5.3.1 Configure a Sine Wave ........................................................................................... 93
5.3.2 Configure a Square Wave ...................................................................................... 94
5.3.3 Configure a Ramp Wave ........................................................................................ 94
5.3.4 Configure a Pulse Wave ......................................................................................... 95
Appendix A ..................................................................................................................................... 96
Appendix B ................................................................................................................................... 103
HDG2000 Series Arbitrary Waveform Generator ii
Copyright Declaration
Copyright Declaration
All rights reserved; no part of this document may be reproduced or transmitted in any form or by
any means, electronic or mechanical, without prior written permission from Hantek Technologies
Co., Ltd (hereinafter referred to as ‘Hantek’).
Hantek reserves all rights to modify this document without prior notice. Please contact Hantek for
the latest version of this document before placing an order.
Hantek has made every effort to ensure the accuracy of this document but does not guarantee the
absence of errors. Moreover, Hantek assumes no responsibility in obtaining permission and
authorization of any third party patent, copyright or product involved in relation to the use of this
document.
HDG2000 Series Arbitrary Waveform Generator iii
General Safety Summary
General Safety Summary
Read the following safety precautions to avoid injury and prevent damage to this product or any
products connected to it. To evade potential hazards, use this product only as specified.
Only qualified personnel should perform maintenance.
Avoid fire or personal injury.
Use suitable power cord. Use only the power cord specified for this product and certified for the
country of use.
Ground the product. This product is provided with protective earth terminals. To minimize shock
hazard, the instrument must be connected to the AC power mains through a grounded power
cable, with the ground wire firmly connected to an electrical ground (safety ground) at the power
outlet. Any interruption of the protective (grounding) conductor or disconnection of the protective
earth terminal will cause a potential shock hazard that could result in personal injury.
Check all terminal ratings. To avoid fire or shock hazard, check all ratings and markings on the
product. Refer to the product manual for detailed information about ratings before making
connections to the product.
Do not operate without covers. Do not operate this product with covers or panels removed.
Avoid exposed circuitry. Do not touch exposed connections and components when power is
present.
Do not operate with suspected failures. If you suspect there is damage to this product, have it
inspected by qualified service personnel.
Assure good ventilation.
Do not operate in wet/damp environments.
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry.
7 inch, 16K true color TFT display, WVGA(800X480)
16 channels digital output, together with the analog channel can rebuild the more mixed signals
in daily practice
TCXO timebase standard, OCXO optional for ultra-high stability
Plenty of interfaces: USB Host, USB Device. LAN optional
Support AM, FM, PM, ASK, FSK, PSK, BPSK and PWM modulations
LXI Class C Compliant
SCPI (Standard Commands for Programmable Instruments) compatibility
Online Help
SD card
Models
This section describes the models about the HDG2000 Series of instruments.
Front Panel can be divided into "F0 to F6 softkeys menu area", "digital keypad area", "Function
keypad area", "Hot keypad area", "Mode keypad area", "Trigger keypad area", "Waveform Output
area", "knob and cursor arrows keypad area", and “interface of the signal and peripheral devices area”. You can press keys and buttons on the front panel to input or output signals.
HDG2000 Series Arbitrary Waveform Generator 7
Basic Operation
1 2 4
5
6
7
12
14
15
16
17
8
1. Menu softkeys F0-F6
2. Function waveform button
3. Channel 1 and Channel 2 Switch Button
4. Function menu
5. Knob and cursor arrows
6. Numeric Keypad
7. Channel 1 and Channel 2 output buttons
8. Trigger button
9. Channel 1 and Channel 2 output Connector
10. Sync Connector
11. Modulation signal Input Connector (for all kinds of External Modulation Signal Source)
12. High frequency Connector(24MHz to 2.7GHz) HDG2000C
13. Low frequency (24MHz or lower)/External trigger Connector
14. Digital channel
15. USB port
16. SD card port
17. On/Off Switch
Note: Press and hold any front-panel key or softkey to get context-sensitive help.
Front-Panel Number Entry
You can enter numbers from the front panel in two ways:
Use the knob and cursor keys to modify the number. Rotate the knob to change a digit (clockwise
increases). The arrows below the knob move the cursor.
HDG2000 Series Arbitrary Waveform Generator 8
Basic Operation
Use the keypad to enter numbers and the softkeys to select units. The +/- key changes the
number's sign.
CE: Delete Key
BASP: Backspace, backspace and remove the last digit
ENT: Enter Key
Warning: For protection from electrical shock, the power cord ground must not be defeated. If
only a two-contact electrical outlet is available, connect the instrument’s chassis ground screw to a
good earth ground.
Online Help
Signal generator online help provides two kinds of view. One is content view and the other is index
view.
1. The index view is the subject of the online help content according to the ascending order of the
view. Choose index by knob, enter to go into the subject.
2. Content view shows a view of specific topic content.go into the content linked by the "link
words".
3. Press ‘Exit’ menu softkey to exit.
Screen Area
The signal generator can be divided into three regions: the status panel at the top of the screen,
the menu panel at the right of the screen, the window area at the left corner of the screen.
HDG2000 Series Arbitrary Waveform Generator 9
Basic Operation
Waveform
parameters
Softkey
labels
CH1
Information
States Panel
CH2
Information
System Time
Waveform
display
Status panel indicates whether external usb device is connected with this device, whether usb
device in the rear panel is connected with PC, whether the Internet is connected and the current
time.
The menu panel and window area are corresponding. The content of the window area can be
modified by controlling the menu panel. There are three main ways of style menu panel item.
One way is operation directly. As is shown, press it, then the operation will be modified, or enter
submenu to select, or page to next page to select.
One way is for choosing. It provides a topic and some params. Topic is showed at the top of the
menu item and params showed below. Press the Softkeys F0~F6 corresponded to the item to
choose different params.
The other way is altering. It appears like choosing. But it don’t have topic item, current params is
showed on the front of the menu item and the optional item is at below.
Use F6 button to page down or page up. Press F0 to return the parent menu.
1.3 Rear Panel
1. External 10 MHz Reference Input
2. Internal 10 MHz Reference Output
3. USB Interface Connector, communication to PC software
4. LAN Interface Connector, communication to PC software (optional)
HDG2000 Series Arbitrary Waveform Generator 10
Basic Operation
1.4 Prepare Instrument for Use
This section describes basic procedures to help you get started quickly with the instrument.
Prepare Instrument for Use
Connect the power cord as desired. Turn the instrument on by pressing the power switch in the
lower left corner of front panel.
The instrument's default function is a 1 kHz, 200 mVpp sine wave. At power-on, the channel
output connectors are disabled. To enable output on a channel connector, press the key above the
channel connector and then press the Output button.
If the instrument does not turn on, verify that the power cord is firmly connected (power-line
voltage is automatically sensed at power-on). Also make sure that the instrument is connected to
an energized power source.
Power Switch:
To turn off the instrument, please press power switch.
Attention: After turning on, the process of system initialization will take 1~2 minutes; if you turn it
off before the initialization is done, the next time you turn it on, it will prompt that the system is not
properly shut-down. This is necessary to have the system restorable when it collapses, like the
"safe mode" of Windows.
First, press F1, the device will be restart.
If not, press F2 to recover the system, the system will be recovered to the factory default setting.
If not, contact the customer service to get new recover partition program, and press F3 to update
the recover partition.
HDG2000 Series Arbitrary Waveform Generator 11
Chapter 2 Basic Operation
Front-Panel Menu Reference
Utility Function
Save and Recall
Remote Control
Basic Operation
HDG2000 Series Arbitrary Waveform Generator 12
Basic Operation
Sine
Square
Ramp
Pulse
Noise
Arb
Harmonic
CH1and CH2 switch button
Frequency/Time hot key
Amplitude hot key
Offset hot key
2.1 Front-Panel Menu Reference
Overview of the front-panel menus. The remainder of this chapter contains examples of using the
front-panel menus.
Configures waveform-specific parameters
Period/Frequency
Amplitude or High and Low Voltage
Offset
Phase
Duty Cycle
Symmetry
Pulse Width
Edge Time
Arbitrary Waveform
Bandwidth
HDG2000 Series Arbitrary Waveform Generator 13
Specifies unit and parameter preferences
Frequency or Period
Voltage as Amplitude/Offset or High/Low
Voltage units
Pulse Width or Duty Cycle
Frequency sweep as Center/Span or Start/Stop
Mode:
Configures modulation parameters
Modulation on or off
Modulation type: AM, FM, PM, 2ASK, 2FSK, 2PSK, BPSK and PWM
Modulation source
Modulation parameters
Basic Operation
Sweep:
Configures frequency sweep parameters
Sweep on or off
Linear
Sweep time
Start/stop frequencies or center/span frequencies
Hold and return times
Burst:
Configures burst parameters
Burst on or off
Burst mode: triggered or externally-gated
Cycles per burst
Burst period
Save/Recall:
Stores and recalls instrument states
Help:
HDG2000 Series Arbitrary Waveform Generator 14
Shows list of Help topics
Display help topic list
Display help index list
Get help on any key
Utility:
1. System setting
Basic Operation
Configures system-related parameters
Language setting: Select local language for front-panel messages and help text
Display setting: Enable or disable screen saver; Adjust display brightness
Sound setting: Enable or disable beeper
Network setting: Set network
Time and Date setting: Set date and time
Clock source setting: Select clock source internal or external.
2. System Information
Device Information
Hardware information
Software information
Law information
3. SysStatus
Check system status
Save and recall system status
4. Update
Update firmware
5. Counter
Turn on/off counter or frequency meter
HDG2000 Series Arbitrary Waveform Generator 15
6. Self Test
Perform self test
7. Calibration
Perform calibration
8. Digital Generator
Digital Generator
Trigger:
Configures trigger settings
Perform a manual trigger, when illuminated.
Specify the trigger source for sweep, burst or arbitrary waveform advance.
Specify the trigger count and delay.
Basic Operation
Specify the slope (rising or falling edge) for an external trigger source.
Specify the slope (rising or falling edge) of the trigger output signal.
Enable / disable the signal output from the "Sync" connector.
Default setting:
Default Settings
Press the Default button in the front panel, and the factory status will be recalled.
2.2 Utility
There are some features to be selected in Utility menu, such as: <System Settings>, <System
Information>, <System State >, <Counter>, <Self Test>, <Calibration>.
To choose these features, use the knob and left/right direct button, then push ENT button to
confirm.Also you can push F1 and F2 button to choose function, then press ENT button or F4
button to confirm. Press F5 button to exit.
2.2.1 System Settings
The followings are settings:
Language Settings
Display Settings
HDG2000 Series Arbitrary Waveform Generator 16
Basic Operation
Sound Settings
Network Settings
Time Settings
Reference Source Settings
Please follow the steps below to enter each setting:
1. Press <Utility> button, select “System Setting”, then press ENT button to do System Settings;
2. In the main menu of System Settings, push F1 to F5 button to go into desired settings
3. Also, user can choose these settings by the konb key.
Language Settings
English and Chinese Language are supported; Please follow the steps below to change current
language:
1. Press<Utility> button, select “System Setting”, then press ENT button to go into System
Settings;
2. Choose “Language Settings”, press ENT button to open Language Settings interface;
3. Push F1 button to choose different language;
4. Push F2 button to exit.
Display Settings
Adjust display brightness.
1. Press<Utility> button, and press ENT button to go into System Settings, then choose <Display>;
2. To adjust brightness, please press F1 button, then the menu is highlighted, then rotate the knob
on front panel to adjust the parameter values.
Disable or enable saver:
1. Press<Utility> button, and press ENT button to go into System Settings, then choose <Display>;
3. Push F2 button to enable or disable the saver;
4. When the saver is enable, push F3 button to determine after what time, the saver will be started;
5. Push F4 button to exit;
Screenshot function:
1. Press<Utility>button, and press ENT button to go into System Settings, then choose <Display>;
2. Press F4 button to set <Default> key to <Print>.
3. Then press <Default> to save the screenshot to external USB device.
Sound Settings
If turn on keypad tone, we can hear sound when press a button.
HDG2000 Series Arbitrary Waveform Generator 17
Basic Operation
Please follow the steps below to enable or disable sound:
1. Press<Utility> button, and press ENT button to go into System Settings, then choose “Sound
Settings”;
2. Push F1 to enable or disable beep.
3. Press F2 to exit.
Network Settings
The system can communicate with PC by Ethernet interface.
1. Press<Utility> button, and press ENT button to go into System Settings, then choose “Network Settings”, press ENT button to open “Network Settings” interface;
2. Press F1 to change IP address, input right format IP values by using digital keypad in the front
panel.
4. Press F2 to change subnet mask, input right format subnet mask values by using digital keypad
in the front panel.
5. Press F3 to change gateway, enter right format gateway values by using digital keypad in the
front panel.
6. Press F4 to change MAC, enter right format MAC values using digital keypad in the front panel.
7. Press F5 to exit.
Note: Not all mode of the machine has Ethernet interface.
Time Date Setting
Please follow the steps to change time and date:
1. Press<Utility> button, and press ENT button to go into System Settings, then choose “Time and date Settings”, press ENT button to open “Time and date Settings” interface;
3. Press F1 to change year.
4. Press F2 to change month.
5. Press F3 to change day.
6. Press F4 to change hour.
7. Press F5 to change minute.
8. Press F6 to enter page 2 of main menu, and press F1 to change second value.
9. Press F6 to enter page 2 of main menu, and press F2 to exit.
Reference Source Settings
The default is 10MHz internal source. For the synchronization among different devices, please
choose external 10MHz crystals as clock source.
Please follow the steps to change reference clock source:
1. Press<Utility> button, and press ENT button to go into System Settings, then choose “Clock
HDG2000 Series Arbitrary Waveform Generator 18
Basic Operation
Source”, press ENT button to open “Clock Source” interface.
2. Press F1 to select internal source.
3. Press F1 twice to select external source.
4. Press F2 to exit.
2.2.2 System Infomation
It contains information about the device, hardware, software and law information.
2.2.3 System State
Please follow the steps to view the system's main state:
1. Press<Utility> button, and press F2-> F2->ENT button to go into Systatus interface.
2. Press F5 to exit.
Please follow the steps bellow to save the system state file:
1. Press<Utility> button, and press F2-> F2->ENT button to go into Systatus interface.
2. Press F1 to go into <File Brower>, then select dir (File Win or Disk Win) of the file to be saved.
3. Press F3 in the <File Brower>main menu, and then name the file.
4. Press F3 again to actually save the file.
5. Then it returns to the system state window.
6. Press F5 to exit.
Please follow the steps bellow to recall the system state file:
1. Press<Utility> button, and press F2-> F2->ENT button to go into Systatus interface.
2. Press F1 to go into <File Brower>, then select dir of the saved file.
3. Press F4 to recall the file selected to recall the state.
4. Then it returns to the system state window.
5. Press F5 to exit.
Please press <Utility> button, select <SysStatus>, and press F3 to select <last> to save last time
setting. When start up next time, the system is set by last time setting.
2.2.4 Update
Please follow the steps to update the system:
1. Copy .upk file to U disk. And insert U disk into device’s USB port.
2. Press “Save/Recall” button, then press F1 to switch to “Disk Win” window, rotate the knob to
“USB” and press “Enter” button to confirm.
HDG2000 Series Arbitrary Waveform Generator 19
File Window
Disk Window
Basic Operation
3. Select “Recall->Update” softkey to update latest firmware.
4. Finally, please restart the device.
2.2.5 Counter
Counter function and frequency meter are also built-in the series waveform generator, including
low frequency counter, high frequency counter, low frequency meter, high frequency meter.
HDG2000 Series Arbitrary Waveform Generator 20
Basic Operation
To measure the frequency low than 100KHz, please select low freq counter and low freq meter.
To measure frequency higher than 10MHz, please select high freq counter and high freq meter.
Please follow these steps to count low frequency signal:
1. Connect the signal to the COUNT connector in the front panel.
2. Press<Utility> button, and press F2-> F2-> F2-> F2->ENT button to go into count interface.
3. Choose low freq counter type.
4. Press F3 to start counter
5. Press F3 again to stop counter
Please follow these steps to measure the frequency of low frequency signal:
1. Connect the signal to the COUNT connector in the front panel.
2. Press<Utility> button, and press F2-> F2-> F2-> F2->ENT button to go into count interface.
3. Choose low freq meter type.
4. Press F3 to start
5. Press F3 again to stop
6. The default gate time is 1s. Please press F4 button and use knob button and digital keypad to
modify the value.
Please follow these steps to count high frequency signal:
1. Connect the signal to the COUNT connector in the front panel.
2. Press<Utility> button, and press F2-> F2-> F2-> F2->ENT button to go into count interface.
3. Choose high freq conter type.
4. Press F3 to start counter.
5. Press F3 again to stop counter.
Please follow these steps to measure the frequency of high freqency:
1. Connect the signal to the COUNT connector in the front panel.
2. Press<Utility> button, and press F2-> F2-> F2-> F2->ENT button to go into count interface.
3. Choose high freq meter type.
4. Press F3 to start
5. Press F3 again to stop
6. The default gate time is 1s. Please press F4 button and use knob button and digital keypad to
modify the value.
2.2.6 Self Test
To make sure the device can be used normally, please do self test procedure.
During the operation, real-time clock, ADC model, FPGA and the front analog circuit will be tested.
HDG2000 Series Arbitrary Waveform Generator 21
Basic Operation
Please follow these steps to start the operation:
1. Press<Utility> button, and press F2-> F2-> F2-> F2-> F2->ENT button to go into self test
interface.
2. Press F1 to start.
3. The result of self test will be displayed on the screen.
All the whole self test will cost 15 second.
2.2.7 Calibration
Under the premise of well self test, user can do calibration manually. But it needs to administrator
privilege.
Calibration will calculate correction factor and try to correct error of the related parameters. During
this operation, frequency, ADC offset and DC offset will be calibrated.
Please follow these steps to do calibration:
1. Press<Utility> button, and press F2-> F2-> F2-> F2-> F2-> F2->ENT button to go into
calibration interface.
2. Press F1 button and input the password.
3. Press F1 again to start. During the operation, input specific values.
4. Input a value with digital keypad.
5. Press F3 button to save result.
6. Press F4 to exit.
Virtual Keyboard
Virtual keyboard mainly used to input the characters.
Virtual Keyboard's Features:
1. Provide two languages: English and Chinese
2. Choose different characters by using knobs, then press enter button to confirm.
3. Modify characters
Please follow these steps to use Virtual Keyboard:
1. Select input method English or Chinese
2. Input character
3. Modify character
4. Confirm to return
HDG2000 Series Arbitrary Waveform Generator 22
Basic Operation
Select input method
Firstly, Press F1 to select input method English or Chinese.
Input characters
1. If input method is English, shift knob to choose a character, press ENTER button to input it, then
the character appears in the text edit box in the window.
2. If input method is Chinese, rotate knob to choose Chinese pinyin, then the pinyin characters
show in the text edit box in the window. At the same time candidate of Chinese characters are
showed. Press digital num 1 to 9 to choose or to see next or previous group of candidates by using
knob.
Modify characters
1. Press F3 to select the focus widget as Text Edit box.
2. Using left/right direction button and backspace or CE button to modify operation.
3. If continue to input character, press F3 again to select soft keyboard as the focus widget.
Confirm and return
Press F4 to exit.
Please follow these steps o check the password:
1. Input the password by the digital keypad;
2. Press F1 button to confirm. If the password is right, function interface will be visited. If not, the
content will be empty and wait for inputting again.
3. Press F3 to exit.
USB Storage Device
The series generator can connect with USB storage device with fat32 format. Most identifiable
USB equipment memory is 32G. Wave File generated by <Wave Editor> or State file generated
by<System State> can be stored or recalled from the USB Storage Device.
SD Card
The series generator can connect with SD card with Fat32 format. Most identifiable equipment
memory is 32G. Wave File generated by <Wave Editor> or status file generated by<System
State> can be stored or recalled from the SD card.
2.3 Save and Recall
File Brower
File Brower is a storage device to browse files and save configuration information, consist of
internal flash, external USB device and SD card.
Features:
HDG2000 Series Arbitrary Waveform Generator 23
Basic Operation
1. Browse files and dirs
2. Create or delete files and dirs
3. Create a new dirs folder
4. Open files identified by the device
Brower files and dirs:
1. Choose a storage media. Press F1 to choose media area in the window as focus area, rotate
the knob button and press enter button to change the current media.
2. Choose the file. Press F1 to choose file area in the window as focus area, change the current
file by rotating the knob button.
Delete files:
1. Choose the file. Then press F5 button to delete it.
2. After delete all files in a dir, choose the dir and press F5 to delete it.
Create dirs floder:
Press F4 to create a new dir under parent dir.
Open files:
1. Choose the file.
2. Press Enter button to recall the file, if the file can be recognized, it will be opened automatically.
Otherwise, nothing happens.
Save and load in internal flash, external USB or SD card:
1.Click “Utility->SysStatus->Save” to save system status as .hsf file.
2.Click “Utility->SysStatus->Recall” to recall saved system status from .hsf file.
3.In arb menu, click “Type->User->Recall” or click “Editor->File->Open” to recall arb waveform
data from .hwf file.
4.In arb menu, click “Editor->File->Save” to save arb waveform data which is edited by HDG2000
to .hwf file. Especially, the size of the data file must be less than 2M. When user uses software
to edit and download arb wavefrom to HDG2000, the data can not be saved by HDG2000
again.
2.4 Remote Control
The device can communicate with PC by USB port.
The software on the PC can create more waves data and carry out the operation for waveform
signal at the same time.
HDG2000 Series Arbitrary Waveform Generator 24
Features and Function
Chapter 3 Features and Function
This section contains details on instrument features, including front panel and remote interface
operation. You may want to read Front-Panel Menu Referencefirst. See SCPI Programming
Referencefor details on SCPI commands and queries. This section covers:
Output Configuration
Pulse Waveforms
Amplitude Modulation (AM) and Frequency Modulation (FM)
Phase Modulation (PM)
Frequency-Shift Keying (FSK) Modulation
Pulse Width Modulation (PWM)
Frequency Sweep
Burst Mode
Triggering
HDG2000 Series Arbitrary Waveform Generator 25
Features and Function
Carrier
AM
FM
PM
ASK
FSK
PSK
BPSK
PWM
Burst
Sweep
Sine
· · · · · · · ·
·
Square
· · · · · · · · ·
·
Ramp
· · · · · · · · · · Pulse
· · · · · · · ·
·
Gaussian
Noise
· ·a
Arbitrary
Waveform
· · ·
·
3.1 Output Configuration
This section describes output channel configuration. Many commands associated with output
configuration start with SOURce1: or SOURce2: to indicate a certain channel. User can’t omit.
Output Function
The instrument includes six standard waveforms: sine, square, ramp, pulse, noise and Harmonic.
There are also nine built-in arbitrary waveforms, and you can create custom waveforms with the
embedded waveform editor.
The table below shows which functions are allowed (•) with modulation, sweep, and burst.
Selecting a function that is not allowed with a modulation or mode disables the modulation or
mode.
(a) Gated burst only
Frequency Limitations: Changing functions may change the frequency to meet the new
function's frequency limits.
Amplitude and offset cannot combine to exceed the instrument’s capability. The one you set
last may be changed to stay within limits.
You may protect a device under test (DUT) by specifying upper and lower output voltage limits.
Front Panel
SCPI:
SOURce<n>: FUNCtion <SINusoid|SQUare|RAMP|PULSe|NOISe|DC|SINC|EXPFall|
HAVErsine|LOREntz| DUALtone|GAUSe|ECG| USER| HARMonic| >
Output Frequency
The output frequency range depends on the function (default frequency 1 kHz for all functions).
Please refer to Appendix B.
HDG2000 Series Arbitrary Waveform Generator 26
Features and Function
Frequency Limitations: Changing functions may change the frequency to meet the new
function's frequency limits.
Burst Limitation: For internally-triggered bursts, the minimum frequency is 1 Hz.
Duty Cycle Limitations: For Square and Pulse, Duty Cycle is limited by the 16ns minimum pulse
width specification. For example, at 1 kHz, Duty Cycle may be set as low as 0.01%, because
that would result in a pulse width of 100 ns. At 1 MHz, the minimum Duty Cycle is 1.6%, and at
10 MHz it is 16%. Changing to a frequency that cannotproduce the current duty cycle will adjust
the duty cycle to meet the minimum pulse width specification.
Autoranging is enabled by default and the instrument selects optimal attenuator settings.
Output Control
By default, channel output is disabled at power on to protect other equipment. To enable a
channel's output, see below. When channel output is enabled, the corresponding channel button
is lit.
Front Panel:
CH1=CH2: The output signal of CH1 is the same as the output signal of CH2.
CH1+CH2: When thisbutton is on, the output signal of current channel will be the superposition
of CH1 and CH2.
SCPI:
OUTPut<n> ON|OFF,
OUTPut<n>?
Sync Output Signal
A sync output is provided on the front-panel Sync connector. All of the standard output functions
(except DC and noise) have an associated Sync signal. For applications where you may not want
to output the Sync signal, you can disable the Sync connector (Trig Menu->Syns->OFF). The
Sync signal may be derived from either output channel in the instrument.
General Behavior
By default, the Sync signal is derived from channel 1 and is routed to the Sync connector
(enabled).
When the Sync signal is disabled, the output level on the Sync connector is at a logic "low."
For sine, pulse, ramp, square, and triangle waves, the Sync signal is a square wave that is
"high" in the first half of the cycle and "low" in the last half.
The amplitude of Sync signal is not adjustable and fixed at TTL level.
Front Panel:
SCPI:
OUTPut:SYNC:SOURce CH1|CH2
HDG2000 Series Arbitrary Waveform Generator 29