Dear Customer,
Thank you for choosing a Hanna Product.
Please read this instruction manual carefully before using the instru-
ment. This manual will provide you with the necessary information for
a correct use of the instrument, as well as a more precise idea of its
versatility. If you need more technical information, do not hesitate to
e-mail us at tech@hannainst.com.
These instruments are in compliance with the
EN 50081-1, 50082-1 and 61010-1.
CE DECLARATION OF CONFORMITY .............................................. 35
directives
ISO 9000 Certified
Company since 1992
2
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it
carefully to make sure that no damage has occurred during shipping. If there is any noticeable damage, notify your Dealer.
Note: Save all packing materials until you are sure that the
instrument functions correctly. Any defective item must be
returned in the original packaging together with the supplied
accessories.
IMPORTANT:
1. Read the instructions before using the instrument.
2. The instrument should be connected to a mains socket.
3. Never install the controller outdoors, in a wet or humid area or
under direct sun light. Nor install the controller where liquids
may be sprayed or poured on it.
4. The instrument’s main power line as well as the dosage and
alarm terminals are protected by separate 2A fuses. Use only 2A
fuses for replacement.
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
Hanna's wall-mounted conductivity/TDS and pH controllers
designed to meet a variety of process control requirements, especially
those in horticultural, hydroponics and agricultural applications on
the one hand and boilers and cooling towers on the other.
All controllers provide for two separate relays, one for pH and another
for Conductivity/TDS. HI 9913 and HI 9935 also come with proportional control to save on use of chemicals and fertilizers.
All models operate with the optional Hanna 4-ring probes to provide
a linear and repeatable measurement. The conductivity/TDS probes
can be installed quickly and easily. Simply plug the DIN connector
into the socket and tighten the retainer ring.
The EC/TDS probes incorporate a temperature sensor and the controllers will automatically compensate for the temperature effect. Accurate
measurements are displayed on a large LCD.
Similarly, the pH electrode can be installed quickly and easily. Simply
plug the universal BNC connector into the socket and twist it into a
secured position. In order to avoid electrical noise and interference,
the pH circuitry also provides for a ground probe (differential input).
3
are
The controllers come equipped with relays operating at a maximum of
2A (240V).
They incorporate a triple contact alarm system. When activated, the
alarm contacts will open or close, triggering the mechanism of your
choice, whether a buzzer, light or any other electrical device.
The controllers are housed in a rugged, modular, fiber-reinforced ABS
housing.
All models can be wired to work with 110/115V or 220/240V at
50/60 Hz power supplies.
MECHANICAL LAYOUTSMECHANICAL LAYOUTS
MECHANICAL LAYOUTS
MECHANICAL LAYOUTSMECHANICAL LAYOUTS
BNC CONNECTOR
MATCHING PIN FOR GROUND PROBE
DIN CONNECTOR
WIRING ACCESS
PORTS
Fig. 1
Figure 1 illustrates the connector for pH electrode, matching pin
connection, conductivity/TDS probe and the wiring access ports.
MATCHING PIN
LCD DISPLAY
CALIBRATION
TRIMMERS
CONTROL KNOBS
AND SWICHES
pH
OFFSET SLOPE
0.5 1.5
0
EC
mS
ZERO CAL
0.5 1.5
0
2.0
pH
PROPORTIONAL SETTINGS
2.0 0
mS
PROPORTIONAL SETTINGS
HI 9913
pH
1
ALARM
pH
FEED
ACID
0
2
6
5
ALARM
ALARM
ACID
ON
47
012
SET
OFF
5
mS
ALK
7.5
2.5
ALARM
90
10
1
0
SECONDS
MINUTES
TIMER
SET pH
ALARM
mS
FEED
FERT
SET EC
3
2
SLOPE CAL
ON
14
ALARM
SET
5
OFF
2.5
7.5
10
90
1
SECONDS
MINUTES
LINE
TIMER
ALARM CONTACTS
DOSING TERMINALS
POWER SUPPLY
TERMINALS
Fig. 2
Figure 2 illustrates the controls and terminals.
4
3.8“
96.9 mm
1“
25 mm
75 mm
Fig. 3
Figure 3 is a dimensioned, bottom view of the wall mounted
controller. The modular design isolates the control circuitry from the
contacts making it possible to make the connections and then close
the compartment. Adjustments can then be made only in the control
area, without having to open the contacts compartment.
227.8 mm
9“
3“
7.4“
188.6 mm
221.8 mm
8.7“
6.9“
174.8 mm
Fig. 4
Figure 4 is a dimensioned front view of the wall mounted controller.
The molded, mounting holes in the corners provide for quick and
secure installation. No additional hardware is needed for mounting.
All electrical connections and controls are located on the front of the
instrument so that adjustment can be made without having to
remove the unit.
5
FUNCTIONAL DIAGRAM HI 9913FUNCTIONAL DIAGRAM HI 9913
FUNCTIONAL DIAGRAM HI 9913
FUNCTIONAL DIAGRAM HI 9913FUNCTIONAL DIAGRAM HI 9913
FRONT PANEL
1
17
2
3
4
5
6
7
8
9
2
10
11
12
13
14
pH
OFFSET SLOPE
0.51.5
0
pH
PROPORTIONAL SETTINGS
EC
mS
ZERO CAL
0.51.5
0
mS
PROPORTIONAL SETTINGS
HI 9913
pH
ACID
5
ALARM
ACID
ON
47
OFF
5
ALK
2.5
2.0
90
1
0
SECONDS
MINUTES
TIMER
mS
FERT
SECONDS
ON
ALARM
OFF
90
16
2
14
5
2.5
1
MINUTES
TIMER
9
SLOPE CAL
2.0 0
pH
1
ALARM
FEED
0
2
6
ALARM
012
SET
mS
7.5
ALARM
10
SET pH
ALARM
FEED
SET EC
3
SET
7.5
10
LINE
25
Left panel
1. Alarm LED signal for pH
2. Liquid Crystal Display for pH and Conductivity (EC)
3. Acid feed LED and dial to adjust the pH setpoint
4. Selection switch for Acid or Alkaline dosage
5. Alarm disable switch for pH
6. Slope calibration trimmer for pH
7. Offset calibration trimmer for pH
8. Proportional pH band and time cycle settings
9. Two independent overdosage timers
10. Alarm LED signal for Conductivity (EC)
11. Fertilizer feed LED and dial to adjust the Conductivity setpoint
12. Zero calibration trimmer for Conductivity (EC)
13. Slope calibration trimmer for Conductivity (EC)
14. Proportional Conductivity band and time cycle settings
16. Alarm disable switch for Conductivity (EC)
18
19
20
21
22
23
24
Right panel
17. pH alarm setting from 0.0 to 2.0 pH
18. Short the terminals if a pH ground probe is not in use, or
connect the ground probe wire to the Matching Pin terminal
19. Conductivity alarm setting from 0 to 2.0 mS/cm (EC)
6
20. Triple contact alarm in a Normally Closed (NC) or a Normally
Open (NO) position.
21. Powered dosage terminals (Relay) for pH correction
22. Powered dosage terminals (Relay) for EC correction
23. 110/115V or 220/240V power configuration
24. Incoming power terminals
25. Fuses
BOTTOM VIEW
26
27
28
26. Female BNC connector for a combination pH electrode
27. 4-mm Banana socket for the pH ground probe
28. Female DIN connector for Conductivity probe
Unplug the instrument from the power supply before
wiring or replacing the fuses.
7
SpecificationsHI 9913
RANGE0.00 to 14.00 pH and 0.00 to 10.00 mS/cm (EC)
RESOLUTION0.01 pH and 0.01 mS/cm
ACCURACY±0.02 pH and ±2% F. S.
(@20°C/68°F)
TYPICAL EMC±0.1 pH and ±2% F. S.
DEVIATION
CALIBRATION
SETPOINT RANGEFrom 4.0 to 7.0 pH and 1.0 to 4.0 mS/cm (EC)
PROPORTIONAL
CONTROL
ALARM
CONTACT
PROBE
DOSING
TERMINALS
POWER SUPPLY 220/240V or 110/115V at 50/60Hz
ENVIRONMENT-10 to 50°C (14 to 122°F)
WEIGHT1.6 Kg (3.5 lb.)
ENCLOSURE181 x 221 x 142mm (7.1 x 8.7 x 5.6")
CASE MATERIALFiber-reinforced, self-extinguishing ABS
Through "OFFSET" and "SLOPE" trimmers for pH, and
“ZERO CAL” and “SLOPE CAL” for Conductivity (EC)
Two independent controls: pH from 0.0 to 2.0 and
Conductivity (EC) from 0.0 to 2.0 mS/cm with two
separate time cycles from 0 to 90 seconds
Terminals can be configured as normally open or
normally closed (isolated output Max. 2A, Max. 240V,
resistive load, 1,000,000 strokes). The alarm is
activated if pH falls below the setpoint by the userselectable interval (0.0 to 2.0 pH), or conductivity
exceeds the setpoint by more than the user-selectable
interval (0 to 2.0 mS/cm) or due to overdosage
Any combination pH electrode with a universal BNC
connector and Hanna Conductivity 4-ring
potentiometric probe with built-in temperature sensor
and DIN connector (optional)
Two sets of independent terminals (115 to 240V,
Max.2A, 1,000,000 strokes) are activated whenever
pH exceeds the pH setpoint and/or conductivity falls
below the EC setpoint
max. 95% RH non-condensing
8
FUNCTIONAL DIAGRAM HI 9923FUNCTIONAL DIAGRAM HI 9923
FUNCTIONAL DIAGRAM HI 9923
FUNCTIONAL DIAGRAM HI 9923FUNCTIONAL DIAGRAM HI 9923
FRONT PANEL
17
1
2
3
5
pH
6
7
OFFSET SLOPE
9
10
11
12
13
15
mS
0
ZERO CAL
DEAD BAND
mS
HI 9923
ALK
AUTO
510
45
MANUAL
1
MINUTES
pH
TIMER
mS
2
AUTO
SLOPE CAL
0.50
1
45
MANUAL
1
MINUTES
S
m TIMER
pH
1
ALARM
pH
0
2
FEED
ALARM
7.5
012
mS
SET
ALARM
90
SET pH
ALARM
BLEED
SET EC
4
3
5
6
SET
90
LINE
18
19
20
21
22
23
24
9
16
25
Left panel
1. Alarm LED signal for pH
2. Liquid Crystal Display for pH and Conductivity
3. Alkaline feed LED and dial to adjust the pH setpoint
5. Alarm disable switch for pH
6. Slope calibration trimmer for pH
7. Offset calibration trimmer for pH
9. Two independent overdosage timers
10. Alarm LED signal for Conductivity
11. Bleed LED and dial to adjust the Conductivity setpoint
12. Zero calibration trimmer for Conductivity
13. Slope calibration trimmer for Conductivity
15. Conductivity Dead Band setting
16. Alarm disable switch for Conductivity
Right panel
17. pH alarm setting from 0.0 to 2.0 pH
18. Short the terminals if a pH ground probe is not in use, or
connect the ground probe wire to the Matching Pin terminal
19. Conductivity alarm setting from 0 to 2.0 mS/cm
20. Triple contact alarm in a Normally Closed (NC) or a Normally
Open (NO) position.
21. Powered dosage terminals (Relay) for pH correction
22. Powered dosage terminals (Relay) for Conductivity correction
9
23. 110/115V or 220/240V power configuration
24. Incoming power terminals
25. Fuses
BOTTOM VIEW
26
27
28
26. Female BNC connector for a combination pH electrode
27. 4-mm Banana socket for the pH ground probe
28. Female DIN connector for conductivity probe
Unplug the instrument from the power supply before
wiring or replacing the fuses.
10
SpecificationsHI 9923
RANGE 0.00 to 14.00 pH and 0.00 to 10.00 mS/cm (mmho/cm)
RESOLUTION0.01 pH and 0.01 mS/cm (mmho/cm)
ACCURACY±0.02 pH and ±2% F. S.
(@20°C/68°F)
TYPICAL EMC±0.1 pH and ±2% F. S.
DEVIATION
CALIBRATION
SETPOINT RANGE From 5.0 to 10.0 pH & 1.00 to 6.00 mS/cm (mmho/cm)
DEAD BANDFrom 0.0 to 0.5 mS/cm (mmho/cm)
ALARM
CONTACT
PROBE
DOSING
TERMINALS
POWER SUPPLY 220/240V or 110/115V at 50/60Hz
ENVIRONMENT-10 to 50°C (14 to 122°F)
WEIGHT1.6 Kg (3.5 lb.)
ENCLOSURE181 x 221 x 142mm (7.1 x 8.7 x 5.6")
CASE MATERIALFiber-reinforced, self-extinguishing ABS
Through "OFFSET" and "SLOPE" trimmers for pH, and
“ZERO CAL” and “SLOPE CAL” for Conductivity
Terminals can be configured as normally open or
normally closed (isolated output Max. 2A, Max. 240V,
resistive load, 1,000,000 strokes). The alarm is activated
if pH exceeds the setpoint by the user-selectable interval
(0 to 2 pH), or Conductivity falls below the setpoint by
more than the user-selectable interval (0 to 2.0 mS/
cm) or due to overdosage
Any combination pH electrode with a universal BNC
connector and Hanna Conductivity 4-ring
potentiometric probe with built-in temperature sensor
and DIN connector (optional)
Two sets of independent terminals (115 to 240V, Max.
2A, 1,000,000 strokes) are activated whenever pH falls
below the pH setpoint or the conductivity exceeds the
“BLEED” setpoint
max. 95% RH non-condensing