HBM ML71BS6 Operating Manual

Page 1
Operating manual
MGCplus Measuring amplifier system CAN bus interface
ML71BS6
A0851-3.1 en
Page 2
Page 3
MGCplus−ML71BS6
3
Contents Page
Safety instructions 4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1 Introduction 6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2 Connections 7 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.1 Pin assignment 7 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3 Front panel 8 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4 Parameterization 9 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.1 Settings CAN1 /CAN2 9 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.1.1 Panel output signals CAN 1 11 . . . . . . . . . . . . . . . . . . . . . .
5 Menu structure in set-up mode 13 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6 CAN protocol 14 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.1 Measurement signal transmission 14 . . . . . . . . . . . . . . . . . . . . . . . . .
6.1.1 Absolute values in Float 14 . . . . . . . . . . . . . . . . . . . . . . . . .
6.1.2 In digits as Long 14 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.1.3 In digits as Word 14 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.1.4 Identifier assignment 14 . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7 Specifications 16 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Page 4
4
MGCplus−ML71BS
6

Safety instructions

Use in accordance with the regulations
The ML71BS6 CAN bus module is to be used exclusively for measurement tasks and directly related control tasks. Use for any purpose other than the above shall be deemed to be not in accordance with the regulations.
To ensure safe operation, the device may only be operated in accordance with the information given in the Operating Manual. It is also essential to comply with the legal and safety requirements for the application concerned during use. The same applies to the use of accessories.
General dangers of failing to follow the safety instructions
The ML71BS6 CAN bus module complies with the state of the art and is fail-safe. The device may give rise to further dangers if it is inappropriately installed and operated by untrained personnel.
Any person instructed to carry out installation, commissioning, maintenance or repair of the device must have read and understood the Operating Manual and in particular the technical safety instructions.
Remaining dangers
The scope of performance and supply of the ML71BS6 only covers part of the range of measurement technology. In addition, equipment planners, installers and operators should plan, implement and respond to the safety engineering considerations of measurement technique in such a way as to minimise remaining dangers. Prevailing regulations must be complied with at all times. There must be reference to the remaining dangers connected with measurement technique. In this manual, the following symbols are used to draw your attention to any remaining dangers that might occur when working with the ML71BS6:
Symbol:
WARNING
Meaning: Dangerous situation Warns of a potentially dangerous situation in which failure to comply with
safety requirements can lead to death or serious physical injury.
A0851-3.1 enHBM
Page 5
MGCplus−ML71BS6
Symbol: Meaning: CE mark The CE mark enables the manufacturer to guarantee that the product com-
plies with the requirements of the relevant EC directives (the declaration of conformity is available at http://www.hbm.com/HBMdoc).
Working safely
Error messages must only be acknowledged when the cause of the error has been removed and no further danger exists.
The device complies with the safety requirements of DIN EN 61010-Part 1 (VDE 0411-Part 1); Protection Class I.
To ensure adequate immunity from interference, use only Greenline shielded ducting (see HBM offprint ”Greenline shielding design, EMC-compliant mea­suring cable; G36.35.0)
5
Conversions and modifications
No modifications that affect the design or the technical safety of the ML71BS6 CAN bus module may be carried out without our express agreement. Any modification shall exclude all liability on our part for any resulting damage.
In particular any repair or soldering work on motherboards is prohibited. When exchanging complete modules, use only original parts from HBM.
Qualified personnel
This instrument must only to be installed and used by qualified personnel, strictly in accordance with the technical data and the safety requirements and regulations listed. It is also essential to comply with the legal and safety requirements for the application concerned during use. The same applies to the use of accessories.
Qualified personnel means persons entrusted with the installation, assembly, commissioning and operation of the product who possess the appropriate qualifications for their function.
Maintenance and repair work on an open device with the power on must only be carried out by trained personnel who are aware of the danger involved.
A0851-3.1 en HBM
Page 6
6
MGCplus−ML71BS
6

1 Introduction

The ML71BS6 is an MGCplus system module that takes measurement data from the MGC and outputs it as CAN messages. It records up to 128 measured values per CAN port, scales them and outputs them at the CAN bus.
Two independent CAN interfaces per channel and two different physical layers (ISO 11898−24V and a ”low-speed” link based on the Philips TJA1053 transceiver module) are supported.
The measurement data to be transmitted can be selected either by using the AB22A/AB32 display and control panel or by using the “MGCplus Assistant” software from HBM.
A0851-3.1 enHBM
Page 7
MGCplus−ML71BS6

2 Connections

WARNING
Before starting the device, read the safety instructions.

2.1 Pin assignment

The CAN bus is connected to the AP71 connection board by a 9-pin Sub-D connector. You can connect two independent CAN field bus systems (female connectors CAN1 and CAN2). Both the CAN interfaces are used to output the MGCplus measurement data. Each of the interfaces can be assigned up to 128 measured values.
7
Connectors (male)
CAN1/CAN2
1
CAN_L
CAN_GND
CAN Shield
Physical layer: High: Standard CAN speeds up to 1MBaud acc. to ISO11898
(24 V dielectric strength). Low: Low−speed bus for in-car communication up to 100 kBaud
(CAN_V+ must be connected to 12 V)
5
GND
6
CAN_H
CAN_V+
9
Fig. 2.1: CAN connection
A0851-3.1 en HBM
Page 8
8
MGCplus−ML71BS
6

3 Front panel

The front panel LEDs have a dual function: In Level mode, the signal level at the analogue output is displayed. In Status mode, the LEDs are assigned as shown below:
ML71 BS6
LED caption Colour Meaning in Status mode
CHAN. yellow Channel selected ERROR/WARN. red error/warning Rx1 not used Tx1/OVRN1 red/yel-
low BUSERR1 red Bus error at CAN1 Rx2 yellow not used Tx2/OVRN2 red/yel-
low BUSERR2 red Bus error at CAN2
CAN protocol transmitted
CAN protocol transmitted
Analog output (BNC connector (female))
Any voltage from −10 V to +10 V can be output through the female BNC connector by interface command.
A0851-3.1 enHBM
Page 9
MGCplus−ML71BS6
Format
9

4 Parameterization

4.1 Settings CAN1 /CAN2

The baud rate, the transmission formats, the number of signals and the output rate of the 1st CAN bus are set here. The same menu is applied correspondingly to the 2nd CAN bus.
The maximum number of signals for a bus is 128 signals. The following output rates are offered for selection depending on the number of output signals. In the case of the 2nd CAN port, please note that the maximum possible output rates can only be achieved if no signals have been selected at CAN port1. If a selected rate cannot be achieved, the messages will be output as quickly as possible and an error message will be generated. Transmission will then not be at a fixed frequency.
Number of signals
CAN1 CAN2
4 4 INT16 2400 800 4 4 INT32 2400 600 4 4 Float32 2400 400 8 8 INT16 1200 800 8 8 INT32 1200 400
8 8 Float32 1200 300 16 16 INT16 600 600 16 16 INT32 600 300 16 16 Float32 600 300
Output rate [Hz]
Port1 Port2
The first ID of a range can be set. The last ID is derived from this and from the number of signals.
A0851-3.1 en HBM
Page 10
10
MGCplus−ML71BS
6
1 Use the shift key to change to set-up mode and then press the
function key confirm with
Settings CAN1 CHANNEL3
Baud rate Format Number of signals
Output rate ID range used First ID Last ID
System Display Parameters Options
. Use the cursor keys to select “Settings CAN1” and
.
1000 kBaud Float 8
1200 1/2
640 641
Reset CAN1 port
Float 32−bit integers 16−bit integers
1 1/s 10 1/s
. .
1200 1/s
1000 kBaud 500 kBaud
. . 10 kBaud
Fig. 4.1: CAN1 / CAN2 dialogue window
2 Use the cursor keys to choose the baud rate selection field. Use to
call the selection list and choose the desired baud rate. Confirm with
3 Select the output rate in the same way as for 2. 4 Use the cursor keys
to select the “Number of signals” edit field and
enter the desired figure. 5 Select the output rate in the same way as for 2. 6 Use the cursor keys
to select the “First ID” edit field and enter the
desired figure. The “Last ID” edit field is set automatically. 7 Use the shift key
follow-up question with
to change to measuring mode and confirm the
.
.
A0851-3.1 enHBM
Page 11
MGCplus−ML71BS6

4.1.1 Panel output signals CAN 1

This is where to set the sources for the signals to be transmitted and if necessary, scaling for the output values.
1 Use the shift key to change to set−up mode and then press the
11
function key confirm with
. Use the cursor keys to select “Sources CAN1” and
.
Format 1; Absolute values Float
Sources CAN1 CHANNEL 3
Source for MGC channel
Signal
System Display Parameters Options
Fig. 4.2: CAN1 / CAN 2 output signals dialogue window
CAN signal 1 8
Gross
Gross Net Store1 Store2 Combined PV
CAN signal 1 CAN signal 2 . . CAN signal n
2 Use the cursor keys to select the “Source for MGC channel” selection
field. Use Confirm with
3 Use the cursor keys
to call the selection list and choose the desired CAN signal.
.
to select the MGC channel field and enter the
desired channel (e.g. 8.1).
4 Use the cursor keys
selection list and choose the desired signal. Confirm with
to select the “Signal” edit field. Use to call the
.
A0851-3.1 en HBM
Page 12
12
6
Formats 2 and 3; 32-bit integers and 16-bit integers
You must also specify scaling here.
MGCplus−ML71BS
1 Use the shift key
function key
confirm with
Sources CAN1 CHANNEL 3
Source for MGC channel
Signal Scaling
Point 1
Point 2
System Display Parameters Options
7680000
to change to set−up mode and then press the
. Use the cursor keys to select “Sources CAN1” and
.
CAN signal 1 8
Gross
10
1000
0
dig kN
dig kN
Gross Net Store1 Store2 Combined PV
Channel 8 to 128
Fig. 4.3: CAN1 / CAN 2 output signals dialogue window
CAN signal 1 CAN signal 2 . . CAN signal n
2 Use the cursor keys to select the “Source for MGC channel” selection
field. Use
Confirm with
to call the selection list and choose the desired signal.
.
3 Use the cursor keys to select the MGC channel field and enter the
desired channel (e.g. 8.1). 4 Use the cursor keys
selection list and choose the desired signal. Confirm with
to select the “Signal” edit field. Use to call the
.
5 Use the cursor keys to select the “Point 1” edit field and enter the digit
value. 6 Use the cursor keys
to select the next edit field and enter the value of
the accompanying physical unit. 7 Use the cursor keys
to select the “Point 2” edit field and proceed in the
same way as for points 5 and 6. 8 Use the shift key
follow-up question with
to change to measuring mode and confirm the
.
A0851-3.1 enHBM
Page 13
MGCplus−ML71BS6

5 Menu structure in set-up mode

The set−ups for the MGCplus unit are subdivided into groups according to
13
their function. Pressing the shift key the selection bar is displayed.
Display in Setup mode
Setup for channel 3
Display
Parameters
DisplaySystem Parameters Options
System−specific set­tings that in most ca­ses are only carried out during the initial start-up or for a new measurement task.
User−defiened dis­play settings such as the required method of representing va­lues, function key as­signment or the spe­cification of channel names.
Output signal CAN1 Output signal CAN2
takes you to the set-up dialogue, and
OptionsSystem
Setup menu
Settings CAN1 Settings CAN2 Version
You will find more notes on working with the AB22A display and control panel in the “MGCplus with AB22A/AB32” manual.
A0851-3.1 en HBM
Page 14
14
MGCplus−ML71BS
6

6 CAN protocol

Up to 128 measured values can be transmitted at each bus. A number of signals is assigned to each bus. This determines the signal
transmission rate at the particular CAN port (in the case of CAN port 2, the operating mode of CAN port 1 is also a factor).

6.1 Measurement signal transmission

The operating priority for the two CAN ports is different. CAN port 1 always takes priority, to ensure that the transmission pattern of the data sent from here is exactly periodic. When sending a synch signal, the values sent thus far are valid here. CAN port 2 has the lower priority. Measurement signals can be transmitted in the following three formats.

6.1.1 Absolute values in Float

In each 8−byte CAN message, 2 signals are transmitted as Float values (32−bit). This is dealing with scaled measured values.

6.1.2 In digits as Long

In each CAN message, 2 of these values are transmitted as Long (32-bit),
1)
with the MSByte
always being 0. Digital value scaling can be preset by using
two point calibrations.

6.1.3 In digits as Word

In each CAN message, 4 of these values are transmitted as Words (16-bit). Digital value scaling can be preset by using two point calibrations.

6.1.4 Identifier assignment Default assignment

The possible formats are sent out for the same signal under different identifiers, so that fixed scaling can be assigned in the receiver of a specific ID. The synch message at CAN port 1 has the identifier 80h
1)
Most significant byte
2)
This hexadecimal number corresponds to the decimal value 128
2)
and does not contain any data bytes.
A0851-3.1 enHBM
Page 15
MGCplus−ML71BS6
Measurement data messages CAN port 1:
Absolute values in Float (identifier: default values)
Identifier 8−byte data field
15
0x200−0x23F 4−byte measured value 1 (absolute
Float)
In digits from 7680000 (= 24−bit) as Long
Identifier 8−byte data field
0x240−0x27F 4−byte measured value 1 (signed
integers in digits from 7680000)
In digits from 30000 (= 16−bit) as Word
Identifier 8−byte data field
0x280−0x29F
2−byte measured
value 1 (signed in-
tegers in digits
from 30000)
2−byte measured
value 1 (signed in-
tegers in digits
from 30000)
Measurement data messages CAN port 2:
Absolute values in Float
Identifier 8−byte data field
4−byte measured value 2 (absolute
Float
4−byte measured value 1 (signed
integers in digits from 7680000)
2−byte measured
value 1 (signed in-
tegers in digits
from 30000)
2−byte measured
value 1 (signed in-
tegers in digits
from 30000)
0x300−0x33F 4−byte measured value 1 (absolute
Float)
4−byte measured value 2 (absolute
Float)
In digits from 7680000 (= 24−bit) as Long
Identifier 8−byte data field
0x340−0x37F 4−byte measured value 1 (signed
integers in digits from 7680000)
4−byte measured value 1 (signed
integers in digits from 7680000)
In digits from 30000 (= 16−bit) as Word
Identifier 8−byte data field
0x380−0x39F
2−byte measured
value 1 (signed in-
tegers in digits
from 30000)
2−byte measured
value 1 (signed in-
tegers in digits
from 30000)
2−byte measured
value 1 (signed in-
tegers in digits
from 30000)
2−byte measured
value 1 (signed in-
tegers in digits
from 30000)
Adjusting the ranges
As a variant to the default setting, you can choose the start of the ID range you are using for the selected format at the particular CAN bus.
A0851-3.1 en HBM
Page 16
16
MGCplus−ML71BS
6

7 Specifications

CAN interface
Number of CAN interfaces 2 Protocol CAN 2.0B Baud rate Baud 10k 20k 50k 125k 250k 500k 1M Line length m 1000 1000 1000 500 250 100 25 Hardware bus link individually selecta-
ble for each CAN interface Connection technique 2x 9−pin DSUB, individual potential separation
Measurement recording
Number of signals at 1 Mbaud, 11 bit Format INT16; Signals
1 ... 8 9 ... 16 1200 17 ... 24 800 25 ... 32 600 33 ... 48 400 49 ... 64 300 65 ... 96 200 97 ... 128 100
Format Float, INT32; Signals
1 ... 4 5 ... 8 1200 9 ... 12 800 13 ... 16 600 17 ... 24 400 25 ... 32 300 33 ... 48 200 49 ... 96 100 97 ... 120 80 121 ... 128 50
Database with parameter information via the CAN signals
Size of databases Byte 2x 100 k Database storage non−volatile in ML71B flash memory
Hz
Hz 2400
Standard High Speed ISO 11898−24 V
Fault Tolerant Low Speed
of supply and ground
Output rate
2400
2 (one database per CAN interface)
A0851-3.1 enHBM
Page 17
MGCplus−ML71BS6
Mechanical values
Nominal (rated) temperature range Operating temperature range Storage temperature range Operating voltage V +14.6 ... +17.0 (<90 mA)
Card format mm Europe 100 x 160 Width mm 20.3 (4 divs) Receptacle DIN 41612 indirectly
Analogue output
The analogue output can display one of the max. 128 input signals at a time
Nominal (rated) voltage V "10 asymmetrical Permissible load resistance k > 5 Internal resistance < 5 Non−linearity % < 0.05 Effect of a 10 K−change of the ambient
temperature on the sensitivity Effect of a 10 K−change of the ambient
temperature on the zero point
_C [°F] _C [°F] _C [°F]
%
mV
−20 ...+ 60 [−4 ... +140]
−20 ...+ 60[−4 ... +140]
−25 ... +70[−13 ... +158]
−14.6 ... –17.0 (<100 mA)
−7 ... −9 (<10 mA)
< 0.08 typically 0.04
3 typ. 2
17
A0851-3.1 en HBM
Page 18
18
MGCplus−ML71BS
6
A0851-3.1 enHBM
Page 19
Page 20
Email: support@hbm.com Internet: www.hbm.com
Modifications reserved.
All details describe our products in general form only.They are
A0851−3.1 en
not to be understood as express warranty and do not constitute any liability whatsoever.
Hottinger Baldwin Messtechnik GmbH
Postfach 10 01 51, D-64201 Darmstadt Im Tiefen See 45, D-64293 Darmstadt Tel.: +49 6151 803-0 Fax: +49 6151 8039100
Loading...