Instruction Manual
(HI 993301 - HI 993302)
These Instruments are in
www.hannainst.com
Compliance with CE
Directives
WARRANTY
HI 993301 and HI 993302 are warranted for two years against
defects in workmanship and materials when used for their intended purpose and maintained according to instructions. The probe is warranted
for a period of six months. This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed maintenance are not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of purchase,
serial number and the nature of the failure. If the repair is not covered by
the warranty, you will be notified of the charges incurred. If the instrument
is to be returned to Hanna Instruments, first obtain a Returned Goods
Authorization Number from the Customer Service department and then
send it with shipment costs prepaid. When shipping any instrument, make
sure it is properly packaged for complete protection.
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner, Hanna Instruments
Inc., Woonsocket, Rhode Island, 02895 , USA.
Dear Customer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for a correct
operation. Please read it carefully before using the meter.
If you need additional technical information, do not hesitate
to e-mail us at tech@hannainst.com.
These instruments are in compliance with the directives.
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and
examine it carefully. If any damage has occurred during
shipment, immediately notify your Dealer or the nearest
Hanna Customer Service Center.
The meters are supplied with:
• HI 70031 1413 µS/cm calibration solution, for HI 993301;
• HI 70030 12.88 mS/cm calibration solution, for HI 993302;
• Hanna 12 VDC power adapter;
• Instruction manual.
Note: Conserve all packing material until the instrument
has been observed to function correctly. Any defective
item must be returned in its original packing.
GENERAL DESCRIPTION
EC/TDS is a combined EC/TDS/temperature
meter specially designed to meet the needs of growers in
greenhouses and hydroponic applications.
You can simply install the meter above the sample to be
tested for continuous measurement.
The EC/TDS/Temperature probe has been molded with pipe
thread (1/2’’ NPT), which allows the user to attach it to an
in-line system.
Measurements are accurate and the meter can be calibrated
at one point.
You no longer need to worry about battery changes either:
the unit runs without interruption on 12 VDC power supply.
FUNCTIONAL DESCRIPTION
1. Liquid Crystal Display
2. ON/OFF/MODE button
3. Power supply connector
4. 12 VDC power adapter
5. EC/TDS/Temperature probe, pipe thread 1/2’’ NPT
6. SET/HOLD button
SPECIFICATIONS
HI 993301
EC/TDS Range 0 to 3999 µS/cm - 0 to 2000 ppm
Resolution 1 µS/cm - 1 ppm
Accuracy (@20ºC/68ºF) ±2% f.s.
Typical EMC Deviation ±2.5% f.s.
HI 993302
EC/TDS Range 0.00 to 20.00 mS/cm - 0.00 to 10.00 ppt
Resolution 0.01 mS/cm - 0.01 ppt
Accuracy (@20ºC/68ºF) ±2% f.s.
Typical EMC deviation ±2.5% f.s.
Common specifications
Temperature Range 0.0 to 60ºC (32.0 to 140ºF)
Resolution 0.1ºC or 0.1ºF
Accuracy (@20ºC/68ºF) ±0.5 ºC or ±1ºF
Typical EMC Deviation ±1 ºC or ±2 ºF
Temp.Compensation β=0.0 to 2.4%/ºC
TDS Conversion factor 0.45 to 1.00 (CONV)
Calibration Automatic at 1 point
Power supply 12VDC power adapter (included)
Environment 0 to 50ºC (32 to 122ºF);
RH 95% non-condensing
Dimensions (meter only) 160 x 105 x 31 mm
(6.2 x 4.1 x 1.2’’)
Weight (meter only) 190 g (6.7 oz.)
Recommendations for Users
Before using this product, make sure that it is entirely suitable for the
environment in which it is used. Operation of this instrument in residential areas could cause unacceptable interferences to radio and TV equipment.
Any variation introduced by the user to the supplied equipment may degrade
the instrument’s EMC performance.
To avoid electrical shock, do not use this instrument when voltages at the
measurement surface exceed 24 VAC or 60 VDC. To avoid damages or burns,
do not perform any measurement in microwave ovens.
Hanna Instruments reserves the right to modify the design, construction and
appearance of its products without advance notice.
ISTR993301 10/01
OPERATIONAL GUIDE
• Press the MODE button again to show the current
conversion factor. Eg. 0.50 CONV.
CALIBRATION PROCEDURE
ACCESSORIES
To turn the meter on
Connect the 12VDC adapter to the meter and to the mains;
the display will lit. Press and hold the MODE button for 23 seconds. All the used segments on the LCD will be visible
for a few seconds.
To change the temperature unit
To change the temperature unit (from °C to °F), from
measurement mode, press and hold the MODE button until
TEMP and the current temperature unit are displayed on
the lower LCD. Eg. TEMP °C.
Use the SET/HOLD button to change the temperature unit,
and then press MODE button twice to return to normal
measurement mode.
To freeze the display
Press and hold the SET/HOLD button for 2-3 seconds until
HOLD appears on the secondary display.
Press either button to return to normal mode.
Taking measurements
Select either EC or TDS mode with the SET/HOLD button.
Immerse the probe in the solution to
be tested. In order to ensure better
accuracy, probe should not touch or
stand close to the walls or bottom of
the sample vessel.
• Press the SET/HOLD button to change the conversion
factor.
• Press the MODE button to show the current temperature
compensation coefficient β. Eg. 2.1 BETA.
• Press the SET/HOLD button to change the temperature
compensation coefficient β.
• Press the MODE button to return to normal measuring
mode.
To turn the meter off
Press the MODE button while in normal measurement mode.
OFF will appear on the lower part of the display. Release
the button. The display still lit, until the power supply is
connected.
Notes:
• Before taking any measurement make sure the meter
has been calibrated.
• To clear a previous calibration, press the MODE button
after entering the calibration mode. The lower LCD will
display ESC for 1 second and the meter will return to
normal measurement mode. The CAL symbol on the LCD
will disappear. The meter will be reset to the default
calibration.
• From measurement mode, press and hold the MODE
button until CAL is displayed on the lower LCD.
• Release the button and immerse the probe in the proper
calibration solution: HI 70031 (1413 µS/cm) for HI 993301
and HI 70030 (12.88 mS/cm) for HI 993302.
• Once the calibration has been automatically performed,
the LCD will display OK for 1 second and the meter will
return to normal measurement mode.
• Since there is a known relationship between EC and TDS
readings, it is not necessary to calibrate the meter in TDS
Note: for storing calibration data in the non-volatile memory,
turn the meter OFF and then ON again through the MODE
button.
The CAL symbol on the LCD means that the meter is
calibrated.
HI 70030P 12.88 mS/cm @25ºC cal. solution, 20 mL sachet (25 pcs)
HI 70031P 1413 µS/cm @25ºC cal. solution, 20 mL sachet (25 pcs)
HI 70032P 1382 ppm @25ºC cal. solution, 20 mL sachet (25 pcs)
HI 70038P 6.44 ppt @25ºC cal. solution, 20 mL sachet (25 pcs)
HI 70442P 1500 ppm @25ºC cal. solution, 20 mL sachet (25 pcs)
HI 7061M Electrode cleaning solution, 230 mL bottle
HI 70300M Electrode storage solution, 230 mL bottle
HI 710005 12VDC power adapter, US plug
HI 710006 12VDC power adapter, European plug
HI 710012 12VDC power adapter, Australian plug
HI 710013 12VDC power adapter, Southern Africa plug
HI 710014 12VDC power adapter, UK plug
The EC or TDS value automatically
compensated for temperature is shown
on the primary LCD while the secondary LCD shows the temperature of the
sample.
The measurements should be taken when the stability
symbol on the top left of the LCD disappears.
To change the EC/TDS conversion factor (CONV)
and the temperature compensation coefficient
(BETA)
• From measurement mode, press and hold the MODE
button until TEMP and the current temperature unit are
displayed on the lower LCD. Eg. TEMP °C.
• If measurements are taken in different samples successively, rinse the probe thoroughly to eliminate
cross-contamination; and after cleaning, rinse the probe
with some of the sample to be measured.