HBM AE101, AE301, AE501 Operating Manual

Page 1
Operating manual
Clip Electronic
Amplifier for installation onto mounting rails
Clip IG
Industrial amplifier in cast housing
A0114-6.3 en
Page 2
Clip

Contents Page

Safety instructions 4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1 Application 7 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2 Mounting / Dismounting 8 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3 Connection 9 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.1 Measuring amplifiers AE101, AE301, AE501 10 . . . . . . . . . . . . . . .
3.2 GR201, EM201, EM201K2, TS101additional units 12 . . . . . . . . . . . .
3.3 NT101A, NT102A Power supply 14 . . . . . . . . . . . . . . . . . . . . . . . . . .
4 Setup 15 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.1 AE101 Measuring amplifiers 15 . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.2 AE301 Measuring amplifier 18 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.3 AE501 Measuring amplifier 22 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.4 GR201 Double limit-value switch 25 . . . . . . . . . . . . . . . . . . . . . . . . . .
4.5 EM201 / EM201K2 Output stage 28 . . . . . . . . . . . . . . . . . . . . . . . . . .
5 TS101 Automatic tare and store unit 30 . . . . . . . . . . . . . . . . . . . . . . . . . .
5.1 General 30 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.2 Function 30 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.2.1 Taring 30 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.2.2 Store unit 31 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.3 Connection 33 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.3.1 Voltage supply 34 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.3.2 Inputs / Outputs 34 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.3.3 Control inputs / Control output 34 . . . . . . . . . . . . . . . . . . . .
5.4 Adjustment 36 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.4.1 Factory setup 36 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.4.2 Tare unit adjustment 36 . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.4.3 Store unit adjustment 38 . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.5 Zero-point balance 40 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.6 Component position diagram 41 . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6 Safety barriers 42 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7 CLIP IG 42 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8 Dimensions 43 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
9 Specifications 45 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
10 Accessories 51 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
11 Copy of Declaration of Conformity 52 . . . . . . . . . . . . . . . . . . . . . . . . . . . .
HBMA0114-6.3 en
Page 3
Clip
Safety instructions
The NT101A and NT102A Power Supplies Conform to Protection Class I. The
other Clip components correspond to Protection Class III (Symbol
III
)
when they are operated with safety extra-low voltage (SELV circuits). In order to ensure sufficient immunity from disturbance only use Greenline
screening (see HBM‘s special publication ”Greenline” Screening Concept, electromagnetic compatibility of measuring cable, G36.35.0)
Appropriate use
The Clip Electronic with the connected transducers may be used for measure­ment and directly related control and regulation tasks, only. Any other use is not appropriate.
To ensure safe operation, the Clip Electronic may only be used according to the specifications given in this manual. When using the transducer, the legal and safety regulations for the respective application must also be observed. The same applies if accessories are used.
General dangers in the case of non-observance of the safety instructions
The Clip Electronic complies with the state of the art and is operationally reli­able. If the device is used and operated inappropriately by untrained person­nel, residual dangers might develop.
Any person charged with device installation, operation, maintenance or repair must in any case have read and understood the operating manual and the safety instructions, in particular.
Conditions on site
Protect the device from moisture or atmospheric influences such as rain, snow, etc.
Maintenance and cleaning
The Clip Electronic are maintenance-free. Please note the following points when cleaning the housing:
Remove the mains plug from the socket before cleaning.
Clean the housing with a soft, slightly damp (not wet!) cloth. Never use sol-
vents, since they may damage the labelling on the front panel.
When cleaning, please ensure that no liquid finds its way into the device or onto the contacts.
HBM A0114-6.3 en
Page 4
Clip
Residual dangers
The Clip Electronic scope of performance and supply covers part of the mea­suring-technology, only. The plant designer/constructor/operator must in addi­tion design, realize and take responsibility for the measuring-system’s safety such that potential residual dangers are minimized. The respective regulations must in any case be observed. Residual dangers regarding the measuring system must be specified explicitly.
If there is any risk of remaining dangers when working with the, it is pointed out in this introduction by means of the following symbols:
Symbol:
Meaning: Maximum danger level
Warns of an imminently dangerous situation in which failure to comply with safety requirements will result in death or serious physical injury.
Symbol:
Meaning: Dangerous situation
Warns of a potentially dangerous situation in which failure to comply with safety requirements can result in death or serious physical injury.
Symbol:
Meaning: Potentially dangerous situation
Warns of a potentially dangerous situation in which failure to comply with safety requirements could result in damage to property or some form of physical injury.
DANGER
WARNING
CAUTION
HBMA0114-6.3 en
Page 5
Symbols for using advices and helpful information:
Clip
Symbol: Means that important information about the product or its handling is being
given.
Symbol:
Meaning: CE mark
The CE mark enables the manufacturer to guarantee that the product com­plies with the requirements of the relevant EC directives (see Declaration of Conformity at the end of this document).
Safe operation
Do only quit error messages if the reason for the error has been eliminated and there is no more danger.
Reconstruction and modifications
NOTE
HBM’s express consent is required for modifications regarding the Clip Elec­tronic construction and safety. HBM does not take responsibility for damage resulting from unauthorized modifications.
In particular, repair and soldering works on the boards are prohibited. If com­plete componentry is replaced use original HBM components, only.
Qualified personnel
The device may be used by qualified personnel, only; the technical data and the special safety regulations must in any case be observed. When using the device, the legal and safety regulations for the respective application must also be observed. The same applies if accessories are used.
Qualified personnel means: personnel familiar with the installation, mounting, start-up and operation of the product, and trained according to their job.
Maintenance and repair work on an open device with the power on should only be undertaken by trained personnel who are aware of the above-men­tioned dangers.
HBM A0114-6.3 en
Page 6
Clip

1 Application

The measuring amplifiers are suitable for the measurement of mechanical quantities such as force, weight, torque, pressure, displacement, strain and acceleration. Appropriate transducers complying with the international stan­dards can be connected to the measuring amplifiers.
Transducers can be operated in potentially explosive areas, if safety barriers are used.
The individual CLIP components can be arranged as desired and are wired by terminals. Switches and potentiometers for adjustment have been provided on the circuit boards. The accuracy class is 0.1.
The following units are available:
AE101 (DC), AE301/301S6/301S7 (600 Hz CF), AE501 (4.8 kHz CF), Measuring amplifier
GR201 Double limit-value switch
EM201, EM201K2 Output-stage modules with current output
NT101A, NT102A Power supply units
TS101 Automatic Tare and Store Unit
HBMA0114-6.3 en
Page 7
Clip

2 Mounting / Dismounting

The housings are installed onto mounting rails to DIN EN 50 022; they are hooked onto the upper edge and snapped into the spring plate on the lower edge.
For dismounting, use a screwdriver to push down the spring plate and unhook the housing.
Mounting Dismounting
Hook in
Mounting rail
Lock
Fig. 2.1: Mounting / Dismounting
HBM A0114-6.3 en
Page 8
Clip

3 Connection

Use the 12-pin terminal strip to connect the transducers, to wire the modules with each other and to connect the voltage supply. Individual wires can be
2
clamped in a range of 0.13 mm connected to one terminal, e.g. with internal and external connections, the conductor cross-sections must be matched accordingly. End sleeves (without plastics hoop, length: 10 mm) should be used to connect the cores to the ter­minals. Cores with or without end sleeves must not be tin-plated. When con­necting the lines, measures should be taken to prevent electrostatic dis­charge.
The subsequent figures give the respective connection diagrams that are printed onto the housing cover.
... 1.5 mm2. If two conductor lines are to be
NOTE
The Clip modules have been designed for installation in closed metallic housings (e. g. control cabinet); they can also be operated without any additional housings.
The transducer connection lines and the analog signal lines (inputs and outputs) have to be screened.
At the control cabinet, the screening must lie on a screen rail. Supply lines and lines connected to the GR201 relay contacts must be
screened if the cable length exceeds 30 m or if the cables are routed out­side closed buildings.
NOTE
Only after tightening the clamping screws (ensure good contact) will the output signal be available on the terminals.
HBMA0114-6.3 en
Page 9
10

3.1 Measuring amplifiers AE101, AE301, AE501

Clip
11 11 11 11 11 11 11 11 11 11 11 11
122’33’4589101112
11 11 11 11 11 11 11 11 11 11 11 11
122’33’4589101112
122’33’4589101112
11 11 11 11 11 11 11 11 11 11 11 11
Fig. 3.1: Measuring-amplifier connections
HBM A0114-6.3 en
Page 10
Clip
Type AE101, AE301, AE501
Terminal Function Color ( HBM - cable)
1 Measuring signal WH (white)
2 Bridge excitation voltage BK (black)
2’ Sensor line GY (grey)
3 Bridge excitation voltage BU (blue)
3’ Sensor line GN (green)
4 Measuring signal RD (red)
5 Screen/Ground YE (yellow)
8 Synchronization (not with AE101)
9 Operating-voltage zero
10 Output voltage
11 Supply voltage zero
12 Supply voltage
Tab. 3.1: Amplifier connection
*)
*)
11
*)
Operating-voltage zero and supply-voltage are internally connected.
Terminal 5 of the Clip amplifier and the transducer’s screen connection should be grounded (e. g. connected to the screen rail). If the transducers are installed in a metallic housing with PG glands (e. g. control cabinet or ClipIG), a large area of the transducer screen must be in contact with the PG gland.
HBMA0114-6.3 en
Page 11
12

3.2 GR201, EM201, EM201K2, TS101 additional units

Clip
11 11 11 11 11 11 11 11 11 11 11 11
123456789101112
11 11 11 11 11 11 11 11 11 11 11 11
123456789101112
123456789101112
11 11 11 11 11 11 11 11 11 11 11 11
Fig. 3.2: Connection of the additional units
HBM A0114-6.3 en
Page 12
Clip
Type GR201 EM201, EM201K2 TS101
Terminal Function
1 Relay 1; Break contact Input 1; Ground*) Input; Ground
2 Relay 1; Make contact Input 1; "10 V Input ; "10 V
3 Relay 1; Center Output 1; Ground*) Output; Ground
13
4 Input 1; Ground Output 1;
"20 mA / 4 ... 20 mA
5 Input 1; "10 V Input 2; Ground*) (only with
EM201K2)
6 Relay 2; Break contact Input 2; "10 V
(only with EM201K2)
7 Relay 2; Make contact Output 2; Ground*) (only with
EM201K2)
8 Relay 2; Center Output 2;
"20 mA / 4 ... 20 mA
(only with EM201K2)
9 Input 2; Ground Ground
10 Input 2; "10 V Ground
11 Supply zero Supply zero
12 Supply voltage Supply voltage Supply voltage
Tab. 3.2: Connection of additional units
Output; PVCS "10 V
Output; TAR "10 V
Output; TAR (24 V DC)
Control; Ground
Control; Run/Hold (24 V DC)
*)
Control; Peak/Clear (24 V DC)
*)
Control; TAR (24 V DC)
*)
Supply zero
*)
Ground and supply zero are internally connected. GR201 has differential inputs.
HBMA0114-6.3 en
Page 13
14

3.3 NT101A, NT102A Power supply

Clip
Fig. 3.3: Connection of the power supply
The mains voltage (230 V for NT101A, 115 V for NT102A) must be connected to N and L (phase); the protective ground wire must be connected to . The D.C..-voltage outputs are fed to pairs of internally connected terminals (0 V and 15 V). The max. output current is 650 mA.
HBM A0114-6.3 en
Page 14
Clip
15

4 Setup

If all cables are connected, proceed as follows:
Connect transducer cable
Switch on supply voltage.
Slacken screw on upper right housing edge and hinge down housing cover.

4.1 AE101 Measuring amplifiers

Fig. 4.1: Stick-on label in housing
Connection method
The standard transducer connection-method is the 6-wire technique (with two sensor lines). If the transducer is to be connected using a 4-wire circuit, the terminals 2 and 2’ and 3 and 3’ must be provided with jumper wires.
HBMA0114-6.3 en
Page 15
16
Bridge excitation voltage
Clip
Use switch S26 to select the bridge excitation voltage V
according to the
E
strain-gauge full-bridge’s resistance RB.
Bridge excitation voltage V
10 V 340 ... 5000
5 V 170 ... 5000
2.5 V 85 ... 5000
E
Transducer resistance R
B
Measuring frequency range
Use S28 to adjust the measuring frequency range to 10 Hz or 6 kHz.
Zero point
- Discharge the transducer or charge with preload.
- Use digital voltmeter to measure output voltage, and use rotary switch S23 to adjust minimum indication (if necessary, use S22 to switch polarity); fine balancing with P21 to 0.000 V.
Measuring range
To obtain the maximum measurement resolution, the max. amplifier output­voltage ("10 V) should be used.
AE101
VE = 2.5 V VE = 5 V VE = 10 V switch position S21
mV/V mV/V mV/V 1 2 3 4 5 6 7 8
0.34-0.48 0.17-0.24 0.09-0.12 X X
0.44-0.64 0.22-0.32 0.11-0.16 X X
0.60-0.84 0.30-0.42 0.15-0.21 X X
0.78-1.1 0.39-0.55 0.20-0.28 X X
1.04-1.46 0.52-0.73 0.26-0.37 X X
1.36-1.94 0.68-0.97 0.34-0.49 X X
1.80-2.56 0.90-1.28 0.45-0.64 X X
2.36-3.38 1.18-1.69 0.59-0.85 X X
3.12-4.46 1.56-2.23 0.78-1.12 X X
4.14-5.88 2.07-2.94 1.04-1.47 X X
5.46-7.76 2.73-3.88 1.37-1.94 X X
7.22-10.26 3.61-5.13 1.81-2.57 X X
Tab. 4.1: Nominal measured value depending on the bridge excitation voltage V
HBM A0114-6.3 en
E.
Page 16
Clip
Formula:
17
Part load
Nominal load
Measuring range in V
10 V
Sensitivity in mVńV +
Nominal measurement
value (Range) in
mVńV
Example 1
Force transducer 100 N ¢ 2 mV/V; measuring range wanted 60N ¢ 10 V
60 N
100 N
10 V
10 V
2mVńV + 1.2 mVńV
Load transducer with 60 N. Use S21 to select appropriate interval as given in Tab. 4.1 (page 16) (switch
on 4 + 7), and use P22 to adjust the output voltage to 10 V exactly. The mea­suring range can also be adjusted with no transducer connected by means of a calibration unit (e.g. K 3607) that simulates the transducer signal.
Additive calibration signal
The measuring range can also be adjusted using the additive calibration sig­nal (0.2 mV/V).
Example 2
Force transducer 100 N ¢ 2 mV/V; measuring range wanted: 60 N ¢ 10 V
60 N
100 N
10 V
10 V
2mVńV + 1.2 mVńV
Calculate nominal measurement value, execute zero balance (see above) and use S21 to select the corresponding measuring-range interval. Connect addi­tive calibration signal (set S25 to CAL) and use P22 to adjust with the trans­ducer under no load. The amplifier has now been calibrated. Adjust S25 to MEAS.
0.2 mVńV
1.2 mVńV
10 V + 1.667 V
HBMA0114-6.3 en
Page 17
18

4.2 AE301 Measuring amplifier

Clip
Fig. 4.2: Stick-on label in housing
Connection method
The standard transducer connection-method is the 6-wire technique (with two sensor lines). Cable lengths > 50 m require one resistor each with half the bridge resistance value (R
/2) to be connected to the sensor line instead of
B
the feedback bridges. If the transducer is to be connected using a 4-wire cir­cuit, the terminals 2 and 2’ and 3 and 3’ must be provided with jumper wires.
Bridge excitation voltage
Use switch S26 to select the bridge excitation voltage V
according to the
E
strain-gauge full-bridge’s resistance RB.
Bridge excitation voltage V
5 V 170 ... 5000
2.5 V 85 ... 5000
E
Transducer resistance R
B
HBM A0114-6.3 en
Page 18
Clip
19
Measuring frequency range
The measuring frequency range is 10 Hz.
Zero point
- Discharge the transducer or charge with preload.
- Use digital voltmeter to measure output voltage, and use rotary switch S23 to adjust minimum indication (if necessary, use S22 to switch polarity); fine balancing with P21 to 0.000 V.
Measuring range
To obtain the maximum measurement resolution, the max. amplifier output­voltage ("10 V) should be used.
AE301
VE = 2.5 V VE = 5 V switch position S21
mV/V mV/V 1 2 3 4 5 6 7 8
0.34-0.48 0.17-0.24 X X
0.44-0.64 0.22-0.32 X X
0.60-0.84 0.30-0.42 X X
0.78-1.1 0.39-0.55 X X
1.04-1.46 0.52-0.73 X X
1.36-1.94 0.68-0.97 X X
1.80-2.56 0.90-1.28 X X
2.36-3.38 1.18-1.69 X X
3.12-4.46 1.56-2.23 X X
4.14-5.88 2.07-2.94 X X
5.46-7.76 2.73-3.88 X X
7.22-10.26 3.61-5.13 X X
Tab. 4.2: Nominal measured value depending on the bridge excitation voltage V
Factory setting: VE= 5 V; MR = 2 mV/V
E
HBMA0114-6.3 en
Page 19
20
AE301S6 Measuring frequency range: 2 Hz Calibration signal: 0.1 mV/V
VE=5 V VE=2.5 V switch S21
mV/V mV/V 1 2 3 4 5 6 7 8
0.09-0.12 0.17-0.24 X X
0.11-0.16 0.22-0.32 X X
0.15-0.21 0.30-0.42 X X
0.20-0.28 0.39-0.55 X X
0.26-0.37 0.52-0.73 X X
0.34-0.49 0.68-0.97 X X
0.45-0.64 0.90-1.28 X X
0.59-0.85 1.18-1.69 X X
0.78-1.12 1.56-2.23 X X
1.04-1.47 2.07-2.94 X X
1.37-1.94 2.73-3.88 X X
1.81-2.57 3.61-5.13 X X
Clip
Factory setting: VE= 5 V; MR= 1 mV/V
AE301S7 Measuring frequency range: 60 Hz Calibration signal: 1 mV/V
VE=5 V VE=2.5 V switch S21
mV/V mV/V 1 2 3 4 5 6 7 8
0.09-0.12 0.17-0.24 X X
0.11-0.16 0.22-0.32 X X
0.15-0.21 0.30-0.42 X X
0.20-0.28 0.39-0.55 X X
0.26-0.37 0.52-0.73 X X
0.34-0.49 0.68-0.97 X X
0.45-0.64 0.90-1.28 X X
0.59-0.85 1.18-1.69 X X
0.78-1.12 1.56-2.23 X X
1.04-1.47 2.07-2.94 X X
1.37-1.94 2.73-3.88 X X
1.81-2.57 3.61-5.13 X X
Factory setting: VE= 5 V; MR= 1 mV/V Formula:
Part load
Nominal load
HBM A0114-6.3 en
Measuring range in V
10 V
Sensitivity in mVńV +
Nominal measurement
value (Range) in
mVńV
Page 20
Clip
Example 3
Force transducer 100 N ¢ 2 m V/V; measuring range wanted: 60 N ¢ 10 V
21
60 N
100 N
10 V
10 V
2mVńV + 1.2 mVńV
Load transducer with 60 N. Use S21 to select appropriate interval as given in Tab. 4.2 (page 19) (switch
on 4 + 7), and use P22 to adjust the output voltage to 10 V exactly. The mea­suring range can also be adjusted with no transducer connected by means of a calibration unit (e.g. K 3607) that simulates the transducer signal.
Additive calibration signal
The measuring range can also be adjusted using the additive calibration sig­nal (0.2 mV/V).
Example 4
Force transducer 100 N ¢ 2mV/V; measuring range wanted: 60 N ¢ 10 V
60 N
100 N
10 V
10 V
2mVńV + 1.2 mVńV
Calculate nominal measurement value, execute zero balance (see above) and use S21 to select the corresponding measuring-range interval. Connect addi­tive calibration signal (set S25 to CAL) and use P22 to adjust with the trans­ducer under no load. The amplifier has now been calibrated. Adjust S25 to MEAS.
0.2 mVńV
1.2 mVńV
10 V + 1.667 V
Synchronization
For synchronization of several CF amplifiers, one master device must be de­fined by adjusting S24 to MASTER. The other devices must be set to SLAVE using S24. Then, interconnect terminals 8. Synchronization is also possible with AE501 measuring amplifiers.
HBMA0114-6.3 en
Page 21
22

4.3 AE501 Measuring amplifier

Clip
Fig. 4.3: Stick-on label in housing
Connection method
The standard transducer connection-method for inductive halfbridges is the 5-wire technique for inductive fullbridges the 6-wire technique. If the trans­ducer is to be connected using a 3-wire circuit (inductive half bridge) or a 4-wire circuit (inductive full bridge), the terminals 2 and 2’ and 3 and 3’ must be provided with jumper wires.
Bridge type
Use switch S27 to select the bridge type (inductive half bridge, inductive full bridge).
Bridge excitation voltage
Use switch S26 to select the bridge excitation voltage V ductance L
of the half-bridge or full-bridge transducer. When using SI01
B
according to the in-
E
safety barriers, the bridge excitation voltage must in any case be adjusted to 1 V.
HBM A0114-6.3 en
Page 22
Clip
23
Bridge excitation voltage V
2.5 V 2.5 ... 20 mH
1 V 6 ... 19 mH
E
Transducer inductance L
B
Measuring frequency range
The measuring frequency range is 10 Hz.
Zero point
- Bring transducer into initial position (e.g. mechanical center position).
- Use digital voltmeter to measure output voltage, and use rotary switch S23 to adjust minimum indication (if necessary, use S22 to switch polarity); fine balancing with P21 to 0.000 V.
Measuring range
To obtain the maximum measurement resolution, the max. amplifier output­voltage ("10 V) should be used.
AE501
VE = 1 V VE = 2.5 V switch position S21
mV/V mV/V 1 2 3 4 5 6 7 8
17.0-24.0 6.8-9.6 X X
22.0-32.0 8.8-12.8 X X
30.0-42.0 12.0-16.8 X X
39.0-55.0 15.6-22.0 X X
52.0-73.0 20.8-29.2 X X
68.0-97.0 27.2-38.8 X X
90.0-128.0 36.0-51.2 X X
118.0-169.0 47.2-67.6 X X
156.0-223.0 62.4-89.2 X X
207.0-292.5 82.8-117.0 X X
272.0-387.5 109.0-155.0 X X
360.0-512.5 144.0-205.0 X X
Tab. 4.3: Nominal displacement depending on the bridge excitation voltage V
E
Formula:
Part load
Nominal load
Measuring range in V
10 V
Sensitivity in mVńV +
Nominal measurement
value (Range) in mVńV
HBMA0114-6.3 en
Page 23
24
Example 5
Displacement transducer, nominal displacement " 20 mm ¢ " 80 mV/V Measuring range wanted: 16 mm ¢ 10 V
Clip
16 mm 20 mm
10 V 10 V
80 mVńV + 64 mVńV
Position transducer to 16mm displacement, e.g. using gauge blocks. Use S21 to select the appropriate interval as given in Tab. 4.3 (page 23) (switch on 4 + 8), and use P22 to adjust the output voltage to 10V exactly. The measur­ing range can also be adjusted with no transducer connected by means of a calibration unit (e.g. BN 4800) that simulates the transducer signal.
Additive calibration signal
The measuring range can also be adjusted using the additive calibration sig­nal (8mV/V).
Example 6
Displacement transducer, nominal displacement " 20mm ¢ " 80mV/V; Measuring range wanted: 16mm ¢ 10V.
16 mm
20 mm
10 V 10 V
80 mVńV + 64 mVńV
Calculate nominal measurement, execute zero balance (see above) and use S21 to select the corresponding measuring-range interval. Activate additive calibration signal (set S25 to CAL) and use P22 to adjust with the transducer in displaced position. The amplifier has now been calibrated. Adjust S25 to MEAS.
8mVńV
64 mVńV
10 V + 1.250 V
Synchronization
The maximum number of modules that can be synchronized is 16. For synchronization of several CF amplifiers, one master device must be de-
fined by adjusting S24 to MASTER; the other devices must be set to SLAVE using S24. Then, interconnect terminals 8. Synchronization is also possible with AE301 measuring amplifiers.
HBM A0114-6.3 en
Page 24
Clip

4.4 GR201 Double limit-value switch

M 4
M 3
25
Fig. 4.4: Stick-on label in housing
Reference voltage
Use the potentiometers P21 (coarse) and P22 (fine) for limit 1 or P23 (coarse) and P24 (fine) for limit 2 to adjust the relay response point (reference voltages V
Ref1
and V
). The reference voltages V
Ref2
Ref1
and V
are available at the
Ref2
measuring points M4 (for limit1) and M5 (for limit 2). Connect a digital voltme­ter to adjust the reference voltages as follows:
Limit 1 to measuring point M4 and terminal 4 Limit 2 to measuring point M3 and terminal 9
CAUTION
When the amplifier and limit value switch are supplied from several power supply units their ground connections have to be interconnected.
HBMA0114-6.3 en
Page 25
26
Clip
Switching direction
There are two possible switching directions.
1. S23/24 and S27/28 in upper position
The make contact (terminal 2 or 7) closes upon exceeding of reference volt­age V teresis voltage V
V
V
Ref
V
Hys
and opens upon falling below reference voltage V
Ref
.
Hys
I
minus the hys-
Ref
Contact
open
Contact
closed
Contact
open
Fig. 4.5: GR201 Switching behavior, S23/24 and S27/28 in upper position
2. S23/24 or S27/28 in lower position
The make contact (terminal 2 or 7) closes upon falling below reference volt­age V teresis voltage V
V
V
Ref
V
Hys
and opens upon exceeding of reference voltage V
Ref
.
Hys
I
plus the hys-
Ref
t
Contact
open
Contact
closed
Contact
open
t
Fig. 4.6: GR201 Switching behavior, S23/24 and S27/28 in lower position
HBM A0114-6.3 en
Page 26
Clip
NOTE
Under normal operating conditions, the relays switch as described at V
or V
Ref
- V
Hys
or V before V switching states are always unambiguous.
Changing the hysteresis voltage
The hysteresis voltage is 220 mV. It can be varied by changing R43 resp. R48. The following applies:
respectively. With signal levels in the hysteresis range (V
Hys
Ref
Hys
or V
- V
), EMC influences may cause the relays to switch
Ref
has been reached. Outside the hysteresis range the
Hys
27
Ref
R43 (R48) +
670 mV
[mV]
V
Hys
k
Changing the response and releasing times
, t
The response/releasing time (t
rise
) is less than 5ms. It can be varied by
fall
changing C31a/C34a and C27/C28 (max. 6.8µF). The capacitor’s dielectric strength must be ≥35 V. The following applies:
C31a (C34a)
) 3ms
µF
C31a (C34a)
) 3ms
µF
t
t
fall
rise
18 ms
+
710 ms
+
w Changing the releasing time only
Change C27 (C28) to vary the relay releasing times (max. 47µF). The follow- ing applies:
C27 (C28)
) 3ms
µF
HBMA0114-6.3 en
t
fall
6ms
+
Page 27
28

4.5 EM201 / EM201K2 Output stage

Clip
Fig. 4.7: Stick-on label in housing
The EM201 Output stage has been equipped with one EM002 module, the EM201K2 Output stage has been equipped with two EM002 modules. The EM002 module converts the amplifier output-voltage into a proportional cur­rent signal. For both versions the same housing has been used. Depending on the switch positions, the following current outputs are available:
"20 mA current output
S21/22 (S25/26) in upper position
S23 (S27) in lower position
S24 (S28) in OFF position
4 ... 20 mA current output
S21/22 (S25/26) in upper position
S23 (S27) in upper position
S24 (S28) in ON/OFF position
HBM A0114-6.3 en
Page 28
Clip
29
With S24 (S28) ON, the minimum output current is 3 mA. Even with negative input voltages, this value will not be fallen below. Consequently, a following device that monitors an interruption of cable, for example, will not respond er­roneously, or an externally supplied following device will be supplied continu­ously.
Gain
Use P22 (P23) to adjust the gain for both output stages in the range of 0.9 ... 1.1 Instead of the EM002 module, the EM001 module can also be used. It makes available an additional "10 V/20 mA voltage output. A sepa­rate zero balance is also possible.
Zero balance
Since a zero balance is not provided in the EM002 module, the zero point has to be adjust on the amplifier.
HBMA0114-6.3 en
Page 29
30
Clip

5 TS101 Automatic tare and store unit

5.1 General

The TS101 automatic tare and store unit is an additional Clip electronics mod­ule for further signal processing. The TS101 enables signals supplied by an amplifier connected in series to be tared and stored.

5.2 Function

The TS101 enables the below functions to be activated at the same time:
D tare and save a current value or a peak value (Minimum value or maximum
value or peak-to-peak-value or envelope curve)
or
D save minimum value and maximum value (no taring)

5.2.1 Taring

Use the ”Tare” control input (terminal 10) to start taring. To ensure correct ac­quisition of the tare value even in the case of unstable input signals, an ad­justable low-pass can be used to filter the tare unit’s input signal. In addition, the net signal can be amplified.
Input
"10 V
Supply 15 ... 30 V DC
Output ("10 V)
Tare or peak value (max.)
Output ("10 V)
Peak value (min, max, peak-to-peak, envelope curve)
RDY
Valid tare
0V
24 VDC (15...30 V)
Clear/Peak TareRun/Hold
Control inputs
Fig. 5.1: Block diagram
HBM A0114-6.3 en
Page 30
Clip
31

5.2.2 Store unit

Upon storage, you can select as input signal for the store unit either the ampli­fier signal (gross signal) or the tare unit’s output signal (net signal, not ampli­fied). The selected input signal as well can be filtered and amplified. Use ter­minals 8 and 9 to control the store unit’s operating mode (current value/peak value) and the Run/Hold function. With the ”High” level the control output (RDY) signals a valid tare value.
Current value operating-mode
When the store unit is used to acquire current values, the store contents is permanently kept up to date (”Run” function). Use the ”Hold” control contact to hold the store contents; then, the value is present at the output as an analog signal.
NOTE
With the ”Run” function, the TS101 output signal does correspond to the amplifier’s output signal up to a frequency of 15 Hz.
Meas. signal
o
TS101 output
Run Run
t
Hold
Current value
Function
Operating mode
V
, V
i
Fig. 5.2: Current value operating-mode (Clear)
HBMA0114-6.3 en
Page 31
32
Peak value operating-mode
The peak value operating-mode enables the below values to be acquired:
minimum value or
maximum value or
peak-to-peak values
Use the ”Hold” control contact to hold the store contents.
Store contents
TS101 output
Vi, V
o
Clip
t
Function Operating
mode
Peak value
HoldRun Run
Hold
Current value
Fig. 5.3: Peak values operating-mode (PEAK)
Upon acquisition of peak values, enabling and adjusting the discharge rate enables an envelope-curve function to be prepared (Fig. 5.4 and Fig. 5.5). By adjusting the discharge rate you determine the speed of linear discharging of the peak-value store to the current value.
Discharge rate: too highDischarge rate: too lowDischarge rate: correct
Fig. 5.4: Envelope-curve function
HBM A0114-6.3 en
Page 32
Clip
V
, V
i
o
Discharge rate=0 V/s
Discharge rate=1 V/s
t
Fig. 5.5: Adjusting the discharge rate

5.3 Connection

Before starting up the device, read the safety instructions on page 4.
33
12 series terminals for wire diameters from 0.13 ... 1.5 mm (10 mm end sleeves for strands) are used to connect the TS101.
1
11121111111011911811711611511411311211
Voltage
supply
inputs
Fig. 5.6: TS101 connection terminals
InputOutputsControl
HBMA0114-6.3 en
Page 33
34
Clip
Terminal
1 Operating-voltage zero
2 Input voltage "10 V
3 Operating-voltage zero
4 Peak-value store output-voltage "10 V
5 Tare unit output-voltage "10 V
6 Control output: valid tare (24 V DC)
7 Ground (external for control inputs)
8 Run/Hold peak-value store (24 V DC) control input
9 Peak/Clear peak-value store (24 V DC control input)
10 Tare (24 V DC control input)
11 Supply voltage 0 V
12 Supply voltage 15 ... 30 V DC (unregulated)
Function
Tab. 5.1: Terminal assignment

5.3.1 Voltage supply

The automatic tare and store unit must be supplied with an external supply voltage of 15 V to 30 V. Terminals 11 and 12 are used for connection.

5.3.2 Inputs / Outputs

Inputs
Connect to terminals 1 and 2 the output signal supplied by the amplifier con­nected in series.
Outputs
The TS101 output signals can be picked up at terminal 4 (peak value) and ter­minal 5 (net value). They can be used as the input signal for an indicator (load resistance >5 k) as shown in Fig. 5.7.
NOTE
Screened cable should be used for the analog input and output signals to ensure proper functioning even if the components are exposed to high EMC-levels.

5.3.3 Control inputs / Control output

The control inputs (terminals 7 ... 1 0) are electrically isolated from the proces­sor. They must be connected to an external ground (e.g. PLC) and 24 V (as control signal).
HBM A0114-6.3 en
Page 34
Clip
35
Open control contacts correspond to the ”Low” state (”Peak values” operating mode; ”Run” function).
The RDY control output (terminal 6) signals a valid tare value. The signal level corresponds to the level of the supplied voltage (15 ... 26 V).
Fig. 5.7 shows the connections required for TS101 operation. For the terminal assignment refer to table Tab. 5.1.
1234567891011122’3’ 124589101112 3
11 11 11 11 11 11 11 11 11 11 11 11
11 11 11 11 11 11 11 11 11 11 11 11
Output
"10 V/5 k
Fig. 5.7: TS101 connection (example)
HBMA0114-6.3 en
Page 35
36
S100/6

5.4 Adjustment

5.4.1 Factory setup

See below for the automatic tare and store unit’s factory setup:
Tare unit: Tare value will not be stored upon power failure
Low-pass filter active (0.4 Hz ... 12.5 Hz)
Amplification v=1
Store unit: Peak-value store (maximum value) on
Low-pass filter off
Amplification v=1
Envelope curve off
Clip

5.4.2 Tare unit adjustment

Taring is started when the voltage level on terminal 10 changes over from 0 V to 24 V (positive edge). The tare value is held during voltage supply, even if the voltage level goes down to 0 V again (also refer to next chapter).
NOTE
Taring with input voltages greater than "10.5 V is not possible due to amplifier overflow. A valid tare value (<10.5 V) is signalled by an LED on the board and the ”High” voltage level on terminal 6.
Save tare value and protect it from power failure
The tare value can be stored for reuse after a power failure or after power-off. For this, set switch S100/6 to ”ON” position before taring.
Switch Switch position Effect
OFF Tare value will be lost upon power-off
ON Tare value will be permanently stored in EEPROM
*)
Factory setup
HBM A0114-6.3 en
*)
Page 36
Clip
37
Signal amplification
The net signal can be amplified to obtain a sufficiently high resolution even with measurement tasks with great initial load. Use switch S101 to adjust the amplification in four steps:
S101/2 S101/3 S101/4 S101/5 Amplification
ON OFF OFF OFF v=1
OFF ON OFF OFF v=2
OFF OFF ON OFF v=5
OFF OFF OFF ON v=10
Example
A load cell with 10 kg nominal load is used with an initial load of 4 kg. The mass to be weighed is 1 kg. To avoid amplifier overload the amplifier’s measuring range must be adjusted to maximum signal (here 5 kg=10 V output voltage). The automatic tare unit connected in series is used to tare the initial load (4 kg=8 V). The remaining residual signal of 1 kg corresponds to an output voltage of 2 V. Subsequent amplification by factor 5 enables the residual signal to be raised to 10 V.
Low-pass filter adjustment
Use switch S100/5 to activate the filter for the automatic tare unit, only, or also for the peak-value store. Upon taring, the low-pass filter influences the input signal. However, it has no effect on the tare output’s bandwidth.
S100/5 Position P1 (cutoff frequency) Influences
OFF
ON
Tab. 5.2: Low-pass filter
5 Hz
12.5 Hz
0.4 Hz
1.7 Hz
3.1 Hz
0.1 Hz
tare unit
tare and
store unit
HBMA0114-6.3 en
Page 37
38
8
9
Clip
Procedure
Set switch S100/7 to ”OFF” position
To activate the filter for the peak value as well, set switch S100/5 to
”ON”position
Push and hold switch S6 and turn potentiometer P1. Refer to Tab. 5.2 for the cutoff frequency and the appropriate potentiometer position.
NOTE
The filter frequencies for taring and peak value always bear a constant relationship and cannot be adjusted independently of each other.

5.4.3 Store unit adjustment

Adjust operating mode
Use the signal levels on terminals 8 and 9 to select the operating mode (cur­rent value/peak value) and the Run/Hold function.
Terminal Level Function Effect
High Run Store contents is permanently kept up to date
Low Hold Store contents is held
High Peak Peak value operating-mode
Low Reset Current value operating-mode
High=24 V; Low=0 V
NOTE
The peak-value store is cleared by switching from the peak value operat­ing-mode to current value.
HBM A0114-6.3 en
Page 38
Clip
Select peak value and amplification
Use the switches S100/1... 4 and S101/1 to define the type of peak value to be stored (min/max, peak-peak) and the amplification.
S101/1 S100/1 S100/2 S100/3 S100/4 Function
ON OFF OFF OFF OFF Peak value minimum ON ON OFF OFF OFF Peak value maximum ON OFF ON OFF OFF Peak-to-peak ON ON ON OFF OFF (Peak-to-peak) x 0.5 ON OFF OFF ON OFF Tared minimum value (v=1) ON ON OFF ON OFF Tared maximum value (v=1) ON OFF ON ON OFF Tared minimum value (v=2) ON ON ON ON OFF Tared maximum value (v=2) ON OFF OFF OFF ON Tared minimum value (v=5) ON ON OFF OFF ON Tared maximum value (v=5) ON OFF ON OFF ON Tared minimum value (v=10) ON ON ON OFF ON Tared maximum value (v=10)
OFF OFF OFF ON ON Minimum value (peak-value output)
and maximum value (tare output)
*)
39
Tab. 5.3: Adjust peak value and amplification
*)
Factory setup
NOTE
With the minimum and maximum value setup selected, no taring is pos­sible due to the fact that the tare output is used for output of the maxi­mum value. Please make sure that for the maximum value (tare output) amplification v=1 has been adjusted (switches S101/2...5, see page 37).
Adjust envelope curve
Use switch S100/7 and potentiometer P1 to adjust the envelope curve operat­ing-mode.
Procedure:
First make sure that the ”Peak value” operating mode has been adjusted (see page 38).
Make sure that the type of peak value to be stored has been adjusted (see Tab. 5.3).
Set switch S100/7 to ”ON” position.
Push and hold switch S6 and turn potentiometer P1.
HBMA0114-6.3 en
Page 39
40
Clip
Refer to Tab. 5.4 for the discharge rate and the appropriate potentiometer position. For more precise acquisition of the change in discharge rate use an oscilloscope.
Switch S100/7 P1 Envelope-curve function
ON At right-hand stop Off
Discharge rate 0.005...1 V/s
ON
OFF No effect Off
Turn anticlockwise
from right-hand stop
0.5 V/s
1 V/s
0.005 V/s
0 V/s
Tab. 5.4: Envelope-curve adjustment
Low-pass filter
It is possible to filter the peak-value store’s input signal. Refer to Tab. 5.2, page 37 for the required setups. Please remember that the filter frequencies cannot be adjusted independently of each other because they bear a constant relationship.

5.5 Zero-point balance

A zero-point balance has been made at the factory. Should the output signals for tare unit and store unit vary from zero, it is possible to correct them. Use switches S100/1... 4 and S101/1 to define the output to be corrected.
S101/1 S100/1 S100/2 S100/3 S100/4 Function
ON ON OFF ON ON Tare output (terminal 5) ON OFF ON ON ON Peak-value output (terminal 4)
Procedure
Push and hold key S6.
Use potentiometer Pl to adjust the zero point.
Should the balancing range available not be sufficient:
Release key S6.
Turn potentiometer to initial position.
Restart the balancing procedure.
HBM A0114-6.3 en
Page 40
Clip

5.6 Component position diagram

S6
LED
41
S10
0
P1
11121111111011911811711611511411311211
Fig. 5.8: TS101 tare and store unit
S10
1
1
HBMA0114-6.3 en
Page 41
42
Clip

6 Safety barriers

Connect HBM safety barriers to obtain intrinsically safe measuring circuits [EEx ia] IIC. A test certificate is required for transducers used in such applica­tions.
Measuring amplifier Safety barrier
AE101 SD01A
AE301 SD01A
Due to the safety barriers’ high internal resistance, the bridge excitation volt­age V transducer inductance LB respectively. Refer to the below table to obtain the minimum permissible transducer resistance.
must be reduced according to the transducer resistance RB or the
E
Measuring amplifier V
AE101 10 V not permissible
AE301 5 V 320
*)
With longer cables the supply-line resistance must be added.
E
5 V 320
2.5 V 130
2.5 V 130
R
(minimum)
B
*)
, L
B

7 CLIP IG

For use in industrial environments, the CLIP modules can be installed in an aluminium-die-cast wall housing. The degree of protection is IP65. The hous­ing enables max. 4 modules incl. power supply unit to be inserted; it is also possible to insert two measuring amplifiers with safety barriers.
The wall-housing version is particularly suitable for rough environments. It is protected from electromagnetic interference in the surroundings.
The CLIP IG devices are mounted at the factory and wired internally. All combinations are available on request.
Attached to CLIP IG devices accessory bags 2-9278.0339 are delivered in­cluding a standard set of earthing bushes, end sleeves for strands (0.5 mm
2
,
10mm long) and non-buckling bushes that fit HBM cables of type
2
KAB8/00-2/2/2 (6 x 0.14 mm
Order No. 4-3301.0082, blue (6 x 0.14 mm
4-3301.0071, gray (6 x 0.14 mm
HBM A0114-6.3 en
) or
2
) or
2
)
Page 42
Clip

8 Dimensions

AE101 AE301 AE501 EM201 EM201K2 GR201 TS101
77
111.3
112.9
35
43
42
15
110.1
NT101A NT102A
113.5
102
69
53
56
10.5
10
37
HBMA0114-6.3 en
Page 43
44
Clip
100
hole pattern for screw / 6
Clip IG
HBM A0114-6.3 en
Weight of housing: 3.2kg
Page 44
Clip

9 Specifications

Clip IG
Type of protection IP65
Wight approx. kg 4.3
Wight (empty) kg 3.2
Ambient temperature °C -20 ... +50
Operating voltage
with Power supply unit 101A (Type WG 010) with Power supply unit NT 102A (Type WG 011) without Power supply unit (Type WG 012)
Mechanical strain (examination similar DIN IEC 68)
Oscillate (30 min each direction)
Impact (3 times each direction, impact duration 6ms)
V V
V DC
m/s
m/s
2
2
230 "10 % 115 "10 %
15 ... 30
50 (5 ... 65 Hz)
350
45
HBMA0114-6.3 en
Page 45
46
AE101, AE 301, AE501 Measuring amplifiers
Type AE101 AE301 AE501
Accuracy class 0.1 0.1 0.1
Transducers that may be connected strain gauge full bridge
V
E
V
E
V
E
=10 V = 5 V =2.5 V
Ω Ω Ω
Inductive half/full bridge
=5 V
V
E
=2.5 V
V
E
V
=1 V
E
Bridge excitation voltage VE (sym­metrical)
mH mH mH
V V V
Permissible cable length
between transducer and amplifier
m 500
Carrier frequency Hz DC 600 4800
Bridge zero balance
coarse approx. fine approx.
mV/V mV/V
Measuring ranges
V
E
V
E
V
E
V
E
=10 V = 5 V = 2.5 V = 1 V
mV/V mV/V mV/V mV/V
Calibration signal, in addition to the meas. signal
mV/V + 0.2 " 1 % + 0.2 " 1 % + 8 " 1 %
Input impedance M >10 / 2 nF > 1 / 3 nF > 1 / 1.5 nF Common mode voltage. max. perm. Vpp " 10 V " 5V
Common mode rejection
0... 300 Hz > 300 Hz
dB dB
Linearity deviation % full
scale
Output voltage
Slew rate. max.
V
V/µs
Load resistance k 4
Internal resistance < 2
1)
AE301S6 and AE301S7: 0.1...2
2)
AE301S6 and AE301S7: 0.2...4
3)
AE301S6 and AE301S7: 0.2...4
340 ... 5000 170 ... 5000
85 ... 5000
-
-
-
10
5
2.5
" 2
" 0.08
0.1 ... 2
0.2 ... 4
0.4 ... 8
-
> 100
>85
typ 0.03
" 10
0.4
170 ... 5000
< 0.05
-
85 ... 5000
-
-
-
5
2.5
-
" 1
" 0.05
-
0.2 ... 4
0.4 ... 8
1)
2)
-
-
-
-
5 ... 20
2.5 ... 20 1 ... 20
2.5 1
-
" 80
" 3.2
-
-
8 ... 160
20 ... 400
> 100
­< 0.1
typ 0.05
" 10
-
Clip
HBM A0114-6.3 en
Page 46
Clip
Technical data continued
Type AE101 AE301 AE501
Measuring frequency range
3rd-order changeover low-pass filter. Bes­sel (-1 dB) 3rd-orderlow-pass filter Bessel(-1 dB)
Hz
kHz
Hz
Phase velocity
with 0...10 Hz filter with 0...6 kHz filter
ms µs
Rise time with 0...10 Hz filter ms 25
Overshoot in the case of voltage surge
with 0 ... 10 Hz filter with 0 ... 6 kHz filter
% %
Noise voltage
measuring range 0.2 mV/V (10 Hz) measuring range 2 mV/V (10 Hz) measuring range 1 mV/V (10 Hz) measuring range 10 mV/V (10 Hz) measuring range 8 mV/V (10 Hz) measuring range 80 mV/V (10 Hz) measuring range 0.2 mV/V (6 kHz) measuring range 2 mV/V (6 kHz)
mV mV mV mV mV mV mV mV
rms rms rms rms rms rms rms rms
Long term drift over 48 hours (after 1 h warm-up time)
µV/V
Effect of a 10 K-change in ambient tem­perature
on sensitivity
% full
scale
on zero point
measuring range 0.2 mV/V measuring range 2 mV/V meas. range 8 mV/V (1 mV/V) meas. range 80 mV/V(10 mV/V)
mV mV mV mV
Effect of a +15 ... 30 V change in operat­ing voltage
on sensitivity zero point (350 bridge resistance)
mV mV
5V-synchronization (square wave) kHz - 76.8
Residual carrier voltage mV - < 5
Operating voltage (DC) V +15 ... 30
Power consumption mA 125 100
Nominal temperature range °C -20 ... +60
Service temperature range °C -20 ... +60
Storage temperature range °C -25 ... +70
Degree of protection IP20
Weight g 200
3)
AE301S6: 0...2 (-1dB) AE301S7: 0...60 (-1dB)
4)
AE301S6: <80 (filter frequency 2 Hz) AE301S7: <2,8 (filter frequency 60 Hz)
5)
Rise time with AE301S7 6 ms Rise time with AE301S6 200 ms
0 ... 10
0 ... 6
-
< 18 < 20
0
<10
< 4 < 4
-
-
-
-
< 30
< 6
< 0.2
< 0.1
typ 0.05
< 60 < 10
-
-
-
-
0 ... 10
4)
< 16
-
5)
< 4 < 4
-
-
-
-
-
-
< 0.1
< 0.1
typ 0.05
< 10
< 4
-
-
< 1 < 1
-
-
3)
0 ... 10
< 17
-
<2
-
-
-
-
­< 4 < 4
-
-
< 0.8
< 0.1
typ 0.05
-
-
<10
< 4
47
HBMA0114-6.3 en
Page 47
48
GR201 Limit value switch
Accuracy class 0.1
Differential input
Voltage Impedance
Reference voltage
coarse approx. fine approx.
V
k
V V
" 10
> 50
" 10
" 0.5
Clip
Switching hysteresis
Factory setting: R43, R48 to be changed by R43 and R48
Influence of a 10K-change in ambient tem­perature on the switching point
Switching-point error % < 0.05 full scale
Relay capacity
max. voltage max. current max. power
Switching times (Factory setting)
Response time
Releasing time Use C31a. C34a to change the response or releasing time
Response time
Releasing time
Use C27. C28 to change the releasing time only
Releasing time
mV
k k
% < 0.05 full scale
V A
W
ms ms
ms
ms
ms
45 (Protective low voltage)
18 . C31a (C34a) +3
710
220
3.01
670 mV
V
Hyst.
1
30 (25 VA)
< 5 < 5
µF
.
C31a (34a) +3
µF
.
C27 (C28) +3
6
µF
Operating voltage V DC +15 ... 30
Power consumption mA < 100 (20 mA no relay closed)
Nominal temperature range °C - 20 ... + 60
Service temperature range °C - 20 ... + 60
Storage temperature range °C - 25 ... + 70
Degree of protection IP20
Weight g 200
HBM A0114-6.3 en
Page 48
Clip
EM201 Output stage module (with one module EM002) EM201K2 Output stage module (with two modules EM002)
Accuracy class 0.1
Input
Voltage Impedance
V
k
Operating voltage V DC +15 ... 30
Power consumption
(fully assembled with 2 x EM002)
mA < 180
" 10 (0 ... + 10 V)
> 11.5
49
Nominal temperature range
Service temperature range
C - 20 ... + 60
°
C - 20 ... + 60
°
Storage temperature range °C - 25 ... + 75
Weight g 200
EM002
Output signal. selectable mA " 20 4 ... 20
Output current
with V with V
=10 V
E
= 0 V
E
mA mA
20 " 0.02
< " 0.04
20 " 0.5
4 " 0.2
Output current limit - - > 3 (switchable)
Permissible load resistance < 500
Linearity deviation % full scale < 0.05
Internal resistance k > 100
Measuring frequency range kHz 3 (-1 dB)
Degree of protection IP20
NT 101A, NT 102A*) power supply
Type NT101A NT102A
Input voltage V 230 "10 % 115 " 10 %
Permissible frequency range Hz 47 ... 63
Output voltage V
Output current at >25° ... +60° A
DC
DC
15.3" 2 %
0.4
Output power W 6.75
Efficiency approx. % 60
Current limiter (protected against sustained
1.2 x
(permanently adjusted)
I
n
short circuit) at
Residual ripple mV
Maximum permissible cable length between
m 3
pp
< 10
NT101A or NT102A and Clip module
Ambient temperature
Excess-temperature protection
Test voltage kV
°
C - 20 ... + 60
C typ.105 (trafo temperature)
°
eff
3.75 (prim/sec and prim/housing)
Degree of protection IP20
Weight g 420
*)
Version to DIN -VDE0551, EN60742 Protection class 1. The maximum permissible continuous current is 450 mA.
HBMA0114-6.3 en
Page 49
50
TS101 tare and store unit
Accuracy class 0.1
Input voltage V "10
Input impedance k 100
Output voltage V " 10
Permissible load resistance k 5
Linearity deviation % < 0.04 of full scale
Effect of a 10 K-change of the ambient temperature
Effect of a 15 ... 26 V-change of the op­erating voltage
Long-term drift over 48 h (after 1 hour warm-up time)
Noise voltage of the output mV
Control inputs (floating)
High signal level Low signal level
Control output
High signal level Low signal level
Output current mA <500
% < 0.1 of full scale
% < 0.01 of full scale
% < 0.02 of full scale
pp
V V
V V
11...30 (24 V nominal)
< 20
0 ... 5
Vb-2
<1
Clip
Tare unit
Output ms Net value (alternatively pos. peak val.)
Net-value amplification 1, 2, 5, 10-fold, selectable in steps, for tar-
ing of major initial loads
Tare error (with v=1) mV <4
transient time for the output voltage
after taring
Low-pass filter (before taring) Hz 0.1 ... 12.5; adjustable
Transmission bandwidth kHz >10
Storage time for tare value Unlimited as long as V
Control input Taring with rising edge
Delay time for taring ms <1
Control output Valid taring
ms 40 (to 99.9 %)
is present (alterna-
tively, storage in EEPROM)
b
HBM A0114-6.3 en
Page 50
Clip
Technical data continued
Peak-value store unit
Output Peak value (alternatively, pos./neg. peak,
peak/peak 0.5 x peak/peak or current
value or envelope-curve value, tared and
amplified (1, 2, 5, 10-fold))
Peak-value store update-rate ms <1.3
51
Accuracy %
%
Transmission bandwidth Hz 15 (-1 dB)
Transient time for the output voltage ms 40 (to 99.9 %)
Discharge rate for envelope curve mV/s 5 ... 1000, adjustable
Control inputs Run/Hold;
Delay time for the control signals ms <8
Connection 12 series terminals for
wire 0.13 ... 1.5 mm2; 10 mm end
Operating voltage V
Power consumption mA <90
Nominal temperature range °C -20 to +60
Service temperature range °C -20 to + 60
Storage temperature range °C - 25 to +70
Weight g ca. 200
Protection to EN60529 IP20
Mounting On mounting rails to EN 50022
b
V DC 15...30, unstabilized
0.25 (in 6 ms)
0.05 (in 20 ms)
(clear/peak)
sleeves for strands

10 Accessories

To be ordered separately:
Brackets 3-6450.0001
HBMA0114-6.3 en
Loading...