HBM MP85, MP85DP Operating Manual

Operating manual
Two−channel Amplifier
Module MP85
A0897-3.4 en
PME-MP85
3
Safety instructions 5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1 Introduction 8 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.1 Scope of supply and accessories 8 . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.2 General 8 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.3 Valuable information about PME MP85 documentation 9 . . . . . . . .
2 Commissioning 10 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.1 Philosophy of operation 11 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.1.1 Settings on the device 11 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.1.2 Overview of all groups and parameters 14 . . . . . . . . . . . . . . . . .
2.1.3 Configuring parameters on the device 15 . . . . . . . . . . . . . . . . . .
2.2 Configuring the hardware 16 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.2.1 Voltage supply / Transducer 16 . . . . . . . . . . . . . . . . . . . . . . . . . .
2.2.2 Connecting the CAN adapter (LPT and USB) 17 . . . . . . . . . . . .
2.3 Installing the PME Assistant software 18 . . . . . . . . . . . . . . . . . . . . . . .
2.3.1 System requirements 18 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.4 Starting the software and setting up 19 . . . . . . . . . . . . . . . . . . . . . . . . .
2.4.1 Using the LPT interface 19 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.4.2 Using the USB interface 21 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3 Switch settings / Replacing the battery 22 . . . . . . . . . . . . . . . . . . . . . . . .
4 MP85/MP85DP mounting/disassembly (schematic diagrams) 24 . . .
4.1 Linking several modules 25 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5 Connections 26 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.1 MP85 functional overview 26 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.2 MP85DP functional overview 27 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.3 Supply voltage and control inputs/outputs 28 . . . . . . . . . . . . . . . . . . . .
5.3.1 External supply voltage for control outputs 29 . . . . . . . . . . . . . .
5.3.2 Reference potential for control inputs 29 . . . . . . . . . . . . . . . . . . .
5.4 Transducer 30 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.4.1 Transducers with carrier frequency excitation 30 . . . . . . . . . . . .
5.4.2 Active encoders 33 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.5 CAN interface 35 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.6 Profibus interface (MP85DP only) 36 . . . . . . . . . . . . . . . . . . . . . . . . . . .
A08973.4 en HBM
4
PME-MP85
6 Communication with the control system 37 . . . . . . . . . . . . . . . . . . . . . .
6.1 Timing of the test operations 37 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.2 Transducer test 39 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.3 Zero balance 39 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.4 Parameter set switching 40 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7 Error messages/operating status (LED) 41 . . . . . . . . . . . . . . . . . . . . . . .
8 Specifications 43 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
9 Declaration of conformity 49 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
10 Keyword index 51 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
A08973.4 enHBM
PME-MP85
5

Safety instructions

Appropriate use
The MP85/MP85DP module and its connected transducers are to be used exclusively for measurement tasks and directly related control tasks. Use for any purpose other than the above shall be deemed to be not in accordance with the regulations.
To ensure safe operation, the device may only be operated in accordance with the information given in the Operating Manual. It is also essential to comply with the legal and safety requirements for the application concerned during use. The same applies to the use of accessories.
The device must not be connected directly to the mains supply. The supply voltage may be 18 to 30 V.
General dangers of failing to follow the safety instructions
The MP85/MP85DP module corresponds to the state of the art and is failsafe. The device may give rise to further dangers if it is inappropriately installed and operated by untrained personnel.
Any person instructed to carry out installation, commissioning, maintenance or repair of the device must have read and understood the Operating Manual and in particular the technical safety instructions.
Conditions on site
Protect the device from direct contact with water (IP20).
Maintenance and cleaning
The MP85/MP85DP module is maintenance-free. Please note the following points when cleaning the housing:
Before cleaning, disconnect the devices from the power supply.
Clean the housing with a soft, slightly damp (not wet!) cloth. You should
never use solvent, since this could damage the labelling on the front panel and the display.
When cleaning, ensure that no liquid gets into the device or connections.
Remaining dangers
The scope of supply and list of components provided with the MP85/MP85DP cover only part of the scope of measurement technology. In addition, equipment planners, installers and operators should plan, implement and respond to the safety engineering considerationsof measurement technology in such a way as to minimize remaining dangers. Prevailing regulations must be complied with at all times. There must be reference to the remaining dangers connected with measurement technology.
A08973.4 en HBM
6
Any risk of residual dangers when working with the MP85/MP85DP is pointed out in this introduction by means of the following symbols:
PME-MP85
Symbol:
Meaning: Dangerous situation
Warns of a potentially dangerous situation in which failure to comply with safety requirements can result in death or serious physical injury.
Symbol:
Meaning: Potentially dangerous situation
Warns of a potentially dangerous situation in which failure to comply with safety requirements could result in damage to property or some form of physical injury.
Symbols indicating application notes and useful information:
Symbol:
WARNING
CAUTION
NOTE
Indicates that important information is given about the product or how to handle it.
Symbol:
Meaning: CE mark
With the CE mark, the manufacturer guarantees that his product complies with the requirements of the relevant EC guidelines (see Declaration of conformity at the end of this operating manual).
Working safely
Error messages must only be acknowledged when the cause of the error has been removed and no further danger exists.
The instrument complies with the safety requirements of DIN EN 61010, Part 1 (VDE 0411, Part 1) protection class I.
The device must be mounted on a supporting rail connected to the protective earth potential. Make sure, that at this spot both the supporting rail and the MP85/MP85DP module are free from dirt and other foreign material.
To guarantee adequate immunity from interference, the buses (CAN and in the case of MP85DP, Profibus DP) should be run with shielded and twisted 2-wire lines. The transducer lines should also be run using shielded cables.
A08973.4 enHBM
PME-MP85
The shield of each transducer cable should be connected at the PME end by as short a line as possible (t5 cm) and a blade connector (4.8 mm; “Faston”).
The lines to connect the power supply and the digital control inputs and outputs only need to be shielded if the cable length exceeds 30 m or if the lines are installed outside closed buildings.
When connecting the lines (clip on and pull off clamps) and when replacing the MMC, measures must be taken against electrostatic discharge to prevent damage to the electronics.
The MP85/MP85DP module should be operated at a separated extralow voltage (SELV) (supply voltage 18 − 30 V DC) that usually supplies one or more consumers within a control panel.
1)
If the device is operated on a direct voltage network
, additional precautions
must be taken to discharge excess voltages.
7
Conversions and modifications
The MP85/MP85DP module must not be modified from the design or safety engineering point of view except with our express agreement. Any modification shall exclude all liability on our part for any resulting damage.
In particular, any repair or soldering work on motherboards is prohibited. When exchanging complete modules, use only original parts from HBM.
Qualified personnel
This instrument is only to be installed and used by qualified personnel strictly in accordance with the technical data and with the safety rules and regulations which follow. It is also essential to comply with the legal and safety requirements for the application concerned during use. The same applies to the use of accessories.
Qualified personnel means persons entrusted with the installation, assembly, commissioning and operation of the product who possess the appropriate qualifications for their function.
Maintenance and repair work on an open device with the power on must only be carried out by trained personnel who are aware of the danger involved.
1)
Distribution system for electrical power with greater spatial expansion (e.g. via several control panels) that may possibly even supply consumers with high nominal (rated) currents.
A08973.4 en HBM
8

1 Introduction

1.1 Scope of supply and accessories

Scope of supply:
D 1 module MP85 or MP85DP
D 4 plugin type screw terminals, coded
1x CAN, 6pin Order no.: 3.33120250 2x transducers, 8pin Order no.: 3.33120299 1x I/O 1, 8pin Order no.: 3.33120301
in addition, for MP85:
1x I/O 2, 8pin Order no.: 3.33120301
D ribbon cable connector 10pin
D 1 Operating Manual module MP85/MP85DP
PME-MP85
Accessories:
D multimedia card (MMC), e.g. Sandisk (www.sandisk.com) D standard ribbon cable, 10pin, grid 1.27 mm

1.2 General

The MP85/MP85DP module from the PME product line is a twochannel measuring amplifier suitable for connecting the transducers of a great variety of technologies. In the case of the MP85DP, there is also a Profibus interface available for CAN-Interface.
Carrier-frequency
excitation voltage 4.8 kHz
Supply for active transducer
5V or 24V
A
D
Intelligent signal conditioning
and powerful algorithms for
A
press-fit monitoring
D
Voltage supply,
electrically isolated
µP
24 V
Control inputs and control outputs
CANopen interface
"10V
SSI
Keyboard &
display
MP85 / MP85DP
MMC memory
card
Fig. 1.1: Block diagram of the MP85/MP85DP module
Profibus DP interface (for MP85DP)
A08973.4 enHBM
PME-MP85
The PME Assistant provides a simple user interface under MS-Windows for parameterizing the modules.
The PME Assistant is the sole means of parameterizing and setting up the MP85/MP85DP module (described in the Online Help).
To do this you need an LPT CAN or USB CAN interface converter, that has to be ordered separately (1PMESETUP; 1PMESETUP-USB).
The software allows you to set all the device parameters apart from the one for the CAN-BUS interface used for setup.
The software also allows you to set up the other devices of the PME family.

1.3 Valuable information about PME MP85 documentation

MP85 documentation comprises
9
this Operating Manual,
which describes mainly the configuration of the hardware (transducer, device and PC with the software).
the Online Help of the PME Assistant
which describes how to use the software to set up the device
a separate Operating Manual for the MP85DP with a CAN object index,
CAN interface description and Profibus communication
A08973.4 en HBM
10
PME-MP85

2 Commissioning

To prepare the MP85 measurement system for a measurement task, the following conditions must be met or the relevant steps taken:
You need a PME module MP85 or MP85DP and the PME Assistant,
including software
The hardware must be configured (transducer, device, PC with the
software) (from Page 16)
The PME Assistant software must be installed (from Page 18)
To connect the PME module to a PC, you need a CAN adapter (from Page
17)
PCAN parallel interface LPT to CAN or
PCAN USB to CAN or
PCAN card (PCI or ISA) to CAN
The PME module must be connected to the CAN adapter
Once you have started the software, you can set up the PME module with
the PME Assistant (Online Help) (Page 19)
A08973.4 enHBM
PME-MP85

2.1 Philosophy of operation

2.1.1 Settings on the device

The display of the MP85/85DP shows you the measured value and status information. The device is actually set up by means of the “PME Assistant” software (described in Online Help).
Display in measuring mode:
11
18.0024
status field
!
kg
x
Function of the keys:
SET
1. Switch from measuring mode to input mode
2. Choose the first parameter within the group.
3. Confirm your Input
4. Back to measuring mode (press for 2 secs.)
measured value unit
+
Select Parameters/Group
+
Flashes in the status field if the parameter value can be edited
Keys :
+
Keep key pressed value runs continuously Short press of the key advance one value at a time
A08973.4 en HBM
12
During measurement you can press
1. Measured values
MVx measured value channel x
MVy measured value channel y
2. Process status, “ProcStat”
Alarm the last process was finished via the alarm window
OK the last process was OK
Not OK the last process was not OK
Started a process has been started and not yet finished
+
to view in the display:
PME-MP85
3. Status of the digital inputs and outputs
The MP85 has 5 inputs and 8 outputs. The MP85DP has 1 input and 4 outputs.
Input
set, not set
Output
set, not set
4. Profibus status
BD_SEAR (baud rate search) WT_PARM (waiting for parameterization) WT_CONF (waiting for configuration) DATA_EX (cyclical data traffic) ERROR (bus error)
A08973.4 enHBM
PME-MP85
5. MMC-DisplayState (MMCState)
Display Explanation
notinuse No data storage on MMC
with any other display: data storage on MMC no MMC There is no MMC in the device
Init... The MMC is initialized automatically after insertion
SET ³ STOP MMC ready for storing. Press SET to reset to Stop state.
Then the MMC can be removed.
Storing Data is written to the MMC. Upon completion of data sto-
rage, the state changes to “Ready”; SET ³ STOP is dis­played.
Stopped Before removing the MMC from the device,
you must press the SET key to close all files and write the FAT.
13
The message “MMC Disk Closed” or in case of an error “MMC DiskClose Err” is displayed for a short time.
Then the MMC changes to the “Stopped” state and can be removed from the device.
If there is no more free memory available on the MMC, the status automatically changes to “Stopped”.
The only way to exit this state is to remove the MMC from the device. Afterwards, there will be an automatic reinitializa­tion.
5. Error types
During measurement, the character ! in the status field (whilst in measuring mode) indicates a module error.
+
The errors are displayed one after the other (obtainable by using
).
ERROR x (related to screw terminal SENSOR X)
ERROR y (related to screw terminal SENSOR Y)
SENSOR X
SENSOR Y
Possible error messages are summarized in
Chapter 7 ”Error messages”,
Page 41.
A08973.4 en HBM

2.1.2 Overview of all groups and parameters

+
SET
+
Up
Down
CAN-BUS PROFIBUS ADDITIONAL FUNCTION
Baud rate Address AmplType
Address
OutpR. ms SrNo
MAINGRP
Overview of parameters
Groups
MAINGRP
Explanation of parameters:
Load MMC: Allows you to load a parameter set that has already been stored on the MMC using the PME Assistant.
PrgVers
HW vers
Load MMC
MAINGRP
We recommend to format the MMC from time to time for optimizing MMC access times.
MAINGRP with
A08973.4 en HBM
SET
back to Group
14

2.1.3 Configuring parameters on the device

Measured value
+
2sec
SET
Groups
10kB 20kB 50kB 100kB
+
+/-
+
PROFIBUS
SET
ADDITIONAL FUNCTION
AmplTyp
e
PrgVers
CAN bus
SET SET SET
Baud rate
SET
125kB
Address
OutpR. ms
SET
SET
250kB 500kB 1000kB
1127
000
+/-
+/-
SETSET
Address
back MAINGRP
SET
3 123
+/-
SrNo
+
HW vers
Load MMC
+
SET
Measured value
MEAS MODE
Save ?
0 39
+/-
SET
Yes
No
+/-
SET
back MAINGRP
A08973.4 en HBM
Parameter values
+
Select Parameters
Flashes if the parameter value can be edited
SET
Confirm your Input
+/-
=
+
or
Back to measuring mode:
Print
SET
2sec
SET
back MAINGRP
15
16
PME-MP85

2.2 Configuring the hardware

2.2.1 Voltage supply / Transducer

Connect the power supply cable and the transducer to the module
(Chapters 5.3 and 5.4).
CAUTION
Be sure to comply with the safety instructions!
Activate the power supply.
The device will run a self-test (approx. 10 secs) and will then, if working
properly, be in measuring mode. During the self-test, the control outputs stay at 0V.
If ! appears in the display or the status LED shows red, please consult Chapter 7 ”Error messages”.
Connecting the bus system
To find out how to connect several devices to a CAN bus, refer to Chapter 5.5. How to set up the CAN address and the baud rate is described in Chapter
2.1.3.
NOTE
A08973.4 enHBM
PME-MP85
17

2.2.2 Connecting the CAN adapter (LPT and USB)

To enable PME devices to communicate with the PC, you first have to link the device to the PC.
1. Connecting LPT to CAN (parallel interface)
Switch off the PC and attach the LPT to CAN adapter to the parallel
interface port of the PC.
Unfortunately it is not possible to operate a printer from this parallel interface at the same time. You need a second parallel interface.
Attach the cable on the adapter between the male connector of the PS2
mouse and the female connector of the PC. This cable is needed to supply power for the adapter. Alternatively you can attach this power supply cable between the keyboard connector and the relevant PC socket.
Attach the CAN connector cable to the MP85 to the adapter.
Activate the PC.
2. Connecting USB to CAN
First install the PME Assistant.
Attach the USB to CAN adapter to a free USB interface on your PC.
The PC will immediately detect the adapter (Plug and Play).
A floppy disk is supplied with the adapter. It includes the driver required for installation.
A08973.4 en HBM
Loading...
+ 39 hidden pages