Hanna Instruments HI 9931 User Manual

Instruction Manual
HI 9931 - HI 9934
Wall Mounted
Conductivity & TDS
Controllers
EC
ALARM
1
0
mS
2
4
/
2
0
0
m
/
2
A
0
O
U
T
P
U
T
A
L
A
R
M
SET EC
L
E
A
EC
mS
L
A
HI 9
S
E
T
P
O
IN
T
Z
E
R
O
C
C
O
A
A
L
R
S
E
F
IN
E
S
L
O
P
E
C
A
R
M
AU
T
O
M
A
N
U
A
0
.8
0
m
S P
R
O
P
O
L
S
E
T
L
1
.6
R
T
F
R
E
E
R
A
T.
D
F
E
E
D
4
5
5
2
.5
0
S
9
IO
E
0
C
N
O
A
N
L
D
S
1
S
E
T
T
IN
G
S
M
IN
U
T
E
S
T
IM
E
R
http://www.hannainst.com
These Instruments are in
Compliance with the CE Directives
Dear Customer, Thank you for choosing a Hanna Product. Please read this instruction manual carefully before using the instru-
ment. This manual will provide you with the necessary information for a correct use of the instrument, as well as a more precise idea of its versatility. If you need more technical information, do not hesitate to e-mail us at tech@hannainst.com.
These instruments are in compliance with the EN 50081-1, 50082-1 and 61010-1.
TABLE OF CONTENTSTABLE OF CONTENTS
TABLE OF CONTENTS
TABLE OF CONTENTSTABLE OF CONTENTS
PRELIMINARY EXAMINATION........................................................ 3
GENERAL DESCRIPTION ................................................................3
MECHANICAL LAYOUTS ................................................................. 4
FUNCTIONAL DIAGRAM HI 9931 .................................................. 6
FUNCTIONAL DIAGRAM HI 9934 .................................................. 9
CONNECTIONS & WIRING ..........................................................12
NORMAL OPERATION & MEASUREMENT ..................................... 15
CALIBRATION.............................................................................16
ADJUSTEMENT OF SETPOINT......................................................17
PROPORTIONAL CONTROL.......................................................... 18
OVERDOSAGE TIMER ..................................................................20
PROBE CLEANING & MAINTENANCE ............................................ 20
SUGGESTED INSTALLATIONS FOR EC/TDS PROBES .......................21
ACCESSORIES............................................................................. 23
WARRANTY ................................................................................ 25
OTHER PRODUCTS FROM HANNA ............................................... 26
CE DECLARATION OF CONFORMITY .............................................. 27
directives
ISO 9000 Certified Company since 1992
2
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipping. If there is any noticeable damage, notify your Dealer.
Note: Save all packing materials until you are sure that the
instrument functions correctly. Any defective item must be returned in the original packaging together with the supplied accessories.
IMPORTANT:
1. Read the instructions before using the instrument.
2. The instrument should be connected to a mains socket.
3. Never install the controller outdoors, in a wet or humid area or under direct sun light. Nor install the controller where liquids may be sprayed or poured on it.
4. The instrument’s main power line as well as the dosage and alarm terminals are protected by separate 2A fuses. Use only 2A fuses for replacement.
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
Hanna's wall-mounted conductivity and TDS controllers proportional control are designed to meet a variety of process control requirements, especially those in horticultural, hydroponics and agricultral applications.
The controllers operate with the optional Hanna 4-ring probes to provide a linear and repeatble measurement. The conductivity/TDS probes can be installed quickly and easily. Simply plug the DIN connector into the socket and tighten the retainer ring.
The probes incorporate a tempertaure sensor and the contoller will automatically compensate for the temperature effect. Accurate mea­surements are displayed on a large LCD.
The controllers come equipped with relays operating at a maximum of 2A (240V).
The controllers incorporate a triple contact alarm system. When activated, the alarm contacts will open or close, triggering the mechanism of your choice, whether a buzzer, light or any other electrical device.
3
with
The recorder output terminals are isolated from the controller circuitry to avoid any interference and are user-switchable between 0 to 20 mA or 4 to 20 mA.
The controller is housed in a rugged, modular, fiber-reinforced ABS housing.
All models can be wired to work with 110/115V or 220/240V at 50/ 60 Hz power supplies.
MECHANICAL LAYOUTSMECHANICAL LAYOUTS
MECHANICAL LAYOUTS
MECHANICAL LAYOUTSMECHANICAL LAYOUTS
DIN connector
WIRING ACCESS
PORTS
Fig. 1
Figure 1 illustrates the connector for probe and the wiring access ports.
LCD DISPLAY
CALIBRATION
TRIMMERS
CONTROL KNOBS
AND SWICHES
EC
mS
ALARM
0
COARSE
ZERO CAL
AUTO
MANUAL
0.8
1.6
mS
PROPORTIONAL SETTINGS
SET POINT
45
0
SECONDS
EC
HI 9931
ALARM
1
0
2
mS
4/20 0/20 mA OUTPUT
FINE
SLOPE CAL
ALARM
SET
SET EC
FERT. FEED
READ
2.557.5
90
10
1
MINUTES
TIMER
LINE
Fig. 2
Figure 2 illustrates the controls and terminals.
4
OUTPUT CONTACTS
RECORDER OUTPUT
SWITCH
ALARM CONTACTS
DOSING TERMINALS
POWER SUPPLY
TERMINALS
3.8“
96.9 mm 1“
25 mm
75 mm
Fig. 3
Figure 3 is a dimensioned, bottom view of the wall mounted controller. The modular design isolates the control circuitry from the contacts making it possible to make the connections and then close the compartment. Adjustments can then be made only in the control area, without having to open the contacts compartment.
227.8 mm 9
3“
7.4
188.6 mm
221.8 mm
8.7
6.9
174.8 mm
Fig. 4
Figure 4 is a dimensioned front view of the wall mounted controller. The molded, mounting holes in the corners provide for quick and secure installation. No additional hardware is needed for mounting. All electrical connections and controls are located on the front of the instrument so that adjustment can be made without having to remove the unit.
5
FUNCTIONAL DIAGRAM HI 9931FUNCTIONAL DIAGRAM HI 9931
FUNCTIONAL DIAGRAM HI 9931
FUNCTIONAL DIAGRAM HI 9931FUNCTIONAL DIAGRAM HI 9931
FRONT PANEL
13
EC
ALARM
0
mS
4/20 0/20 mA OUTPUT
ALARM
SET EC
1
2
SET POINT
FINE SLOPE
SET
READ
HI 9931
CAL
FERT. FEED
1
EC
2
mS
3 4 5 6
7
COARSE
ZERO
CAL
AUTO
ALARM
MANUAL
8
45
9
0.8
0
mS
PROPORTIONAL SETTINGS
10
11 12
2.557.5
1.6
90
10
1
0
MINUTES
SECONDS
TIMER
LINE
20
Left panel
1. Liquid Crystal Display
2. Slope calibration trimmer
3. Fine setpoint trimmer
4. Coarse setpoint trimmer
5. Zero calibration trimmer
6. Alarm LED warning signal
7. Alarm disable switch
8. READ for actual measurement and SET for setpoint adjustment
9. Dosing LED signal
10. Proportional conductivity band setting
11. Proportional time cycle setting
12. Overdosage timer
14
15
16
17
18
19
Right panel
13. Conductivity alarm setting from 0 to 2.0 mS/cm (EC)
14. Recorder output contacts
15. 0 to 20 or 4 to 20 mA output selector
16. Triple contact alarm in a Normally Closed (NC) or a Normally Open (NO) position.
6
17. Powered dosage terminals (Relay)
18. 110/115V or 220/240V power configuration
19. Incoming power terminals
20. Fuses
BOTTOM VIEW
21
21. Female DIN connector for conductivity probe
Unplug the instrument from the power supply before wiring or replacing the fuses.
7
Specifications HI 9931 RANGE 0.00 to 10.00 mS/cm (EC)
RESOLUTION 0.01 mS/cm ACCURACY ±2% F. S.
(@20 C/68 F) TYPICAL EMC ±2% F. S.
DEVIATION CALIBRATION Through "ZERO CAL" and "SLOPE CAL" trimmers
SETPOINT RANGE From 0 to 10.00 mS/cm PROPORTIONAL Conductivity from 0.0 to 1.6 mS/cm and
CONTROL time cycle from 0 to 90 seconds ALARM
CONTACTS
DOSING TERMINALS PROBE (optional) 4-ring potentiometric with built-in temperature sensor
POWER SUPPLY 220/240V or110/115V at 50/60Hz ENVIRONMENT -10 to 50°C (14 to 122°F)
WEIGHT 1.6 Kg (3.5 lb.) ENCLOSURE 181 x 221 x 142mm (7.1 x 8.7 x 5.6") CASE MATERIAL Fiber-reinforced, self-extinguishing ABS
Terminal can be configured as normally open or normally closed (isolated output Max. 2A, Max. 240V, resistive load, 1,000,000 strokes). The alarm is activated if conductivity exceeds by more than the user­selectable interval (0 to 2.0 mS/cm) from the setpoint or due to overdosage
Relay (115 to 240V, Max.2A,1,000,000 strokes) is activated whenever conductivity falls below the setpoint
max. 95% RH non-condensing
8
FUNCTIONAL DIAGRAM HI 9934FUNCTIONAL DIAGRAM HI 9934
FUNCTIONAL DIAGRAM HI 9934
FUNCTIONAL DIAGRAM HI 9934FUNCTIONAL DIAGRAM HI 9934
FRONT PANEL
13
TDS
ALARM
200
0
ppm
4/20 0/20 mA OUTPUT
ALARM
SETTDS
400
SET POINT
FINE SLOPE
SET
READ
HI 9934
CAL
FERT. FEED
1
TDS
2
ppm
3 4 5 6
7
COARSE
ZERO
CAL
AUTO
ALARM
MANUAL
8
45
9
200
0
ppm
PROPORTIONAL SETTINGS
2.557.5
10
1
400
90
0
MINUTES
SECONDS
TIMER
LINE
10
11 12
20
Left panel
1. Liquid Crystal Display
2. Slope calibration trimmer
3. Fine setpoint trimmer
4. Coarse setpoint trimmer
5. Zero calibration trimmer
6. Alarm LED warning signal
7. Alarm disable switch
8. READ for actual measurement and SET for setpoint adjustment
9. Dosing LED signal
10. Proportional TDS band setting
11. Proportional time cycle setting
12. Overdosage timer
14
15
16
17
18
19
Right panel
13. TDS alarm setting from 0 to 400 ppm (mg/L)
14. Recorder output contacts
15. 0 to 20 or 4 to 20 mA output selector
16. Triple contact alarm in a Normally Closed (NC) or a Normally
9
Open (NO) position.
17. Powered dosage terminals (Relay)
18. 110/115V or 220/240V power configuration
19. Incoming power terminals
20. Fuses
BOTTOM VIEW
21
21. Female DIN connector for TDS probe
10
Specifications HI 9934 RANGE 0.00 to 1999 ppm (mg/L)
RESOLUTION 1 ppm (mg/L) ACCURACY ±2% F. S.
(@20 C/68 F) TYPICAL EMC ±2% F. S.
DEVIATION CALIBRATION Through "ZERO CAL" and "SLOPE CAL" trimmers SETPOINT RANGE From 0 to 1999 ppm (mg/L) PROPORTIONAL TDS from 0 to 400 ppm and
CONTROL time cycle from 0 to 90 seconds ALARM
CONTACTS
DOSING TERMINALS
TDS RATIO 0.65 ppm (mg/L) = 1 µS/cm PROBE (optional) 4-ring potentiometric with built-in temperature sensor POWER SUPPLY 220/240V or110/115V at 50/60Hz
ENVIRONMENT -10 to 50°C (14 to 122°F)
WEIGHT 1.6 Kg (3.5 lb.) ENCLOSURE 181 x 221 x 142mm (7.1 x 8.7 x 5.6") CASE MATERIAL Fiber-reinforced, self-extinguishing ABS
Terminals can be configured as normally open or normally closed (isolated output Max. 2A, Max. 240V, resistive load, 1,000,000 strokes). The alarm is activated if TDS exceeds by more than the user­selectable interval (0 to 400 ppm) from the setpoint or due to overdosage.
Relay (115 to 240V, Max.2A, 1,000,000 strokes) are activated whenever TDS falls below the setpoint
max. 95% RH non-condensing
11
CONNECTIONS & WIRINGCONNECTIONS & WIRING
CONNECTIONS & WIRING
CONNECTIONS & WIRINGCONNECTIONS & WIRING
GENERAL POINTS
relay terminals relay terminals
• The
relay terminals of the controller are
relay terminals relay terminals means that you can simply hook up your pump or electrovalve directly to the controller and do not need additional power supply.
• Unscrew the 4 screws on the right hand panel and remove the cover and the gasket. Thread the wires through the access ports on the right hand side of the controller.
BeforeBefore
connecting the controller to the mains, wire connecting the controller to the mains, wire
Before
connecting the controller to the mains, wire
BeforeBefore
connecting the controller to the mains, wire connecting the controller to the mains, wire
the controller completelythe controller completely
the controller completely and make all the connections for
the controller completelythe controller completely pumps, alarm, probe, set the alarm threshold and adjust the settings. Upon completion
replace the coverreplace the cover
replace the cover. Only then
replace the coverreplace the cover
connect the controller to the power supply.
poweredpowered
powered. This
poweredpowered
PROBE CONNECTION
DIN connector
WIRING ACCESS
PORTS
• Attach the conductivity/TDS probe (HI 3002, HI 3001D or HI 7638) to the DIN socket located on the bottom of the casing. Align the guide on the connector with the socket, push the connector in and tighten the retainer ring. (HI 3002 is more suitable for direct immersion in the tank, vat or pipes. HI 3001D can be mounted directly into a pipe and HI 7638 is recom­mended for high temperature and pressure applications).
• The probes incorporate a temperature sensor and the controller’s circuitry automatically compensates for the temperature effect.
RELAY CONNECTIONS
• Wire the external devices (pumps or electrovalves) directly to the relay terminal strips of the controller (see 17 - Functional Dia-
12
gram). The need an external power supplyneed an external power supply
need an external power supply for the pumps or valves.
need an external power supplyneed an external power supply
terminals are poweredterminals are powered
terminals are powered and hence you
terminals are poweredterminals are powered
do notdo not
do not
do notdo not
BL DOSING PUMP
ALARM
EC
mS
0.8
0
mS PROPORTIONAL SETTINGS
EC
HI 9931
ALARM
1
0
2
mS
SET POINT
COARSE
ZERO CAL
AUTO
MANUAL
1.6
0
SECONDS
4/20 0/20 mA OUTPUT
FINE
SLOPE
CAL
ALARM
SET
SET EC
FERT. FEED
READ
45
2.557.5
90
10
1
MINUTES TIMER
LINE
ALARM CONNECTIONS
• The operator can select an alarm thresh­old of 0.0 to 2.0 mS/cm for HI 9931 and 0 to 400 ppm for HI 9934 by turning the alarm knob (see 13 ­Functional Diagram). If the actual mea­surements are above the setpoint by a margin greater than the user-select­able alarm threshold, the alarm terminal is activated. The alarm can be selected as normally closed ("NC") by connecting the external device to the C and NC terminals or normally open ("NO") by connecting the external device to the C and NO terminals.
EC
HI 9931
ALARM
SLOPE
CAL
2.557.5
1
MINUTES
1
0
2
mS
4/20 0/20 mA OUTPUT
ALARM
SET EC
FERT. FEED
10
TIMER
LINE
EC
mS
ALARM
0
COARSE
ZERO
CAL
AUTO
MANUAL
0.8
1.6 mS PROPORTIONAL SETTINGS
SET POINT
FINE
SET
READ
45
90
0
SECONDS
TDS
ALARM
200
0
400
ppm
4/20 0/20 mA OUTPUT
EXTERNAL ELECTRICAL DEVICE ACTIVATED BY AN ALARM CONDITION
EC
ALARM
1
0
mS
4/20 0/20 mA OUTPUT
2
• When activated, the alarm contacts will open or close, triggering the mechanism of your choice. When the all other terminals all other terminals
all other terminals (such as dosing relay etc.)
all other terminals all other terminals disactivateddisactivated
disactivated. The alarm LED light also comes on.
disactivateddisactivated
alarm is activatedalarm is activated
alarm is activated
alarm is activatedalarm is activated
areare
are
areare
• The alarm AUTO/MANUAL switch (see 7 - Functional Diagram) is
disable the alarm terminaldisable the alarm terminal
only to
disable the alarm terminal (e.g. the
disable the alarm terminaldisable the alarm terminal buzzer will not sound). However, all other functions such as disactivation of the dosing relay remain
AUTO
ALARM
MANUAL
unvaried, i.e. the pump ceases to dose until the alarm condition is alleviated.
13
• The contoller provides for
automatic fail-safe secruityautomatic fail-safe secruity
automatic fail-safe secruity by
automatic fail-safe secruityautomatic fail-safe secruity activating the alarm if there is a power failure, regardless of whether the NC or NO configurations were chosen.
• The alarm is also activated if the overdosage time is exceeded. The maximum time that the relay contacts remain active continuously can be set from 1 to 10 minutes.
• Once in an alarm condition, the alarm contact
5
2.5
1
MINUTES
TIMER
7.5
10
remains activated until the switch is put in the manual position or the measurements return to normal values.
RECORDER OUTPUT CONNECTIONS
• The recorder output contacts are iso­lated from the controller circuitry to avoid interference. Select 0 to 20 mA or 4 to 20 mA with the switch (see 15 - Functional
Output mA
0/204/20
selector
Diagram) before wiring the recorder to the terminals. The output mA value is proportional and is the measured conductivity or TDS over the entire range. For example, when HI 9931 measures
2.00 mS/cm (EC), the output values are 4 mA or 7.2 mA based on whether the 0-20 or 4-20 output was respectively selected. Likewise, when HI 9934 measures 1500 ppm (mg/L), the output values are 15 mA or 16 mA based on whether the 0-20 or 4-20 output was selected.
RECORDER
EC
mS
ALARM
0
COARSE
ZERO
CAL
AUTO
MANUAL
0.8
1.6 mS PROPORTIONAL SETTINGS
EC
HI 9931
ALARM
1
0
2
mS
FINE
SET
READ
90
SLOPE
CAL
2.557.5
1
MINUTES
TIMER
4/20 0/20 mA OUTPUT
ALARM
SET EC
FERT. FEED
10
LINE
SET POINT
45
0
SECONDS
TEMPERATURE COMPENSATION
Temperature affects conductivity and TDS mesurements considerably (approx. 2% per °C). You however do not need to worry about having to compensate for this or go through complicated calculations since all three probes recommended in this manual (HI 7638, HI 3002 and HI 3001D) compensate for temperature automatically.
14
MAIN POWER SUPPLY CONNECTION
BeforeBefore
• For 220-240V, short the L1 and N1 termi-
connecting the unit to the mains connecting the unit to the mains
Before
connecting the unit to the mains, make sure
BeforeBefore
connecting the unit to the mains connecting the unit to the mains
that the
controller is completely wiredcontroller is completely wired
controller is completely wired and that all
controller is completely wiredcontroller is completely wired
connections for pump, alarm, probe, etc. have been made.
nals. Then wire the external power supply to the three terminals as shown.
220 VAC
Configuration
Neutral
Line
POWER SUPPLY
L L1 N1
• For 110-115V, short the L and L1 termi­nals and the N1 and Neutral. Then wire the external power supply to the three
110 VA C
Configuration
110 VA C
Configuration
Neutral
Line
POWER SUPPLY
terminals as shown.
• Replace the cover with the gasket and screw it tight with the 4 screws provided. Only then connect the controller to the mains.
NORMAL OPERATIONNORMAL OPERATION
NORMAL OPERATION
NORMAL OPERATIONNORMAL OPERATION
& MEASUREMENT& MEASUREMENT
& MEASUREMENT
& MEASUREMENT& MEASUREMENT
Make sure that the controller has been properly calibrated before commencing and that the conductivity/TDS setpoint has been ad­justed (see following pages).
The conductivity/TDS probe must be properly connected and wired to the controller (see preceding pages).
The protective sleeve must not be removed and the holes on the sleeve should be near the top (the cable end). The probe must be immersed in the solution above the air-vent holes on the external sleeve. The probe must be installed in such a way as to minimize presence of air bubbles (see probe installation tips at the end of the manual).
SET
Ensure that the selector (see 8 - Functional Diagram) is in the “READ” mode.
READ
The controller provides for a visual dosing status through the “FERT FEED” LED (see 9 - Functional Diagram).
The actual conductivity (EC) or TDS value will be dis-
FERT. FEED
played on the LCD in mS/cm or ppm, respectively.
L L1 N1
EC
mS
15
TDS
ppm
12401.20
CALIBRATIONCALIBRATION
CALIBRATION
CALIBRATIONCALIBRATION
Make sure that the probe has been properly connected and wired to the controller (see preceding pages) and that the meter is plugged to the mains.
Calibration should be ideally performed at a temperature similar to that of the liquid to be monitored.
SET
Ensure that the controller is in the “READ” position.
READ
ZERO ADJUSTMENT:
• Leave the probe in the air and make sure that it is dry.
• The LCD should show 0. If neces­sary adjust the "ZERO CAL" trimmer to display zero.
ZERO
COARSE
CAL
SET POINT
0
FINE
SLOPE
CAL
SLOPE ADJUSTMENT:
• Pour sufficient amount of a known conductivity or TDS solution in a beaker. The solution should be close to the sample stream to be monitored. For example, when using the HI 9931, choose HI 7031L with a 1.41 mS/cm value at 25 °C if the measure­ments are in the 1.2 to 2.5 EC range. Similarly, utilize HI 70442L with a value of 1500 ppm at 25 °C if controlling your procress with HI 9934.
• Immerse the probe in the beaker esuring that the holes on the sleeve of the probe are completely covered.
• Stir the probe and gently tap it on the bottom of the beaker to ensure that any air bubbles trapped inside the sleeve escape. Best results are obtained when the probe is not too close to the walls of the beaker nor lying on the bottom.
• Wait for the reading to stabilize. Adjust the "SLOPE CAL" trimmer to display the same value as the cali­bration solution at 25 °C. For example, with HI 7031L, you should adjust the dispaly to 1.41 mS/cm and with HI 70442 to 1500 ppm.
EC
mS
Min7cm
ZERO
CAL
(2.8“)
1.41
COARSE
SET POINT
FINE
SLOPE
CAL
16
• Once the controller is calibrated to a solution by referring to its value at 25 °C, all subsequent measurements are temperature compensated to 25 °C. You can also obtain temperature compen­sation to a different temperature reference point by calibrating the meter to that value. For example, the conductivity and TDS value of HI 7031L and HI 70442L at 20 °C are 1.28 mS/cm and 1358 ppm, respectively. By adjusting the display to these two values, all subsequent measurements are compensated to a base temperature of 20 °C.
Calibration is now complete.
ADJUSTEMENT OF SETPOINTADJUSTEMENT OF SETPOINT
ADJUSTEMENT OF SETPOINT
ADJUSTEMENT OF SETPOINTADJUSTEMENT OF SETPOINT
Make sure that the probe is properly installed and calibrated (see the preceding pages).
Turn the switch to the "SET" position (see 8 - Functional Diagram). The display will show the previously adjusted (e.g. 1.60 mS/cm or 1400 ppm).
SET
READ
EC
mS
1.60
Using a small screwdriver, first adjust the setpoint through the COARSE trimmer and then fine tune it to your required level with the FINE trimmer (see 4 and 3 - Functional Diagram) until the desired value is displayed (e.g. 2.1 mS/cm or 1700 ppm).
EC
1.60
mS
SET POINT
ZERO
COARSE
CAL
FINE
SLOPE
CAL
EC
2.10
mS
SET POINT
ZERO
COARSE
CAL
17
FINE
SLOPE
CAL
PROPORTIONAL CONTROLPROPORTIONAL CONTROL
PROPORTIONAL CONTROL
PROPORTIONAL CONTROLPROPORTIONAL CONTROL
In order to optimize the controlling process and reduce the amount of chemicals and fertilizers used, it is recommended to use a proportional dosage appropriate for the system.
0.8
45
45200
0
mS
PROPORTIONAL SETTINGS
1.6
0
SECONDS
90
0
400
ppm
PROPORTIONAL SETTINGS
0
SECONDS
HI 9931 and HI 9934 allow for a proportional band (delta) of 0 to
1.6 mS/cm and 0 to 400 ppm for conductivity (EC) and TDS,
respectively. Each controller also provides for a time cycle from 0 to 90 seconds. The proportional dosage is obtained by personalizing a curent pulse whose height equals the EC or TDS proportional delta and the length corresponds to the selected time cycle.
The controller will enter proportional dosage at the setpoint minus the preselected delta. It will then keep the dosing relay activated for a period proportional to the difference between the actual measurement less the setpoint over the cycle.
If the setting is left at 0 mS/cm or 0 ppm, the controller will operate as an ON/OFF control with no proportional dosage. The hysteresis of such an ON/OFF control will be 0.1 mS/cm or 15 ppm, respectively.
e.g. HI 9931 proportional control
Setpoint = 2.2 mS/cm (EC) Measured value = 1.6 mS/cm Delta = 2.2 - 1.6 = 0.6 mS/cm Proportional settings = EC set to 0.8 and time cycle to 60 seconds
0.8
45
90
0
mS
PROPORTIONAL SETTINGS
1.6
0
SECONDS
90
The controller will be dosing fertilizers to increase the EC to the desired
18
D mS
0.80
0.60
0.40
0.20
60
15 30 45 60
Seconds15 30 45
limit. Since it is 0.6/0.8 = 75% away from the ideal setting, it will keep the dosing relay activated for 75% of the time over the preselected 60 seconds. The terminals are hence theoretically activated for 45 seconds
and off for 15 seconds. In order to avoid over dosage of highly
concentrated chemicals or fertilizers, the controller provides even a more accurate control.
As the graphs show, it does that by stopping the dosage as soon as the cur­rent pulse curve intersects the dosage curve. This means shortening the dosage period if the chemicals have reacted quickly or lengthening it if the measured conductiv­ity continues to drift from the ideal setpoint as can be seen from the graphs.
D mS
0.80
0.60
0.40
0.20
15 30 45 60
15 30 45 60
Seconds15 30 45
Seconds15 30 45
60
D mS
0.80
0.60
0.40
0.20
60
e.g. HI 9934 proportional control
Setpoint = 1600 ppm TDS) Measured value = 1550 ppm Delta = 1600 - 1550 = 50 ppm Proportional settings = ppm set to 200 and time cycle to 60 seconds
45200
0
400
ppm
PROPORTIONAL SETTINGS
0
SECONDS
90
The controller will be dosing fertilizers to reach the desired value. Since it is 50/200 = 25% away from the ideal setting, it will keep the dosing relay activated for 25% of the time over the 60 second period. The terminals are hence activated for 15 seconds and off for 45 seconds until the next cycle.
19
D ppm
200
150
100
50
60
15 30 45 60
Seconds15 30 45
OVERDOSAGE TIMEROVERDOSAGE TIMER
OVERDOSAGE TIMER
OVERDOSAGE TIMEROVERDOSAGE TIMER
HI 9931 and HI 9934 provide for an overdosage alarm system. The operator can set the maximum amount of time that the dosage terminals should continuously remain activated from 1 to 10 minutes . Should this period elapse, the alarm terminals are activated (and dosage
5
2.5
7.5
disactivated to ensure that fertilizers or chemi­cals have not run out or pumps or electrovalves have not ceased to function
1
MINUTES
10
TIMER
properly).
PROBE CLEANINGPROBE CLEANING
PROBE CLEANING
PROBE CLEANINGPROBE CLEANING
& MAINTENANCE& MAINTENANCE
& MAINTENANCE
& MAINTENANCE& MAINTENANCE
PREPARATION
Make sure that the protective sleeve is on the probe shaft and that the air-vent holes are on the top (nearer to the cable).
STORAGE
The conductivity/TDS probes should be stored dry. If the probes are not to be used for a while, clean and dry them throroughly (see below) before storing them in a dry store room.
PERIODIC MAINTENANCE
Inspect the probe and the cable. The cable used for the connection to the controller must be intact and there must be no points of broken insulation on the cable or cracks on the probe sleeve.
Connector must be clean and dry. If any cracks are present, replace the probe and cable.
CLEANING PROCEDURE
Remove the sleeve and soak the probe in a Hanna HI7061 General Cleaning Solution for 1 hour. Wash the sleeve with plenty of water and remove any dirt.
If the probe has been left in highly concentrated fertilizer solution and does not seem to become clean, repeat the cleaning procedure.
The rings can also be cleaned with a cloth. Make sure that the cloth is made of a soft and non-abrasive material and does not scratch the rings.
20
IMPORTANT: After performing the cleaning procedures rinse the probe thoroughly with distilled or tap water. Dry it and replace the sleeve ensuring that the air-vents are nearer to the cable end and recalibrate the controller.
TROUBLESHOOTING
If the controller does not respond properly or constantly reads zero or a value close to it, or is in over range, check the probe and cable for cracks. The cable may be shorted or the probe broken, in which case they need to be replaced. If the response seems sluggish, follow the cleaning procedure above.
If there are anomolies such as numbers fluctuating, enure that the probe has been properly mounted and that it is in a pipe or well continuously filled with the solution. Air bubbles also disturb the measurements (especially with HI 3001D) and the probe should be installed in such a way as to minimize their presence (see Suggested Installations below).
Note: With industrial applications, it is always recommended to keep
a spare probe handy. When anomalies are not resolved with a simple cleaning or maintenance, change the probe (and recalibrate the controller) to see if the problem is alleviated.
SUGGESTED INSTALLATIONSSUGGESTED INSTALLATIONS
SUGGESTED INSTALLATIONS
SUGGESTED INSTALLATIONSSUGGESTED INSTALLATIONS
for EC/TDS PROBESfor EC/TDS PROBES
for EC/TDS PROBES
for EC/TDS PROBESfor EC/TDS PROBES
As with any other liquid contact measurement, ensure that the probe is installed in such a way that it permanently lies in the solution, whether in the tank or the discharge pipe.
HI 3001D FOR IN-LINE INSTALLATION
HI 3001D is suitabe for this purpose due to its external 1/2” NPT thread, 6 Bar (87 psi) pressure rating and the short immersion depth of 20 mm (0.8”). It can be installed in 1/2” pipes (with the aid of a “T”). For larger pipe sizes, HI 3001D can be mounted upside down or vertically.
1/2Thread
21
HI 3002 FOR IN-LINE INSTALLATION
The drawing illustrates the ideal installation system since the stream pressure in the pipe forces the air bubbles out automatically. HI 3001D illustrated previously can also be ,mounted in this fashion. However the longer stem of HI 3002 facilitates an easier escape route for any air bubbles. HI 3002 comes with external 1/2” NPT and thread 60 mm (2.4”) immersion level.
HI 3002
>1/2
GAS
HI 7638 FOR IN-LINE INSTALLATION
HI 7638 comes with stainless steel 3/8” NPT external thread and 60 mm (2.4”) immersion level. It can work at tempertaures of up to 120°C and should be used in high temperature/pressure applica­tions.
HI 3002 FOR TANK INSTALLATION
Due to its Ultem® and sealed body, HI 3002 can be installed directly in a tank or vat without requiring a probe holder.
22
ACCESSORIESACCESSORIES
ACCESSORIES
ACCESSORIESACCESSORIES
EC/TDS PROBE
HI 3001D 1/2” NPT platinum 20 mm (0.8”) long, 4-ring probe
with temperature sensor and external threads for pipes
HI 3002 1/2” NPT platinum 60 mm (2.4”) 4-ring probe with
temperature sensor and external threads for pipes and tanks
HI 7638 3/8” NPT platinum 60 mm (2.4”) 4-ring probe with
temperature sensor and stainless steel external threads for high temperature and pressure
Hanna manufactures an extensive range of probes for process, water treatment and fertilizer applications. Consult the specific handbooks for process instrumentation, or simply call the Hanna Office nearest to you for a complete list.
EC CALIBRATION SOLUTIONS
HI 7031L 1.41 mS/cm EC calibration solution, 460 mL HI 7039L 5.00 mS/cm EC calibration solution, 460 mL
TDS CALIBRATION SOLUTION
HI 70442L 1500 ppm (mg/L) TDS calibration solution, 460 mL
PROBE CLEANING SOLUTION
HI 7061L General purpose cleaning solution, 460 mL
PUMPS
BL PUMPS Dosing pumps (several models are available with
flow rates from 1.5 to 18.3 lph / 0.4 to 4.8 gph)
23
24
WARRANTYWARRANTY
WARRANTY
WARRANTYWARRANTY
All Hanna controllers are warranted for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions.
Damages due to accident, misuse, tampering or lack of prescribed maintenance are not covered. This warranty is limited to free of charge repair or replacement of the meter only, if any malfunctioning is due to manufacturing defects.
If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization Number from the Cus­tomer Service department and then send it with shipment costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protection.
To validate your warranty, fill out and return the enclosed warranty card within 14 days from the date of purchase.
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc., 584 Park East Drive, Woonsocket, Rhode Island, 02895 , USA.
Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.
25
OTHER PRODUCTS FROM HANNAOTHER PRODUCTS FROM HANNA
OTHER PRODUCTS FROM HANNA
OTHER PRODUCTS FROM HANNAOTHER PRODUCTS FROM HANNA
• CABLES AND CONNECTORS
• CALIBRATION AND MAINTENANCE SOLUTIONS
• CHEMICAL TEST KITS
• CHLORINE METERS
• CONDUCTIVITY/TDS METERS
• DISSOLVED OXYGEN METERS
• HYGROMETERS
• ION SPECIFIC METERS (Colorimeters)
• MAGNETIC STIRRERS
• Na/NaCl METERS
• pH/ORP/Na ELECTRODES
• pH METERS
• PROBES (DO, µS/cm, RH, T, TDS)
• PUMPS
• REAGENTS
• SOFTWARE
• THERMOMETERS
• TITRATORS
• TRANSMITTERS
• TURBIDITY METERS
• Wide Range of ACCESSORIES
Most Hanna meters are available in the following formats:
• BENCH-TOP METERS
• POCKET-SIZED METERS
• PORTABLE METERS
• PRINTING/LOGGING METERS
• PROCESS METERS (Panel and Wall-mounted)
• WATERPROOF METERS
• METERS FOR FOOD INDUSTRY
For additional information, contact your dealer or the nearest Hanna Customer Service Center. You can also e-mail us at tech@hannainst.com.
26
CE DECLARATION OF CONFORMITYCE DECLARATION OF CONFORMITY
CE DECLARATION OF CONFORMITY
CE DECLARATION OF CONFORMITYCE DECLARATION OF CONFORMITY
DECLARATION OF CONFORMITY
We
Hanna Instruments Srl V.ledelle industrie 12 35010 Ronchi di Villafranca (PD) ITALY
herewith certify that the wall-mounted instrument:
HI 9931 HI 9934
has been tested and found to be in compliance with the following regulations:
IEC 801-2 IEC 801-3 IEC 801-4 EN 55022 EN 61010-1
Electrostatic Discharge RF Radiated Fast Transient Radiated, Class B User Safety Requirement
Date of Issue: 07-06-1999
D.Volpato- Engineering Manager
On behalf of
Hanna Instruments S.r.l.
Recommendations for Users
Before using these products, make sure that they are entirely suitable for the environ­ment in which they are used.
Operation of these instruments in residential areas could cause unacceptable interfer­ence to radio and TV equipment.
Any variation introduced by the user to the supplied equipment may degrade the instruments' EMC performance.
Unplug the instruments from power supply before replacing the fuse or making any electrical connections.
27
HANNA LITERATUREHANNA LITERATURE
HANNA LITERATURE
HANNA LITERATUREHANNA LITERATURE
Hanna publishes a wide range of catalogs and hand­books for an equally wide range of applications. The reference literature currently covers areas such as:
• Water Treatment
• Process
• Swimming Pools
• Agriculture
• Food
• Laboratory
• Thermometry
and many others. New reference material is constantly being added to the library.
For these and other catalogs, handbooks and leaflets, contact your dealer or the Hanna Customer Service Center nearest to you. To find the Hanna Office in your vicinity, check our home page at www.hannainst.com.
http://www.hannainst.com
28
PRINTED IN PORTUGAL
MAN9931 10/99
Loading...