Hanna Instruments HI 991001 User Manual

Instruction Manual
HI 991001
Extended Range Waterproof
pH / Temperature Meter
www.hannainst.com
WARRANTY
All Hanna Instruments meters are warranted for two years against defects in workmanship
and materials when used for their intended pur­pose and maintained according to instructions.
The probes are warranted for a period of six months.
This warranty is limited to repair or replacement free of charge. Damage due to accidents, mis­use, tampering or lack of prescribed maintenance are not covered.
If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of pur­chase, serial number and the nature of the prob­lem. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instru­ments, first obtain a Returned Goods Authoriza­tion Number from the Customer Service depart­ment and then send it with shipment costs pre­paid. When shipping any instrument, make sure it is properly packaged for complete protection.
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner.
Dear Customer, Thank you for choosing a Hanna Instruments
Product. Please read this instruction manual care-
This instrument is in compliance with the
directives.
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipment. If noticeable damage is evident, notify your dealer.
Note: Save all packing material until you
are sure that the instrument func­tions correctly. All defective items must be returned in the original packing together with the supplied accessories.
GENERAL DESCRIPTION
HI 991001 is a portable, microprocessor-
based pH and temperature meter. Main features include: extended pH and
temperature ranges; waterproof and com­pact casing; large dual-line display; low battery detection; automatic pH calibration at one or two points with two memorized buffer sets (standard or NIST); selectable temperature unit (°C or °F).
The pH electrode has a built-in tempera­ture sensor for simultaneous pH and tem­perature readings, and also contains a mini amplifier to render measurements imper­vious to noise and electrical interferences.
Each meter is supplied complete with
HI 1296D pH/temperature probe with DIN
connector and 1 m (3.3’) cable
calibration & cleaning solution sachets
3x1.5V AA alkaline batteries
instruction manual
rugged carrying case.
LCD DESCRIPTION
1. Automatic Temperature Compensation indicator
2. Stability indicator
3. Battery percentage
4. pH calibration indicator
5. Low battery indicator
6. Selectable temperature unit
7. Secondary display
8. Primary display
9. Measuring unit for primary display
ACCESSORIES
HI 1296D Combination amplified pH/
temperature probe with DIN connector and 1 m (3.3’) cable
HI 7004L pH 4.01 buffer solution, 500 mL HI 7006L pH 6.86 buffer solution, 500 mL HI 7007L pH 7.01 buffer solution, 500 mL HI 7009L pH 9.18 buffer solution, 500 mL HI 7010L pH 10.01 solution, 500 mL HI 70300L Storage solution, 500 mL HI 7061L Cleaning solution, 500 mL HI 710007 Blue shockproof rubber boot HI 710008 Orange shockproof rubber boot HI 721312 Rugged carrying case HI 76405 Electrode holder
SPECIFICATIONS
Range (*) -2.00 to 16.00 pH
-5.0 to 105.0°C / 23.0 to 221.0°F
Resolution 0.01 pH
0.1°C / 0.1°F
Accuracy
Typical EMC ±0.02 pH Deviation ±0.2°C or ±0.4°F
Temperature Automatic Compensation
pH Calibration Automatic, 1 or 2 point
Probe
probe with DIN connector and 1 m (3.3’) cable
Battery Type 3 x 1.5V AA / IEC LR6 Battery Life Approx. 2000 hours Auto-off After 8 minutes of non-use Environment 0 to 50°C (32 to 122°F);
Dimensions
Weight 210 g (7.4 oz.)
(*)
To clean the meter, use water only.
modify the design, construction and appear-
ance of its products without advance notice.
(@20°C) ±0.02 pH
±0.5°C up to 60°C; ±1°C outside
±1°F up to 140°F; ±2°F outside
with 2 sets of memorized buffers
(pH 4.01/7.01/10.01 or 4.01/6.86/9.18)
(included) HI 1296D
amplified pH/temperature
RH max 100%
150 x 80 x 36 mm (5.9 x 3.1 x 1.4”)
the temperature range is limited to 80°C (176°F) if using the HI1296D probe.
Hanna Instruments reserves the right to
IST991001R2 07/05
OPERATIONAL GUIDE METER SETUPpH MEASUREMENT & CALIBRATION
To connect the probe
With the meter turned off, connect the HI 1296D probe to the DIN socket on the top of the meter by aligning the pins and pushing in the plug. Tighten the nut to ensure a good connection. Remove the protective cap from the probe before taking any measurements.
To turn the meter ON and check the bat­tery status
Press the ON/OFF/MODE button until the dis­play lights up. At start-up, all the LCD segments are displayed for 1 second, then the percent in­dication of the remaining battery life is shown for another second (E.g. % 100 BATT). The meter then enters the normal measuring mode.
Note: If the display needs to be checked, keep
the ON button pressed while turning the meter on. The meter will display all segments as long as the button is pressed.
To freeze the display
While in measurement mode, press the SET/ HOLD button, HOLD appears on the secondary display and the reading will be frozen on the LCD (E.g. pH 5.73 HOLD). Press any button to return to normal mode.
To turn the meter OFF
While in normal measurement mode, press the ON/OFF/MODE button. OFF will appear on the secondary display. Release the button.
Note: The meter is provided with an acoustic
signal feature, which can be disabled using the switch located in the battery compartment.
Note: When the meter detects the absence of a
temperature probe at its input, the Automatic Temperature Compensation is turned off, and the meter uses a default value of 25°C (77°F) for the temperature measurement and com­pensation. In this condition, the secondary LCD shows 25.0°C (77.0°F) blinking. When a probe is connected, the meter automatically returns to the A TC mode, the A TC tag is turned on, and the temperature is shown on the sec­ondary display.
• Make sure the meter has been calibrated be­fore use.
• If the probe is dry, soak it in HI 70300 storage solution for one hour to reactivate it.
• Place the tip of the probe into the sample to be tested.
• Stir briefly and wait until the symbol on the top left of the LCD is turned off.
• The LCD shows the pH value (automatically compensated for temperature) on the primary LCD, while the secondary LCD shows the tem­perature of the solution.
• If measurements are taken in different samples successively, rinse the probe tip thor­oughly to eliminate cross-contamination. After cleaning, rinse the probe tip with some of the sample to be measured.
pH calibration
For better accuracy, frequent calibration of the instrument is recommended. In addition, the instrument must be recalibrated whenever:
a) The pH electrode is replaced. b) After testing aggressive chemicals. c) Where high accuracy is required. d) At least once a month.
• From normal mode, press and hold ON/OFF/ MODE until OFF on the secondary display is replaced by CAL. Release the button.
• The LCD enters the calibration mode, display­ing “pH 7.01 USE” (or “pH 6.86 USE” if the NIST buffer set was selected). After 1 second the meter activates the automatic buffer rec­ognition feature. If a valid buffer is detected, then its value is shown on the primary display, and REC appears on the secondary LCD. If no valid buffer is detected, the meter keeps the USE indication active for 12 seconds, and then replaces it with WRNG , indicating that the sample being measured is not a valid buffer.
• For a
single-point calibration
with buffers pH
4.01, 9.18 or 10.01, the meter automatically accepts the calibration when the reading is stable; the meter will show on the primary dis­play the accepted buffer, with the message “OK 1” on the secondary display, and an au­dible signal is produced.
After 1 second the meter automatically returns to the normal measuring mode. If a single-point calibration with buffers pH 7.01 or 6.86 is desired, then after the calibration point has been accepted the ON/OFF/MODE button must be pressed in order to return to the normal measuring mode. After the button is pressed, the meter shows "7.01" (or "6.86")
- "OK 1", and an audible signal is produced. After 1 second, the meter automatically returns to the normal measuring mode.
Note: It is always recommended to carry out a
two-point calibration for better accuracy.
• For a
two-point calibration
, place the probe in pH 7.01 (or pH 6.86) buffer . After the calibration point has been accepted, the “pH 4.01 USE” message appears. The message is held for 12 seconds, unless a valid buffer is recog­nized. If no valid buffer is recognized, then the WRNG message is shown. If a valid buffer (pH 4.01, pH 10.01 or pH 9.18) is detected, then the meter completes the calibration pro­cedure. When the buffer is accepted, the LCD shows the accepted value with the “OK 2” message on the secondary display. The meter then returns to the normal measuring mode.
Note: When the calibration procedure is com-
pleted, the CAL tag is turned on.
To quit calibration and to reset to the de­fault values
• After entering the calibration mode and before the first point is accepted, it is possible to quit the procedure and return to the last calibration data by pressing the ON/OFF/MODE button. The secondary LCD displays ESC for 1 sec­ond and the meter returns to normal mode.
• To reset to the default values and clear a pre­vious calibration, press the SET/HOLD button after entering the calibration mode and before the first point is accepted. The secondary LCD displays CLR for 1 sec­ond, the meter resets to the default calibration and the “CAL” tag on the LCD disappears.
Setup mode allows the selection of the tempera­ture unit and of the pH buffer set. To enter the Setup mode, press and hold the ON/OFF/MODE button until CAL on the second­ary display is replaced by TEMP and the current temperature unit (E.g. TEMP °C). Then:
for °C/°F selection
, use the SET/HOLD button. After the temperature unit has been selected, press ON/OFF/MODE to enter the buffer set selection mode; press ON/OFF/MODE twice to return to the normal measuring mode.
to change the calibration buffer set
, after set­ting the temperature unit, the meter will show the current buffer set: “pH 7.01 BUFF” (for 4.01/
7.01/10.01) or “pH 6.86 BUFF” (for 4.01/6.86/
9.18). Change the set with the SET/HOLD but­ton, then press ON/OFF/MODE to return to nor­mal mode.
BATTERY REPLACEMENT
The meter displays the remaining battery percentage when turned on. When the level is below 5%, the left of the LCD blinks to indicate a low battery condition. If the battery level is low enough to cause erroneous readings, the Battery Error Prevention System (BEPS) turns the meter off. Unscrew the 4 screws located on the back of the meter and carefully replace the 3 AA batteries located in the battery compartment, while paying attention to their polarity. Reat­tach the back making sure that the gasket is in place and tighten the 4 screws to ensure a watertight seal.

symbol on the bottom
Loading...