Hardy WS100 User Manual

HI WS100 CONFIGURATION SOFTWARE
OPERATION AND INSTALLATION
MANUAL
Corporate Headquarters
3860 Calle Fortunada San Diego, CA 92123-1825 Phone: (858) 278-2900 FAX: (858) 278-6700 Web-Site: http://www.hardyinst.com
Hardy Instruments Document Number: 0596-0241-01 REV A-1
Copyright November 1999 H ard y Inst ruments, Inc. All Rights Reserved . Printed in the U.S.A. (941028)
1-2

Table of Contents

Table of Contents
TABLE OF CONTENTS - - - - - - - - - - - - - - - - - - - - - -I
TABLE OF ILLUSTRATIONS - - - - - - - - - - - - - - - - - - -I
CHAPTER 1 - OVERVIEW - - - - - - - - - - - - - - - - - - - - -1-1
A Brief Description of Chapter 1 - - - - - - - - - - - - - - - - - -1-1 About Hardy Manuals - - - - - - - - - - - - - - - - - - - - - - -1-2 Description - - - - - - - - - - - - - - - - - - - - - - - - - - - - -1-2 C2® Calibration - - - - - - - - - - - - - - - - - - - - - - - - - -1-3 IT - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -1-3
Weighing System Tests - Optional - - - - - - - - - - - - - - -1-4
CHAPTER 2 - REQUIREMENTS - - - - - - - - - - - - - - - - - -2-1
General Introduction to Chapter 2 - - - - - - - - - - - - - - - - -2-1 Operating System Requirements - - - - - - - - - - - - - - - - - -2-1 Minimum Hardware Requirements - - - - - - - - - - - - - - - - -2-1 Software Requirements - - - - - - - - - - - - - - - - - - - - - -2-1 RSLogix OR EQUAL (FOR REMOTE MODE OPERATION ONLY) -2-1
CHAPTER 3 - INSTALLATION - - - - - - - - - - - - - - - - - - -3-1
Installing the Hardy Instruments Configuration Software - - - - - - -3-1 Installation of RSLOGIX - - - - - - - - - - - - - - - - - - - - - -3-1 Installation of RSLINX or RSLINX Lite - - - - - - - - - - - - - - -3-1
Wiring Diagram for the HI 215IT Junction Box - - - - - - - - -3-2
     
CHAPTER 4 - SETUP - - - - - - - - - - - - - - - - - - - - - - -4-1
A Brief Description of Chapter 4 - - - - - - - - - - - - - - - - - -4-1 Presetup Procedures - - - - - - - - - - - - - - - - - - - - - - - -4-1 Setup Procedures - - - - - - - - - - - - - - - - - - - - - - - - -4-3
Comm Setup for SLC Models (Except SLC 5/02) - - - - - - - -4-3 Comm Setup for the SLC 5/02 - - - - - - - - - - - - - - - - -4-5 Comm Setup for the HI 1746WS Remote Rack - - - - - - - - -4-7
  
i
HI WS100 CONFIGURATION SOFTWARE
CHAPTER 5 - OPERATING PROCEDURES - - - - - - - - - - - -5-1
A Brief Description of Chapter 5 - - - - - - - - - - - - - - - - - -5-1 Reading Live Weight, - - - - - - - - - - - - - - - - - - - - - - -5-1
Taring the HI 1746-WS - - - - - - - - - - - - - - - - - - - - -5-1 Zeroing HI 1746-WS - - - - - - - - - - - - - - - - - - - - - -5-2 Relay 1 and Relay Status - - - - - - - - - - - - - - - - - - -5-2 Setting the ROC Units - - - - - - - - - - - - - - - - - - - - -5-2 Errors - - - - - - - - - - - - - - - - - - - - - - - - - - - - -5-2
Calibration Procedures - - - - - - - - - - - - - - - - - - - - - - -5-2
C2 Calibration - - - - - - - - - - - - - - - - - - - - - - - - -5-2 Hard Calibration Procedures - - - - - - - - - - - - - - - - - -5-3
Soft Calibration - - - - - - - - - - - - - - - - - - - - - - - -5-4 Setting Parameters - - - - - - - - - - - - - - - - - - - - - - - - -5-5 Set Point Setup Procedures - - - - - - - - - - - - - - - - - - - -5-8
About Set Points - - - - - - - - - - - - - - - - - - - - - - - -5-8
Three General Rules for Set Points - - - - - - - - - - - - - - -5-8 Reading C2 Load Cell Serial Numbers - - - - - - - - - - - - - - -5-9 Remote Screen - - - - - - - - - - - - - - - - - - - - - - - - - -5-10
CHAPTER 6 - TROUBLESHOOTING - - - - - - - - - - - - - - -6-1
A Brief Description of Chapter 6 - - - - - - - - - - - - - - - - - -6-1 IT Troubleshooting - - - - - - - - - - - - - - - - - - - - - - - - -6-1
ii

Table of Illustrations

Table of Illustrations
CHAPTER 3 - INSTALLATION - - - - - - - - - - - - - - - - - - -3-1
FIG. 3-1 C2 LOAD CELL CONNECTION/HI 215IT
JUNCTION BOX - - - - - - - - - - - - - - - - - - - - -3-2
FIG. 3-2 NON C2 LOAD CELL CONNECTION/HI 215IT
JUNCTION BOX - - - - - - - - - - - - - - - - - - - - -3-3
CHAPTER 4 - SETUP - - - - - - - - - - - - - - - - - - - - - - -4-1
FIG. 4-1 INTRO PAGE - - - - - - - - - - - - - - - - - - - - - -4-3
FIG. 4-2 COMM SETUP - - - - - - - - - - - - - - - - - - - - - -4-4
FIG. 4-3 REMOTE SELECTED/ENTER FILE#, STARTING WORD -4-5
FIG. 4-4 SAMPLE OF REQUIRED LADDER LOGIC
FOR INTEGER TYPE FILES SLC 5/02 - - - - - - - - - -4-6 FIG. 4-5 CHECKING “USE INTEGER FILE. . . “/SLC 5/02 - - - - -4-7 FIG. 4-6 HI 1746WS REMOTE RACK LADDER LOGIC - - - - - -4-11
CHAPTER 5 - OPERATING PROCEDURES - - - - - - - - - - - -5-1
FIG. 5-1 LIVE WEIGHT PAGE - - - - - - - - - - - - - - - - - - -5-1 FIG. 5-2 C2 CALIBRATION - - - - - - - - - - - - - - - - - - - -5-3 FIG. 5-3 HARD CALIBRATION - - - - - - - - - - - - - - - - - -5-4 FIG. 5-4 SOFT CALIBRATION - - - - - - - - - - - - - - - - - - -5-5 FIG. 5-5 PARAMETERS SETTINGS - - - - - - - - - - - - - - - -5-6 FIG. 5-6 SETPOINT SCREEN - - - - - - - - - - - - - - - - - - -5-8 FIG. 5-7 C2 SERIAL NUMBER SCREEN - - - - - - - - - - - - -5-10 FIG. 5-8 REMOTE DISPLAY - - - - - - - - - - - - - - - - - - - -5-11
CHAPTER 6 - TROUBLESHOOTING - - - - - - - - - - - - - - -6-1
FIG. 6-1 INTEGRATED TECHNICIAN TEST - - - - - - - - - - - -6-1
I

CHAPTER 1 - OVERVIEW

Chapter 1 - Overview

A Brief Description of Chapter 1

This manual provides the user and service personnel with a description of the specifications, installation, setup, operation, communication procedures for the Hardy Instrument’s HI 100WS Configuration Soft­ware to configure the HI 1746-WS Weigh Scale Mod-
ule, designed for use in Allen-Bradley’s SLC 5/02
©
, 5/03, 5/04, 5/05 programmable controllers. Chapter 1 provides a general intro duction to the HI WS 100 Co n-
figuration Software. The HI 100WS Software is C2 and I
NTEGRATED TECHNICIAN
(IT)® enabled. The HI
100WS Software i s desi gned t o run on th e W i ndows
®
®
95/98/NT/2000 platforms in conjunction with Allen-
©
Bradley’s RS Linx gramming Software for Windows
or RS Linx Lite© Industrial Pro-
®
95/98/NT/2000. To get the maximum service from this product, users should operate this software in accordance with rec­ommended practices either implied or expressed in this manual. Before using the HI WS100 Software, all users and maintenance personnel should read and understand all cautions, warnings, and safety proce­dures, referenced or explicitly stated in this manual. Hardy Instruments appreciates your business. Should you not understand any information in this manual or experience any problems with the product, please contact our Customer Support Department at:
Phone: (858) 278-2900 FAX: (858) 278-6700 e-mail: support@hardyinst.com Web Address: hardyinst.com
NOTE: C2
marks of Hardy Instruments Inc. Windows
®
, I
NTEGRATED TECHNICIAN
®
are registered trade-
®
is a regis-
tered trademark of the Microsoft Corporation.
©
RSLinx
and RSLinx Lite© SLC 500©are registered
trademarks of the Rockwell Corporation.
NOTE: Hardy Instru m ent s bas es all procedures on the
assumption that th e us er has an adequate underst an d-
1-1
HI WS100 CONFIGURATION SOFTWARE
ing of all Allen-Bradley SLC 500 products. In addi­tion the user should understand process control and be able to interpret ladder logic instructions neces­sary to generate the electronic signals that control your application(s).

About Hardy Manuals Every Hardy Installation and Operation manual is

organized into easily referenced chapters, that are almost always the same:
Chapter 1 - Provides an introduction to the
instrument and an Overview of the equip­ment and its capabilities.
Chapter 2 - Provides a complete list of
Specifications.
Chapter 3 - Contains information needed to
install the HI 1746-WS (both standard and optional equipment) and the HI 215IT series Junction Box.
Chapter 4 - Provides complete hardware
Configuration instructions for setting dip switches and jumpers.
Chapter 5 - Pertains to the firmware/soft-
ware Setup and preparation procedures to calibrate and operate the instrument.
Chapter 6 - Provides all Calibration
instructions.
Chapter 7 - Pertains to the Operating Pro-
cedures of the HI 1746-WS.
Chapter 8 - Pertains to the Troubleshoot-
ing procedures for repair of the instrument.
Hardy Instruments hopes that this manual meets your needs for information and operation. All corrections or suggestions for improvements of this manual are welcome and can be sent to the Technical Publica­tions Department or Cust omer Supp ort De partment at Hardy Instruments Inc.

Description The HI WS100 Configuration Software is a program

that enables the user to configure a remote HI 1746­WS Weigh Scale Module. The HI 1746-WS Weigh Scale Module is a self contained microprocessor based I/O module with control inputs and outputs, that is designed to be eas ily plu gged int o the back plan e of
1-2
Chapter 1 - Overview
an Allen-Bra dley SLC 5/02, 5/03, 5/04 or 5/05 pro­grammable controller and/or Remote Rack. The HI 100WS is IT enabled which means you can perform the I
NTEGRATED TECHNICIAN
®
C2
Calibration C2 Second Generation Calibration enables a scale
(IT) Tests.
system to be calibrated electronically without using certified test weights which equal the systems load capacity. A C2 weighing system consists of up to eight (8) C2 load sensors, a junction box, interconnect cable and an instrument with C2 capabilities such as the HI 1746-WS Weight Scale Module. All Hardy Instruments C2 certified load sensors contain digital information detailing its unique performance charac­teristics. If the system is fitted with C2 type load cells the HI WS100 software displays the serial numbers programmed into each load cell. From the serial num­bers the operator can look up the parameters of each loadcell by going to the Hardy Instruments Internet Site. http://www.hardyinst.com
IT I
®
NTEGRATED TECHNICIAN
is a system diagnostics
utility which continuously monitors the weighing sys­tem (with up to 4 load sensors) for possible malfunc­tions. For full functionality the HI 215IT junction box (J Box) should be used. Full IT functionality allows the operator to rapidly troubleshoot a weighing sys­tem from the HI 100WS Configuration Software, Parameters Edit Display. The test are:
Loadcell Impedance - The user can see the
impedance determined at calibration (Base R), the impedance found during the test (Read R), and the test result (Good or Bad). also shown are the volts at the loadcells at calibration (Sense V), and the current going to the load cells at the time of the reading.
RTZ - (Return to Zero) This test gives a
pass/fail on whether there is a zero load. If you have a “J Box”, indi vidual load cell s can be tested. If you do not have a :”J Box” the test is done on the combined loadcells.
1-3
HI WS100 CONFIGURATION SOFTWARE
DVM - (Digital Volt Meter) If you have a “J
Box”, the voltage outputs from the individ­ual loadcells can be read. If you do not have a “J Box” the voltage is a combined voltage output from all loadcells in the system.
System Test - If you have a “J Box” you
can read the weight being read on individual load cells and the internal reference.

Weighing System Tests - Optional

Requires the HI 215IT Junction Box. This te st i s u sed to diagnose drifting or unstable weight reading prob­lems. The Weighing System Test does the following:
1. Disconnects the controller and engages an inter­nal reference signal to see if the problem is within the instrument.
1. Disconnects the load sensors and engages an internal (in the junction box) reference signal to see if the cable between the instrument and the Junction Box is causing the problem.
2. Reads the weight of each load sensor to s ee if the load sensor might be causing the problem.
The ability to read the weight seen by each individual load sensor allows use of this test to make cornering, leveling and load sharing adjustment s to the weighing system.
1-4

CHAPTER 2 - REQUIREMENTS

Chapter 2 - Requirements

General Introduction to Chapter 2

Operating System Requirements

Minimum Hardware Requirements

All the information in Chapter 2 pertains to the hard­ware and software requirements for HI WS100 Con­figuration Software. These requirements are the minimum requirements for loading and operating HI WS100 Configuration Software. Check to see if the personal computer on which you want to install HI WS100 Configuration Software, meets these require­ments. It is very important that the user be familiar with this section before installing or operating this software.
HI WS100 Configuration Software runs on the fol­lowing operating systems:
Windows NT 4.0/2000
Windows 95/98
HI WS100 Configuration Software requires the following hardware:
CPU - Pentium® 133 MHz or better
3 MB free space on your hard drive
•RAM - 16 MB
Monitors - SVGA, configured for 640 x 480 resolution or better
Floppy Drive - 3.5”, 1.44 MB drive or better
NOTE: The HI WS100 must run in conjunction with Allen
Bradley’s RS LINX or RS LINX LITE therefore the HI WS100 hardware requirements are in addition to hardware requirements for RS LINX or RS LINX LITE.

Software Requirements

RSLogix OR EQUAL (FOR REMOTE MODE OPERATION ONLY)

NOTE: RSLOGIX
Allen Bradley’s RS LINX ware.
Allen Bradley’s RSLOGIX
Corporation.
®
is a registered trademark of the Rockwell
®
or RS LINX® LITE soft-
®
software or equivalent.
2-1
HI WS100 CONFIGURATION SOFTWARE
2-2

CHAPTER 3 - INSTALLATION

Chapter 3 - Installation

Installing the Hardy Instruments Configuration Software

Contact your local Hardy Representative or Hardy Instruments Sales Department for information about how to purchase the Hard y WS 100 Configuration Software for Windows.
Step 1. The SETUP program copies the required
files to your hard disk.
Step 2. Run Windows NT/95/98/2000 and insert
the Configuration Installation System Disk
into the 3.5” floppy drive. Step 3. Click on START. Step 4. Click on Run. Step 5. In the Run field type the command:
a:setup. If your floppy drive is drive b
enter: b:setup. Step 6. Press the Enter key. Step 7. You can also use the Add/Remove Func-
tion in the Control Panel dialogue box.
1 Click on Start
1 Move the cursor to Settings.
1 Select Control Panel.

Installation of RSLOGIX

Installation of RSLINX or RSLINX Lite

1 Double click on the Add/Remove Programs
icon.
1 Click on the Install button.
Step 8. The SETUP program will lead you
through the installation pro cess . Please refer to your Allen Bradley Installation and
Operation manual for instructions. Please refer to your Allen Bradley Installation and
Operation manual for instructions.
3-1
HI WS100 CONFIGURATION SOFTWARE

Wiring Diagram for the HI 215IT Junction Box

C2 Loadcell Cable Connection
When connecting the HI 215IT Junction box using C2 loadcell/point cable (6020-0001) use the following color code:
(J1)
Model EXC + SEN + SIG + SIG - SEN - EXC - C2 + C2 -
J-BOX RED B LUE GRN WHT BRWN BLK GREY VIO
Table 1: C2 CABLE COLOR CODE/HI 215IT J-BOX
FIG. 3-1 C2 LOAD CELL CONNECTION/HI 215IT JUNCTION BOX
1 Recommended load cell cable, Hardy Instru-
ments (Prt. #6020-0001)
2 Attach the load shield to the terminal block
mounted next to the J1 connector on the HI 200DNWM Module.
3-2
Chapter 3 - Installation
3 Do not run load cell cable in parallel with or
in the same conduit with power w iring, relay cables or any other high energy cables.
4 Remove the factory installed jumpers for C2
wire load cell connection.
5 IT summing Card (Prt. #0535-0465)
Non-C2 Loadcell Cable Connection (J1)
1 Attach the load cell cable to the terminal
block mounted next to the J1 connector on the HI 200DNWM Module.
2 Factory installed jumpers to remain in place
for 4 wire load cell connection.
3 Do not run load cell cable in parallel with or
in the same conduit with power w iring, relay cables or any other high energy cables.
4 Eight (8) conductor load cell cable required
when:
Cable runs are greater than 50 feet.
With Soft Calibration
FIG. 3-2 NON C2 LOAD CELL CONNECTION/HI 215IT JUNCTION BOX
3-3
HI WS100 CONFIGURATION SOFTWARE
3-4

CHAPTER 4 - SETUP

Chapter 4 - Setup

A Brief Description of Chapter 4

Presetup Procedures Step 1. Make sure that the HI 1746-WS is

All information contained in Chapter 4 pertains to firmware and software settings to prepare the HI 1746WS Weight Module controller for calibration and operation. Alternatives to these procedures explicit or implied, contained in this section are not recommended. It is very important that the user and service personnel be familiar with the pro cedures co n­tained in this chapter, before going through the setup procedures. The Setup procedures require either Allen-Bradley DOS PLC 5 00 AI Se ries Ladder Logic Editor*, or Allen-Bradley RSLogix 500* and Allen­Bradley RSLinx™ or RSLinx™ Lite.
installed in the rack correctly. Step 2. Make sure that the HI 1746-WS has been
setup correctly and that all data necessary
to run the HI 1746-WS module has been
entered into RSLogix 500 configuration
software. Step 3. Check the module ID code to verify that
the jumper has been set for the proper
mode of operation (Local ID Code =
13635, Remote ID Code = 3235). If you
have put the HI 1746-WS module in a
local rack and get the 3235 ID number the
jumper is set incorrectly. Step 4. If you have set the jumper for the HI 1746-
WS module for local mode of operation
and placed the module in a rem ote rack the
1747 ASB will display an invalid module
error. Step 5. If these errors occur, remove the HI 1746-
WS and reset the jumper to the proper
mode of operation. (See the HI 1746-WS
Weigh Module Installation and Operation
Manual (Prt. #0596-0234-01), Chapter 3,
Page 3-2 for instructions) Step 6. Reinstall the module. Step 7. Repeat Steps 3 & 4.
4-1
HI WS100 CONFIGURATION SOFTWARE
NOTE: If you are using RSLogix 500, the HI 1746-WS (in
local mode) is configured as an “Other” card with an I/O module ID code of 13 635, M0 and M1 le ngth s are
128.
NOTE: If you are using RSLogix 500, the HI 1746-WS (in
remote mode) is configured as an “Other” card with an I/O module ID code of 3235, with two input words and two output words. In remote mode the HI 1746­WS has no M files. Instead, it has files inside that cor­respond to the 32 word I and O files of the local mode HI 1746-WS, and to its M files. These files are called “logical” files, meaning that they perform the func­tion of the I,O, and M files in the local mode but are not directly accessible through the SLC backplane.
Step 8. Double click on the HI WS100 Icon on
your Windows Desktop. The HI WS100 will automatically launch RSLinx.
NOTE: If you are st arting HI WS10 0 for the first time you will
get a “Communi cation Failure” message. Don’t worry about this message at this time. The Comm Setup has not been configured yet. Once the HI WS100 has been configured correctly the “Communi­cation Failure” message goes away.
4-2
Step 9. The Intro Sheet displays with the setup
instructions in a text field. (See Fig. 4-1) Scroll through the Text field and read the instructions for setting up the HI 100WS Configuration Softwar e before star ting th e setup procedures.

Setup Procedures

Chapter 4 - Setup
FIG. 4-1 INTRO PAGE

Comm Setup for SLC Models (Except SLC 5/02)

Step 1. Click on the Comm Setup Tab. The Comm
Setup Page appears. (See Fig 4-2)
4-3
HI WS100 CONFIGURATION SOFTWARE
FIG. 4-2 COMM SETUP
Step 2. Enter the SLC slot number of the installed
HI 1746-WS Weigh Module in the Slot textfield.
Step 3. Enter the Station number of your SLC in
the Station textfield. This number must be an Octal number for the Station number.
Step 4. Select a driver (supplied by RSLinx) by
clicking on the down arrow on the “Driver” pull down menu. Click on the driver you want to use to communicate with the HI 1746-WS.
CAUTION:
4-4
DO NOT CLICK IN THE CHECK BOX NEXT TO THE “USE DTSA.TXT FILE”.
Step 5. If the HI 1746-WS module is installed in a
remote rack, click in the check box next to the word “Remote?”. (See Fig. 4-3)
Chapter 4 - Setup
FIG. 4-3 REMOTE SELECTED/ENTER FILE#, STARTING WORD
Step 6. Enter the integer number used in the Lad-
der Logic for remote operation of the HI
1746-WS Weigh Module in the textfield
next to “File#”. In our example we entered
the number 7. Step 7. Enter the starting number us in the Ladder
Logic for remote operation of the HI 1746-
WS Weigh Module in the textfield next to
“starting word”. Step 8. Click on the Save button to save the set-
tings.

Comm Setup for the SLC 5/02

Normally the HI WS100 talks to the HI 1746-WS via the “M Files”. The SLC 5/ 02 does not su pport com­munication commands for direct reading and writing of M Files. The SLC 5/02 suppor ts comm ands to read and write “integer” type files.
Step 1. Write the ladder logic provided (See Fig.
4-4) to copy the integer file to the HI
12746-WAS’s M files. The ladder logic
4-5
HI WS100 CONFIGURATION SOFTWARE
enables the HI 1746-WS to read the infor­mation in the integer file.
FIG. 4-4 SAMPLE OF REQUIRED LADDER LOGIC FOR INTEGER TYPE
FILES SLC 5/02
NOTE: Our example has the HI 1746-WS weigh module
located in slot #3 of the SLC and the integer file #9 (N9) is chosen with the starting word = 0.
Step 2. Click in the check box next to the “Use
integer file (required by 5/02, needs ladder logic support)”. (See Fig. 4-5) The pro­gram will write to the specified integer file number (7,8,9, etc. for N7, N8, N9,. . .) starting at the specified offset (starting word) within that file. The program
4-6
Chapter 4 - Setup
requires a total of 43 integer words: 21 for
the M0 file, 21 for the M1 file, and 1 as a
flag.
FIG. 4-5 CHECKING “USE INTEGER FILE. . . “/SLC 5/02

Comm Setup for the HI 1746WS Remote Rack

Legend
The ladder logic example below must be running so that the WS100 Configuration Software will work with the 1746 module installed in a Remote Rack. The addresses shown in the ladder logic example are arbi­trary locations. If the file is added to an existing pro­gram, these address locations can be changed for the system to insure there are no conflicts. The color cod­ing of the locations is as follows:
Green: Slot # for scanner module in local
rack and portion of the M file to be used.
Yellow: Integer files used for data defining
the M file, (read/write, length, address). Also status/error bits per­taining to the transfers of data.
4-7
HI WS100 CONFIGURATION SOFTWARE
Pink: Integer file specified in the WS 100
program COMM SETUP screen that it will use to read/write data. The WS 100 Configuration Pro­gram also specifies the starting word in this file. The WS 100 will 17 words starting at the specified word of th e file. The first word is used as a trigger to start the write process. Words 2-9 are the write data, and words 10-17 are the read data.
No Color: Status bits used for pro-
gram control.
4-8
Chapter 4 - Setup
4-9
HI WS100 CONFIGURATION SOFTWARE
4-10
Chapter 4 - Setup
FIG. 4-6 HI 1746WS REMOTE RACK LADDER LOGIC
4-11
HI WS100 CONFIGURATION SOFTWARE
4-12

Chapter 5 - Operating Procedures

CHAPTER 5 - OPERATING PROCEDURES

A Brief Description of Chapter 5

NOTE: The HI WS100 will not read non-C2 load cell serial

Reading Live Weight, Step 1. Click on the Live Weight Tab. The Live

All information contained in Chapter 5 pertains to the operation of the HI WS100 Configurat ion Software. The Operating Procedures include, Reading Live Weight, Calibration, Creating or modifying Set­point(s) Parameters, reading C2 load cell serial num­bers and quickly monitoring Gross Wei ght (remot e modules only) for the HI 1746-WS Weigh Module.
numbers.
Weight page appears. (See Fig. 5-1) The
Gross, Net and Tare weights are displayed.
In addition the Rate of Change (ROC) and
the ROC Units are displayed. Relay 1 and
Relay 2 status are displayed.
FIG. 5-1 LIVE WEIGHT PAGE
Taring the HI 1746-WS Step 2. To Tare the HI 1746-WS click on the Tare
button.
5-1
HI WS100 CONFIGURATION SOFTWARE
Zeroing HI 1746­WS

Relay 1 and Relay Status

Setting the ROC Units

Errors Step 6. The lights next to Motion, Excite Error, A/

Calibration Procedures

Step 3. To Zero the HI 17 46-WS click o n the Zero
button.
Step 4. The Relay 1 and Relay 2 status fields dis-
play only two values:
Low = O voltage output
High =5 volts output.
Step 5. Click on the radio button next to the unit
you want to use for the Rate of Change. (Default is Minute)
D Conversion Error and A/D Convertor Failure should all be green. A red light indicates an error in the weight readings.
Step 1. Click on the Calibrate Tab. The Calibra-
tion page appears. (See Fig. 5-2)
Step 2. HI WS100 gives you three choices for
Calibration:

C2 Calibration

Hard Cali bration
Soft Calibration
C2 Calibration Step 3. Click in the radio button next to C2. The
C2 Cal parameters appear. (See Fig. 5-2)
Step 4. Follow the instructions in the Text Area
Field for C2 Calibration. For additional information consult your HI 1746-WS manual (Prt. #0596-0234-01) Chapter 5 ­Calibration.
5-2
Chapter 5 - Operating Procedures
FIG. 5-2 C2 CALIBRATION

Hard Calibration Procedures

Step 1. Click on the radio button n ext to Hard Cal-
ibration. The Hard Calibration parameters
appear. (See Fig. 5-3) Step 2. Follow the instructions in the Text Area
Field for Hard Calibration. For additional
information consult your HI 1746-WS
manual (Prt. #0596-0234-01) Chapter 5 -
Calibration.
5-3
HI WS100 CONFIGURATION SOFTWARE
FIG. 5-3 HARD CALIBRATION

Soft Calibration Step 1. Click on the radio button next to Soft Cali-

bration. The Soft Calibration parameters appear. (See Fig. 5-4)
Step 2. Follow the instructions in the Text Area
Field for Soft Calibration. For additional information consult your HI 1746-WS manual (Prt. #0596-0234-01) Chapter 5 ­Calibration.
5-4
Chapter 5 - Operating Procedures
FIG. 5-4 SOFT CALIBRATION

Setting Parameters Step 1. Click on the Parameter Tab. The Parame-

ter page appears. (See Fig. 5-5) Step 2. Set the Tare weight - Double click in the
Tare Weight field and enter the Tare
Weight Value. The Tare Value is an artifi-
cial zeroing of the weight hopper so that a
new weight can be displayed. Also, the
action of ad j usting out the known weight
of the container from the total indicated
weight, so that the indicator reads net
weight directly. You must click in the
check box next to Tare Enabled to enable
the Tare Function. Step 3. Set the Weight Unit of measure - The
Weight Unit of Measure can be set to
either kilograms or pounds. Any weight
value input to the module (e.g. CAL-LO,
CAL-HI setpoints) are in the currently
selected units. The unit of measure can be
set at any time, not just at calibration. Set-
ting the unit of measure before calibrating
reminds the user what unit of measure is
5-5
HI WS100 CONFIGURATION SOFTWARE
being displayed. It is important to note that the weigh scale module does not need to be calibrated again after changing the unit of measure.
Step 4. Set the WAVERSAVER level - There are
5 selectable levels. 0 provides the least vibration immunity with the fastest response time. 4 provides the most vibra­tion immunity with the slowest response time. The Default setting is 2. Click on the pull down menu next to WAVERSAVER and click on your selection.
5-6
FIG. 5-5 PARAMETERS SETTINGS
Step 5. Set the Number of Averages (NumAver-
ages) - Click on the up or down arrow to
reach the value you want or double click in the field and enter the value. The range is 1-255. The Number of Averages sets the number of weight readings which will be used to compute the displayed weight. Th e average is a sliding average so that a new average reading is available for display at every reading.
Chapter 5 - Operating Procedures
Step 6. Set the ROCTimeBase - Click on the up
or down arrow to reach the value you want
or double click in the field and enter the
value. The Rate of Change Time Base is
the time period you want for weight sam-
pling to determine the Rate of Change,
either loss in weight or gain in weight. The
range is 1 - 1800 seconds. Step 7. Set the Zero Tolerance - Double click in
the Zero T olerance field. The default value
is 10 units of measure. Ent er th e new Zer o
Tolerance value. The range is 1 to 32766.
Zero Tolerance sets the range of weights
so that the Zero Command works, as an
offset of the calibrate d Zero. Zero Toler-
ance is the number of graduations from
zero that will be accepted as zero by the
module. To enable Zero Tolerance click in
the check box next to “Zero Enabled” so
that a check mark is visible. The operator
can reset this value at any time. Step 8. Set the Auto Zero Tolerance - The
default value is 10 units of measure. The
range is 1 to 32766.
Double click in the Auto Zero Tol­erance field.
Enter a positive value that does not exceed the maximum zero tolerance value or is not a negative value.
Click in the check box next to ”Zero Track Enabled” to activate Auto Zero Tolerance tracking.
The Auto Zero Tolerance setting automatically sets the Zero Toler­ance so that the Zero Command works, as an offset of the calibrated Zero.
Step 9. Set the Motion Tolerance Value - Click
in the Motion T olerance field and enter the motion value. The motion tolerance is the tolerance value used to determine if the scale is in motion. If the motion is outside
5-7
HI WS100 CONFIGURATION SOFTWARE
the motion value entered, the light next to Motion Tolerance will turn to red. When the scale motion is within the tolerance the light turns green.
Step 10. Click on the Save button to save the new
parameters.

Set Point Setup Procedures

About Set Points The set point value is the target we ight or level. It may

be set in net, gross or Rate of Change (ROC). When the set point is reached a relay is energized or de-ener­gized depending what you want the relays to do. The HI 1746-WS has two output relays which can be s etup in HI WS100 Configuration Software.
Step 1. Click on the Setpoint Tab. The Setpoint
page appears. (See Fig. 5-6)

Three General Rules for Set Points

5-8
FIG. 5-6 SETPOINT SCREEN
1. Set points activate at the set point plus the preact.
2. Set points deactivate at the set point plus the deadband.
Chapter 5 - Operating Procedures
3. The deadband should be numerically larger than
the preact to prevent relay chatter.
Step 2. Double click in the Value field. Step 3. Enter the desir ed set point value. Step 4. Double click in the Preact field. Step 5. Enter the desired Preact value. Preact is
the number of units above or below the set point value of which the relay will trip.
Use as an “in flight” compensation value. Step 6. Double click in the Deadband field. Step 7. Enter the desired Deadband value. Dead-
band is a value used to prevent relay chat-
ter once the setpoint is reached. You can
use positive or negative values. The nega-
tive value can be used as a low trip limit,
the positive value can be used as a high
trip limit. Step 8. Click in the radio button next to the Set-
point Mode you want. Step 9. Click in the radio button next to the Value
units you want to read the weight. Gross,
Net, or ROC. Step 10. Click on the Save button to save the set-
tings.

Reading C2 Load Cell Serial Numbers

Step 1. Click on the C2 Tab. The C2 page
appears. (See Fig. 5-7) Step 2. Go to the Hardy Internet Site (http://
www.hardyinst.com) where you can look
up the parameters of your load cell(s).
5-9
HI WS100 CONFIGURATION SOFTWARE
FIG. 5-7 C2 SERIAL NUMBER SCREEN

Remote Screen Step 1. Click on the Remote Tab. The Remote

page appears. (See Fig. 5-8)
NOTE: The Remote Display will only appear if the HI 1746-
WS module is installed in a remote chassis, with the jumper set to the remote mode of operation.
Step 2. The Remote screen is a quick way to mon-
itor gross weight of an HI 1746-WS mod­ule, installed in a remote chassis and operating in the remote mode. This screen gives the user a faster Gross Weight read­ing than the Live Weight display.
Step 3. The indicator lights for Motion, and A/D
Conversion and A/D Converter should all be green in normal operation. A motion red light means the scale exceeds the Motion To lerance setting. A/D Conversion and A/D Converter red lights means a mal­function.
5-10
Chapter 5 - Operating Procedures
FIG. 5-8 REMOTE DISPLAY
5-11
HI WS100 CONFIGURATION SOFTWARE
5-12

CHAPTER 6 - TROUBLESHOOTING

Chapter 6 - Troubleshooting

A Brief Description of Chapter 6

Chapter 6 provides instructions on how to use Hardy Instruments I
NTEGRATED TECHNICIAN
™ to perform various tests for fault isolation troubleshooting of the load cells in your weighing system. I T
ECHNICIAN
used in conjunction with the HI 2215IT
NTEGRATED
Junction Box can tell you which load cell is not func­tioning properly, thereby eliminating the necessity of having to physically go to the load cell for testing. You can see from your PC desktop which cell is not working properly.

IT Troubleshooting Step 1. Click on the Integrated Technician tab. the

Integrated Technician page appears. (See Fig. 6-1)
FIG. 6-1 INTEGRATED TECHNICIAN TEST
Step 2. In the “Number of Sensors (JBox only)”
field you will see an up and down arrow. A few things need to be said about this field.
If you have all C2 load cells and an HI 215IT Junction Box, w hen y ou cal ibrate
6-1
HI WS100 CONFIGURATION SOFTWARE
the instrument, HI WS100 automatically loads the number of C2 load cells that are incorporated in your weight system. The number will automatically display in the Number of Sensors field. Having the HI 215IT Junction Box enables the configuration software to list the test results for each individual load cell in your weighing system.
If you have non-C2 load cells and an Hi 215JT Junction Box, then you need to add the number of load cells in your sys­tem by clicking on the up or down arrows or double clicking in the field and typing the number, to enter the number of load cells in your weight sys­tem. Having the HI 215IT Junction Box enables the configuration software to list the test results for each individual load cell in your weighing system.
If you do not have the HI 215IT Junc­tion Box, the configuration software will give you a combined test result of all the load cells in your weighing system. (See Fig. 6-1)
Step 3. Click on the Test b utton. After a seco nd or
two the results of the test will appear in the fields. The tests are:
1. Loadcell Impedance
Base R is the impedance at the time of the last Calibration.
Read R is the impedance reported during the test.
Test R indicates whether the cell is good or bad.
NOTE: Keep in mind that if you have the HI 215IT Junction
Box the Tes t R result will appear for each load cell. Therefore you will know exactly which load cell failed. If you do not have an HI 215I T Jun c ti on i t wi ll
6-2
Chapter 6 - Troubleshooting
give a combined fail result but not for the specific load cell that failed. You will still have to determine which Load Cell failed.
2. Return to Zero Test (RTZ) - This test returns a pass/fail result on whether there is a zero load.
If you have an HI 215IT Junct ion
Box, individual load cells can be tested.
If you do not have an HI 215IT
Junction box the result will be for the combined load cells in your weighing syst em.
3. Digital Volt Meter (DVM) - This test returns the voltage outputs from the load cells.
If you have an HI 215IT Junct ion
Box, the voltage output for each individual load cell is read.
If you do not have an HI 215IT
Junction box the voltage output read will be for the combined load cells in your weighing sys­tem.
4. System Test - Reads the weight for the load cells in your system.
If you have an HI 215IT Junct ion
Box, the weight for each individ­ual load cell can be read as well as the internal reference.
If you do not have an HI 215IT
Junction box the weight read will be combined weight for all the load cells in your weighing sys­tem.
6-3
Loading...