Harvard Apparatus CMA 31 User Manual

CMA 31 Linear Microdialysis Probe User’s Manual
TECHNICAL INFORMATION
Membrane
Material Polyethersulfone (PES)
Molecular Cut-Off 55,000 Daltons
Outer Diameter 0.26 mm
Length 10 mm
Material Polyimide
Length Inlet Tubing 350 mm
Length Outlet Tubing 100 mm
Total Length (tubing + membrane) 460 mm
Outer Diameter 0.19 mm
Double Tubing Outer Diameter 0.63 mm
Double Tubing Length 12 mm
The probe can be sterilized in its inner package with ethylene oxide.
Cutaneous Implantation, Anesthetized Animal
For CMA 31 Linear Microdialysis Probe: To avoid ultra ltration over the membrane
the probe should be used with a Perfusion Fluid consisting of a Ringer solution containing 30g Dextran 60/1000mL.
1. Anesthetize the animal. Be sure to keep the animal’s body temperature normal during the surgical procedure and the study period by using a CMA 450 Temperature Controller.
2. Clean and shave the animal.
3. Mark the skin with a felt tip marker where the introducer needle should go in and exit, leaving at least an 18 mm window if drug transport is studied. Tunnel the introducing needle at the appropriate depth required for the experiment.
4. Load the CMA pump with a Perfusion Fluid-lled syringe (either a Luer lock
single use syringe or a glass syringe with a xed needle). Make sure there are
no air bubbles.
5. Connect the inlet Luer lock connector to the single use syringe or cut off the Luer lock connector from the probe inlet with a sharp cut, and connect it to the glass syringe needle via a blue Tubing Adapter.
6. Start the pump at 5 µL/min. Conrm that liquid is owing from the outlet of
the probe. Maintain this ushing rate for approximately 2-3 min to remove air
bubbles.
7. Reduce the ow rate to operational rates, usually 1-3 µL/min.
8. Remove the protective tubing from the probe. Slide the probe outlet tubing through the needle end of the introducer carefully so the membrane sits in the middle of the introducer. Make sure not to kink the tubing.
9. Check that the ow rate is the same as before the insertion. Hold the probe in place against the skin while carefully withdrawing the introducer needle.
10. Secure the probe position with surgical tape or surgical glue.
11. Wait at least 30 min for stabilization before starting the sample collection.
WARRANTY
The probes manufactured by CMA Microdialysis are warranted to be free from defects in material and workmanship for a period of two years from the manufacturing date if stored in the original package. Claims should be forwarded without delay to CMA Microdialysis or to your local distributor.
CMA Microdialysis AB
Head Ofce, Sweden
Box 2, 171 18, Solna, Sweden • Tel: 46-8-470 10 00 • Fax: 46-8-470 10 50
E-mail: cma@microdialysis.se • www.microdialysis.com
Harvard Apparatus
84 October Hill Road Holliston, MA 01746 USA • Tel: 508-893-8999 • Fax: 508-429-5732
E-mail: support@hbio.com • www.harvardapparatus.com
ORDER INFORMATION Ref No.
CMA 31 Linear Microdialysis Probe, 4/pkg CMA 8010631
Tubing Adapter, 10/pkg CMA 3409500
FEP Tubing 1m, 1/pkg CMA 3409501
FEP Tubing 1m, 10/pkg CMA 8409501
Vial, plastic, 300µL, 1000/pkg CMA 7431100
Caps, plastic, 1000/pkg CMA 7431101
Vial, Glass 300 µL CMA 7431007
Cap/Seal Non-Reclosing, 1000/pkg CMA 7432175