Hanna Instruments HI 98601 User Manual

FUNCTIONAL DESCRIPTION
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it to make sure that no damage has occurred during shipping. If there is any damage, notify your dealer.
NOTE: Save all packing material until you are sure that the instrument functions correctly. All defective items must be returned to us in the original carton with the supplied accessories.
GENERAL DESCRIPTION
The HYGROCHECK is a lightweight, pocket-sized tester designed for on-the­spot testing of relative humidity.
The rugged ABS housing has a built-in LCD that displays the reading in large digits.
Measurement are simple with HYGROCHECK. Just turn it on, wait a few seconds, and read the display. The relative humidity is expressed in percents of 100 (10%-90%).
www.hannainst.com
NOTE:
The RH probe sensor must never come into contact with water or other liquids.
USING THE INSTRUMENT
CALIBRATION
SPECIFICATIONS
Using the HYGROCHECK is simple. However, the recommended procedure is as follows:
The end of the humidity detector should be exposed to a current of air moving at 0.5 m (20") per second or more.
In the absence of air movement, the response can be accelerated by mov­ing the probe.
The probe sensor must never come into contact with water or other liq­uids. If this should happen, or if condensation causes drops to form on the surface of the humidity sen­sor, turn off the instrument and wait until this water has evaporated com­pletely. In order to accelerate the evapora­tion process, the humidity sensor could be exposed to a current of air.
If you experience any problems in taking measurements, please contact your dealer or the nearest Hanna Instruments Customer Service Department.
All Hanna hygrometers are calibrated with reference to NIST certified instruments. In addition, the Hanna Service Centers are equipped with state-of-the-art cali­bration chambers. For an accurate recalibration of your meters, contact your nearest Hanna Service Center.
BATTERY REPLACEMENT
In order to replace the batteries, simply slide the cover off and replace it with new batteries (Note: attention must be given to assure the correct polarity).
Range 10.0 to 90.0%
Resolution 0.1% RH
Accuracy ±3% RH of F.S.
Operating 0 to +50ºC Conditions (+32 to +122ºF)
95% RH
Dimensions 180 x 30 x 15mm
(7.1 x 1.2 x 0.6") L x W x H
Battery 3 x 1.5V, 100 hours
of continuous use
Weight 62 g (2.2 oz)
Response Time Up to 20 seconds
for 95% accuracy
ISTRHYGCHK
5/93