is displayed when the temperature being
when lower
is displayed when the thermometer is exposed to
is
displayed when the ambient temperature of the
.
In both cases you should allow plenty of time (minimum
to the
For all other error messages it is necessary to reset the
thermometer. To reset it, turn the instrument off, remove
the battery and wait for a minimum of one minute, reinsert
. If the error message remains
e Service Department for further
Battery OK
:
Battery Low
: battery needs to
Battery
Depleted’
:
Measurement Range
-55 to +250
°C (-67 to 482°F)
Ambient Operating Range
0 to 50°C
(32 to 122°F
)
Accuracy
±
2% of reading or
±
2°C
(4°F)
Resolution
0.1°C
/ 0.1°F (selectable
)
Response Time
1 second
Distance
-to-
Spot ratio
6:1
(i.e., measures a 1
-
inch spot at a
Emissivity
Range
0.95 default
– adjustable
from
0.05 to 1.00
Battery Life
Typ. 40hr, minimum 30hr
Battery
2 CR2032 batteries
Dimensions
Weight
65 grams (2.29 oz)including
batteries
The instrument is a non-contact infrared laser thermometer. Simply aim the thermometer at the target and press the ‘Scan key’ to display the surface temperature.
Laser Information: By partially pressing the ‘Scan key’, the device will measure the temperature of the target. When the ‘Scan key’ is fully depressed, the laser operates.
EMISSIVITY
The infrared thermometer has a factory default emissivity setting of 0.95. The emissivity of the thermometer can be changed from 0.05 (5E) to 1.00 (100E), but changes should only be made by experienced personnel. To change emissivity, turn the instrument ON by pressing the ‘Scan key’, then press the ‘Mode key’ five times to select the emissivity function. The 95E will flash on the LCD screen, then press the ‘Scan key’ to adjust the emissivity value, press the ‘Mode key’ again to set the desired emissivity value and exit the set-up screen. For information relating to the emissivity of specific materials, please contact the nearest retailer. Caution: non-contact infrared thermometers are not recommended for measuring the temperature of shiny or polished metals.
LCD ERROR MESSAGES
The thermometer incorporates visual diagnostic messages as follows:
BATTERY REPLACEMENT
When the ‘Low Battery’ icon indicates the battery is low, the battery should be replaced immediately with a CR2032 lithium cell. The battery is located under the twist cover on the back of the thermometer. Please note: It is important to turn the instrument OFF before replacing the battery otherwise the thermometer may malfunction.
Dispose of used batteries properly and keep away from children.
STORAGE & CLEANING
The sensor lens is the most delicate part of the thermometer. The lens should be kept clean at all times, care should be taken when cleaning the lens using only a soft cloth or cotton swab with water or medical alcohol, allowing the lens to fully dry before using it, do not submerge any part of the thermometer. The thermometer should be stored at room temperature between -20 to 65°C (4 to
149°F).
CAUTION!
1. WHEN DEVICE IS IN USE, DO NOT LOOK DIRECTLY INTO THE LASER BEAM–PERMANENT EYE DAMAGE MAY RESULT.
2. USE EXTREME CAUTION WHEN OPERATING THE LASER.
3. NEVER POINT THE DEVICE TOWARD ANYONE’S EYES.
4. KEEP OUT OF REACH OF CHILDREN.
°C OR °F MODE
To change the units from °C to °F or from °F to °C, first turn the instrument ON by pressing the ‘Scan key’, then press the ‘Mode key’ four times, the ‘°C’ or ‘°F’ symbol will flash, press the ‘Scan key’ to change to temperature units.
MINIMUM OR MAXIMUM MODE
To utilize the thermometer’s Minimum or Maximum mode, turn the instrument ON by pressing the ‘Scan key’, press the ‘Mode key’ once for Minimum, twice for Maximum. The ‘MIN’ or ‘MAX’ icon will flash, then press the ‘Scan key’ to set operation for the MIN or MAX mode. The thermometer will display the MIN or MAX temperature value measured while the Scan key was pressed.
LOCK MODE
The LOCK mode continuously monitors temperature without pressing the SCAN key.
To utilize the thermometer’s LOCK mode, turn the instrument ON by pressing the ‘Scan key’, then press the ‘Mode key’ three times for the LOCK mode function. The LOCK icon will flash, then press the ‘Scan key’ to confirm and set the LOCK mode. The thermometer will continuously display temperature for up to 60 minutes or until the ‘Scan key’ is pressed again.
‘Hi’ or ’Lo’ measured is outside of the range of the instrument, ‘Hi’ when higher than +250°C (482°F) and ‘Lo’ than –55°C (-67°F).
‘Er2’ rapid changes in the ambient temperature. ‘Er3’
thermometer EXCEEDS 0°C (32°F) OR +50°C (122°F)
30 minutes) for the thermometer to stabilize working/room temperature.
the battery and turn ON please contact th assistance.
BATTERIES
The thermometer has two separate batteries. The battery closer to the side of the laser beam output is for laser operation only. The other battery is dedicated to temperature measurements. The thermometer can measure temperature accurately, even without the laser battery.
The thermometer provides a visual indication of battery status as follows:
measurements
are possible
be replaced, measurements
are possible
measurements are
not possible
SPECIFICATIONS
greater
(-9.9 to 199.9°C)
(90%)
(field-of-view)
6-inch distance)
(auto power OFF after 15 seconds)
22.5 x 50 x 103mm(0.9×2.0×4.0 inch)
EMC/RFI
Readings may be affected if the instrument is operated within radio frequency electromagnetic field strength of approximately 3 volts per meter, but the performance of the instrument will not be affected.
WARRANTY
The thermometer is guaranteed for a period of one year from the date of purchase against mechanical and electrical manufacturing defects. There are no user serviceable parts inside the instrument. Any attempted repair by unauthorized persons voids the warranty.