Hansol Multitech 3000 User Manual

INSTALLATION INSTRUCTIONS Uromat 3000
________________________________________________________________________
________________________________________________________________________
English Version 06/00 - 1 of 90 - 0116 7201
Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru
CONTENTS Page
1. Technical Data
1.1 General Safety Information 4
1.2 Component Weights 4
1.3 Component Dimensions 4
1.4 Packaging Dimensions and Weights 4
1.5 Component Designations 5
1.6 Dimensional Drawing 6
1.7 Cable Lengths of the High Voltage, Collimator, Adjustable Bucky and Image lntensifier Cables 10
1.8 Power Line Connection Data 11
1.9 Required Special Tools 11
1.10 Required Test Equipment 11
1.11 Environmental Conditions 11
1.12 Operating Conditions 11
1.13 Circuit Diagram Overview 12
1.14 Component Designations for the Circuit Diagram 27
1.15 Function Description of the Controller 31
2. Installation
2.1 Preparations for Installation 35
2.2 Unpacking the Unit Parts 35
2.3 Taking the Unit Off the Pallet 36
2.4 Installing the Unit 37
2.5 Aligning the Unit 38
2.6 Temporarily Connecting Line Power 39
2.7 Installing the Tube Unit Support Arm 39
2.8 Preinstallation of the Tube Unit - Collimator 40
2.9 Installing the Tube Unit - Collimator on the Tube
Unit Support Arm 40
2.10 Installing the Image Intensifier (I.I.) 41
2.11 Laying the Collimator Cables 41
2.12 Laying the High Voltage Cables 42
2.13 Laying the I.I. Cables 44
2.14 Connecting the Power Supply via the Generator 44
2.15 Installing the Measuring Chamber 45
2.16 Installing the Grid 46
2.17 Installing the Tilt Angle Indicator - Error Display 46
2.18 Installing the Cover Panels 47
2.19 Installing the Monitor Support Arm 47
2.20 Installing the Holder for the Flush Bowl 47
2.21 Sealing the Table Frame Cover Panel 48
Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru
3. Adjustments
3.1 Installing the Tube Unit Support Arm 48
3.2 Central Beam to Center of Bucky 50
3.3 Central Beam to Center of I.I. 50
3.4 Tabletop, Longitudinal Movement Path Potentiometer 51
3.5 Tabletop, Transverse Movement Path Potentiometer 52
3.6 Bucky Movement Path Potentiometer 52
3.7 I.I. Carriage Movement Path Potentiometer 52
3.8 Table Tilt Movement Path Potentiometer 1 53
3.9 Table Tilt Movement Path Potentiometer 2 53
3.10 Table, Longitudinal Safety Switches 54
3.11 Tabletop, Transverse Safety Switches not configured 54
3.12 Tabletop, Longitudinal Safety Switches 55
3.13 I.I. Carriage Movement Path Safety Switches 55
3.14 Table Tilt Up Movement Path Switch Strike Plate 55
3.15 Table Tilt Down Movement Path Switch Strike Plate 55
3.16 +88° Tilt Movement Safety Switches 56
3.17 -20° Tilt Movement Safety Switches 56
3.18 Spindle Nut Safety Switches 56
3.19 Accessory Sensor Switches 57
3.20 Longitudinal Tabletop Belt Tension 57
3.21 Raise - Lower Table Belt Tension 57
3.22 Cassette Carriage Drive Belt Tension 58
3.23 Operation in the Service Mode 59
3.24 Startup 61
3.25 Collimator Basic Setting 62
3.26 Collimation to Cassette and I.I. Formats 63
3.27 Saving the Parameters in the PC 64
3.28 Setting Parameters - Electronics 65
4. Technical Maintenance
4.1 Mechanical and Electrical Checks 67
4.2 Function Test 70
4.3 Spare Parts - Overall View 72
4.4 List of Spare Parts 79
5. Troubleshooting 84
6. Replacing Boards 86
7. Version Updates 86
8. Maintenance Verification 87
9. Location of Identification Labels 88
06/00 - 3 of 90 - 0116 7201
Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru
1. TECHNICAL DATA
1.1 General Safety Information
Maintenance and repair may be performed only by an office that is authorized by the manufacturer.
In the Federal Republic of Germany, electrical installation of medically used rooms must conform to VDE Regulation 0107. In all other countries, the particular applicable national regulations take precedence and must be observed. These can be found in the system project plan.
During installation, it must be assured that all ground wire connections provided by the manufacturer must be connected prior to starting up the unit. The ground wire between the individual components and the power supply must be connected as shown in the Wiring Diagram.
The regulations provided by the trade unions for occupational safety and prevention of accidents must be observed.
No work may be performed on parts that are under power (above 42 V). This condition applies for both measurement and adjustment work steps. However, special care is required when performing these worksteps.
If the operating instructions call for voltage to be switched on for unit movements, the system must be shut down immediately following such tests.
The radiation regulations must be observed while making settings and checks that must be performed under X-radiation; radiation protective clothing must be worn.
1.2 Component Weights
Unit column approx. 320 kg Table approx. 230 kg Tube unit support arm approx. 50 kg Footswitch approx. 5 kg Total weight, incl. tube unit, collimator, adjustable Bucky and I.I.: approx. 730 kg
1.3 Component Dimensions
Unit column with table and tube unit support arm on the pallet. 2050 mm x 1015 x 1800 mm
1.4 Packaging Dimensions and Weights
Unit column with table, incl. accessories 2050 mm x 1015 mm x 1890 mm approx. 900 kg
Transport carriage 1800 mm x 200 mm x 200 mm approx. 85 kg
Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru
1.5 Component Designations
(the illustration shows the right-handed version, the left-handed version is the mirror
image)
A X-ray tube unit - collimator B Tube unit support arm, adjustable C Unit table with four-way table movement D Unit column E Manual control unit F Leg supports G Elbow supports H Flush bowl and rinse bag holder J Table extension K Footrest for table extension L Cassette shaft cover M Micturation seat N Emergency stop switch O Head cushion with holder P Paper roll with holder R Patient handgrips S Footswitch for exposure and fluoroscopy T Multi-function footswitch U Grip handle V Tilt angle display / position memory display / error display
06/00 - 5 of 90 - 0116 7201
Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru
1.6 Dimensional Drawing, Right-handed Version
min.
650
1750min.
1200
The mentioned centre of gravity applies
without X-ray components and accessories.
Table horizontal in highest position,
tube support arm at the head end.
* gemÌB DIN EN 349
according to DIN EN 349
min.680 - max.1225
1150FFA
113
BV
65
Film
360
650
640
130 130
1940max.
1400
880
700
inchmm
inchmm
26,772
27,559700
680
0,787
2,559
65
20
30,709
29,921
780
760
5,118
4,449
113
130
34,646
36,220
880
920
5,709
7,087
180
145
45,276
47,244
1150
1200
7,874
9,449
240
200
48,228
52,559
1335
1225
11,811
12,795
300
325
55,118
52,756
1400
1340
14,173
19,488
495
360
57,087
68,898
1450
1750
19,685
25,197
640
500
76,378
102,362
1940
2600
26,063
25,591
650
662
2600min.
200min.
Der angegebene Schwerpunkt gilt
ohne RÎntgenkomponenten und ZubehÎr.
Tisch horizontal in oberster Stellung,
RÎhrentragarm auf der Kopfseite.
Gewicht ca. 670 kg
weight approx. 670 kg
1335
920
760
500min. *
240
495
20°
88°
240
200
325
1450max. ca.
300max.
20
20
145min.
TischverlÌngerung
+ FuBbank table extension
+ foot rest
780
662
*
495min. ca.
500
1340max. ca.
Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru
1.6.1 Dimensional Drawing, Left-handed Version
1200
1750min.
650
min.680 - max.1225
1150FFA
113
BV
65
Film
130130
360
650
640
700
1940max.
1400
880
mm inch
mm inch
27,559
26,772
700
680
0,787
2,559
65
20
30,709
29,921
780
760
5,118130
4,449
113
36,220
34,646
920
880
5,709
7,087
145
180
47,244
45,276
1150
1200
9,449
7,874
200
240
52,559
48,228
1225
1335
11,811
12,795
300
325
55,118
52,756
1340
1400
19,488
14,173
360
495
68,898
57,087
1750
1450
25,197
19,685
500
640
76,378
102,362
2600
1940
26,063
25,591
650
662
495
20°
88°
2600min.
1340max. ca.
495min. ca.
200min.
500min. *
Der angegebene Schwerpunkt gilt
ohne RÎntgenkomponenten und ZubehÎr.
Tisch horizontal in oberster Stellung,
RÎhrentragarm auf der Kopfseite.
The mentioned centre of gravity applies
without X-ray components and accessories.
Table horizontal in highest position,
tube support arm at the head end.
Gewicht ca. 670 kg
weight approx. 670 kg
* gemÌB DIN EN 349
according to DIN EN 349
662
1450max. ca.
240
200
325
20
20
145min.
TischverlÌngerung
+ FuBbank table extension
+ foot rest
780
300max.
240
920
760
1335
06/00 - 7 of 90 - 0116 7201
Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru
500min. *
1.6.2 Dimensional Drawing for Floor Mounting, Right-handed Version
450 mm über dem Boden
345
90
min. Wandabstand
Drehachse
Fußauflagfläche
1420
74
90 74
**
900
45 800
90 470
P1
*
200
P2
180
140 205205
* Cable intake through the floor ** End of flexible hose for wall connection of 2000 mm flexible hose length
Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru
Kabelanschluß
690
mit Rillenschlauch
1.6.3 Dimensional Drawing for Floor Mounting, Left-handed Version
450 mm über dem Boden
345
180
min. Wandabstand
Drehachse
Fußauflagefläche
P1
1420
P2
900
*
200
90 470
90 74
45 800
74
140205 205
**
* Cable intake through the floor ** End of flexible hose for wall connection of 2000 mm flexible hose length
06/00 - 9 of 90 - 0116 7201
Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru
690
Kabelanschluß mit Rillenschlauch
1.7 Cable Lengths of the High Voltage, Collimator, Bucky and Image
Intensifier Cables
High voltage cables A, C, D, E, F, 700 cm
Collimator cables B, C, D, E, G, 650 cm
Bucky, measuring chamber A, K, L, M, P, S, 490 cm
Image intensifier cables A, K, L, M, N, 630 cm
Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru
1.8 Power Line Connection Data
Power lead-in must be routed over a 30 mA fault current interrupter that is provided by the customer. The room installation must comply with VDE 0107. In all other countries outside the Federal Republic Germany, the legally specified country regulations take precedence and must be maintained. Prerequisite: The unit is designed for single-phase DC current with a fixed installation and for a fixed connection using an all-pole separator from the Network (IEC 601, Chapter 57.1). During installation, it must be possible to adapt the power line voltage and frequency to correspond to the order.
Power connection: 1 N 115 /200/208/ 230/240 V AC Frequency: 50 / 60 Hz Nominal current (fuse): 13/7.5/7.2/6.5/6.25 A Nominal line power: 1500 VA Heat dissipation: 240 W
1.9 Required Special Tools
Torque wrench 50 Nm (5 mkp) Masonry drill bit 12 mm dia. Sista sanitary caulking F 101 Special grease - Tunap Tungrease BS Pa. Nr.: 9026 0001
1.10 Required Test Equipment
PC or laptop with color display (black/white is also sufficient, but operation is made more difficult); min. 486 processor; 16 MB RAM; with a 3 ½" diskette drive and hard drive. Windows 3.x ; Win 95 operating system (Windows NT can cause difficulties with the Com interface). Serial connection cable (9-pole, Sub-D connector to 9-pole Sub-D socket, 1-1 connection), length > 1m. Spirit level Tape measure Multimeter test instrument
1.11 Conditions for Transport and Storage
Ambient temperature range -25 C to 70 C Relative humidity in the range 5% to 95% Air pressure in the range 700 hPa to 1100 hPa
1.12 Operating Conditions
Ambient temperature range 10 C to 40 C Relative humidity in the range 20% to 80% Air pressure in the range 700 hPa to 1100 hPa
06/00 - 11 of 90 - 0116 7201
Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru
1.13 Circuit Diagram, Page 1
1.13.1 Description of Circuit Diagram
....... Circuit Diagram, Page 2
= AX Control unit = system, designation
...... Circuit Diagram, Page 3
= AX + A2 Display unit = system, + location, designation
1.13.2 Contents of Circuit Diagram
1.13 Circuit Diagram, Page 2 Page 13
1.13 Circuit Diagram, Page 3 Page 14
1.13 Circuit Diagram, Page 4 Page 15
1.13 Circuit Diagram, Page 5 Page 16
1.13 Circuit Diagram, Page 6 Page 17
1.13 Circuit Diagram, Page 7 Page 18
1.13 Circuit Diagram, Page 8 Page 19
1.13 Circuit Diagram, Page 9 Page 20
1.13 Circuit Diagram, Page 10 Page 21
1.13 Circuit Diagram, Page 11 Page 22
1.13 Circuit Diagram, Page 12 Page 23
1.13 Circuit Diagram, Page 13 Page 24
1.13 Circuit Diagram, Page 14 Page 25
1.13 Circuit Diagram, Page 15 Collimator Page 26
Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru
1.13 Circuit Diagram, Page 2
06/00 - 13 of 90 - 0116 7201
Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru
1.13 Circuit Diagram, Page 3
Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru
1.13 Circuit Diagram, Page 4
06/00 - 15 of 90 - 0116 7201
Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru
1.13 Circuit Diagram, Page 5
Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru
1.13 Circuit Diagram, Page 6
06/00 - 17 of 90 - 0116 7201
Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru
1.13 Circuit Diagram, Page 7
Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru
1.13 Circuit Diagram, Page 8
06/00 - 19 of 90 - 0116 7201
Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru
1.13 Circuit Diagram, Page 9
Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru
1.13 Circuit Diagram, Page 10
06/00 - 21 of 90 - 0116 7201
Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru
1.13 Circuit Diagram, Page 11
Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru
1.13 Circuit Diagram, Page 12
06/00 - 23 of 90 - 0116 7201
Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru
1.13 Circuit Diagram, Page 13
Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru
1.13 Circuit Diagram, Page 14
06/00 - 25 of 90 - 0116 7201
Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru
1.13 Circuit Diagram, Page 15 Collimator connection
Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru
1.14 Component Designations for the Circuit Diagram
System Location Component Function =AU Uromat 3000
=AU +A1 CPU board =AU +A1 -D9 LED, footswitch power supply =AU +A1 -D15 LED, display power supply =AU +A1 -D18 LED, control console power supply =AU +A1 -D29 LED, CPU power supply =AU +A1 -D30 LED, Analog power supply =AU +A1 -D31 LED, + 24 V power supply =AU +A1 -D33 LED, K19 relay activated =AU +A1 -D40 LED, CPU status =AU +A1 -D41 LED, 20VAC power supply =AU +A1 -D50 LED, +24V DC power supply =AU +A1 -D51 LED, +10V DC power s =AU +A1 -D55 LED, emergency stop
=AU +A1 -F1 Fuse, CPU power supply =AU +A1 -F2 Fuse, analog power supply =AU +A1 -F3 Fuse, control console power supply =AU +A1 -F4 Fuse, footswitch power supply =AU +A1 -F5 Fuse, display power supply
=AU +A1 -K8 Button fuse relay =AU +A1 -K19 Enable table transverse movement relay =AU +A1 -K20 Table, transverse, left, right movement relay =AU +A1 -K23 Emergency stop interface relay =AU +A1 -K24 Emergency stop interface relay
=AU +A1 -Q2 Transistor for K19 pull-in relay =AU +A1 -S1 Service switch =AU +A1 -X1 Power supply plug-in connector
=AU +A1 -X2 Footswitch plug-in connector =AU +A1 -X3 Control console plug-in connector =AU +A1 -X4 Display connector strip =AU +A1 -X5 Display connector strip =AU +A1 -X6 RS 485 plug-in connector =AU +A1 -X7 Service PC plug-in connector =AU +A1 -X8 Collimator plug-in connector =AU +A1 -X9 Interface plug-in connector =AU +A1 -X10 Tube unit system plug-in connector =AU +A1 -X11 Digital table plug-in connector =AU +A1 -X12 Collimator plug-in connector =AU +A1 -X13 Analog table plug-in connector =AU +A1 -X14 Analog column plug-in connector =AU +A1 -X15 Power supply terminal strip =AU +A1 -X17 Emergency stop plug-in connector =AU +A1 -X19 Display fiber-optic cable plug-in connector
06/00 - 27 of 90 - 0116 7201
Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru
Loading...
+ 63 hidden pages