OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDEOPERATIONAL GUIDE
INFRARED SENSOR MEASUREMENT
Switch on the instrument and be sure
that it is in the IR (infrared) mode.
Note: If a “P” appears on the LCD, it means that the
thermometer is in the external probe mode. Press the
IR/CONTACT button to switch to IR (infrared) measurement mode.
To obtain the temperature
of an object, simply aim
the sensor in the direction
of the object.
Reading Distance
The distance between the sensor and the object is an
important factor in infrared measurement.
The relation between “distance/object area” is 3:1. This
means that the diameter of the measured area is one third
of the distance between the sensor and the object.
EMISSIVITY
The instrument measures the temperature of objects contain-
ing an emissivity of 0.95, which is the characteristic of most
substances, such as, organic material, water, plastic, fabrics,
etc.
To measure substances that have a polished surface and are
very reflective, the emissivity needs to be compensated. It is
advisable to cover the object with black paint or black
adhesive tape to compensate for the reflectivity.
WARNING
To avoid damage, do not hold the sensor too close to a
heat source for a long period.
SENSOR CLEANING
To maintain the instrument’s accuracy, it is important that
the sensor lens is kept clean. To clean the sensor, gently pass
a humid and non-abrasive cloth (HI 731318) on the lens.
EXTERNAL PROBE
Connect the temperature probe
HI 765PWL to the connector
on top of the instrument.
O
N
OFF
IR
C
O
N
T
A
C
T
MEASUREMENT MODE
Press the IR/CONTACT button to select measurement mode.
“P” on the LCD indicates external probe.
WEAK BATTERY
HI 99556HI 99556
HI 99556 continuously checks the battery status. If the
HI 99556HI 99556
battery level is too low to guarantee the stated accuracy, it
shuts off the LCD completely. If the meter cannot be switch
on, or switches itself off after being switched on, replace the
battery.
BATTERY REPLACEMENTBATTERY REPLACEMENT
BATTERY REPLACEMENT
BATTERY REPLACEMENTBATTERY REPLACEMENT
Battery replacement must only take place in a non-hazardous area using a 9V alkaline battery.
Simply slide off the battery cover on the back of the meter.
Detach the battery from the terminals and attach a fresh
9V battery while paying attention to the correct polarity.
Replace the battery and the cover.
ACCESSORIESACCESSORIES
ACCESSORIES
ACCESSORIESACCESSORIES
HI 710004 Soft carrying case
HI 710007 Blue shockproof rubber boot
HI 710008 Orange shockproof rubber boot
HI 721316 Rugged carrying case
HI 731318 Wiping tissue (4 pcs)
HI 740016 Battery cover
HI 740029 9 V Battery
HI 740123 Wrist-strap
HI 765PWL Reinforced penetration thermistor probe
CE DECLARATIONCE DECLARATION
CE DECLARATION
CE DECLARATIONCE DECLARATION
OF CONFORMITYOF CONFORMITY
OF CONFORMITY
OF CONFORMITYOF CONFORMITY
HI 99556
Recommendations for Users
Before using these products, make sure that they are entirely suitable for the environment in which they are used.
Operation of these instruments in residential area could cause unacceptable interference
to radio and TV equipment, requiring the operator to take all necessary steps to correct
interferences.
Any variation introduced by the user to the supplied equipment may degrade the
instruments' EMC performance.
To avoid damage or burns, do not perform any measurement in microwave ovens.
For example, if the sensor is 15 cm away from the object,
the instrument measures the temperature of an area with a
5cm diameter.
Dip the probe in the substance which you intend to measure
(4cm immersion level is recommended) and wait for the
reading to stabilize.
PRINTED IN PORTUGAL
IST99556R2 02/00
Instruction Manual
HI 99556
Contact and Infrared
Thermometer
http://www.hannainst.com
WARRANTYWARRANTY
WARRANTY
WARRANTYWARRANTY
HI 99556 is warranted for two years against defects in workmanship and materials when used for its intended purpose and maintained
according to instructions.
This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not
covered by the warranty, you will be notified of the charges incurred. If
the instrument is to be returned to Hanna Instruments, first obtain a
Returned Goods Authorization Number from the Customer Service department and then send it with shipment costs prepaid. When shipping
any instrument, make sure it is properly packaged for complete protection.
To validate your warranty, fill out and return the enclosed warranty card
within 14 days from the date of purchase.
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895 , USA.
This Instrument is in
Compliance with the CE Directives
Dear Customer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for the correct
operation of the meter. Please read it carefully before using
the meter. If you need additional technical information, do
not hesitate to e-mail us at tech@hannainst.com.
This instrument is in compliance with directives
EN 50081-1 and EN 50082-1.
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
Remove the instrument from the packing material and
examine it carefully to make sure that no damage has
occurred during shipment. If there is any damage, notify
your Dealer.
The instrument is supplied complete with
• 9V Battery
• Instruction manual
Note: Conserve all packing material until the instrument
has been observed to function correctly. Any defective
item must be returned in its original packing.
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
Infrared radiation emitted from any object depends on its
temperature. HI 99556 is a thermometer based on the
infrared radiation technology.
The most evident advantage of infrared thermometry is the
non-contact nature of the measurement.
HI 99556 also accepts an external thermistor temperature
probe. The selection of measurement between the external
probe and infrared sensor is simply through a key.
The different models are:
HI 99556-00 measures in Celsius,
HI 99556-01 measures in Fahrenheit,
HI 99556-10 measures in Celsius.
The thermometers also provide for a low battery indication.
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
HI 99556-00 HI 99556-01 HI 99556-10
Range
IR: -10 to 300 °C 14 to 572 °F -20 to 199.9°C
Probe: -40 a 150 °C -40 a 302 °F -40 to 150.0°C
Resolution
IR: 1 °C 1 °F 0.1 °C
Probe: 0.1 °C 1 °F 0.1 °C
Accuracy
IR: ±2% of reading or ±2°C/ ±3°F
Probe: ±0.5°C ( -20 to +120°C) / ±1°F (0 to +250°F)
±0.5°C / ±1°F +1% reading (outside)
EMC Deviation
IR: ±1 °C ±2 °F ±1 °C
Probe: ±0.5 °C ±1 °F ±0.5 °C
Temperature probe HI 765PWL (included)
Emissivity (IR) 0.95 (fixed)
Wave length (IR) 6 a 14 µm
Optic coefficient 3 : 1
(ratio of distance to diameter of target)
Battery type 9V battery
Environment 0 to 50 °C (32 to 122 °F)
0 to 95% RH non-condensing
Dimensions 143x80x38 mm (5.6x3.2x1.5”)
Weight 320 g (11.3 oz.)
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
1. Infrared sensor
2. Temperature probe connector
3. Liquid Crystal Display (LCD)
4. ON/OFF button
5. IR/CONTACT: Selects Infrared sensor or external probe
6. Wrist-strap
Hanna Instruments reserves the right to modify the design, construction