HARVARD 50-9513 Operating And Maintenance Instructions Manual

Page 1
Publication # Revision 0 Date 12/28/94
84 October Hill Road M
assachusetts 01746
Tel.: (508) 893-8999 Fax: (508) 429-5732 e
-mail: bioscience @harvardapparatus.com
W
eb Site: harvardapparatus.com
Operating and Maintenance
Instructions
Harvard
Isolated Preamplifier
Catalog Number:
50-9513
Page 2
Table of Contents
Subject Page(s)
Caution/Warning Statement ............................................ 2
Introduction ....................................................................... 2
Specifications ..................................................................... 3
Operating Procedure:
Recording with Electrodes .......................................... 4
Siting and Fixing the Electrodes ................................ 5 - 7
Use with the Harvard Modular
Universal Oscillograph ................................................. 8 - 9
Use with the Harvard
Student Oscillograph .................................................... 10 - 11
Problems .............................................................................. 12
Service Information .......................................................... 12 - 13
Customer Contact Addresses ........................................... 13
Diagrams:
Figure 1 – Isolated Preamplifier Controls ................. 3
Figure 2 – ENG Electrode Positions ............................ 5
Figure 3 – EMG Electrode Positions ........................... 5
Figure 4 – ECG Electrode Positions ........................... 6 - 7
Figur
e 5 –
Modular Universal Oscillograph Controls
........................................................ 9
Figur
e 6 – Student Oscillograph Contr
ols
..................
11
Figur
e 7 – Back Panel of Pr
eamplifier
.......................
12
Page 3
2
Important Notice
This unit is intended for use as a research or teaching tool ONLY and is not authorized for use as a component in a life support device or other clinical system. Therefore, neither we nor our agents can be held responsible if this notice is dis­regarded.
WARNING
50-6808 Harvard Surface Electrodes
When using Coupland-Davis type electrodes or ANY type of needle electrode, ALWAYS STERILIZE THE ELECTRODES BEFORE AND AFTER USE ON EACH SUBJECT to reduce the chance of skin infections or blood-carried diseases being transmitted.
Introduction
The Harvard Isolated Preamplifier has been designed for ECG (EKG), EMG, and ENG (EOG) recording with subject isolation. The front end of the preamplifier is a battery operated opto-isolator. The remainder of the circuit obtains its power from the host unit.
There are two types of input: a 3-pin socket for ordinary differen­tial applications, and a “5-lead” input for ECG limb selection, cou­pled with a 7-position selector switch. There is a 3-position gain selector and 3-position filter selector.
NOTE: When this Isolated Preamplifier is to be used with our Universal range of Oscillographs or with either the 50-4027 or 50-4028 4-Channel Interface Adapter, the 50-5966 Isolator Interface Unit is also required.
Page 4
Specifications
Batteries 2 standard 9 V “transistor radio” type (PP3 or
IEC designation 6LR61)
Inputs 5 standard color coded ECG 2 mm pin type
sockets LA, RA, LL, RL, C. Also 3-pin socket for general differential input applications
Lead Selector Standard 7-position ECG lead selector: I, II,
III, AVR, AVL, AVF, V
Filtering 3-position filter switch:
30 Hz 3 dB down 150 Hz 3 dB down 15 kHz 3 dB down
Gain Selectable x10, x100, x1000
Other Facilities DC level control; battery level indicator;
choice of DC or AC coupling
Rear Panel Coupling lead for use with Harvard instru-
ments; battery housings
3
Figure 1
I s o l a t e d P r e a m p l i f i e r
INPUT
R.A. L.A. R.L. L.L. V
RED YELLOW BLACK GREEN WHITE
I
II
III
AVR
AVL
AVF
V
D.C. LEVEL
FREQ
CUT
GAIN
15 xHz
150 Hz
30 Hz
A. C.
D.C.
x 1000
x 100
x 10
HARVARD
(4)(3)(2)(1)
(6)
(5)
Page 5
4
Operating Procedure
Please note the “Important Notice” on page 2
The instructions that follow are for use with the Modular Universal Oscillograph and the Student Oscillograph. To use the Isolated Preamplifier with the Harvard Electrophysiological Teaching Unit, please refer to the instruction manual supplied with that unit.
Points About Recording with Electrodes
Human Subjects: First, always clean the area of electrode applica-
tion with surgical spirit or similar, to remove grease, dirt, grime etc.. When using Coupland-Davis type electrodes or ANY type of needle electrode ALWAYS STERILIZE THE ELECTRODES BEFORE AND
AFTER USE ON EACH SUBJECT
to reduce the chance of skin
infections or blood-carried diseases being transmitted.
Always ensure that the electrodes are clean. This reduces noise on your chart recordings.
Choose the siting of the electrodes carefully. Avoid areas of excess fat (fat will reduce the amplitude of your signals).
Animal Subjects: If you are using needle electrodes, there is a chance that a small DC voltage will develop at the electrode site which may well “swamp” the input of the amplifier, causing the sig­nal to get progressively smaller until it finally disappears. This can be countered by connecting the 50-8390 Harvard Capacitor Box between the electrodes and the Isolated Preamplifier.
Page 6
Siting and Fixing the Electrodes
Fig. 2 below shows a suggested positioning of electrodes for ENG (EOG). Note the positioning of the reference electrode (or zero volt electrode). This electrode (within reason) can be placed anywhere.
5
For EMG, just place the electrodes over the group of muscles to be monitored, see Fig. 3 below.
Figure 3
Figure 2
+
To 3-pin socket on Isolated Preamplifier
To 3-pin socket on Isolated Preamplifier
Bicep
Ticep
Earth
Page 7
}
6
Siting and Fixing the Electrodes (continued)
Fig. 4 on this page shows the conventional positions of electrodes for recording ECG. You will notice that the patient’s RIGHT LEG is the 0 volt reference.
Essentially an ECG can be recorded from any part of the body, but unless the conventional positions are used, the interpretation of the results would be difficult. For more guidance, please refer to one of the many books published which cover the recording of ECG wave­forms.
2
1
3
4
5
6
Figure 4
0 volt reference
black
green
red
blue
yellow
To Isolated Preamplifier
Page 8
7
AVL
I
AVR
II
III
AVF
V
1
+
+
V
2
+
+
V
4
+
+
V
5
V
6
V
3
Page 9
Recording Waveforms Using the Harvard Modular Universal Oscillograph
Numbers in brackets, e.g. (1), (2) etc. refer to Fig. 1 on page 3. Letters in brackets, e.g. [A], [B] etc. refer to Fig. 5 on page 9 and Fig. 12 in the Modular Universal Oscillograph Instruction Manual.
1) Have you read the complete Operating Procedure section?
2) Check that the Isolated Preamplifier is fitted with batteries (see page 12).
3) Refer to Fig. 2, 3 or 4 on pages 5, 6 and 7 to apply the elec­trodes to your subject.
4) Connect the electrodes to the Isolated Preamplifier using either the 3-lead input cable or the ECG limb cable.
5) With reference to the Harvard Modular Universal Oscillograph Instruction manual section 3.4.0, follow steps 1 to 13 inclu­sive.
6) Plug the Isolated Preamplifier to the Isolator Interface.
7) Switch the Isolator Interface on and check that the battery indicator (1) on the Preamplifier is in the green zone. If it is not, refer to page 12.
8) Set the Preamplifier’s AC/DC switch (5) to the DC position and set the filter switch to 15 kHz.
9) Set the Preamplifier's GAIN switch (6) to 10 and select posi­tion I on switch (2).
10) Set the Oscillograph's Direct-OFF-Interface switch [AC] to INTERFACE, the AC/DC switch [AA] to DC, the chart speed control [T] to 2.5, and start chart drive by pushing chart drive lever [M] to ON.
11) Adjust the Oscillograph’s GAIN control [W] to about 1. The pen will move on the chart.
12)
Adjust the Pr
eamplifier’s DC LEVEL control (3) until the pen
r
etur
ns to its central position on the chart. At this point you
should see a small waveform on the chart.
8
Page 10
The amplitude of the waveform can be increased or decreased by using the Oscillograph's GAIN control [W] and the Preamplifier's GAIN control (6).
If either of the GAIN controls are re-adjusted, it may be necessary to adjust the Preamplifier's DC LEVEL control (3) to bring the pen back to the center of the chart.
Drift caused by the polarization of the electrodes can be overcome by selecting AC on the Preamplifier's AC/DC control (5).
Filtering can be introduced using the Preamplifier's FREQ CUT control (4).
If you are performing a 5-lead ECG you may select the various limb configurations using switch (2).
Calibrate the waveform,using the Oscillograph's the CAL 0.1 switch [Z].
This will cause the pen to move on the chart. The distance moved by the pen when this switch is operated, is equivalent to 0.1 V (100 mV). You can now calculate how many mV equals 1 mm of pen deflection. Let this equal d.
Measure the height of the waveform in mm. Let this value equal h. Let g equal the gain selected on control (6).
Therefore the waveform (mV) = d x h
g
9
Figure 5
PEN ZERO
GAIN
2
3
4
5
6
7
8
9
10
1
0
MONITOR
CAL. 0-1V.
DIRECT INPUT
D.C.
A.C.
DIRECT
INTERFACE
POLARITY
OFF
+
[W] [Z] [AA]
[AC]
Page 11
Recording Waveforms with the Harvard Student Oscillograph
Numbers in brackets, e.g. (1), (2) etc. refer to Fig. 1 on page 3. Letters in brackets, e.g. [A], [B] etc. refer to Fig. 6 on page 11 and Fig. 13 in the Student Oscillograph Instruction Manual.
1) Have you read the complete Operating Procedure section?
2) Check that the Isolated Preamplifier is fitted with batteries (see page 12).
3) Refer to Fig. 2, 3 or 4 on pages 5, 6 and 7 to apply the elec­trodes to your subject.
4) Connect the electrodes to the Isolated Preamplifier using either the 3-lead input cable or the ECG limb cable.
5) With reference to the Harvard Student Oscillograph Instruction manual section 3.4.0, follow steps 1 to 13 inclu­sive.
6) Connect the Isolated Preamplifier to the INTERFACE socket [S] on the Student Oscillograph.
7) Check that the battery indicator (1) on the Preamplifier is in the green zone. If it is not, refer to page 12.
8) Set the Preamplifier’s AC/DC switch (5) to the DC position and set the filter switch to 15 kHz.
9) Set the Preamplifier’s GAIN switch (6) to 10 and select posi­tion I on switch (2).
10) Set Oscillograph’s AC/DC switch [N] to DC, the chart speed control [L] to 2.5 and the chart drive lever [A] to ON.
11) Adjust the Oscillograph’s GAIN control [M] to about 1. The pen will move on the chart.
12) Adjust the Preamplifier’s DC LEVEL control (3) until the pen returns to its central position on the chart. At this point you should see a small waveform on the chart.
10
Page 12
The amplitude of the waveform can be increased or decreased by using the Oscillograph’s GAIN control [M] and the Preamplifier’s GAIN control (6).
If either of the GAIN controls are re-adjusted, it may be necessary to adjust the Preamplifier’s DC LEVEL control (3) to bring the pen back to the center of the chart.
Drift caused by the polarization of the electrodes can be overcome by selecting AC on the Preamplifier’s AC/DC control (5).
Filtering can be introduced using the Preamplifier’s FREQ CUT control (4).
If you are performing a 5-lead ECG you may select the various limb configurations using switch (2).
Calibrate the waveform, using the Oscillograph’s the CAL 0.1 switch [Q].
This will cause the pen to move on the chart. The distance moved by the pen when this switch is operated, is equivalent to 0.1 V (100 mV). You can now calculate how many mV equals 1 mm of pen deflection. Let this equal d.
Measure the height of the waveform in mm. Let this value equal h. Let g equal the gain selected on control (6).
Therefore the waveform (mV) = d x h
g
11
Figure 6
PEN ZERO
8
GAIN
2
3
4
5
6
7
9
10
1
0
CAL. 0-1V.
DIRECT INPUT
D.C.
A.C.
EVENT
INTERFACE
OFF
CHART SPEED
2
3
4
5
6
7
9
10
1
0
POWER
HARVARD
[L]
[Q]
[N]
[M]
Page 13
Problems
If “noise” is seen superimposed on the waveform, then check the following:
1) Have you sited the electrodes correctly? Check that the elec-
trodes are not sited on excessive fat. Check that they are not sited over large muscles, otherwise, EMG will also be displayed.
2) Check the earth or reference electrode.
3) Check that the correct leads are attached to the correct elec-
trode and correct socket on the Isolated Preamplifier.
4) If the Isolated Preamplifier is sited close to a main power lead,
then move one or the other or preferably both as far apart as possible.
Service Information
Fitting or Replacing Batteries
The type of batteries required are 9 volt, size PP3 (IEC designation 6LR61) of the alkaline type.
Locate the battery holders on the back panel of the amplifier (see Fig. 7 below). Lift the plastic covers. Remove spent batteries and connect new batteries to the connectors inside. Place the batteries inside the holders and close the lids.
12
Figure 7
Lift here to gain access to battery
Page 14
Customer Contact Addresses
Customers located in the U.K., U.S.A., CANADA and FRANCE
In the event of a system breakdown, please contact our local office
(address etc. below) stating the nature of the fault and the approxi­mate date of purchase, if possible.
Customers located in places other than listed above
In the event of a system breakdown, please contact our local agent or, if one is not available, contact our U.K. office direct (telephone number, fax number, telex number and address below) stating the nature of the fault and the approximate date of purchase, if possible.
ENGLAND
Harvard Apparatus Ltd. Telephone: (0732) 864 001 Fircroft Way Fax: (0732) 863 356 Edenbridge Telex: 95292 SRI G. Kent TN8 6HE ENGLAND
U.S.A.
Harvard Apparatus, Inc. Telephone: 508-893-8999 84 October Hill Road Toll Free: 800-272-2755 Holliston, MA 01746 USA Fax: 508-429-5732
CANADA
Harvard Apparatus Canada Telephone: 514-335-0792 6010 Vanden Abeele Street Toll Free: 800-361-1905 Saint-Laurent, Quebec Fax: 514-335-3482 CANADA H4S 1R9 Telex: 05-824064
FRANCE
Ealing S.A.R.L. Telephone: (1) 64 46 25 00 6 Avenue Des Andes Fax: (1) 64 46 94 38 Miniparc – Bât. 8 91952 LES ULIS Cédex FRANCE
13
Loading...