Force transducers of type C18 are to be used for precise force
measurements. Use for any additional purpose shall be deemed to be not in
accordance with the regulations.
In the interests of safety, the transducer should only be operated as described
in the Mounting Instructions. It is also essential to observe the appropriate
legal and safety regulations for the application concerned during use. The
same applies to the use of accessories.
The transducer is not a safety element within the meaning of its use as
intended. Proper and safe operation of this transducer requires proper
transportation, correct storage, assembly and mounting and careful operation
and maintenance.
General dangers of failing to follow the safety instructions
C18 force transducers comply with the state of the art and are fail-safe. The
transducers can give rise to remaining dangers if they are inappropriately
installed and operated by untrained personnel.
Everyone involved with the installation, commissioning, maintenance or repair
of a force transducer must have read and understood the Mounting
Instructions and in particular the technical safety instructions.
Remaining dangers
The scope of supply and performance of the transducer covers only a small
area of force measurement technique. In addition, equipment planners,
installers and operators should plan, implement and respond to the safety
engineering considerations of force measurement technique in such a way as
to minimize remaining dangers. Prevailing regulations must be complied with
at all times. There must be reference to the remaining dangers connected with
force measurement technique.
HBMA0887-1.1 en
Page 5
C18
In these mounting instructions remaining dangers are pointed out using the
following symbols:
5
Symbol:
Meaning:Highest level of danger
Warns of a directly dangerous situation in which failure to comply with safety
requirements will lead to death or serious physical injury.
Symbol:
Meaning:Possibly dangerous situation
Warns of a potentially dangerous situation in which failure to comply with
safety requirements can result in death or serious physical injury.
Symbol:
Meaning:Possibly dangerous situation
Warns of a potentially dangerous situation in which failure to comply with
safety requirements could lead to damage to property, slight or moderate
physical injury.
DANGER
WARNING
CAUTION
Symbols indicating application notes and useful information:
Symbol:
Means that important information about the product or its handling is being
given.
Symbol:
Meaning:CE mark
The CE mark is the manufacturer’s guarantee that his product meets the
requirements of the relevant EC directives (see Declaration of Conformity at
the end of this document).
NOTE
HBMA0887-1.1 en
Page 6
6
C18
Prohibition of own conversions and modifications
The transducer must not be modified from the design or safety engineering
point of view except with our express agreement. Any modification shall
exclude liability on our part for any resultant loss or damage.
Qualified personnel
These transducers are only to be installed by qualified personnel strictly in
accordance with the specifications and with the safety rules and regulations
which follow. It is also essential to observe the appropriate legal and safety
regulations for the application concerned. The same applies to the use of
accessories.
Qualified personnel means persons entrusted with the installation, fitting,
commissioning and operation of the product who possess the appropriate
qualifications for their function.
Conditions on site
Protect the transducer from damp and weather influences such as rain, snow,
etc.
Maintenance
The C18 force transducer is maintenance free.
Accident prevention
Although the specified nominal (rated) force in the destructive range is several
times the full scale value, the relevant accident prevention regulations from
the trade associations must be taken into consideration.
HBMA0887-1.1 en
Page 7
C18
1Scope of supply
•C18 measuring body
•Thrust piece
•Base
•Test report
•Operating Manual C18
Accessories (not included in the scope of supply)
•DKD calibration certificate according to EN10002-3, ISO376
Order no. K-CAL-FD...
7
2Application information
Force transducers of the C18 type series are suitable for measuring
compressive forces.
Do always install the transducer including base and thrust piece to
avoid faulty measurement.
They measure static and quasi-static forces with great accuracy and
reproducibility and therefore require careful handling.
You must take particular care when transporting and installing the devices. If
you knock or drop the transducers, this could permanently damage them.
The force transducers can be used with both DC amplifiers and CF amplifiers
with a maximum carrier frequency of 600 Hz.
C18 force transducer with DKD certificate from HBM:
Additional force calibration according to EN 10002-3 or ISO376, guarantees a
transducer rating to class 0.5.
For a DKD calibration, we recommend to calibrate the complete measuring
chain including the transducer and the connected amplifier to ensure optimum
accuracy.
The limits for the permissible mechanical, thermal and electrical stresses are
stated in the Specifications. It is essential that these are taken into
consideration in planning the measuring set-up, during installation and finally,
during operation.
HBMA0887-1.1 en
Page 8
8
C18
3Structure and mode of operation
3.1Measuring body
The measuring body comprises a measuring beam system made of
rust-resistant steel with strain gages (ring torsion principle).
3.2Base and thrust piece
The base is made from stainless steel for nominal (rated) forces from 10 kN to
500 kN and from hardened tempering steel for nominal (rated) forces from
1 MN to 4.5 MN. They must be oiled to protect them from corrosion. All thrust
pieces are made of stainless steel.
3.3Disturbances
Torsion, bending, and lateral loads are disturbances and must be avoided.
The temperature effects on the zero signal (SG bridge and housing) and on
the sensitivity are compensated.
HBMA0887-1.1 en
Page 9
C18
4Conditions on site
4.1Ambient temperature
The effects of temperature on the zero signal and on the sensitivity are
compensated. To achieve optimal measurement results the nominal (rated)
temperature range must be maintained. Temperature-induced measurement
errors can be caused by heating (for example radiant heat) or cooling on one
side. A radiation shield and all-round heat insulation bring about marked
improvements. They must not form a force shunt.
4.2Moisture
Externed humidity and tropical climatic do not affect the transducer‘s function.
9
4.3Air pressure
Barometric variations have similar effects on the force transducer as force variations. Please take into account that barometric variations result in zero offset.
Nominal (rated) loadkN102050u 50
Zero effect at a barometric variation of 10mbar
(related to the nom. (rated) load)
%0.010.0050.002t0.001
HBMA0887-1.1 en
Page 10
10
C18
5Mechanical installation
5.1Important measures for installation
− treat the transducer gently
− when measuring, make sure that the support structure is rigid
− the force-introduction surfaces must be scrupulously clean and carry in full
− do not overload the transducer.
WARNING
If there is a risk of breakage through overload on the transducer and
thus a risk to persons, additional safety measures are to be taken.
5.2General installation guidelines
The measurement direction in which forces act on the transducer must be as
precise as possible.
WARNING
Torsion and bending moments, eccentric loading and transverse forces
may result in measurement errors and if the limit values are exceeded,
could destroy the transducer.
HBMA0887-1.1 en
Page 11
C18
5.3Installing the transducer
NOTE
Do always install the transducer including base and thrust piece to
avoid faulty measurement (both included in scope of supply).
Thrust
piece
Measuring
body
11
Base
Fig. 5.1:Schematic diagram of the force transducer
C18/10kN...50kN
For measuring ranges from 10kN to 50kN, the base is firmly attached to the
underside of the measuring body by four screws.
Tightening torque: 4Nm
C18/100kN...4.5MN
For measuring ranges from 100kN to 4.5MN, the measuring body is placed on
the base. There is a centering device in the middle of the base.
HBMA0887-1.1 en
Page 12
12
C18
6Electrical connection
6.1Instructions for cabling
•Use only shielded and low-capacitance measurement cables (HBM cables
meet these conditions).
•Do not install measurement cables parallel to power lines or control
circuits. If this is not possible (e.g. in cable ducts) protect the measurement
cable, e.g. with armoured steel tube and maintain a minimum distance of
50 cm from the other cables. High-voltage power lines and control lines
should be twisted (15 turns per meter).
•Avoid stray fields of transformers, motors and contactors.
•Do not earth transducer, amplifier and display device more than once. All
the devices in the measurement chain are to be connected to the same
earthed conductor.
•The force transducers have been designed for four-wire connection, the cable must not be shortened.
•For optimum precision, we recommend to realize the cable extension using
the six-wire technique.
•The screen of the connection cable is connected to the transducer housing.
CAUTION
Under no circumstances must the screwed cable gland of the force
transducer be opened. Should this occur by mistake, the transducer
must be returned to the factory for repair.
HBMA0887-1.1 en
Page 13
C18
13
6.2Wiring pin assignment
The 5m long transducer connection cable has color-coded pigtails.
If the transducer is connected in accordance with the following connection
diagram, when compressive loading is applied to the transducer, the output
voltage at the measuring amplifier is positive.
The transducers are fitted with a four-core connection cable and calibrated as
standard using the four-wire circuit
wh (white)Measuring signal (+)
bk (black)Bridge excitation voltage (−)
rd (red)Measuring signal (−)
bu (blue)Bridge excitation voltage (+)
ye (yellow)Cable screen
Fig. 7.1:Transducer with four-core cable
wh (white)Measuring signal (+)
Sensor circuit
bk (black)Bridge excitation voltage (−)
rd (red)Measuring signal (−)
bu (blue)Bridge excitation voltage (+)
Sensor circuit
ye (yellow)Cable screen
Fig. 7.2:Transducer with four-core cable, amplifier in six wire technique
HBMA0887-1.1 en
Page 14
14
7Specifications
C18
Nominal (rated) forceF
nom
kN10 −
200
300500 −
1000
2000 −
4500
Accuracy class0.05
Nominal (rated) sensitivityC
Rel. sensitivity deviation compressive forced
mV/V2
nom
C
%0.1
Rel. zero signal deviation (zero signal
tolerance)
d
s,o
%1
Rel. zero point compensation (zero signal
return)
1
f
o
%0.0120.024
Rel. range (0.2Fnom to Fnom) at:
unchanged mounting position, typically
different mounting positions, typically
Hysteresis (0.2F
nom
to F
nom
1
)
1
1
Linearity deviationd
b
rg
b
rv
%
%
u%0.08
lin
%0.05
0.04
0.08
Effect of temperature on sensitivity/10 K rel.
to sensitivity
TK
%0.01
c
Effect of temperature on zero signal/10 K rel.
to sensitivity
TK
%0.01
0
Effect of lateral forces (lateral force 10 %
F
nom
)
)*
d
Q
%0.0350.10.15
Eccentricity effect / mm%0.02
Rel. creep over 30 mind
cr,F+E
Input resistanceR
Output resistanceR
Isolation resistanceR
Reference excitation voltageU
Operating range of the excitation voltageB
e
a
is
ref
U,G
%0.03
Ω4450"100
Ω4010"5
Ωu50 x 10
V5
V5 to 30
9
Carrier frequency of the excitation voltageHzv600
Nominal (rated) temperature rangeB
Operating temperature rangeB
Storage temperature rangeB
Reference temperaturet
t,nom
t,G
t,S
ref
o
C+10 to +40
o
C−30 to +80
o
C−50 to +85
o
C+22
Max. operational force(FG)%170150
Limit force(FL)%170150
Breaking force(FB)%400320
Static lateral limit force(FQ)%0.3⋅F
(F
; (to Fz≤ 0.5F
nom
0.5⋅(F
(for F
= Force in direction of
z
−0.8⋅Fz);
nom
> 0.5F
z
nom
nom
)
measurement)
1)
class 0.5 acc. to EN10002−3 or ISO376,
Classification only guaranteed in conjunction with a DKD calibration certificate acc. to EN10002−3
resp. ISO376
2)
relative to a force introduction point on the force-introduction surface of the measuring body
)
HBMA0887-1.1 en
Page 15
C18
15
Nominal
(rated)
force
Nominal
(rated)
displace
ment
Weightkg1.21.21.22.32.33.910.4 15.345.652.690.4
Rel.
permissible
vibrational
stress
Degree of protection
acc. to DIN EN60529
Cable length, four-wire
connection
Measuring body
material
FnomkN102050100200300500100020003000450
0
Snom mm 0.13 0.12 0.13 0.17 0.19 0.23 0.26 0.450.620.790.98
Frb%70
IP68
5
stainless steel
HBMA0887-1.1 en
Page 16
16
8Dimensions C18
10 kN − 50 kN
C18
∅A
Thrust piece
4 x screws M5 x 12
100 kN − 4.5 MN
D
D
approx. 15
approx. 15
∅C
∅B
∅A
∅C
Measuring
body
H
G
Base
H
G
∅25
∅B
TypejAjBjCDGH
C18 / 10 kN ... 50kN357060721543
C18 / 100 kN ... 200 kN457575892050
C18 / 300 kN5895951122065
C18 / 500 kN851301301573785
C18 / 1 MN1001501501714090
C18 / 2 MN13523022523950130
C18 / 3 MN13523022525450145
C18 / 4.5 MN16027527030360170
HBMA0887-1.1 en
Page 17
C18
9Declaration of conformity
17
HBMA0887-1.1 en
Page 18
18
C18
HBMA0887-1.1 en
Page 19
HBM
Page 20
pp
A0887−1.1 en
Modifications reserved.
All details describe our products in general form only.They are
not to be understood as express warranty and do not constitute
any liability whatsoever.
Hottinger Baldwin Messtechnik GmbH
Postfach 10 01 51, D-64201 Darmstadt
Im Tiefen See 45, D-64293 Darmstadt
Tel.: +49/61 51/ 8 03-0; Fax: +49/61 51/ 8039100
E−mail: su
ort@hbm.comwww.hbm.com
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.