Hbm AED9101C User Manual

Page 1
User Manual
AED9101C
Digital Transducer Electronics Basic Device for AD103C
with RS232, RS422, RS485
I1692-2.0 en
Page 2
Page 3
Content
Typographical conventions................................................................................................3
Important information................................................................................................ ......... 4
Safety instructions.............................................................................................................. 5
1 Introduction and appropriate use......................................................................................6
2 Mechanical construction....................................................................................................7
3 Electrical connections........................................................................................................8
1
3.1 Transducer connection......................................................................................................................... 8
3.2 Connecting the supply voltage.............................................................................................................15
3.3 Connection to a computer ...................................................................................................................16
3.4 Multi-channel measurements (bus mode, RS485) ...............................................................................19
3.5 Connecting the diagnostic bus.............................................................................................................22
3.6 Trigger input........................................................................................................................................24
3.7 AED9101C cable connection via a PG gland.......................................................................................25
Index...................................................................................................................................26
AED9101C en HBM
Page 4
2
HBM AED9101C en
Page 5

Typographical conventions

Typographical conventions
For clear identification and improved legibility, the following conventions have been used in this documentation:
Important paragraphs are marked with a symbol to draw attention to them.
CE Designation
Statutory marking requirements for waste disposal
3
Italics
“File Open“ All menus and menu commands appear in quotes, here the “File” menu and the “Open”
“Start”
MSV
Points out external documents and files
sub-menu. Quotes and italics are used for buttons, input fields and user input. All commands are set out in a bold font style or as a link to the command description.
AED9101C en HBM
Page 6
4

Important information

Neither the design of the device nor any technical safety aspects may be modified without the express permission of Hottinger Baldwin Messtechnik GmbH. Any modification ex­cludes Hottinger Baldwin Messtechnik GmbH from any and all liability for any damage re­sulting therefrom.
It is strictly forbidden to carry out any repairs and soldering work on the motherboards or to replace any components. Repairs may only be carried out by persons authorized thereto by Hottinger Baldwin Messtechnik GmbH.
All the factory defaults are stored safe from power failure at the factory, not in the measur­ing amplifier where they can be deleted or overwritten. They can be reset at any time by using the command TDD0. For more information, see aed_help_e, AD103C; “Description of the basic commands”.
The production number is set at the factory and cannot be changed.
The transducer connection must always be assigned.
It is essential for a transducer or a bridge model to be connected up for operation.
Important information
HBM AED9101C en
Page 7
Safety in
structions

Safety instructions

 There are not normally any hazards associated with the product, provided the notes and
instructions for project planning, assembly, appropriate operation and maintenance are observed.
 Each time, before starting up the modules, you must first run a project planning and risk
analysis that takes into account all the safety aspects of automation technology. This particularly concerns personal and machine protection.
 It is essential to comply with the safety and accident prevention regulations applicable to
each individual case.
 Installation and start-up must only be carried out by suitably qualified personnel.  Do not allow the equipment to become dirty or damp.
5
 During installation and when connecting the cables, take action to prevent electrostatic
discharge as this may damage the electronics.
 The required power supply is an extra-low voltage with safe disconnection from the
mains.
 When connecting additional devices, comply with the local safety requirements.  All the interconnecting cables must be shielded cables. The screen must be connected
extensively to ground on both sides. The power supply and digital I/O connection cables only need to be shielded if the ca­bles are longer than 30 m (32.81 yd) or are routed outside closed buildings (EN 61326-1).
 The CE mark enables the manufacturer to guarantee that the product complies with the
requirements of the relevant EC directives (the declaration of conformity is available at http://www.hbm.com/HBMdoc).
 In accordance with national and local environmental protection and material recovery
and recycling regulations, old devices that can no longer be used must be disposed of separately and not with normal household garbage. If you need more information about waste disposal, please contact your local authorities or the dealer from whom you purchased the product.
AED9101C en HBM
Page 8
6

1 Introduction and appropriate use

AED9101C digital transducer electronics are part of the AED component family that digitally conditions signals from mechanical measurement sensors and networks them with bus ca­pability. These include digital amplifier motherboards, basic devices with serial interfaces and intelligent sensors with integrated signal processing. The purpose of these components is to directly digitize and condition the measurement signals at the transducer location. Using digital transducer electronics, you can connect SG1)transducers in a full-bridge circuit di­rectly to a computer or a PC. This enables you to configure complete measurement chains quickly and with little extra work.
Basic device AED9101C can contain the AD103C amplifier board. It provides mechanical protection, shields the amplifier board (EMC protection) and allows you to select the serial interfaces RS422 (factory default), RS485 or RS232.
The AD103C amplifier mother board is not included in the scope of supply of the basic de­vice and must be ordered separately.
Introduction and appropriate use
The AD101B amplifier is only required for replacement and will not be considered when new AED basic devices are ordered.
The PC software AED PANEL 32 is available to facilitate parameter settings, to display dy­namic measurement signals and for comprehensive analysis of the dynamic system. The HBM display unit DWS2103 can be connected to all AED basic devices.
All basic devices of the AED family can be connected with the digital display unit DWS2103. This unit supports all implemented functions of the AED.
All commands are described in the help file aed_help_e.
The abbreviation AED is also used for transducer electronics in the following text.
1)
Strain Gage
HBM AED9101C en
Page 9
Mechanical
construction

2 Mechanical construction

The basic device extends the functionality of the AD amplifier boards and provides:
 mechanical protection (IP65)  the power supply for the amplifier motherboard and transducer excitation  total transducer bridge resistance 40…4000   a choice of serial interfaces RS422, RS485, RS232  EMC-tested  Diagnostic bus
(The AED9101A basic device can be replaced by the AED9101C basic device)
7
The amplifier motherboard is designed as a plug-in board that can be plugged into the carrier board of the basic device via a 25-pin D-connector. The basic device contains terminals for the transducer, power pack and interface connections, slide switches for interface selection and the voltage stabilizer. The connection cables exit the casing via PG glands on the side.
Fig. 2-1: Mechanical construction
AED9101C en HBM
Page 10
8

3 Electrical connections

A connection diagram is attached inside the lid of the AED9101C basic device.
When making the connections, please ensure that the wires of the cable do not protrude beyond the connection terminals (risk that loops may form). Please make sure that the cable shielding is properly connected to the PG gland.
If it should be necessary, a separate cable can be used to establish potential equalization between the transducer and the AED and between the AED and the Master control unit (grounding concept). The cable shielding must not be used for this potential equalization.
Electrical connections

3.1 Transducer connection

The transducer connection must always be assigned (connect the transducer).
AED9101C with AD103C
You can connect SG transducers in a full-bridge circuit with a total bridge resistance of RB= 40...4000 . With a transducer resistance of > 1000 , increased noise (measurement
ripple) must be taken into account. The bridges are supplied with power in the AED9101C basic device (5 VDC).
HBM AED9101C en
Page 11
Transducer
connection
9
Fig. 3.1-1: Transducer connection in 6-wire circuitry (HBM color-coding)
The 6-wire connection avoids the effect of a long cable on the measured value. When sev­eral transducers and a junction box are used, the 6-wire circuitry is routed to the junction box.
Fig. 3.1-2: Transducer connection in the AED9101C basic device for a 6-wire connection
AED9101C en HBM
Page 12
10
transducer electronics
Cable shield
Transducer connection
There are two methods of connection for transducers implemented in four-wire circuitry:  Connection via a 6-core extension cable; bridged sensor circuit in the transducer con-
nector.
Plug-in connection
Transducer
Fig. 3.1-3: Transducer connection in 6-wire circuitry via a 6-core cable extension
bu rd wh
bk
bu gn
wh gy
bk
Cable to
HBM AED9101C en
Page 13
Transducer
connection
11
 Connection without an extension cable; sensor circuit bridged at the transducer electron-
ics.
Fig. 3.1-4: Transducer connection in 4-wire circuitry without a cable extension (jumpers 2 – 2‘
and 3 – 3‘)
Connection without an extension cable; sensor circuit at transducer electronics. When con­necting several transducers, it is advisable to use an HBM junction box VKKx. In general, the feed lines running to the AED should be shielded cables.
When connecting several transducers to the AED, the number of load cells that can be con­nected (and the resultant bridge resistance) must be taken into consideration with regard to the external supply voltage, so that the maximum power loss in the basic device is not ex­ceeded.
AED9101C en HBM
Page 14
12
U in (V)
30 24
21 16
13
Transducer connection
O O
23 C [73.4 F]
O O
70 C [158 F]
Impedance ( )
350 175 87 58
1 2 3 4 5 6 7 8
Fig. 3.1-5: Maximum operating voltage for the AED9101C basic device, with regard to the number of
transducers and the ambient temperature
43
Number of transducers
Notes on type of connection, length and cross-section of cables:
Depending on the bridge resistance of the load cell being used and the length and cross­section of the load cell connection cable, there may be voltage drops that can reduce the bridge excitation voltage. The voltage drop at the connection cable is also dependent on temperature (copper resistance). Likewise, the output signal of the load cell changes in pro­portion to the bridge excitation voltage.
This is balanced out when connecting in 6-wire circuitry.
6-wire circuit (standard mode of operation):
This will correct all the effects of the load cell cabling up to the feedback points. Even chang­ing the length of a cable after calibration will not make any difference to the measurement results.
For load cells with a 6-wire connection, feedback lines 2´ and 3´ are bridged in the load cell with excitation 2 and 3 (Fig. 3.1-2). For load cells with a 4-wire connection, the feedback bridges must be implemented directly at the load cell connection (Fig 3.1-3 or 3.1-4).
HBM AED9101C en
Page 15
Transducer
connection
13
4-wire circuit:
As correction through AUTOCAL can only ever take place up to feedback points 2´, 3´, all the changes of cable resistances affect the measurement result. This means that even if no further changes are made to the 4-wire cable used for calibration, there will still be meas­urement errors when there are temperature changes, because the cable resistance and pos­sibly the contact resistances at the connectors are temperature-dependent. With the 4-wire circuit, feedback lines 2´ and 3´ are directly connected at connection terminals 2 and 3 in the AED (see Fig. 3.1-4).
Equivalent circuit of the bridge with bridge resistance RBand supply lines with line resistances R
and RL2:
L1
R = R = (4 p / ) (I [m] / A [mm ])
R
L1
R
B
R
U
L2
BR
L1 L2 CU
p = 0,0178 [ mm /m] for copper
CU
= 3,14 I = Length of cable, A = Cross-section of cable
R = R = 1,6 at I = 10 m and A = 0,14 mm
Fig. 3.1-6: Bridge equivalent circuit diagram
L1 L2
2
2
2
AED9101C en HBM
Page 16
14
Transducer connection
The voltage drop over the bridge feeder cables can be determined from bridge resistance RB, cable length l, cable cross-section A and the bridge excitation voltage:
UB+ U
RL1
+ U
RL2
= U
BR
For RB= 80 , R
= RL2= 1,6 (l = 10 m) and UBR= 5 V
L1
there is an excitation current of IBR= UBR/(RL1+ RL2+ RB) = 60 mA
and thus a voltage drop over the two line resistances totaling approx. 0.2V (U
Bridge
= 4.8 V).
For RB= 80 , R
= RL2= 16 (l = 100 m) and UBR= 5 V
L1
there is an excitation current of IBR= UBR/(R
+ RL2+ RB) = 45 mA
L1
and thus a voltage drop over the two line resistances totaling approx. 1.4 V (U
= 3.6 V = 80 0.045).
Bridge
This is irrelevant for the 6-wire circuit, as the voltage drop over the sensor lines is taken into account in the measurement signal.
But with a 4-wire circuit, the dependency of the copper resistance of the cables on tempera­ture goes directly into the measurement result, as the bridge excitation voltage U
Bridge
changes: RL(t) = RL20 (1 +   (t –20 °C)), where R
is the line resistance at 20 °C and  is the temperature coefficient of the cop-
L20
per. R
– for calculation, see page 23, CU: = 0.00392 [1/K]
L20
With a cable length of l = 100m and a temperature differential of 10 °C, there is a line resis­tance of
RL1(t) = RL2(t) = 16 (1 + 0.00392 10) = 16.6 This changes the bridge excitation voltage of U
= 3.6 V (at 20 °C) to U
Bridge
Bridge
= 3.53 V.
This change in bridge excitation voltage directly at the transducer changes the measurement signal of the bridge by 1.9 % (= 100 % (1 – 3.53 V / 3.6 V)).
This typical calculation shows that if long cables are involved, only 6-wire circuitry should be used.
HBM AED9101C en
Page 17
Connecting
the supply voltage

3.2 Connecting the supply voltage

The power supply must meet the following requirements:
15
AED9101C with AD103C
Calculation the total current consumption:
DC voltage +6...+30 V current consumption
250 mA (for a 40 bridge)
Fig. 3.2-1: Power supply connection
The voltage feed must be shielded. It can be applied within the interface cable or be imple­mented as a separate cable.
When supplying several AEDs via one cable, the voltage drop over the cable must be taken into consideration. The voltage drop depends on the supply current required and on the line resistance.
AED9101C en HBM
Page 18
16

3.3 Connection to a computer

The basic device can be set up for several interface variants:
Connection to a computer
Fig. 3.3-1: Pin assignment and settings for the interfaces
HBM AED9101C en
Page 19
Connection
to a computer
17
No bus mode
The RS232 interface allows the AED to be connected directly to a PC. The cable length is limited to 15 m and bus mode is not possible.
Fig. 3.3-2: Connecting an AED to a computer via the RS232 interface
For the RS232 interface setting, the “bus termination switch must be set to OFF”.
The RS422 interface is a 4-wire interface also with a maximum cable length of 1000 m. To connect the AED to the COM port of a PC (RS232), you need an interface converter. This in­terface is not designed for bus mode, as the transmitter is always active.
AED9101C en HBM
Page 20
18
RS 485-Bus AED 9101C
Connection to a computer
Bus mode
The RS485 interface with a 4-wire connection allows the full range of AED functions with a maximum cable length of 1000 m. To connect the AED to the COM port of a PC (RS232), you need an interface converter.The RS485 interface with a 2-wire connection also allows cable lengths of 1000 m. To connect the AED to the COM port of a PC (RS232), you need an interface converter (1-SC232/422B from HBM, with 2-wire mode selected). However, the AED must be operated in half-duplex mode (command COFxxx must be set). The command MSV?0; must not be used, as it is not possible to interrupt data output with the STP; command.
Basically, shielded cables should be used for the interface wiring, with the cable shielding being connected to the AED housing via the PG (see AED9101C cable connection via the PG gland). The power supply can also be connected via this cable, with a 6-core, shielded cable being necessary.
If it should be necessary, a separate cable can be used to establish potential equalization between the bus nodes. The cable shielding must not be used for this potential equalization. For reasons of electromagnetic compatibility, it is advisable to use a double-shielded cable (from the HBM program, for example: 3 2 0.14 m², 4-3301.0071).
The shielding of the physical circuit is connected to the AED line (not to the power supply ground).
U : 8 V...30 V
B DC
Interface converter
PC COM port
RxD TxD GND
SUB-D connector 9-pin (PC) PIN 2 = RxD 3 = TxD 5 = GND
TxD
RxD
GND
RS232 - RS485
HBM Wägetechnik GmbH
CONVERTER
RS232
SUB-D female connector 9-pin (converter) PIN 2 = TxD 3 = RxD 5 = GND
SC232/422A
8...30V
RS485
2/4-wire
TA TB RA RB GND U
B
Equivalent Interface marking:
TA = T- TB = T+ RA = R- RB = R+
TA TB RA RB GND U
B
(blue) (black) (green) (grey) (white) (red, 8...30 V )
DC
HBM AED9101C en
Fig. 3.3-3: Connecting an AED (RS422 or RS485 4-wire) to a computer via the interface converter
Page 21
Mult
i-channel measurements (bus mode, RS485)

3.4 Multi-channel measurements (bus mode, RS485)

With the RS485 interface, several AEDs can be connected to a common bus line. The bus cabling for 4-wire mode is shown in Fig. 3.4-1.
With the RS485 interface, up to 32 AEDs can be connected to a common bus line. With the aid of the RS485 bus driver, it is possible to implement long cables (up to 1000 m (1,093.61 yd) in length).
AED bus mode is set out as a Master-Slave configuration, with the AED implementing a slave.
This means that all the AED activities are initiated from the control computer. Each AED is given a separate communication address (00...89) and can then be activated by an Sii (ii = 00...89) Select command.
A broadcast command (S98) is implemented for specific communication situations. This means that after a command of this type, all the AEDs execute the command of the Master, but none of the AEDs respond. All these communication commands are described with rele­vant examples in file aed_help_e, AD103C; “Description of the basic commands”.
19
Fig. 3.4-1: Bus structure 4-wire bus (general)
AED9101C en HBM
Page 22
20
Multi-channel measurements (bus mode, RS485)
The termination resistances of 500 marked in the above diagram are important for the electrical function of the bus system. These resistances safeguard the quiescent level for the receiver on the bus line. The physical circuit must only be connected with these resistances at the line ends.
For the local bus termination distribution shown in the diagram, the Master and the AED with address 89 should include the termination resistances. Which is why in this AED, the bus termination is activated using the “bus termination ON” switch. This bus termination in the AD103C must also be activated using the commands STR1; and TDD1 (see file aed_help_e, AD103C; “Description of the basic commands”). If the “bus termination switch is set to OFF”, the command STR will have no effect, which means that bus termination is deactivated.
The HBM interface converter also includes these bus termination resistances (do not deacti­vate them).
These terminations must not be activated more than twice in one bus. The ground of the interface driver is related to the GND terminal. The interface driver of
the master should be also connected to this GND.
The quiescent level on the RS485 physical circuit is produced in 4-wire mode at:
TB - TA > 0.35 V RB - RA > 0.35 V
(quiescent level though AED termination resistances) (quiescent level though Master termination resistances)
As the RS485 is a differential bus interface, the quiescent level is also specified as a differ­ential voltage between the lines (and not ground-related). Furthermore, please note that this interface tolerates a maximum common-mode range of ±7 V. If it should be necessary, a separate cable can be used to establish potential equalization between the bus nodes. The cable shielding must not be used for this potential equalization.
HBM AED9101C en
Page 23
Mult
i-channel measurements (bus mode, RS485)
RS485 2-wire mode
21
Fig. 3.4-2: Bus structure RS485 2-wire bus (general, TB/RB = T/R+, TA/RA = T/R-)
Termination resistances are also necessary for the electrical function of this bus system. These resistances safeguard the quiescent level for the receiver on the bus line. These ter­mination resistances are already included in the AED and should be set with the “bus termi- nation switch” (see 4-wire circuit). The quiescent level on the RS485 physical circuit is pro­duced in 2-wire mode at:
TB/RB - TA/RA 0.35 V
For the local bus termination distribution shown in the diagram, the Master and the AED with address 89 should include the termination resistances. Which is why in this AED, the bus termination is activated using the “bus termination ON” switch. This bus termination in the AD103C must also be activated using the commands STR1; and TDD1 (see file aed_help_e, AD103C; “Description of the basic commands”). If the “bus termination switch is set to OFF”, the command STR will have no effect, which means that bus termination is deactivated.
AED9101C en HBM
Page 24
22

3.5 Connecting the diagnostic bus

The diagnostic bus is used to analyze dynamic processes. The bus is set out as an RS485 2-wire bus (lines: TB/RB and TA/RA, GND).
Connecting the diagnostic bus
Fig. 3.5-1: Connecting the diagnostic bus via terminal KL4
The interfaces setting of the bus is defined and cannot be changed (38400 bit/s, 8E1). External bus termination resistances are not necessary for this bus. The HBM interface converter (1-SC232/485B) can be used to connect the RS485 bus to an
(RS232) COM port of the PC.
HBM AED9101C en
Page 25
Connecting
the diagnostic bus
23
Fig. 3.5-2: Diagnostic RS485 bus
The ground of the interface driver is related to the GNDext terminal. The interface driver of the master should be also connected to this GNDext.
Only a connecting cable with a screen grounded on two sides should be used as the inter­connecting cable between the AED 9101C and the bus and the master (see also: AED9101C cable connection via a PG gland)
The functions and commands of the diagnostic channel are described in the help file aed_help_e Diagnosis. The address corresponds to the address of the AD103C amplifier, command ADR (00...89, factory setting: 31), see aed_help_e, Basic Commands). This ad­dress is independently from the CANOpen address.
The following functions can also be executed via this bus:
Parameters
Read only (changes are not possible)
Measured values Reading individual measured values MSV?; (MSV?i not possible) Results
Trigger results and dosing results can be read
The diagnostic functions can be executed using the HBM AED_Panel32 program (as from Version V3.0.0).
The HBM display unit DWS2103 can be connected with this interface. Than all implemented functions and parameters are accessible. This is independent from the main communication channel.
AED9101C en HBM
Page 26
24

3.6 Trigger input

An external sensor (light barrier, contact, etc.), can be connected to the trigger input of the AD103C, to drive the trigger measurement function (see file aed_help_e, AD103C; “Descrip­tion of the signal processing”).
The input is activated as an external trigger by the TRC command.
Trigger input
Fig. 3.6-1: Trigger input
The trigger input has the following properties: Quiescent level: Low Active edge: High – Low High level: 2...30 V Low level: 0...1 V Input current: If the input is not required, the input remains unassigned. The GND of the trigger input is
connected to the GND of the supply voltage.
3 mA (for 30 V), 10 kinput resistance
HBM AED9101C en
Page 27
AED9101C
cable connection via a PG gland
Slidecablescrew
connectionwithsealing
ringandclampingring
s
overtheendofthecable.
3.7 AED9101C cable connection via a PG gland
Only a connecting cable with a screen grounded on both sides (and metal connectors) should be used as the connecting cable between the AED9101C and its partner device. Bring the screen extensively into contact on both sides at the PG gland (and at the metal shell of the connector). If the partner device does not have a metal connector, connect the cable shielding extensively to ground. If there are vast differences between the ground po­tential of the AED9101C and its partner device, a potential equalization line must be pro­vided in addition.
Remove outer sheathofcable to expose required length wire L.
25
Fig. 3.7-1: Connecting the transducer, supply voltage and computer to the PG gland
AED9101C en HBM
Page 28
26

Index

Index
4
4-wire circuitry.................................................................................................................................................10, 11
6
6-wire connection..............................................................................................................................................9, 12
B
basic box ............................................................................................................................................................... 7
bridge excitation voltage .................................................................................................................................12, 14
bridge resistance ...................................................................................................................................... 12, 13, 14
bus
bus line ............................................................................................................................................................ 19
bus structure ...............................................................................................................................................19, 21
bus mode........................................................................................................................................................16, 19
RS232.............................................................................................................................................................. 17
RS485.......................................................................................................................................16, 18, 19, 20, 21
bus termination...............................................................................................................................................20, 21
bus termination RS485..................................................................................................................................... 19
C
cable connection.................................................................................................................................................. 17
cable connection - serial interface..................................................................................................................... 16
cable connection supply voltage
cable connection AED9101C.......................................................................................................................15, 25
cable length....................................................................................................................................................12, 14
computer connection............................................................................................................................................ 16
connection to a computer, AED9101C.............................................................................................................. 16
connecting the diagnostic bus.............................................................................................................................. 22
D
diagnostic bus...................................................................................................................................................... 22
E
electrical connections............................................................................................................................................. 8
I
interface............................................................................................................................................................... 16
interface converter..........................................................................................................................................16, 18
HBM AED9101C en
Page 29
Index
27
L
load cell connection ............................................................................................................................................. 12
M
manufacturer’s notes ............................................................................................................................................. 6
mechanical construction......................................................................................................................................... 7
multi-channel measurements............................................................................................................................... 19
P
PG gland ....................................................................................................................................................8, 16, 25
potential equalization........................................................................................................................................... 16
Q
quiescent level
RS485.............................................................................................................................................................. 19
R
RS232 ..........................................................................................................................................................6, 7, 16
RS232 serial interface.......................................................................................................................................... 16
RS422 ..........................................................................................................................................................6, 7, 16
RS485 with 2-wire................................ ................................................................................................................ 16
RS485 with 4-wire................................ ................................................................................................................ 16
S
serial interface................................................................................................................................................... 6, 7
RS232............................................................................................................................................................ 6, 7
RS422.......................................................................................................................................................6, 7, 16
RS485.............................................................................................................................................................. 16
supply voltage...................................................................................................................................................... 15
supply voltage AED9101C................................................................................................................................ 25
T
trigger input.......................................................................................................................................................... 24
U
use........................................................................................................................................................................ 6
AED9101C en HBM
Page 30
HBM AED9101C en
Page 31
AED9101C en HBM
Page 32
Modifications reserved. All details describe our products in general form only. They are not to be understood as express warranty and do not constitute any liability whatsoever.
I1692-2.0 en
Hottinger Baldwin Messtechnik GmbH
Postfach 100151 D-64201 Darmstadt Im Tiefen See 45 D-64293 Darmstadt Tel.: +49/6151/803-0 Fax: +49/6151/8039100 E-mail: support@hbm.com · www.hbm.com
Loading...