Hanna Instruments HI 99551 User Manual

Instruction Manual
HI 99551 • HI 99556
Infrared Thermometers
for the Food Industry
www.hannainst.com
WARRANTY
HI 99551 and HI 99556 are warranted for two years against defects in workmanship and materials when used for its intended purpose and maintained according to instructions. Probes are guaranteed for
This warranty is limited to repair or replacement free of charge. Damages due to accident, misuse, tampering or lack of prescribed maintenance are not covered. If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charges in­curred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization Number from the Customer Ser­vice department and then send it with shipment costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protection.
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner.
Hanna Instruments reserves the right to modify the design, construction
and appearance of its products without advance notice.
Dear Customer, Thank you for choosing a Hanna product.
This manual will provide you with the necessary information for the correct operation of the meter. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com. These instruments are in compliance with the directives.
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and exam­ine it carefully to make sure that no damage has occurred during shipment. If there is any damage, notify your Dealer.
Each meter is supplied complete with:
HI 765PW temperature probe (HI 99556 only)
• 9V battery
• instruction manual
Note: Conserve all packing material until the instrument has
been observed to function correctly. Any defective item must be returned in its original packing.
GENERAL DESCRIPTION
HI 99551 and HI 99556 thermometers employ infrared technol- ogy to measure the surface temperature. Infrared measurements are extremely practical with a fast re­sponse time, typically around 1 second. Another main advantage is the non-intrusive nature of the measurement, which is particularly attractive in the food indus­try since it translates itself into substantial savings by leaving products intact, especially those sealed or pre-wrapped. This type of non-intrusive measurement is also useful when the surface temperature is high, for difficult-to-reach places or due to hy­giene requirements. Simply point to the product or the spot and hold down the measurement key. The measured value will be immediately dis­played on the LCD. If in addition to surface measurement, you must check the core temperature too, then HI 99556 is the ideal solution: simply attach the external probe to the meter and you will have a 2-in­1 infrared-thermistor thermometer. Different models are available, with different temperature ranges.
FUNCTIONAL DESCRIPTION
1. Infrared sensor
2. Temperature probe connector (HI 99556 only)
3. Liquid Crystal Display (LCD)
4. ON/OFF button
5. IR/CONTACT button to select infrared sensor or external
probe (HI 99556 only)
6. Wrist-strap
SPECIFICATIONS
HI 99551-00 HI 99551-01 HI 99551-10 HI 99556-00 HI 99556-01 HI 99556-10
Range IR Sensor
-10 to 300°C 14 to 572°F -20 to 199.9°C
Range Probe (HI 99556 only)
-40 a 150°C -40 a 302°F -40 to 150.0°C
Resolution IR Sensor
1°C 1°F 0.1°C
Resolutio Probe (HI 99556 only)
0.1°C 1°F 0.1°C
Accuracy IR Sensor
±2% of reading or ±2°C (-00 and -10 models);
±2% of reading or ±3°F (-01 models)
Accuracy Probe (HI 99556 only)
±0.5°C (-20 to 120°C) /±1°F (0 to 250°F);
±0.5°C / ±1°F +1% reading (outside)
Typical EMC Deviation IR Sensor
±1°C ±2°F ±1°C
Typical EMC Deviation Probe (HI 99556 only)
±0.5°C ±1°F ±0.5 °C
IR Sensor Response Time 1 second
Optic Coefficient 3:1 (distance/target diameter) Minimum Distance 30 mm (1.2’’) Emissivity 0.95 (fixed) Wave Length 6 a 14 µm
Probe HI 765PW (included) (HI 99556 only) Battery Type 1 x 9V alkaline (IEC 6LR61) Battery Life approx. 150 hours of continuous use Environment 0 to 50°C (32 to 122°F);
RH max 95%
Dimensions 143 x 80 x 38 mm (5.6 x 3.2 x 1.5”) Weight 320 g (11.3 oz.)
OPERATIONAL GUIDE
INFRARED SENSOR MEASUREMENT
Switch the instrument on. For HI 99556 models, check that the
instrument is in IR (infrared) mode. If necessary (a “P” tag appears on the LCD to indicate
“external probe mode”), press the IR/CONTACT button to switch to IR (infrared) measurement mode.
To measure the temperature of an object, simply aim the sensor in the direction of the object.
EMISSIVITY
The instruments measure the temperature of objects with an emissivity of 0.95, which is the characteristic of most sub­stances, such as organic material, water, plastic, fabrics, etc.
To measure substances that have a polished surface and are very reflective, the emissivity needs to be compensated. It is advisable to cover the object with black paint or black adhesive tape to compensate for the reflectivity.
WARNING
To avoid damage, do not hold the sensor too close to a heat source for a long period.
SENSOR CLEANING
To maintain the instrument’s accuracy, it is important that the sensor lens is kept clean. To clean the sensor, gently wipe the lens with a humid and non-abrasive cloth (HI 731318).
EXTERNAL PROBE (HI 99556 only)
Connect the HI 765PW tem- perature probe to the connector on top of the instrument.
ON
OFF
IR
CONTACT
BATTERY REPLACEMENT
Battery replacement must only take place in a safe area and using a 9V alkaline battery.
Simply slide off the battery cover on the back of the meter, detach the battery from the terminals and attach a new 9V battery, while paying attention to the correct polarity. Re­place the battery cover.
CE DECLARATION OF CONFORMITY
READING DISTANCE
The distance between the sensor and the object is an important factor in infrared measurement.
The relation between “distance/object area” is 3:1. This means that the diameter of the measured area is one third of the distance between the sensor and the object.
For example, if the sensor is 15 cm away from the object, the instrument measures the temperature of an area with a 5 cm diameter.
MEASUREMENT MODE
Press the IR/CONTACT button to select the external probe measurement mode: “P” appears on the LCD to indicate that the mode has been entered.
Dip the probe in the substance which to be measured (an immersion level of approximately 4 cm is recommended), and wait for the reading to stabilize.
LOW BATTERY
These instruments continuously check the battery status. If the battery level is too low to ensure realiable reading, the meter shuts off the LCD completely. In this case and if the meter can not be switched on, it is recommended to replace the battery immediately.
ACCESSORIES
HI 710004 Soft carrying case HI 710007 Blue shockproof rubber boot HI 710008 Orange shockproof rubber boot HI 721316 Rugged carrying case HI 731318 Wiping tissue (4 pcs) HI 740016 Battery cover HI 740123 Wrist-strap HI 765PW Penetration thermistor probe with white handle
Recommendations for Users
Before using these products, make sure that they are entirely suitable for the environment in which they are used.
Operation of these instruments in residential area could cause unacceptable interference to radio and TV equipment, requiring the operator to take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the instruments' EMC performance.
To avoid damage or burns, do not perform any measurement in microwave ovens.
IST99556R3 07/05