Hanna Instruments HI 9931 User Manual

Instruction Manual
HI 9931 - HI 9934
Wall Mounted
Conductivity & TDS
Controllers
EC
ALARM
1
0
mS
2
4
/
2
0
0
m
/
2
A
0
O
U
T
P
U
T
A
L
A
R
M
SET EC
L
E
A
EC
mS
L
A
HI 9
S
E
T
P
O
IN
T
Z
E
R
O
C
C
O
A
A
L
R
S
E
F
IN
E
S
L
O
P
E
C
A
R
M
AU
T
O
M
A
N
U
A
0
.8
0
m
S P
R
O
P
O
L
S
E
T
L
1
.6
R
T
F
R
E
E
R
A
T.
D
F
E
E
D
4
5
5
2
.5
0
S
9
IO
E
0
C
N
O
A
N
L
D
S
1
S
E
T
T
IN
G
S
M
IN
U
T
E
S
T
IM
E
R
http://www.hannainst.com
These Instruments are in
Compliance with the CE Directives
Dear Customer, Thank you for choosing a Hanna Product. Please read this instruction manual carefully before using the instru-
ment. This manual will provide you with the necessary information for a correct use of the instrument, as well as a more precise idea of its versatility. If you need more technical information, do not hesitate to e-mail us at tech@hannainst.com.
These instruments are in compliance with the EN 50081-1, 50082-1 and 61010-1.
TABLE OF CONTENTSTABLE OF CONTENTS
TABLE OF CONTENTS
TABLE OF CONTENTSTABLE OF CONTENTS
PRELIMINARY EXAMINATION........................................................ 3
GENERAL DESCRIPTION ................................................................3
MECHANICAL LAYOUTS ................................................................. 4
FUNCTIONAL DIAGRAM HI 9931 .................................................. 6
FUNCTIONAL DIAGRAM HI 9934 .................................................. 9
CONNECTIONS & WIRING ..........................................................12
NORMAL OPERATION & MEASUREMENT ..................................... 15
CALIBRATION.............................................................................16
ADJUSTEMENT OF SETPOINT......................................................17
PROPORTIONAL CONTROL.......................................................... 18
OVERDOSAGE TIMER ..................................................................20
PROBE CLEANING & MAINTENANCE ............................................ 20
SUGGESTED INSTALLATIONS FOR EC/TDS PROBES .......................21
ACCESSORIES............................................................................. 23
WARRANTY ................................................................................ 25
OTHER PRODUCTS FROM HANNA ............................................... 26
CE DECLARATION OF CONFORMITY .............................................. 27
directives
ISO 9000 Certified Company since 1992
2
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipping. If there is any noticeable damage, notify your Dealer.
Note: Save all packing materials until you are sure that the
instrument functions correctly. Any defective item must be returned in the original packaging together with the supplied accessories.
IMPORTANT:
1. Read the instructions before using the instrument.
2. The instrument should be connected to a mains socket.
3. Never install the controller outdoors, in a wet or humid area or under direct sun light. Nor install the controller where liquids may be sprayed or poured on it.
4. The instrument’s main power line as well as the dosage and alarm terminals are protected by separate 2A fuses. Use only 2A fuses for replacement.
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
Hanna's wall-mounted conductivity and TDS controllers proportional control are designed to meet a variety of process control requirements, especially those in horticultural, hydroponics and agricultral applications.
The controllers operate with the optional Hanna 4-ring probes to provide a linear and repeatble measurement. The conductivity/TDS probes can be installed quickly and easily. Simply plug the DIN connector into the socket and tighten the retainer ring.
The probes incorporate a tempertaure sensor and the contoller will automatically compensate for the temperature effect. Accurate mea­surements are displayed on a large LCD.
The controllers come equipped with relays operating at a maximum of 2A (240V).
The controllers incorporate a triple contact alarm system. When activated, the alarm contacts will open or close, triggering the mechanism of your choice, whether a buzzer, light or any other electrical device.
3
with
The recorder output terminals are isolated from the controller circuitry to avoid any interference and are user-switchable between 0 to 20 mA or 4 to 20 mA.
The controller is housed in a rugged, modular, fiber-reinforced ABS housing.
All models can be wired to work with 110/115V or 220/240V at 50/ 60 Hz power supplies.
MECHANICAL LAYOUTSMECHANICAL LAYOUTS
MECHANICAL LAYOUTS
MECHANICAL LAYOUTSMECHANICAL LAYOUTS
DIN connector
WIRING ACCESS
PORTS
Fig. 1
Figure 1 illustrates the connector for probe and the wiring access ports.
LCD DISPLAY
CALIBRATION
TRIMMERS
CONTROL KNOBS
AND SWICHES
EC
mS
ALARM
0
COARSE
ZERO CAL
AUTO
MANUAL
0.8
1.6
mS
PROPORTIONAL SETTINGS
SET POINT
45
0
SECONDS
EC
HI 9931
ALARM
1
0
2
mS
4/20 0/20 mA OUTPUT
FINE
SLOPE CAL
ALARM
SET
SET EC
FERT. FEED
READ
2.557.5
90
10
1
MINUTES
TIMER
LINE
Fig. 2
Figure 2 illustrates the controls and terminals.
4
OUTPUT CONTACTS
RECORDER OUTPUT
SWITCH
ALARM CONTACTS
DOSING TERMINALS
POWER SUPPLY
TERMINALS
3.8“
96.9 mm 1“
25 mm
75 mm
Fig. 3
Figure 3 is a dimensioned, bottom view of the wall mounted controller. The modular design isolates the control circuitry from the contacts making it possible to make the connections and then close the compartment. Adjustments can then be made only in the control area, without having to open the contacts compartment.
227.8 mm 9
3“
7.4
188.6 mm
221.8 mm
8.7
6.9
174.8 mm
Fig. 4
Figure 4 is a dimensioned front view of the wall mounted controller. The molded, mounting holes in the corners provide for quick and secure installation. No additional hardware is needed for mounting. All electrical connections and controls are located on the front of the instrument so that adjustment can be made without having to remove the unit.
5
FUNCTIONAL DIAGRAM HI 9931FUNCTIONAL DIAGRAM HI 9931
FUNCTIONAL DIAGRAM HI 9931
FUNCTIONAL DIAGRAM HI 9931FUNCTIONAL DIAGRAM HI 9931
FRONT PANEL
13
EC
ALARM
0
mS
4/20 0/20 mA OUTPUT
ALARM
SET EC
1
2
SET POINT
FINE SLOPE
SET
READ
HI 9931
CAL
FERT. FEED
1
EC
2
mS
3 4 5 6
7
COARSE
ZERO
CAL
AUTO
ALARM
MANUAL
8
45
9
0.8
0
mS
PROPORTIONAL SETTINGS
10
11 12
2.557.5
1.6
90
10
1
0
MINUTES
SECONDS
TIMER
LINE
20
Left panel
1. Liquid Crystal Display
2. Slope calibration trimmer
3. Fine setpoint trimmer
4. Coarse setpoint trimmer
5. Zero calibration trimmer
6. Alarm LED warning signal
7. Alarm disable switch
8. READ for actual measurement and SET for setpoint adjustment
9. Dosing LED signal
10. Proportional conductivity band setting
11. Proportional time cycle setting
12. Overdosage timer
14
15
16
17
18
19
Right panel
13. Conductivity alarm setting from 0 to 2.0 mS/cm (EC)
14. Recorder output contacts
15. 0 to 20 or 4 to 20 mA output selector
16. Triple contact alarm in a Normally Closed (NC) or a Normally Open (NO) position.
6
17. Powered dosage terminals (Relay)
18. 110/115V or 220/240V power configuration
19. Incoming power terminals
20. Fuses
BOTTOM VIEW
21
21. Female DIN connector for conductivity probe
Unplug the instrument from the power supply before wiring or replacing the fuses.
7
Specifications HI 9931 RANGE 0.00 to 10.00 mS/cm (EC)
RESOLUTION 0.01 mS/cm ACCURACY ±2% F. S.
(@20 C/68 F) TYPICAL EMC ±2% F. S.
DEVIATION CALIBRATION Through "ZERO CAL" and "SLOPE CAL" trimmers
SETPOINT RANGE From 0 to 10.00 mS/cm PROPORTIONAL Conductivity from 0.0 to 1.6 mS/cm and
CONTROL time cycle from 0 to 90 seconds ALARM
CONTACTS
DOSING TERMINALS PROBE (optional) 4-ring potentiometric with built-in temperature sensor
POWER SUPPLY 220/240V or110/115V at 50/60Hz ENVIRONMENT -10 to 50°C (14 to 122°F)
WEIGHT 1.6 Kg (3.5 lb.) ENCLOSURE 181 x 221 x 142mm (7.1 x 8.7 x 5.6") CASE MATERIAL Fiber-reinforced, self-extinguishing ABS
Terminal can be configured as normally open or normally closed (isolated output Max. 2A, Max. 240V, resistive load, 1,000,000 strokes). The alarm is activated if conductivity exceeds by more than the user­selectable interval (0 to 2.0 mS/cm) from the setpoint or due to overdosage
Relay (115 to 240V, Max.2A,1,000,000 strokes) is activated whenever conductivity falls below the setpoint
max. 95% RH non-condensing
8
FUNCTIONAL DIAGRAM HI 9934FUNCTIONAL DIAGRAM HI 9934
FUNCTIONAL DIAGRAM HI 9934
FUNCTIONAL DIAGRAM HI 9934FUNCTIONAL DIAGRAM HI 9934
FRONT PANEL
13
TDS
ALARM
200
0
ppm
4/20 0/20 mA OUTPUT
ALARM
SETTDS
400
SET POINT
FINE SLOPE
SET
READ
HI 9934
CAL
FERT. FEED
1
TDS
2
ppm
3 4 5 6
7
COARSE
ZERO
CAL
AUTO
ALARM
MANUAL
8
45
9
200
0
ppm
PROPORTIONAL SETTINGS
2.557.5
10
1
400
90
0
MINUTES
SECONDS
TIMER
LINE
10
11 12
20
Left panel
1. Liquid Crystal Display
2. Slope calibration trimmer
3. Fine setpoint trimmer
4. Coarse setpoint trimmer
5. Zero calibration trimmer
6. Alarm LED warning signal
7. Alarm disable switch
8. READ for actual measurement and SET for setpoint adjustment
9. Dosing LED signal
10. Proportional TDS band setting
11. Proportional time cycle setting
12. Overdosage timer
14
15
16
17
18
19
Right panel
13. TDS alarm setting from 0 to 400 ppm (mg/L)
14. Recorder output contacts
15. 0 to 20 or 4 to 20 mA output selector
16. Triple contact alarm in a Normally Closed (NC) or a Normally
9
Loading...
+ 19 hidden pages