max. Hz 2400 Analysis Max. number of analysis windows 9 Type of window Oblique or straight Analysis methods per window Plot analysis
Mean value of x or y analysis in window
x coordinates for tolerance windows absolute or relative to the start position, or
relative to the end position
Y coordinates for tolerance windows absolute or
relative to F
of tolerance window 2
min
or
relative to F
of tolerance window 2
max
Stop conditions External stop signal
Target value y + decay time Target value x + decay time
Standstill recognition
Number of independent parameter sets 31 Switching parameter sets, max. ms 500 Duration of offline evaluation straight
ms 15 + 0.15/data pair in window
window Duration of offline evaluation oblique
ms 15 + 0.3/data pair in window
window Statistics (for each parameter set, separately) Maximum number of fitting processes 4 x 10 Number of histogram classes for 2
, x
values (x
max
min
, y
max
, y
min
)
9 per tolerance window
9
A08973.4 enHBM
Page 49
PME-MP85

9 Declaration of conformity

49
A08973.4 en HBM
Page 50
50
PME-MP85
A08973.4 enHBM
Page 51
PME-MP85

10 Keyword index

51
A
active encoders, 22 , 33
active parameter set, 40
C
CAN bus, 15 , 26 , 27 , 35
connecting, 26 , 27
CAN interface, 35
CANopen, 35
CANopen interface, connecting, 35
coding pin, 28
coding tab, 28
commissioning, 10
configuring parameters, 15
connecting
bus system, 16 CAN adapter, 17 CAN adapters, 9 CAN interface, 35 LPT to CAN, 17 supply voltage, 28 transducers, 30 , 33 USB to CAN, 17
D
DC voltage source, 33
declaration of conformity, 49
digital input, 28
digital inputs, 12
digital output, 28
disassembly, 24
display mode, 13
dongle, 19
E
error message, 41
F
feedback bridges, 31
fourwire technology, 31
I
interface, connecting, 35
interface converter, 9
connecting the bus system, 16
connecting transducers, 30
active encoders, 33 S.G. full and half bridges, inductive full
and half bridges, potentiometric, pie­zoelectric, LVDT, 30 , 33
transducers with CF excitation, 30
connection, pulse counter, incremental
transducer, symmetric, 33
control inputs, 26 , 27
control inputs and outputs, 26 , 27
control outputs, 26 , 27
A08973.4 en
L
LED, 42
LPT interface, 19
M
measured value channel x, measured va-
lue channel y, 12
MMC load, 14
MMCState, 13
mounting, 24
multimedia card, 8
HBM
Page 52
52
PME-MP85
P
parameters, 14
configuring, 15
PLC connection, 29
PME Assistant, 18
power failure, 28
process status, 12
Profibus interface, 36
Profibus status, 12
R
replacing the battery, 22 , 23
ribbon cable, 32
S
screw terminal assignment, 28
selftest, 16
supply voltage, 28
synchronization, 26 , 27 , 32
T
terminating resistance, 22 , 35
Test timing, 39
U
USB interface, 21
V
voltage supply, 26 , 27 , 28
screw terminal, 26 , 27
voltage supply, CAN bus, synchroniza-
tion, control inputs, control outputs, 26 , 27
Z
Zero balance, 39
HBM
A08973.4 en
Page 53
PME-MP85
53
A08973.4 en
HBM
Page 54
54
PME-MP85
HBM
A08973.4 en
Page 55
Page 56
pp
A08973.4 en
Modifications reserved. All details describe our products in general form only.They are not to be understood as express warranty and do not constitute any liability whatsoever.
Hottinger Baldwin Messtechnik GmbH
Postfach 10 01 51, D-64201 Darmstadt Im Tiefen See 45, D-64293 Darmstadt Tel.: +49/61 51/ 8 03-0; Fax: +49/61 51/ 8039100 Email: su
ort@hbm.com www.hbm.com
Loading...