In order to ensure sufficient immunity from disturbance only use Greenline
shielding (see HBM’s special publication ”Greenline” Screening Concept, electromagnetic compatibility of measuring cable, G36.35.0)
The connection lines’ ( ≤ 50 V) minimum insulation strength must be
350 V (AC).
Appropriate use
The ME50S6 with the connected transducers may be used for measurement
and directly related control and regulation tasks, only. Any other use is not appropriate.
To ensure safe operation, the ME50S6 may only be used according to the
specifications given in this manual. When using the transducer, the legal and
safety regulations for the respective application must also be observed. The
same applies if accessories are used.
General dangers in the case of non-observance of the safety instructions
The ME50S6 complies with the state of the art and is operationally reliable. If
the device is used and operated inappropriately by untrained personnel, residual dangers might develop.
Any person charged with device installation, operation, maintenance or repair
must in any case have read and understood the operating manual and the
safety instructions, in particular.
Conditions on site
Protect the device from moisture or atmospheric influences such as rain,
snow, etc.
HBMA0260-1.1 en
ME50S6
5
Maintenance and cleaning
The Clip Electronic are maintenance-free. Please note the following points
when cleaning the front panel:
•Remove the mains plug from the socket before cleaning.
•Clean the front panel with a soft, slightly damp (not wet!) cloth. Never use
solvents, since they may damage the labelling on the front panel.
•When cleaning, please ensure that no liquid finds its way into the device or
onto the contacts.
Residual dangers
The ME50S6 scope of performance and supply covers part of the measuringtechnology, only. The plant designer/constructor/operator must in addition design, realise and take responsibility for the measuring-system’s safety such
that potential residual dangers are minimized. The respective regulations
must in any case be observed. Residual dangers regarding the measuring
system must be specified explicitly.
HBMA0260-1.1 en
6
If there is any risk of remaining dangers when working with the, it is pointed
out in this introduction by means of the following symbols:
ME50S6
Symbol:
Meaning: Maximum danger level
Warns of a decidedly dangerous situation in which failure to comply with
safety requirements can lead to death or serious physical injury.
Symbol:
Meaning:Possibly dangerous situation
Warns of a potentially dangerous situation in which failure to comply with
safety requirements can lead to death or serious physical injury.
Symbol:
Meaning:Dangerous situation
Warns of a possibly dangerous situation in which failure to comply with safety
requirements can cause damage to property or lead to some form of physical
injury.
DANGER
WARNING
CAUTION
Symbols indicating application notes and useful information:
Symbol:
Means that important information about the product or its handling is being
given.
Symbol:
Meaning: CE mark
The CE mark is a guarantee by the manufacturer that his product meets the
requirements of the relevant EG Directives (see declaration of conformity at
the end of these operating instructions).
NOTE
HBMA0260-1.1 en
ME50S6
Safe operation
Do only quit error messages if the reason for the error has been eliminated
and there is no more danger.
Reconstruction and modifications
HBM’s express consent is required for modifications regarding the ME50S6
construction and safety. HBM does not take responsibility for damage resulting from unauthorized modifications.
In particular, repair and soldering works on the boards are prohibited. If complete componentry is replaced use original HBM components, only.
Qualified personnel
The device may be used by qualified personnel, only; the technical data and
the special safety regulations must in any case be observed. When using the
device, the legal and safety regulations for the respective application must
also be observed. The same applies if accessories are used.
7
Qualified personnel means: personnel familiar with the installation, mounting,
start-up and operation of the product, and trained according to their job.
Maintenance and repair work on an open device with the power on should
only be undertaken by trained personnel who are aware of the above-mentioned dangers.
HBMA0260-1.1 en
8
ME50S6
1Measuring Amplifiers on Eurocards ME50S6
The Eurocards ME50S6 are 4.8KHz carrier frequency measuring amplifiers
for strain gauge transducers.
All plug-in cards are delivered without housing or rack frame and without
power supply unit so that an individual setting in an existing 19” rack mounting
system is possible. Owing to the narrow width of the plug-in cards (4 divs =
20.32mm) up to 21 cards can be inserted into a 19” rack. The dimensions and
connecting details correspond to the IEC standards 48D and 297.
All controls necessary for easy operation are accessible from the front.
Individual adjustments are described in chapter four; ex-works settings are
specified on the back of the Eurocard.
St 21
Notes for the diagram on the right
S21:
calibration − measuring − zero
S22/23:
zero balancing (coarse)
S24/1+2:
MR1 fine balancing internal/external
S24/3:
switch for measuring frequency
range 60 Hz to 2 Hz
S24/4:
MR changeover from MR1 to
MR2
S25:
alteration to bridge supply voltage
S26/27:
setting of measuring range
(coarse)
P21/22:
fine balancing of MR
P23:
zero balancing (fine)
S26/27S22/23
P23
Cal. 1mV/V
zero
0..250Hz
0..40Hz
S24/3
S21
G
4.8kHz
S25
5V/1V
P21/22
S24/4
S24/1+2
1c
7c
8c
9c
26c
21c
24c
21a
25a
25c
30c
30a
32c
32a
28a
28c
31a+c
6c
7a
15c
4a
5a
15a
16a
19c
18c
19a
20a
11c
5kOhm
50kOhm
wh
rd
gy
bk
bu
gn
chassis
MR1
fine balancing
external
zero balancing
external
zero
calibr.
MR1−2
+14.5...15.5V
−14.5...15.5V
option DC/DC-con-
+
verter
−
Ub +14.5...+15.5V
Ub −14.5...−15.5V
Ua ±10V >5kOhm
option Ia ±20mA
2.33V synchr.-voltage
or synchr. input
Fig. 1.1:Block diagram
HBMA0260-1.1 en
ME50S6
2Electrical Connection
2.1Transducer Connection
Strain gauge transducers are connected by means of 6-wire circuits.