Dear Customer,
Thank you for choosing a Hanna Instruments product. This manual will provide you with the
necessary information for correct use of the instrument.
Please read this instruction manual carefully before using the instrument.
If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com
or see the back side of this manual for our worldwide sales and technical service contacts.
This instrument is in compliance with directives.
WARRANTYWARRANTY
WARRANTY
WARRANTYWARRANTY
HI 98713 is warranted for two years against defects in workmanship and materials when used
for its intended purpose and maintained according to instructions. This warranty is limited to
repair or replacement free of charge.
Damage due to accidents, misuse, tampering or lack of prescribed maintenance is not covered.
If service is required, contact the dealer from whom you purchased the instrument. If under
warranty, report the model number, date of purchase, serial number and the nature of the failure.
If the repair is not covered by the warranty, you will be notified of the charges incurred. If the
instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization
number from the Technical Service Department and then send it with shipping costs prepaid.
When shipping any instrument, make sure it is properly packed for complete protection.
To validate your warranty, fill out and return the enclosed warranty card within 14 days from the
date of purchase.
TABLE OF CONTENTSTABLE OF CONTENTS
TABLE OF CONTENTS
TABLE OF CONTENTSTABLE OF CONTENTS
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
Please examine this product carefully. Make sure the instrument is not damaged. If any damage
has occurred during the shipment, please notify your dealer.
This HI 98713 Portable Turbidimeter is supplied complete with:
GENERAL DESCRIPTION .......................................................................................................................................... 4
TAG IDENTIFICATION SYSTEM .................................................................................................................................. 5
PRINCIPLE OF OPERATION ....................................................................................................................................... 6
MEASUREMENT UNITS ............................................................................................................................................ 7
GOOD LABORATORY PRACTICE (GLP) ...................................................................................................................... 29
TAG INSTALLATION ............................................................................................................................................... 35
LED REPLACEMENT ............................................................................................................................................... 36
PC INTERFACE ..................................................................................................................................................... 38
RECOMMENDATIONS FOR USERS ........................................................................................................................... 39
23
Note:Save all packing material until you are sure that the instrument works correctly. Any
defective item must be returned in the original packing with the supplied accessories.
GENERAL DESCRIPTION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
TAG IDENTIFICATION SYSTEMTAG IDENTIFICATION SYSTEM
TAG IDENTIFICATION SYSTEM
TAG IDENTIFICATION SYSTEMTAG IDENTIFICATION SYSTEM
HI 98713 is a high accuracy ISO compliant portable turbidimeter that benefits from Hanna’s years
of experience as manufacturer of analytical instruments. The HI 98713 meets and exceeds the
requirements of the ISO 7027 for water quality.
The instrument is specially designed for water quality measurements, providing a reliable and
accurate reading on low turbidity values.
The HI 98713 instrument measures the turbidity of a sample in the 0.00 to 1000 FNU (Formazin
Nephelometric Units) range. An effective algorithm calculates and converts the detectors output in FNU.
Depending on the measured probe and needed accuracy, normal measurement, continuous
measurement, or signal averaging measurement can be selected.
The instrument is based on a state-of-the-art optical system, which guarantees accurate results. The
optical system, consisting in an infrared LED and two detectors (scattered and transmitted light),
assures long term stability and minimizes stray light and color interferences. It also compensates for
variations in intensity of the LED, minimizing the need of frequent calibration.
The 25 mm round cuvets made from special optical glass guarantee the repeatability and consistency
of the measurements.
Calibration can be easily performed at any time in two, three or four points (<0.1, 15, 100 and
750 FNU-adjustable calibration points), using the supplied or user prepared standards.
HI 98713 has complete GLP (Good Laboratory Practice) functions that allows traceability of the
calibration conditions. The last calibration points, time and date can be checked at any time by a
single touch.
HI 98713 has a very user-friendly interface, with an easy to read, large LCD (Liquid Crystal Display).
The displayed codes guide the user step by step with routine operation and through calibration.
Confirmation and acoustic signals help the user during instrument operation.
The HI 98713 turbidimeter is a truly splash proof portable instrument. It is supplied with a rigid
carrying case that offers protection for harsh environments.
One battery set is enough for at least 1500 measurements. The battery charging percentage and low
battery condition are displayed on the LCD to avoid unexpected battery failure. In order to save the
battery life, the instrument has an auto shut-off feature and will turn off after 15 minutes of non-use.
In addition, the instrument is equipped with backlight and the current time is continuously displayed
on the LCD.
The instrument also provides a logging function. Up to 200 measurements can be stored in the
internal memory and consulted at any time. Data can be downloaded to a PC for storing or further
analysis through one of the two available ports: RS232 or USB.
For advanced field applications, the HI 98713 turbidimeter is equipped with Tag Identification
System (TIS) that makes data collecting and management simpler than ever.
Hanna is the first manufacturer of turbidity instruments that has decided to add the unique T.I.S.
- Tag Identification System to our Portable Turbidimeters, to meet the more restrictive needs of
the users and fit all advantages of this system to the turbidity measurements and data management.
The system is designed for scientific and industrial applications, or to prove during safety audits
and inspections that samples have been truly taken on pre-established locations.
The system is as easy to install as to operate. Just fix the so-called
iButton® tags near your
sampling sites that need to be checked often, and with this the T.I.S. is setup. The tag contains
a computer chip embedded in a durable stainless steel can. It is designed to withstand the harsh
environments, indoors or outdoors. The number of tags that can be installed is practically
unlimited, because each tag has a unique identification code.
Immediately after tags installation you can start collecting data. Use the Portable Turbidimeter to
take measurements and memorize the test result by pressing the Log-on-Demand key. Then, the
instrument will ask for the tag identification. Simply touching the iButton® with the matching
connector on the Portable Turbidimeter does identify and authenticate logging, by storing the
iButton® serial number, time and date stamp.
The power of the T.I.S. features resides in the PC application. Download all test data to your PC
and use our HI 92000 Windows® compatible application software for further data management.
You can sort or filter all your collected test data on different criteria like on a specific sampling
location, parameter, date and time intervals, or fix range to filter measured values. The data can
be plotted in a graph, exported to other common Windows® applications or printed for reporting
purpose.
It is also possible to add new tags later on, thus increasing an already existing database. Each
time the PC software recognizes a new added tag, it will ask for a description of the new sampling
location.
ABBREVIATIONSABBREVIATIONS
ABBREVIATIONS
ABBREVIATIONSABBREVIATIONS
FNUFormazin Nephelometric Units
LCDLiquid Crystal Display
RTCReal Time Clock
RHRelative Humidity
TISTag Identification System
IDIdentification
ISOInternational Standard Organization
®®
®
®®
is registered Trademark of “MAXIM/DALLAS semiconductor Corp.”
iButton
45
PRINCIPLE OF OPERATIONPRINCIPLE OF OPERATION
UTJUTN/UTF/UNFOiS
2
)L/gm(
UTJ19105.2
UNF
UTF
UTN
350.0131.0
OiS
2
)L/gm(4.05.71
PRINCIPLE OF OPERATION
PRINCIPLE OF OPERATIONPRINCIPLE OF OPERATION
MEASUREMENT UNITSMEASUREMENT UNITS
MEASUREMENT UNITS
MEASUREMENT UNITSMEASUREMENT UNITS
Turbidity of the water is an optical property that causes light to be scattered and absorbed, rather than
transmitted. The scattering of the light that passes through a liquid is primarily caused by the
suspended solids. The higher the turbidity, the greater the amount of scattered light. Because even
the molecules in a very pure fluid scatter light to a certain degree, no solution will have zero
turbidity.
The ISO 7027 standard specifies the key parameters for the optical system to measure turbidity for
drinking and surface water, using the formazin based metric method.
The HI 98713 Portable Turbidimeter is designed to meet or exceed the criteria specified by the
ISO 7027 standard.
The light beam that passes through the sample is scattered in all directions. The intensity and pattern
of the scattered light is affected by many variables like wavelenght of the incident light, particle size,
shape, refractive index and color.
The Hanna’s HI 98713 is based on a state-of-the-art optical system that guarantee both high
performance and reliable results.
This optical system includes an infrared LED, a scattered light detector (900) and a transmitted light
detector (1800). The microprocessor of the instrument calculates from the signals that reaches the
two detectors, the FNU value, using an effective algorithm. This algorithm corrects and compensates
for interferences of color, making the HI 98713 turbidimeter color-compensated.
Many methods were used to measure turbidity over the years. The Jackson Candle Turbidimeter was
used to measure turbidity as Jackson turbidity units (JTU). The Secchi Disk is commonly used to
measure turbidity in lakes and other deep waters (mg/L SiO
). Both methods are visual and are not
22
2
22
considered very accurate. To obtain more accurate readings a formazin based meter should be used
as a turbidity reading instrument.
The HI 98713 turbidimeter reports the measurements only in FNU (Formazin Nephelometric
Units). The conversion table between these measurement units is shown bellow:
The optical system and measuring technique allow the compensation of LED intensity fluctuations,
minimizing the need of frequent calibration.
The lower detection limit of a turbidimeter is determined by the so called “stray light”. Stray light
is the light detected by the sensors, that is not caused by light scattering from suspended particles.
The optical system of HI 98713 turbidimeter is designed to have very low stray light, providing
accurate results for low turbidity samples. However, special care must be taken when measuring low
turbidities (see General Tips for an Accurate Measurement for sample preparation and measuring
techniques).
67
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
INSTRUMENT DESCRIPTION
1) Cuvet Lid. Close the cuvet lid prior to start a measurement.
2) Cuvet Holder. Insert the cuvet into the holder with the cuvet mark matching the case mark.
3) Liquid Crystal Display (LCD). The LCD has backlight for better visibility in dark environments.
4) Keypad. Splash proof resistant.
5) LED connector. Connect the new LED using a screwdriver during LED changing procedure.
6) LED. Replaceable infrared LED.
7) Battery Lid. Remove the battery lid in order to change batteries or replace the LED.
KEYPAD DESCRIPTION
8) ON/OFF, press to turn the instrument ON/OFF. If no key is pressed for more than 15
minutes, the instrument automatically shuts off.
9) GLP , press to enter/exit GLP feature. In SETUP it is used to increase the set values. In Log
Recall it is used to select a newer record (scroll up).
10) AVG , press to set the average reading mode ON/OFF. In SETUP it is used to decrease the
set values. In Log Recall it is used to select an older record (scroll down).
11) CAL, press to enter/exit calibration. During setup it is used to start/stop editing a parameter.
12) LIGHT, press to turn ON/OFF the backlight.
13) RCL, press to enter/exit viewing log content.
14) LOG/CFM, press to save the log records or to confirm the selected option.
15) READ , press to start a measurement. Press and hold READ to make a continuous
measurement. In Log Recall it is used to see the content of a record. In GLP it is used to
see all available informations. In SETUP, during date or time editing, it is used to move the
focus on the next setting item.
16) SETUP/DEL, press to enter/exit setup. The DEL function is available in Log Recall to delete
one or all records. In GLP it is used to delete the user calibration.
CONNECTORS DESCRIPTION
17) AC adapter connector, used to connect an external AC Adapter.
18) RS232 connector, used to transfer data through the RS232 connection. Use HI 920011
serial cable to connect to the PC.
19) Tag reader connector. Touch the tag with the connector to read the location identification
number during logging.
20) USB connector, used to transfer data to the PC.
89
DISPLAY DESCRIPTION
1) Battery icon. When it is ON, it shows that the instrument works on battery. When blinking,
the batteries are almost empty and need to be replaced.
2) Wait icon. It is displayed when the instrument performs an internal checkup.
3) Measurement icon. The icon shows the measuring scheme of the instrument.
4) LED icon. The LED icon is shown when the LED is turned on.
5) Four digit main display. The main display shows the turbidity value after one measurement.
Depending on the instrument working mode, other values or messages are displayed.
6) Measurement units. The turbidity is measured in FNU. When average or continuous mode is
selected, the “FNU” tag blinks for each new displayed value. For conversions in other units
see Measurement Units section.
7) AVG icon. When selected, the measurement will be taken in average mode. The “FNU” tag
will blink for each new displayed value.
8) Four digit secondary display. The secondary display shows the current time (if selected) or
other values/messages.
BEEPER
A beeper is used to make the user interface more friendly. An error or invalid key press is
signaled by a long beep. A confirmation beep is signaled by a short beep. The beeper is
selectable as ON or OFF in Setup Menu.
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
Range0.00 to 9.99 FNU
10.0 to 99.9 FNU
100 to 1000 FNU
Range SelectionAutomatically
Resolution0.01 FNU from 0.00 to 9.99 FNU
0.1 FNU from 10.0 to 99.9 FNU
1 FNU from 100 to 1000 FNU
Accuracy±2% of reading plus 0.1 FNU
Repeatibility±1% of reading or 0.01 FNU, whichever is greater
Stray Light< 0.1 FNU
Typical EMC Deviation±0.05 FNU
Light Source860 nm infrared LED
IR DetectorSilicon Photocell
MethodAdaptation of ISO 7027, ratio method with 90° and 180° detector.
Display60 x 90 mm LCD with backlight
Standards< 0.1, 15, 100 and 750 FNU
CalibrationTwo, three or four point calibration
LOG Memory200 records
Serial InterfaceRS232 and USB 1.1
Environment0 to 50°C (32 to 122°F); max 95% RH non-condensing
Power supply4 x 1.5V AA alkaline batteries or AC adapter
Auto Shut-offAfter 15 minutes of non-use
Dimensions224 x 87 x 77 mm (8.8 x 3.4 x 3.0”)
Weight512 g (18 oz.)
1011
GENERAL TIPS FOR AN ACCURATE MEASUREMENTGENERAL TIPS FOR AN ACCURATE MEASUREMENT
GENERAL TIPS FOR AN ACCURATE MEASUREMENT
GENERAL TIPS FOR AN ACCURATE MEASUREMENTGENERAL TIPS FOR AN ACCURATE MEASUREMENT
HI 98713 is a highly accurate turbidimeter. To meet the instrument performance and fully
benefit of its properties it is very important that the analyst perform accurate, precise and
repeatable readings using proper measurement techniques. Special care must be taken during
sample preparation and handling.
The instructions listed below should be carefully followed during measurement and calibration to
ensure best accuracy.
CUVET
The cuvet is part of the optical system in all measurements. The light reaches the sample by
passing through the cuvet glass. As a result, the measurement can be affected by the glass
imperfections, dirt, dust, scratches, or fingerprints present on the cuvet surface.
CUVET HANDLING
The cuvets should be free of scratches or cracks. Any cuvet with visible scratches will be
discarded. The cuvets should be periodically washed with acid. After washing, the cuvets should
be well rinsed many times with distilled or deionized water. Allow cuvets to air-dry and store
them for long periods of time with caps, to avoid dirt entering inside. Always handle the cuvet
by touching only the cap or its top side (over the horizontal line).
Always store the cuvets in separate boxes or with separators between them to avoid scratches on
the surface.
CUVET PREPARATION
Whenever a cuvet is used, it must be clean inside and outside. When it is placed into the
instrument, it must be dry outside, completely free of fingerprints or dirt.
If the cuvet is not indexed, put the cuvet with the factory mark aligned with the sign on the
instrument top.
CUVET OILING
To hide minor imperfections and scratches, the cuvets should be oiled outside with the supplied
silicone oil. This is very important, especially for low turbidity samples (< 1 FNU), otherwise
scratches can contribute and alter turbidity readings.
The silicone oil has the same refractive index as the glass and will not alter the turbidity readings.
It is important to apply only a thin layer of silicone oil.
Warning: Do not apply the oil in excess, it may retain dirt or contaminate the cuvet holder of the
instrument, altering the turbidity readings.
It is very important to apply the silicone oil on a clean, dry cuvet. Apply a few drops of oil and wipe
the cuvet thoroughly with a lint-free cloth. Wipe off the excess of oil till you obtain a thin, uniform
layer. If the procedure is correctly followed, the cuvet should appear nearly dry with no visible oil.
Note: The supplied cloth for oiling should be stored together with the silicone oil bottle and cuvets,
taking care to avoid contamination with dirt. After a few oiling procedures the cloth will
contain enough oil to wipe the bottle with it without adding more oil. From time to time add
some drops of oil on the cuvet to provide the necessary oil quantity in the cloth.
INDEXING A CUVET
It is very important for low turbidity readings to always insert the cuvet into the instrument in the
same position.
All cuvets are factory indexed. This index can be used to put the cuvet with the factory mark on the
cuvet aligned with the sign on the instrument top.
To further reduce the effect of glass imperfections, the cuvet can be indexed and use this new index
as the position mark.
For indexing one cuvet or matching multiple cuvets the continuous reading mode is suggested. In this
mode if READ is kept pressed, multiple successive readings are taken without turning off the LED.
After first reading is displayed, it is possible to open the cuvet lid and rotate the cuvet without
generating an error condition. The turbidity is immediately displayed, reducing drastically the
measurement time. The LED of the instrument will turn off only when READ is released.
Note: The instrument can not perform continuous readings if the average mode is on.
1213
Loading...
+ 14 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.