To ensure proper operation of this product it is essential that the service
personnel is familiar with the "Operating Instructions" which should be studied
carefully before use.
Special attention is to be given to the chapter "Safety Notes"
The equipment must be used in accordance with the safety procedures
described below, and must not be used for purposes other than those for which it
was designed. The equipment may only be used by persons having recognized
qualification, including adequate training in radiation protection, authorizing them to
perform the examination or treatment carried out.
It is the responsibility of the user to ensure that the government regulations are
observed in the installation and operation of the equipment.
Technical safety note:
Regulations
If legal regulations govern the operation of the above equipment, it is the
responsibility of the operator to observe them.
For the safety of patients, operators and others, as well the efficient functioning
of the equipment it is necessary to have periodic service inspections at 12-month
intervals according to the maintenance schedule. Please apply to your service
organisation for inspection and maintenance.
Inspections intervals must by all means meet the requirements of the respective
legislation or government regulations.
Changes and additions to the product must comply with the relevant legislation
as well as with the accepted standards of good manufacturing practice.
As manufacturer of electromedical systems, we assume responsibilityfor
the safety of the equipment only if maintenance, repairs and changes are carried
out exclusively by us or third parties expressly authorised by us to do so, and if
defective parts relating to the safety of the equipment are replaced by genuine
spare parts.
We recommend that the service personnel is being asked to issue a certificate
specifying the kind and extend of things or work ranges. Also the certificate
should show the date of repair, the name of the service company and the
signature of the technician.
Before operating the equipment, the operator must check all devices concerning
the safe and efficient functioning.
If the user of this equipment likes to combine the unit with other units,
components or assemblies and this can not be made clear from the technical
data, he must question us as manufacturer or another expert to make sure that
the safety of the patients and operator is given by the planned combination.
Product Safety
Electrical safety
Only trained service personnel are permitted to remove covers and panels from
the x-ray equipment .
In the Federal Republic of Germany, the electrical installation of rooms used for
medical purposes must conform to the provisions of the VDE Standard 0107. In all
other countries, the provisions of the applicable local laws and regulations have
priority.
The unit is only prepared for solid installation with an all poled separation from the
power (ICE 601, Kap. 57.1).
Mechanical safety
It is the responsibility of the operator to ensure safety of patient while the unit is
in operation by visual check, proper patient positioning, and use of devices that
are provided.
Danger of injury
The solid black arrows and dotted lines in the illustration show areas which present
potential Danger of Injury to operating personnal and patient from the equipment
motion.
The unit has no controls with which radiation could be triggerd.
Exposure is triggerd only from the radiation-protected location of the generator.
The general radiation-protection measures must be observed.
In addition, we recommend the following:
1 Set the tube current as low as possible .
2 Limit the radiation field to the maximum possible extent.
3 Keep as fare away as possible.
4 Provide radiation protection for the patient.
This equipment is not designed for use in areas where explosion hazard can take
place.
Only skin cleansing agents may be used which form non-explosive mixtures with
air.
Interference Suppression
The equipment complies with the EMC-requirements of the guideline 89/336 EWG.
of
* Special board International Electronic Commission (IEC) This unit complies to
EN 55011 and the reference value is according EN 55011 Group 1 Class B the
international electrotechnical committee (IEC).
Classification of product
The equipment complies to the protection degree of Class 1 and for protection
against electric shock Type B.
EC Conformity
The COMBI ELEVATOR 2 to which this declaration relates fulfills the essential
requirements for safety of medical electrical equipment and follows the provisions
of Medical Device Directive 93/42 EEC part 11 para. 3 according the procedure in
annex II.
The CE-Mark is only applicable for the product without X-ray components and
Bucky.
The declaration of EC-conformity can be sent to you by request:
Write to:
Hans Pausch
Röntgengerätebau
c/o Quality Assurance Sys. Mgr.
Postfach 28 60
D-91016 Erlangen
Fax #: ..49 9131 99 24 22
Enviroment Condition
Surrounding temperatur range10 C to 40 C
Humidity20% to 80%
Atmospheric pressure700 hPa to 1100 hPa
Disposal of equipment
Legal waste disposal regulations may apply to the disposal of this product. To
avoid causing damage to the environment and personal injury, we recommend that
you contact your Customer Services representative before permanently removing
this product from service.
A Table top, floating, manually movable, scratch-proof
B Profile rail with trim cover, smooth, accepts accessories
C Upper table frame
D Bucky unit, movable
E Table base, solid, vibration-free
F Guide rails for tube stand
G Foot treadle
H Vertical carriage with X-ray tube support arm
I Column, rotatable
K X-ray tube
L Control handle
M Protractor
The equipment consists of a patient table with coordinate table top and diaphragm carriage,
and of a column with X-ray tube, collimator and control arm.
The solid, vibration-free table base and the rail stand with column from one unit.
The large and 2.20 m long, floating table top is manually movable and locks
electromagnetically. The table top is provided for a maximum patient weight of 136 kg. For
fast and easy positioning of the patient, it allows spacious lateral travel: 60 cm to the left, 50
cm to the right and ± 12 cm transversely.
Especially for patient comfort and easy cleaning, the table top offers a scratch-proof surface
(Resopal) and trim-covered, smooth profile rails on both sides, which can accept
accessories.
The Bucky carriage takes up cassette trays as supplied by any reknowned manufacturer. In
longitudinal direction, it is manually moved and electromagnetically locked. The shortest
possible FFD of 70 mm guarantees images of superior geometrical proportion. Low
radiation absorption by the table top reduces the X-ray amount. The brake of the Bucky is
released by a push-button on the operating handle.
As accessory is a automatic mechanical coupler availabe for automatic connection from the
Bucky and tube stand.
The foot treadle along the table base allows the release of the elctromagnetic table top
brakes.
The tube stand for the X-ray tube support arm and tube is moved on the rail stand parallel
to the longitudinal table axis.
The X-ray tube with collimator and control arm for the tube stand is supported by a lateral
arm fixed to the vertical carriage. Vertical movement and rotation of the tube around the
longitudinal axis of the lateral arm permit vertical, horizontal or oblique beam projections.
Each position is electromagnetically locked.
The X-ray tube with column stand rotates around the vertical column axis and catches
mechanically at angles of 0° (basic position), ± 90° (for lateral exposures).
The standard tube support is intended for DIN 6836 flanged plate C.
The max. weight of the X-ray tube and collimator must not exceed 40 kg.
The equipment is an universal X-ray examination system for doctors´practices as well as for
hospitals. It provides high-quality gridded exposures on lying patients. Large travel of the
table top and uncomplicated functioning facilitate operation and increase patient comfort. A
special cassette holder is available for lateral exposures. With a Bucky wall stand or a
cassette stand, exposures can also be made with patients standing or sitting. The total tube
stand rotation range of 90° allows exposures with patients on stretchers or in bed.
Without requiring wall and ceiling fixtures, the equipment easily adjust to changes in
location.
Prerequisition
For safe and efficient operation of this product the personnel must be familiar with the
operating instructions. The chapter on „Safety procedures“ deserves special attention.
The system is designed for stationary operation. The approximate floor space required
measures 330 cm x 152 cm. An additional 20 cms minimum distance must be kept between
tube stand and wall.
Height of room
The tube stand height is 234 cms. Installation requires a minimum room height of 245 cms.
The table top of the equipment has a working height of 75 cms above floor.
Power
The system is equipped for single-phase alternating current with fixed installation. Two
versions are available, depending on order. The unit is only prepared for solid installation
with an all poled separation from the power (ICE 601, Kap. 57.1).
Without transformer, the system corresponds to nominal ratingsas follow:
Nominal voltage:230 V ACoder115/208 VAC
Nominal current:5 A10 / 6,25 A
Rated frequency:50/60 Hz50 / 60 Hz
Nominal capacity:1150 VA1150 VA
Mains
The mains connection requires a 30 mA circuit breaker to be installed by the customer. The
electrical installation must meet the relevant legislation, e.g. VDE 0107, IEC/SC 62A.
AL-equivalent
The weakening equivalent of the table top (patient pos. table top) is £ 0,7 mm.
According to:
DIN EN 60601-1-3 mit 100 kV und HWS 3,7 mm AL
and FDA 21 CFR § 1020.30 (n) mit 100 kV und HWS 2,7 mm AL.
Foot switch 14 and security switch 6a for motorised lowering of
the table top. The table is moving down as long as the switches
are pressed.
The lowering speed with soft start and break is constant.
Automatic shutdown of the down motion in end position and in
exposure height. Continue of movement after interruption in
exposure height.
Release foot switch/security switch. Push again.
Foot switch 15 and security 6a for motorised raising of the
table top. The table raises as long as the switches are pressed.
The raising speed with soft start and break is constant.
Automatic shutdown of the up motion in end position and
exposure height.
Release foot switch/security switch. Push again.
Foot switch 13 to release the brakes of the floating table top.
The table top can be manually shifted longitudinally and
transversely as long as the switch is depressed. The table top
locks in place when the switch is released.
Security switch 6a
for motorised raise respectively lowering of the table top.
Switch 6
to release the brakes for the rotation movement of the X-ray
tube. Release of the switch locks the tube in position.
By turning the x-ray tube close to the 0° and 90° position, the
LED blinks. Release of the switch causes a exactly motorized
positioning of the x-ray tube.
unlocks the column brake for longitudinal movement. Release
of the switch locks the column in place.
Switch 5
unlocks the vertical carriage. Release the switch and the vertical
carriage will be locked in place again. The motion stops
automatically at the preferred vertical SID and can be continued
by pressing the switch agin. The motion stops automatically at
the preferred vertical SID and can be continued by pressing the
switch again.
Switch 7
unlocks the column brakes. Release of the switch locks all
movements in place.
Protractor 9
indicates the angle of the tube vs. Exposure object.
Indicator LED (green) 1
lights up to indicate that the Bucky is centered to the column .
Turn of X-ray tube 10
The column stand with X-ray tube can be turned round its
longitudinal axle to the left or right side by 90 ° with a strong
start on both hand grips of the control arm. It catchs at 0° , +
90° each.
Indicator LED (green) 2a
lights when the table top is in exposure height.
Indicator LED (green) 2
indicates an exactly 0° or 90° positioned x-ray tube
Indicator LED (green) 3
indicates the adjusted vertical SID
Operation
The equipment is ready for operation when
switched on.
Emergency-off
Provided an emergency switch has been installed in the
examination room, the red switch button must be pushed
immediately in case of danger for patients, operating personnel,
or equipment.
After positive elimination of the danger only, turn the red
emergency-off switch clockwise to resume operation.
Patient Positioning/ Centering of Bucky, Object and Tube Unit.
Patient Positioning
Note: To prevent injury from the patient sitting down or beeing
positioned, move the tube stand and tube out of the critical
range before positioning the patient.
Press the foot switch 13 to unlock the table top brakes.
Move the table top manually floating to the rear stop.
Release the foot switch.
Press the foot switch 14 and the safety switch 6a
simultaneously. The table top is lowered by the motor in a
telescopic way.
Release the switchwhen comfortable working height for
patient access or transfer is reached.
Exposure
Press the foot switch 15 and the safety switch 6a. The table
top is lifted up by the motor to exposure position. Release the
switch after the automatic stop.
Note
When reaching the preset exposure position of the table top, the
downward movement is interrupted automatically. To continue
movement, release the foot switch. Press again.
Centering of the Bucky
The centering can also be achieved by moving the Bucky
carriage into the exposure range.
Note
This moving is necessary when the green lamp of the automatic
tube stand/Bucky coupler does not lighten.
Move the object into the central beam by adjusting the table
top.
For this purpose, press the foot treadle 13 fully down. The
brakes of the floating table top are now unlocked. Adjust the
table top. Release the foot treadle. The table top is locked in
position.
Centering of the radiation field
Use the cassette size sensing for optimal localization of the radiation field of the collimator (operating instruction of the collimator).
Exposure preparation
Insert cassette. Chosse SID (FFD). Set exposure dates on
control desk. Control readiness for exposure. Command
patient to „hold your breath“.
Note
Do not forget radiation protection protection devices for the
patient (lead rubber apron, gonad protection, etc.)
Oblique Exposure
Move tube stand beyond the Bucky range of the automatic
range/Bucky coupler. Move the Bucky under the patient. Tilt
and center the tube over the Bucky with the indicator light of
the collimator switched on.
Move the table top in the direction away from the wall stand.
Slide the tube stand towards the wall stand. Rotate the tube to a
90° angle (indication of angle!), switch on the indicator light, then
center the tube and adjust the wall stand Bucky to the patient by
vertical movement.
Lateral Exposure
For lateral exposures with the cassette holder sideways (cf.
page with accessories), rotate the tube stand to a 90° angle.
Then rotate the tube to a 90° angle (indication of angle!).
Proceed as described above.
The lateral cassette holder permits lateral exposures if
the tube unit is mounted to a tube swivelling device.
The lateral cassette holder is slipped in one of the profile
rails.
Grip screw (A): secures the holder at the table top
Gripscrew (B): locks the holder setting
Grip screw (C): fixes the lateral position of the cassette
clamps.
Head Supports
The head supports slide into profile rails of the table or
Bucky. The supports can be clamped in any position
desired. The patient`s head is fixed to the appropriate
exposure position by cushioned plates on adjustable
bars.
Hand screw A: Clamping head supports to table top or
Bucky.
Hand screw B: Clamping of head holder
Hand grips
The hand grips are slipped in the profile rails of the
table. They may be fixed at any position and offer a
reliablehold for the patient.
Grip screw (A): secures the grip in place
Important note:
The positioned patient may only hold on to the hand
grips. In no case may he put his hands around the
edge of the table top.